columns Protein-Pak Hi Res IEX Columns and Standards [ CARE AND USE MANUAL ] I. INTRODUCTION CONTENTS I. INTRODUCTION

Size: px
Start display at page:

Download "columns Protein-Pak Hi Res IEX Columns and Standards [ CARE AND USE MANUAL ] I. INTRODUCTION CONTENTS I. INTRODUCTION"

Transcription

1 CONTENTS I. INTRODUCTION II. CONFIGURING AN ACQUITY SYSTEM FOR USE IN IEX PROTEIN SEPARATIONS III. GETTING STARTED a. ecord Installation b. Column Connectors c. Column Installation d. Column Equilibration e. Useful Functional Tests for Benchmarking a New Column I. INTRODUCTION Thank you for choosing a Protein-Pak ion-exchange column. Waters Protein-Pak Hi Res IEX family of columns were developed to assist in the UPLC characterization of recombinant proteins and monoclonal antibodies found in many of today s novel biopharmaceutical therapeutics. The non-porous, high-ligand density particles overcome some of the high molecular weight limitations present using traditional porous, IEX particles. In addition, use of a multi-layered network of ion-exchange groups (SP, CM, or Q) on non-porous particles assist the effective diffusion and binding of charged proteins or other biomolecules onto the particles to deliver relatively high sample loading capacities and component resolution while minimizing column fouling. IV. COLUMN SPECIFICATIONS AND USE a. Specifications b. Sample Preparation c. Operating ph d. Solvents e. Pressure f. Flow Rate g. Temperature The available ion-exchange chemistries include a strong-anion and weak- and strong-cation exchangers (Table 1), all of which are stable, over a wide ph range (3 10), high salt concentrations and standard pressures (see Column Specifications and Use section) via use of polymer-based particles. Only when the Protein-Pak Hi Res IEX column series is combined with the ACQUITY UPLC system will the full performance benefits of this column be realized. V. TROUBLESHOOTING Table 1 VI. COLUMN CLEANING, REGENERATION, AND STORAGE a. Cleaning and Regeneration b. Storage Abbreviation SP CM Q Ion-Exchange Ligand Sulfopropyl Carboxymethyl Quaternary ammonium VI. INTRODUCING ECORD INTELLIGENT CHIP TECHNOLOGY a. Introduction b. Installation c. Manufacturing Information d. Column Use Information VIII. ORDERING INFORMATION columns

2 II. CONFIGURING AN ACQUITY SYSTEM FOR USE IN IEX PROTEIN SEPARATIONS Although Waters Protein-Pak Hi Res IEX columns do not require the high pressure capability of an ACQUITY UPLC system, the full performance benefits of this column will only be realized with the low system volume and low detector dispersion of the optimized ACQUITY UPLC system. However, ion-exchange chromatography may require modifications to an existing ACQUITY UPLC system. Please refer to Size Exclusion and Ion-Exchange Chromatography of Proteins using the ACQUITY UPLC System, ( A) for specific recommendations that can be obtained at com/chemcu. III. GETTING STARTED An Analysis Report and an Inspection Data Sheet are generated and available for each shipped Protein-Pak Hi Res IEX column. The Analysis Report is specific to each batch of packing material and includes the batch number and chromatographic separation obtained using defined protein standards. This document is available upon request htm?id= &ev= &locale=en_us. The Inspection Data Sheet is included with each shipment and is specific to each individual column and contains the following information: batch number, column serial number, plate count, tailing factor, retention factor, and chromatographic separation conditions. These data should be stored for future reference. b1. The ACQUITY UPLC system utilizes tubing and gold plated compression screws that have been designed to meet stringent tolerance levels and to minimize extra column volumes. The columns should be attached to the ACQUITY UPLC injector with a column stabilizer. b2. Waters Protein-Pak Hi Res columns require end fittings that have a 2 mm depth between the ferrule and the end of the extending stainless steel tubing (Figure A) that must perfectly seat to the end of the Protein-Pak Hi Res column. Connecting a different style connector to a Waters Protein-Pak Hi Res IEX column may leave a gap between the end of the stainless steel tubing and the column that will result in undesired peak broadening (Figure B). To correct this problem, disassemble the finger-tight fitting of the stabilizer and remake the connection (i.e., 2 mm depth) to the Waters Protein-Pak Hi Res IEX column. Figure A Figure B 2 mm a. ecord Installation The ecord button should be attached to the side of the Waters ACQUITY UPLC column heater module. The ecord button is magnetized and does not require specific orientation. b. Column Connectors There are two options for connecting the Protein-Pak Hi Res column to an ACQUITY system. The first option is to use the reusable fitting on the 150 mm and 100 mm column stabilizers (sect b1). This fitting/tubing is supplied with ACQUITY UPLC column heaters. A second alternative is to remove this fitting and replace it with a non removable fitting (i.e. steel). If the second option is employed, specific directions must be followed (sec b2). In a proper tubing/column connection, the tubing touches the bottom of the column end fitting with no void between them. (Figure C). Figure C 2

3 c. Column Installation Note: The flow rates given in the procedure below are for a Protein- Pak Hi Res CM or SP, 7 µm, 4.6 mm i.d. column. 1. Purge the solvent delivery system of any organic or waterimmiscible mobile phases and connect the inlet end of the column to the injector outlet. An arrow on the column identification label indicates the correct direction of solvent flow. 2. Flush column with 100% aqueous buffer (line A1) by setting the pump flow to 0.2 ml/min. 3. When the buffer is flowing freely from the column outlet attach the column outlet to the detector. Monitor the system pressure to ensure the column is within its pressure limitations. 4. Gradually increase the flow rate, by not more than 0.5mL/min at a time (as described in step 2). 5. Once a stable backpressure has been achieved, proceed to the next section. d. Column Equilibration Protein-Pak Hi Res columns are shipped in 20 mm sodium sulfate with 0.05% sodium azide. It is important to ensure mobile-phase compatibility before changing to a different mobile-phase system. Equilibrate the column with a minimum of 10 column volumes of the buffer to be used (refer to Table 2 for column volumes). Table 2. Empty Column Volumes in ml (multiply by 10 for flush solvent volumes). Column Dimension Approximate Volume 4.6 x 100 mm 1.66 ml e. Useful Functional Tests for Benchmarking a New Column It may be useful to benchmark the column performance with a sample that is representative of the intended application. A separation of common proteins with an appropriate method is suggested for your new column. Two standards were specially designed for this purpose. The IEX Cation Test Standard is specifically designed with proteins that resolve well on cationexchange columns. The IEX Anion Test Standard is specifically designed with proteins that resolve well on anion-exchange columns. This test can also be used to monitor the condition of your column after extended use. The following sets of conditions and chromatograms show examples of using the IEX Standards for benchmarking and monitoring column and system health. The exact results observed in any laboratory will depend on the instrument in use. Similar results can only be expected with a Waters ACQUITY UPLC system. These tests are valuable for monitoring the life of the column and for troubleshooting separation difficulties that may arise. Protein-Pak Hi Res Cation-Exchange Columns and Standards Sample Diluent: 20 mm Sodium Phosphate Buffer, ph 6 7 The IEX Cation Test Standard is packaged in a Waters TruView Max Recovery Vial (p/n cv) and is lyophilized for prolonged storage and stability. The standard is shipped at room temperature but it upon arrival, it is recommended to store this standard at -20 C. Protein Sigma P/N Conc. mg/ml Bovine Ribonuclease A R Cytochrome C, bovine heart C Chicken Lysozyme L Column: Protein-Pak Hi Res CM, 4.6 x 100 mm, or Protein-Pak Hi Res SP, 4.6 x 100 mm Injection Volume: 10.0 μl Flow Rate: 1 ml/min Mobile Phase A: 20 mm sodium phosphate buffer, ph 6.6 Mobile Phase B: 20 mm sodium phosphate buffer, 0.5 M NaCl, ph 6.6 Weak Needle Wash: Mobile phase A Strong Needle Wash: Mobile phase B Seal Wash: 90/10 H 2 0/MeOH Temperature: 30 C Detection: 280 nm Gradient: Time %A %B Curve

4 Protein-Pak Hi Res CM Column AU Cytochrome C Ribonuclease A Protein-Pak Hi Res SP Column AU Ribonuclease A Lysozyme min Cytochrome C Lysozyme min Protein-Pak Hi Res Anion-Exchange Column and Standard Sample Diluent: Lyophilized in 20 mm Tris Buffer, ph 8 9, Reconstituted in 1ml 100% Milli-Q Water, vortex and inspect for undisssolved solids. Protein Sigma P/N Conc. mg/ml Horse Myoglobin M Conalbumin from Chicken Egg White C Soybean Trypsin Inhibitor T Column: Inj Vol: 10.0 µl Flow Rate: Mobile Phase A (Acid): Mobile Phase B (Base): Mobile Phase C (Salt): Mobile Phase D (Aqueous): Protein Pak Hi Res Q, 4.6 x 100 mm 0.6 ml/min 100 mm TRIS HCl (aq) 100 mm TRIS (aq) 1000 mm NaCl (aq) Water Sample Manager Wash: 25 mm TRIS, 150 mm NaCl, ph 8.3 Sample Manager Purge: 25 mm TRIS, 150 mm NaCl, ph 8.3 Seal Wash: Column Temp.: 30 C Sample Manager Temp.: 4 C Detection: Auto Blend Plus Gradient: Time Flow (min) (ml/min) 90/10 H 2 O/MeOH 280 nm ph ph Curve Salt (mm) Salt Curve Initial Actual Gradient (with no Auto Blend capability): Flow Time %A %B %C %D (ml/min) Initial Protein-Pak Hi Res Q Column Myoglobin Conalbumin* Trypsin Inhibitor *Contains Isoform A and B which could be resolved under various ph conditions) 4

5 IV. COLUMN SPECIFICATIONS AND USE a. Specifications Description Protein-Pak Hi Res Q Protein-Pak Hi Res CM Protein-Pak Hi Res SP Ion Exchange Strong Anion Weak Cation Strong Cation Functional Group Matrix Quaternary ammonium Hydrophilic polymer Carboxymethyl Sulfopropyl Hydrophilic polymer Hydrophilic polymer Particle Size (µm) Pore Size: i.d. x L (mm) Non porous 4.6 x 100 Non porous 4.6 x 100 Non porous 4.6 x 100 Counter Ion Cl- Na+ Na+ ph Range Small Ion Capacity (µeq/g dry gel) pk a Approximate Protein Binding Capacity in mgs per column (i.e., BSA for Hi Res Q column; Lysozyme for Hi Res CM and Hi Res SP columns) Flow Rates 2. Max Pressure across column ml/min 2175 psi (15MPa) ml/min 1450 psi (10Mpa) ml/min 1450 psi (10Mpa) Salt Concentration No limit No limit No limit Organic Concentration <50%. When switching from aqueous buffers to organic, lower flow rates to <0.25 ml/ min. <50%. When switching from aqueous buffers to organic, lower flow rates to <0.5 ml/min. <50%. When switching from aqueous buffers to organic, lower flow rates to <0.5 ml/min. Temperature ( C) For optimal resolution of complex samples, do not exceed 20% of the column s protein binding capacity. 2. See section e. pressure for details. To ensure the continued high performance of Protein-Pak Hi Res IEX columns, follow these guidelines: b. Sample Preparation 1. It is preferable to prepare the sample in the operating mobile phase or a mobile phase that has a higher ph (anion exchange) or lower ph (cation exchange) than the mobile phase to ensure complete loading of the sample onto the column. The ionic strength of the sample should also be lower or equivalent to that of the starting buffer. 2. If the sample is not dissolved in the mobile phase, ensure that the sample, solvent and mobile phases are miscible in order to avoid sample and/or buffer precipitation. c. Operating ph The recommended operating ph range for Protein-Pak Hi Res IEX columns is A listing of commonly used buffers and additives is given in Table 3. The column lifetime will vary depending upon the operating temperature as well as the type and concentration of buffer used. Table 3. Buffers Commonly Used for Ion Exchange Anion-Exchange Buffers ph Range Additive/Buffer pk a (25 C) Counter-ion Conc. (mm) N-Methylpiperazine 4.75 Cl Piperazine 5.68 Cl-/HCOO bis-tris 6.48 Cl Bis-tris propane 6.80 Cl MOPS 7.28 Cl Triethanolamine 7.76 Cl-/HCH 3 COO Tris 8.06 Cl Diethanolamine 8.88 Cl Ethanolamine 9.50 Cl CAPS Cl- 20 Cation-Exchange Buffers ph Range Additive/Buffer pk a (25 C) Counter-ion Conc. (mm) Lactic acid 3.81 Na Formic acid 3.75 Na+/Li Acetic acid 4.76 Na+/Li Malonic acid 5.68 Na+/Li MES 6.10 Na+/Li Phosphate 7.20 Na HEPES 7.55 Na+ or Li BICINE 8.35 Na+ 20 a Adapted from: 1. P Stanton, Ion-Exchange Chromatography, HPLC of Proteins and Peptide: Methods and Protocols; Aguilar, M.-I. Ed; Method in Molecular Biology, Humana Press, Totowa, NJ, Vol. 251, Ch Buffer Reference Center, Sigma Aldrich, com/life-science/core-bior agents/biological-buffers/learning-center/bufferreference-center.html 5

6 d. Solvents To maintain maximum column performance, use high quality chromatography grade solvents. Filter all aqueous buffers prior to use. Solvents containing suspended particulate materials can damage the fluidic components of the UPLC system and may clog the inlet distribution frit of the column. This will result in higher operating pressure and poor performance. The maximum organic concentration must be <50%. When the solvent is replaced by distilled or ion exchanged water, the flow rate must be less than 0.25 ml/min for Protein-Pak Hi Res Q, 5 µm column or less than 0.5mL/min for Protein-Pak Hi Res CM or SP, 7 µm columns. e. Pressure Protein-Pak Hi Res IEX columns can tolerate pressures across the column of up to 2175 psi for Protein-Pak Hi Res Q, 5 µm column and up to 1450 PSI for Protein-Pak Hi Res SP, 7 µm and Protein- Pak Hi Res CM, 7 µm columns. These suggested limits are for the pressure drop across the column. The actual recorded pressure on an HPLC or UPLC system is a sum of the pressure generated from solvent flow through the LC tubing and detector cell combined to the pressure generated from solvent flow through the packed IEX column. To determine the actual pressure drop across the IEX column, subtract the above value from the system pressure obtained (using the same eluent, flow, and temperature selected for the IEX separation method) when a union is substituted in place of the column on the LC System. Note: Working at the extremes of pressure, ph and/or temperature may result in shorter column lifetimes. f. Flow Rate Standard flow rate: ml/min: Protein-Pak Hi Res Q, 5 µm column ml/min: Protein-Pak Hi Res CM and SP, 7 µm columns Under certain conditions (viscous buffers, low temperatures) the maximum flow rate may have to be reduced so as to not exceed the maximum column pressure limits. g. Temperature Temperatures between 10 C 60 C are recommended for operating Protein-Pak Hi Res IEX columns. V. TROUBLESHOOTING The first step in systematic troubleshooting is comparison of the column, in its current state, to the column when it was functioning properly. The functional tests with the protein mixture may reveal subtle changes in surface chemistry that affect the application. There are several common symptoms of change in the column. 1. An increase in pressure is often associated with lost performance in the application. The first step in diagnosis is to ensure that the elevated pressure resides in the column rather than somewhere else in the system. This is determined by measuring pressure with and without the column attached to the instrument. If the system is occluded, the blockage should be identified and removed. If the pressure increase resides in the column, it is helpful to know whether the problem was associated with a single injection or whether it occurred over a series of injections. If the pressure gradually built up, it is possible that the column can be cleaned as described in Section VI. For future stability, it may be useful to incorporate a stronger regeneration step in the method. If a single sample caused the pressure increase, it likely reflects particulates or insoluble components, such as lipids. Cleaning is still an option, but using the more aggressive options. The sudden pressure increase suggests that the user should consider some sample preparation, such as filtration or high-speed centrifugation. 2. Loss of retention can reflect a change in the column surface chemistry. Before proceeding with diagnostic or corrective measures, check that the mobile phases have been correctly prepared and the correct method has been selected. Then repeat the functional or protein test. If the proteins show loss of retention, the column may require replacement. If the changes are small and reflected only for some proteins, one of the cleaning procedures may be effective. 3. Change in peak shape, resolution, or relative retention of peaks. Follow the same steps as for loss of retention (Symptom 2). 4. Carryover and memory effects are defined as the appearance of the constituents of one sample in the next gradient analysis. First determine whether the column or the system is the source of carryover. Define a gradient method that includes an internal gradient. That is, the analytical gradient is repeated within a single method. If the protein peaks appear in both gradients, at the same time after start, the protein came from the column in what is often described as a memory effect. If the protein peaks only appear when an injection is 6

7 made, they likely originate from adsorption to some system component. In that case follow the instrument manufacturer s recommendations. Memory effects as a source of carryover may be reduced or eliminated in several ways. First, raising the temperature of the separation reduces the possibility of nonspecific protein adsorption. Second, memory effects may be more pronounced with steep gradients. Keep the gradient slope at 10 column volumes or greater. Third, memory effects may be exacerbated by high flow rates. Reduce the flow rate by one half while doubling the gradient time to maintain a constant slope. Finally, apparent memory effects may actually reflect the solubility of the protein in the mobile phase. Reducing the amount injected may eliminate the effect. Note: Useful general information on column troubleshooting problems may be found in HPLC Columns Theory, Technology and Practice, U.D. Neue, (Wiley-VCH, 1997), the Waters HPLC Troubleshooting Guide (Literature code # EN) or visit the Waters Corporation website for information on seminars VI. COLUMN CLEANING, REGENERATION AND STORAGE a. Cleaning and Regeneration Changes in peak shape, peak splitting, shoulders on the peak, shifts in retention, change in resolution, carryover, ghost peaks, or increasing backpressure may indicate contamination of the column. Choose a cleaning option that may be expected to dissolve the suspected contaminant. 3. Flow reversal or back flushing is often suggested as part of a cleaning procedure. This should be reserved as a last resort. It may further damage the column or provide a short-lived improvement in performance. b. Storage For overnight storage, the column can remain in the IEX separation eluent while maintaining a low flow rate (e.g., 0.1 ml/min). For long-term storage, replace the eluent at the suggested flow rate shown below with filtered, high purity water containing 0.05% sodium azide that will inhibit undesired microbial growth. <0.25 ml/min for Protein-Pak Hi Res Q, 5 µm columns <0.5 ml/min for Protein-Pak Hi Res CM and SP, 7 µm columns Completely seal column to avoid evaporation and drying of the bed. VII. INTRODUCING ECORD INTELLIGENT CHIP TECHNOLOGY a. Introduction The ecord intelligent chip is a new technology that will provide the history of a column s maximum backpressure, maximum temperature and the number of injections performed throughout its lifetime. The ecord will be permanently attached to the column to assure that the column s performance history is maintained in the event that the column is moved from one instrument to another. 1. It may be useful to conduct cleaning procedures at one-half the flow rate typically used with that column. In this way the possibility of high pressure events is reduced. 2. Several different cleaning solutions may be injected to strip strongly adsorbed material or particulates from the column. Make the largest injection possible with the system configuration. With such strong cleaning solutions, it is best to disconnect the detector from the column and to direct the flow to waste. a. 0.1mol/L NaOH, or b. 20 ~ 40% acetic acid, or c. Solution containing aqueous organic solvent such as methanol or acetonitrile d. Solution containing a solubilizer such as urea and non-ionic surfactants Figure D. Waters ecord Intelligent Chip 7

8 At the time of manufacture, tracking and quality control information will be downloaded to the ecord. Storing this information on the chip will eliminate the need for a paper Certificate of Analysis. Once the user installs the column, the software will automatically download key parameters into a column history file stored on the chip. In this manual, we explain how the ecord will provide a solution for easily tracking the history of the columns, reduce the frustration of paperwork trails, and give customers the reassurance that a well performing column is installed onto their instruments. d. Column Use Information Figure E. ecord Inserted into Side of Column Heater. b. Installation Install the column into the column heater. Plug the ecord into the side of the column heater. Once the ecord is coupled to the reader, the overall column usage information will be available in the ACQUITY console, allowing the user to access column information on their desktop. The ecord chip provides the customer with column use data. The top of the screen identifies the column including chemistry type, column dimensions and serial number. The overall column usage information includes the total number of samples, total number of injections, total sample sets, date of first injection, date of last injection, maximum pressure, and temperature. The information also details the column history by sample set including date started, sample set name, user name, system name, number of injections in the sample set, number of samples in the sample set, maximum pressure, and temperature in the sample set and if the column met basic system suitability requirements. c. Manufacturing Information The ecord chip provides the user with an overview of the bulk material QC test results. The ecord chip provides the user with QC test conditions and results on the column run by the manufacturer. The information includes mobile phases, running conditions and analytes used to test the columns. In addition, the QC results and acceptance is placed onto the column. 8

9 VIII. ORDERING INFORMATION Product Description Part Number Protein-Pak Hi Res CM, 7 µm, 4.6 x 100 mm Protein-Pak Hi Res SP, 7 µm, 4.6 x 100 mm Protein-Pak Hi Res Q, 5 µm, 4.6 x 100 mm IEX Cation Test Standard IEX Anion Test Standard Note: Only when Protein-Pak Hi Res IEX columns are combined with the ACQUITY UPLC system are the full performance benefits realized. See Waters service notes P/N A for ACQUITY UPLC system configuration guidelines for ion-exchange chromatography. Waters, The Science of What s Possible, ACQUITY UPLC, ACQUITY, and UPLC are registered trademarks of Waters Corporaiton. Protein-Pak, TruView, and ecord are trademarks of Waters Corporation. All other trademarks are the property of their respective owners Waters Corporation. Produced in the U.S.A. February 2014 Rev C EN KP-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A. T: F:

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION

More information

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml GE Healthcare Life Sciences Instructions 71-7146-00 AI Ion Exchange Columns RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml Introduction RESOURCE Q and S are pre-packed columns for separating biomolecules

More information

[ Care and Use Manual ]

[ Care and Use Manual ] ACQUITY UPLC CSH Columns Contents I. Getting started a. Column Connectors b. Column Installation c. Column Equilibration d. ecord Installation e. Initial Column Efficiency Determination f. VanGuard Pre-Columns

More information

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual Catalog numbers 780-1070 780-1650 Please read these instructions before you use ENrich SEC high-resolution size exclusion

More information

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10

More information

INSTRUCTIONS 56-1190-98. Edition AC

INSTRUCTIONS 56-1190-98. Edition AC Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography

More information

[ Care and Use Manual ]

[ Care and Use Manual ] Analytical columns and cartridges i. Introduction Waters offers a large selection of analytical columns and cartridges packed with silica-based packing materials. Follow this instruction sheet to obtain

More information

The Theory of HPLC. Gradient HPLC

The Theory of HPLC. Gradient HPLC The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims

More information

Affi-Prep Protein A Matrix Instruction Manual

Affi-Prep Protein A Matrix Instruction Manual Affi-Prep Protein A Matrix Instruction Manual Catalog Numbers 156-0005 156-0006 Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules, CA 94547 LIT-230 Rev B Table of Contents Section 1 Introduction...1

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

Oasis HLB Cartridges and 96-Well Plates

Oasis HLB Cartridges and 96-Well Plates CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION

More information

HiPer Ion Exchange Chromatography Teaching Kit

HiPer Ion Exchange Chromatography Teaching Kit HiPer Ion Exchange Chromatography Teaching Kit Product Code: HTC001 Number of experiments that can be performed: 5 Duration of Experiment: Protocol: 5-6 hours Storage Instructions: The kit is stable for

More information

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C HIS-Select Nickel Affinity Gel Catalog Number P6611 Storage Temperature 2 8 C TECHNICAL BULLETIN Product Description HIS-Select Nickel Affinity Gel is an immobilized metalion affinity chromatography (IMAC)

More information

Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS Automating the Determination of Drug-to-Antibody Ratio (DAR) of Antibody Drug Conjugates (ADCs) Based on Separation by Hydrophobic Interaction Chromatography (HIC) Robert Birdsall, Eoin Cosgrave, Henry

More information

Gel Filtration Standard

Gel Filtration Standard Gel Filtration Standard Instruction Manual Catalog # 151-1901 Table of Contents Section 1 Introduction... 1 1.1 Instructions... 1 1.2 Recommended Volume of Standard... 2 1.3 Shelf Life... 5 1.4 Storage...

More information

HiTrap Heparin HP, 1 ml and 5 ml

HiTrap Heparin HP, 1 ml and 5 ml Instructions 71-7004-00 AU HiTrap affinity columns HiTrap Heparin HP, 1 ml and 5 ml HiTrap Heparin HP is a prepacked ready to use, column for preparative affinity chromatography. The special design of

More information

Protocol: HPLC (amino acids)

Protocol: HPLC (amino acids) Page 1 of 6 1. Chemicals Composition M MW gram volume (ml) product code 0-pthaldialdehyde C 8H 6O 2 134.13 Sigma P0657 Perchloric acid HClO 4 3.5 100.46 Dilute 70% stock solution 1:1 with demi to obtain

More information

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) IonSwift MAC-100 & MAC-200 Product Manual Page 1 of 16 PRODUCT MANUAL for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) MAC-100, 0.5 x 80mm (P/N 074702)

More information

Data File. Sephadex G-25 media and pre-packed columns. Introduction. Sephadex G-25 Bead structure. Desalting/buffer exchange and gel filtration

Data File. Sephadex G-25 media and pre-packed columns. Introduction. Sephadex G-25 Bead structure. Desalting/buffer exchange and gel filtration P H A R M A C I A B I O T E C H Sephadex G-25 media and pre-packed columns Data File Desalting/buffer exchange and gel filtration Reproducible desalting and buffer exchange in minutes with 90% 100% recovery

More information

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns Product Manual for SolEx TM HRP Cartridges and SolEx TM HRP RSLC Columns Page 1 of 15 Product Manual for SolEx TM HRP Cartridges SolEx HRP (P/N 074400)

More information

High-Throughput 3-D Chromatography Through Ion Exchange SPE

High-Throughput 3-D Chromatography Through Ion Exchange SPE High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample

More information

AFFINITY CHROMATOGRAPHY

AFFINITY CHROMATOGRAPHY 92 AFFINITY CHROMATOGRAPHY Tosoh bioscience ANALYSIS 93 AFFINITY CHROMATOGRAPHY AFFINITY CHROMATOGRAPHY afc PRODUCTS TSKgel BORONATE-5PW TSKgel CHELATE-5PW TSKgel TRESYL-5PW TOSOH FACT The Tosoh logo symbolizes

More information

[ Care and Use Manual ] styragel column. Contents. I. Introduction. II. Installing the Column Bank. III. Preparing Solvent and Samples

[ Care and Use Manual ] styragel column. Contents. I. Introduction. II. Installing the Column Bank. III. Preparing Solvent and Samples styragel column I. INTRODUCTION This manual covers the care and use of the Waters Styragel HR, HT, and HMW families of Gel Permeation Chromatography (GPC) columns. Please take a few moments to read this

More information

[ Care and Use Manual ]

[ Care and Use Manual ] The Waters GPC Cleanup System uses two Envirogel GPC Cleanup Columns to perform routine cleanup of environmental samples as specified by EPA Method 3640A, GPC Cleanup. The columns, 19 x 150 mm and 19 x

More information

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322. i.chianella.1998@cranfield.ac.

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322. i.chianella.1998@cranfield.ac. SPE and HPLC Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322 i.chianella.1998@cranfield.ac.uk Solid-Phase Extraction- SPE Simple, fast and efficient sample preparation

More information

Aurum Ion Exchange Mini Kits and Columns. Instruction Manual

Aurum Ion Exchange Mini Kits and Columns. Instruction Manual Aurum Ion Exchange Mini Kits and Columns Instruction Manual Catalog # 732-6710 Aurum AEX Mini Kits, 2 pk 732-6705 Aurum AEX Mini Kits, 10 pk 732-6706 Aurum AEX Mini Columns, 25 pk 732-6707 Aurum AEX Mini

More information

Protein Separation Technology

Protein Separation Technology [ ] Protein Separation Technology Protein Separation Technology The analysis and characterization of protein samples requires the detection of small chemical differences between large molecules. Most often

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

Solid Phase Extraction Products PAGE: 1. Introduction of Solid Phase Extraction (SPE) Why Choose Nano-Micro Tech SPE

Solid Phase Extraction Products PAGE: 1. Introduction of Solid Phase Extraction (SPE) Why Choose Nano-Micro Tech SPE Solid Phase Extraction Products PAGE: 1 Introduction of Solid Phase Extraction (SPE) SPE has been used increasingly in chemical analysis and purification broadly, and become the most popular technology

More information

SUCRALOSE. White to off-white, practically odourless crystalline powder

SUCRALOSE. White to off-white, practically odourless crystalline powder SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA

More information

Purification of reaction mixtures using flash chromatography.

Purification of reaction mixtures using flash chromatography. Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?

More information

Efficient Multi-Well Protein Purification Strategies

Efficient Multi-Well Protein Purification Strategies Application Note PN 33576 Efficient Multi-Well Protein Purification Strategies Introduction Many tools and techniques are available today for protein purification. Development of a purification process

More information

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,

More information

Bio-Gel P Polyacrylamide Gel Instruction Manual

Bio-Gel P Polyacrylamide Gel Instruction Manual Bio-Gel P Polyacrylamide Gel Instruction Manual Table of Contents Section 1 Introduction...1 Section 2 Technical Description...3 Section 3 Instructions for Use...6 3.1 Column Selection...6 3.2 Eluant Selection...6

More information

Dionex IonPac AS19-4µm

Dionex IonPac AS19-4µm Total Thermo Scientific Dionex IonPac AS9-4µm Column Product Manual P/N: September 203 Part of Thermo Fisher Scientific Product Manual for Dionex IonPac AS9-4µm Column Dionex IonPac AG9-4µm Capillary Guard

More information

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for

More information

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary Application Note Separation of three monoclonal antibody variants using MCSGP Category Matrix Method Keywords Analytes ID Continuous chromatography, Biochromatography; FPLC Protein A-purified monoclonal

More information

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS Development of Integrated Informatics Workflows for the Automated Assessment of Comparability for Antibody Drug Conjugates (ADCs) Using LC-UV and LC-UV/MS Henry Shion, Robert Birdsall, Steve Cubbedge,

More information

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Analysis of Vegetable Oils by High Performance Liquid Chromatography Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia

More information

An In-Gel Digestion Protocol

An In-Gel Digestion Protocol An In-Gel Digestion Protocol This protocol describes the digestion of a protein present in an SDS-PAGE gel band with trypsin. The band can be taken from either a 1D or 2D electrophoresis gel. Reagents

More information

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary Application Note Purifying common light-chain bispecific antibodies using MCSGP Category Matrix Method Keywords Analytes ID Continuous chromatography, biochromatography Antibodies MCSGP Bispecific antibody,

More information

How To Use An Acquity Qda Detector

How To Use An Acquity Qda Detector Mass-Directed Isolation of a Synthetic Peptide Using the ACQUITY QDa Detector Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY QDa Detector

More information

[ Care and Use Manual ]

[ Care and Use Manual ] WATERS IC-PAK ION EXCLUSION COLUMNS Contents I. Introduction II. Installation AND EQUILIBRATING THE COLUMn a. Preparation b. Installing the Column c. Equilibration III. PREPARATION OF ELUENTS AND SAMPLES

More information

Anion chromatography using on-line recycled eluents

Anion chromatography using on-line recycled eluents Journal of Chromatography A, 1089 (2005) 82 86 Anion chromatography using on-line recycled eluents Takashi Yokoyama, Hitoe Maekubo, Asami Sakai, Michio Zenki Department of Chemistry, Faculty of Science,

More information

Tamsulosin Hydrochloride Capsules

Tamsulosin Hydrochloride Capsules . nal Revision Bulletin Official October 1, 2011 Tamsulosin 1 standard solution, and shake well. Centrifuge at 1500 rpm for 10 min, and use the supernatant, passing it if Tamsulosin Hydrochloride Capsules

More information

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins APPLICATION NOTE POROS CaptureSelect affinity chromatography columns POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins Introduction POROS columns, containing highperformance

More information

Innovative vs Traditional

Innovative vs Traditional And in the red corner, introducing the challenger NASDAQ April 4 2011 243.47 $ 1.38 $ 0.58 % 0.07 $ 0.02 $ 39.4 % Technology DVD Digital Streamline Instant Play Internet Access Unlimited Selection No Late

More information

UNO Q&S Continuous Bed Ion Exchange Column. Instruction Manual

UNO Q&S Continuous Bed Ion Exchange Column. Instruction Manual UNO Q&S Continuous Bed Ion Exchange Column Instruction Manual Catalog Numbers 720-0001, 720-0003, 720-0005, 720-0021, 720-0023 and 720-0025 For Technical Service Call Your Local Bio-Rad Office or in the

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

Size Exclusion Chromatography

Size Exclusion Chromatography Size Exclusion Chromatography TOYOPEARL Resins for SEC TOYOPEARL TOYOPEARL TOYOPEARL HW-55 TOYOPEARL HW-65 TOYOPEARL TOSOH BIOSCIENCE TOSOH BIOSCIENCE LLC 56 Keystone Drive Montgomeryville, PA 896-967

More information

Blue Sepharose 6 Fast Flow

Blue Sepharose 6 Fast Flow GE Healthcare Life Sciences Instructions 71-7055-00 AI Affinity chromatography Blue Sepharose 6 Fast Flow Blue Sepharose 6 Fast Flow is Cibacron Blue 3G covalently attached to the Sepharose 6 Fast Flow

More information

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA

More information

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 Simultaneous determination of Telmisartan

More information

Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary

Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary Application Note Increasing the activity of monoclonal antibody isoforms by MCSGP Category Matrix Method Keywords Countercurrent chromatography, FPLC Antibodies MCSGP FPLC, Biobetters, MCSGP, countercurrent

More information

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

ION EXCHANGE FOR DUMMIES. An introduction

ION EXCHANGE FOR DUMMIES. An introduction ION EXCHANGE FOR DUMMIES An introduction Water Water is a liquid. Water is made of water molecules (formula H 2 O). All natural waters contain some foreign substances, usually in small amounts. The water

More information

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo

More information

Gas Chromatography Liner Selection Guide

Gas Chromatography Liner Selection Guide Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing

More information

Covalent Conjugation to Cytodiagnostics Carboxylated Gold Nanoparticles Tech Note #105

Covalent Conjugation to Cytodiagnostics Carboxylated Gold Nanoparticles Tech Note #105 Covalent Conjugation to Cytodiagnostics Carboxylated Gold Nanoparticles Tech Note #105 Background Gold nanoparticle conjugates have been widely used in biological research and biosensing applications.

More information

Marmara Üniversitesi Fen-Edebiyat Fakültesi Kimya Bölümü / Biyokimya Anabilim Dalı PURIFICATION AND CHARACTERIZATION OF PROTEINS

Marmara Üniversitesi Fen-Edebiyat Fakültesi Kimya Bölümü / Biyokimya Anabilim Dalı PURIFICATION AND CHARACTERIZATION OF PROTEINS EXPERIMENT VI PURIFICATION AND CHARACTERIZATION OF PROTEINS I- Protein isolation and dialysis In order to investigate its structure and properties a protein must be obtained in pure form. Since proteins

More information

Capillary HPLC Introduction

Capillary HPLC Introduction Capillary HPLC Introduction Capillary HPLC Liquid chromatography/mass spectrometry, LC/MS, is a revolutionary tool in the chemical and life sciences. LC/MS is accelerating chemical research by providing

More information

Analytical Test Report

Analytical Test Report Analytical Test Report Customer: Address (City, State): Purchase Order: Report Number: Project Number: Date Received: Date of Report: Test Location: Boulder, CO. Assay: Part Number: Amino Acids by HPLC

More information

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines

More information

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in

More information

Micromass LCT User s Guide

Micromass LCT User s Guide Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project

More information

Thermo Scientific Syncronis HPLC Columns. Technical Manual

Thermo Scientific Syncronis HPLC Columns. Technical Manual Thermo Scientific Syncronis HPLC Columns Technical Manual Thermo Scientific Syncronis HPLC Columns When developing a new method, one of the most important goals for the chromatographer is to achieve a

More information

Pipette Tips PACKARD BIOSCIENCE MultiPROBE II COMPATIBLE

Pipette Tips PACKARD BIOSCIENCE MultiPROBE II COMPATIBLE User Guide for Reversed-Phase Pipette Tips PACKARD BIOSCIENCE MultiPROBE II COMPATIBLE Notice The information in this document is subject to change without notice and should not be construed as a commitment

More information

Corning Spin-X UF Concentrator Selection and Use Guide

Corning Spin-X UF Concentrator Selection and Use Guide Corning Spin-X UF Concentrator Selection and Use Guide Contents Introduction.......................................................... 2 Choosing the Right Concentrator......................................

More information

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for

More information

Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns

Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns Jonathan P. Danaceau, Kenneth J. Fountain, and Erin E. Chambers Waters

More information

6 Characterization of Casein and Bovine Serum Albumin

6 Characterization of Casein and Bovine Serum Albumin 6 Characterization of Casein and Bovine Serum Albumin (BSA) Objectives: A) To separate a mixture of casein and bovine serum albumin B) to characterize these proteins based on their solubilities as a function

More information

HPLC Basics, Equipment, and Troubleshooting

HPLC Basics, Equipment, and Troubleshooting HPLC Basics, Equipment, and Troubleshooting HPLC Basics, Equipment, and Troubleshooting explains chromatography in practical terms from the ground up HPLC Basics, Equipment, and Troubleshooting HPLC Basics,

More information

PrepTip. Reverse Phase PrepTip User Guide

PrepTip. Reverse Phase PrepTip User Guide PrepTip Reverse Phase PrepTip User Guide All text, photographs and illustrations are copyrighted by Harvard Apparatus, Inc. 2004. PrepTip is a trademark of Harvard Apparatus, Inc. Harvard Apparatus 84

More information

Genomic DNA Extraction Kit INSTRUCTION MANUAL

Genomic DNA Extraction Kit INSTRUCTION MANUAL Genomic DNA Extraction Kit INSTRUCTION MANUAL Table of Contents Introduction 3 Kit Components 3 Storage Conditions 4 Recommended Equipment and Reagents 4 Introduction to the Protocol 4 General Overview

More information

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC The Essential CHROMacademy Guide Mobile Phase Optimization Strategies for Reversed Phase HPLC Webcast Notes Type your questions in the Submit Question box, located below the slide window You can enlarge

More information

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for

More information

Bringing Downstream Productivity into Phase with Upstream Antibody Production

Bringing Downstream Productivity into Phase with Upstream Antibody Production Bringing Downstream Productivity into Phase with Upstream Antibody Production Pete Gagnon, Validated Biosystems 3 rd International Monolith Symposium, Portoroz, May 30 June 4, 2008 The need for speed When

More information

ÄKTA system Training Guide

ÄKTA system Training Guide ÄKTA system Training Guide Explorer 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bioscience ÄKTAexplorer ÄKTAexplorer Buffers Injector Box amersham bioscience Monitor Conductivity

More information

ProteoMiner Protein Enrichment Kits

ProteoMiner Protein Enrichment Kits ProteoMiner Protein Enrichment Kits Instruction Manual Catalog # 163-3003 163-3006 163-3007 163-3008 163-3009 163-3010 163-3011 163-3012 For Technical Support, contact your local Bio-Rad office, or in

More information

Automation of Sample Preparation in Ion Chromatography

Automation of Sample Preparation in Ion Chromatography Automation of Sample Preparation in Ion Chromatography Praha, nd October 04 Ing. Roman Repáš ACO Sales Manager (IC and SP) Agenda Thermo Scientific Dionex Sample Preparation Products Automation of Sample

More information

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid

More information

A Practical Guide to Ion Chromatography AN INTRODUCTION AND TROUBLESHOOTING MANUAL. SeQuant. Innovators in Chemical Analysis

A Practical Guide to Ion Chromatography AN INTRODUCTION AND TROUBLESHOOTING MANUAL. SeQuant. Innovators in Chemical Analysis A Practical Guide to Ion Chromatography AN INTRODUCTION AND TROUBLESHOOTING MANUAL SeQuant Innovators in Chemical Analysis An introduction and troubleshooting manual 1 A PRACTICAL GUIDE TO ION CHROMATOGRAPHY

More information

Separation by Solvent Extraction

Separation by Solvent Extraction Experiment 3 Separation by Solvent Extraction Objectives To separate a mixture consisting of a carboxylic acid and a neutral compound by using solvent extraction techniques. Introduction Frequently, organic

More information

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate

More information

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.

More information

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.

More information

MabSelect SuRe. GE Healthcare Life Sciences

MabSelect SuRe. GE Healthcare Life Sciences GE Healthcare Life Sciences Instructions 11-0026-01 AD Affinity Chromatography MabSelect SuRe MabSelect SuRe is an alkali-tolerant protein A-derived medium for capturing monoclonal antibodies from large

More information

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Chapter 28: High-Performance Liquid Chromatography (HPLC) Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion

More information

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities ICMS and LCMS Determination of Ionic Liquids, Counterions, and Impurities Leo Wang, Detlef Jensen, and William Schnute Dionex Corporation, Sunnyvale, CA, USA; Dionex Corporation, Olten, Switzerland INTRODUCTION

More information

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,

More information

Waters 2695 Separations Module

Waters 2695 Separations Module Waters 2695 Separations Module Quick Start Guide 34 Maple Street Milford, MA 01757 71500269503, Revision A NOTICE The information in this document is subject to change without notice and should not be

More information

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -

More information

Process-scale purification of monoclonal antibodies polishing using Capto Q

Process-scale purification of monoclonal antibodies polishing using Capto Q GE Healthcare Life Sciences Application note 28-937-16 AB Ion exchange chromatography Process-scale purification of monoclonal antibodies polishing using Capto Q Summary Anionic exchange media are an industry

More information

TECHNICAL BULLETIN. FluoroTag FITC Conjugation Kit. Product Number FITC1 Storage Temperature 2 8 C

TECHNICAL BULLETIN. FluoroTag FITC Conjugation Kit. Product Number FITC1 Storage Temperature 2 8 C FluoroTag FITC Conjugation Kit Product Number FITC1 Storage Temperature 2 8 C TECHNICAL BULLETIN Product Description The FluoroTag FITC Conjugation Kit is suitable for the conjugation of polyclonal and

More information

ChromQuest 5.0 Chromatography Data System

ChromQuest 5.0 Chromatography Data System ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks

More information