1 ABUTMENT/ LOADING MEMBER, CONNECTION SCREW AND DENTAL IMPLANT SYSTEM WITH INTERNAL MORSE TAPER CONNECTION. RESULTS OF STATIC RESISTANCE TO COMPRESSION BENDING On request of: Medical Precision Implants, S.A. JULY 2011

2

3

4

5 INSTITUTO DE BIOMECÁNICA DE VALENCIA Abutment/loading member, connection screw and implant system with internal Morse taper connection. Results of static resistance to compression bending PROY10/0519 C o n t e n t s SIGNATURES AND AGREEMENT CONDITIONS 1. INTRODUCTION, ANTECEDENTS AND OBJECTIVES 2. MATERIAL AND METHODS 3. RESULTS 4. ANALYSIS OF RESULTS 5. CONCLUSIONS 6. REFERENCES ANNEX 1. INTRODUCTION, ANTECEDENTS AND OBJECTIVES This report presents the results related to static tests of compression bending for an abutment/loading member, connection screw and dental implant system with internal Morse taper connection (henceforth SYSTEM), according to the standard ISO 14801:2007 Dentistry Implants Dynamic fatigue test for endosseous dental implants. The tests have been requested by the company MEDICAL PRECISION IMPLANTS, S.A. sited in C/ Solidaridad, 2 (P.I Los Olivos), Madrid (SPAIN). 2. MATERIAL AND METHODS The description of the tested specimens is presented in Table 1. Dimensions of tested specimens are bold marked. Picture Part Reference Description Material Dimensions Straight abutment/ loading member CM551 CM558 Straight conical abutment Grade 5 - Ti D=3,5 mm / 4,0 mm / 5,0 mm L=6,5 mm/10,5 mm Final screw CM535 / CM536 Abutment-implant connection screw Grade 5 - Ti Torque=30 N cm Dental implant IPCM MPI Privilege Internal Connection - Conical Seal Grade 4 - Ti D=3,5 mm / 4,0mm / 5,0 mm L=7 mm/8,5 mm/10 mm/11,5 mm/13 mm/15 mm M1,6 / M2 Table 1. Description of the tested specimens. 5/11

6 INSTITUTO DE BIOMECÁNICA DE VALENCIA Abutment/loading member, connection screw and implant system with internal Morse taper connection. Results of static resistance to compression bending PROY10/0519 Below is shown a picture of the system. Figure 1. Abutment/loading member and dental implant system with internal Morse taper connection. The loading geometry for the tests (Figure 2) is described in section 5 of the standard ISO 14801:2007. Figure 3 shows a picture of the test set-up. 30º±1º Load A 30º±1º E B E C B A C D D F F 3±0,1mm 3±0,1mm 8±0,1mm 8±0,1mm A. A. Loading device device B. Nominal bone level level C. Connecting part part D. Hemispherical loading member D. E. Hemispherical Dental implant body loading member E. F. Dental Specimen implant holder body F. Specimen holder Figure 2. Loading geometry for the tests (ISO 14801:2007). 6/11

7 Load (N) INSTITUTO DE BIOMECÁNICA DE VALENCIA Abutment/loading member, connection screw and implant system with internal Morse taper connection. Results of static resistance to compression bending PROY10/0519 Figure 3. Mechanical testing of the system. The static resistance to compression bending of the system has been evaluated with the aim of determining the load and displacement at the rupture point, the load and displacement at the yield point and the stiffness (Figure 4). Testing environment conditions have been the ones indicated by the standard ISO 14801:2007. Yield point Rupture point Stiffness Displacement (mm) Figure 4. Parameters of the static resistance test. 7/11

8 INSTITUTO DE BIOMECÁNICA DE VALENCIA Abutment/loading member, connection screw and implant system with internal Morse taper connection. Results of static resistance to compression bending PROY10/0519 Figure 5. Test sample after static resistance test. 3. RESULTS Table 2 presents the results of the static tests for the system. The values of this table correspond to the mean and standard deviation of the five specimens tested. The results, as well as mode of failure of the specimens, are described in detail in the ANNEX. System Stiffness (N/mm) Load at the yield point (N) Displacement at the yield point (mm) Load at the rupture point (N) Displacement at the rupture point (mm) MPI Privilege Implant System Internal Morse taper connection 1119 ± ± ± ± ± 0.23 Table 2. Results of static tests. 4. ANALYSIS OF RESULTS The failure to static resistance of the tested system has been produced due to the material yielding. The load at the yield point for the system with internal Morse taper connection has been 520 N (mean of five specimens), with the loading geometry indicated by the standard ISO 14801:2007. Due to the value of the diameter, the tested systems are only used in lower incisors and upper lateral incisors. Scientific studies about maximum bite forces with natural teeth in several materials have been found, which have observed values between 50 and 370 N for the maximum bite force with incisive teeth (Paphangkorakit and Osborn, 1997; 8/11

9 INSTITUTO DE BIOMECÁNICA DE VALENCIA Abutment/loading member, connection screw and implant system with internal Morse taper connection. Results of static resistance to compression bending PROY10/0519 Fontijn-Tekamp et al., 2000; Regalo et al., 2008). The measured values in the static tests widely exceed these load values. 5. CONCLUSIONS In conclusion, it can be stated that the results obtained in the static tests for the system with internal Morse taper connection are satisfactory, as the systems have withstood static load values higher than the usual biting forces. In the static tests, the system with internal Morse taper connection has reached a load value at the yield point of 520 N. The maximum bite force with incisive teeth ranges from 50 to 370 N. Therefore, results of the static tests are satisfactory because the measured static loads are higher than the ones expected during the usual activity of the system. 9/11

10 INSTITUTO DE BIOMECÁNICA DE VALENCIA Abutment/loading member, connection screw and implant system with internal Morse taper connection. Results of static resistance to compression bending PROY10/ REFERENCES FONTIJN-TEKAMP FA, SLAGTER AP, VAN DER BILT A, VAN T HOF MA, WITTER DJ, KALK W, JANSEN JA. Biting and chewing in overdentures, full dentures, and natural dentitions. Journal of Dental Research 2000; 79(7); ISO 14801:2007. Dentistry Implants Dynamic fatigue test for endosseous dental implants. PAPHANGKORAKIT J, OSBORN JW. The effect of pressure on a maximum incisal bite force in a man. Archives of Oral Biology 1997; 42(1); REGALO SC, SANTOS CM, VITTI M, REGALO CA, DE VASCONCELOS PB, MESTRINER WJR, SEMPRINI M, DIAS FJ, HALLAK JE, SIÉSSERE S. Evaluation of molar and incisor bite force in indigenous compared with white population in Brazil. Archives of Oral Biology 2008; 53(3); /11

11 INSTITUTO DE BIOMECÁNICA DE VALENCIA Abutment/loading member, connection screw and implant system with internal Morse taper connection. Results of static resistance to compression bending PROY10/0519 ANNEX REPORT. Dental implants. Evaluation of static resistance to compression bending. 11/11

12 IBV- ENSAYOS DENTAL IMPLANTS. EVALUATION OF STATIC RESISTANCE TO COMPRESSION BENDING PROY10/0519. R E P O R T DENTAL IMPLANTS. EVALUATION OF STATIC RESISTANCE TO COMPRESSION BENDING Direct to: Prepared by: MEDICAL PRECISION IMPLANTS, S.A. INSTITUTO DE BIOMECÁNICA DE VALENCIA Universidad Politécnica de Valencia Edificio 9C Camino de Vera s/n E VALENCIA (ESPAÑA) Tel.: Fax: JULY /6

13

14 IBV- ENSAYOS DENTAL IMPLANTS. EVALUATION OF STATIC RESISTANCE TO COMPRESSION BENDING PROY10/0519 Contents SIGNATURES AND AGREEMENT CONDITIONS 1. INTRODUCTION AND OBJECTIVES 2. MATERIAL AND METHODS 3. RESULTS 3/6

15 IBV- ENSAYOS DENTAL IMPLANTS. EVALUATION OF STATIC RESISTANCE TO COMPRESSION BENDING PROY10/ INTRODUCTION AND OBJECTIVES The objective of this test is to evaluate the static resistance to compression bending of a dental implant with morse cone internal conection configured by pillar-load body and connection screw. The tests have been requested by the company MEDICAL PRECISION IMPLANTS, S.A., sited in: Pol. Ind. Los Olivos Getafe (MADRID). 2. MATERIAL AND METHODS The test sample and its codification appear in the following table: CODE DESCRIPTION UNITS Dental implant with morse cone internal conection, pillar-load body and connection screw MU The test samples will be given back to the client after the tests. The tests have been performed using an INSTRON 8874/429 universal testing device. The tests have been done on 8 th July 2011, with an ambient temperature of 23ºC and humidity of 45%. Before testing, the pillar-load body is screwed on the connection screw by means of a torque wrench supplied by the customer (torque 30Ncm). For the test, the implant is inserted between the actuators of the testing machine and compression loads are applied as shown in the following picture. 4/6

16 IBV- ENSAYOS DENTAL IMPLANTS. EVALUATION OF STATIC RESISTANCE TO COMPRESSION BENDING PROY10/0519 Load A. Loading device B. Nominal bone level C. Connecting part D. Hemispherical loading member E. Dental implant body F. Specimen holder The conditions of test have been: The test has been performed by displacement control. The displacement velocity of the actuator has been 0.01 mm/s. The test has been finished when the implant failure has been reached. 3. RESULTS The following table presents the individual results, mean values and standard deviations of: stiffness, load at the yield point, displacement at the yield point and load and displacement at the rupture point of the tested dental implant. Code MU Stiffness (N/mm) Load Yield point (N) Displ. Yield point (mm) Load Rupture point (N) Displ. Rupture point (mm) Observations [1] [1] [1] [1] [1] 1119 ± ± ± ± ± /6

17 IBV- ENSAYOS DENTAL IMPLANTS. EVALUATION OF STATIC RESISTANCE TO COMPRESSION BENDING PROY10/0519 [1] All systems fail because of rupture on morse cone-screw conection. (Figure 1). Figure 1 6/6

Dentistry Fatigue test for endosseous dental implants. Report No.: 1 ASTRA System Ti Design

Dentistry Fatigue test for endosseous dental implants. Report No.: 1 ASTRA System Ti Design Dentistry Fatigue test for endosseous dental implants Report No.: 1 ASTRA System Ti Design Client: NT - Trading Time of execution: Mai 2011 Subject of work: - - S - 24285 Abutment Ti DesignTM - RESULTS

More information

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare.

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare. Use of CP Titanium and Titanium alloys for Dental Implants Steve Hurson, Nobel Biocare U.S.A. Yorba Linda CA Abstract Professor Per Ingvar Branemark of Sweden published his results on 15 years of research

More information

NARROW DIAMETER implant

NARROW DIAMETER implant ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball

More information

Implants of high-performance ceramic. vk-lange 12.03.2007 14:44 Uhr Seite 1. One-piece precision implant. Made in Germany

Implants of high-performance ceramic. vk-lange 12.03.2007 14:44 Uhr Seite 1. One-piece precision implant. Made in Germany OMNIS Implants of high-performance ceramic vk-lange 12.03.2007 14:44 Uhr Seite 1 One-piece precision implant C M Y CM MY CY CMY K Dieter Lange Verkaufsleiter Labor Made in Germany creamed nge 12.03.2007

More information

Supervisors: Dr. Farhan Raza Khan

Supervisors: Dr. Farhan Raza Khan 1 Presenter: Dr. Sana Ehsen Supervisors: Dr. Farhan Raza Khan 2 A dental implant (also known as an endosseous implant or fixture) is a surgical component that interfaces with the bone of the jaw to support

More information

Information on the product

Information on the product Information on the product Information on the product -Implantat.doc Page 1 of 15 This document composes of the following items 1. Name of the product 2. Raw materials 3. Structure and dimensions 3.1.

More information

ALL-CERAMIC DENTAL IMPLANT SOLUTIONS

ALL-CERAMIC DENTAL IMPLANT SOLUTIONS ALL-CERAMIC DENTAL IMPLANT SOLUTIONS Scientific Evidence Bone-to-implant contact of 78% at 3 months. One piece implant = no prosthetic connections References 1. One-year follow-up of first consecutive

More information

Mechanical joining Tensile shear testing of single joint specimens Specimen dimensions and test procedure

Mechanical joining Tensile shear testing of single joint specimens Specimen dimensions and test procedure Mechanical joining Tensile shear testing of single joint specimens Specimen dimensions and test procedure Assemblage mécanique Essais quasi statiques de traction-cisaillement des échantillons à jonction

More information

DESS. Screws. Tijuana Ventas: (664) 685 6294/95 hirambogarin@dabocorp.com. For all major implant systems!! www.dabocorp.com

DESS. Screws. Tijuana Ventas: (664) 685 6294/95 hirambogarin@dabocorp.com. For all major implant systems!! www.dabocorp.com Screws Screws for definitive use. Made from medical grade 5 ELI Titanium Integrity and soundness guaranteed by stress tests of up to 150% of their nominal torques. Their designs ensure a perfect fit with

More information

CAD/CAM technology supporting successful implant therapy

CAD/CAM technology supporting successful implant therapy CAD/CAM technology supporting successful implant therapy Suheil M. Boutros, DDS, MS, Manuel Fricke, DT Modern implantology opens up new treatment options for individuals with only minimal or no remaining

More information

Product Catalog. ERA Mini TM Dental Implant System

Product Catalog. ERA Mini TM Dental Implant System Product Catalog ERA Mini TM Dental System How to Order Return Procedure To order, call Zimmer Dental Customer Service between 6:00 a.m. and 4:30 p.m. PST, Monday through Friday, or fax your order to 1

More information

Convertible Abutment. Introduction. Prosthetic Flow Diagram for GS Convertible Abutment. Configuration of GoldCast Cylinder

Convertible Abutment. Introduction. Prosthetic Flow Diagram for GS Convertible Abutment. Configuration of GoldCast Cylinder Prosthetic Flow Diagram for GS Cement & Screw & Combination Retained Restoration for Mini, Regular, Wide - Abutment Level Impression Taking Introduction Cylinder Screw(WCC) Three types of cylinder (combination,

More information

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment by Dr. Ronald Jung and Master Dental Technician Xavier Zahno Initial situation

More information

for Kitchen KanaStone Marble series Granite series Natural stone composite panel

for Kitchen KanaStone Marble series Granite series Natural stone composite panel for Kitchen Marble series Granite series Natural stone composite panel for Kitchen Marble series Granite series Natural stone composite panel Achieved light weight, strength and workability exceeding common

More information

Ozone, mineral oil HLP to DIN , triglyzeride (rape seed oil) HETG, synthetic ester HEES to VDMA

Ozone, mineral oil HLP to DIN , triglyzeride (rape seed oil) HETG, synthetic ester HEES to VDMA Company standard Flexible rotary Type KD AB-E - 2001-08-27 Replaces: Issue 2000-01-27 1 Description Flexible rotary coupling are used to compensate for radial, angular and axial misalignment of the shafts

More information

Implant Systems MAX LOGIC MONO. Today s name in tomorrow s implant technology

Implant Systems MAX LOGIC MONO. Today s name in tomorrow s implant technology Implant Systems MAX LOGIC MONO Today s name in tomorrow s implant technology Implant Systems Etgar Medical Implant Systems brings to the market an entire spectrum of dental implant products based on its

More information

PERIPROSTHETIC IMPLANTS

PERIPROSTHETIC IMPLANTS PERIPROSTHETIC IMPLANTS PRODUCT OVERVIEW CLINICAL SOLUTIONS Periprosthetic fractures present unique challenges, such as how to gain fixation when the medullary canal is occupied. Special techniques and

More information

Online Vocational Training course on laparoscopy's ergonomics for surgeons and laparoscopic instruments' designers

Online Vocational Training course on laparoscopy's ergonomics for surgeons and laparoscopic instruments' designers laparoscopic instruments' designers 527985-LLP-2012-1-ES-LEONARDO-LMP http://www.adam-europe.eu/adam/project/view.htm?prj=9889 1 LEONARDO-LMP) Project Information Title: Project Number: Project Type: Year:

More information

The advantages of computerized 3D dentistry Dental Architecture

The advantages of computerized 3D dentistry Dental Architecture The advantages of computerized 3D dentistry Dental Architecture Dr. Ehud Teperovich Philip Segal Raanana Israel Part one: Digital Planning Drawing Board enhanced possibilities with SimPlant Clinic for

More information

September 13-16, 2009

September 13-16, 2009 September 13-16, 2009 Waikoloa, HI September 13-16, 2009, Waokoloa, Waikoloa, HI Biomimetic i Tooth Implants by Electron Beam Melting Mari Koike, DDS, PhD 1 ; Gilbert Chahine 2 ; Radovan Kovacevic, PhD

More information

IMPLANT DENTISTRY EXAM BANK

IMPLANT DENTISTRY EXAM BANK IMPLANT DENTISTRY EXAM BANK 1. Define osseointegration. (4 points, 1/4 2. What are the critical components of an acceptable clinical trial? (10 points) 3. Compare the masticatory performance of individuals

More information

MAY 2 1 2913. Attachment4. 5 10(k) Summary 21 CER 872.3640

MAY 2 1 2913. Attachment4. 5 10(k) Summary 21 CER 872.3640 MAY 2 1 2913 Traditional 510(k) Submission DIO tjf HSA Internal Sub-Merged Implant System Attachment4 5 10(k) Summary This summary of 5 10(K) safety and effectiveness information is being submitted in

More information

Improving Esthetics with Sequential Treatment Planning and Implant-Retained Dentures

Improving Esthetics with Sequential Treatment Planning and Implant-Retained Dentures Improving Esthetics with Sequential Treatment Planning and Implant-Retained Dentures by Timothy F. Kosinski, DDS, MAGD While oral function is the primary concern for most patients, the importance of esthetics

More information

RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS

RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education. CONTENTS Product

More information

USA RUSSIA GREECE ROMANIA OFFERING EXCITING DIGITAL OPPORTUNITIES AROUND THE GLOBE

USA RUSSIA GREECE ROMANIA OFFERING EXCITING DIGITAL OPPORTUNITIES AROUND THE GLOBE EXPERIENCE USA RUSSIA GREECE OFFERING EXCITING DIGITAL OPPORTUNITIES AROUND THE GLOBE ROMANIA FEB. 2016 DIGITAL DIGITAL DENTISTRY Technology is Changing Dentistry Dentistry is evolving, starting from diagnosis

More information

Contents. Cement retained restoration. Screw retained restoration. Overdenture retained restoration. TS Implant System. 70 ComOcta Gold Abutment

Contents. Cement retained restoration. Screw retained restoration. Overdenture retained restoration. TS Implant System. 70 ComOcta Gold Abutment Contents TS Implant System Cement retained restoration Screw retained restoration 06 Cement-retained bridges with the Solid abutment system (non- 72 Screw retained crown with the ComOcta Gold abutment

More information

Don t Let Life Pass You By Because Of Oral Bone Loss

Don t Let Life Pass You By Because Of Oral Bone Loss Don t Let Life Pass You By Because Of Oral Bone Loss Ask For Dental Implant Solutions From BIOMET 3i Scan With Your Smartphone! In order to scan QR codes, your mobile device must have a QR code reader

More information

MINIATURE METAL BELLOWS COUPLINGS

MINIATURE METAL BELLOWS COUPLINGS VERSATILE AND PRECISE. MINIATURE METAL BELLOWS COUPLINS SERIES MK 0.05 10 Nm THE ULTIMATE COUPLIN FROM 0.05 10 Nm BACKLASH FREE MINIATURE BELLOWS COUPLINS Areas of application: Ideal for precise transmission

More information

Ridge Reconstruction for Implant Placement

Ridge Reconstruction for Implant Placement Volume 1, No. 5 July/August 2009 The Journal of Implant & Advanced Clinical Dentistry Ridge Reconstruction for Implant Placement 2 Hours of CE Credit Oral Implications of Cancer Chemotherapy Immediate

More information

BICON DENTAL IMPLANTS

BICON DENTAL IMPLANTS BICON DENTAL IMPLANTS The Bicon Dental Implant System, since 1985, has offered discerning dentists the ability to provide secure implant restorations that look, feel, and function like natural teeth. With

More information

SKY Titanium Implants

SKY Titanium Implants SKY Titanium Implants Easy & safe prosthetically versatile 3D osseo connect surface (ocs ) Depending on the implant s positioning, a corresponding coronal implant design is needed for optimally fulfilling

More information

E-learning tool about spine surgery techniques for orthopaedic surgeons and biomedical engineers

E-learning tool about spine surgery techniques for orthopaedic surgeons and biomedical engineers biomedical engineers ES/07/LLP-LdV/TOI/149037 1 Project Information Title: Project Number: Project Type: Year: 2007 Status: Marketing Text: Summary: E-learning tool about spine surgery techniques for orthopaedic

More information

DNV BUSINESS ASSURANCE MANAGEMENT SYSTEM CERTIFICATE

DNV BUSINESS ASSURANCE MANAGEMENT SYSTEM CERTIFICATE DNV BUSINESS ASSURANCE MANAGEMENT SYSTEM CERTIFICATE Certificate No. 118727-2012-AQ-ARG-NA This is to certify that the Management System of: STEELDEC S.A Valparaiso 4100 Parque Ind. Tortuguitas, (1667)

More information

Dental Morelli Ltda.

Dental Morelli Ltda. K Dental Morelli Ltda. JAN 82014 OR TOOON tia SECTION 7 510(k) SUMMARY Proprietary Name Orthodontic Miniscrew for Absolute Anchorage Date Prepared December 1, 2013 Submitter Official Contact DENTAL MORELLI

More information

Duran. Tubing, Capillary and Rod of Borosilicate Glass 3.3

Duran. Tubing, Capillary and Rod of Borosilicate Glass 3.3 Duran Tubing, Capillary and Rod of Borosilicate Glass 3.3 Duran (borosilicate 3.3): Tubing Outside Diameter Wall Thickness per Tube Carton Contents mm mm g/1500 mm Number of Tubes 3 ±0.14 0.7 ±0.04 17

More information

INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS. Published: April 1999

INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS. Published: April 1999 INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS Published: April 1999 CONTENTS 1 Introduction 2 Test Procedures General 3 Standards and Specific Test Procedures

More information

PLQE. The base of the economy gearboxes with square output flange. Economy Line. Formerly PLE

PLQE. The base of the economy gearboxes with square output flange. Economy Line. Formerly PLE 20 ormerly PL The base of the economy gearboxes with square output flange The PL series with square output flange. A powerful alternative for additional higher radial and axial loads. Low backlash High

More information

Dental Implant Options in Atrophic Jaws

Dental Implant Options in Atrophic Jaws Dental Implant Options in Atrophic Jaws Orthopedic Application Jay B. Reznick, D.M.D., M.D. Diplomate, American Board of Oral and Maxillofacial Surgery Tarzana, CA Endopore Dental Implant System Screw-Type

More information

ITI Small Diameter (SD) external hex 3.25 / 3.3 external hex. ITI Standard 4.8 internal non-octagon Morse taper 4.

ITI Small Diameter (SD) external hex 3.25 / 3.3 external hex. ITI Standard 4.8 internal non-octagon Morse taper 4. ITI Standard Implant Dental Implant System The ITI Dental Implant System was developed in Switzerland by the Institute Straumann and Dr. ndre Schroeder, President of the University ern. ITI implants are

More information

Straumann Dental Implant System. Implant Selection Guide.

Straumann Dental Implant System. Implant Selection Guide. Straumann Dental Implant System. Implant Selection Guide. STRAUMANN's IMPLANT PORTFOLIO The Straumann Dental Implant System offers two implant lines with diverse body and neck designs ranging from the

More information

PROSTHETIC PROCEDURE. for HG IMPLANT SYSTEM

PROSTHETIC PROCEDURE. for HG IMPLANT SYSTEM PROSTHETIC PROCEDURE for HG IMPLANT SYSTEM PROSTHETIC PROCEDURE for HG IMPLANT SYSTEM HG Implant System Contents Cement retained restoration Rigid abutment When abutment reduction is unnecessary When abutment

More information

SCD Case Study. Treatment Considerations for Implant Rehabilitation

SCD Case Study. Treatment Considerations for Implant Rehabilitation SCD Case Study Treatment Considerations for Implant Rehabilitation Multiple surgical and restorative factors play a role in the treatment planning of implant restorations for the edentulous patient (Ali

More information

CARES Technical Approval Report TA1-B 5025

CARES Technical Approval Report TA1-B 5025 CARES Technical Approval Report TA1-B 5025 Issue 2 December 2010 G Tech Splicing Pvt. Ltd. Spliceman Barbreak TM Couplers Assessment of the Spliceman Barbreak TM Coupler Product and Quality System for

More information

STATIC COEFFICIENT OF FRICTION MEASUREMENT USING TRIBOMETER. Static COF 0.00 0.0 0.5 1.0 1.5 2.0. Time(min) Prepared by Duanjie Li, PhD

STATIC COEFFICIENT OF FRICTION MEASUREMENT USING TRIBOMETER. Static COF 0.00 0.0 0.5 1.0 1.5 2.0. Time(min) Prepared by Duanjie Li, PhD STATIC COEFFICIENT OF FRICTION MEASUREMENT USING TRIBOMETER 0.20 Static COF Coefficient of Friction 0.15 0.10 0.05 0.00 0.0 0.5 1.0 1.5 2.0 Time(min) Prepared by Duanjie Li, PhD 6 Morgan, Ste156, Irvine

More information

Technical Data. 7. Bearing Fits. 7.1 Interference. 7.2 Calculation of interference F B LLLLLLLLL( A-54

Technical Data. 7. Bearing Fits. 7.1 Interference. 7.2 Calculation of interference F B LLLLLLLLL( A-54 Technical Data 7. Bearing Fits 7.1 Interference For rolling s the rings are fixed on the or in the housing so that slip or movement does not occur between the mated surface during operation or under. This

More information

Tooth Loss and Therapy: Eating and Smiling with confidence again

Tooth Loss and Therapy: Eating and Smiling with confidence again Tooth Loss and Dental Implant Therapy: Eating and Smiling with confidence again Tooth loss is an extremely common problem for adults in our society. The average mouth of a person above the age of 40 is

More information

Multi-unit abutment bar overdenture

Multi-unit abutment bar overdenture This prosthetic technique module may contain references to the complete Prosthetic Manual (L02015). To download the full Prosthetic Manual, please visit www.biohorizons.com/prostheticmanual.aspx Use this

More information

Advanced bolt assessment through process automation

Advanced bolt assessment through process automation Titelmasterformat Bolt Assessment inside durch ANSYS Klicken bearbeiten Advanced bolt assessment through process automation Contents Short presentation of VDI 2230 Workflow with Bolt Assessment inside

More information

Implant Parts. A Radford Heath Guide http://www.radfordheath.com 1

Implant Parts. A Radford Heath Guide http://www.radfordheath.com 1 Implant Parts A Radford Heath Guide http://www.radfordheath.com 1 Disclaimer The information given in this document has been provided in good faith for basic information purposes only and the information

More information

Milling tools made by JONGEN! Face Milling FP 554

Milling tools made by JONGEN! Face Milling FP 554 Milling tools made by JONGEN! Face Milling FP 554 The Tool Especially efficient face milling cutter for the roughing and finishing machining Axial depth of cut max. 7 mm, with 8 effective cutting edges

More information

High Pressure Gear Pumps KP 0

High Pressure Gear Pumps KP 0 High Pressure Gear Pumps KP 0 2 KRACHT GmbH Gewerbestr. 20 58791 Werdohl, Germany fon +49(0)23 92/935-0 fax +49(0)23 92/935 209 mail info@kracht.eu web www.kracht.eu Construction According to its design,

More information

In Problems #1 - #4, find the surface area and volume of each prism.

In Problems #1 - #4, find the surface area and volume of each prism. Geometry Unit Seven: Surface Area & Volume, Practice In Problems #1 - #4, find the surface area and volume of each prism. 1. CUBE. RECTANGULAR PRISM 9 cm 5 mm 11 mm mm 9 cm 9 cm. TRIANGULAR PRISM 4. TRIANGULAR

More information

JIS G3472 Electric Resistance Welded Carbon Steel Tubes for Automobile Structural Purposes

JIS G3472 Electric Resistance Welded Carbon Steel Tubes for Automobile Structural Purposes JIS G3472 Electric Resistance Welded Carbon Steel Tubes for Automobile Structural Purposes 1. Scope This Japanese Industrial Standard specifies the electric resistance welded carbon steel tubes, hereinafter

More information

Two most common lock nut groups: 1-Prevailing Torque Type Lock Nuts:

Two most common lock nut groups: 1-Prevailing Torque Type Lock Nuts: Two most common lock nut groups: 1. PREVAILING TORQUE a design feature of the lock nut produces friction between threads of mated components thereby increasing the force needed to tighten as well as loosen

More information

Strong teeth with prosthesis! ANKYLOS SynCone

Strong teeth with prosthesis! ANKYLOS SynCone Strong teeth with prosthesis! ANKYLOS SynCone Implants for life vorher: Initial situation: The bone in the toothless jaw degenerates and the prosthesis jiggle. ANKYLOS SynCone gives you that secure feeling

More information

METHOD OF STATEMENT FOR STATIC LOADING TEST

METHOD OF STATEMENT FOR STATIC LOADING TEST Compression Test, METHOD OF STATEMENT FOR STATIC LOADING TEST Tension Test and Lateral Test According to the American Standards ASTM D1143 07, ASTM D3689 07, ASTM D3966 07 and Euro Codes EC7 Table of Contents

More information

. Simplification. High Performance. Safety and Longevity. The fundamental goal is to restore the quality of life of patients.

. Simplification. High Performance. Safety and Longevity. The fundamental goal is to restore the quality of life of patients. Catalog 2 P-I product line was developed by te Osseointegration pioneer, Professor Per-Ingvar Brånemark, jointly wit experienced scientists in world recognized entities to meet modern Implant Dentistry

More information

ABOUT US. The Complete and Comprehensive solution for implants! Edison Medical. Edison Medical s Quality Strategy:

ABOUT US. The Complete and Comprehensive solution for implants! Edison Medical. Edison Medical s Quality Strategy: IMPLANTS CATALOG ABOUT US Edison Medical The Complete and Comprehensive solution for implants! Edison Medical Ltd. is engaged in the import, export, marketing and distribution of dental Implants and other

More information

Don t miss life because of missing teeth, say...

Don t miss life because of missing teeth, say... Don t miss life because of missing teeth, say... Consequences of missing teeth Bone loss is responsible for many of the problems encountered after tooth loss. When one or more teeth are lost, the bone

More information

Straumann Dental Implant System. Fax Order Form Collection.*

Straumann Dental Implant System. Fax Order Form Collection.* Straumann Dental Implant System. Fax Order Form Collection.* * The NNC Fax Order Form is still a stand alone document due to the unique restorative portfolio Straumann Roxolid SLActive Tissue Level Implants

More information

ExpertEase Surgery Step-by-Step. new. Surgery

ExpertEase Surgery Step-by-Step. new. Surgery ExpertEase Surgery Step-by-Step new Surgery ExpertEase computer-guided planning and surgical guides In the first stage the future denture is designed with reference to esthetic, functional and phonetic

More information

Forces in Permanent Magnets Team Workshop Problem 23

Forces in Permanent Magnets Team Workshop Problem 23 Forces in Permanent Magnets Team Workshop Problem 23 N. Ida* and J.P.A. Bastos** * Department of Electrical Engineering The University of Akron, Akron, OH. 44325-3904, USA ** Universidade Federal de Santa

More information

Basic information on the. Straumann Variobase Abutment.

Basic information on the. Straumann Variobase Abutment. Basic information on the Straumann Variobase Abutment. 1 Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education. Contents.

More information

Mandrel system MIS. www.microcentric.de

Mandrel system MIS. www.microcentric.de Mandrel system MIS MicroCentric GmbH Berblingerstr. 16 71254 itzingen / Germany Tel.: +49 (0) 7156 17819-0 Fax: +49 (0) 7156 17819-20 E-Mail: info@microcentric.de www.microcentric.de Products in this catalog:

More information

PRECISION, RELIABILITY, SIMPLICITY. simeda. customized. & solutions. Digital performance

PRECISION, RELIABILITY, SIMPLICITY. simeda. customized. & solutions. Digital performance PRECISION, RELIABILITY, SIMPLICITY simeda customized & solutions Digital performance 2 Simeda Scan The high-precision scanner for implantology! With, a company specialised in industrial metrology, Simeda

More information

Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment.

Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment. Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment. Dr. med. dent. David McFadden, Dallas County, USA Initial situation (single X-ray) Tooth 16

More information

Digital input modules

Digital input modules 8 172 TX-I/O Digital input modules TXM1.8D TXM1.16D Two fully compatible versions: TXM1.8D: 8 inputs, each with a three-color LED (green, yellow or red) TXM1.16D: As TXM1.8X, but 16 inputs, each with a

More information

BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT

BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT B K Biswas 1 S Bag 2 & S Pal 3 1. Dept. of Dental Surgery, Associate Professor, KPC Medical College & Hospital, Kolkata,

More information

Lecture Slides. Chapter 8. Screws, Fasteners, and the Design of Nonpermanent Joints

Lecture Slides. Chapter 8. Screws, Fasteners, and the Design of Nonpermanent Joints Lecture Slides Chapter 8 Screws, Fasteners, and the Design of Nonpermanent Joints The McGraw-Hill Companies 2012 Chapter Outline Reasons for Non-permanent Fasteners Field assembly Disassembly Maintenance

More information

LOMAS / MONDEFIT ORTHODONTICS. The most innovative anchorage method for tooth correction

LOMAS / MONDEFIT ORTHODONTICS. The most innovative anchorage method for tooth correction LOMAS / MONDEFIT The most innovative anchorage method for tooth correction No extraction of healthy teeth Completion of tooth correction without gaps Symmetrical, beautiful smile For young people and young

More information

Edentulous patients. Straumann Dental Implant System

Edentulous patients. Straumann Dental Implant System Edentulous patients treatment solutions Straumann Dental Implant System Restoration solutions for all patient requirements Basic Basic solutions: standard retentive elements hold the denture securely in

More information

North American Stainless Long Products

North American Stainless Long Products North American Stainless Long Products 1 North American Stainless North American Stainless is part of the most competitive stainless steel manufacturing group in the world, Acerinox, S.A. The development

More information

A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS

A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS Evandro Afonso Sartori, evandroasartori@yahoo.com.br Josué Ricardo Broilo Luiz Oscar

More information

Self-Aligning Lowest Profile

Self-Aligning Lowest Profile Self-Aligning Lowest Profile * Lowest Profile * Self-Aligning * Dual Retention * Long Lasting * Choice of Retention * Easy Solution for Divergence * Money-Back Guarantee *LOCATOR Patent No. 6,030,219 SELF

More information

Dynamic Measurement of Brake Pads Material Parameter

Dynamic Measurement of Brake Pads Material Parameter VDA Dynamic Measurement of Brake Pads Material Parameter 315 This non-binding recommendation by the German Association of the Automotive Industry (VDA) has the following objectives: This standard treats

More information

All-on-4 treatment concept with NobelSpeedy Groovy

All-on-4 treatment concept with NobelSpeedy Groovy All-on-4 treatment concept with NobelSpeedy Groovy Product overview Immediate Function for high patient satisfaction Immediately loaded fixed provisional prosthesis on the day of surgery. Immediate improvement

More information

HAHN TAPERED IMPLANT SYSTEM. Product Catalog. April

HAHN TAPERED IMPLANT SYSTEM. Product Catalog. April Product Catalog April 2016 HAHN TAPERED IMPLANT SYSTEM www.hahnimplant.com 1 HAHN TAPERED IMPLANT SYSTEM 2 Introducing the Hahn Tapered Implant System Combining decades of clinical experience with cutting-edge

More information

Technical data machine connector M12 1, 4-pole according to IEC ; FIXCON-capable

Technical data machine connector M12 1, 4-pole according to IEC ; FIXCON-capable Data Sheet 902815 Page 1/7 JUMO Dtrans T100 Screw-in RTD temperature probe with/without transmitter for temperatures from -50 to +260 C EHEDG certification complies with ROHS for EU and China configuration

More information

FABRICATING CUSTOM ABUTMENTS

FABRICATING CUSTOM ABUTMENTS FABRICATING CUSTOM ABUTMENTS LUC AND PATRICK RUTTEN How much should a Dental Technician know about the clinical aspects of implantology? The answer is clear: as much as possible. This is the distinction

More information

Development of U-shaped Steel Damper for Seismic Isolation System

Development of U-shaped Steel Damper for Seismic Isolation System UDC 624. 042. 7 Development of U-shaped Steel Damper for Seismic Isolation System Kazuaki SUZUKI* 1 Atsushi WATANABE* 1 Eiichiro SAEKI* 1 Abstract Seismic isolation system was widely admitted after Hanshin-Awaji

More information

March 20, 2015. Dear Mr. Chen:

March 20, 2015. Dear Mr. Chen: DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002 March 20, 2015 Biomate

More information

ASTM D 1599 Standard Test Method for Resistance to Short-Time Hydraulic Pressure of Plastic Pipe, Tubing, and Fittings

ASTM D 1599 Standard Test Method for Resistance to Short-Time Hydraulic Pressure of Plastic Pipe, Tubing, and Fittings ASTM D 1599 Standard Test Method for Resistance to Short-Time Hydraulic Pressure of Plastic Pipe, Tubing, and Fittings This test method establishes the short-time hydraulic failure pressure of pipe and

More information

HUS-HR,CR. Anchor Fastening. Screw anchor. Technology Manual HUS-HR HUS-CR. Screw anchor. Version /

HUS-HR,CR. Anchor Fastening. Screw anchor. Technology Manual HUS-HR HUS-CR. Screw anchor. Version / Anchor Fastening Technology Manual HUS-HR HUS-CR Version 2015-09 09 / 2015 1 HUS-HR, CR, stainless steel Anchor version HUS-HR 6 / 8 / 10 / 14 Stainless steel concrete Screw with hexagonal head HUS-CR

More information

Dental Updates. Excerpted Article e-mail: re777@comcast.net. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS

Dental Updates. Excerpted Article e-mail: re777@comcast.net. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS ¼ ½ ¾ µ mw/cm 2 Volume 17; 2007 Dental Updates "CUTTING EDGE INFORMATION FOR THE DENTAL PROFESSIONAL " 200 SEMINARS AND 30 JOURNALS REVIEWED YEARLY FOR THE LATEST, CUTTING EDGE INFORMATION Excerpted Article

More information

Pilot Operated Check Valves Model SV/SL (Series 4X)

Pilot Operated Check Valves Model SV/SL (Series 4X) R 68/06.98 Size 0 to 0 Pilot Operated Check Valves Model SV/SL (Series ) 600 PSI (5 bar)... 5 GPM (550 L/min) R 68/06.98 Replaces: 05.9 Pilot operated check valve, with leak-free closure in checked direction

More information

Series 6000 Torque measured metal bellow coupling

Series 6000 Torque measured metal bellow coupling Properties Free of float metal bellow coupling with integrated torque measurement Non-contact measurement system, high robustness High torsional stiffness Limited torque of inertia Performance Measurement

More information

While the prosthetic rehabilitation of

While the prosthetic rehabilitation of Restoring Mandibular Single Teeth with the Inclusive Tooth Replacement Solution Go online for in-depth content by Bradley C. Bockhorst, DMD While the prosthetic rehabilitation of full-arch cases provides

More information

Gas-actuated thermometer Model 73, stainless steel version

Gas-actuated thermometer Model 73, stainless steel version Mechanical temperature measurement Gas-actuated thermometer Model 73, stainless steel version WIKA data sheet TM 73.01 for further approvals see page 15 Applications General process instrumentation in

More information

Anchor HSA Fastening. Technology Manual. Hilti. Stud anchor. Carbon steel and stainless steel HSA HSA-BW HSA-R2 HSA-R. Hot dipped galvanized HSA-F

Anchor HSA Fastening. Technology Manual. Hilti. Stud anchor. Carbon steel and stainless steel HSA HSA-BW HSA-R2 HSA-R. Hot dipped galvanized HSA-F Anchor HSA Fastening Stud anchor Technology Manual Hilti Stud anchor Carbon steel and stainless steel HSA HSA-BW HSA-R2 HSA-R Hot dipped galvanized HSA-F M6 M20 Version 2012-11 11 / 2012 1 HSA Stud anchor

More information

Patient information on dental implants. More than just beautiful teeth. A new quality of life with dental implants.

Patient information on dental implants. More than just beautiful teeth. A new quality of life with dental implants. Patient information on dental implants More than just beautiful teeth. A new quality of life with dental implants. More than tooth replacement. Today's standard of care. Dental implant treatment can improve

More information

Switched Reluctance Machine (SRM)

Switched Reluctance Machine (SRM) Switched Reluctance Machine (SRM) 3D FEA Analysis using MagNet Jerry Foshage - RCT Systems 1 Wind and Water Power Technologies Office eere.energy.gov Acknowledgment & Disclaimer Acknowledgment This material

More information

Dynamic Loading Fluid Leakage Characterization of Dental Implant Systems

Dynamic Loading Fluid Leakage Characterization of Dental Implant Systems Dynamic Loading Fluid Leakage Characterization of Dental Systems Published by Dynamic Loading Fluid Leakage Characterization of Dental Systems Introduction The integrity of the dental implant-abutment

More information

What is it? How long does it take?

What is it? How long does it take? Dental Implants at What is it? An implant is a titanium post that is placed into and biologically fuses with the jawbone to replace the root or roots of 1 or more missing teeth. This is as close as you

More information

Measuring Luxation of Dental Implants In Vitro

Measuring Luxation of Dental Implants In Vitro Measuring Luxation of Dental Implants In Vitro Horea T. Ilieş ilies@engr.uconn.edu Dennis Flanagan DDS dffdds@charter.net Matthew Raby and Richard Stevenson Abstract Loading of dental implants immediately

More information

What Dental Implants Can Do For You!

What Dental Implants Can Do For You! What Dental Implants Can Do For You! Putting Smiles into Motion About Implants 01. What if a Tooth is Lost and the Area is Left Untreated? 02. Do You Want to Restore Confidence in Your Appearance? 03.

More information

HYDAC Compact Power Units CA

HYDAC Compact Power Units CA 1007 HYDAC Compact Power Units CA Up to 250 bar Up to 12.6 l/min Suitable for: Short-term operation S2 Intermittent operation S3 1. TECHNICAL SPECIFICATIONS 1.1 GENERAL For short-term operation S2 or intermittent

More information

VdS 2100-09en. VdS Guidelines for water extinguishing systems. Non-return valves. Requirements and test methods. VdS 2100-09en : 2011-05 (01)

VdS 2100-09en. VdS Guidelines for water extinguishing systems. Non-return valves. Requirements and test methods. VdS 2100-09en : 2011-05 (01) VdS Guidelines for water extinguishing systems VdS 2100-09en Requirements and test methods VdS 2100-09en : 2011-05 (01) Publishing house: VdS Schadenverhütung GmbH Amsterdamer Str. 172-174 50735 Köln,

More information

Prosthetic Manual. Implant System. Contact Us: Smarter thinking. Simpler design. Phone: 866-902-9272 781-328-3490. Fax: 866-903-9272 781-328-3400

Prosthetic Manual. Implant System. Contact Us: Smarter thinking. Simpler design. Phone: 866-902-9272 781-328-3490. Fax: 866-903-9272 781-328-3400 Contact Us: Implant System Phone: 866-902-9272 781-328-3490 Fax: 866-903-9272 781-328-3400 Prosthetic Manual Mail: Keystone Dental, Inc. 144 Middlesex Turnpike Burlington, MA 01803 USA www.keystonedental.com

More information

Finite Element Analysis of a Total Knee Replacement

Finite Element Analysis of a Total Knee Replacement Finite Element Analysis of a Total Knee Replacement Simone Machan Supervisor: Ass. Prof. Dennis Bobyn Company: Australian Surgical Design and Manufacture Research Goals: Static Analysis STANCE Aim: To

More information

Medium External Fixator Tibial Shaft Box Frame. For small-statured adults and pediatric patients, using outrigger posts.

Medium External Fixator Tibial Shaft Box Frame. For small-statured adults and pediatric patients, using outrigger posts. Medium External Fixator Tibial Shaft Box Frame. For small-statured adults and pediatric patients, using outrigger posts. Technique Guide Part of the Medium External Fixation System MRI Information Synthes

More information