SAKASHITA UNDERGROUND COSMIC RAY TELESCOPE
|
|
- Corey Reed
- 7 years ago
- Views:
Transcription
1 SAKASHITA UNDERGROUND COSMIC RAY TELESCOPE Cosmic-Ray Research Section, Solar-Terrestrial Environment Laboratory, Nagoya University Nagoya JAPAN Sakashita Tunnel 1
2 I. INTRODUCTION Sakashita Underground Multi-directional Telescope was planned as one of projects of Japan program for International Magnetospheric Study (IMS, ) to study solar modulation of cosmic rays in high energy region. For the Telescope Sakashita Underground Cosmic Ray Observatory, Nagoya University was constructed in 1976 in a disused railway tunnel of Japan National Railways (now Central Japan Railway Co.) at Sakashita-cho, Gifu prefecture, about 100 km north-east of Nagoya. The Underground Telescope was built, extending detectors with three steps in 1977 to 1979 to 4 x 15 m 2 upper and 3 x 14 m 2 lower finally, and full observations of 10 directional cosmic ray intensities started in Dec Fig.1 Map near Sakashita Observatory II. SAKASHITA UNDERGROUND OBSERVATORY The observatory was constructed in the middle of a tunnel at Sakashita (35 35 N and E in geographic coorditnates, 334 m above sea level), which runs 14 westward from the north as shown in Fig. 1. Topography above the observatory is fairly flat as can be seen from the vertical cross sections of the soil in Fig. 2. The vertical depth is estimated at 80 mwe from the topography and also from a comparison between cosmic ray fluxes inside and outside the tunnel (Ueno et al., 1976; Fujii and Jacklyn, 1979). The detector room is 3.6 x 33 m 2 in floor area and has a cross-section of horseshoe shape as shown in Fig. 3. Inside the room, there constructed two layers, upper and lower, with iron-frame and plate to lay out detectors. Thin plastic plate was built along with the coved ceiling of tunnel to protect detector room from water drop out of the brick wall. The humidity is kept 70% by seven 300 W dehumidifiers. The seasonal change of the room temperature is about 1 C from the average 22 C and its daily variation is about 0.2 C. Fig.2 Vertical cross section of the ground 2
3 III. MULTI-DIRECTIONAL UNDERGROUND TELESCOPE a. Detectors and their arrangements The multi-directional telescope consists of two layers of detectors as shown in Fig. 3. Each of detectors is composed of plastic scintillators of 1 x 1 m 2 in area and 10 cm in thickness, which is set at the bottom of a pyramid shaped box of 1.6 mm iron plate and is viewed by a 5" photomultiplier (Hamamatsu Photonics R877). These detectors are set respectively on the upper and lower layers separated by 1.75 m in height without absorber. Configuration of the detector layout was designed in considering the maximum availability of space in the tunnel. Centre of each detector on the lower layer is placed just right under one of the corners of each detector on the upper layer. Each detector is called by the number with initial U or L for the upper or lower layer, as U1, 02,, U60 and L1. L2,, L42. b. Detectors and the adjustments Pulses from each photomultiplier are amplified by about 500 times, and then discriminated from noises at a fixed voltage of 0.5 V by an amplifier/comparator installed in each detector box. These signal pulses are sent to a coincidence circuit board by a coaxial cable (25 m), and reshaped to uniform pulses (5 volts and 200 nsec) to trigger the coincidence circuits. Fig. 4 shows one of plateau curves of detectors, i.e., the dependence of counting rate of the detector on high voltage supplied to the photomultiplier. A plateau is seen at around 850 to 900 volts with a counting rate of about 350 counts per minute, indicating a clear separation of cosmic ray signals from background noises. Based on this character, high voltage of each photomultiplier is adjusted so that the counting rate of every detector is nearly equal to 400 counts per minute. These adjustments are made by inserting a proper resister in series between the photomultiplier and the high voltage supply set at 950 volts. Fig.3 Setting of Telescope in the tunnel. Fig.4 Plateau curve (variation of counting rate with high volate) 3
4 c. Directional telescopes and their counting rates Directional telescopes are composed of selected 2-fold coincidences of signals from upper and lower layer detectors. The selection of the coincidences is illustrated in the top view of the telescope in Fig. 5, where the upper and lower layer detectors are shown by squares of thin solid and thin broken lines respectively. Coincidences for the directional telescopes are indicated by the relative position of the upper layer detectors (thick solid lines) to one representative lower detector or detectors. Central viewing directions and average counting rates of these directional telescopes are summarized in Table I. Two directional telescopes V1 and V2 in the Table are a sum of (NE, SE, SW and WN) and that of (N, E, S and W) respectively. Effective depth and effective median primary rigidity were calculated by Fujimoto et al. (1977) from the yield function of cosmic rays (Murakami et al., 1979) in considering the geometrical configuration of detectors and particle s path length in the rock with a mean density of 2.5 g/cm 3. Fig.5 Configuration of upper and lower layer detectors showing two-fold coincidences for directional telescopes. Table 1. Viewing directions, effective depths, median rigidities and counting rates for multidirectional telescopes at Sakashita 4
5 d. Recording system The output pulses from the coincidence system are fed into 24 channel recorder. Each channel of the recorder consists of 12 bits binary counter with 12-bit gate and 4 digits decimal counter with 4 digits buffer memory. Each binary counter is designed as a sacling factor of 1 to 4096 for decimal counter and is set so that hourly count of the decimal counter becomes about 5000/hour. The accumulated counts in the decimal counter are transferred to the buffer memory once every hour and then the counter is cleared up for the next hour. The contents of the buffer memories as well as the time code are punched every hour on a paper tape. The atmospheric pressure and the room temperature are measured by a digital barograph and thermometers with V-F converters, and are recorded also on paper tape by the recorder. e. Power supplies In order to avoid interruption of observations due to AC power line failure, an AC-DC-AC converter system with a deck of floating batteries (96V, 35AH) is used for the AC power supply. This system can supply 500 W power to the telescope for 3 hours without AC power input. High voltage supply and DC power supplies are highly stabilized against changes of AC voltage and room temperature. The conventional stabilized power supplies are used for the coincidence system and the recording system, as they are not sensitive to the change of the supply voltage,. f. Stability of the telescope All the instrumental gain drift due to the variations of the power supply voltage and room temperature are reflected in the counting rate through slopes of the plateau curve. Fig. 6 shows the plateau curves for five representative coincidence components together with those for total single components of upper and lower layers. Slopes of the plateaus for coincidence components are less than 0.2 % counts per 1 volt at 950 volts. The change of high voltage due to that of AC power line voltage has negligible effect to the counting rate. The main instrumental drift is due to the Fig.6 Plateau curves of various telescopes variation of room temperature. The room temperature causes gain changes of the photomultiplier and amplifier-discriminator system of -0.1 % Gain/ C. Change of 0.1% gain is equivalent to change of -0.1 V of high voltage, and the room temperature effect of the Telescope is estimated to be % counts/ C. As the temperature variation is kept within 0.2 C on daily basis and 2 C on yearly basis, the instrumental drift of the coincidence rate is less than % in a day and % in a year. The other instrumental drift is due to the fatigue of the photomultiplier. The local rainfall has also effect on counting rate by the change of effective ground mass. As no correction for these effects was made, the long term drift of about 1 %/year were observed in the counting rate. g. Automatic check systems To check the malfunction of the Telescope and the long term gain change due to fatigue of photomultiplier, the automatic check system by a PC records daily means of single and coincidence counting rates from each 5
6 detector. This system measure also differential pulse height distribution of each detector by a built-in 256 channel pulse height analyzer, automatically scanning all detectors once every day. Some of the data are checked once every day in Nagoya through a telephone line to the observatory. This automatic check system has been operated since the middle of IV. REDUCTION OF DATA The paper tape is collected once a month and the hourly outputs on paper tape are processed by a computer system in the Institute, to derive the relative cosmic ray intensity variation. Hourly counts N s in the paper tape are converted into the barometer corrected relative intensity (Wp) using the Natural Logarithmic Representation (Wada, 1957) as follows, W P = 10 4 x (In N - In N 0 + WL) - β(p-p 0 ) in unit of 0.01 % where N 0 is average counting rate divided by the scaling factor and WL is artificially added to make the values of Wp around (%). P is the atmospheric pressure, P 0 is 980 hpa and β is the barometric pressure effect coefficient (-0.03 %/hpa for all components). Hourly values of Wp, daily sum and average as well as the 1st, 2nd and 3rd harmonics coefficients of the daily variation of each component are computed and tabulated on daily basis. In the same table, the residual dispersion of the cosmic ray intensities for one day after removing the daily variation (up to 3rd harmonics) is tabulated. Further, these hourly values of Wp are used to produce hourly value plot for every solar rotation period using the X-Y plotter attached to the computer system. Check of the processed data is made, utilizing the hourly data table and inspecting visually the plotting diagrams. The final data are written into a magnetic tape and used as the input to the data processing for final tabulation and plotting for the publication at NAGOYA University Computation Center. REFERENCES Fuji, Z., and R. M. Jacklyn Proc. 16th Coemic Ray Conf. Kyoto (1979) Fujimto, K., K. Murakami, I. Kondo, and K. Nagashitna Proc. 15th Cosmic kay Conf. Plovdiv (1977) Murakami, K., K. Nagasbima, S. Sagisaka, Y. Mishima, and A. Inoue Nuovo Cimento 1-2C 635 (1979) Sagisaka, S., K. Murakami. A. Inoue, Y. Mishima, and K. Nagashima Proc. 16th Int. Cosmic Ray Conf., Kyoto (1979) Ueno, H., K. Fujinroto, Z. Fujii, I. Kondo, and K. Nagashima Proc. Cosmic Ray Symp. of High Energy C.R. Modulation 91, (1976) Wada M. J. Sci. Res. Inst. 51, 201 (1957) 6
7 Appendix 1 EPILOGUE Cosmic-Ray Research Laboratory, Nagoya University found in 1958 as a research institute with cosmic ray telescopes, was reorganized with Research Institute of Atmospherics, Nagoya University to Solar-Terrestrial Environment Laboratory (STEL) in June, Sakashita multi-directional underground telescope continued observations by cosmic ray research section of STEL until closing of observations of about 23 years in March, Appendix 2 FOR USE OF DATA Hourly value data of SAKASHITA Multi-directional underground telescope are available from Web Site of Solar-Terrestrial Environment Laboratory ( via network. One record of the hourly value files (92 bytes) consist of 24 data of 4 digits each, WP in 0.01% and Pressure in 0.1 hpa, shown below SSS none Pressure Z (=N-S) Digit content 1-4 Last digit of year + Day of Year 5 8 Hour and minute (start time of record) 9-12 Upper Single + Lower Single none (for monitoring Telescope) V V NE ES SW WN N E S W NN SS none none NNN 7
ISTITUTO NAZIONALE DI FISICA NUCLEARE
ISTITUTO NAZIONALE DI FISICA NUCLEARE Sezione di Catania INFN/TC-11/02 February 28, 2011 PRELIMINARY TESTS OF A SCINTILLATOR-BASED MINI-STATION FOR EXTENSIVE AIR SHOWERS MEASUREMENTS S.Aiola 1, P. La Rocca
More informationSound absorption and acoustic surface impedance
Sound absorption and acoustic surface impedance CHRISTER HEED SD2165 Stockholm October 2008 Marcus Wallenberg Laboratoriet för Ljud- och Vibrationsforskning Sound absorption and acoustic surface impedance
More informationThe data acquisition system of the XMASS experiment
The data acquisition system of the XMASS experiment Katsuki Hiraide 1,2, for the XMASS Collaboration 1 Kamioka Observatory, Institute for Cosmic Ray Research, the University of Tokyo, Higashi-Mozumi, Kamioka,
More information(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier
(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (no PiN and pinned Diodes) Peter Fischer P. Fischer, ziti, Uni Heidelberg, Seite 1 Overview Reminder: Classical Photomultiplier
More informationABB i-bus EIB / KNX Analogue Input AE/S 4.2
Product Manual ABB i-bus EIB / KNX Analogue Input AE/S 4.2 Intelligent Installation Systems This manual describes the functionality of Analogue Input AE/S 4.2. Subject to changes and errors excepted. Exclusion
More informationAccurate Measurement of the Mains Electricity Frequency
Accurate Measurement of the Mains Electricity Frequency Dogan Ibrahim Near East University, Faculty of Engineering, Lefkosa, TRNC dogan@neu.edu.tr Abstract The frequency of the mains electricity supply
More informationDDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS
DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display
More informationQFP910. Water detector. Building Technologies. CE1N2732en. Synco living
2 732 Synco living Water detector QFP910 Wireless radio sensor to record water puddles RF communication based on KNX standard (868 MHz unidirectional) Battery operation with standard 1.5 V batteries Use
More informationCopyright 2011 Casa Software Ltd. www.casaxps.com. Centre of Mass
Centre of Mass A central theme in mathematical modelling is that of reducing complex problems to simpler, and hopefully, equivalent problems for which mathematical analysis is possible. The concept of
More informationApplication Note. So You Need to Measure Some Inductors?
So You Need to Measure Some nductors? Take a look at the 1910 nductance Analyzer. Although specifically designed for production testing of inductors and coils, in addition to measuring inductance (L),
More informationManual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
More informationGAMMA-RAY SPECTRA REFERENCES
GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.
More informationAdvanced Natural Gas Storage (ANGAS) Project and Verification Tests of Experimental Lined Rock Cavern in Japan
Advanced Natural Gas Storage (ANGAS) Project and Verification Tests of Experimental Lined Rock Cavern in Japan Toru Komatsubara THE JAPAN GAS ASSOCIATION Taku Watanabe TOKYO GAS CO.,LTD. Satoshi Murase
More informationSelecting IHLP Composite Inductors for Non-Isolated Converters Utilizing Vishay s Application Sheet
VISHAY DALE www.vishay.com Magnetics Selecting IHLP Composite Inductors for Non-Isolated Converters INTRODUCTION This application note will provide information to assist in the specification of IHLP composite
More informationMeasurement of Muon Lifetime and Mass Using Cosmic Ray Showers
Measurement of Muon Lifetime and Mass Using Cosmic Ray Showers Angela Hansen Physics 4052 School of Physics and Astronomy, University of Minnesota May 4, 2001 Abstract In this experiment, we used a scintillation
More informationSensor Network for HACCP Food Safety Management
Hitachi Review Vol. 59 (2010), No. 5 229 Sensor Network for HACCP Food Safety Management Hiroshi Hanyu Takanori Shimura Takuya Fukui OVERVIEW: An increasing number of companies are incorporating HACCP
More informationG. Karasinski, T. Stacewicz, S.Chudzynski, W. Skubiszak, S. Malinowski 1, A. Jagodnicka Institute of Experimental Physics, Warsaw University, Poland
P1.7 INVESTIGATION OF ATMOSPHERIC AEROSOL WITH MULTIWAVELENGTH LIDAR G. Karasinski, T. Stacewicz, S.Chudzynski, W. Skubiszak, S. Malinowski 1, A. Jagodnicka Institute of Experimental Physics, Warsaw University,
More informationCoating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
More informationCHAPTER 6 PHOTON COUNTING
CHAPTER 6 PHOTON COUNTING 1) 2) 4) - 12) Photon counting is an effective technique used to detect very-lowlevel-light such as Raman spectroscopy, fluorescence analysis, and chemical or biological luminescence
More informationEdmund Li. Where is defined as the mutual inductance between and and has the SI units of Henries (H).
INDUCTANCE MUTUAL INDUCTANCE If we consider two neighbouring closed loops and with bounding surfaces respectively then a current through will create a magnetic field which will link with as the flux passes
More informationHarmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies
Soonwook Hong, Ph. D. Michael Zuercher Martinson Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies 1. Introduction PV inverters use semiconductor devices to transform the
More informationLab 9: The Acousto-Optic Effect
Lab 9: The Acousto-Optic Effect Incoming Laser Beam Travelling Acoustic Wave (longitudinal wave) O A 1st order diffracted laser beam A 1 Introduction qb d O 2qb rarefractions compressions Refer to Appendix
More informationELECTRON SPIN RESONANCE Last Revised: July 2007
QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron
More informationDENSITY MEASURING SYSTEMS
SHIBAYAMA SCIENTIFIC CO., LTD. DENSITY MEASURING SYSTEMS Density Gradient Tube Method Direct Reading Type A / Type B Comply with the standards below: JIS K-0061-1992, K-7112 1980 ASTM D1505 Type A /Left
More informationUltrasound Distance Measurement
Final Project Report E3390 Electronic Circuits Design Lab Ultrasound Distance Measurement Yiting Feng Izel Niyage Asif Quyyum Submitted in partial fulfillment of the requirements for the Bachelor of Science
More information2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
More informationHow to measure absolute pressure using piezoresistive sensing elements
In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article
More informationChapter 6: From Digital-to-Analog and Back Again
Chapter 6: From Digital-to-Analog and Back Again Overview Often the information you want to capture in an experiment originates in the laboratory as an analog voltage or a current. Sometimes you want to
More informationA System for Capturing High Resolution Images
A System for Capturing High Resolution Images G.Voyatzis, G.Angelopoulos, A.Bors and I.Pitas Department of Informatics University of Thessaloniki BOX 451, 54006 Thessaloniki GREECE e-mail: pitas@zeus.csd.auth.gr
More informationMETU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING
METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING Met E 206 MATERIALS LABORATORY EXPERIMENT 1 Prof. Dr. Rıza GÜRBÜZ Res. Assist. Gül ÇEVİK (Room: B-306) INTRODUCTION TENSION TEST Mechanical testing
More informationInformation about the T9 beam line and experimental facilities
Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions
More informationSmart Electromagnetic Flowmeter Open channel Flowmeter Detector
Magne3000 PLUS Smart Electromagnetic Flowmeter Open channel Flowmeter Detector Model NNK150/951 OVERVIE The Magne3000 PLUS Electromagnetic Flowmeter is submersible type of flowmeter mainly used for flow
More informationSpeed Control Methods of Various Types of Speed Control Motors. Kazuya SHIRAHATA
Speed Control Methods of Various Types of Speed Control Motors Kazuya SHIRAHATA Oriental Motor Co., Ltd. offers a wide variety of speed control motors. Our speed control motor packages include the motor,
More informationA METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS Joseph J. Stupak Jr, Oersted Technology Tualatin, Oregon (reprinted from IMCSD 24th Annual Proceedings 1995) ABSTRACT The
More informationA PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION
35'th Annual Precise Time and Time Interval (PTTI) Systems and Applications Meeting San Diego, December 2-4, 2003 A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION Józef Kalisz and Ryszard Szplet
More informationEFC 3600. Frequency converters
2 Bosch Rexroth AG Electric Drives and Controls Documentation Compact and complete: space saving side-by-side assembly, plug-in I/O terminals, with brake chopper and mains filter for ultra-simple installation
More informationWind Sensor W-RS485 with RS485 Interface
Wind Sensor W-RS485 with RS485 Interface Technical data and notes for installation Elsner Elektronik GmbH Steuerungs- und Automatisierungstechnik Herdweg 7 D-75391 Gechingen Germany Phone: +49 (0) 70 56/93
More informationThe accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
More informationLAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS
LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS Objective In this experiment you will study the i-v characteristics of an MOS transistor. You will use the MOSFET as a variable resistor and as a switch. BACKGROUND
More informationBasic RTD Measurements. Basics of Resistance Temperature Detectors
Basic RTD Measurements Basics of Resistance Temperature Detectors Platinum RTD resistances range from about 10 O for a birdcage configuration to 10k O for a film type, but the most common is 100 O at 0
More informationQFP910. Water detector. Building Technologies. CE1N2732en 2016-03-01. Synco living
2 732 Synco living Water detector QFP910 Wireless radio sensor to record water puddles RF communication based on KNX standard (868 MHz unidirectional) Battery operation with standard 1.5 V batteries Use
More informationDecimal Number (base 10) Binary Number (base 2)
LECTURE 5. BINARY COUNTER Before starting with counters there is some vital information that needs to be understood. The most important is the fact that since the outputs of a digital chip can only be
More informationEvaluation Tools for the Performance of a NESTOR Test Detector
Evaluation Tools for the Performance of a NESTOR Test Detector G. Bourlis, A. Leisos, S. E. Tzamarias and A. Tsirigotis Particle and Astroparticle Physics Group School of Science and Technology Hellenic
More informationTOTALLY SOLID STATE NON-DIRECTIONAL RADIO BEACONS 190-535 khz
TOTALLY SOLID STATE NON-DIRECTIONAL RADIO S 190-535 khz This family of radio transmitters has been developed as extremely efficient, highly reliable Non Directional Beacons. ND2000A/4000A» MODULAR CONSTRUCTION»
More informationLEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS
LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS March 2010 Lattice Semiconductor 5555 Northeast Moore Ct. Hillsboro, Oregon 97124 USA Telephone: (503) 268-8000 www.latticesemi.com
More informationManual Analysis Software AFD 1201
AFD 1200 - AcoustiTube Manual Analysis Software AFD 1201 Measurement of Transmission loss acc. to Song and Bolton 1 Table of Contents Introduction - Analysis Software AFD 1201... 3 AFD 1200 - AcoustiTube
More informationCHAPTER 2 HYDRAULICS OF SEWERS
CHAPTER 2 HYDRAULICS OF SEWERS SANITARY SEWERS The hydraulic design procedure for sewers requires: 1. Determination of Sewer System Type 2. Determination of Design Flow 3. Selection of Pipe Size 4. Determination
More informationE190Q Lecture 5 Autonomous Robot Navigation
E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator
More informationE. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE
E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter
More informationLayers of the Earth s Interior
Layers of the Earth s Interior 1 Focus Question How is the Earth like an ogre? 2 Objectives Explain how geologists have learned about the interior of the Earth. Describe the layers of the Earth s interior.
More informationCC-Link Ver.1.10 Compatible
Model Title CC-Link Ver.1.10 Compatible CC-Link Flexible Cable Conformance Test Specifications Management number : BAP-C0401-009 CC-Link Partner Association (1/12) Revisions Revision item Revision Description
More information12 Appendix 12 Earth Electrodes And Earth Electrode
12 Appendix 12 Earth Electrodes And Earth Electrode Testing 12.1 Introduction This appendix provides guidance and background information an earth electrode testing and some limited information on earth
More information4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:
4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing
More informationObjectives 200 CHAPTER 4 RESISTANCE
Objectives Explain the differences among conductors, insulators, and semiconductors. Define electrical resistance. Solve problems using resistance, voltage, and current. Describe a material that obeys
More informationINSTALLING A/C ADAPTER
WS 6502 INTRODUCTION 1. SNOOZE/LIGHT BUTTON 2. BASE 3. ALM SET BUTTON 4. + / C/F BUTTON 5. HISTORY BUTTON 6. MODE BUTTON 7. CHANNEL BUTTON 8. - / RCC BUTTON 9. MAX/MIN BUTTON 10. BATTERY COMPARTMENT 11.
More informationNUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706
(revised 4/21/03) NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 Abstract This experiment studies the Nuclear Magnetic Resonance of protons
More informationMounting Instructions for SP4 Power Modules
Mounting Instructions for SP4 Power Modules Pierre-Laurent Doumergue R&D Engineer Microsemi Power Module Products 26 rue de Campilleau 33 520 Bruges, France Introduction: This application note gives the
More informationDigital to Analog Converter. Raghu Tumati
Digital to Analog Converter Raghu Tumati May 11, 2006 Contents 1) Introduction............................... 3 2) DAC types................................... 4 3) DAC Presented.............................
More informationA Multichannel Storage Capacitor System for Processing of Fast Simultaneous Signals. Atsushi Ogata. IPPJ-T-16 Oct. 1973
A Multichannel Storage Capacitor System for Processing of Fast Simultaneous Signals Atsushi Ogata IPPJ-T-16 Oct. 1973 A Multichannel Storage Capacitor System for Processing of Fast Simultaneous Signals
More informationINTERNATIONAL ASSOCIATION OF CLASSIFICATION SOCIETIES. Interpretations of the FTP
INTERNATIONAL ASSOCIATION OF CLASSIFICATION SOCIETIES Interpretations of the FTP CONTENTS FTP1 Adhesives used in A or B class divisions (FTP Code 3.1, Res A.754 para. 3.2.3) June 2000 FTP2 Pipe and duct
More informationAddress for Correspondence
International Journal of Advanced Engineering Technology E-ISSN 0976-3945 Research Paper DEVELOPMENT OF LOW COST SHAKE TABLES AND INSTRUMENTATION SETUP FOR EARTHQUAKE ENGINEERING LABORATORY C. S. Sanghvi
More informationTechnical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity
More informationConstructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.
Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist. Abstract. I have been asked to put together a detailed article on a SWR meter. In this article I will deal
More informationNumerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope Rakesh Sidharthan 1 Gnanavel B K 2 Assistant professor Mechanical, Department Professor, Mechanical Department, Gojan engineering college,
More informationMars Atmosphere and Volatile EvolutioN (MAVEN) Mission
Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission MAVEN Science Community Workshop December 2, 2012 Particles and Fields Package Solar Energetic Particle Instrument (SEP) Davin Larson and the SEP
More informationTransistor Characteristics and Single Transistor Amplifier Sept. 8, 1997
Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997 1 Purpose To measure and understand the common emitter transistor characteristic curves. To use the base current gain
More informationJNIOR. Overview. Get Connected. Get Results. JNIOR Model 310. JNIOR Model 312. JNIOR Model 314. JNIOR Model 410
The INTEG is an Ethernet I/O (digital, analog) device that monitors and controls a small set of process signals. functions as both basic I/O for integration with another application or system AND as a
More informationApplication Report: Running µshape TM on a VF-20 Interferometer
: Running µshape TM on a VF-20 Interferometer General This report describes how a fiber interferometer from Arden Photonics Ltd was used together with the µshape TM Generic software package. The VF-20
More informationBulk Flow Meter LB 442 P R O C E S S C O N T R O L
Bulk Flow Meter LB 442 P R O C E S S C O N T R O L Mass flow determination using the LB 442 Versatile uses Contactless mass flow determination of solid matter on continuous conveyor systems Conveyor belts
More informationProcon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS
Procon Engineering Technical Document PELR 1002 TERMS and DEFINITIONS The following terms are widely used in the weighing industry. Informal comment on terms is in italics and is not part of the formal
More informationMATHEMATICS FOR ENGINEERING BASIC ALGEBRA
MATHEMATICS FOR ENGINEERING BASIC ALGEBRA TUTORIAL 4 AREAS AND VOLUMES This is the one of a series of basic tutorials in mathematics aimed at beginners or anyone wanting to refresh themselves on fundamentals.
More informationOptimizing VCO PLL Evaluations & PLL Synthesizer Designs
Optimizing VCO PLL Evaluations & PLL Synthesizer Designs Today s mobile communications systems demand higher communication quality, higher data rates, higher operation, and more channels per unit bandwidth.
More informationDDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS
DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display
More informationSingle Channel Loop Detector
Single Channel Loop Detector Model - LD100 Series The LD100 is a single channel inductive loop detector designed for parking and access control applications. The detector is connected to an inductive loop
More informationUsing Thermocouple Sensors Connecting Grounded and Floating Thermocouples
Connecting Grounded and Floating Thermocouples For best performance, Thermocouple sensors should be floating. This will ensure that no noise currents can flow in the sensor leads and that no common-mode
More informationAmplifier for Small Magnetic and Electric Wideband Receiving Antennas (model AAA-1B)
Amplifier for Small Magnetic and Electric Wideband Receiving Antennas (model AAA-1B) 1. Description and Specifications Contents 1.1 Description 1.2 1.2 Specifications 1.3 1.3 Tested parameters in production
More informationThe quadrature signals and the index pulse are accessed through five 0.025 inch square pins located on 0.1 inch centers.
Quick Assembly Two and Three Channel Optical Encoders Technical Data HEDM-550x/560x HEDS-550x/554x HEDS-560x/564x Features Two Channel Quadrature Output with Optional Index Pulse Quick and Easy Assembly
More informationEncoders for Linear Motors in the Electronics Industry
Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy
More informationLEAD CRYSTAL. User Manual. Valve-regulated lead-crystal batteries Energy storage Cells
Engineering Production Sales LEAD CRYSTAL Valve-regulated lead-crystal batteries Energy storage Cells User Manual www.axcom-battery-technology.de info@.axcom-battery-technology.de Chapter 1: 1. Introduction
More informationImpedance 50 (75 connectors via adapters)
VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:
More informationSIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION
1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.
More informationELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.
More informationFundamentals of Mass Flow Control
Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of
More informationMaking Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz
Author: Don LaFontaine Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Abstract Making accurate voltage and current noise measurements on op amps in
More informationCalibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement
Chapter 2 Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement 2.1 Introduction and Objectives This laboratory exercise involves the static calibration
More informationNuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics
More informationExperiment 10. Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado
1 Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Introduction Some radioactive isotopes formed billions of years ago have half- lives so long that they are
More informationTHE CMS PIXEL DETECTOR: FROM PRODUCTION TO COMMISSIONING
THE CMS PIXEL DETECTOR: FROM PRODUCTION TO COMMISSIONING VINCENZO CHIOCHIA On behalf of the CMS Collaboration University of Zürich, Physik-Institut, Winterthurerstr. 190 Zürich, CH-8057, Switzerland The
More information104 Practice Exam 2-3/21/02
104 Practice Exam 2-3/21/02 1. Two electrons are located in a region of space where the magnetic field is zero. Electron A is at rest; and electron B is moving westward with a constant velocity. A non-zero
More informationModifying the Yaesu FT-847 External 22.625 MHz Reference Input
Modifying the Yaesu FT-847 External 22.625 MHz Reference Input David Smith VK3HZ Introduction This document describes the modification of an FT-847 to allow an external 22.625 MHz Reference oscillator
More informationAC3200 INTELLIGENT BROADBAND AMPLIFIER
Kari Mäki 9.4.2013 1(7) AC3200 INTELLIGENT BROADBAND AMPLIFIER AC3200 is the latest leading-edge addition to AC family with extended frequency and gain ranges and integrated electrical controls in both
More information1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.
CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change
More informationCalibration and performance test of the Very Front End electronics for the CMS electromagnetic calorimeter
Calibration and performance test of the Very Front End electronics for the CMS electromagnetic calorimeter Jan Blaha IPN Lyon 10th Topical Seminar on Innovative Particle and Radiation Detectors (IPRD06)
More informationMichelin North America
www.centecinc.com SC Telephone: 864.527.7750 Outside SC: 800.227.0855 Michelin North America Industrial Maintenance Technical Interview Outline Industrial Maintenance Technical Interview Outline The Technical
More informationA novel sharp beam-forming flat panel loudspeaker using digitally driven speaker system
Audio Engineering Society Convention Paper Presented at the 131st Convention 2011 October 20 23 New York, NY, USA This Convention paper was selected based on a submitted abstract and 750-word precis that
More informationwww.klmtechgroup.com TABLE OF CONTENT
Page : 1 of 24 Project Engineering Standard www.klmtechgroup.com KLM Technology #03-12 Block Aronia, Jalan Sri Perkasa 2 Taman Tampoi Utama 81200 Johor Bahru Malaysia S TABLE OF CONTENT SCOPE 2 DEFINITIONS
More informationS P E C I F I C A T I O N F O R L I T H I U M B A T T E R Y. M o d e l : C R 2 0 3 2
S P E C I F I C A T I O N F O R L I T H I U M B A T T E R Y M o d e l : C R 2 0 3 2 Approved By Department Name Title Signature/Date (Remarks: The above table shall be filled by customer) POWER GLORY BATTERY
More informationWireless Data Logger Monitoring Systems
Wireless Data Logger Monitoring Systems Table of Contents About MadgeTech, Inc. Wireless Systems & Components Wireless Data Loggers 2000A Wireless Data Logger Models MadgeTech Cloud Services MadgeTech
More informationIron Powder Cores for Switchmode Power Supply Inductors. by: Jim Cox
HOME APPLICATION NOTES Iron Powder Cores for Switchmode Power Supply Inductors by: Jim Cox Purpose: The purpose of this application note is to cover the properties of iron powder as a magnetic core material
More information