Independent End User Training. Curriculum and Supporting Materials. DHVI BD LSRII (L01/L02) and BD LSRFortessa (F01)
|
|
- Jocelin Tyler
- 7 years ago
- Views:
Transcription
1 Shared Resources: Flow Cytometry and Cell Sorting Independent End User Training Curriculum and Supporting Materials DHVI BD LSRII (L01/L02) and BD LSRFortessa (F01) Revised 1/2016, PMM, GDS
2 1.0 Purpose / Scope The purpose of this training packet is to provide materials to support and enhance your hands-on training experience. 2.0 Expectations Trainees must have been recommended by their manager/mentor to undergo training, have at least 10 instrument contact hours with an experienced flow cytometer user, have a need to use DHVI Flow Facility instruments at least 5 hours/month and have a strong desire to learn how to safely and correctly operate a flow cytometer to acquire comprehensive polychromatic flow data. Trainees must complete two BD Biosciences online courses, Introduction to Flow Cytometry and BD FACSDiva Software v8 Overview, PRIOR to initiating hands-on training. Trainees to will be expected to complete all three components of the DHVI Flow Facility Training program: Classroom group orientation (~2hr), Hands on session (~2 hr) and a Practical Exam (~2 hr). If the trainee misses one of the sessions or fails the practical exam, they will not be certified by the assigned training Operator and will have to repeat the training program at a later date (see attachment #2). The Pin Code to access Bay 5 will be given to the Trainee at the classroom session. If the Independent End User is the last person to leave the room they need to make sure that the door(s) are closed and locked. All approved Independent End Users of the DHVI Flow Facility instruments are expected to fully comply with the Policies and Procedures for DHVI Flow Facility Users detailed in SOP Flow_010 (on website) that is covered in the classroom session of the training. 3.0 Cytometer Overview Your trainer will orient you to the critical components of the cytometer: Fluidics, Optics, Computers, and the BDFACS Diva software (see attachments #3-5). 4.0 Operation and Maintenance 4.1 Instrument Startup Turn on Fluidics Cart (L02 only) Turn on Cytometer (needs 30 minute laser warm up) Be sure Sheath tank is full: Filling L02 Sheath Tank: Unscrew large cap and carefully pour sheath into the tank up to the top line on the tank Filling L01/F01 Sheath Tank: Depressurize the sheath tank by slowly pulling up on the pressure relief valve Remove the lid of the sheath tank, and fill up to the top weld line inside the tank. Revised 1/2016 Page 2 of 26
3 DO NOT OVER FILL Replace the lid making sure that you lock it down from side to side, not top to bottom Making up Sheath Fluid (PBS): x sheath containers are located under the sink between Bay 5 and Bay 6. Fill the sheath container with dh 2 O to the Fill Line on the back of the container Put a large X on the cap and the sides of the sheath container to indicate that it has been diluted. DO NOT X THE LABEL If you spill the sheath while pouring, please clean it up Additional sheath is located under the desk behind the L Empty Waste tank if full: Emptying Waste Tank: There are two (2) waste tanks used with the L01/L02/F01 because the waste needs to sit in the bleach solution for at least 24 hours before emptying down the sink L02: Bring the spare waste tank to the sink. Remove the large cap. Turn on the cold water and pour the 10% bleach waste down the drain letting the water run for a couple minutes when finished to flush the pipes Put 1 L of bleach and 2 L of water into the waste tank, so that when the tank is full it will be a 10% final bleach solution The bleach is located under the sink between Bay 5 and Bay Additional bleach is located under the desk behind the L Remove the RED filter waste cap from the full waste tank and put it on the empty waste tank going onto the fluidics cart. Unscrew the sensor and carefully pull it out of the tank and put it in the empty waste tank. Be careful not to break the filter on the tubing NOTE: The red filter waste cap ALWAYS goes on the waste tank on the fluidics cart!! Put the large white cap and the small cap on the full waste tank and put the empty after date on the tape. Put the waste tank in the bucket to the side of the fluidics cart. Revised 1/2016 Page 3 of 26
4 L01/F01: Disconnect the orange waste line and the black waste sensor line from the white cap on the full waste tank Bring the spare waste tank to the sink. Remove the large cap. Turn on the cold water and pour the 10% bleach waste down the drain letting the water run for a couple minutes when finished to flush the pipes Remove the waste tank from the holder. Put the empty waste tank in the holder Put the white waste cap with the sensor connections on it on the empty waste tank Reconnect the waste and sensor lines Put the large white cap on the full waste tank and put the empty after date on the tape Put the full waste tank in the bucket to the side under the L01/F01 table L02 Only. Be sure the Sheath Reservoir has ~1/4 of sheath in it. See Attachment #8 for troubleshooting the sheath reservoir Filling Sheath Reservoir: Put fluidics in Standby Unscrew the white cap and take sensor out of the tank Using the funnel (located at the sink), put ~1/4 of sheath fluid in the tank and replace the sensor Low Sheath/Sheath Alarm Put the cytometer in standby and refill the sheath tank After you refill the tank, you need to restart the fluidics cart NOTE: If you do not restart the cart, the sensor doesn t sense the sheath and will drain the reservoir causing what can be a lengthy troubleshooting procedure that you will be billed for at Operator rates Check Sheath Filter for air bubbles Hold the sheath filter so that the tubing connected to the stopcock/vent cap is facing you Tap the filter a couple of times to dislodge any bubbles that might be in there Remove air bubbles from sheath filter: L02: Using the stopcock device. Revised 1/2016 Page 4 of 26
5 Using a container to catch the sheath that comes out, roll back the stopcock to un-pinch the tubing Discard sheath into container until all bubbles are gone, and then roll back the stopcock to pinch the tubing to stop the flow of sheath L01/F01: Using the vent port on the top of the filter Using a container to catch the sheath that comes out, pinch off the tubing right below the vent cap, and remove the cap Discard sheath into container until all bubbles are gone, and then replace the vent cap so that it pinches off the tubing to stop the flow of sheath Priming Cytometer to remove Air Bubbles NOTE: This has to be done every time the cytometer is turned on because of the flow cell emptying and refilling with sheath Remove the tube of dh 2 O from SIP and press prime times or more if necessary to remove any air inside the flow cell Note: Never prime the cytometer with a tube on the SIP to prevent a potential clog going into the flow cell. 4.2 Instrument Shutdown If you are the last User of the day, you need to shut down the cytometer, fluidics cart (L02) and log off the computers Bleach: Run 5 minutes of 10% Bleach on HI dh 2 0: Run 5 minutes of dh 2 0 on HI. You need to run water at least as long as you run bleach, so that you get all the bleach out of the tubing to prevent corrosion. Revised 1/2016 Page 5 of 26
6 4.2.3 Standby: Set the fluidics control to Standby and Low and leave a tube of NO MORE THAN 1 ml of dh 2 0 on the SIP The tube of water is to keep the sheath from crystallizing in the injection probe You CANNOT have more than 1 ml in the tube because when the cytometer is turned off, or left in standby mode, there is a small amount of fluid that back drips into the tube. Too much fluid in the tube could over flow into the air line causing a LENGTHY service call to be placed Shut Off the Fluidics Cart (L02) and Cytometer as well as logging off of the computers NOTE: Please DO NOT shutdown the computers, we need access Top Off the Sheath tank (see above) Empty Waste if necessary (see above). 5.0 Cytometer Setup and Tracking (CS&T) CS&T needs to be run EVERY time the cytometer is turned on, to ensure consistent laser performance from day to day. Lasers are light bulbs, with use they can blow out at any time. 5.1 Log into the Cytometer Controller computer => launch the FACSDiva software => click OK in the Administrator window: L01: Username: userl01; Password: Msrb L02: Username: userl02; Password: Msrb F01: Username: userf01; Password: Msrb If you start the software before the cytometer, you will need to manually connect them by selecting the Instrument menu => Connect. 5.3 The instrument Status will probably say Fluidics Not Ready, just click clear. 5.4 Making CS&T Beads: Beads are located in the refrigerator behind the L01. Vortex the beads and place 3-4 drops of beads/500 ul of PBS. Please make sure that you are using the correct CS&T bead lot number which can be found in the CS&T software Label tube with: CS&T, lot # (found on CS&T vial), date and your three initials. 5.5 Running CS&T Beads: Select Cytometer in Workspace menu => CS&T. Revised 1/2016 Page 6 of 26
7 5.5.2 The cytometer is now disconnecting from the Diva software and connecting to the CS&T software Check the Load Tube Manually box (if not already checked) Make sure the CS&T Lot ID matches the CS&T Beads being used Put the Cytometer in Low and Run, Vortex and Load the CS&T beads Click Run in the CS&T software, click OK. The CS&T will take about 8 minutes to complete When completed, remove the CS&T beads and put back into the refrigerator If CS&T: Passed: Exit CS&T Program => after cytometer connects to the Diva software, proceed as usual Passed with Warnings (PWW): Same as step , but notify the Flow Facility about the warning Failed: Try making fresh beads, double check Lot ID and /or shutdown and restart the computer and cytometer and repeat process. If still failed, please contact the Flow Facility before continuing If no Flow Facility personnel are available: Exit the CS&T software and run the CS&T beads in the manual CS&T experiment located in the Diva Browser Browser => Manual CS&T folder => Double click CS&T experiment to open it => Expand the specimen => Copy and paste the prior tube and rename it with the correct date and lot number Run the CS&T beads to make sure that the dim and bright bead populations show up in the FCS/SSC plot (indicates that the blue laser and FCS/SSC PMTs are working) and that all three fluorescent peaks show up in their relative histograms (indicates all lasers/pmts are working). If all look good, record 1,000 events and continue on as usual with your experiment NOTE: It is important that you report any warnings/failures to the Flow Facility so that we can monitor the lasers to prevent a blowout that could shut down the cytometer for use for a lengthy period of time NOTE: To see if CS&T has been performed for that day: Click on the CS&T Reports folder on the desktop => Config folder => Year Folder => Month Folder => Day Folder =>.pdf of the CS&T Report. Revised 1/2016 Page 7 of 26
8 6.0 Data Collection NOTE: The Reset Target Values and Define Baseline procedures will always be done by the Flow Facility. If you get a notice saying that the Baseline has expired, please contact the Flow Facility. 6.1 Before Running Samples: Log onto the Workstation computer (posted instructions on Quick Fact sheet hanging on the wall next to the computer) Open Internet Explorer Open Fac Man => Log in using your DHE => Start your Fac Man session Open Biotrue Uploader and log in (posted instructions on Quick Fact sheet hanging on the wall next to the computer). Click continue through the security windows if applicable) Open the BDExort folder => FCS folder => Day s folder. 6.2 Creating a New Experiment in Diva Browser: Make a Yearly, Monthly and/or Daily folder if necessary and name accordingly in the Browser Frame => right click on Administrator => add a New Folder for the Year and name it accordingly => right click on the Yearly folder => add a folder for the Month and name it accordingly => right click on the Monthly folder => add a folder the Day and name it accordingly Right click the Daily Folder => New Experiment => 10 Color Template (L01/L02) or 8 Color Template (F01) => OK The Experiment should default open, but if it does not, double click on the Experiment Book to open it Right click on the Experiment and rename it with the date (yymmdd), run number, machine number, your initials and experiment number (e.g L02 PMM005) Right click on the Specimen and rename it with the date, your initials and experiment number (e.g PMM005) Expand the Specimen and right click on the Tube_001 and rename it with your initials_001 (e.g. PMM_001) Click the Acquisition Tube Pointer Box to the left of the tube in the Browser Frame => click the Parameters Tab in the Cytometer Frame => delete any unwanted parameters by holding down the control key => select the dot next to the parameter you want to delete => click on delete (this saves on memory) NOTE: We do suggest collecting one empty parameter for gating out autofluorescence if needed. Revised 1/2016 Page 8 of 26
9 6.3 Creating Global Worksheets: A new experiment (created from the icon and not the menu) will default with an empty Normal Worksheet. Normal Worksheets are tube specific, so we recommend using a Global Worksheet which is experiment specific To create a Global Worksheet, right click the Experiment book => New Global Worksheet Click on the worksheet to add plots Make the number of plots needed to satisfy your experimental design/panel, and change the labels accordingly You will need at least one dot plot of FSC/SSC, one dot plot with two parameters for compensation and a histogram for each parameter you are using (including the dump parameter if you are using one) to set your voltages Click on the Dot plot icon in the Worksheet toolbar, and then click on the worksheet Click on the parameter description to change the parameter labels. (i.e. FSC/SSC) NOTE: If you have FSC-H (Height) checked, the first dot plot will always default to FSC-A vs FCS-H Note: To make all your plots the same size, select all the plots making sure that the plot of the size you want is selected last and has the yellow sizing boxes on it. Go to the icon in the Worksheet toolbar that says make plots the same size (3 rd icon from the left). 6.4 Optimizing Instrument Settings: Voltage/Threshold/Compensation In the Browser Frame, highlight the Unstained tube (should be tube 1 on your protocol sheet) and position the Acquisition Tube Pointer in the box at the left of the tube Select Low and Run Put the Unstained tube on the SIP and click Acquire in the Acquisition Dashboard Frame In the Cytometer Frame, click on the Parameter Tab and adjust the FSC/SSC voltage to place your population of interest in view on the FSC/SSC dot plot If you are unsure what your population of interest is, you can back gate using one of your positive controls. Revised 1/2016 Page 9 of 26
10 6.4.6 Put on your positive control => draw a gate on the positive population => right click on the FSC/SSC plot => Show Population => select the gate around your positive population => look for the color of your gate in the FSC/SSC plot Adjust the FSC/SSC voltage to place your population of interest on scale So that you don t waste your positive control sample, when your population of interest is in view you can delete the gate and put your unstained tube back on to finish the adjustments Adjust the Threshold in the Cytometer Frame to exclude most of the debris (each tick mark is equal to 10,000 channels) If you are running Kappa beads, the threshold has to be at 5000 or you will not see the beads Draw a Gate around your population of interest (cells or beads) Select all the other plots by holding down the shift key => right click on one plot => Show Population => select the gate in the FCS/SSC plot Using the optimal voltages from the CS&T Report and your single color control tubes, adjust the Voltages of the Fluorescence channels in the Cytometer Frame => Parameter Tab so that you have an optimal voltage while making sure that your positive population is on scale For a weak signal (AB labeling), adjust the voltage to get the best signal to noise ratio For a strong signal (AB labeling), adjust the voltage so that the positive population is on scale After all the voltages have been set, select the Stopping Gate for your target population and the number of Events to Record in the Acquisition Dashboard Frame LEAVE THE STORAGE GATE AT ALL EVENTS!!! You can now setup your experimental panel and name the parameters accordingly using the Experiment Layout tool in the Workspace menu Experiment Menu => Experiment Layout => Label => label your parameters with the correct antibody name and fluorochrome you are using (e.g. CD3-FITC) => ok Put your Unstained tube on the SIP and click Record You can now put the fluidics in High if you choose After the sample is finished recording, remove it from the SIP. Revised 1/2016 Page 10 of 26
11 Expand the unstained tube in the Browser Frame => right click on its Cytometer Settings and Copy Paste the Cytometer Settings into the Global Cytometer Settings under the experiment book by right clicking on the Cytometer Settings => Paste Select any remaining tubes in your specimen => right click and paste the cytometer settings into them as well Click Next Tube to go to the next tube Put your first Compensation tube on the SIP and click Acquire and then Record When the sample if finished recording, remove it from the SIP Put the fluidics in Standby to stop the flow of sheath In your Fluorescent Parameter dot plot, change the parameter on the X axis to the positive control you are working with and the parameter on the Y axis to your next positive control (every time you change a parameter in the plot, you will need to redraw the gates) Click on the Rectangle Gate in the Worksheet Toolbar, and draw it around the Positive population. Make sure it encloses the whole population including those events below axis (use the biexponential view if needed) Draw another Rectangle Gate around the Negative population, making sure that you get those events below the x- and y-axis If you have not already created a Population Hierarchy View and a Statistics View box, do so now Right click on the plot => Show Population Hierarchy Right click on the plot => Create Statistics View The only information you need in the statistics view at this time, are the Populations you are matching up and the Parameters you are looking at with their Means (MFI) To Edit the information you do not need, right click on the Statistics View Box => Edit Statistics => Header Tab => deselect all the header information by unchecking the All box at the top of the column => Populations Tab deselect all the populations you are not looking at as well as the % and # columns by deselecting the box at the top of the column => Statistics Tab select all the fluorescent parameters you are using and their means => OK To compensate your control tubes, you need to match up the Means of the Positive population to that of the Negative Revised 1/2016 Page 11 of 26
12 population of all of the other parameters by adjusting the percent compensation values to add or subtract the Spectral Overlap NOTE: The positive population value should not be smaller than the negative value or be a negative number when compensation is applied because you will be over comping that population In the Browser, highlight the tube you want to compensate (it should turn Blue) => click the Cytometer Settings Tab in the Inspector Frame => click the Compensation Tab and make sure that the Enable Compensation box is checked Make sure that the tube you are trying to compensate is highlighted. If it is not, you will be adjusting compensation for the wrong tube Adjusting the compensation can be done a number of ways. You can use the Arrows to make Small Increments of.10, you can hold down the Control Key and use the Arrows for Medium Increments of 1 or you can Type in the Values for Large Increments Note: I DO NOT suggest using the Slide Ruler, you will be just chasing your tail To help you remember what fluorochrome you need to subtract/add compensation from, look at the top of the compensation table. It has Fluorochrome - % Fluorochrome The positive control will ALWAYS be in the - % Fluorochrome column Rule of Thumb: if unsure what numbers you need to match up in your statistics box, place your thumb over the positive control and match up all the other fluorochrome columns When you finish with the compensation of your first control tube, right click its Cytometer Settings and Copy Spectral Overlap Go up to the Global Cytometer Settings (under your experiment book) and right click and Paste Spectral Overlap With Zeros It is IMPORTANT that you paste with zeros because if you subtract a value to a zero, the software will keep the value Go to your next tube and expand it and Paste Spectral Overlap With Zeros into its cytometer settings Repeat steps thru for the remaining tubes When you are finished with the last compensation tube, Paste the Spectral Overlap With Zeros into the Global Cytometer Settings Revised 1/2016 Page 12 of 26
13 and to all of the previous as well as prior tubes. Click Yes To All. This will guarantee that all tubes have the same cytometer settings. 6.5 Data Recording: Running your Experiment Samples To keep the software from seizing up on you during acquisition, please make up additional experiments for runs with more than 100 samples. If additional experiments are needed to complete the run, after optimizing the first experiment duplicate the experiment without data Put your sample tube on the SIP Click Acquire, wait until the sample gets going and then click Record When the sample is finished recording, remove it from the SIP and click Next Tube to go on to the next sample If you need to Stop Recording before the set number of events have been reached, just click Stop Recording and then Stop Acquiring There are a number of ways to tell when you sample has finished recording: a bell will ding (if not muted), the blue bar in the acquisition control frame will have reached the end and the acquisition tube pointer box as well as the tube in the browser will turn green Repeat steps thru with the rest of your samples Standardizing your Experiments: To standardize your experiments, you need to run a bead after your first experiment to capture where the peaks of your fluorescence parameters fall. This bead should then be the first tube run with future experiments to set your voltages to the mean peak of the first experiment. (Ask Flow Facility staff for help with this approach if needed). 7.0 When finished Running Samples 7.1 Determine if you are, or are not, the last person scheduled to run for that day by looking at the calendar in FacMan. If there is a session behind you, please click on that session to see if the User has put CANCEL in the experiment box IF YOU ARE NOT the last person, then on high run 2 minutes 10% bleach and then 2 minutes dh 2 O => leave the tube of dh 2 O on the SIP => put the fluidics in Low and Standby => top off sheath tank => empty Waste (if necessary) IF YOU ARE the last person, then on high run 5 minutes 10% bleach and then 5 minutes dh 2 O => leave the tube of dh 2 O on the SIP => put the fluidics in Low and Standby => turn off the cytometer => turn off the wet cart (L02 only) => close Diva software and log Revised 1/2016 Page 13 of 26
14 7.2 Exporting Data out of the cytometer controller computer (do not shutdown computer) => top off sheath tank => empty waste (if necessary). When acquisition is completed you need to Export both your Experiment and your FCS files. We also suggest exporting your Experimental Template if you plan on using it more than once If you get an error message when trying to export your data, make sure that the F drive is connected to the controller by clicking My Computer. If F drive says disconnected, click on the Map_Drive icon on the desktop and follow the prompts If F drive still doesn t connect, try rebooting both the Controller computer and the Workstation computer If F drive still fails to connect, export to the local D drive and let the Flow Facility staff know so that they can trouble shoot and transfer your data to the correct drive. 7.3 Exporting Experiment: Right click on the Experiment Book => Export => Export Experiment => OK NOTE: The experiment should export to the default location: F:\BDExport\Experiments If not, Browse => F drive => BDExport => Experiments. 7.4 Exporting FCS Files: Right click on the Experiment Book => Export => FCS => 3.0 format => OK => Save NOTE: The FCS should export to the default location: F:\BDExport\FCS If not, Browse => F drive => BDExport => FCS. 7.5 Exporting Experiment Template: Right click on the Experiment Book => Export => Experiment Template => Type => Select your Initials => Name the Template so that you will know what template it is in the future => Finish. (e.g. 7 color PBMC) NOTE: If you check the Lock Template box, this ensures that the template cannot be overwritten by a template with the same name NOTE: We suggest that you keep a copy of every different template you make up, because to manage the Data Base, we need to delete experiments off of the Browser as well as off of the export drive. These templates stay on the D drive until the User Revised 1/2016 Page 14 of 26
15 leaves the lab or an upgrade to the software is made which will no longer read the older formatted templates. 7.6 Transferring Data to CDMS/Biotrue: After you have exported your Experiment and FCS files, you need to Transfer your FCS folder of data to our Collaborative Data Management System (CDMS)/Biotrue from the Workstation computer Go to the Workstation computer and log in if you haven t already done so (log in information is on the Quick Fact sheet on the wall next to the Workstation computer) Open the Biotrue Uploader located on the desktop and log in (log in information is on the Quick Fact sheet on the wall next to the Workstation computer) Open the BDExport folder located on the desktop => FCS => Days folder => Drag your FCS folder into your Lab s Biotrue dropbox => click Yes, please to upload NOTE: If you are unable to Export your data or transfer it to Biotrue, please contact the Flow Facility via or phone to let us know so that we can assist you NOTE: It is the end Users responsibility to export and upload their FCS files to Biotrue. Facility staff will no longer be checking behind you to make sure you have uploaded your FCS Files NOTE: If you have recorded a LARGE amount of data, please do a property check on your FCS folder to see how big the file is. If it is greater than 500MB, you need to divide the files into two folders before uploading to Biotrue to prevent corruption of any of the files. Please name the second folder the same as the original and put the word cont. at the end. 7.7 Closing FacMan Session: The Fac Man session needs to have the protocol sheet uploaded and the session completed with survey immediately after running the experiment to keep from delaying data management of the Flow Facility staff Update your protocol sheet, and upload it to your Fac Man session BEFORE completing the session End your session in Fac Man and fill out the required survey completely (see attachment #7) NOTE: If you leave ANY question in the survey blank, your session gets red flagged until that question has been answered. Until the survey is completed, the session cannot be moved to invoicing status. So, please remember that you need to go back into that session and answer the question after you have completed the task at hand. Revised 1/2016 Page 15 of 26
16 7.8 If you are the last person, log off of the Workstation computer, please do not shut down the computer, just log off. 7.9 Spray your cooler with 70% Ethanol before leaving the lab Accessing Biotrue in your Lab: URL: Username: flowguest Password: DHVIflow Select DHVI Flow Cytometry Facility and your lab flow data folder 8.0 Attachments 8.1 #1: DHVI Flow Facility Contact List 8.2 #2: Training Checklist / Homework 8.3 #3: Overview of BD LSRII/Fortessa Instruments 8.4 #4: Overview of BD FACSDiva Software 8.5 #5: Cytometer Specific Protocol Form/Template 8.6 #6: Example Job/Session Details Survey 8.7 #7: Troubleshooting Sheath Reservoir Revised 1/2016 Page 16 of 26
17 Attachment #1: DHVI Flow Facility Contact List DHVI Flow Contact List General User Support: FACMAN HELP: MSRBII HELP GHRB HELP Labs/Instruments: A01: GHRB, A02: MSRBII, N01: MSRBII, F01: MSRBII, Bay 5/6 Rm L01: MSRBII, Bay 5/6 Rm L02: MSRBII, Bay 5/6 Rm /29/15 Role Faculty/Staff FacMan Registration Peters Julie Analyst (A01) Cumming Ian Analyst (CHAVI-ID) Jones-Marshall Dawn Analyst (A02) McDermott Patti Scientific Advisor, CHAVI-ID Moody Tony Billing/Finance Hope Tony Facility Director Sempowski Greg Director for Instrumentation Whitesides John Office: Cell: Home: Pager: Revised 1/2016 Page 17 of 26
18 Attachment #2: Training Checklist Prerequisites for Classroom Session! Complete the Introduction to Flow Cytometry web-based training on BD Bioscience s website Will orient the trainee to the basic principals of Flow Cytometry. (~1 hour to complete; required quizzes). 1. URL: 2. Take the quizzes at the end of each section. 3. Record quiz results and bring to Classroom session.! Review the BD FACS Diva Software Overview web-based training on BD Bioscience s website Will orient the trainee to the basic components and function of the acquisition software. (~4 hours to complete). 1. URL: Reference%20Manual.pdf Focus on sections: BD FACSDiva Workspace, Tools for Data Analysis and Data Management.! Print and Review DHVI SOP Flow_010: Policies and Procedures for Flow Facility Users. PDF available on DHVI Flow Facility website: 1. URL: 2. Bring printed copy to Classroom session. Prerequisites for Hands-On Session! Print and Review the appropriate manual: DHVI Flow Facility Training Manual_L01/L02/F01. PDF available on DHVI Flow Facility website: 1. URL: 2. Be sure to review the attachment materials introducing you to the LSRII/LSRFortessa instruments and the Diva Software. 3. Bring printed copy to Hands-On session and Exam session.! Review BD LSRII User's Guide. Please orient yourself with this reference guide and its content. PDF available on DHVI Flow Facility website: 1. URL: Revised 1/2016 Page 18 of 26
19 Attachment #3: Overview of BD LSRII Instrument Figures/text captured from BD LSRII User s Guide See full PDF for details, on Facility website. Components Revised 1/2016 Page 19 of 26
20 Control Panel Revised 1/2016 Page 20 of 26
21 Fluid Control Revised 1/2016 Page 21 of 26
22 Sample Injection Port (SIP) Revised 1/2016 Page 22 of 26
23 Attachment #4: Overview of BD FACSDiva Software Online Course: rence%20manual.pdf Workspace Components Menu Bar Workspace Toolbar Save Browser Frame - Search Field - Browser Toolbar -Folders -Experiments -Specimens -Tubes - Global Instrument Settings - Global Worksheet - Current Tube Pointer Plate Frame Cytometer Frame -Status of the Instrument -Laser Delay/Area Scaling -Instrument Settings for Current Tube Inspector Frame -Experiment Information -Tube Information -Acquisition Information -Instrument Setting Information -Worksheet Information -Plot Information -Plate Information -Keyword Information Worksheet Frame -Worksheet Toolbar -Toggle between Global and Normal Worksheets -Select Arrow -Plots -Magnification Tools -Gating Tools -Editing Tools -Alignment Tools -Population Hierarchy -Statistics View Acquisition Dashboard Frame Revised 1/2016 Page 23 of 26
24 Attachment #5: Cytometer Specific Protocol Form/Template Download CURRENT version from Facility website: Revised 1/2016 Page 24 of 26
25 Attachment #6: EXAMPLE FacMan Job/Session Details Survey Revised 1/2016 Page 25 of 26
26 Attachment #7: Troubleshooting Sheath Reservoir (L02 Only) 1.0 Removing Air from System When The Sheath Reservoir Runs dry: 1.1 Put a tube of dh 2 O on SIP. 1.2 Press prime, and when you see bubble in the water tube press standby (this empties the flow cell). 1.3 While in standby, depressurize the sheath reservoir by loosening the white cap on the reservoir. 1.4 Toggle between prime and run waiting 2 seconds between toggles (repeat a dozen times). 1.5 Put it in standby, and tighten the white cap. 1.6 Wait 40 seconds and then put the machine in run. 1.7 Put the CS&T beads on to make sure that the FSC/SSC are the same as they were during the QC in the morning. If not, repeat the above steps. Revised 1/2016 Page 26 of 26
FACSCanto RUO Special Order QUICK REFERENCE GUIDE
FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW
More informationGETTING STARTED. 6. Click the New Specimen icon.
BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified
More informationHow to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
More informationLSR II Users Guide Stanford Shared FACS Facility 06/10/13
LSR II Users Guide Stanford Shared FACS Facility 06/10/13 A configuration sheet is located on the Website in the Instrument Overview area Do I need to run beads when I collect data using Diva Digital?
More informationBD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done
More informationUCHC Guide to FACS DiVa. LSR II instruments
UCHC Guide to FACS DiVa LSR II instruments Start Up: Turn on the instrument using the Green power button. Fluidics pump power, LSR II-B & C LSR II-B and LSRII-C-> Turn on the fluidics pump and ensure that
More informationBecton-Dickinson FACSCalibur Flow Cytometer Instructions
Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If
More informationNavios Quick Reference
Navios Quick Reference Purpose: The purpose of this Quick Reference is to provide a simple step by step outline of the information needed to perform various tasks on the system. We begin with basic tasks
More informationBD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin
BD LSR II and FACSDiVa Software, Basel Dr. Norbert Leclere,, Berlin Overview Electronics Covers Fluidics Connectors Controlpanel Sample Port On/Off The Control Panel Ease of Use The basic procedures to
More informationTROUBLESHOOTING FLOW CHART
TROUBLESHOOTING FLOW CHART Revised and updated on January 16, 2003, by Tony Leger and Automation Lab Technology. FACScan(FC)\FACSort(FST) FACScalibur (FCB) Fluidics/Pneumatics troubleshooting guide DCM
More informationInstitut für Experimentelle Immunologie und Bildgebung
Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check
More informationAnalyzer Experiment setup guide for LSRII and Canto s
Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,
More informationInstructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835
Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that
More informationBD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
More informationUCHC Guide to CellQuest Pro. FACS Calibur instruments
UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn
More informationLEGENDplex Data Analysis Software
LEGENDplex Data Analysis Software Version 7.0 User Guide Copyright 2013-2014 VigeneTech. All rights reserved. Contents Introduction... 1 Lesson 1 - The Workspace... 2 Lesson 2 Quantitative Wizard... 3
More information1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8
Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms
More informationBulkSMS Text Messenger Product Manual
BulkSMS Text Messenger Product Manual 1. Installing the software 1.1. Download the BulkSMS Text Messenger Go to www.bulksms.com and choose your country. process. Click on products on the top menu and select
More informationImmunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...
Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and
More informationSTEPfwd Quick Start Guide
CERT/Software Engineering Institute June 2016 http://www.sei.cmu.edu Table of Contents Welcome to STEPfwd! 3 Becoming a Registered User of STEPfwd 4 Learning the Home Page Layout 5 Understanding My View
More informationAfter completing this module, you will be able to: Describe the BD FACScan system and how it works.
BD FACScan System After completing this module, you will be able to: Describe the BD FACScan system and how it works. Perform instrument startup and shutdown procedures. Describe instrument maintenance
More informationWebSphere Business Monitor V6.2 KPI history and prediction lab
Copyright IBM Corporation 2009 All rights reserved IBM WEBSPHERE BUSINESS MONITOR 6.2 LAB EXERCISE WebSphere Business Monitor V6.2 KPI history and prediction lab What this exercise is about... 1 Lab requirements...
More informationCyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version 6.0.626)
CyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version 6.0.626) *Amended from Sampler Manual created by Zach Bjornson Because CyTOF computer is on a private network, you will be bombarded with
More informationAttune Acoustic Focusing Cytometer
USER GUIDE Attune Acoustic Focusing Cytometer USER GUIDE Publication Part Number 4453328 Rev. F Revision Date June 2012 Copyright 2012, Life Technologies Corporation. All rights reserved. For Research
More informationBIGPOND ONLINE STORAGE USER GUIDE Issue 1.1.0-18 August 2005
BIGPOND ONLINE STORAGE USER GUIDE Issue 1.1.0-18 August 2005 PLEASE NOTE: The contents of this publication, and any associated documentation provided to you, must not be disclosed to any third party without
More informationedgebooks Quick Start Guide 4
edgebooks Quick Start Guide 4 memories made easy SECTION 1: Installing FotoFusion Please follow the steps in this section to install FotoFusion to your computer. 1. Please close all open applications prior
More informationUser Guide for TASKE Desktop
User Guide for TASKE Desktop For Avaya Aura Communication Manager with Aura Application Enablement Services Version: 8.9 Date: 2013-03 This document is provided to you for informational purposes only.
More informationGetting Started with BD FACSDiva Software
Getting Started with BD FACSDiva Software For In Vitro Diagnostic Use bdbiosciences.com Part No. 642213 Rev. A June 2007 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 USA Tel (877) 232-8995 Fax
More informationScience is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening
More informationCell Sorting. S3e Cell Sorter Quick Guide
Cell Sorting S3e Cell Sorter Quick Guide STARTUP Checking Fluidics STARTUP There are three bulk fluid containers (Figure 1) that may require attention before starting the system. If fluidics need to be
More informationBeckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab
Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Room 213 Beckman Coulter DTX 880 Multimode Detector Information The Beckman Coulter DTX 880 Multimode Detector is
More informationInternet Explorer 7. Getting Started The Internet Explorer Window. Tabs NEW! Working with the Tab Row. Microsoft QUICK Source
Microsoft QUICK Source Internet Explorer 7 Getting Started The Internet Explorer Window u v w x y { Using the Command Bar The Command Bar contains shortcut buttons for Internet Explorer tools. To expand
More informationFrontDesk Installation And Configuration
Chapter 2 FrontDesk Installation And Configuration FrontDesk v4.1.25 FrontDesk Software Install Online Software Activation Installing State Related Databases Setting up a Workstation Internet Transfer
More informationCM HOST CM CardTransporter Fuel Communication and Management Software 10.10.06 Software version up to 3.1
CM HOST CM CardTransporter Fuel Communication and Management Software 10.10.06 Software version up to 3.1 CM Host Manual For CardMaster Fuel Control www.cardlockvending.com customers call 888-487-5040
More informationDIIMS Records Classifier Guide
DIIMS Records Classifier Guide Featuring Content Server 10 Second Edition, November 2012 Table of Contents Contents 1. DIIMS Overview... 3 1.1 An Overview of DIIMS within the GNWT... 3 1.1.1 Purpose of
More informationUsing the enclosed installation diagram, drill three holes in the wall with the lower hole 1150mm from the floor.
Terminal Installation When choosing the location of the terminal, care should be taken to select an area with consistent light levels throughout the day and avoid areas where the unit may be subjected
More informationSalesforce Classic Guide for iphone
Salesforce Classic Guide for iphone Version 37.0, Summer 16 @salesforcedocs Last updated: July 12, 2016 Copyright 2000 2016 salesforce.com, inc. All rights reserved. Salesforce is a registered trademark
More informationCreating a Poster in PowerPoint 2010. A. Set Up Your Poster
View the Best Practices in Poster Design located at http://www.emich.edu/training/poster before you begin creating a poster. Then in PowerPoint: (A) set up the poster size and orientation, (B) add and
More informationCity of De Pere. Halogen How To Guide
City of De Pere Halogen How To Guide Page1 (revised 12/14/2015) Halogen Performance Management website address: https://global.hgncloud.com/cityofdepere/welcome.jsp The following steps take place to complete
More informationuser guide the conference call people Simple Effective Collaboration
user guide the conference call people Simple Effective Collaboration Introduction MeetingZone is a fully automated conferencing service that lets you set up and manage audio and web conferences. Your MeetingSpace
More informationCustomer admin guide. UC Management Centre
Customer admin guide UC Management Centre June 2013 Contents 1. Introduction 1.1 Logging into the UC Management Centre 1.2 Language Options 1.3 Navigating Around the UC Management Centre 4 4 5 5 2. Customers
More informationIntegrated Invoicing and Debt Management System for Mac OS X
Integrated Invoicing and Debt Management System for Mac OS X Program version: 6.3 110401 2011 HansaWorld Ireland Limited, Dublin, Ireland Preface Standard Invoicing is a powerful invoicing and debt management
More informationTable of Contents. Manual for Core Staff - Equipment/Scheduling Core Facilities
Table of Contents 1. Overview 2. How do I manage my account? 3. Equipment Scheduling Workflow Overview 4. Equipment Scheduling Walk Through a. How do I access the list of calendars available for scheduling?
More informationFrog VLE Update. Latest Features and Enhancements. September 2014
1 Frog VLE Update Latest Features and Enhancements September 2014 2 Frog VLE Update: September 2014 Contents New Features Overview... 1 Enhancements Overview... 2 New Features... 3 Site Backgrounds...
More informationFACS-CALIBUR. USER-Guide
FACS Calibur User guide Version 2.1-03/2006 AXS 1 FACS-CALIBUR Version 2.1 USER-Guide Note: This guide does note compensate for a personal introduction from an operator before first usage Draft mainly
More informationPassport installation Windows 8 + Firefox
Passport installation Windows 8 + Firefox OS Version Windows 8 System Type 32-bit and 64-bit Browser name & version Firefox 23 Before you begin Check you have Anti Virus software installed on your computer
More informationFebruary 2013 Copyright 2013 by CTB McGraw-Hill Education. 1
February 2013 Copyright 2013 by CTB McGraw-Hill Education. 1 OCCT & OMAAP Welcome to the Record Editing System (RES) Utility... 3 About Your Task... 3 Security Concerns... 3 Before You Begin... 3 Contacting
More informationUtilities. 2003... ComCash
Utilities ComCash Utilities All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including photocopying, recording, taping, or
More informationCOMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
More informationTable of Contents. Page 3
Welcome to Exchange Mail Customer Full Name Your e-mail is now being delivered and stored on the new Exchange server. Your new e-mail address is @rit.edu. This is the e-mail address that you should give
More informationLaser Cutter User Manual
Laser Cutter User Manual frequently asked questions... the laser is cutting weird! it s cutting a thick line or not cutting through at all! Table Of Contents Section 1: Materials Guide 1.1 Can I Cut this?
More informationPassport Installation. Windows 8 + Internet Explorer 10
Passport Installation Windows 8 + Internet Explorer 10 OS Version Windows 8 System Type 32bit and 64bit Browser Name & version Internet Explorer 10 Before you begin Make sure you have Anti Virus software
More informationUSAccess System- Registrar. Help Guide. Prepared for
USAccess System- Registrar Help Guide Prepared for United States Department of Agriculture Office of Security 300 7th Street SW, Washington DC 20024 Version 1.0 December 12, 2007 USAccess System Registrar
More informationLinux provides two basic types of interface for you to use when working with your
CHAPTER 6 Using Your Desktop Linux provides two basic types of interface for you to use when working with your computer: GUI (graphical user interface) and CLI (command-line interface). An overview of
More informationUSER GUIDE MANTRA WEB EXTRACTOR. www.altiliagroup.com
USER GUIDE MANTRA WEB EXTRACTOR www.altiliagroup.com Page 1 of 57 MANTRA WEB EXTRACTOR USER GUIDE TABLE OF CONTENTS CONVENTIONS... 2 CHAPTER 2 BASICS... 6 CHAPTER 3 - WORKSPACE... 7 Menu bar 7 Toolbar
More informationDecision Support AITS University Administration. Web Intelligence Rich Client 4.1 User Guide
Decision Support AITS University Administration Web Intelligence Rich Client 4.1 User Guide 2 P age Web Intelligence 4.1 User Guide Web Intelligence 4.1 User Guide Contents Getting Started in Web Intelligence
More informationOnce you have obtained a username and password you must open one of the compatible web browsers and go to the following address to begin:
CONTENT MANAGER GUIDELINES Content Manager is a web-based application created by Scala that allows users to have the media they upload be sent out to individual players in many locations. It includes many
More informationHow To Use Senior Systems Cloud Services
Senior Systems Cloud Services In this guide... Senior Systems Cloud Services 1 Cloud Services User Guide 2 Working In Your Cloud Environment 3 Cloud Profile Management Tool 6 How To Save Files 8 How To
More informationJROTC Unit Management System (JUMS) User s Guide
JROTC Unit Management System (JUMS) User s Guide 1 March 2016 Summary of Changes JUMS User Guide This revision, dated 1 March 2016 o Updates Chapter 3, paragraph 3-2, Import New Cadets from a File into
More informationWhat Do You Think? for Instructors
Accessing course reports and analysis views What Do You Think? for Instructors Introduction As an instructor, you can use the What Do You Think? Course Evaluation System to see student course evaluation
More informationithenticate User Manual
ithenticate User Manual Updated November 20, 2009 Contents Introduction 4 New Users 4 Logging In 4 Resetting Your Password 5 Changing Your Password or Username 6 The ithenticate Account Homepage 7 Main
More informationSAP Business Intelligence ( BI ) Financial and Budget Reporting. 7.0 Edition. (Best Seller At Least 43 copies Sold)
SAP Business Intelligence ( BI ) Financial and Budget Reporting 7.0 Edition (Best Seller At Least 43 copies Sold) November 2011 Table of Contents Log In... 3 Initial Variable Screen... 5 Multiple / Single
More informationMcAfee Endpoint Encryption Reporting Tool
McAfee Endpoint Encryption Reporting Tool User Guide Version 5.2.13 McAfee, Inc. McAfee, Inc. 3965 Freedom Circle, Santa Clara, CA 95054, USA Tel: (+1) 888.847.8766 For more information regarding local
More informationGetting Started with Mamut Online Desktop
// Mamut Business Software Getting Started with Mamut Online Desktop Getting Started with Mamut Online Desktop Contents Welcome to Mamut Online Desktop... 3 Getting Started... 6 Status... 23 Contact...
More informationDoes the GC have an online document management solution?
This FAQ contains: Web File Services definition Access web file services Create a WFS folder Upload a document Email a link to a document folder Create WFS subscription Cancel existing subscription Create
More informationTimeless Time and Expense Version 3.0. Copyright 1997-2009 MAG Softwrx, Inc.
Timeless Time and Expense Version 3.0 Timeless Time and Expense All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including
More informationBD CellQuest Pro Software Analysis Tutorial
BD CellQuest Pro Analysis Tutorial This tutorial guides you through an analysis example using BD CellQuest Pro software. If you are already familiar with BD CellQuest Pro software on Mac OS 9, refer to
More informationIntegrated Accounting System for Mac OS X
Integrated Accounting System for Mac OS X Program version: 6.3 110401 2011 HansaWorld Ireland Limited, Dublin, Ireland Preface Standard Accounts is a powerful accounting system for Mac OS X. Text in square
More information13 Managing Devices. Your computer is an assembly of many components from different manufacturers. LESSON OBJECTIVES
LESSON 13 Managing Devices OBJECTIVES After completing this lesson, you will be able to: 1. Open System Properties. 2. Use Device Manager. 3. Understand hardware profiles. 4. Set performance options. Estimated
More informationTABLE OF CONTENTS. Do not click to Remember my password, so that unauthorized persons will not access your gradebook.
TABLE OF CONTENTS 1. Logging into etac 2. Creating a Short Cut on Your Desktop 3. Navigating Around Your Home Page 4. Taking Attendance 5. Printing a Class Roster with Columns 6. Grading Templates/Categories
More informationWebSphere Business Monitor V6.2 Business space dashboards
Copyright IBM Corporation 2009 All rights reserved IBM WEBSPHERE BUSINESS MONITOR 6.2 LAB EXERCISE WebSphere Business Monitor V6.2 What this exercise is about... 2 Lab requirements... 2 What you should
More informationShimadzu UV-1700 UV/Vis Spectrometer Updated July 13, 2010
Shimadzu UV-1700 UV/Vis Spectrometer Updated July 13, 2010 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the
More informationIncreasing Productivity and Collaboration with Google Docs. Charina Ong Educational Technologist charina.ong@nus.edu.sg
Increasing Productivity and Collaboration with Google Docs charina.ong@nus.edu.sg Table of Contents About the Workshop... i Workshop Objectives... i Session Prerequisites... i Google Apps... 1 Creating
More informationHow To Create A Powerpoint Intelligence Report In A Pivot Table In A Powerpoints.Com
Sage 500 ERP Intelligence Reporting Getting Started Guide 27.11.2012 Table of Contents 1.0 Getting started 3 2.0 Managing your reports 10 3.0 Defining report properties 18 4.0 Creating a simple PivotTable
More informationUniFLOW Print and Copy Instructions for Students
UniFLOW Print and Copy Instructions for Students Getting Started, Setting up your PIN Page 2 Purchasing Credits for printing and Photocopying Page 4 How to use the MFD Page 8 Photocopying Page 10 Printing
More informationSMART BOARD USER GUIDE FOR PC TABLE OF CONTENTS I. BEFORE YOU USE THE SMART BOARD. What is it?
SMART BOARD USER GUIDE FOR PC What is it? SMART Board is an interactive whiteboard available in an increasing number of classrooms at the University of Tennessee. While your laptop image is projected on
More informationeinstruction CPS (Clicker) Instructions
Two major approaches to run Clickers a. Anonymous b. Tracked Student picks any pad as s/he enters classroom; Student responds to question, but pad is not linked to student; Good for controversial questions,
More informationWebSphere Business Monitor V7.0 Business space dashboards
Copyright IBM Corporation 2010 All rights reserved IBM WEBSPHERE BUSINESS MONITOR 7.0 LAB EXERCISE WebSphere Business Monitor V7.0 What this exercise is about... 2 Lab requirements... 2 What you should
More informationDigital Signage with Apps
Version v1.0.0 Digital Signage with Apps Copyright 2012 Syabas Technology, All Rights Reserved 2 Digital Signage with Apps Project...6 New Project...6 Scheduler...6 Layout Panel...7 Property Panel...8
More informationBanner Document Management Suite (BDMS) Web Access Help
May 10 th, 2011 Banner Document Management Suite (BDMS) Web Access Help Division of Information Technology AppXtender Web Access Help: For questions regarding AppXtender Web Access, please contact the
More informationThe FlexiSchools Online Order Management System Installation Guide
The FlexiSchools Online Order Management System Installation Guide FlexiSchools May 2012 Page 1 of 16 Installation Pack Welcome to the FlexiSchools system. You will have received a disc containing: Sato
More informationAppendix C. Vernier Tutorial
C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer
More informationHow To Write A Cq5 Authoring Manual On An Ubuntu Cq 5.2.2 (Windows) (Windows 5) (Mac) (Apple) (Amd) (Powerbook) (Html) (Web) (Font
Adobe CQ5 Authoring Basics Print Manual SFU s Content Management System SFU IT Services CMS Team ABSTRACT A summary of CQ5 Authoring Basics including: Setup and Login, CQ Interface Tour, Versioning, Uploading
More informationUsing an Edline Gradebook. EGP Teacher Guide
Using an Edline Gradebook EGP Teacher Guide Table of Contents Introduction...3 Setup...3 Get the Gradebook Web Plugin... 3 Using Your Web Gradebook... 4 Using the Web Gradebook on a Shared Computer...
More informationMail Merge Microsoft Word and Excel Queries Scott Kern Senior Consultant
Mail Merge Microsoft Word and Excel Queries Scott Kern Senior Consultant What We ll Cover 1. Enabling database connections through Microsoft Excel 2. Accessing the data stored in the SQL Database via the
More informationAttune NxT Acoustic Focusing Cytometer
USER GUIDE Attune NxT Acoustic Focusing Cytometer Catalog Number A24858, A24859, A24860, A24861, A24862, A24863, A24864 Publication Number 100024235 Revision B.0 For Research Use Only. Not for use in diagnostic
More informationSet Top Box User Guide. Global Reach with a Local Touch
Set Top Box User Guide Global Reach with a Local Touch Table of Contents REMOTE SET UP 3 SET TOP BOX DIAGRAM 4 SET TOP BOX SETTINGS 5 TURN ON/OFF 5 TV INPUT SETTINGS 5 ADJUSTING THE VOLUME 5 RESETTING
More informationAccessing Citrix on a MAC using OS X (Mountain Lion and Newer)
These instructions are specifically designed for OS X Mountain Lion and Mavericks. Earlier versions of the OS will require a different installation. These instructions assume that there are no Java, Citrix
More informationHosted VoIP Phone System. Admin Portal User Guide for. Call Center Administration
Hosted VoIP Phone System Admin Portal User Guide for Call Center Administration Contents Table of Figures... 4 1 About this Guide... 6 2 Accessing the Hosted VoIP Phone System Administration Portal...
More informationMicrosoft SharePoint 2010 End User Quick Reference Card
Microsoft SharePoint 2010 End User Quick Reference Card Microsoft SharePoint 2010 brings together the people, documents, information, and ideas of the University into a customizable workspace where everyone
More informationExcel Tutorial. Bio 150B Excel Tutorial 1
Bio 15B Excel Tutorial 1 Excel Tutorial As part of your laboratory write-ups and reports during this semester you will be required to collect and present data in an appropriate format. To organize and
More informationXythos on Demand Quick Start Guide For Xythos Drive
Xythos on Demand Quick Start Guide For Xythos Drive What is Xythos on Demand? Xythos on Demand is not your ordinary online storage or file sharing web site. Instead, it is an enterprise-class document
More informationithenticate User Manual
ithenticate User Manual Version: 2.0.2 Updated March 16, 2012 Contents Introduction 4 New Users 4 Logging In 4 Resetting Your Password 5 Changing Your Password or Username 6 The ithenticate Account Homepage
More informationClip Art in Office 2000
Clip Art in Office 2000 In the process of making a certificate, we will cover: Adding clipart and templates from the Microsoft Office Clip Gallery, Modifying clip art by grouping and ungrouping, Flipping
More informationUsing SANDBOXIE to Safely Browse the Internet (verified with ver 4.20) Jim McKnight www.jimopi.net Sandboxie.lwp revised 5-25-2016
Using SANDBOXIE to Safely Browse the Internet (verified with ver 4.20) Jim McKnight www.jimopi.net Sandboxie.lwp revised 5-25-2016 GENERAL NOTES: These tips are based on my personal experience using Sandboxie.
More informationBD FC Beads 8-Color Kit for BD OneFlow Assays
BD FC Beads 8-Color Kit for BD OneFlow Assays 7/2014 23-15785-00 IVD 5 tests per kit Catalog No. 658621 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2014 BD Becton,
More informationModule 2 Using the Database
Health Information System (HIS) Part Three: Data Management Module 2 Using the Database CONTENTS 2.1 Before you start...................................................11 2.2 The first time you start..............................................12
More informationOPERATION MANUAL. MV-410RGB Layout Editor. Version 2.1- higher
OPERATION MANUAL MV-410RGB Layout Editor Version 2.1- higher Table of Contents 1. Setup... 1 1-1. Overview... 1 1-2. System Requirements... 1 1-3. Operation Flow... 1 1-4. Installing MV-410RGB Layout
More informationHow to use SURA in three simple steps:
How to use SURA in three simple steps: Most of SURA s functionality can be accessed through these three steps. 1) Download SURA to your computer Go to the SU Downloads page to obtain the SURA utility.
More informationGetting Started Guide
Getting Started Guide Introduction... 3 What is Pastel Partner (BIC)?... 3 System Requirements... 4 Getting Started Guide... 6 Standard Reports Available... 6 Accessing the Pastel Partner (BIC) Reports...
More information