The Laser Diode. Figure 1. Internal structure of a typical diode laser.

Size: px
Start display at page:

Download "The Laser Diode. Figure 1. Internal structure of a typical diode laser."

Transcription

1 The Laser Diode 1 Introduction This set of laboratory experiments is designed to have you become familiar with the properties of laser diodes and to become familiar with optical spectrum analyzers typically used in fiber and laser instrumentation to asses spectral characteristics of sources. In addition you will become familiar with the technique of using an optical grating to stabilize and to tune the frequency of a laser diode. The discussion and figures in this section are for the Hitachi 633 nm, HL6314MG diode laser. Figure 1 shows the diode laser with a cut-away of the housing where the chip mounts. Figure 2 shows the laser chip structure, typical chip dimensions, the direction of the forward current, and the radiation pattern. The radiation is produced in the active layer, which is a small fraction of the height of the chip, hence the radiation is diffracted when it emerges from the active region analogously to the diffraction of radiation passing through a narrow slit. The diffraction produces the radiation pattern shown in Fig. 2. The polarization of the emitted light is in the plane of the active junction, and therefore parallel to the short axis of the elliptically patterned light field. Figure 1. Internal structure of a typical diode laser. Figure 2. Chip structure of a typical diode laser. Figure 3 shows the package type and the internal circuit of HL6314MG. Notice that the package includes a built-in photodiode for the power-monitoring purpose. The common pin of the laser diode and the photodiode is usually designated as the ground pin that is connected to the case. Figure 4 shows the actual package dimensions and the pin layout. These two figures allow you to determine the Diode Laser Experiment Page 1 of 10

2 appropriate mechanical mounting configuration and electric connection for the laser diode. Figure 5 displays a number of important operating characteristics of the HL6314MG laser diode. The panel in the upper left corner shows the optical output power versus the forward bias current, measured under various temperatures. You should notice two important things. First, the laser diode output power rises sharply after the forward current exceeds a certain value, namely the lasing threshold current (25 ma). Second, the value of the threshold current increases with the rising temperature of the laser diode. The maximum operating forward current (50 ma) and power output (3 mw) are specified by the manufacturer for each diode laser. The upper right corner in Fig. 5 shows the relationship between the output power of the laser diode and the current of the monitoring photodiode (under a reverse bias of 5V). A calibrated external power meter would allow you to verify this relationship. The ellipticity of the emitted light field pattern is shown in the left lower corner of Fig. 5. The right lower corner of Fig. 5 shows that the laser diode operates in an increasingly coherent fashion when its output power is increased, with a more effective suppression of the side-modes and a narrower linewidth of the main mode. Figure 3. Package type and internal circuit of HL6314MG laser diode. Figure 4. Package dimensions of HL6314MG laser diode. Diode Laser Experiment Page 2 of 10

3 Figure 5. Important laser diode operating characteristics. The case temperature of the diode laser plays an important role in laser operations and Figure 6 shows the various effects. In general, when the case temperature rises, the diode laser operates less efficiently, with a rising threshold and decreasing slope efficiency (a conversion factor from the injection current to the actual optical output power). These scenarios are clearly shown in the top two panels in Fig. 6. Case temperature can be used to tune the operating wavelength of the laser diode. However, the wavelength tuning curve is not smooth over a large range of temperature changes, rather it will display a staircase type of tuning characteristics, as shown in the right lower corner of Fig. 6. This is because the laser diode actually jumps to different longitudinal modes when the temperature changes over a large range. While it is possible to change the laser wavelength by a couple of nanometers when the case temperature changes by 10 degree for example, the temperature tuning coefficient for a particular longitudinal lasing mode (without mode-hopping) is usually much smaller. The capability of continuous tuning of the diode laser wavelength without mode-hop is important for a number of applications in the AMO physics, including precision spectroscopy, laser cooling, and chemical sensing, etc. Combined adjustment of temperature and injection current can accomplish wavelength tuning to a certain degree, but usually within a fairly limited range. External cavity laser Diode Laser Experiment Page 3 of 10

4 diodes help solve this problem. Ordinarily, for a solitary laser diode, the laser cavity is formed between the two cleaved facets of the semiconductor chip. The facets are smoother and flatter than any mechanically polished mirror. If there are no coatings on the end surfaces of the laser chip, then the reflectivity R of a surface is given by: R = n n 2 c a, where n c and n a, are the indices of refraction of n c + n a the chip and air, respectively. Sometimes a diode laser manufacturer provides a reduced reflective coating on the output facet of the chip; therefore, the reflectivity of this facet is less than that calculated in Exercise 1. When the output of the diode laser is being returned with the first-order diffracted beam from an external grating, an external laser cavity is formed between the grating and the far chip facet. This would be particularly true if the grating diffraction efficiency (40% or more) dominates over the residual reflectivity of the front facet and the external cavity laser will operate more reliably and stably. The laser wavelength can now be tuned by displacing and/or rotating the grating. A carefully designed external cavity diode laser can pull the wavelength of the original laser diode by more than 10 nm without adjusting the injection current or chip temperature. It can also tune a single longitudinal laser mode by tens or even hundreds of GHz without mode-hop. Figure 6. The effects of case temperature on laser diode operations. Diode Laser Experiment Page 4 of 10

5 2 Experiment 2.1 PREPARATION You will be using a Hitachi laser diode from ThorLabs No. HL6316G. Begin by looking over the specifications sheet for the diode. In particular you will want to notice the operating current. You would normally also note the maximum allowed operating current, but this specification sheet does not provide it. Instead note the Typical Characteristics Curves where they show up to about 50 ma. The typical operating current is about 30 ma. Please do not exceed the maximal output power of 3 mw. Carefully handle the laser diode and avoid any static electric charge that can easily damage the diode. The diode needs to be mounted onto a TCLDM9 ThorLabs laser diode head with correct polarity. When you are ready to begin turning on equipment, first find the ThorLabs LDC 500 laser diode controller. Briefly read through the instruction manual. When you first turn on the power it enters a safe mode. Check that the LED for the correct polarity is lit, otherwise make a change in the polarity switch located at the rear panel of the controller. At this point you should also notice that the OPEN CKT LED should be off, indicating the laser diode has been connected to the controller. Set the Display indicator to show the current limit (ILIM). Located a small hole labeled LIM I. There is a small trimmer potentiometer that can be adjusted with a jeweler s screwdriver. You must set this trimmer for the maximum output current. The digital LED indicator will indicate the limit level in milliamps. Set it to be between 40 and 45 ma. Check to see that the mode indicator is on I for current. If not, press it once and the mode should change from P or power to I. When the mode switch is set to current, the diode laser driver will control the current to the diode so that it remains at a fixed level determined by the level knob. When the mode switch is set to power, the driver controls the laser diode to provide fixed output power. For the entire experiment leave the mode switch set to current. Turn the level knob all the way counter-clockwise. Now set the display indicator to ILD. As the level knob is increased in the clockwise direction the current indicator will increase. At this time no current is actually going to the laser, it only indicates the current that will be supplied when the laser is enabled. The laser will be turned on after a few other items are attended to. 2.2 THE TEMPERATURE CONTROLLER. Make a cable connection between the temperature controller (ThorLabs TEC 2000) and the laser diode head TCLDM9. Read the TEC 2000 instruction manual briefly and then turn on the temperature controller. As we now know, a stable case temperature is critical for reliable operations of the laser diode. In a research laboratory, a temperature controller for a laser diode is normally kept active all the time. The temperature controller consists of three parts, a temperature sensing unit, a feedback control loop, and a servo transducer for heating or cooling. The temperature sensor can be a standard thermistor or an IC temperature transducer such as an AD590. Both types of temperature sensing elements are included in the ThorLabs diode mount TCLDM9, one to provide a temperature error signal to feed into the feedback loop located inside the temperature controller, the other to provide an independent readout of the diode's temperature. The temperature controller provides current to the thermoelectric cooler, also built in with TCLDM9, to activate temperature corrections. When TEC 2000 is first turned on, the front panel LCD display should become active and you can display the desired measurement value, such as the actual temperature (Tact), the set temperature (Tset), the current limit to TEC (Ilim), and the actual TEC current (Itec). Select a suitable current limit Diode Laser Experiment Page 5 of 10

6 (Ilim) using the small trimmer potentiometer, for example, at 1.5 A. Toggle the temperature sensorsetting switch located at the rear panel and observe the front panel display. The OPEN CKT LED should be off. The temperature of the laser diode is controlled by the set temperature. Set the display mode to Tset and use the big adjustment knob to set Tset at about 25 degrees C. Turn on the On/off switch for temperature control loop, the associated LED ( Enable ) should turn on. Set the display mode to Tact and monitor how the actual temperature approaches the set value. Connect a BNC cable between the CTL OUT BNC Jack located at the rear panel and a digital scope. This output port allows additional monitoring of the temperature settling behavior on the digital scope. You need to experiment with the feedback control loop. Read in the TEC 2000 manual (page 16) on the procedure to adjust the Proportional (P), Integrator (I), and Derivative (D) gains. It will be good if you have some rudimentary feedback control understandings to appreciate the adjustment procedure to be performed here. Monitor the temperature-settling curve on the digital scope. You may need to adjust the time base on the scope to accommodate a global view of the settling behavior. Adjust the PID gains until you achieve an optimized compromise between the settling time and the number of error signal overshoots. Of course to record a new settling curve you will need to set a new temperature value, say 2 degrees away from the previous one. Please sketch several of these settling curves on your lab book and record the important time scales (setting time, number and size of overshoots, etc.). At this time, you can also turn on the signal generator, the Ando spectrum analyzer, the external photodetector or optical power-meter, and the PZT driver, etc. 2.3 LASER DIODE Now preset the laser diode current to about 15 ma on LDC 500 and push the Enable button to activate the injection current to the laser diode. Turn up the level knob so that the current goes up to about 30 ma. Notice the output from the laser (*DO NOT LOOK DIRECTLY INTO THE LASER BEAM!*) The laser should have a collimating lens in front of it, and nothing else. If there is other optics in the way, such as grating, move it so that you can see the laser output on the wall of the room. Adjust the collimating lens to obtain the best possible collimation judged by your eyes. This is an important step to achieve the best external cavity alignment later on. Using the photodetector module or the optical power meter you will measure the power output as a function of current and temperature. Connect a BNC cable from the photodiode output to the oscilloscope. Place the photodetector in front of the laser beam, note the signal on the oscilloscope and optimize the oscilloscope settings. Distinguish the laser power from the ambient room light. With the diode laser case temperature set at 25 degree C, make a measurement of the detected output power (proportional to the voltage on the oscilloscope) as a function of laser diode current. Start considerably below threshold. Then increase current in steps of 3-5 ma to the maximum current. Then go back and fill in more data points near the threshold value of current. Make sure you keep track of the oscilloscope voltage setting as you take measurements. Now retake the power versus current curve at two more temperatures. Take one at nearly the coldest temperature that the controller can manage (try 5 or 10 degree C). Then take another at about 40 degrees C. Finally, set the current to about 20-25% above the threshold value for 25 degrees C and take a power versus temperature curve. Do not exceed about 45 degrees C. (The specification sheet says 50, Diode Laser Experiment Page 6 of 10

7 but you must account for the fact that the thermistor does not really measure the diode temperature but the temperature of the diode mount. A much more efficient way to take the threshold curve is to use an external modulation input to the current drive. Set the temperature at 25 degree C and the current at 25 ma (or the threshold of your particular laser diode). Set the waveform on the external signal generator to be Triangle or Sawtooth ramp, with a frequency of about 5 Hz. Verify on the scope that the output signal from the signal generator is symmetric around 0 V and the amplitude does not exceed 1 V. Connect a BNC cable between the signal generator and the MOD IN BNC jack on the rear panel of LDC 500. Connect another BNC cable from the CTL OUT of the LDC 500 to the x- channel of the scope. Connect the photodetector output to the y-channel of the scope. Set the scope in X-Y mode. What you will observe on the scope is a direct display of the laser diode output power versus current, similar to that shown in Fig. 5. Record the trace on your notebook and compare with what you have measured earlier. When you are done, make sure you first disconnect the external signal input to the current driver before you turn the laser diode current to zero. 2.4 THE OPTICAL SPECTRUM ANALYZER. In this next portion of the lab you will be taking the wavelength spectrum of the diode. The optical spectrum analyzer is at first intimidating but in the end rather user friendly. Familiarize yourself with the controls with the help of the manual. To obtain a spectrum you must inject the laser diode light into a multi-mode fiber. One end of the fiber is already mounted in an optical claw. Take the short (15mm) focal length lens and place it after the collimating lens of the laser diode. Use a post-it note to determine the position of the focal point. Then place the fiber at this position. The easiest way to align the fiber more precisely is to look directly at the output of the other end of the fiber. (Yes your eye will be safe at these power levels.) When the fiber is correctly positioned the light through the fiber is quite bright (you should stop looking into the beam before it gets too bright), and you can easily use the fiber to illuminate a nearby object. You can use the power meter to maximize the output power through the fiber. When the fiber is aligned, screw the output end into the spectrum analyzer. Note that the fiber is keyed, and correctly fits into the receptacle in only one orientation. The threaded collar is annoyingly difficult to screw into the spectrum analyzer but it can be done. Set the spectrum analyzer center wavelength to 635 nm and the span to 20 nm. Set the wavelength resolution to the highest value (0.1 nm or 0.5 nm depending upon units). Set the laser current above threshold. Set the vertical scale to be logarithmic. You should easily observe a peak in the spectrum with no substantial noise for +/- 5 nm or so around the peak. At this point center the peak and change the span to 10 nm. Adjust the vertical scale accordingly. Now observe the spectrum for a few values of current both above and below the threshold. Record any interesting features and phenomena in the spectrum as a function of power. Note in particular the sinusoid-like pattern in the "pedestal" near the peak. Measure the spacing (in nm) between successive peaks. What do you conclude about the length of this laser diode? Now measure the position of the peak as a function of temperature for a current about 20-30% above threshold. Also at 25 degrees C measure the wavelength of the central peak as a function of current. Record your findings. Diode Laser Experiment Page 7 of 10

8 2.5 EXTERNAL CAVITY LASER DIODE For this portion of the experiment you will employ an optical grating of 1800 lines per mm. The grating is placed on a mirror mount with a piezo mounted under the horizontal adjustment screw, which can be used to tilt the grating by minute amounts. Place the grating as close as possible to the collimating lens mount. Tilt the grating so that the first order diffracted beam is reflected back into the diode. The zeroth order is available for measurement. It is a bit tricky to ensure that the 1st order is indeed retro-reflected back into the laser. Follow this procedure: Place the photodetector on the zeroth order beam. A.C. couple the oscilloscope input and turn up the gain almost all the way. Center the trace on the screen. As you move the vertical and horizontal adjustment screws on the grating mount observe the scope trace. Tap lightly on the mount. When the beam is properly retro-reflected the output power will appear to be very sensitive to your tapping. Maximize the tapping noise observed on the scope with the two grating adjustment knobs. At this point the grating is adequately adjusted. You can also directly monitor the diode laser threshold curve on the digital scope, as described at the end of section 2.3. As you adjust the grating to achieve different amounts of optical feedback, you will notice threshold reduction and optical power enhancement. Now align the fiber again to obtain a spectrum of the laser diode. Adjust the temperature to 25 degrees and the current to about 15% above threshold. Hook up the piezo to the pzt drive. Turn the voltage fully counter-clockwise and turn on the power supply. Slowly turn up the piezo voltage and record its value from the pzt driver monitor. Note the peak on the spectrum analyzer. Does it shift? If so, by how much? You may want to decrease the wavelength span. Be careful, sometimes the laser mode-hops between two modes. If you see the spectrum change, pin down whether or not a mode hop is responsible for the change. We will now proceed to calibrate the piezo wavelength control using an optical interferometer. 2.6 UNBALANCED MICHELSON INTERFEROMETER Remove the fiber from the beam path. Construct an unbalanced-path Michelson interferometer. Use Fig. 7 as a guide though your optical arrangement may look rather different in detail. Start with a path length difference as large as you can -- place an extra mirror in the longer arm if you wish to fold it to make it still larger. Record the distances from the beamsplitter to the respective end mirrors for later use. Hook up the piezo to the signal generator rather than the power supply. Send the output of the signal generator to the oscilloscope channel 2 as well. Connect the synch out output of the generator to the external trigger input of the scope. Set the sweep to triangular. Set the scope to external trigger. Put the output of the generator to about halfway. Observe the triangle wave on the oscilloscope. Diode Laser Experiment Page 8 of 10

9 L 1 M 1 Beamsplitter Photodiode Output signal L 2 Diode laser Grating M 2 Figure 7. Unbalanced Michelson interferometer for calibrating the piezo controlled grating tilt. Adjust the output amplitude and the dc offset of the signal generator so that the triangle output does not go negative, and so that the upper portion of the waveform does not distort. Set the signal generator frequency to between 40 and 100 Hz. Place the photodetector at the output of the Michelson interferometer and set the oscilloscope to view the photodetector and triangle waveforms simultaneously. You should see a sinusoid output from the detector that is synchronized with the triangle wave. If the laser is oscillating in more than one mode then the photodetector may show a beat envelop on the sinusoid. Try changing the laser current a bit if you run into this. If you can t get rid of it, then tolerate it. Your goal is simply to measure the change in piezo voltage that causes one full fringe at the photodetector. One full fringe corresponds to c/2(l2-l1) change in frequency. Observe the geometry used to tilt the grating with the small green piezo attached to the optical mount. Using your above measurement, the grating specs of 1800 lines per mm, and the fact that it is used in littrow configuration, estimate the displacement of the piezo in microns/volt. Congratulations! You re done! 3 Pre-lab Questions 1. The index of refraction of GaAs and air are about 3.6 and l.0. What is the reflectivity of the chip facet? Compare your answer with the reflectivity of the mirrors used in the HeNe laser, which is about Diode Laser Experiment Page 9 of 10

10 2. Assuming the length of the diode laser cavity is 250 µm and the index of refraction of the cavity is 3.6, show that the frequency spacing ν of the axial modes is 167 GHz. (You may want to refer to the discussion of longitudinal modes in the mode matching experiment.) 3. What does the axial mode spacing correspond to in wavelength spacing? (Give your answer in nm) 4. You will be working with a grating with 1200 groves per millimeter. Calculate the littrow, or blaze, condition for 635 nm. That is, calculate the angle of incidence of the laser beam such that the first order is retro-reflected. 5. The grating above is tilted by 1x10-3 rad. What is the change in wavelength, in nm, of the laser diode? 6. You are using an unbalanced Michelson with a path length difference of 1.12 m. You observe that 8.3 V on the piezo that tilts the grating causes one complete fringe to appear on the oscilloscope (that is, for example, the photodetector output changes by one peak to the next). What is the frequency tuning constant in MHz/Volt of the piezo? Diode Laser Experiment Page 10 of 10

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

Lab 9: The Acousto-Optic Effect

Lab 9: The Acousto-Optic Effect Lab 9: The Acousto-Optic Effect Incoming Laser Beam Travelling Acoustic Wave (longitudinal wave) O A 1st order diffracted laser beam A 1 Introduction qb d O 2qb rarefractions compressions Refer to Appendix

More information

Lab 1: The Digital Oscilloscope

Lab 1: The Digital Oscilloscope PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix

More information

Alignement of a ring cavity laser

Alignement of a ring cavity laser Alignement of a ring cavity laser 1 Introduction This manual describes a procedure to align the cavity of our Ti:Sapphire ring laser and its injection with an Argon-Ion pump laser beam. The setup is shown

More information

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1) Interferometers The true worth of an experimenter consists in his pursuing not only what he seeks in his experiment, but also what he did not seek. Claude Bernard (1813-1878) OBJECTIVES To examine the

More information

Interference. Physics 102 Workshop #3. General Instructions

Interference. Physics 102 Workshop #3. General Instructions Interference Physics 102 Workshop #3 Name: Lab Partner(s): Instructor: Time of Workshop: General Instructions Workshop exercises are to be carried out in groups of three. One report per group is due by

More information

AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified High Speed Fiber Photodetector. This user s guide will help answer any questions you may have regarding the

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,

More information

Introduction to Diode Lasers

Introduction to Diode Lasers Introduction to Diode Lasers Simon L. Cornish (Dated: August 2, 2007) The goal of this document is to guide you through a series of simple experiments and measurements that will help you become familiar

More information

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS Objective In this experiment you will study the i-v characteristics of an MOS transistor. You will use the MOSFET as a variable resistor and as a switch. BACKGROUND

More information

A Guide to Acousto-Optic Modulators

A Guide to Acousto-Optic Modulators A Guide to Acousto-Optic Modulators D. J. McCarron December 7, 2007 1 Introduction Acousto-optic modulators (AOMs) are useful devices which allow the frequency, intensity and direction of a laser beam

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 2011 Interference - 1 INTERFERENCE OF SOUND WAVES The objectives of this experiment are: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves. To observe interference phenomena

More information

Amplified High Speed Fiber Photodetectors

Amplified High Speed Fiber Photodetectors Amplified High Speed Fiber Photodetectors User Guide (800)697-6782 sales@eotech.com www.eotech.com Page 1 of 7 EOT AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified

More information

GAERTNER SCIENTIFIC CORPORATION 3650 Jarvis Ave. Skokie, Illinois 60076 U.S.A. tel: 1 847 673-5006 fax: 1 847 673-5009 email@gaertnerscientific.

GAERTNER SCIENTIFIC CORPORATION 3650 Jarvis Ave. Skokie, Illinois 60076 U.S.A. tel: 1 847 673-5006 fax: 1 847 673-5009 email@gaertnerscientific. 7109-C-244E-R1 Field Installation of HeNe Laser in B, C, and D-Type Auto Gain Ellipsometers GAERTNER SCIENTIFIC CORPORATION 3650 Jarvis Ave. Skokie, Illinois 60076 U.S.A. tel: 1 847 673-5006 fax: 1 847

More information

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity

More information

Lab E1: Introduction to Circuits

Lab E1: Introduction to Circuits E1.1 Lab E1: Introduction to Circuits The purpose of the this lab is to introduce you to some basic instrumentation used in electrical circuits. You will learn to use a DC power supply, a digital multimeter

More information

Reading assignment: All students should read the Appendix about using oscilloscopes.

Reading assignment: All students should read the Appendix about using oscilloscopes. 10. A ircuits* Objective: To learn how to analyze current and voltage relationships in alternating current (a.c.) circuits. You will use the method of phasors, or the vector addition of rotating vectors

More information

# 2. Selecting and Using Thermistors for Temperature Control

# 2. Selecting and Using Thermistors for Temperature Control # 2 Selecting and Using Thermistors for Temperature Control Selecting and Using Thermistors for Temperature Control Thermally sensitive resistors (thermistors) are used widely in laser diode and detector

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.

More information

Atomic Force Microscope

Atomic Force Microscope Atomic Force Microscope (Veeco Nanoman) User Manual Basic Operation 4 th Edition Aug 2012 NR System Startup If the system is currently ON To start the NanoScope software, double-click the NanoScope startup

More information

FIRERAY 2000 Installation Guide

FIRERAY 2000 Installation Guide FIRERAY 2000 Installation Guide Features Range 33ft to 330 ft. 24Vdc operation Selectable alarm thresholds Low current consumption Ground level electronics Manual or Automatic reset System Description

More information

Lab 3 - DC Circuits and Ohm s Law

Lab 3 - DC Circuits and Ohm s Law Lab 3 DC Circuits and Ohm s Law L3-1 Name Date Partners Lab 3 - DC Circuits and Ohm s Law OBJECTIES To learn to apply the concept of potential difference (voltage) to explain the action of a battery in

More information

Cross-beam scanning system to detect slim objects. 100 mm 3.937 in

Cross-beam scanning system to detect slim objects. 100 mm 3.937 in 891 Object Area Sensor General terms and conditions... F-17 Related Information Glossary of terms... P.1359~ Sensor selection guide...p.831~ General precautions... P.1405 PHOTO PHOTO Conforming to EMC

More information

ITA3D 710. RGB DMX laser with 3D effect. User manual

ITA3D 710. RGB DMX laser with 3D effect. User manual ITA3D 710 RGB DMX laser with 3D effect User manual 1 Safety precautions WARNING: This unit may cause serious injury to eyesight when used incorrect. It therefore is strongly advised to read this user manual

More information

22.302 Experiment 5. Strain Gage Measurements

22.302 Experiment 5. Strain Gage Measurements 22.302 Experiment 5 Strain Gage Measurements Introduction The design of components for many engineering systems is based on the application of theoretical models. The accuracy of these models can be verified

More information

Polarization of Light

Polarization of Light Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights

More information

HITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE

HITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE HITACHI INVERTER SJ/L1/3 SERIES PID CONTROL USERS GUIDE After reading this manual, keep it for future reference Hitachi America, Ltd. HAL1PID CONTENTS 1. OVERVIEW 3 2. PID CONTROL ON SJ1/L1 INVERTERS 3

More information

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 (revised 4/21/03) NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 Abstract This experiment studies the Nuclear Magnetic Resonance of protons

More information

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control

More information

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz 1 Sine wave with a DC offset f = frequency in Hz A = DC offset voltage (average voltage) B = Sine amplitude Vpp = 2B Vmax = A +

More information

ECEN 1400, Introduction to Analog and Digital Electronics

ECEN 1400, Introduction to Analog and Digital Electronics ECEN 1400, Introduction to Analog and Digital Electronics Lab 4: Power supply 1 INTRODUCTION This lab will span two lab periods. In this lab, you will create the power supply that transforms the AC wall

More information

Scaling and Biasing Analog Signals

Scaling and Biasing Analog Signals Scaling and Biasing Analog Signals November 2007 Introduction Scaling and biasing the range and offset of analog signals is a useful skill for working with a variety of electronics. Not only can it interface

More information

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b DIODE CIRCUITS LABORATORY A solid state diode consists of a junction of either dissimilar semiconductors (pn junction diode) or a metal and a semiconductor (Schottky barrier diode). Regardless of the type,

More information

Lab Exercise 1: Acoustic Waves

Lab Exercise 1: Acoustic Waves Lab Exercise 1: Acoustic Waves Contents 1-1 PRE-LAB ASSIGNMENT................. 2 1-3.1 Spreading Factor: Spherical Waves........ 2 1-3.2 Interference In 3-D................. 3 1-4 EQUIPMENT........................

More information

ILX Lightwave Corporation www.ilxlightwave.com

ILX Lightwave Corporation www.ilxlightwave.com # 14 Optimizing TEC Current Optimizing TEC Drive Current This Application Note discusses methods for optimizing the TEC drive current for optimum operation in heating and cooling applications. BACKGROUND

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2006 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

SpectraTec II. Polarized Multi-Laser Source BLUE SKY RESEARCH WAVELENGTHS. The SpectraTec II

SpectraTec II. Polarized Multi-Laser Source BLUE SKY RESEARCH WAVELENGTHS. The SpectraTec II BLUE SKY RESEARCH The SpectraTec II, two wavelength laser module is a highly integrated system comprised of two lasers, individual driving and temperature control electronics, wavelength combining, and

More information

Little LFO. Little LFO. User Manual. by Little IO Co.

Little LFO. Little LFO. User Manual. by Little IO Co. 1 Little LFO User Manual Little LFO by Little IO Co. 2 Contents Overview Oscillator Status Switch Status Light Oscillator Label Volume and Envelope Volume Envelope Attack (ATT) Decay (DEC) Sustain (SUS)

More information

APPLICATION NOTES: Dimming InGaN LED

APPLICATION NOTES: Dimming InGaN LED APPLICATION NOTES: Dimming InGaN LED Introduction: Indium gallium nitride (InGaN, In x Ga 1-x N) is a semiconductor material made of a mixture of gallium nitride (GaN) and indium nitride (InN). Indium

More information

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997

Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997 Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997 1 Purpose To measure and understand the common emitter transistor characteristic curves. To use the base current gain

More information

Fraunhofer Diffraction

Fraunhofer Diffraction Physics 334 Spring 1 Purpose Fraunhofer Diffraction The experiment will test the theory of Fraunhofer diffraction at a single slit by comparing a careful measurement of the angular dependence of intensity

More information

Figure 1: Multiple unsynchronized snapshots of the same sinusoidal signal.

Figure 1: Multiple unsynchronized snapshots of the same sinusoidal signal. 1 Oscilloscope Guide Introduction An oscilloscope is a device used to observe and measure time-dependent electronic signals. It is essentially an enhanced voltmeter which displays a graph of potential

More information

A More Efficient Way to De-shelve 137 Ba +

A More Efficient Way to De-shelve 137 Ba + A More Efficient Way to De-shelve 137 Ba + Abstract: Andrea Katz Trinity University UW REU 2010 In order to increase the efficiency and reliability of de-shelving barium ions, an infrared laser beam was

More information

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

WAVELENGTH OF LIGHT - DIFFRACTION GRATING PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant

More information

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law.

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law. 260 17-1 I. THEORY EXPERIMENT 17 QUALITATIVE STUDY OF INDUCED EMF Along the extended central axis of a bar magnet, the magnetic field vector B r, on the side nearer the North pole, points away from this

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2009 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab

EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab EXPERIMENT O-6 Michelson Interferometer Abstract A Michelson interferometer, constructed by the student, is used to measure the wavelength of He-Ne laser light and the index of refraction of a flat transparent

More information

PASSIVE INFRARED INTRUSION DETECTOR PASSIVE INFRAROOD DETECTOR DETECTEUR D INTRUSION PASSIF INFRAROUGE

PASSIVE INFRARED INTRUSION DETECTOR PASSIVE INFRAROOD DETECTOR DETECTEUR D INTRUSION PASSIF INFRAROUGE MODEL HAA51 PASSIVE INFRARED INTRUSION DETECTOR PASSIVE INFRAROOD DETECTOR DETECTEUR D INTRUSION PASSIF INFRAROUGE OPERATING MANUAL HAA51 PASSIVE INFRARED INTRUDER DETECTOR INTRODUCTION HAA51 Passive Infrared

More information

SENSORS. Miniature Sensors - S3Z. Advanced line of miniature Asian style of photoelectric sensors. 40-300 mm background suppression

SENSORS. Miniature Sensors - S3Z. Advanced line of miniature Asian style of photoelectric sensors. 40-300 mm background suppression Advanced line of miniature Asian style of photoelectric sensors 4-3 background suppression.7 m proximity, 15 with narrow beam 4 m polarized retroreflective 3 m LASER through beam Standard 3-wire output

More information

DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES)

DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES) Features: DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES) PATENT NUMBERS: CANADA 2,494,133, USA 7095931, 7295731 AND CHINA 1672073 Telcordia GR-468 qualified Available in versions for any wavelength

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions

Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions Description The Flame Integrated Sampling System (FLAME-DA-CUV-UV-VIS) is a snap-on, direct-attach 1 cm cuvette

More information

More colors, more power, more applications... Solid state laser Diode laser Line laser Multicolor laser Laser diode driver Cooling modules

More colors, more power, more applications... Solid state laser Diode laser Line laser Multicolor laser Laser diode driver Cooling modules PEGASUS Linienlaser Diodenlaser Festkörperlaser PEGASUS LASERSYSTEME GMBH More colors, more power, more applications... Solid state laser Diode laser Line laser Multicolor laser Laser diode driver Cooling

More information

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment. Helium-Neon Laser Experiment objectives: assemble and align a 3-mW HeNe laser from readily available optical components, record photographically the transverse mode structure of the laser output beam,

More information

Multipurpose Analog PID Controller

Multipurpose Analog PID Controller Multipurpose Analog PID Controller Todd P. Meyrath Atom Optics Laboratory Center for Nonlinear Dynamics University of Texas at Austin c 00 March 4, 00 revised December 0, 00 See disclaimer This analog

More information

OPERATION MANUAL VALVE CHECKER G040-123

OPERATION MANUAL VALVE CHECKER G040-123 OPERATION MANUAL VALVE CHECKER AUSTRALIA PTY. LTD. L:\DWG_NO\DOCUMENT\MANUALS\Word File\A OPERATION MANUAL CN 20.4.99 VALVE CHECKER 1 of 20 CONTENTS Chapter Title Page 1. Description 3 2. Specification

More information

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz AN1200.04 Application Note: FCC Regulations for ISM Band Devices: Copyright Semtech 2006 1 of 15 www.semtech.com 1 Table of Contents 1 Table of Contents...2 1.1 Index of Figures...2 1.2 Index of Tables...2

More information

Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015

Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015 Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) The standard tuning values used in ADVANCED Motion Controls drives are conservative and work well in over 90%

More information

Diffraction and Young s Single Slit Experiment

Diffraction and Young s Single Slit Experiment Diffraction and Young s Single Slit Experiment Developers AB Overby Objectives Preparation Background The objectives of this experiment are to observe Fraunhofer, or far-field, diffraction through a single

More information

Physics 441/2: Transmission Electron Microscope

Physics 441/2: Transmission Electron Microscope Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This

More information

Dash 18X / Dash 18 Data Acquisition Recorder

Dash 18X / Dash 18 Data Acquisition Recorder 75 Dash 18X / Dash 18 Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 3.1 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH

More information

PCM Encoding and Decoding:

PCM Encoding and Decoding: PCM Encoding and Decoding: Aim: Introduction to PCM encoding and decoding. Introduction: PCM Encoding: The input to the PCM ENCODER module is an analog message. This must be constrained to a defined bandwidth

More information

4.3.5: High Temperature Test 3

4.3.5: High Temperature Test 3 temperature and 800 degrees Celsius is made by matching the optical path lengths of the measurement and sensing arms at both temperatures. By marking the relative distance between the GRIN lens and mirror

More information

1. Learn about the 555 timer integrated circuit and applications 2. Apply the 555 timer to build an infrared (IR) transmitter and receiver

1. Learn about the 555 timer integrated circuit and applications 2. Apply the 555 timer to build an infrared (IR) transmitter and receiver Electronics Exercise 2: The 555 Timer and its Applications Mechatronics Instructional Laboratory Woodruff School of Mechanical Engineering Georgia Institute of Technology Lab Director: I. Charles Ume,

More information

How To Use A Power Supply On A Powerline 2.2 (Ai)

How To Use A Power Supply On A Powerline 2.2 (Ai) KNX/EIB Product documentation Issue: 05.08.2010 629x1220 Push button sensor 3 comfort 1-gang Push button sensor 3 comfort 2-gang (1+1) Push button sensor 3 comfort 3-gang Push button sensor 3 comfort 4-gang

More information

TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER

TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER (PWM Input, 1.2A or 2A Output, Metal Box or PCB) PCB Board - P/N: PWMC-PCB-2A, PWMC-PCB-1.2A Packaged Driver (metal box with 1.5 m (5 ft.) cable)

More information

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL FIBRE-OPTICS POWER METER INSTRUCTION MANUAL FIBRE-OPTICS POWER METER ELLMAX ELECTRONICS LTD. Unit 29, Leyton Business Centre, Etloe Road, Leyton, London E10 7BT, England Tel: 020-8539 0136 Fax: 020-8539

More information

Multi-Range Programmable DC Power Supplies 9115 Series

Multi-Range Programmable DC Power Supplies 9115 Series Data Sheet Multi-Range Programmable DC Power Supplies 1200 W / 3000 W Multi-Range DC Power Supplies Features & Benefits Any model can replace several supplies on your bench or in your rack. Unlike conventional

More information

Germanium Diode AM Radio

Germanium Diode AM Radio Germanium Diode AM Radio LAB 3 3.1 Introduction In this laboratory exercise you will build a germanium diode based AM (Medium Wave) radio. Earliest radios used simple diode detector circuits. The diodes

More information

Laserlyte-Flex Alignment System

Laserlyte-Flex Alignment System Laserlyte-Flex Alignment System LaserLyte-Flex The LaserLyte-Flex Alignment System is a unique, interchangeable, low cost plug and play laser system. Designed specifically for aligning and positioning

More information

Joule Thief 3.0 Kit. June 2012, Rev 1 1 http://www.easternvoltageresearch.com Joule Thief 3.0

Joule Thief 3.0 Kit. June 2012, Rev 1 1 http://www.easternvoltageresearch.com Joule Thief 3.0 Kit Instruction Manual Eastern Voltage Research, LLC June 2012, Rev 1 1 http://www.easternvoltageresearch.com HIGH BRIGHTNESS LED THIS KIT USES A 1W CREE, HIGH BRIGHTNESS LED. DO NOT STARE AT THIS (OR

More information

Projects. Objective To gain hands-on design and measurement experience with real-world applications. Contents

Projects. Objective To gain hands-on design and measurement experience with real-world applications. Contents Projects Contents 9-1 INTRODUCTION...................... 43 9-2 PROJECTS......................... 43 9-2.1 Alarm Radar Sensor................ 43 9-2.2 Microwave FM Communication Link....... 46 9-2.3 Optical

More information

OPTICAL FIBERS INTRODUCTION

OPTICAL FIBERS INTRODUCTION OPTICAL FIBERS References: J. Hecht: Understanding Fiber Optics, Ch. 1-3, Prentice Hall N.J. 1999 D. R. Goff: Fiber Optic Reference Guide (2 nd ed.) Focal Press 1999 Projects in Fiber Optics (Applications

More information

Agilent AEDB-9140 Series Three Channel Optical Incremental Encoder Modules with Codewheel, 100 CPR to 500 CPR Data Sheet

Agilent AEDB-9140 Series Three Channel Optical Incremental Encoder Modules with Codewheel, 100 CPR to 500 CPR Data Sheet Agilent AEDB-9140 Series Three Channel Optical Incremental Encoder Modules with Codewheel, 100 CPR to 500 CPR Data Sheet Description The AEDB-9140 series are three channel optical incremental encoder modules

More information

Measuring Laser Power and Energy Output

Measuring Laser Power and Energy Output Measuring Laser Power and Energy Output Introduction The most fundamental method of checking the performance of a laser is to measure its power or energy output. Laser output directly affects a laser s

More information

Example of SOP for IR Laser

Example of SOP for IR Laser Example of SOP for IR Laser Standard Operating Procedures (SOP) for Class 4 HPD semiconductor Laser Manufacturer HPD Building & Room Number EH&S 101 Model Number HPD-1/100/975 Laser Class 4 Serial Number

More information

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction

More information

INFRARED PARTS MANUAL

INFRARED PARTS MANUAL INFRARED PARTS MANUAL PIR325 FL65 GLOLAB CORPORATION Thank you for buying our Pyroelectric Infrared components. The goal of Glolab is to produce top quality electronic kits, products and components. All

More information

The Sonometer The Resonant String and Timbre Change after plucking

The Sonometer The Resonant String and Timbre Change after plucking The Sonometer The Resonant String and Timbre Change after plucking EQUIPMENT Pasco sonometers (pick up 5 from teaching lab) and 5 kits to go with them BK Precision function generators and Tenma oscilloscopes

More information

Designing Fiber Optic Systems David Strachan

Designing Fiber Optic Systems David Strachan Designing Fiber Optic Systems David Strachan Everyone knows that fiber optics can carry a huge amount of data. There are more benefits to using fiber optics in broadcast applications than you might realize.

More information

Blackbody Radiation References INTRODUCTION

Blackbody Radiation References INTRODUCTION Blackbody Radiation References 1) R.A. Serway, R.J. Beichner: Physics for Scientists and Engineers with Modern Physics, 5 th Edition, Vol. 2, Ch.40, Saunders College Publishing (A Division of Harcourt

More information

GRID AND PRISM SPECTROMETERS

GRID AND PRISM SPECTROMETERS FYSA230/2 GRID AND PRISM SPECTROMETERS 1. Introduction Electromagnetic radiation (e.g. visible light) experiences reflection, refraction, interference and diffraction phenomena when entering and passing

More information

Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006

Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006 Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006 1 Purpose To measure and understand the common emitter transistor characteristic curves. To use the base current gain

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

Acousto-optic modulator

Acousto-optic modulator 1 of 3 Acousto-optic modulator F An acousto-optic modulator (AOM), also called a Bragg cell, uses the acousto-optic effect to diffract and shift the frequency of light using sound waves (usually at radio-frequency).

More information

Alignment Laser System.

Alignment Laser System. - O T 6 0 0 0 Alignment Laser System. The OT-6000. Multi -Target,Two Dimensional Alignment. Introducing the most powerful way to measure alignment at distances up to 300 feet. The OT-6000 Alignment Laser

More information

Self-leveling Rotary Laser. Operating Manual. surface. Emitting direction of rotary laser-beam illustrated as follows:

Self-leveling Rotary Laser. Operating Manual. surface. Emitting direction of rotary laser-beam illustrated as follows: Self-leveling Rotary Laser Operating Manual 1. Functions This instrument is equipped with the semiconductor diode with wavelength of 635nm(green:532nm), which the laser beam has supreme visibility. And

More information

User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR

User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR Content General safety summary...1 Introduction...2 Chapter 1 Getting started...3 System Requirements...4 Installing Hardware...5 Installing Software...8

More information

Application Report: Running µshape TM on a VF-20 Interferometer

Application Report: Running µshape TM on a VF-20 Interferometer : Running µshape TM on a VF-20 Interferometer General This report describes how a fiber interferometer from Arden Photonics Ltd was used together with the µshape TM Generic software package. The VF-20

More information

TECHNICAL DATASHEET #TD1000AX DUAL SOLENOID MULTIFUNCTION VALVE DRIVER With Joystick, Potentiometer and Control System Interface P/N: DSDA-SMB Series

TECHNICAL DATASHEET #TD1000AX DUAL SOLENOID MULTIFUNCTION VALVE DRIVER With Joystick, Potentiometer and Control System Interface P/N: DSDA-SMB Series TECHNICAL DATASHEET #TD1000AX DUAL SOLENOID MULTIFUNCTION VALVE DRIVER With Joystick, Potentiometer and Control System Interface P/N: DSDA-SMB Series Description: The Dual Solenoid Multifunction Valve

More information

INTRODUCTION. Please read this manual carefully for a through explanation of the Decimator ProRackG and its functions.

INTRODUCTION. Please read this manual carefully for a through explanation of the Decimator ProRackG and its functions. INTRODUCTION The Decimator ProRackG guitar noise reduction system defines a new standard for excellence in real time noise reduction performance. The Decimator ProRackG was designed to provide the maximum

More information

LDG M-7700 External Meter for Icom IC-7700

LDG M-7700 External Meter for Icom IC-7700 M-7700 OPERATIONS MANUAL MANUAL REV A LDG M-7700 External Meter for Icom IC-7700 LDG Electronics 1445 Parran Road St. Leonard MD 20685 Phone: 410-586-2177 Fax: 410-586-8475 ldg@ldgelectronics.com www.ldgelectronics.com

More information

Servo Info and Centering

Servo Info and Centering Info and Centering A servo is a mechanical motorized device that can be instructed to move the output shaft attached to a servo wheel or arm to a specified position. Inside the servo box is a DC motor

More information