MODEL IQ240 ph Meter Getting Started TWO-POINT CALIBRATION--QUICK REFERENCE GUIDE

Size: px
Start display at page:

Download "MODEL IQ240 ph Meter Getting Started TWO-POINT CALIBRATION--QUICK REFERENCE GUIDE 1. 2. 3. 4."

Transcription

1 1

2

3 MODEL IQ240 ph Meter Getting Started TWO-POINT CALIBRATION--QUICK REFERENCE GUIDE Connect probe to meter. Put probe in first calibrating buffer. Press ON/OFF to turn on meter If necessary, press ph until ph status indicator arrow points to ph. Status indicator arrow should NOT point to BUFFER. Press CAL, then CLEAR to clear the previous calibration. Status indicator arrow will point to 1PT. Meter will automatically identify the first buffer solution. If necessary, press SELECT BUFFER until display matches first buffer value. Press ENTER and wait until the large number display stops flashing. Rinse probe in deionized water and place in second calibrating buffer Press CAL. Status indicator arrow will point to both 1PT and 2PT. Meter will automatically identify the second buffer solution. If necessary, press SELECT BUFFER until display matches second buffer value. Press ENTER and wait until the large number display stops flashing. Rinse probe in deionized water and place in sample. 1

4 INTRODUCTION IMPORTANT NOTICE IF THE METER IS USED IN A MANNER OTHER THAN AS DESCRIBED, THE SAFETY AND PERFORMANCE OF THE METER CAN BE IMPAIRED. This system uses an Ion Sensitive Field Effect Transistor (ISFET) ph sensor. There is no glass to break and the probe should be stored dry. No electrode storage solution is required. The 3-in-1 probe contains the ph sensor, reference electrode, and temperature sensor. ESTIMATED LIFE OF AN ISFET SENSOR ph PROBE Depending upon the application, the expected life of an ISFET probe is approximately 18 months. The reference electrode contains a KCl gel that becomes increasingly diluted with usage and time. The reference is sealed and is not refillable. When it becomes difficult to calibrate (ph buffer value does not stop flashing during calibration or takes an excessively long time to stop flashing) the probe should be replaced. A probe that is dirty, miscalibrated, or has reached the end of its useful life will give an error message E11 on the meter display. Your ISFET probe is warranted to be free from manufacturing defects for six (6) months from the date of purchase. 2

5 Table of Contents Getting Started--Quick Reference Guide for Two Point Calibration Techniques and Tips For Best Performance Display Features and Meter Connections Main Display Meter Connections Status Indicators Keypad Functions Cleaning and Reconditioning the Probe ph Calibration and Measurement Important Notes for Best Calibration Results Calibration With Two Buffers Calibration With Three Buffers Calibration With One Buffer How to Abort Calibration Automatic Buffer Recognition Automatic Temperature Compensation of Automatic Buffer Recognition Values Changing Stored Automatic Buffer Recognition Values Restoring Automatic Buffer Recognition Values to Factory Default. 18 mv Measurement Titrations Storing and Recalling Data Storing Sample Readings Recalling Sample Readings Clearing Data from One Storage Address Clearing Data from All Storage Addresses Output Modes Analog Output Serial Output Output of Calibration Data to the Serial Port Output of Current Sample to the Serial Port Output of the Contents of a Single Memory Address Output of the Contents of all Memory Addresses Setting the Baud Rate Temperature Calibration User Temperature Calibration Restoring Temperature Calibration to Factory Default Battery Replacement Troubleshooting Error Codes

6 Techniques and Tips for Best Performance DO DO: Soak the probe in ph 7.00 buffer for at least five minutes if the probe is new or has been stored dry for an extended period. This will hydrate the sensor and dissolve any reference material, which may have crystallized on the sensor surface. DO: Clean the probe regularly with a toothbrush and mild detergent. DO: STORE THE PROBE DRY with the black protective cap covering the probe tip. No electrode storage solution is required. DO: Always begin each measuring session with a two-point calibration. Update often with one-point or two-point calibrations. DO: Use fresh buffers and deionized rinse solution. DO: Use buffer solutions with ph values no greater than three ph units apart. Buffers should bracket the anticipated values of the samples to be measured. DO: Choose buffer solutions that most closely mimic the chemical nature and conductivity level of the sample solutions to be measured. DO: Use deionized water to rinse residual buffer and sample solutions from the probe after calibration and measurement. DO: Calibrate at the same temperature as the sample solution. Although the meter has automatic temperature compensation, best results will be achieved if the calibration buffers and sample are the same temperature. DO: Keep the connectors clean and dry. Dirty or damp connectors can cause unstable readings. DO: Be sure the surface of the sensor in the probe is free from any deposits or films. See the cleaning instructions in this manual. DO: Contact IQ Scientific Instruments if your ph testing involves monitoring small ph changes over time (i.e. titrations). 4

7 DO NOT DO NOT: change probes while the meter is turned on. Always turn off the meter before changing probes. DO NOT: store the probe in solution or use for long-term ph measuring applications. Extended immersion will shorten the life of the probe. DO NOT: leave probe uncapped for long periods of time. Always place the protective cap over the probe tip when finished measuring. DO NOT: allow oil, fat, food particles, starch, protein, or other materials to remain on the probe tip after use. DO NOT: use a sharp metal object (needle, pin, etc.) to clean the sensor surface. DO NOT: take readings in direct sunlight. Direct sunlight may cause unstable readings or difficulty in calibration. Call IQ Scientific Instruments or your Authorized Dealer for optional light shields. DO NOT: use the probe in an environment that will damage the ph sensor, such as hydrofluoric acid or abrasive samples. DO NOT: use the probe in environments that will damage the epoxy materials used in probe construction, such as, acetone, toluene, methylene chloride, xylene, and other strong organic solvents. DO NOT: use the probe in environments with static electricity. Electrostatic discharge may permanently damage the probe. DO NOT: expose the ISFET probe to repeated temperature extremes (60 o C or higher). Do not use the probe for applications that require the probe to cycle between hot and room temperature samples. DO NOT: put any other active measurement devices in buffers or samples while measuring ph. Even if another device is not actually turned on, AC power interference can still occur, especially if the ph meter AC battery adapter is in use. 5

8 Display Features and Meter Connections 1. Main Display 2. Temperature Display. 3. AUTO HOLD Mode Indicator 4. Status Indicator Arrows 5. Serial Interface (RS-232) 6. AC Adapter Connector 7. Recorder Output 8. Probe Connector Main Display The large digits show the following: 1. Readout of the measured value, either ph or millivolt. 2. Memory address display AXX indicates memory location, where XX is a number from 01 to 20. For example: A03 indicates memory address 03 is in use. 3. Error Codes EXX where XX = a number. For example: E11 indicates a calibration error. See the troubleshooting section of this manual for details on error codes. Temperature Display The small digits display temperature in o C. The display is accurate only when the probe is in solution. The temperature sensor can be calibrated by the user (See Page 24). AUTO HOLD (eye icon) Indicator This mode allows a reading to remain automatically locked on the display for future reference. The eye symbol indicates that the meter is in the automatic hold AUTO HOLD mode. The eye symbol starts flashing when the meter begins to take a reading and stops flashing (locks) when the reading has stabilized. The reading remains locked until the ph key is pressed to unlock the reading or the HOLD key is pressed to exit the AUTO HOLD mode. 6

9 Meter Connections RS-232 Standard 9 pin connector for serial output to printer or computer. AC Power Adapter to attach the provided AC adapter. Analog Recorder Output for a chart recorder. Probe Connector Jack for connection of ph probe to meter. ph mv REL mv BUFFER Indicates that the meter is in the ph mode. Flashing ph arrow indicates that the meter is still seeking a stable reading. Stable ph arrow indicates that a stable reading endpoint has been reached. Indicates that the meter is in the Millivolt mode. Indicates that the meter is in the Relative Millivolt mode. Indicates that the meter is in the Select Buffer mode. 1 PT Indicates that one buffer value has been used for calibration. If flashing, first buffer has been selected but not read. 7

10 2 PT Indicates that two buffer values have been used for calibration. If flashing, second buffer has been selected but not read. 3 PT Indicates that three buffer values have been used for calibration. If flashing, third buffer has been selected but not read. RECALL STORE PROBE Indicates that the meter is in Memory Recall mode. Indicates that the meter is in Memory Storage mode. Indicates damaged probe and/or calibration problem. Indicates meter is on battery power. Flashing indicates battery should be replaced. 8

11 Keypad Functions ON/OFF CLEAR ph Turns meter ON and OFF. When turned off, the meter retains the calibration in memory. If the meter is on battery power and no key is pressed for 10 hours, the meter will turn off automatically. If the meter is operating on battery power and an error message is displayed for 10 minutes, the meter will turn off automatically. If the meter is operating on AC power, it will not turn off automatically. Used to clear the meter of the current value. This key has various uses, as described in these instructions. Selects ph mode..1/.01/.001 Selects the resolution of the ph reading. mv Selects the mv mode. When pressed successively, toggles between mv and the Relative mv mode. 9

12 SELECT CAL HOLD STORE Selects the Select Buffer mode. Repeatedly pressing the SELECT BUFFER key scrolls through Automatic Buffer Recognition values. Selects the Calibration mode. Selects the Auto Hold mode. When Auto Hold mode is ON, the eye symbol is displayed. When a stable reading is obtained, the display remains locked until the ph key is pressed or the HOLD key is pressed to turn off the Auto Hold function Selects the Memory Storage mode. Pressing a second time selects the Memory Recall mode. fl Used to increment through memory addresses or manually set buffer values. ENTER Various functions depending upon the mode. Generally used to store information (calibrations and sample readings) and to send data to the serial port. 10

13 Cleaning and Reconditioning ISFET Probes Cleaning With Detergent Regularly clean the ISFET probe with detergent (a few drops of dish detergent in a cup of warm water). Rinse well with deionized water. ISFET probes can be cleaned with a toothbrush and detergent Reconditioning with Warm Buffer ISFET probes are designed to be stored dry and have a virtually unlimited shelf life; however extended period of dry storage can crystallize the KCl gel at the reference junction. Heat ph 7.00 buffer to approximately 45 o C and 60 C (115 o F and 140 F). Soak the probe for 2 minutes. Place the probe in room temperature ph 7.00 buffer and allow to cool. CAUTION: Do not use the probe as a thermometer to determine if the buffer is above 60 o C. Sudden immersion in excessively hot water may permanently damage the probe. 11

14 Cleaning With Alcohol or Methanol The ISFET sensor (recessed shiny blue-green spot at the probe tip) can be gently cleaned using alcohol or methanol and a wooden toothpick tipped with a small amount of cotton wool. Be sure that the cotton tip is small enough to actually make contact with the recessed sensor. CAUTION: Cleaning with alcohol is not recommended if the sample you are testing contains proteins. Use Sodium Hypochlorite (see below). CAUTION: Do not use metal objects such as pins or paperclips to clean the sensor. The sensor can be permanently damaged with aggressive abrasion of the sensor surface. Cleaning with Sodium Hypochlorite (Bleach) If the ISFET probe has been exposed to fat or protein, the response time may begin to slow. Soak the probe for five minutes in a 10:1 dilution of deionized water to household bleach (0.5% solution of sodium hypochlorite). Brush with a soft bristle brush or toothbrush to remove deposits. The sensor is slightly recessed, be sure to make gentle contact with the sensor surface. Rinse well with deionized water. CAUTION: Do not soak the probe in sodium hypochlorite for more than 5 minutes. Extended immersion will permanently damage the probe. 12

15 ph Calibration and Measurement Range The IQ240 can measure from ph 0 to 14. Temperature compensation is automatic from -5 o C to + 80 o C. The meter can perform a one, two or three point calibration using any buffer listed below. Calibration Buffer Options 1.68, 4.00, 6.86, 7.00, 7.41, 9.18, 10.01, and It is recommended to avoid temperatures above 60 o C (140 o F) when possible. Elevated temperatures shorten the life of ph probes. Important Notes for Best Calibration Results Always clear the previous calibration before beginning the calibration procedure (See Step 6 of Calibration with Two Buffers on Page 14). Always begin each measuring session with a two-point calibration. Calibrate no greater than 3 ph units apart. If probe is new or has not been in regular use, soaking the probe in ph 7.00 buffer for at least five minutes is recommended. For best accuracy, when working at non-ambient temperatures, allow the probe to equilibrate to the temperature of the buffer or sample solution before starting calibration or ph measurement. Erroneous results will occur if the meter is incorrectly calibrated; for example, setting the Buffer Value to ph 7.00 when the probe is in ph 4.00 buffer. If unusual behavior occurs, recheck the buffer values, clear the previous calibration, and recalibrate the meter. If calibration times are excessive (over 2 minutes) follow the cleaning and reconditioning instructions on Page 11. Optimizing for Accuracy or Response Time The normal operating resolution is 0.01 ph. You can change resolution by pressing the.1/.01/.001 key to optimize for accuracy or response time. Use ph resolution for maximum accuracy. Use 0.1 ph resolution to optimize for maximum speed; the response time in 0.1 resolution will be approximately three times the speed of resolution. 13

16 Calibration with Two Buffers Remember: Always use buffers that bracket the sample ph range. (For example, if the sample is about 8.5, use 7.00 and buffers). The narrowest calibration range gives the most accurate reading. 1. Remove the black protective cap from the probe tip and connect the probe to the meter. A small amount of clear or blue gel may be visible at the probe tip. 2. If the probe has not been in regular use, hydrate the probe for five minutes in ph 7.00 buffer. 3. Place the probe in the first buffer solution. Stir briefly to remove any bubbles from the probe surface. 4. Press ON/OFF to turn on the meter. 5. The ph status indicator arrow (top left corner of the LCD display) should be pointing to ph. If not, press the ph button until the status indicator points to ph. The BUFFER status indicator should NOT be displayed. 6. Press CAL then CLEAR to clear previous calibration. 7. The meter will automatically identify the ph of the first buffer solution. If the display does not match the ph of the buffer, press SELECT BUFFER until the display matches the first buffer solution ph. The 1 PT status indicator arrow will be flashing. (If using a buffer other than those automatically recognized by the IQ240, refer to Page 17, Changing Stored Automatic Buffer Recognition Values). 8. Press ENTER. The buffer ph value will begin flashing to indicate calibration is underway. Wait until the buffer ph value and 1 PT status indicator arrow stop flashing. NOTE: When calibration is complete, the ph reading displayed will be the automatically temperature compensated (ATC) ph value of the buffer (e.g. ph 7.00 buffer will be corrected to 7.01 if the buffer is at 20 o C, ph buffer will be corrected to at 20 o C). 14

17 9. Rinse the probe in deionized water and place in the second buffer. Stir briefly to remove any bubbles from the sensor surface. 10. Press CAL. 11. The meter will automatically identify the ph of the second buffer solution. If the display does not match the ph of the buffer, press SELECT BUFFER until display matches the second buffer solution ph. The 2 PT status indicator arrow will be flashing. 12. Press ENTER. The ph buffer value will begin flashing to indicate calibration is underway. Wait until the ph buffer value and 2 PT status indicator arrow stop flashing. 13. When the ph buffer value stops flashing, two point calibration is complete. 14. Rinse the probe in deionized water and place in sample. Read sample ph. Calibration with Three Buffers Three point calibration is the most accurate method of ph measurement, especially if the samples cover a wider range. Use buffers that bracket the sample ph range. The narrowest range gives the most accurate readings. 1. Follow steps 1 through 13 of the Calibration with Two Buffers procedure. 2. Rinse the probe in deionized water and place in the third buffer. Stir briefly to remove any bubbles from the sensor surface. 3. Press CAL. 4. The meter will automatically identify the ph of the third buffer solution. If the display does not match the ph of the buffer, press SELECT BUFFER until display matches the third buffer solution ph. The 3 PT status indicator arrow will be flashing. 5. Press ENTER. The ph buffer value will begin flashing to indicate calibration is underway. Wait until the ph buffer value and 3 PT status indicator arrow stop flashing. 6. When the ph buffer value stops flashing, three point calibration is complete. 7. Rinse probe in deionized water and place in sample. Read sample ph. 15

18 Calibration with One Buffer One point calibration is recommended only if the sample ph is very close to the calibration buffer ph and an accuracy of ±0.1 ph is acceptable. 1. Follow steps 1 through 8 of the Calibration With Two Buffers procedure. 2. Rinse the probe in deionized water and place in sample. Read sample ph. How to Abort Calibration If calibration times are excessive (ph buffer value continues to flash) or you wish to stop during calibration for any other reason, press CLEAR then ph. The meter will enter the ph reading mode. If no previous calibration is in memory, an error code (E10 or E11) may result. Please refer to the Troubleshooting Tips on Page 26 and the Cleaning and Reconditioning Instructions on Page 11 if trouble is experienced in calibration. Automatic Buffer Recognition Auto Buffer Recognition can discriminate between buffers that differ by more than 0.5 ph units. However, closely lying buffers (such as 6.86, 7.00 and 7.41) may not be selected correctly. The IQ240 has storage locations for nine buffer values. The factory default values are 1.68, 4.00, 6.86, 7.00, 7.41, 9.18, 10.01, 11.88, If the Automatic Buffer Recognition selects the wrong buffer during calibration, press SELECT BUFFER to select the correct buffer value. Then press ENTER to start the calibration. The user can replace any or all of the factory default values with custom buffer values in increments of 0.01 ph. (See Page 17). 16

19 Automatic Temperature Compensation of Automatic Buffer Recognition Values Factory default Automatic Buffer Recognition values have temperature correction data stored in the meter. The meter displays the nominal buffer value at 25 o C. When the calibration is complete, the meter performs the automatic temperature compensation for the actual temperature of the buffer being used (e.g. ph 7.00 buffer will be corrected to 7.01 if the buffer is at 20 o C, ph buffer will be corrected to at 20 o C). The meter does not perform temperature compensation on buffer values input by the user since temperature correction tables are not available. When user-entered buffer values are displayed, the temperature display shows a flashing temperature. The table below shows the ph values of buffers at various temperatures. Nominal Value 25 o C 0 o C 5 o C 10 o C 20 o C 30 o C 40 o C 50 o C Changing Stored Automatic Buffer Recognition Values To change the factory default buffer values: 1. Press SELECT BUFFER to scroll through the factory default values available. 2. To change any value, press the or fl keys. 3. Press ENTER to store this new value. NOTE: If a buffer value already exists in memory the error code dup (duplicate) will be displayed. To clear the error, press the ph key. REMEMBER: User defined buffer values are not temperature compensated. 17

20 Restoring Factory Default Automatic Buffer Recognition Values. 1. Press SELECT BUFFER repeatedly until the display shows the buffer value to be restored to the factory default value. 2. Press CLEAR to clear the displayed value. Clr is displayed. 3. Press ENTER to restore the factory default value in memory. mv Measurement Absolute mv Measurement 1. Connect the probe to the meter. 2. Press ON/OFF to turn on the meter. 3. Press mv to place the meter in the mv mode. Be sure the status indicator is NOT pointing to REL mv. If necessary, press the mv key until the status indicator arrow points to mv. The display will show the current mv reading. NOTE: Repeatedly pressing mv toggles between the mv and Relative mv modes. The status indicator arrow indicates the current reading mode. When in the AUTO HOLD mode, a single press of the mv key starts a new reading. Relative mv Measurement During calibration in the mv mode, the meter takes a reading on a solution and sets its reading to mv. This solution becomes the standard. The calibration data is stored, so that all subsequent readings taken in the Relative mv mode are relative to the initial standard. 1. Connect the probe to the meter. 2. Press ON/OFF to turn on the meter. 3. Press mv to place the meter in the mv mode. Be sure the status indicator is NOT pointing to REL mv. If necessary, press the mv key until the status indicator arrow points to mv. The display will show the current mv reading. 18

21 4. Place the probe in solution to be used as the standard. Stir briefly to remove any air bubbles. The mv reading of the solution will be displayed. 5. Press CAL. The mv value of the solution will be displayed and the flashing status indicator arrow will point to 1 PT. 6. Press ENTER. The mv value will flash until calibration is complete. Upon completion, the mv value will be set to and the mode will switch to Relative mv. 7. Rinse the probe in deionized water and place in sample. The mv value displayed will be Relative mv to the standard. NOTE: The HOLD function is active in the mv and Relative mv mode. Press the mv key to release the display from a locked reading. Titrations Should you wish to use the IQ240 ph meter in a titration application, please contact IQ Scientific Instruments for instructions. 19

22 Storing and Recalling Data The IQ240 can store and recall 20 sample readings (temperature and ph, mv or Relative mv) in memory for later recall. The memory storage addresses are labeled A01 to A20. Storing Sample Readings 1. Follow the instructions to display the desired ph or mv readings. 2. Press STORE/RECALL to put the meter in Store mode. The status indicator arrow points to STORE. The next available memory storage address is displayed (AXX). 3. To access a different memory storage address, press the or key to scroll through the storage locations. 4. When the desired storage address is displayed, press ENTER to store the reading and temperature. As soon as the reading is stored, the meter returns to the reading mode. Recalling Sample Readings. 1. Press STORE/RECALL to place the meter in the Store mode. The status indicator arrow points to STORE and the next available storage address is displayed. 2. Press STORE/RECALL again to place the meter in the Recall mode. The status indicator arrow points to RECALL. 3. Press the or fl keys to display the desired memory storage address. 4. Press ENTER to display the contents of the displayed address. If the storage address is empty, Clr is displayed. 20

23 Clearing Data from One Storage Address 1. Press STORE/RECALL to place the meter in the Store mode. The status indicator arrow points to STORE and the next available storage address is displayed. 2. Press STORE/RECALL again to place the meter in the Recall mode. The status indicator arrow points to RECALL. 3. Press the or fl keys to display the desired memory storage address. 4. Press CLEAR. The display reads Clr. 5. Press ENTER to clear the storage address. Clearing Data from All Storage Addresses. 1. Press STORE/RECALL to place the meter in the Store mode. The status indicator arrow points to STORE and the next available storage address is displayed. 2. Press STORE/RECALL again to place the meter in the Recall mode. The status indicator arrow points to RECALL. 3. Press and hold down the CLEAR key and press the ENTER key. The display reads ClrAll. All memory storage addresses are cleared. 21

24 Output Modes Analog Output An analog signal is sent to the recorder output continuously whenever the meter is in a reading mode. The output is 0.00 mv at ph 7.00 with a slope of mv/ph over the range of 0 to 14 ph. The analog output is temperature compensated for the temperature of the sensor. To connect the analog recorder to the meter, a cable with a miniature phono jack is required. Serial Output To connect to the serial port on the meter, use a NULL MODEM cable less than 3 meters long, with a 9-pin DB-9 connector. Data are sent to the serial port using the following criteria: There are a maximum of 20 characters per line Each line ends with a carriage return and a line feed. When the transmission is complete a second line feed is sent. The baud rate is user selectable to 300, 1200 or The meter sends 8 bit data, no parity, and one stop bit. NOTE: The use of the serial port can quickly drain the battery. It is recommended that the AC adapter be used whenever the meter is connected to a serial device. To limit the possibility of unintentional battery depletion when the meter is operating on battery power, the serial port is only active (momentarily) when the ENTER key is pressed. The serial port is always active when the AC power adapter is used. Some serial devices require that the serial port be continuously active. For these devices, the momentary feature can be overridden by simultaneously pressing the key and the CLEAR keys. The override is indicated by a flashing battery symbol (similar to a low battery condition). The override is canceled when the meter is turned off. Only use this override if problems with serial communications (such as garbled data) are experienced. Since a mismatch of serial port settings can also cause garbled data transfer, make sure that the meter and the external device are set to the same baud rate; along with 8 data bits, no parity, and 1 stop bit. Recheck the setting before using the override. 22

25 Output the Calibration Data to the Serial Port 1. The calibration data will be automatically sent to the serial port at the end of the calibration process when the ph buffer value stops flashing. Output the Current Sample Reading to the Serial Port 1. With the reading displayed, press ENTER. Any error message(s) related to the reading will also be printed following the reading. Output the Contents of a Single Memory Address to the Serial Port 1. Display the desired memory storage address using the directions on Page Press ENTER to display the contents of the storage address and simultaneously send the data to the serial port. Output the Contents of All Memory Addresses to the Serial Port 1. Press STORE/RECALL to place the meter in the Store mode. The status indicator arrow points to STORE and the next available storage address is displayed. 2. Press STORE/RECALL again to place the meter in the Recall mode. The status indicator arrow points to RECALL. 3. Press and hold down STORE/RECALL and press ENTER. NOTE: Only memory storage addresses containing data are sent to the serial port. Setting the Baud Rate 1. Press and hold down the key and press the.1/.01/.001 key to enter the baud rate selection mode. The current baud rate is displayed. 2. Press the key to cycle between 300, 1200, and The factory default baud rate is When the desired baud rate is displayed, press ENTER. The meter stores the new baud rate and returns to its prior mode. 23

26 Temperature Calibration User Temperature Calibration Use the following procedure to calibrate the temperature sensor. NOTE: The Auto HOLD (Auto eye icon) must be off during temperature calibration. 1. Immerse the probe in a solution of known temperature. 2. Press and hold down the key and press CAL. The meter enters the temperature calibration mode. The temperature display flashes. 3. Press the or fl keys to set the temperature to the nearest 0.1 o C. Each time arrow key is pressed, the temperature is incremented in steps of 0.05 o C. Therefore, it is necessary to press the arrow key twice to see a 0.1 o C change on the display. 4. When the desired temperature is displayed, press ENTER. Restoring Temperature Calibration to Factory Default 1. Press and hold down the key and press CAL. The meter enters the temperature calibration mode. The temperature display flashes. 2. Press CLEAR, then ENTER. CAUTION: Do not expose the ISFET probe to repeated temperature extremes (60 C or higher). Do not use the probe for applications that require the probe to cycle between hot and room temperature samples. 24

27 Battery Replacement A 9v alkaline battery powers the IQ240 meter. The expected battery life is 200 hours of operation. The flashing status indicator arrow on the display indicates low battery voltage. Replace the battery whenever low voltage is indicated. Internal memory is not lost when the battery is replaced. To Replace the Battery: 1. Disconnect all connectors from the meter. 2. Turn the meter upside down. 3. Use a Phillips screwdriver to unscrew the two screws, located at the bottom of the two cylindrical cavities. 4. Carefully lift the bottom case of the meter directly up, allowing the screws to remain in the cavities. 5. Remove the old battery, and install the new battery. 6. Replace the cover and secure by tightening the two screws. 25

28 Troubleshooting INDICATION CAUSE USER ACTION No Display No Power Replace 9v battery, or use AC power adapter. Auto Power Off has cleared display Press ON/OFF key. Battery status indicator arrow flashing PROBE status indicator arrow flashing Unstable reading, (ph status indicator does not stop flashing). Battery is low The meter is in the serial output override mode Probe not connected to meter Dirty probe/meter contacts. Damaged probe Dirty probe. Reference junction not flowing. Miscalibrated probe. Interference from other devices. Interference from direct sunlight. Probe is in a very low ionic strength solution. Follow the directions for battery replacement on Page 25. See Page 22. Check probe connection to meter. Disconnect probe, then reconnect probe to meter (An audible click indicates that the probe is in place). Clean probe contacts on probe cable connector and meter with methanol and a cotton swab and let dry completely. Reconnect probe to meter. If condition persists, replace probe. Follow probe-cleaning procedure on Page 11. Follow warm buffer cleaning procedure on Page 11. Soak probe for three minutes in neutral buffer (ph 7.00). Follow two point calibration procedure on Page 14. Remove other devices from solution. Unplug water baths, stirrers, etc. If working outside, shield the probe from direct sunlight. Stable reading may not be possible. 26

29 Unstable reading, (ph status indicator does not stop flashing) Cont... Trouble calibrating. (ph buffer value does not stop flashing during calibration). ph or temperature of solution is changing. Probe near end of useful life. Low battery. Probe sensor surfaces are dirty or probe needs reconditioning. Reference junction not flowing. E10, E11 or E12 error message displayed. Improper calibration procedure. Buffers may be contaminated or expired. Interference from other devices in solution. Improperly grounded stirrers, water baths other devices. Battery is low. Interference from direct sunlight. If trouble persists, probe may have reached the end of its useful life and may need to be replaced. Display will show changes in ph values due to change in solution ph or temperature (meter is automatically temperature compensated to show actual ph at any given temperature). Replace probe. Replace battery if battery status indicator arrow is flashing. Follow cleaning and reconditioning instructions on Page 11. Follow warm buffer cleaning procedure on Page 7. Follow troubleshooting instructions on Page 28. Carefully follow Two Point Calibration procedure on Page 14. Recalibrate with fresh buffers. Remove any other devices from solution. Unplug water baths, stirrers, or any other electrical devices near the probe. Use AC adapter or replace battery if low battery indicator is flashing. Shield probe from sunlight. Replace probe. 27

30 Error Codes INDICATION CAUSE USER ACTION E01 This error code is displayed after the meter is turned on if any malfunction is detected. Turn the meter off. Disconnect, then reconnect the probe. Disconnect any cables connected to the RS-232 or analog ports. Turn the meter on. If E01 persists, and the meter is used with the AC power adapter, turn the meter off. Disconnect then reconnect the power adapter. Turn the meter on. If EO1 persists, turn the meter off. Disconnect the AC power adapter. Turn the meter on using battery power only. E10 This error code is displayed if a probe malfunction is detected, including probe not in solution. If EO1 persists replace probe. Probe may be dry. Place probe in solution. If E10 persists, follow the probe cleaning and reconditioning instructions in this manual. If the error code persists, turn the meter off. Disconnect, then reconnect to probe. Disconnect any cables connected to the RS-232 or analog ports. Turn the meter on. E11 This error code is displayed after completion of any calibration procedure which indicates an out of specification response slope (<45 mv/ph). If E10 persists, probe may have reached the end of its useful life. Replace probe. Carefully follow the two point calibration instructions using fresh buffers. If E11 persists, follow the probe cleaning and reconditioning instructions in this manual. If E11 persists, probe may have reached the end of its useful life. Replace probe. 28

31 Error Codes continued. E12 This error code is displayed after a calibration error. One common cause is the operator overriding the automatic buffer recognition with the SELECT BUFFER key and then selecting the wrong buffer value. Carefully follow the two point calibration instructions. Use extra caution to be sure that the correct buffer value is displayed before pressing the ENTER key to begin the calibration. If E12 persists, probe may have reached the end of its useful life. Replace probe. 29

32 For Technical Support On-line: Toll Free (USA and Canada) Telephone (858) Fax (858) Revision IQ Scientific Instruments, Inc. 30

ExStik TM ph Waterproof Meters

ExStik TM ph Waterproof Meters User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL

More information

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4

More information

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide

More information

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference

More information

ph 200 ph-/redox-/temperature Meter Operating instructions

ph 200 ph-/redox-/temperature Meter Operating instructions ph 200 ph-/redox-/temperature Meter Operating instructions 2 Table of Contents Page 1. Scope of delivery...4 2. System description...4 3. Connections...5 4. Display...7 5. Keypad...7 6. Start up...8 7.

More information

Instruction Manual Temp 4/5/6 Temperature Meter Series

Instruction Manual Temp 4/5/6 Temperature Meter Series Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.

More information

UltraBasic Benchtop ph Meter Operation Manual

UltraBasic Benchtop ph Meter Operation Manual UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.

More information

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration

More information

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page Care and use guide Contents page Features 2 Quick guide 2 To operate 3 Cleaning 4 Battery replacement 4 Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about

More information

Bluelab Soil ph Meter Instruction Manual

Bluelab Soil ph Meter Instruction Manual Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for

More information

ph Measurement and Control

ph Measurement and Control ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components

More information

SenTix 950 SenTix 980

SenTix 950 SenTix 980 SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the

More information

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600 User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously

More information

Operating manual. ph 3110. ph meter. ba75791e02 11/2008

Operating manual. ph 3110. ph meter. ba75791e02 11/2008 AR Operating manual ph 3110 ph 6 9 9 2 2 4 8 C TP ph meter ph 3110 Accuracy when going to press The use of advanced technology and the high quality standard of our instruments are the result of a continuous

More information

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01 Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery

More information

Milwaukee USER MANUAL

Milwaukee USER MANUAL Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the

More information

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical

More information

AutoRanging Digital MultiMeter

AutoRanging Digital MultiMeter Owner's Manual AutoRanging Digital MultiMeter Model No. 82139 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance

More information

Care and use guide. Contents. page

Care and use guide. Contents. page Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement

More information

Manual Ranging MultiMeter

Manual Ranging MultiMeter Owner s Manual Manual Ranging MultiMeter Model 82345 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product.! Safety! Operation! Maintenance!

More information

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...

More information

Portable Waterproof ph/dissolved Oxygen Meter

Portable Waterproof ph/dissolved Oxygen Meter Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35632-Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ /

More information

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510 User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Thermistor Thermometer MODEL NO. 93210-00

Thermistor Thermometer MODEL NO. 93210-00 Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:

More information

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev. Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE

More information

Instruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X361305 Rev. 0-11/05. Technology Made Easy...

Instruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X361305 Rev. 0-11/05. Technology Made Easy... Instruction Manual DO 110 Hand-held Dissolved Oxygen Meter with RS 232C Technology Made Easy... 68X361305 Rev. 0-11/05 Preface This manual serves to explain the use of the hand-held series meters. The

More information

USER GUIDE. LED Light Meter. Model LT45

USER GUIDE. LED Light Meter. Model LT45 USER GUIDE LED Light Meter Model LT45 Introduction Congratulations on your purchase of the Extech LT45 LED Light Meter that measures light from LED lamps in addition to fluorescent, metal halide, high

More information

Model 4510 Conductivity/TDS Meter Operating Manual

Model 4510 Conductivity/TDS Meter Operating Manual Model 4510 Conductivity/TDS Meter Operating Manual SAFETY Please read this information carefully prior to installing or using this equipment. 1. The unit described in this manual is designed to be operated

More information

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn

More information

Appendix C. Vernier Tutorial

Appendix C. Vernier Tutorial C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer

More information

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination

More information

Operating Instructions

Operating Instructions Operating Instructions FiveEasyPlus ph meter FEP20 Contents Contents 1. Introduction 3 2. Safety measures 3 3. Installation 4 3.1 Installing the electrode arm (optional for FEP20) 4 4. Operating the FiveEasyPlus

More information

CQ-Series. User Manual. American Weigh Scales, Inc. CQ-350 (350x0.1g) CQ-500 (500x0.1g) Digital Pocket Scale . = + CALC AC CE % TARE OFF ON MC MR CALI

CQ-Series. User Manual. American Weigh Scales, Inc. CQ-350 (350x0.1g) CQ-500 (500x0.1g) Digital Pocket Scale . = + CALC AC CE % TARE OFF ON MC MR CALI American Weigh Scales, Inc. CQ-Series Digital Pocket Scale 7 4 1 0. 8 9 5 6 2 3 x = - CALI + CALC MODE M CAL AC CE % TARE OFF ON MC MR CQ-350 (350x0.1g) CQ-500 (500x0.1g) User Manual CQ-Series Manual Thank

More information

Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements

Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

1. SAFETY INFORMATION

1. SAFETY INFORMATION RS-232 Sound Level Meter 72-860A INSTRUCTION MANUAL www.tenma.com 1. SAFETY INFORMATION Read the following safety information carefully before attempting to operate or service the meter. Use the meter

More information

Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986

Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986 Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986 Fann Instrument Company Electrical Stability Tester Instruction Manual 2012 Fann Instrument

More information

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy... Instruction Manual ph 5+ ph/ C ph 6+ ph/ C/mV Ion 6+ ph/ C/mV/Ion Technology Made Easy... 68X243633 Rev 1-1 3/2011 Part of Thermo Fisher Scientific Preface This instruction manual serves to explain the

More information

Hygro-Thermometer + InfraRed Thermometer Model RH101

Hygro-Thermometer + InfraRed Thermometer Model RH101 User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative

More information

Measurement by Ion Selective Electrodes

Measurement by Ion Selective Electrodes Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable

More information

MAINTENANCE & TROUBLESHOOTING

MAINTENANCE & TROUBLESHOOTING MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens clean the fan intake filter replace the projection lamp replace the batteries in the remote control use the Kensington lock feature

More information

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407780 Product Page Integrating Sound Level Datalogger

More information

RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer

RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer EN FR IT DE ES CONTENTS / EN Title CONTENTS Page 1. SAFETY INFORMATION...1 2. INTRODUCTION...2 3. SPECIFICATIONS...3 4. PARTS & CONTROLS...4

More information

Thermocouple Thermometer Models

Thermocouple Thermometer Models Technology Made Easy... Thermocouple Thermometer Models 60010-00 (Type J) 60010-10 (Type K) 60010-20 (Type T) 68X309905 Rev 1 04/04 TABLE OF CONTENTS 1. INTRODUCTION 4 2. SAFETY PRECAUTIONS 6 3. SPECIFICATIONS

More information

PART 1 - INTRODUCTION...

PART 1 - INTRODUCTION... Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART

More information

Pro1030 USER MANUAL. English

Pro1030 USER MANUAL. English Pro1030 USER MANUAL English Contents Warranty...i Introduction...1 Item #605182 Rev A, January 2013 For the latest version of this manual, visit ysi.com 2013 YSI Incorporated. The YSI logo is a registered

More information

XP Series Balances Operation Manual

XP Series Balances Operation Manual XP Series Balances Operation Manual North and South America: 6542 Fig Street Arvada, Colorado 84 1-8-321-1135 Tel: 33-431-7255 Fax: 33-423-4831 U.K. and Ireland: Denver House, Sovereign Way Trafalgar Business

More information

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725 Read this manual thoroughly before using and save it for future reference Purchase MAP capsules at Harvey Norman stores nationwide

More information

USEPA Electrode Method Method 8156

USEPA Electrode Method Method 8156 ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method

More information

Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function Important Safeguards READ ALL INSTRUCTIONS BEFORE USE. Instruction Manual Automatic Dispensing Hot Water Pot with Reboil Function Image of SP-3015 & SP-3815 SP-3015: 3.0L SP-3815: 3.8L SP-3017: 3.0L (Stainless

More information

Ground Resistance Clamp On Tester

Ground Resistance Clamp On Tester USER MANUAL Ground Resistance Clamp On Tester MODEL 382357 Introduction Congratulations on your purchase of Extech s 382357 Ground Resistance Tester. This Clamp on device allows the user to measure ground

More information

4-Channel Thermometer / Datalogger

4-Channel Thermometer / Datalogger USER GUIDE 4-Channel Thermometer / Datalogger RTD and Thermocouple Inputs Model SDL200 Introduction Congratulations on your purchase of the Extech SDL200 Thermometer, an SD Logger Series meter. This meter

More information

COASTAL AMUSEMENTS INC.

COASTAL AMUSEMENTS INC. TUTTI FRUITTI A skill based video game OPERATOR MANUAL VERSION TUTTI.D.1.0.1 COASTAL AMUSEMENTS INC. 1950 Swarthmore Ave, Lakewood NJ 08701 Tel. (732) 905-6662 Fax (732) 905-6815 http://www.coastalamusements.com

More information

Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph

Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or

More information

Relative Humidity Calibration Kit

Relative Humidity Calibration Kit Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration

More information

Contents Contents Contents Overview Charging your headset Wearing the headset Powering on the headset Connecting your headset for dictation

Contents Contents Contents Overview Charging your headset Wearing the headset Powering on the headset Connecting your headset for dictation 1 Contents Contents Contents 2 Overview 3 Charging your headset 4 Wearing the headset 4 Powering on the headset 4 Connecting your headset for dictation 5 The adapter 5 Adjusting your computer's audio settings

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510

User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510 User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure

More information

Refer to page 6 to page 19 for detailed instructions.

Refer to page 6 to page 19 for detailed instructions. CODE:GZ0000297065 Refer to page 6 to page 19 for detailed instructions. Preface This manual describes the operation of the COMPACT CONDUCTIVITY METER, B-771. Be sure to read this manual before using the

More information

GAUGEMASTER PRODIGY EXPRESS

GAUGEMASTER PRODIGY EXPRESS GAUGEMASTER PRODIGY EXPRESS DCC01 USER MANUAL Version 1.1 2011 T A B L E O F C O N T E N T S 1 Getting Started Introduction Specifications and Features Quick Start Connecting to Your Layout Running a Loco

More information

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of the blood 7 Procedure to test glucose 7 Procedure to

More information

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation

More information

Care and use guide. soil ph meter. Contents. page

Care and use guide. soil ph meter. Contents. page soil ph meter Care and use guide Contents page Features Bluelab Soil ph Meter Introduction 3 Preparing for use 3 Calibration 4 Information about measuring the ph of soils/media 6 Measuring the ph value

More information

Professional Plus. Calibration Tips

Professional Plus. Calibration Tips Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting

More information

BODY FAT & HYDRATION % MONITOR SCALE

BODY FAT & HYDRATION % MONITOR SCALE EVALUATION BODY FAT & HYDRATION % MONITOR SCALE INSTRUCTION MANUAL MODEL 12-1190 ver 7/08 BODY FAT & HYDRATION % MONITOR SCALE Introduction and Precaution: MODEL: 12-1190 1. Congratulations for purchasing

More information

User Guide. Temperature and Humidity Datalogger. Model 42280

User Guide. Temperature and Humidity Datalogger. Model 42280 User Guide Temperature and Humidity Datalogger Model 42280 Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger. The 42280 is a wall-mount, tripod

More information

Troubleshooting Salt Generators

Troubleshooting Salt Generators Troubleshooting Salt Generators NOTE Turn off power to unit prior to attempting service or repair. Problems and Corrective Action Problem Possible Cause Corrective Action Low or no chlorine. Low stabilizer

More information

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date: PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA

More information

Instruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter

Instruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter Instruction Manual HI 9146 Portable Waterproof Microprocessor Dissolved Oxygen Meter w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction

More information

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN User's Guide Digital Psychrometer Model RH390 MAX MIN F C WBT Dew Point Introduction Congratulations on your purchase of the Extech RH390 Digital Psychrometer. This device measures Humidity, Air Temperature,

More information

MAINTENANCE & TROUBLESHOOTING

MAINTENANCE & TROUBLESHOOTING MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens replace the projection lamp replace the batteries in the remote use the security lock feature troubleshoot the projector Cleaning

More information

MSFET 3332. MICROSENS Miniature ph Sensor Module

MSFET 3332. MICROSENS Miniature ph Sensor Module Product Data Sheet MSFET 3332 MICROSENS Miniature ph Sensor Module Ta2O5 gate Ion Sensitive Field Effect transistor (ISFET) The MSFET3332 ph sensor module comprises a ph-isfet sensing element, an integrated

More information

Digital Photo Picture Frame. Operation Manual

Digital Photo Picture Frame. Operation Manual Digital Photo Picture Frame Operation Manual 20070309 CONGRATULATIONS on your purchase of a Polaroid 7 LCD Digital Photo Picture Frame. Please read carefully and follow all warnings and instructions in

More information

Back-UPS Pro 1300/1500 Installation and Operation

Back-UPS Pro 1300/1500 Installation and Operation Back-UPS Pro 1300/1500 Installation and Operation Inventory Safety Do not install the Back-UPS in direct sunlight, in excessive heat, humidity, or in contact with fluids. Connect the battery bu059a bu058a

More information

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer User Manual Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer MODEL 42515 Introduction Congratulations on your purchase of the Model 42515 IR Thermometer. The Model 42515 IR thermometer

More information

ELECTRONIC THERMOSTAT AND THERMOMETER With SPEED CONTROL

ELECTRONIC THERMOSTAT AND THERMOMETER With SPEED CONTROL 148 OLD CONCORD TURNPIKE, BARRINGTON NH 03825 USA TEL (603) 868-5720 FAX (603) 868-1040 1-800-435-6708 E-Mail:sales@seafrost.com www.seafrost.com ELECTRONIC THERMOSTAT AND THERMOMETER With SPEED CONTROL

More information

ph-meter 1140 and ph-meter 1140-X Instruction manual

ph-meter 1140 and ph-meter 1140-X Instruction manual ph-meter 1140 and ph-meter 1140-X Instruction manual Mettler-Toledo GmbH Process, Postfach, 8902 Urdorf, Switzerland Phone: +41 (1) 736 22 11 Telefax: +41 (1) 736 26 36 Warranty Defects occurring within

More information

Instruction Manual HI 96801 Refractometer for Sucrose Measurements

Instruction Manual HI 96801 Refractometer for Sucrose Measurements Instruction Manual HI 96801 Refractometer for Sucrose Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information

More information

12 Volt 30 Amp Digital Solar Charge Controller

12 Volt 30 Amp Digital Solar Charge Controller 12 Volt 30 Amp Digital Solar Charge Controller User s Manual WARNING Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions

More information

CARING FOR YOUR WATER HEATER

CARING FOR YOUR WATER HEATER http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html Water Heater Inspections CARING FOR YOUR WATER HEATER Venting System (Direct Vent Only) The venting system should be inspected

More information

digital fever alert ear thermometer suitable from birth instruction manual

digital fever alert ear thermometer suitable from birth instruction manual digital fever alert ear thermometer suitable from birth instruction manual digital fever alert ear thermometer instruction manual Please read these instructions carefully before using the thermometer and

More information

WALL MOUNTED SPLIT TYPE AIR CONDITIONER

WALL MOUNTED SPLIT TYPE AIR CONDITIONER WALL MOUNTED SPLIT TYPE AIR CONDITIONER MANUAL 8 OM-G12-ACSON CONTENTS - Operating Guide page 1 - G12 Remote Controller Indication page 2 - Indicator Lights page 4 - Installation of Aroma page 5 - Auto

More information

Wireless Indoor/ Outdoor Thermometer

Wireless Indoor/ Outdoor Thermometer Wireless Indoor/ Outdoor Thermometer Owner s Manual Please read before using this equipment. ˆ Contents FCC Information... 3 FCC Declaration of Conformity... 5 Preparation... 5 Installing Batteries...

More information

Auto-ranging Digital Multimeter 52-0052-2 INSTRUCTION MANUAL

Auto-ranging Digital Multimeter 52-0052-2 INSTRUCTION MANUAL Auto-ranging Digital Multimeter 52-0052-2 INSTRUCTION MANUAL WARNING: READ AND UNDERSTAND THIS MANUAL BEFORE USING YOUR MULTIMETER. FAILURE TO UNDERSTAND AND COMPLY WITH WARNINGS AND OPERATING INSTRUCTIONS

More information

User Manual. Humidity-Temperature Chart Recorder. Model RH520

User Manual. Humidity-Temperature Chart Recorder. Model RH520 User Manual Humidity-Temperature Chart Recorder Model RH520 Introduction Congratulations on your purchase of the Extech RH520 Temperature + Humidity Chart Recorder. The RH520 measures and displays Temperature,

More information

Instructions: Part I. Charge the Power Bank. Method 1. Charge the Power Bank via Computer USB Terminal.

Instructions: Part I. Charge the Power Bank. Method 1. Charge the Power Bank via Computer USB Terminal. Instructions: Part I. Charge the Power Bank Method 1 Charge the Power Bank via Computer USB Terminal. Connect the provided USB power cable to the micro- USB plug connector tip (the tip marked "M10"). Step

More information

USER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING. Non-Programmable Single Stage Heat/Cool Thermostat

USER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING. Non-Programmable Single Stage Heat/Cool Thermostat Builder MODEL 1010 Series Non-Programmable Single Stage Heat/Cool Thermostat USER MANUAL Compatible with low voltage single stage gas, oil or electric heating or cooling systems, including single stage

More information

Ground Resistance Clamp-On Tester

Ground Resistance Clamp-On Tester User s Manual Ground Resistance Clamp-On Tester MODEL 382357 Warranty EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants the basic instrument to be free of defects in parts and workmanship for one

More information

SINGLE RANGE DS METERS Operation Instructions

SINGLE RANGE DS METERS Operation Instructions SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax

More information

3 WATT LED SPOTLIGHT Model No. SLM - 3801

3 WATT LED SPOTLIGHT Model No. SLM - 3801 3 WATT LED SPOTLIGHT Model No. SLM - 3801 OWNER'S MANUAL Customer Service Tel: 1-800-268-3319 Superex Canada Ltd, Toronto,M2H 3B8 Made in China Table of Contents A). Important Safety Instructions B). Charging

More information

LDG M-7700 External Meter for Icom IC-7700

LDG M-7700 External Meter for Icom IC-7700 M-7700 OPERATIONS MANUAL MANUAL REV A LDG M-7700 External Meter for Icom IC-7700 LDG Electronics 1445 Parran Road St. Leonard MD 20685 Phone: 410-586-2177 Fax: 410-586-8475 ldg@ldgelectronics.com www.ldgelectronics.com

More information

Model 349KLX Remote Display Digital Scale User Instructions

Model 349KLX Remote Display Digital Scale User Instructions Model 349KLX Remote Display Digital Scale User Instructions P/N UM349KLX Rev 3-080713 1 Model 349KLX Thank you for your purchase of this Health o meter Professional product. Please read this manual carefully,

More information

4.3-inch Back-Up Camera

4.3-inch Back-Up Camera TM 4.-inch Back-Up Camera Model No.: PKC0BU4 Owner s Manual and Warranty Information Read these instructions completely before using this product. Retain this Owner s Manual for future reference. INTRODUCTION

More information

DISSOLVED OXYGEN PROBE

DISSOLVED OXYGEN PROBE 51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe

More information

Electric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL

Electric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL Electric Kettle Model: SK-1717 INSTRUCTIONS MANUAL Thank you for your purchase. Please read this Instruction Manual thoroughly before use and keep in a safe place for future reference. READ ALL INSTRUCTIONS

More information

Deluxe Bottle Warmer

Deluxe Bottle Warmer Deluxe Bottle Warmer IMPORTANT SAFEGUARDS This product is for household use only. When using electrical appliances, basic safety precautions should always be followed including the following: Read all

More information

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process

More information

KEYPAD LOCK USERS MANUAL

KEYPAD LOCK USERS MANUAL KEYPAD LOCK USERS MANUAL READ ALL INSTRUCTIONS CLOSELY PRIOR TO USE SAVE THESE INSTRUCTIONS CONTENTS Keypad Parts and Features 2 Programming the Master Code 3 Programming User Codes 4 Using Your Keypad

More information

www.burntec.com User's Guide Integrating Sound Level Datalogger Model 407780

www.burntec.com User's Guide Integrating Sound Level Datalogger Model 407780 User's Guide Integrating Sound Level Datalogger Model 407780 Introduction Congratulations on your purchase of the Extech 407780 Integrating Sound Level Meter. The 407780 with programmable integrating time

More information

Operating manual. ph/cond 3400i. Tref25 nlf C TP CAL. ph/conductivity measuring instrument. ba75687e01 01/2007

Operating manual. ph/cond 3400i. Tref25 nlf C TP CAL. ph/conductivity measuring instrument. ba75687e01 01/2007 Operating manual ph/cond 3400i 5 3 0 Tref25 nlf 2 5 0 S/ C TP M AR STO CAL O AR ph/conductivity measuring instrument ba75687e01 01/2007 Accuracy when going to press The use of advanced technology and the

More information

G-100/200 Operation & Installation

G-100/200 Operation & Installation G-100/200 Operation & Installation 2 Contents 7 Installation 15 Getting Started 16 GPS Mode Setup 18 Wheel Sensor Mode Setup 20 Fuel Calibration 23 Basic Operation 24 Telemetery Screen 27 Entering a Distance

More information