This package insert contains information to run the Lactic Acid assay on the ARCHITECT c Systems and the AEROSET System.

Size: px
Start display at page:

Download "This package insert contains information to run the Lactic Acid assay on the ARCHITECT c Systems and the AEROSET System."

Transcription

1 LACTIC ACID 9D /R6 LACTIC ACID This package insert contains information to run the Lactic Acid assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This package insert must be read carefully prior to product use. Package insert instructions must be followed accordingly. Reliability of assay results cannot be guaranteed if there are any deviations from the instructions in this package insert. Customer Support United States: Canada: International: ABBOTT (English speaking customers) (French speaking customers) Call your local Abbott representative Symbols in Product Labeling Concentration Authorized Representative in the European Community Ingredients In vitro diagnostic medical device Batch code/lot number Reagent 1 Catalog number/list number Serial number Consult instructions for use Manufacturer Temperature limitation Use by/expiration date ABBOTT LABORATORIES Abbott Park, IL 60064, USA ABBOTT Max-Planck-Ring Wiesbaden Germany March , 2009 Abbott Laboratories 1

2 NAME LACTIC ACID INTENDED USE The Lactic Acid assay is used for the quantitation of lactic acid in human plasma. SUMMARY AND EXPLANATION OF TEST Lactic acid, an intermediary in carbohydrate metabolism, is derived predominantly from skeletal muscle, brain, and erythrocytes. The blood lactate concentration is dependent on the rate of production and the rate of metabolism in the liver and kidneys. Approximately 30% of total lactate production is used by the liver. Moderate increase in lactate production results in increased hepatic lactate clearance, but uptake by the liver is saturable when concentrations exceed 2 mmol/l. For example, during strenuous exercise, lactate concentrations may increase significantly, from an average concentration of about 0.9 mmol/l to more than 20 mmol/l within 10 seconds. 1 The review by Stacpoole 2 describes the biochemistry and pathobiology of lactate metabolism. A pivotal characteristic of this metabolism is that lactate can only arise from reduction of pyruvate. Elevated lactate associated with reduced arterial ph is called lactic acidosis. Tissue hypoxia is the most common cause of lactic acidosis and is classified as Type A, while lactic acidosis that is nonhypoxic is classified as Type B. Under normal conditions, the concentration of lactate is about 10-fold higher than that of pyruvate. Shock is perhaps the most widely recognized cause of lactic acidosis; however, in some cases, excess lactate production may precede shock. Such conditions as myocardial infarction, severe congestive heart failure, pulmonary edema, and blood loss are common causes of shock associated with lactic acidosis. Other causes of lactic acidosis include intravenous infusion of substances such as fructose, sorbitol, or epinephrine, and large doses of drugs such as ethanol or acetaminophen. Hepatic necrosis, neoplasms, lymphomas, and various forms of leukemia have been reported to cause lactic acidosis. In diabetic coma, lactic acidosis is common. 3 PRINCIPLES OF PROCEDURE Lactic acid is converted to pyruvate and hydrogen peroxide (H 2 O 2 ) by lactate oxidase. Peroxidase catalyzes the oxidation of chromogen precursor by H 2 O 2 to produce a colored dye. The increase in absorbance at 548 nm is directly proportional to the lactic acid concentration in the sample. Methodology: Lactic Acid to Pyruvate REAGENTS Reagent Kit 9D89 Lactic Acid is supplied as a lyophilized, single reagent kit which contains: 10 x 10 ml Funnels (5) Bar code labeled cartridges (5) Estimated tests per kit: 393 Calculation is based on the minimum reagent fill volume per kit. Reactive Ingredients Concentration Lactate Oxidase 400 U/L Peroxidase (Horseradish) 2,400 U/L Chromogen precursors As required Inactive Ingredients: Contains buffer, fillers, and stabilizers. REAGENT HANDLING AND STORAGE Reagent Handling Remove air bubbles, if present in the reagent cartridge, with a new applicator stick. Alternatively, allow the reagent to sit at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove the bubbles. CAUTION: Reagent bubbles may interfere with proper detection of reagent level in the cartridge, causing insufficient reagent aspiration which could impact results. Reagent Storage Unopened reagents are stable until the expiration date when stored at 2 to 8 C. Do not pool reagents. Reagent stability is 7 days if the reagent is uncapped and onboard. REAGENT HANDLING AND STORAGE (Continued) Instructions for Use 1. Remove the reagent caps from two bottles of. 2. Prepare the Working Reagent by adding 10 ml of sterile water to each of the two bottles. 10 ml H 2 O + R1 = 10 ml H 2 O + R1 = Working Reagent 1 3. Replace the reagent caps and mix each bottle by gentle inversion until completely dissolved. Do Not Shake. 4. Pour contents of both bottles into one of the empty bar code labeled cartridges provided with the reagent kit. Remove air bubbles, if present in the cartridge, with a new applicator stick. 5. Place the cartridge in Reagent Supply Center 1. WARNINGS AND PRECAUTIONS Precautions for Users 1. For in vitro diagnostic use. 2. Do not use components beyond the expiration date. 3. Do not mix materials from different kit lot numbers. 4. CAUTION: This product requires the handling of human specimens. It is recommended that all human sourced materials be considered potentially infectious and be handled in accordance with the OSHA Standard on Bloodborne Pathogens. 4 Biosafey Level 2 5 or other appropriate biosafety practices 6,7 should be used for materials that contain or are suspected of containing infectious agents. 5. is classified per applicable European Community (EC) Directives as: Irritant (Xi). The following are the appropriate Risk (R) and Safety (S) phrases: R36/37/38 Irritating to eyes, respiratory system and skin. S26 In case of contact with eyes, rinse immediately with plenty of water and seek medical advice. S35 This material and its container must be disposed of in a safe way. S46 If swallowed, seek medical advice immediately and show this container or label. SPECIMEN COLLECTION AND HANDLING Suitable Specimens Plasma is the acceptable specimen. Plasma: Use plasma collected by standard venipuncture techniques into glass or plastic tubes without gel barriers. The acceptable anticoagulant is sodium fluoride/potassium oxalate. Venous specimens should be obtained without the use of a tourniquet or immediately after the tourniquet has been applied. If there is any delay in obtaining the specimen, the tourniquet should be removed after the puncture has been performed, and the blood should be allowed to circulate for at least two minutes before sample is withdrawn. Patients should avoid exercise of the hand or arm immediately before and during the procedure. Centrifuge the specimen as soon after collection as possible and immediately remove the plasma from the red blood cells. 1 Ensure centrifugation is adequate to remove platelets. For total sample volume requirements, refer to the instrument-specific ASSAY PARAMETERS section of this package insert and Section 5 of the instrument-specific operations manual. Specimen Storage Plasma: Separated plasma may be analyzed immediately, stored at 2 to 8 C, or frozen. Samples are stable for 14 days at 2 to 8 C or up to 4 weeks at -20 C. Specimen stability was verified in an in-house study of 20 sodium fluoride/potassium oxalate plasma samples. NOTE: Stored specimens must be inspected for particulates. If present, mix and centrifuge the specimen to remove particulates prior to testing. 2

3 PROCEDURE Materials Provided 9D89 Lactic Acid Reagent Kit Materials Required but not Provided 1E75 Lactic Acid Calibrator 1 x 10 ml Control Material Saline (0.85% to 0.90% NaCl) for specimens that require dilution Assay Procedure For a detailed description of how to run an assay, refer to Section 5 of the instrument-specific operations manual. Specimen Dilution Procedures The ARCHITECT c Systems and the AEROSET System have automatic dilution features; refer to Section 2 of the instrument-specific operations manual for additional information. Plasma: Specimens with lactic acid values exceeding mg/dl (13.32 mmol/l) are flagged and may be diluted using the Automated Dilution Protocol or the Manual Dilution Procedure. Automated Dilution Protocol If using the Automated Dilution Protocol, the system performs a dilution of the specimen and automatically corrects the concentration by multiplying the result by the appropriate dilution factor. To set up the automatic dilution feature, refer to Section 2 of the instrument-specific operations manual for additional information. Manual Dilution Procedure Manual dilutions should be performed as follows: Use saline (0.85% to 0.90% NaCl) to dilute the sample. The operator must enter the dilution factor in the patient or control order screen. The system uses this dilution factor to automatically correct the concentration by multiplying the result by the entered factor. If the operator does not enter the dilution factor, the result must be multiplied by the appropriate dilution factor before reporting the result. NOTE: If a diluted sample result is flagged indicating it is less than the linear low limit, do not report the result. Rerun using an appropriate dilution. For detailed information on ordering dilutions, refer to Section 5 of the instrument-specific operations manual. CALIBRATION Calibration is stable for approximately 7 days (168 hours). Calibration is required whenever a new reagent is reconstituted. Verify calibration curve with at least two levels of controls according to the established quality control requirements for your laboratory. If control results fall outside acceptable ranges, recalibration may be necessary. For a detailed description of how to calibrate an assay, refer to Section 6 of the instrument-specific operations manual. For information on calibrator standardization, refer to the Lactic Acid Calibrator package insert. QUALITY CONTROL The following process is the recommendation of Abbott Laboratories for quality control. As appropriate, refer to your laboratory standard operating procedure(s) and/or quality assurance plan for additional quality control requirements and potential corrective actions. Two levels of controls (normal and abnormal) are to be run every 24 hours. If more frequent control monitoring is required, follow the established quality control procedures for your laboratory. If quality control results do not meet the acceptance criteria defined by your laboratory, patient values may be suspect. Follow the established quality control procedures for your laboratory. Recalibration may be necessary. Review quality control results and acceptable criteria following a change of reagent or calibrator lot. RESULTS Refer to the instrument-specific operations manual for information on results calculations. ARCHITECT System Operations Manual Appendix C AEROSET System Operations Manual Appendix A Representative performance data are given in the EXPECTED VALUES and SPECIFIC PERFORMANCE CHARACTERISTICS sections of this package insert. Results obtained in individual laboratories may vary. LIMITATIONS OF THE PROCEDURE Refer to the SPECIMEN COLLECTION AND HANDLING and SPECIFIC PERFORMANCE CHARACTERISTICS sections of this package insert. The performance characteristics of Lactic Acid on an analyzer other than an ARCHITECT c Systems or the AEROSET System must be validated and verified. EXPECTED VALUES Reference Range Plasma 8 Range (mg/dl) Range (mmol/l) Venous 4.5 to to 2.2 To convert results from mg/dl to mmol/l, multiply mg/dl by A study was conducted using 204 plasma samples from 99 female and 105 male volunteers. Data were analyzed as described by Solberg 9 and Clinical and Laboratory Standards Institute (CLSI) protocol NCCLS C28-A. 10 From this study, 95% of male specimens fell within 5.0 to 19.4 mg/dl, with male samples ranging from 3.9 to 20.6 mg/dl. For female specimens, 95% fell within 4.7 to 16.8 mg/dl, with female samples ranging from 4.5 to 18.6 mg/dl. It is recommended that each laboratory determine its own reference range based upon its particular locale and population characteristics. SPECIFIC PERFORMANCE CHARACTERISTICS Linearity Lactic Acid is linear up to mg/dl (13.32 mmol/l). Linearity was verified using CLSI protocol NCCLS EP6-A. 11 Limit of Detection (LOD) The LOD for Lactic Acid is 0.05 mg/dl (0.006 mmol/l). The LOD is the mean concentration of an analyte-free sample + 2 SD, where SD = the pooled, within-run standard deviation of the analyte-free sample. Limit of Quantitation (LOQ) The LOQ for Lactic Acid is 0.18 mg/dl (0.020 mmol/l). The LOQ is the analyte concentration at which the CV = 20%. Interfering Substances Interference studies were conducted on the AEROSET System using CLSI protocol NCCLS EP7-P. 12 Interference effects were assessed by Dose Response and Paired Difference methods, at the medical decision level of the analyte. Interfering Substance Bilirubin Hemoglobin Human triglyceride Intralipid Interferent Concentration N Target Observed (mg/dl) (% of Target) 15 mg/dl (257 μmol/l) mg/dl (513 μmol/l) mg/dl (5.0 g/l) mg/dl (7.5 g/l) mg/dl (8.5 mmol/l) ,000 mg/dl (11.3 mmol/l) mg/dl (1.25 g/l) mg/dl (2.50 g/l) Bilirubin solutions at the above concentrations were prepared by addition of a bilirubin stock to human plasma pools. Hemoglobin solutions at the above concentrations were prepared by addition of hemolysate to human plasma pools. Human triglyceride solutions at the above concentrations were prepared by mixing an elevated triglyceride human plasma pool with a normal triglyceride human plasma pool. Intralipid solutions at the above concentrations were prepared by addition of Intralipid to human plasma pools. Interferences from medications or endogenous substances may affect results. 13 3

4 SPECIFIC PERFORMANCE CHARACTERISTICS (Continued) Precision The imprecision of the Lactic Acid assay is 6.3% Total CV. Representative data from studies using CLSI protocol NCCLS EP5-T2 14 are summarized below. Control Level 1 Level 2 N Mean (mg/dl) Within Run Between Run Between Day Total SD %CV SD %CV SD %CV SD %CV Method Comparison Correlation studies were performed using CLSI protocol NCCLS EP9-A. 15 Plasma results from the Lactic Acid assay on the AEROSET System were compared with those from a commercially available lactate oxidase methodology. Plasma results from the Lactic Acid assay on an ARCHITECT c System were compared with the Lactic Acid assay on the AEROSET System. AEROSET vs. Comparative Method ARCHITECT vs. AEROSET N Y - Intercept Correlation Coefficient Slope Range (mg/dl)* 6.2 to to *AEROSET Range BIBLIOGRAPHY 1. Burtis CA, Ashwood ER, editors. Tietz Textbook of Clinical Chemistry, 2nd ed. Philadelphia, PA: WB Saunders; 1994: Stacpoole PW. Lactic acidosis. In: Ober PK, editor. Endocrinology and Metabolism Clinics of North America. Philadelphia, PA: WB Saunders; 1993: Henry JB, editor. Clinical Diagnosis and Management by Laboratory Methods, 19th ed. Philadelphia, PA: WB Saunders; 1996: US Department of Labor, Occupational Safety and Health Administration. 29 CFR Part , Bloodborne Pathogens. 5. US Department of Health and Human Services. Biosafety in Microbiological and Biomedical Laboratories, 5th ed. Washington, DC: US Government Printing Office, January World Health Organization. Laboratory Biosafety Manual, 3rd ed. Geneva: World Health Organization, Sewell DL, Bove KE, Callihan DR, et al. Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline Third Edition (M29-A3). Wayne, PA: Clinical and Laboratory Standards Institute, Tietz NW, editor. Clinical Guide to Laboratory Tests, 3rd ed. Philadelphia, PA: WB Saunders; 1995: Solberg HE. Establishment and use of reference values. In: Burtis CA, Ashwood ER, editors. Tietz Textbook of Clinical Chemistry, 2nd ed. Philadelphia, PA: WB Saunders; 1994: Sasse EA, Aziz KJ, Harris EK, et al. How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline (C28-A). Villanova, PA: The National Committee for Clinical Laboratory Standards, Tholen DW, Kroll M, Astles JR, et al. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (EP6-A). Wayne, PA: The National Committee for Clinical Laboratory Standards, Powers DM, Boyd JC, Glick MR, et al. Interference Testing in Clinical Chemistry; Proposed Guideline (EP7-P). Villanova, PA: The National Committee for Clinical Laboratory Standards, Young DS. Effects of Drugs on Clinical Laboratory Tests, 4th ed. Washington, DC: AACC Press; 1995: Kennedy JW, Carey RN, Coolen RB, et al. Evaluation of Precision Performance of Clinical Chemistry Devices Second Edition; Tentative Guideline (EP5-T2). Villanova, PA: The National Committee for Clinical Laboratory Standards, Kennedy JW, Carey RN, Coolen RB, et al. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP9-A). Wayne, PA: The National Committee for Clinical Laboratory Standards, TRADEMARKS The ARCHITECT c System family of instruments consists of c 4000, c 8000, and c instruments. AEROSET, ARCHITECT, c 4000, c 8000, c 16000, and c System are trademarks of Abbott Laboratories. All other trademarks, brands, product names, and trade names are property of their respective companies. 4

5 ARCHITECT c SYSTEMS ASSAY PARAMETERS Lactic Acid Plasma Conventional and SI Units Configure assay parameters General General о Calibration о SmartWash о Results о Interpretation Assay: Lact Type: Photometric Version: Number: 1050 Reaction definition о Reagent / Sample о Validity checks Reaction mode: End up Primary Secondary Read times Wavelength: 548 / 700 Main: Last required read: 25 Absorbance range: Sample blank type: None Color correction: о Reaction definition Reagent / Sample о Validity checks R1 Reagent: LACT0 Reagent volume: 200 Diluent: Saline Water volume: Diluent dispense mode: Type 0 Dispense mode: Type 0 Diluted Default Dilution name Sample sample Diluent Water Dilution factor dilution STANDARD 2.0 = 1:1.00 : = о : = о о Reaction definition о Reagent / Sample Validity checks Reaction check: None Maximum absorbance variation: Configure assay parameters Calibration о General Calibration о SmartWash о Results о Interpretation Assay: Lact Calibration method: Linear Calibrators о Volumes о Intervals о Validity checks Calibrator set: Calibrator level: Concentration: Lactic Blank: Water 0 Cal 1: Lactic1 Replicates: 3 [Range 1 3] о Calibrators Volumes о Intervals о Validity checks Calibrator: Lactic Diluted Calibrator level Sample sample Diluent Water Blank: Water 2.0 Cal 1: Lactic1 2.0 о Calibrators о Volumes Intervals о Validity checks Calibration intervals: Full interval: 168 (hours) Calibration type: Adjust type: None о Calibrators о Volumes о Intervals Validity checks Blank absorbance range: Span: Blank Blank Span absorbance range: Expected cal factor: 0.00 Expected cal factor tolerance %: 0 Configure assay parameters SmartWash о General о Calibration SmartWash о Results о Interpretation Assay: Lact COMPONENT REAGENT / ASSAY WASH Volume Replicates Cuvette Trig 10% Detergent B 345 Lactic Acid Plasma Conventional Units Configure assay parameters Results Conventional Units о General о Calibration о SmartWash Results о Interpretation Assay: Lact Assay number: 1050 Dilution default range: Result units: mg/dl Low-Linearity: 0.2 High-Linearity: Gender and age specific ranges: GENDER AGE (UNITS) NORMAL EXTREME Either (Y) Configure result units Conventional Units Assay: Lact Version: Result units: mg/dl Decimal places: 1 [Range 0 4] Correlation factor: Intercept: Lactic Acid Plasma SI Units Configure assay parameters Results SI Units о General о Calibration о SmartWash Results о Interpretation Assay: Lact Assay number: 1050 Dilution default range: Result units: mmol/l Low-Linearity: 0.02 High-Linearity: Gender and age specific ranges: GENDER AGE (UNITS) NORMAL EXTREME Either (Y) Configure result units SI Units Assay: Lact Version: Result units: mmol/l Decimal places: 2 [Range 0 4] Correlation factor: Intercept: Due to differences in instrument systems and unit configurations, version numbers may vary. Displays the number of decimal places defined in the decimal places parameter field. Refer to the concentration specified on calibrator labeling or value sheet. In ARCHITECT software version 5.00 and above, these values are defined on the Configure calibrator set screen. The linear low value (Low-Linearity) is LOQ rounded up to the number of decimal places defined in the decimal places parameter field. 5

6 AEROSET SYSTEM ASSAY PARAMETERS Lactic Acid Plasma Conventional Units Assay Configuration: Outline Page Assay Name Assay # Line Lact 50 A-Line Quantitative Ranges Min Text Min Panic-L L-Reference-H Panic-H Max Max Text * 0.0* * * 0.2** L-Linear Range-H Reference Ranges* Age Male Female Qualitative Ranges N/A Lactic Acid Plasma SI Units Assay Configuration: Outline Page Assay Name Assay # Line Lact 50 A-Line Quantitative Ranges Min Text Min Panic-L L-Reference-H Panic-H Max Max Text * 0.0* * * 0.02** L-Linear Range-H Reference Ranges* Age Male Female Qualitative Ranges N/A Assay Configuration: Base Page Reaction Mode Wavelength-Prim/Sec Read time-main/flex AbsMaxVar END UP 548 / / Sample Blank Test Blank Read Time Abs Window Abs Limits ( ) S.Vol DS.Vol D.Vol W.Vol Standard Rgt Name/Pos Dil Diluent: * Dil Type# 0 Rgt Name/Pos R.Vol W.Vol Type# Reagent 1 LACT011 * Reaction Check Read Time A/B Range Minimum 1 1 / Factor/Intercept Decimal Places Units 1.0 / mg/dl Assay Configuration: Base Page Reaction Mode Wavelength-Prim/Sec Read time-main/flex AbsMaxVar END UP 548 / / Sample Blank Test Blank Read Time Abs Window Abs Limits ( ) S.Vol DS.Vol D.Vol W.Vol Standard Rgt Name/Pos Dil Diluent: * Dil Type# 0 Rgt Name/Pos R.Vol W.Vol Type# Reagent 1 LACT011 * Reaction Check Read Time A/B Range Minimum 1 1 / Factor/Intercept Decimal Places Units 1.0 / mmol/l Assay Configuration: Calibration Page Calib Mode Interval (H) Linear 168 Blank/Calib Replicates Extrapolation % Span Span Abs Range 3 / 3 0 BLK Sample S.Vol DS.Vol D.Vol W.Vol Blk Abs Range BLK Water C1 Lactic Cal Deviation C FAC Limit (%) 10 Assay Configuration: SmartWash Page Rgt Probe Reagent Wash Vol Cuvette Assay Name Wash Vol Sample Probe Wash Assay Configuration: Calibration Page Calib Mode Interval (H) Linear 168 Blank/Calib Replicates Extrapolation % Span Span Abs Range 3 / 3 0 BLK Sample S.Vol DS.Vol D.Vol W.Vol Blk Abs Range BLK Water C1 Lactic Cal Deviation C FAC Limit (%) 10 Assay Configuration: SmartWash Page Rgt Probe Reagent Wash Vol Cuvette Assay Name Wash Vol Sample Probe Wash Refer to Assay Configuration in Section 2 of the AEROSET System Operations Manual for information regarding assay parameters. * User defined or instrument defined. ** The linear low value (L-Linear Range) is LOQ rounded up to the number of decimal places defined in the decimal places parameter field. 6

7 7

8 8

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System. CHOLESTEROL 7D62-20 30-3126/R3 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This

More information

ALKALINE PHOSPHATASE

ALKALINE PHOSPHATASE ALKALINE PHOSPHATASE 7D55-20 and 7D55-30 30-3947/R7 ALKALINE PHOSPHATASE This package insert contains information to run the Alkaline Phosphatase assay on the ARCHITECT c Systems and the AEROSET System.

More information

This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System. GLUCOSE 3L82-20 and 3L82-40 30-3941/R4 GLUCOSE This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This

More information

This package insert contains information to run the Uric Acid assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Uric Acid assay on the ARCHITECT c Systems and the AEROSET System. URIC ACID 7D76-20 30-3928/R4 URIC ACID This package insert contains information to run the Uric Acid assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This package

More information

This package insert contains information to run the Creatine Kinase assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Creatine Kinase assay on the ARCHITECT c Systems and the AEROSET System. CREATINE KINASE 7D63-20 and 7D63-30 30-3938/R4 CREATINE KINASE This package insert contains information to run the Creatine Kinase assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes

More information

This package insert contains information to run the Creatinine assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Creatinine assay on the ARCHITECT c Systems and the AEROSET System. CREATININE 7D64-20 30-3916/R7 CREATININE This package insert contains information to run the Creatinine assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This package

More information

This package insert contains information to run the Ultra HDL assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Ultra HDL assay on the ARCHITECT c Systems and the AEROSET System. ULTRA HDL 3K33-20 30-4129/R5 ULTRA HDL This package insert contains information to run the Ultra HDL assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This package

More information

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL

More information

How To Test For Creatinine

How To Test For Creatinine Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date

More information

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban

More information

Intact PTH. system. Intact PTH 84-6434/R5. Key to symbols used 8K25

Intact PTH. system. Intact PTH 84-6434/R5. Key to symbols used 8K25 system en Intact PTH 8K25 84-6434/R5 Intact PTH Read Highlighted Changes Revised November, 2008 Customer Service United States: 1-877-4ABBOTT International: Call your Abbott Representative This package

More information

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Catalog No: E0479r 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K092353 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY B. Purpose for Submission: This is a new 510k application for a new indication for the MONOLISA Anti-HAV IgM EIA

More information

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2 Uscn Life Science Inc. Wuhan Website: www.uscnk.com Phone: +86 27 84259552 Fax: +86 27 84259551 E-mail: uscnk@uscnk.com ELISA Kit for Human Prostaglandin E1(PG-E1) Instruction manual Cat. No.: E0904Hu

More information

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. IgM ELISA For the quantitative determination of IgM in human serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calibrator

More information

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma Mouse Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use Only. Not For Use

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: Creatine kinase (CPK) Serum Beckman UniCel DxC800 Synchron Method No.: Revised: as performed by: Collaborative Laboratory Services, L.L.C Contact: Important

More information

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory

More information

Biochemistry Validation Form

Biochemistry Validation Form Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August

More information

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 3

More information

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit Catalog No. CSB-E15852c (96T) This immunoassay kit allows for the in vitro quantitative determination of canine CK-MB concentrations in serum and plasma.

More information

Lab 2. Spectrophotometric Measurement of Glucose

Lab 2. Spectrophotometric Measurement of Glucose Lab 2 Spectrophotometric Measurement of Glucose Objectives 1. Learn how to use a spectrophotometer. 2. Produce a glucose standard curve. 3. Perform a glucose assay. Safety Precautions Glucose Color Reagent

More information

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev. K-ASSAY Mouse IgM ELISA For the quantitative determination of IgM in mouse biological samples Cat. No. KT-407 For Research Use Only. 1 Rev. 072309 K-ASSAY PRODUCT INFORMATION Mouse IgM ELISA Cat. No. KT-407

More information

Creatine Kinase Activity Assay Kit (Colorimetric)

Creatine Kinase Activity Assay Kit (Colorimetric) ab155901 Creatine Kinase Activity Assay Kit (Colorimetric) Instructions for Use For the sensitive and accurate measurement of Creatine Kinase activity in various samples. This product is for research use

More information

Hydrogen Peroxide Cell-Based Assay Kit

Hydrogen Peroxide Cell-Based Assay Kit Hydrogen Peroxide Cell-Based Assay Kit Item No. 600050 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION

More information

Human Free Testosterone(F-TESTO) ELISA Kit

Human Free Testosterone(F-TESTO) ELISA Kit Human Free Testosterone(F-TESTO) ELISA Kit Catalog Number. MBS700040 For the quantitative determination of human free testosterone(f-testo) concentrations in serum, plasma. This package insert must be

More information

RayBio Creatine Kinase (CK) Activity Colorimetric Assay Kit

RayBio Creatine Kinase (CK) Activity Colorimetric Assay Kit RayBio Creatine Kinase (CK) Activity Colorimetric Assay Kit User Manual Version 1.0 May 28, 2014 RayBio Creatine Kinase Activity Colorimetric Assay (Cat#: 68CL-CK-S100) RayBiotech, Inc. We Provide You

More information

Malondialdehyde (MDA) ELISA

Malondialdehyde (MDA) ELISA Malondialdehyde (MDA) ELISA For the quantitative determination of MDA in serum, plasma and other biological fluids Cat. No. KT-21493 For Research Use Only. Not for use in diagnostic procedures. Page 1

More information

Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit

Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit Catalog Number. CSB-E08141m For the quantitative determination of mouse glycated hemoglobin A1c(GHbA1c) concentrations in lysate for RBC. This package insert

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.

More information

3.4 TEST FOR BACTERIAL ENDOTOXINS. Final text for revision of The International Pharmacopoeia

3.4 TEST FOR BACTERIAL ENDOTOXINS. Final text for revision of The International Pharmacopoeia Document QAS/11.5 FINAL July 01 3. TEST FOR BACTERIAL ENDOTOXINS Final text for revision of The International Pharmacopoeia This monograph was adopted at the Forty-sixth WHO Expert Committee on Specifications

More information

Thrombin Generation Assay

Thrombin Generation Assay Thrombin Generation Assay Kit insert Version: February 2013 Summary Thrombin is a key enzyme of the coagulation cascade. Its measurement gives direct information about the thrombogenicity of a biomaterial

More information

PRINCIPLE REF 475136. ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE REF 475136. ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Rheumatoid Factor REF 475136 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA

Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA KAMIYA BIOMEDICAL COMPANY Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA For the quantitative determination of mouse CKMB in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-57681

More information

Laboratory 5: Properties of Enzymes

Laboratory 5: Properties of Enzymes Laboratory 5: Properties of Enzymes Technical Objectives 1. Accurately measure and transfer solutions with pipettes 2. Use a Spectrophotometer to study enzyme action. 3. Properly graph a set of data. Knowledge

More information

ENZYME KINETICS ENZYME-SUBSTRATE PRODUCTS

ENZYME KINETICS ENZYME-SUBSTRATE PRODUCTS ENZYME KINETICS INTRODUCTION The study of reaction rates catalyzed by enzymes and the factors affecting them is generally referred to as enzyme kinetics. The basic components of an enzyme catalyzed reaction

More information

Vitamin B12. ADVIA Centaur System

Vitamin B12. ADVIA Centaur System Vitamin B12 Principle of the Test ADVIA Centaur System The ADVIA Centaur VB12 assay is a competitive immunoassay using direct chemiluminescent technology in which vitamin B 12 from the patient sample competes

More information

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit Catalog Number. CSB-E14405r For the quantitative determination of rat creatine kinase MM isoenzyme (CK-MM) concentrations in serum, plasma and tissue

More information

Uscn Life Science Inc. Wuhan

Uscn Life Science Inc. Wuhan Uscn Life Science Inc. Wuhan Website: www.uscnk.com Phone: +86 27 84259552 Fax: +86 27 84259551 E-mail: uscnk@uscnk.com ELISA Kit for Human Creatine Kinase,Mitochondrial 1A (CKMT1A) Instruction manual

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical

More information

Rat Creatine Kinase MB Isoenzyme (CKMB) ELISA

Rat Creatine Kinase MB Isoenzyme (CKMB) ELISA Rat Creatine Kinase MB Isoenzyme (CKMB) ELISA For the quantitative determination of rat CKMB in serum, plasma, tissue homogenates and other biological fluids. Cat. No. KT-12247 For Research Use Only. Not

More information

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Introduction: The determination of protein concentration is frequently required in biochemical work. Several methods

More information

Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions... 1. 4 Definitions...

Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions... 1. 4 Definitions... Vol. 27 No. 26 Replaces H3-A5 Vol. 23 No. 32 Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition This document provides procedures for the collection

More information

Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit

Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit Catalog No. MBS702810 (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of mouse KLH(IgM)antibody concentrations

More information

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Creatine Kinase Activity Colorimetric Assay Kit ABE5487 100 assays; Store at -20 C

Creatine Kinase Activity Colorimetric Assay Kit ABE5487 100 assays; Store at -20 C Creatine Kinase Activity Colorimetric Assay Kit ABE5487 100 assays; Store at -20 C I. Introduction: Creatine Kinase (CK) also known as creatine phosphokinase (CPK) and ATP: creatine N- phosphotransferase

More information

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,

More information

DHEA-S. system DHEA-S 48-3712/R4. Key to symbols used 8K27

DHEA-S. system DHEA-S 48-3712/R4. Key to symbols used 8K27 system en DHEA-S 8K27 48-3712/R4 DHEA-S Read Highlighted Changes Revised May, 2009 Customer Service United States: 1-877-4ABBOTT International: Call your Abbott Representative This package insert must

More information

Quaternary Ammonium Compounds DOC316.53.01128

Quaternary Ammonium Compounds DOC316.53.01128 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower

More information

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit Catalog Number. CSB-E14405r For the quantitative determination of rat creatine kinase MM isoenzyme (CK-MM) concentrations in serum, plasma, tissue homogenates.

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:

More information

Factors Affecting Enzyme Activity

Factors Affecting Enzyme Activity INTRODUCTION Factors Affecting Enzyme Activity The chemical reactions occurring in living things are controlled by enzymes. An enzyme is a protein in the cell which lowers the activation energy of a catalyzed

More information

Albumin ELISA Kit. Albumin ELISA Kit. Zur in vitro Bestimmung des Albumin in Urin und Stuhl

Albumin ELISA Kit. Albumin ELISA Kit. Zur in vitro Bestimmung des Albumin in Urin und Stuhl ELISA Kit Super sensitiv Zur in vitro Bestimmung des in Urin und Stuhl ELISA Kit Super sensitive For the in vitro determination of in urine and stool Gültig ab / Valid from 06.03.2007 +8 C IMM-K 6330 96

More information

NITRIC OXIDE and NITROGEN DIOXIDE 6014

NITRIC OXIDE and NITROGEN DIOXIDE 6014 NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm

More information

Rat Fibronectin ELISA Kit

Rat Fibronectin ELISA Kit AssayMax TM Rat Fibronectin ELISA Kit Assaypro LLC 3400 Harry S Truman Blvd St. Charles, MO 63301 T (636) 447-9175 F (636) 395-7419 www.assaypro.com For any questions regarding troubleshooting or performing

More information

ph and Acidity in Wine and Fruit Juice

ph and Acidity in Wine and Fruit Juice ph and Acidity in Wine and Fruit Juice DOC316.52.93089 Endpoint potentiometric titration Scope and application: Wine and fruit juice 1. Introduction This working procedure refers to the titration of total

More information

Toxoplasma gondii IgM ELISA Kit Protocol

Toxoplasma gondii IgM ELISA Kit Protocol Toxoplasma gondii IgM ELISA Kit Protocol (Cat. No.:EK-310-87) 330 Beach Road, Burlingame CA Tel: 650-558-8898 Fax: 650-558-1686 E-Mail: info@phoenixpeptide.com www.phoenixpeptide.com Toxoplasma IgM ELISA

More information

Epinephrine/Norepinephrine ELISA Kit

Epinephrine/Norepinephrine ELISA Kit Epinephrine/Norepinephrine ELISA Kit Catalog Number KA3767 96 assays Version: 01 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay...

More information

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

The kit is a sandwich enzyme immunoassay for in vitro quantitative measurement of PST in mouse tissue homogenates and other biological fluids.

The kit is a sandwich enzyme immunoassay for in vitro quantitative measurement of PST in mouse tissue homogenates and other biological fluids. FOR IN VITRO AND RESEARCH USE ONLY NOT FOR USE IN CLINICAL DIAGNOSTIC PROCEDURES SEB268Mu 96 Tests Enzyme-linked Immunosorbent Assay Kit For Phenol Sulfotransferase (PST) Organism Species: Mus musculus

More information

Measuring Protein Concentration through Absorption Spectrophotometry

Measuring Protein Concentration through Absorption Spectrophotometry Measuring Protein Concentration through Absorption Spectrophotometry In this lab exercise you will learn how to homogenize a tissue to extract the protein, and then how to use a protein assay reagent to

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506 Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking

More information

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required

More information

TOTAL PROTEIN FIBRINOGEN

TOTAL PROTEIN FIBRINOGEN UNIT: Proteins 16tproteins.wpd Task Determination of Total Protein, Albumin and Globulins Objectives Upon completion of this exercise, the student will be able to: 1. Explain the ratio of albumin and globulin

More information

MEDICAL DEVICE GUIDANCE

MEDICAL DEVICE GUIDANCE May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained

More information

Bovine Vitamin B12 (VB12) ELISA Kit

Bovine Vitamin B12 (VB12) ELISA Kit Bovine Vitamin B12 (VB12) ELISA Kit Catalog Number. CSB-E14089B For the quantitative determination of endogenic bovine vitamin B12 (VB12) concentrations in serum, plasma, tissue homogenates. This package

More information

Reference Range: 0.5-1.6 mmol/l (arterial) 0.7-2.1 mmol/l (venous) CPT Code: 83605

Reference Range: 0.5-1.6 mmol/l (arterial) 0.7-2.1 mmol/l (venous) CPT Code: 83605 LACTIC ACID Fasting, arterial specimen preferred. Please note whether arterial or venous. 0.5 ml heparinized plasma. Green top or PST must be drawn, placed on ice, and spun within 15 minutes. Immediately

More information

RealLine HCV PCR Qualitative - Uni-Format

RealLine HCV PCR Qualitative - Uni-Format Instructions for use PCR KIT FOR EXTRACTION OF RNA AND REAL TIME PCR DETECTION KIT FOR HEPATITIS C VIRUS RNA Research Use Only Qualitative Uni Format VBD0798 48 tests valid from: December 2013 Rev11122013

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater

More information

Lab 2 Biochemistry. Learning Objectives. Introduction. Lipid Structure and Role in Food. The lab has the following learning objectives.

Lab 2 Biochemistry. Learning Objectives. Introduction. Lipid Structure and Role in Food. The lab has the following learning objectives. 1 Lab 2 Biochemistry Learning Objectives The lab has the following learning objectives. Investigate the role of double bonding in fatty acids, through models. Developing a calibration curve for a Benedict

More information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k103555 B. Purpose for Submission: New submission C. Measurand: Platelet aggregation D. Type of Test: Platelet aggregometer

More information

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: Thyroid Stimulating Hormone (TSH) Serum Access 2 (Beckman Coulter) Method No: Revised: as performed by: Collaborative Laboratory Services Ottumwa, Iowa

More information

Human Luteinizing Hormone (LH) Custom Kit

Human Luteinizing Hormone (LH) Custom Kit Human Luteinizing Hormone (LH) Custom Kit 18089-v2-2015Jan 1 MSD Toxicology Assays Human Luteinizing Hormone (LH) Custom Kit This package insert must be read in its entirety before using this product.

More information

for patient monitoring

for patient monitoring Thermo Scientific Indiko Reagents for Immunosuppressant Drug Testing Complete solutions for patient monitoring The cost-effective Thermo Scientific Indiko benchtop analyzer and bar-coded reagent is designed

More information

VACUETTE RNAgard Blood Tubes BioMaxi Blood RNA Purification Kit. For the collection, preservation and purification of RNA from whole blood

VACUETTE RNAgard Blood Tubes BioMaxi Blood RNA Purification Kit. For the collection, preservation and purification of RNA from whole blood RNAgard Blood System VACUETTE RNAgard Blood Tubes BioMaxi Blood RNA Purification Kit For the collection, preservation and purification of RNA from whole blood Instructions For Use Protocol blood collection,

More information

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2 Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method

More information

Carcinoembryonic Antigen (CEA)

Carcinoembryonic Antigen (CEA) Principle of the Test Carcinoembryonic Antigen (CEA) ADVIA Centaur System The ADVIA Centaur CEA assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts

More information

MTT Cell Proliferation Assay

MTT Cell Proliferation Assay ATCC 30-1010K Store at 4 C This product is intended for laboratory research purposes only. It is not intended for use in humans, animals or for diagnostics. Introduction Measurement of cell viability and

More information

3% Sodium Chloride Injection, USP 5% Sodium Chloride Injection, USP

3% Sodium Chloride Injection, USP 5% Sodium Chloride Injection, USP PRESCRIBING INFORMATION 3% Sodium Chloride Injection, USP 5% Sodium Chloride Injection, USP IV Fluid and Electrolyte Replenisher Baxter Corporation Mississauga, Ontario L5N 0C2 Canada Date of Revision:

More information

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry General Medicine Jiřina Crkovská (translated by Jan Pláteník) 2010/2011 1 Spectrophotometry is one of the most widely used instrumental

More information

Creatine Kinase Microplate Assay Kit User Manual

Creatine Kinase Microplate Assay Kit User Manual Creatine Kinase Microplate Assay Kit User Manual Catalog # CAK1045 Detection and Quantification of Creatine Kinase (CK) Activity in Urine, Serum, Plasma, Tissue extracts, Cell lysate, Cell culture media

More information

Immunoglobulin E (IgE) concentrations in Human. Immunoglobulin E (IgE) Human ELISA Kit

Immunoglobulin E (IgE) concentrations in Human. Immunoglobulin E (IgE) Human ELISA Kit ab108650 Immunoglobulin E (IgE) Human ELISA Kit Instructions for Use For the quantitative measurement of Immunoglobulin E (IgE) concentrations in Human serum. This product is for research use only and

More information

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8) Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:

More information

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory

More information

LAB 14 ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA)

LAB 14 ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) STUDENT GUIDE LAB 14 ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) GOAL The goal of this laboratory lesson is to explain the concepts and technique of enzyme linked immunosorbent assay (ELISA). OBJECTIVES

More information

Chapter 14 Urinalysis, Body Fluids and Other Specimens. Objectives:

Chapter 14 Urinalysis, Body Fluids and Other Specimens. Objectives: EXERCISE 15: CHEMICAL EXAMINATION OF URINE Textbook: Skill: Chapter 14 Urinalysis, Body Fluids and Other Specimens 15 points Objectives: 1. Name 10 routine chemical tests performed on urine and list a

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

Assay Qualification Template for Host Cell Protein ELISA

Assay Qualification Template for Host Cell Protein ELISA Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to

More information

Cancer Antigen CA 125

Cancer Antigen CA 125 Principle of the Test Cancer Antigen CA 125 ADVIA Centaur System The ADVIA Centaur CA 125 II assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses two monoclonal

More information