S8 Escape II. Clinical Guide. English POSITIVE AIRWAY PRESSURE DEVICE

Size: px
Start display at page:

Download "S8 Escape II. Clinical Guide. English POSITIVE AIRWAY PRESSURE DEVICE"

Transcription

1 S8 Escape II POSITIVE AIRWAY PRESSURE DEVICE Clinical Guide English ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) Danielson Street Poway CA USA ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK ResMed Offices Australia, Austria, Brazil, China, Finland, France, Germany, Hong Kong, India, Japan, Malaysia, Netherlands, New Zealand, Norway, Singapore, Spain, Sweden, Switzerland, UK, USA (see for contact details). Protected by patents: US , US , US Other patents pending. Protected by design registrations: AU , AU , AU , AU , AU , AU , AU , CN , CN , CN , CN , CN , CN , CN , EP , HK , JP , JP , JP , JP , JP , JP , NZ , NZ , NZ , NZ , NZ , NZ , NZ , US D544598, US D553228, US D557406, US D Others pending. S8, S8 Escape, HumidAire 3i, Bubble Cushion and ResScan are trademarks of ResMed Ltd and Escape is registered in U.S. Patent and Trademark Office ResMed Ltd /

2 Contents S8 Escape II Indications for Use 1 Contraindications 1 Adverse Effects 1 The S8 Escape II System Masks 2 Humidifiers 2 Data Management Products 2 Operating Information System Setup 3 Delivering Therapy 5 Navigating the S8 Escape II Menus S8 Escape II Keypad and LCD 6 S8 Escape II menus 7 Setting Ramp Time 10 EPR 10 Reminder Menu 11 Calibration 13 S8 Data Management Transmitting Data with the Data Card 17 S8 Adapters 18 Cleaning and Maintenance Daily 19 Weekly 19 Monthly 20 Replacing the Air Filter 20 Servicing 20 Troubleshooting System Specifications for S8 Escape II General Warnings and Cautions

3 S8 Escape II Indications for Use The S8 Escape II CPAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients > 66 lb (30 kg). The S8 Escape II CPAP system is intended for home and hospital use. CAUTION In the US, Federal law restricts this device to sale by or on the order of a physician. Contraindications The S8 Escape II is not a life support device and may stop operating with power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy. Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions: severe bullous lung disease pneumothorax pathologically low blood pressure dehydration cerebrospinal fluid leak, recent cranial surgery, or trauma. Adverse Effects Patients should report unusual chest pain, severe headache or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side-effects may arise during the course of therapy with the CPAP device: drying of the nose, mouth, or throat nosebleed bloating ear or sinus discomfort eye irritation skin rashes. S8 Escape II 1

4 The S8 Escape II System A complete S8 Escape II system comprises the following elements: S8 Escape II positive airway pressure device 2 m air tubing Power cord Carry bag. Carry handle Screen Keypad Data Card Module Air inlet Air outlet DC AC Optional components include: ResScan Data Card 3 m air tubing DC-12 converter. Masks Patients will also need a ResMed mask system (supplied separately). For information on using masks, see the relevant mask manual. Recommended masks are available on on the Products page under Service & Support. WARNING Only ResMed mask systems are compatible for use with the S8 Escape II. Please refer to General Warnings and Cautions on page 25. Humidifiers A list of humidifiers compatible with ResMed devices can be found on on the Products page under Service & Support. Data Management Products The S8 Escape II will have a ResScan Data Card module attached. See S8 Data Management on page 15. The following data management products may be purchased separately. ResScan Data Card Serial adapter USB adapter and cable ResMed PC application. 2

5 Notes: Ensure that at least one module or adapter is connected to the S8 Escape II at all times. ResMed regularly releases new products. Please check our website at Operating Information The S8 Escape II delivers pressure within the range 4 to 20 cm H 2 O. To make the beginning of treatment more comfortable, you may set a ramp so the pressure starts low and gradually increases to full treatment pressure over the ramp time. System Setup WARNING Make sure the power cord and plug or adapter cable are in good condition and the equipment is not damaged. Only ResMed air tubing should be used with the device. A different type of air tubing may alter the pressure the patient receives, reducing the effectiveness of the treatment. Blocking the hose and/or air inlet of the device while in operation could lead to overheating of the device. In the clinical environment any personal computer that is used with your CPAP system must be at least 5 (1.5 m) away from, or at least 8 (2.5 m) above, the patient. It must also comply with the relevant test standard. For personal computers the international standard is IEC or equivalent. CAUTION Be careful to place the device where it cannot be bumped, and where no one will trip over the power cord. The S8 Escape II device may be placed on the floor beside or underneath the patient s bed. Ensure that the area is free from dust and clear of bedding, clothes, or any other objects that could block the air inlet. When calibrating the S8 Escape II or setting the treatment pressure, assemble the device as the patient would normally use it, and connect all accessories that are normally used. Operating Information 3

6 Connect the power cord to the socket on the rear of the S8 Escape II. 2 Plug the other end of the power cord or adapter cable into the power outlet. 3 Connect one end of the air tubing firmly onto the air outlet. 4 Connect the assembled mask system to the free end of air tubing. Refer to the relevant mask manual for assembly instructions. 4

7 Delivering Therapy WARNING A mask should not be used unless the S8 Escape II is turned on and operating properly. 1 Power on the S8 Escape II. The LCD will display the standby (Ramp) screen. 2 Ensure patient settings are correct. 3 To start therapy, press. Air will begin flowing slowly and will build up to the operating pressure in about 15 to 25 seconds. 4 Instruct the patient to fit the mask according to the mask user instructions. 5 Instruct the patient to lie down in bed. If desired, arrange the air tubing over the top of the headboard to reduce drag on the mask. WARNING Do not leave long lengths of air tubing around the top of the bed; they may twist around the patient's head or neck while sleeping. Blocking the hose while in operation could lead to overheating of the device. 6 Adjust the mask and headgear to achieve optimal fit. At the end of the ramp period, the LCD will look similar to the screen below: CPAP To stop therapy, remove the mask, and press. Using the S8 Escape II on an aircraft The S8 Escape II complies with US FCC Part 15, Class B requirements. Patients should consult the medical services department of their carrier if they intend to use their S8 Escape II on an aircraft. Note: Do not use the S8 Escape II while the aircraft is taking off or landing. Operating Information 5

8 Navigating the S8 Escape II Menus S8 Escape II Keypad and LCD The control panel of the S8 Escape II unit includes a keypad for navigating through the menus and delivering treatment (Figure 1). LCD screen Keypad Start/Stop key Figure 1: S8 Escape II CPAP Keypad and LCD The LCD is used to display the menus and other data. The arrows and the text on the bottom line of the LCD prompt the use of the keys within the menus. The keypad is used to navigate through the menus and start/stop treatment. The functions of the keys and quick key combinations are listed in Table 1. Both the LCD and keypad are backlit with power-up of the device. When you are navigating through the menus, the LCD backlight times out two minutes after your last menu choice. Once has been pressed to initiate treatment, the LCD backlight stays on for two minutes. The keypad remains illuminated to indicate that the device is on. Table 1: Keypad keys and their functions Key(s) Up key Down key Start/Stop Function(s) Navigates between items within the menu. Allows you to increase settings options. Navigates between items within the menu. Allows you to decrease settings options. Starts and stops treatment. Left Performs the function indicated by the guiding text displayed above it on the LCD (eg, change). 6

9 Key(s) Right Down + Right + Left + Down + Function(s) Performs the function indicated by the guiding text displayed above it on the LCD (eg, exit). Returns you to the standby screen from the Clinical menu screen. Extended hold (for at least three seconds) gives access to the Clinical Menu from the standby screen. Extended hold (for at least three seconds) gives access to the Reminder Menu from the standby screen. S8 Escape II menus Parameters in the patient and clinical menus are described in Table 2 on page 9. Patient Menu Enter the Patient Menu by pressing while in the standby (Ramp) screen. Figure 2 summarizes the Patient Menu screens. S8 ESCAPE II S8 PATIENT CLASSIC MENU S8 RAMP CLASSIC Menu 10min EPR LEVEL: 1 Change USED HRS: 960 USAGE: 120/180 Only displayed if clinician enabled LANG: Change ENGLISH SW: SX Figure 2: Screens in the Patient Menu Clinical Menu Enter the Clinical Menu from the standby screen by holding down + simultaneously for at least three seconds. Figure 3 summarizes the Clinical Menu screens. Navigating the S8 Escape II Menus 7

10 S8 ESCAPE II S8 S8 PATIENT CLASSIC MENU S8 RAMP CLASSIC MENU Menu 10min S8 CLINICAL CLASSICMENU EPR: RAMP ONLY Change EPR LEVEL: 1 Change EPR ACCESS:OFF Change EPR INHALE: MED Change RAMP: Change MAX RAMP: Change 20min 45min START CPAP: 4.0 Change CPAP: 9.0 Change CALIBRATE: 1:00 Change USED HRS: 960 USAGE: 120/180 ERASE:COMPLIANCE Reset LANG: Change ENGLISH TIME: 23:10 Change DATE: SN: SW: SX MOTOR HRS: Figure 3: Screens in the Clinical Menu 8

11 Table 2: Parameters in the Patient and Clinical Menus Parameter EPR EPR Level a EPR Access EPR Inhale Ramp a Max Ramp Start CPAP CPAP Calibrate Used Hrs a Usage a,b Erase Language a Time Date SN SW a Motor Hours Description Sets EPR to occur all the time, during ramping only, or never. Sets the EPR value. Enables or disables patient view of, and ability to change, the EPR level. Sets the inhalation type as either Medium or Fast. Selects the Ramp time. (Available in standby mode.) Sets the maximum ramp time the patient can select. Sets the pressure at the start of the ramp time up to fixed treatment pressure. Sets the fixed treatment pressure. Adjusts the measured pressure at the manometer to match the set CPAP pressure. Use if there is a differential of greater than 0.5 cm H 2 O between the set CPAP pressure and the measured pressure. Changing the calibration will change the pressure gain displayed on the LCD. Entering the Calibrate screen will start airflow. ing this screen will stop airflow. See Calibration on page 13. Displays the total number of treatment hours. (View only.) The number of used sessions greater than 1 minute in length over the number of days elapsed since the first used day. Resets the used hours and used days. The language in which menus and settings are displayed on the LCD. Sets the current time. Displays the current date. Displays the serial number of the device. Displays the current software version. Displays the total number of hours the S8 Escape II motor has been running, to a maximum of 18,250 hours. a. Available on both the Patient Menu and the Clinical Menu. b. More detailed data is available via your PC application. Refer to the S8 Escape II Quick Set-up Guide for details of factory settings. Navigating the S8 Escape II Menus 9

12 Setting Ramp Time EPR If the patient has difficulty falling asleep with full pressure, select a ramp time. The pressure will slowly increase to full operating pressure over the selected ramp time. The Ramp Time screen is displayed after the standby (Ramp) screen, and is also available in the Clinical Menu. The clinician also sets a Maximum Ramp time; the patient may select any value up to the set maximum. When ramping begins the LCD will display the word RAMPING with the current pressure indicated as dashes. Table 3 describes the range of set pressures that the dashes represent. Table 3: Pressure range indicators on S8 LCD No. of dashes Pressure range (cm H 2 O) 2 Set Pressure 4.0 and < Set Pressure 5.0 and < Set Pressure 7.0 and < Set Pressure 9.0 and < Set Pressure 11.0 and < Set Pressure 13.0 and < Set Pressure 15.0 and < Set Pressure 17.0 and < Set Pressure 19.0 and = 20.0 During treatment, you can select EPR. EPR is designed to maintain optimal CPAP treatment for the patient during inhalation and reduce the delivered mask pressure during exhalation. The desired result of EPR is to decrease the pressure the patient must breathe out against, making the overall CPAP therapy more comfortable. The features of EPR are: You can select an EPR pressure drop level of 1, 2, or 3 cm H 2 O. You can set EPR to be off, delivered only during ramping, or delivered throughout therapy. There are two inhalation comfort levels; medium and fast. Pressure drop is limited, to avoid sub-optimal treatment (maximum drop is 3cm H 2 O). When EPR is enabled, the delivered pressure will not drop below a minimum pressure of 4 cm H 2 O, even if, for example, the set pressure is 6 cm H 2 O and the EPR level is 3. Clinicians can restrict a patient s access to their EPR level. 10

13 Reminder Menu You can use the Reminder menu to set a number of reminders to alert the patient to specific events; for example, when to replace their mask, when to insert a Data Card (if their device is Data Card enabled) and so on. When a reminder is due, a message is displayed on the LCD and remains whenever the device is not delivering therapy. The backlight on the LCD flashes when a message is displayed. If more than one reminder for a patient is scheduled for the same date, all scheduled reminders are displayed during that day. A patient can clear a message by pressing (or inserting a Data Card, in the case of the Data Card reminder). The default setting for all reminders is that they are disabled. Enter the Reminder Menu from the standby screen by pressing + for at least three seconds. To move through the menu, see Navigating the S8 Escape II Menus on page 6. Figure 4 summarizes the Reminder Menu screens on page 12. Replace mask You can set a timed reminder to remind a patient when they need to replace their mask. The patient can press (OK) to remove the message from the LCD. Call Provider You can set a timed reminder for your patient to phone you at a certain time; for example, to discuss how their therapy is going. The patient can press (OK) to remove the message from the LCD. Insert Card You can set a timed reminder on the S8 Escape II to remind the patient that they need to insert a Data Card to transfer their data. This enables you to establish compliance. The patient should actually insert the Data Card in order to clear the message from the LCD. (They can also press (OK) to remove the message.) Replace filter You can set a timed reminder to remind your patient when to replace the air filter. The patient can press (OK) to remove the message from the LCD. Service Due You can set a timed reminder to remind your patient when to return the device for service. The patient can press (OK) to remove the message from the LCD. Navigating the S8 Escape II Menus 11

14 S8 ESCAPE II S8 S8 PATIENT CLASSIC CLASSIC MENU S8 RAMP CLASSIC Menu MENU 10min S8 REMINDER CLASSIC MENU MASK: Change MASK RECUR: 6mth Change CALL: Change CALL RECUR: OFF Change CARD: Change CARD RECUR: OFF Change FILT: Change FILT RECUR: 6mth Change SERV: Change SERV RECUR:24mth Change Figure 4: Screens in the Reminder Menu 12

15 Table 4 shows the options and default settings for the types of reminders in the Reminder Menu. Table 4: Reminder type options and default settings Type of reminder Default setting a Options b Recurrence period Replace mask (MASK) OFF 7-day increments, starting from current date Call provider (CALL) OFF 7-day increments, starting from current date Insert Card (CARD) OFF 7-day increments, starting from current date Change filter (FILT) OFF 7-day increments, starting from current date Service due (SERV) OFF 7-day increments, starting from current date 1 24 months 1 24 months 1 24 months 1 24 months 1 24 months Setting Reminder Recurrence You can set each reminder to keep reoccuring at a specific time period, starting from the date set for the original reminder. You can set the recurrence time period from 1 to 24 months. (Each month in the recurrence time period is equal to 30 days.) You have to set a date for a specific reminder before its recurrence time period will take effect. For example, if you set a MASK (replace mask) reminder for 16 June 2008, the patient will see that reminder on 16 June If you have also set a MASK RECUR reminder for six months, the mask reminder will occur again six months (180 days) from the date the reminder is acknowledged by the patient. The mask reminder will continue to recur at six-monthly intervals unless you change the recurrence period, or set the mask reminder to OFF. Calibration a. OFF means that the date set for the reminder is today s date or earlier. b. If you have used a PC application to set a reminder date, and you later change that date using the S8 Escape II Reminder Menu, you can only change it in 7-day increments. The S8 Escape II is factory calibrated before delivery and will not require periodic calibration for home use. If there has been a change in altitude within a given altitude compensation range, or a change in circuit configuration, we recommend you check the calibration and adjust as required. Note: Some countries require all medical equipment to have a periodic service and calibration. Please check the regulations in your country. Navigating the S8 Escape II Menus 13

16 Quick Check To check the calibration of the S8 Escape II: 1. In the Clinical Menu, set the CPAP pressure setting to 12 cm H 2 O. 2. Measure the pressure at the mask using an accurate manometer. If the delivered pressure is less than 11.5 cm H 2 O or more than 12.5 cm H 2 O, you may need to recalibrate the device. Adjusting the Calibration To calibrate the S8 Escape II you will need a reference manometer with a range of 0 to 22 cm H 2 O in intervals of 0.1 cm H 2 O, fitted with a sensor tube to (2.5 mm to 3 mm) internal diameter and a standard Luer taper fitting. 1. Assemble the S8 Escape II as the patient would normally use it, complete with air tubing and the mask system. Connect all accessories (humidifier, filters, etc) that are normally used. 2. Turn on the S8 Escape II to allow the device to warm up. If the motor does not start running, press to start the airflow. Let the device warm up for 15 minutes. 3. While the device is warming up, remove the plug from one of the access ports on the mask, and attach the Luer taper end of the sensor tube. The other end of the sensor tube should be fitted to the manometer. If you have a blind Bubble Cushion, attach it to the mask. 4. When the motor has been running for 15 minutes, press to stop the airflow. 5. Access the Clinical Menu from the S8 Escape II screen by holding down + simultaneously for at least three seconds. 6. Scroll to the CPAP menu. 7. Press to change the pressure setting. Use and to set the pressure to 12 cm H 2 O. Press to apply this pressure. 8. Scroll to the Calibrate screen. Press (Change) to change the calibration setting. The airflow will start. If you are not using a blind Bubble Cushion, you can use the patient's normal mask and block the opening of the mask with the palm of your hand (however, be careful not to block the mask exhalation vent). 9. Read the pressure on the manometer. Use or to change the pressure until the manometer reads 12 ± 0.2 cm H 2 O. Note: You will notice that the pressure gain increases when is pressed and decreases when is pressed. 10.Press (Change) to apply this pressure. The airflow will stop. 14

17 WARNING During calibration, you may not be able to calibrate the S8 Escape II for all patients under extreme conditions (eg, high set pressures and high altitudes). This is because the delivered pressure will reach a peak and will not increase any further, even though the pressure gain may be increasing. In this case, we advise that you do not increase the pressure gain once the pressure has peaked; otherwise the pressure may be too high if operated at lower altitudes. S8 Data Management Before issuing the Data Card, make sure that it has been labeled with the patient s ID and is correctly addressed. Refer to Using the Protective Folder on page 16 for more details. Check that the patient knows: why they are being issued with a Data Card when and how they must use the Data Card how to insert, remove and return the Data Card where to look for information (the Troubleshooting section of the User s Manual for their S8 device) if the data transfer does not work as expected. Labeling the Data Card We strongly advise that you write the name of the patient on the Data Card to prevent any mix-up of patient data. A space is provided on the Data Card for this purpose. Each Data Card is reusable. If you are issuing an existing Data Card to a new patient, make sure that the card is labeled with the new patient s ID. Use the space provided on the Data Card label when issuing a new card. Notes: When reusing a Data Card, make sure that you use a label that does not cover more than the name area on the Data Card, and is not more than 0.06 (0.15 mm) thick. If the label is larger than this, the patient may not be able to insert the Data Card into the Data Card module on their S8 Escape II. If you reuse a Data Card that already has a label on it, remove the old label before attaching a new one. Supplying the Data Card You can monitor a patient s treatment with the Data Card and protective folder supplied with their device. The patient can return the card to you when they have copied their data onto it. If you are mailing the Data Card, you need to advise the patient when the card is being mailed out to them. Ensure that the patient knows that they must remove their Data Card from the device when they have finished copying data to it. Leaving the Data Card in the device during therapy will not result in additional data being copied to the card. If a patient needs to copy additional data to the Dard Card, they will need to insert the card again. S8 Data Management 15

18 If you want a patient to update the settings in their S8 Escape II, you can load the new settings on to a Data Card and either give or mail them the card. The patient can return the card to you when they have applied the settings on their S8 device. The information on the card will allow you to assess whether the patient s settings have been changed as expected. Notes The Data Card should be stored in its protective folder when not in use. Use a postal envelope to mail the protective folder and its contents to the patient. Using the Protective Folder The protective folder should always be used to store the Data Card and for mailing. The protective folder contains all the necessary instructions for a patient to copy data or update their settings (as shown in Figure 5). Dear patient, please follow the directions as marked below. FR Cher/Chère patient(e), veuillez suivre les instructions ci-dessous. ES Estimado paciente, tenga a bien seguir las instrucciones que se indican a continuación. PT Caro paciente, siga as instruções abaixo indicadas. COPY DATA COPIEZ LES DONNEES COPIADO DE DATOS COPIAR DADOS FR ES PT FR CARTE INSEREE Patientez SVP ES TARJ. INSERTADA Espere PT CARTAO INSERI Aguarde p.f. FR ES PT COPIE TERMINEE Retirez carte COPIA COMPLETA Quitar tarjeta COPIA COMPLET Retirar cartão Instructions for copying data 5 Return by mail FR Retournez par courrier ES Devolver por correo PT Devolva o cartão por correio Return by person FR Retournez la carte personnellement ES Devolver en persona PT Devolva o cartão pessoalmente Keep the card FR Gardez la carte ES Conservar la tarjeta PT Fique com o cartão Indicate what you want the patient to do Data Card goes here UPDATE SETTINGS METTEZ LES REGLAGES A JOUR ACTUALIZACIÓ N DE LA CONFIGURACIÓN ACTUALIZAR PARÂMETROS FR ES PT FR ES PT CARTE INSEREE Patientez SVP TARJ. INSERTADA Espere CARTAO INSERI Aguarde p.f. FR ES PT CHARG.REUSSI Retirez carte CONFIG CORRECTA Quitar tarjeta PARAMET SUCESSO Retirar cartão Instructions for updating settings If messages differ, see your manual. Use Data Card as per above before returning. FR Si les messages affichés sont différents, référez-vous à votre manuel. Utilisez la carte de données conformément aux instructions ci-dessus avant de la retourner. ES Si los mensajes son distintos, consulte el manual. Antes de devolverla, utilice la tarjeta tal como se indica. PT Se as mensagens forem diferentes, consulte o seu manual. Use o cartão de dados tal como é descrito acima antes de o devolver. Figure 5: Data Card s protective folder 16

19 Transmitting Data with the Data Card Copying Data onto a Data Card If you want a patient to give you their compliance data, you need to tell them how to copy data onto their Data Card. This information is also supplied in the patient s User Guide. 1. Ensure the S8 Escape II is switched on and displaying the standby (Ramp) screen. 2. Insert the Data Card (arrow facing up) into the Data Card Module until it stops. Copying of data starts automatically. The Card Inserted Please Wait message is displayed on the LCD while data is copied to the Data Card. Copying takes approximately five seconds. The Copy Complete Remove Card message is displayed on the LCD when copying has finished. 3. Remove the Data Card by gripping the end of the Data Card and pulling it out. The Data Card should be stored in its protective folder when not in use. 4. If mailing the Data Card to a patient, use a postal envelope to send the Data Card in its protective folder. Updating Settings on the S8 Escape II You can create new treatment settings and transfer them to a Data Card via your ResMed PC application. A patient can then use the Data Card to update the settings in their S8 Escape II. If you mail the Data Card to the patient, send it in the protective folder. If you need the Data Card sent back to you for verification, use a postal envelope to mail the protective folder and its contents to the patient. Instruct the patient to carry out the following steps to update settings on their S8 Escape II. This information is also supplied in the patient s User Guide. 1. With the device in standby (Ramp) mode, insert the Data Card into the slot on the Data Card module. Updating will start automatically. The Card Inserted Please Wait message will appear on the LCD while updating is in progress. Updating takes approximately five seconds. The Settings Success Remove Card message is displayed on the LCD if the settings were updated successfully. Note: This message only appears once. If you re-insert the Data Card after you have updated your settings, this message is not displayed. 2. Remove the Data Card from the device. 3. Store the Data Card in its protective folder when not in use. When you create a Data Card with therapy settings you can specify the serial number of the device on which the settings need to be updated. This prevents the settings from being applied on the wrong device. Refer to your PC application manual for more details. WARNING Advise the patient to contact you immediately if they are using the Data Card to update the settings on their device and the Settings Success message does not appear. S8 Data Management 17

20 Analyzing the Data Card Use a ResMed PC application to transfer data and settings between an S8 Escape II or a Data Card and your personal computer. Refer to the user guide for your PC application for more information about analysing the information on returned Data Cards. S8 Adapters S8 adapters are designed for use with the ResMed S8 series of devices, to transfer data between the device and a personal computer. They slot onto the device in place of the Data Card Module. Two adapters are available: Serial adapter, which uses a standard serial cable USB adapter (supplied with a USB cable). Data Card Module Serial adapter USB adapter WARNINGS Only use S8 adapters as specified in this manual. S8 adapters are specifically designed for use with ResMed S8 devices. Attach the S8 serial or USB adapter to a ResMed S8 device to create a serial or USB port. You can then connect a cable that enables transfer of data between the device and a personal computer. CAUTION If a cable is not attached when the S8 adapter is attached to an S8 device, either attach the adapter s protective cover or remove the adapter and replace it with the Data Card module. This is to prevent damage to the S8 adapter. Attaching an S8 adapter Notes: Ensure that a module or an adapter is attached to the S8 device at all times. Remove the existing module from the S8 device if the serial or USB adapter is required. The device should be turned off when removing a module and attaching the serial adapter. 18

S8 Elite II System Clinical Guide

S8 Elite II System Clinical Guide S8 Elite II System Clinical Guide English ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857

More information

S9 Series AutoSet & Elite

S9 Series AutoSet & Elite S9 Series AutoSet & Elite POSITIVE AIRWAY PRESSURE DEVICEs User Guide A 3 2 5 4 1 S9 AutoSet indications for use The S9 AutoSet self-adjusting system is indicated for the treatment of obstructive sleep

More information

S9 Series AutoSet & Elite

S9 Series AutoSet & Elite B 1 2 3 4 S9 Series AutoSet & Elite POSITIVE AIRWAY PRESSURE DEVICEs User Guide English Deutsch Français Italiano Español Português Nederlands Ελληνικά Polski Magyar Pусский Türkçe Česky Slovensky C 5

More information

Mirage FX NASAL MASK. User Guide

Mirage FX NASAL MASK. User Guide Mirage FX NASAL MASK User Guide Mirage FX NASAL MASK Fitting// 1 2 3 4 5 6 Mirage FX i Removal// 1 2 Disassembly// 1 2 3 ii Reassembly// 1 2 3 4 iii Mirage FX A 2 3 1 4 7 6 5 B iv Mask components// Item/

More information

Z1 CPAP System CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide

Z1 CPAP System CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide Z1 CPAP System CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide Human Design Medical, LLC User Guide English Contents Welcome to Your Z1 CPAP System.... 2 Owner/User Responsibility... 2 Replacement

More information

Mirage FX. User Guide. English NASAL MASK

Mirage FX. User Guide. English NASAL MASK Mirage FX NASAL MASK User Guide English Did you know? Replacing your mask and supplies is one of the most important steps in maintaining effective therapy. Replace when... The cushion is no longer clear

More information

Please read the entire Information and Welcome Guides before using your device.

Please read the entire Information and Welcome Guides before using your device. S9 AutoSet / S9 AutoSet for Her / S9 Elite / S9 Escape / S9 Escape Auto POSITIVE AIRWAY PRESSURE DEVICES H5i / H5i for Her HEATED HUMIDIFIER ClimateLine / ClimateLine MAX / SlimLine / Standard AIR TUBING

More information

Introduction. Intended use. Contraindications. General warnings and cautions. Patient Instructions WARNING CAUTION. English

Introduction. Intended use. Contraindications. General warnings and cautions. Patient Instructions WARNING CAUTION. English Patient Instructions English Introduction ApneaLinkAir ApneaLink Plus Your physician or healthcare provider has given you the ApneaLink Air device to test if you have sleep-disordered breathing. The device

More information

Z1 Auto System AUTO-ADJUSTING CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide

Z1 Auto System AUTO-ADJUSTING CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide Z1 Auto System AUTO-ADJUSTING CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide Human Design Medical, LLC AN AFFILIATE OF BREAS MEDICAL AB User Guide English Contents Welcome to Your Z1 Auto System....2

More information

SPECIAL NOTES SPECIAL NOTES DANGER/WARNING/CAUTION notices as used in this manual apply to hazards or unsafe practices which could result in personal

SPECIAL NOTES SPECIAL NOTES DANGER/WARNING/CAUTION notices as used in this manual apply to hazards or unsafe practices which could result in personal ASSEMBLY, INSTALLATION, AND OPERATING INSTRUCTIONS Invacare Twilight Nasal Mask and Headgear MODEL NO: ISP2000 STANDARD LATEX FREE DEALER: THIS MANUAL MUST BE GIVEN TO THE END USER. USER: BEFORE USING

More information

CPAP Therapy and You

CPAP Therapy and You CPAP Therapy and You CPAP Treatment CPAP (pronounced "see-pap") is short for "continuous positive airway pressure." Positive airway pressure therapy is the most effective noninvasive treatment for Obstructive

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CCS Customer Support : 1-800-474-2147 Voltage rating: 115V~60Hz Power rating : 1400W For product inquiries or support

More information

No Other Sleep Therapy System Delivers More.

No Other Sleep Therapy System Delivers More. COMFORT STYLE INNOVATION No Other Sleep Therapy System Delivers More. Only the ResMed S9 Series Gives You More Comfort, Style and Sleep. About sleep apnea Obstructive sleep apnea (OSA) is a common form

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MN12CES / MN10CESWW Voltage rating: 120V~60Hz Power rating : 1100W (MN12CES) Power rating : 900W (MN10CESWW) Customer

More information

Owner s Manual. Model No.: 9S-005 series Item No. 7501 Please read the instruction manual before use

Owner s Manual. Model No.: 9S-005 series Item No. 7501 Please read the instruction manual before use Owner s Manual Model No.: 9S-005 series Item No. 7501 Please read the instruction manual before use Contents Important Safeguards....1 1. Introduction.......3 2. Product Description.......4 3. Installation......5

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries

More information

www.respironics.com CPAP System User Manual 1007167 LK 5/12/01

www.respironics.com CPAP System User Manual 1007167 LK 5/12/01 www.respironics.com CPAP System User Manual 1007167 LK 5/12/01 Table of Contents Symbol Key... 2 Intended Use... 3 Specifications... 4 Warnings & Cautions... 5 Unpacking the System... 6 System Features...

More information

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator

More information

User guide User guide English

User guide User guide English User guide English ENGLISH Welcome The AirSense 10 AutoSet and AirSense 10 AutoSet for Her are ResMed's premium auto-adjusting pressure devices. The AirSense 10 Elite and AirSense 10 CPAP are ResMed's

More information

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting

More information

SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / 8353. Instruction Manual. Page 1 of 9

SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / 8353. Instruction Manual. Page 1 of 9 SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / 8353 Instruction Manual Page 1 of 9 TABLE OF CONTENTS Introduction...3 Important Safeguards..3 System Components...5 Setup and Operation...5 Setup...5 Operation....6

More information

Taurus Super-S3 LCM. Dual-Bay RAID Storage Enclosure for two 3.5-inch Serial ATA Hard Drives. User Manual March 31, 2014 v1.2 www.akitio.

Taurus Super-S3 LCM. Dual-Bay RAID Storage Enclosure for two 3.5-inch Serial ATA Hard Drives. User Manual March 31, 2014 v1.2 www.akitio. Dual-Bay RAID Storage Enclosure for two 3.5-inch Serial ATA Hard Drives User Manual March 31, 2014 v1.2 www.akitio.com EN Table of Contents Table of Contents 1 Introduction... 1 1.1 Technical Specifications...

More information

Table of Contents. REMstar Auto User Manual 1

Table of Contents. REMstar Auto User Manual 1 Table of Contents Symbol Key... 2 Intended Use & Contraindications... 3 Specifications... 4 Warnings & Cautions... 5 System Features... 7 First Time Setup... 8 Installing the Filter... 9 Device Settings...

More information

AutoSet & Elite. H5i. Clinical Guide. English POSITIVE AIRWAY PRESSURE DEVICES HEATED HUMIDIFIER

AutoSet & Elite. H5i. Clinical Guide. English POSITIVE AIRWAY PRESSURE DEVICES HEATED HUMIDIFIER AutoSet & Elite POSITIVE AIRWAY PRESSURE DEVICES H5i HEATED HUMIDIFIER Clinical Guide English Contents Welcome... 1 S9 AutoSet indications for use... 1 S9 Elite indications for use... 1 S9 contraindications...

More information

Meaco 30L and Meaco 40L dehumidifier instruction manual

Meaco 30L and Meaco 40L dehumidifier instruction manual Meaco 30L and Meaco 40L dehumidifier instruction manual Please read this instruction manual before using the dehumidifier and keep safe for future reference SAFETY INSTRUCTIONS PLEASE READ ALL INSTRUCTIONS

More information

Use and Care Manual. Model CPA12KH AIR CONDITIONER

Use and Care Manual. Model CPA12KH AIR CONDITIONER Use and Care Manual Model CPA12KH AIR CONDITIONER Introduction Thank you for choosing this air conditioner to provide you and your family with all of the "Home Comfort" requirements for your home, cottage

More information

BiPAP Pro Bi-Flex USER MANUAL

BiPAP Pro Bi-Flex USER MANUAL BiPAP Pro Bi-Flex USER MANUAL 2010 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Important... 2 Warnings... 2 Cautions... 3 Contraindications... 3 Symbol

More information

Compact Nasal CPAP Model 7400. System Patient Operating Instructions. Compact CPAP System

Compact Nasal CPAP Model 7400. System Patient Operating Instructions. Compact CPAP System 7400-00B.QXD 10/21/99 8:03 AM Page 1 The Respironics Tranquility Quest Plus is a medical device prescribed by a physician to assist breathing. Compact CPAP System Compact Nasal CPAP Model 7400 System Patient

More information

BiPAP Auto Bi-Flex USER MANUAL

BiPAP Auto Bi-Flex USER MANUAL BiPAP Auto Bi-Flex USER MANUAL 2010 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Important... 2 Warnings... 2 Cautions... 3 Contraindications... 3 Symbol

More information

308 Nasal CPAP 308 / Page 1 Of 7

308 Nasal CPAP 308 / Page 1 Of 7 308 Nasal CPAP 308 / Page 1 Of 7 Description CPAP via nasal mask may be used to relieve upper airway obstruction. Nasal CPAP is primarily used with patients who suffer from sleep apnea disorders. It is

More information

System One Heated Humidifier USER MANUAL

System One Heated Humidifier USER MANUAL System One Heated Humidifier USER MANUAL 2011 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Warnings... 2 Cautions... 2 Contraindications... 3 Symbol Key...

More information

Portable Air Conditioner

Portable Air Conditioner Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE

More information

FT-100 Fire Training Fog Machine User Manual English

FT-100 Fire Training Fog Machine User Manual English FT-100 Fire Training Fog Machine User Manual English 2015 Antari Lighting and Effects Ltd. User Manual - English Safety Information Please read the following safety information carefully before operating

More information

S8 Elite System USER S MANUAL. English. 338116-Eng/1 05 05

S8 Elite System USER S MANUAL. English. 338116-Eng/1 05 05 S8 Elite System USER S MANUAL English 338116-Eng/1 05 05 B INTEGRATED HUMIDIFIER / HUMIDIFICATEUR INTEGRE / HUMIDIFICADOR INTEGRADO / HUMIDIFICADOR INTEGRADO D SETTING UP THE S8 TM / INSTALLATION DE LA

More information

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE

More information

FOR THE FOLLOWING MODELS: EE-8075W EE-8075O EE-8075R EE-8075BK

FOR THE FOLLOWING MODELS: EE-8075W EE-8075O EE-8075R EE-8075BK FIREPLACE HEATER FOR THE FOLLOWING MODELS: EE-8075W EE-8075O EE-8075R EE-8075BK If you have any questions about the operation of your fireplace heater, please contact Crane Customer Care. Toll Free: 888-599-0992

More information

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS ELECTRICAL SAFETY This appliance is for indoor use only. Always turn off the unit and unplug

More information

CPAP System This CPAP system is intended for the treatment of adult obstructive Sleep Apnea only. NOTE: The C-Flex mark is used under license.

CPAP System This CPAP system is intended for the treatment of adult obstructive Sleep Apnea only. NOTE: The C-Flex mark is used under license. Home Care Provider Setup Instructions Always use these instructions along with the User Manual when assembling or adjusting this equipment. For clinical systems, refer to the setup guide entitled Respironics

More information

REMstar SE USER MANUAL

REMstar SE USER MANUAL REMstar SE USER MANUAL 2012 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Important... 2 Warnings... 2 Cautions... 3 Contraindications... 3 Symbol Key...

More information

ResMed Masks. Clinical Use Only. Disinfection and Sterilization Guide. NASAL MASKS Mirage Activa. Mirage Kidsta. Mirage Vista.

ResMed Masks. Clinical Use Only. Disinfection and Sterilization Guide. NASAL MASKS Mirage Activa. Mirage Kidsta. Mirage Vista. ResMed Masks and Sterilization Guide S Mirage Activa Mirage Kidsta Mirage Vista Ultra Mirage II Ultra Mirage NON-VENTED NASAL PILLOWS SYSTEMS Mirage Swift II NEW NASAL PILLOWS SYSTEM Mirage Swift NASAL

More information

TABLE OF CONTENTS QUICK START... 1 DIAGRAM OF THE DIGITAL PHOTO FRAME... 2 KEYPAD BUTTON FUNCTIONS... 3 REMOTE CONTROL FUNCTIONS..

TABLE OF CONTENTS QUICK START... 1 DIAGRAM OF THE DIGITAL PHOTO FRAME... 2 KEYPAD BUTTON FUNCTIONS... 3 REMOTE CONTROL FUNCTIONS.. TABLE OF CONTENTS QUICK START........ 1 DIAGRAM OF THE DIGITAL PHOTO FRAME... 2 KEYPAD BUTTON FUNCTIONS..... 3 REMOTE CONTROL FUNCTIONS..4 DIGITAL PHOTO FRAME NAVIGATION........5 OSD MENU FUNCTIONS..6

More information

H4i. User Guide HEATED HUMIDIFIER. English

H4i. User Guide HEATED HUMIDIFIER. English H4i HEATED HUMIDIFIER User Guide English English H4i HEATED HUMIDIFIER User Guide English Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia. Distributed by: ResMed Corp

More information

Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan

Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan TABLE OF CONTENTS INTRODUCTION 2 IMPORTANT SAFEGUARDS...2 PACKAGE CONTAINS..2 NAMES OF PARTS.3

More information

DEHUMIDIFIER MODELS: SD-30E / SD-40E / SD-60E INSTRUCTION MANUAL

DEHUMIDIFIER MODELS: SD-30E / SD-40E / SD-60E INSTRUCTION MANUAL DEHUMIDIFIER MODELS: SD-30E / SD-40E / SD-60E INSTRUCTION MANUAL Please read these instructions thoroughly and keep it in a safe place for future reference. CONTENTS FUNCTIONS 2 SAFETY NOTES 2 NAME OF

More information

BiPAP autosv Advanced System One USER MANUAL

BiPAP autosv Advanced System One USER MANUAL BiPAP autosv Advanced System One USER MANUAL 2010 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Warnings... 2 Cautions... 2 Contraindications... 3 Patient

More information

OWNERS MANUAL. PAC 210 U Portable Windowless Air-Conditioner

OWNERS MANUAL. PAC 210 U Portable Windowless Air-Conditioner OWNERS MANUAL PAC 210 U Portable Windowless Air-Conditioner 1 IMPORTANT INFORMATION When using electrical appliances, basic safety precautions should always be followed, including the following: - Read

More information

Portable Evaporative Air Cooler. OWNER S MANUAL Read and save these instructions before use. Model: CL48PM

Portable Evaporative Air Cooler. OWNER S MANUAL Read and save these instructions before use. Model: CL48PM Portable Evaporative Air Cooler OWNER S MANUAL Read and save these instructions before use Model: CL48PM Power rating: 130 Watts Voltage rating: 230 Volt, 50Hz Made in P.R.C. QUICK START GUIDE Fill with

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CHCSCS

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CHCSCS Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CHCSCS Voltage rating: 120V~60Hz Power rating : 1400W(Cooling) Power rating : 1350W(Heating) Customer Support : 1-800-474-21477

More information

HumidAire 2i User Manual

HumidAire 2i User Manual English HumidAire 2i User Manual English ResMed Ltd (Manufacturer) 1 Elizabeth MacArthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857

More information

car + boat Box Contents x2 x2 22350_DC16_CAR_BOAT_OPSMAN.indd 1 29/7/08 15:27:01

car + boat Box Contents x2 x2 22350_DC16_CAR_BOAT_OPSMAN.indd 1 29/7/08 15:27:01 car + boat Box Contents x x 350_DC6_CAR_BOAT_OPSMAN.indd 9/7/08 5:7:0 SAVE THESE INSTRUCTIONS THIS APPLIANCE IS INTENDED FOR HOUSEHOLD USE ONLY IMPORTANT SAFETY INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE

More information

Important Safeguards

Important Safeguards Table of Contents Important Safeguards...2 Product Layout...3 Preparing for Use...4 Air-conditioning without installation...4 Air-conditioning with installation...5 Control Panel...6 Operating from the

More information

GENANO 4500 MEDICAL AIR PURIFICATION DEVICE

GENANO 4500 MEDICAL AIR PURIFICATION DEVICE GENANO AIR PURIFICATION TECHNOLOGY GENANO 4500 MEDICAL AIR PURIFICATION DEVICE OPERATING INSTRUCTION 1 GENANO LTD Kimmeltie 3 FI- 02110 Espoo, Finland Tel + 358 9 7743 870 Fax + 358 9 7533 130 www.genano.fi

More information

NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS BEFORE USE GENERAL SAFETY INSTRUCTIONS: ALWAYS OPERATE THE UNIT IN AN UPRIGHT POSITION AND PLACE

More information

USER S MANUAL FH052EAV1 FH070EAV1. System Air Conditioner (Cooling and Heating) ENGLISH ESPAÑOL FRANÇAIS ITALIANO PORTUGUÊS DEUTSCH E HNIKA

USER S MANUAL FH052EAV1 FH070EAV1. System Air Conditioner (Cooling and Heating) ENGLISH ESPAÑOL FRANÇAIS ITALIANO PORTUGUÊS DEUTSCH E HNIKA USER S MANUAL FH052EAV1 FH070EAV1 E HNIKA PORTUGUÊS ENGLISH ESPAÑOL ITALIANO DEUTSCH FRANÇAIS System Air Conditioner (Cooling and Heating) E S F I P D G DB98-29263A(1) Safety Precautions Register your

More information

REMstar Pro C-Flex+ USER MANUAL

REMstar Pro C-Flex+ USER MANUAL REMstar Pro C-Flex+ USER MANUAL 2012 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Important... 2 Warnings... 2 Cautions... 3 Contraindications... 3 Symbol

More information

Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00

Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00 E Refrigerant Charging Unit ICOGD 020AH1000 Operating Manual FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00 Contents Technical Specifications... 20 Safety... 21 A/C System... 22 Components... 23 Control Panel...

More information

CPAP/BiPAP Patient Instructions

CPAP/BiPAP Patient Instructions CPAP/BiPAP Patient Instructions 115 John Maddox Dr Rome, GA 30165 Phone: 706-266-4086 Kevin Cooper, Owner/Respiratory Therapist Purpose of CPAP/BiPAP Therapy The purpose of Continuous Positive Airway Pressure

More information

Power Pro Elite. - Power Pro Elite - Alternating Pressure Mattress System with True Low Air Loss. User Manual

Power Pro Elite. - Power Pro Elite - Alternating Pressure Mattress System with True Low Air Loss. User Manual - Power Pro Elite - Alternating Pressure Mattress System with True Low Air Loss User Manual 1 Blake Medical Distribution 38 Lincoln Street Hamilton, ON L8L 7L5 Ph.: 1-866-662-3459 Fax: 905-560-1342 Warning:

More information

AW10 All-in-One Printer

AW10 All-in-One Printer START OK All-in-One Printer AW10 User Guide AW10 All-in-One Printer ADVENT AW10 All-in-One Printer Your ADVENT Printer combines ease of use and KODAK Color Technology to print crisp-looking documents and

More information

VALLEY Respiratory Services, Inc. Where Patients Are Our Priority

VALLEY Respiratory Services, Inc. Where Patients Are Our Priority VALLEY Respiratory Services, Inc. Where Patients Are Our Priority CPAP/BIPAP Overview What it is: A CPAP or BIPAP machine delivers air pressure through a mask while a patient sleeps. CPAP is pressure to

More information

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS + Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS Inventory Safety and General Information bu001c This unit is intended for indoor use only. Do not operate this unit in direct sunlight,

More information

Dehumidifier Users manual. For Models: DH45S DH65S

Dehumidifier Users manual. For Models: DH45S DH65S Dehumidifier Users manual For Models: DH45S DH65S 950-0062-revD Jan. 9 2007 FORWARD The appearance of the units that you purchase might be slightly different from the ones described in the Manual, but

More information

REMstar Auto A-Flex USER MANUAL

REMstar Auto A-Flex USER MANUAL REMstar Auto A-Flex USER MANUAL 2012 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Important... 2 Warnings... 2 Cautions... 3 Contraindications... 3 Symbol

More information

Carpet Washer. vax.co.uk VRS5W. Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308. Vax model number: Version 1.0

Carpet Washer. vax.co.uk VRS5W. Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308. Vax model number: Version 1.0 VRS5W Powermax User Guide V1.0.qxd:V1.0 23/7/10 15:35 Page 1 Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308 Carpet Washer Vax model number: VRS5W instruction manual Version 1.0 Please read carefully

More information

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05 RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,

More information

Molecular Imager Gel Doc XR+ Installation Guide

Molecular Imager Gel Doc XR+ Installation Guide Molecular Imager Gel Doc XR+ Installation Guide To install your Gel Doc XR+, perform these seven steps in the following order: 1. Unpack the Gel Doc XR+ system. 2. Install Image Lab software. 3. Set up

More information

Table of Contents. isleep 20i users manual

Table of Contents. isleep 20i users manual Table of Contents 1 Intended Use of the isleep 20i... 3 1.1 What is the isleep 20i?... 3 1.2 Indications for Use... 4 1.3 Contraindications... 4 1.4 About this Manual... 5 2 Safety Information... 7 2.1

More information

Tablet Management Rack 16

Tablet Management Rack 16 Tablet Management Rack 16 User's Guide Table of Contents Hazard Symbols Review... 2 Components... 2 Charging Set-up... 3-4 LED and Syncing... 4-5 Cleaning and Maintenance... 6 Specifications... 6 Service

More information

BiPAP Pro Bi-Flex USER MANUAL

BiPAP Pro Bi-Flex USER MANUAL BiPAP Pro Bi-Flex USER MANUAL 2012 Koninklijke Philips Electronics N.V. All rights reserved. Table of Contents Intended Use... 2 Important... 2 Warnings... 2 Cautions... 3 Contraindications... 3 Symbol

More information

MAKING MODERN LIVING POSSIBLE. UniLynx Indoor User Manual. ULX 1800i ULX 3000i ULX 3600i ULX 5400i SOLAR INVERTERS

MAKING MODERN LIVING POSSIBLE. UniLynx Indoor User Manual. ULX 1800i ULX 3000i ULX 3600i ULX 5400i SOLAR INVERTERS MAKING MODERN LIVING POSSIBLE UniLynx Indoor User Manual ULX 1800i ULX 3000i ULX 3600i ULX 5400i SOLAR INVERTERS Contents Contents 1. Introduction 2 Introduction 2 2. Function Description 3 Definition

More information

microlife PF 100 Electronic Asthma Monitor Instruction Manual (1-15) Moniteur d'asthme électronique Mode d'emploi (16-31) Monitor electrónico para el

microlife PF 100 Electronic Asthma Monitor Instruction Manual (1-15) Moniteur d'asthme électronique Mode d'emploi (16-31) Monitor electrónico para el microlife PF 100 Electronic Asthma Monitor Instruction Manual (1-15) Moniteur d'asthme électronique Mode d'emploi (16-31) Monitor electrónico para el asma Manual de instrucciones (32-47) Elektronischer

More information

HP ProDisplay P17A LED Backlit Monitor. User Guide

HP ProDisplay P17A LED Backlit Monitor. User Guide HP ProDisplay P17A LED Backlit Monitor User Guide 2013 Hewlett-Packard Development Company, L.P. Microsoft, Windows, and Windows Vista are either trademarks or registered trademarks of Microsoft Corporation

More information

Laptop Power Adapter for home/office User Manual

Laptop Power Adapter for home/office User Manual Laptop Power Adapter for home/office User Manual Introduction Congratulations and thank you for purchasing the Laptop Power Adapter for home/office from Belkin. This new Adapter provides a single source

More information

Congratulations! Thank you!

Congratulations! Thank you! TM-47 ORDERCODE D1370 Congratulations! You have bought a great, innovative product from DAP Audio. The DAP Audio Microphone range brings excitement to any venue. Whether you want simple plug-&-play action

More information

Owner s manual. Free 5 year parts, 1 year labour gurantee Register your guarantee at www.dysonairblade.com

Owner s manual. Free 5 year parts, 1 year labour gurantee Register your guarantee at www.dysonairblade.com Owner s manual Free 5 year parts, 1 year labour gurantee Register your guarantee at www.dysonairblade.com Thank you for choosing to buy a Dyson Airblade MK2 hand dryer Dyson customer care Your hand dryer

More information

FCC COMPLIANCE STATEMENT FOR AMERICAN USERS

FCC COMPLIANCE STATEMENT FOR AMERICAN USERS FCC COMPLIANCE STATEMENT FOR AMERICAN USERS This equipment has been tested and found to comply with the limits for a CLASS A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed

More information

USER S MANUAL HSC-24A

USER S MANUAL HSC-24A AIRREX AIR CONDITIONER USER S MANUAL HSC-24A Thank you for purchasing an AIRREX AIR CONDITIONER. BEFORE operation please read this user s manual carefully. Keep this manual readily available. It is ESSENTIAL

More information

Digital Photo Picture Frame. Operation Manual

Digital Photo Picture Frame. Operation Manual Digital Photo Picture Frame Operation Manual 20070309 CONGRATULATIONS on your purchase of a Polaroid 7 LCD Digital Photo Picture Frame. Please read carefully and follow all warnings and instructions in

More information

Operating instructions in the back. www.blackanddecker.co.uk PD1020L

Operating instructions in the back. www.blackanddecker.co.uk PD1020L Operating instructions in the back www.blackanddecker.co.uk PD1020L 2 ENGLISH (Original instructions) Intended use Your Black & Decker Dustbuster handheld vacuum cleaner has been designed for vacuum cleaning

More information

User manual. DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP

User manual. DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP User manual DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP Table of Contents Intended Use... 1 Important... 1 Warnings... 1 Cautions... 2 Contraindications... 2 Symbol Key... 3 System Contents...

More information

Dual Cushion User Manual. Specialist Suppliers to the Healthcare Industry

Dual Cushion User Manual. Specialist Suppliers to the Healthcare Industry Dual Cushion User Manual Specialist Suppliers to the Healthcare Industry Thank you for purchasing this Shelden Healthcare pressure relieving cushion system. PLEASE READ THIS USER MANUAL CAREFULLY BEFORE

More information

Portable Evaporative Air Cooler. OWNER S MANUAL Read and save these instructions before use. Model: CL30XC

Portable Evaporative Air Cooler. OWNER S MANUAL Read and save these instructions before use. Model: CL30XC Portable Evaporative Air Cooler OWNER S MANUAL Read and save these instructions before use Model: CL30XC Power rating: 250 Watts Voltage rating: 230 Volt, 50Hz Made in P.R.C. QUICK START GUIDE Fill with

More information

1 High pressure washer 2 Wheel 3 Inlet connector 4 Outlet connector 5 Handle 6 Trigger gun 7 Detergent bottle. Technical Data:

1 High pressure washer 2 Wheel 3 Inlet connector 4 Outlet connector 5 Handle 6 Trigger gun 7 Detergent bottle. Technical Data: 1 High pressure washer 2 Wheel 3 Inlet connector 4 Outlet connector 5 Handle 6 Trigger gun 7 Detergent bottle Technical Data: Rated pressure Permissible pressure (max.) Inlet (max.) Rated flow Max. flow

More information

M Series with C-Flex

M Series with C-Flex M Series with C-Flex IMPORTANT! Fill in the information below when you receive the REMstar Auto M Series system. Serial No.: (located on the bottom of the device) System Prescribed for: Date of Purchase

More information

HOFFRICHTER DATA box

HOFFRICHTER DATA box HOFFRICHTER DATA box Operating Instructions Serial Number Each HOFFRICHTER device is provided with a serial number which guarantees the product can be traced back. The serial number is on the type identification

More information

User Manual. RK-2d / RK-2t. dedicated KVM switch and rackmount screen technology. Designed and manufactured by Austin Hughes

User Manual. RK-2d / RK-2t. dedicated KVM switch and rackmount screen technology. Designed and manufactured by Austin Hughes dedicated KVM switch and rackmount screen technology User Manual RK-2d / RK-2t 1U Short Depth Keyboard Drawer Designed and manufactured by Austin Hughes 751 Legal Information First English printing, October

More information

ITC-BTTN Cellular Bluetooth Gateway. Owner s Manual 1

ITC-BTTN Cellular Bluetooth Gateway. Owner s Manual 1 ITC-BTTN Cellular Bluetooth Gateway Owner s Manual 1 2 Table of Contents Introduction...3 Package Contents...3 XLink Connections Diagram...4 Setup...5 Pairing your Bluetooth Cell Phone to the XLink...6

More information

When you switch off your system, or mute the sound, the red indicator light appears immediately, indicating that the subwoofer is not in use.

When you switch off your system, or mute the sound, the red indicator light appears immediately, indicating that the subwoofer is not in use. BeoLab 11 Guide WARNING: To reduce the risk of fire or electric shock, do not expose this appliance to rain or moisture. Do not expose this equip ment to dripping or splashing and ensure that no objects

More information

The Bouncer Bluetooth Stereo Speaker

The Bouncer Bluetooth Stereo Speaker The Bouncer Bluetooth Stereo Speaker 1 Welcome to JLab! Thank you for purchasing The Bouncer Bluetooth Speaker! Enjoy your music wirelessly from any smartphone, tablet, laptop, or other Bluetooth enabled

More information

2 YEAR WARRANTY S7200PAU USE & CARE MANUAL ADVANCED CERAMIC TECHNOLOGY. To register your product go to www.remington-products.com.

2 YEAR WARRANTY S7200PAU USE & CARE MANUAL ADVANCED CERAMIC TECHNOLOGY. To register your product go to www.remington-products.com. USE & CARE MANUAL PLEASE READ PRIOR TO USE To register your product go to www.remington-products.com.au ADVANCED CERAMIC TECHNOLOGY 2 YEAR WARRANTY S7200PAU 1 IMPORTANT SAFEGUARDS Warning: When the straightener

More information

Do not use soaps with added scents. We recommend plain Ivory liquid soap or baby shampoo. Rinse well and allow to air dry out of direct sunlight.

Do not use soaps with added scents. We recommend plain Ivory liquid soap or baby shampoo. Rinse well and allow to air dry out of direct sunlight. ResMed CPAP Mask FAQ When and how should I clean my mask? Daily Wash the mask system with warm, soapy water. Do not use soaps with added scents. We recommend plain Ivory liquid soap or baby shampoo. Rinse

More information

Vertical Display and Storage B1350-2. SKOPE Gen2: Three Door Chiller

Vertical Display and Storage B1350-2. SKOPE Gen2: Three Door Chiller Vertical Display and Storage User Manual MAN1227 Rev. 3.0 March 2008 edition CONTACT ADDRESSES Designed and Manufactured by New Zealand SKOPE INDUSTRIES LIMITED PO Box 1091, Christchurch New Zealand Freephone:

More information

Get Cleaning... User Guide Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308. Multifunction. vax.co.uk. 6131 Series. What s your Vax model number?

Get Cleaning... User Guide Vax Careline: (UK) 0844 412 8455 (ROI) 1-800 928 308. Multifunction. vax.co.uk. 6131 Series. What s your Vax model number? 6131 Series User Guide v1.1-06.09.10:user guide 09/09/2010 11:42 Page 1 User Guide Vax Careline: (UK) 0844 412 8455 Multifunction Get Cleaning... What s your Vax model number? 6 1 3 1 What s your serial

More information

GDSB02BT20 20W Soundbar User Guide

GDSB02BT20 20W Soundbar User Guide GDSB02BT20 20W Soundbar User Guide IMPORTANT: Please read all instructions carefully before use and keep for future reference Introduction Thank you for choosing this Goodmans 20W Soundbar To get the most

More information

OPTICAL HEADEND PLATFORM OTOHP-NMS NETWORK MONITORING MODULE INSTRUCTION MANUAL

OPTICAL HEADEND PLATFORM OTOHP-NMS NETWORK MONITORING MODULE INSTRUCTION MANUAL OPTICAL HEADEND PLATFORM OTOHP-NMS NETWORK MONITORING MODULE INSTRUCTION MANUAL Phone: (209) 586-1022 (800) 545-1022 Fax: (209) 586-1026 OTOHP-NMS Rev. X1 E-Mail: salessupport@olsontech.com www.olsontech.com

More information

Installation and Operation Back-UPS 1250, 1300, 1500

Installation and Operation Back-UPS 1250, 1300, 1500 Installation and Operation Back-UPS 1250, 1300, 1500 Inventory bu001a Safety and General Information This unit is intended for indoor use only. Do not operate this unit in direct sunlight, in contact with

More information

Active Personal Programmer G5

Active Personal Programmer G5 Active Personal Programmer G5 Instruction manual and user s guide This manual explains features and provides basic information about how to set up and operate the Active Personal Programmer Tool, including

More information

SL280UHV SERIES GAS FURNACE WARNING

SL280UHV SERIES GAS FURNACE WARNING 2010 Lennox Industries Inc. Dallas, Texas, USA 506677 01 11/2010 Supersedes 506409 01 SL280UHV SERIES GAS FURNACE Litho U.S.A. FIRE OR EXPLOSION HAZARD. Failure to follow safety warnings exactly could

More information

PORTABLE AIR CONDITIONER

PORTABLE AIR CONDITIONER MODEL: DC10RC INSTRUCTION MANUAL PORTABLE AIR CONDITIONER IMPORTANT INFORMATION Carefully read this instruction manual before you install or use your portable air conditioner. Please retain this manual

More information

WF720 Wireless Home Phone User Manual

WF720 Wireless Home Phone User Manual WF720 Wireless Home Phone User Manual Content Getting to Know Your Device... 3 Appearance...3 LED Indicator...4 Device Installation... 5 Before You Begin...5 Installing the SIM Card and the Battery...5

More information