Optimet 3D Dental Multi-Facet Scanner Attainable Accuracy for Dental Application Cases

Size: px
Start display at page:

Download "Optimet 3D Dental Multi-Facet Scanner Attainable Accuracy for Dental Application Cases"

Transcription

1 Optimet 3D Dental Multi-Facet Scanner Attainable Accuracy for Dental Application Cases Optimet Dental 3D Scanner This paper presents an overview of the Optimet multi-facet Dental Scanner used for 3D scanning of dental objects for CAD/CAM procedures and the Scanner accuracy. Optimet Multi-Facet 3D Dental Scanner (Marketed by Nobel Biocare) Optimet s CAD/CAM Dental Scanner provides accurate 3-D scans for an extensive range of dental objects and CAD/CAM products including single to multi unit implant bridge cases, over-dentures, individualized abutments, cast models (gypsum), wax-up s and impressions. The Scanner, developed and manufactured by Optimet, is marketed globally by Nobel Biocare under NobelProcera. The Dental Scanner integrates Optimet s patented conoscopic holography (collinear) measurement technology combined with multi-facet optical technology enabling accurate 3D scanning of undercuts, concavities and complex geometries, including undercut coverage and double sided scans (360 ). Optimet collinear measurement technology Conoscopic holography In the Dental CAD/CAM application, accurate 3D measurements provided by Optimet Dental Scanner are used for planning, design and manufacturing of individualized dental restoration, including crowns, bridges, single unit abutments and implant bridges and frameworks. The design is performed by dedicated dental CAD software based on 3D scan data of the cast models and impressions representing the dental situation. Multi-Facet scanning technology

2 Metrology definitions Measurement Precision, Accuracy, Reproducibility and measurement Certainty (or Uncertainty) are defined in different standards and publications (such as the German DIN). Dental objects under measurement are characterized as free form geometries and surfaces. Relevant methods analyze the accuracy as deviations of the 3D surface and dimension measurement from independent and accurate dimensions (nominal values) of the object in defined coordinate system. The analysis provides the level of certainty or uncertainty stipulated by statistical parameters based on the deviations distribution (such as 2σ or 3σ) and related statistical tests. Reproducibility analysis tests the ability of the Scanner to provide similar measurement results for the same measure and under different and repeated measurement conditions. Accuracy results on metrology objects Optimet Dental Scanner accuracy was tested in several methods and analyzed with Geomagic Qualify 11 software, based on surface output format (STL) of the scanner while additional tests were performed based on the cloud of points output where specified. The first object scanned and tested was a metal (brass) metrology cone with undercuts and an edge radius of 0.150mm. The actual dimensions of the metrology jig were certified by a CMM machine (with an accuracy of 1µm). The scans were performed by several Scanners in standard environmental conditions and the results were compared to the solid model file of the metrology jig. Deviation map metrology cone Green color indicate deviations within ±15µ from nominal dimensions The average deviations between scanned surface and the solid dimensions are within -5µm to +7µm with Std. of 7µm. More than 95% of the surface (~2σ) is within ± 15µm from the nominal surfaces. The results are typical for all object's area s including the sharp margin/edge area. See the deviation map and deviation distribution in the images herein. Volumetric accuracy multi spheres test A volumetric accuracy test was performed by scanning a 3 spheres metrology jig with distances between the spheres of 30mm, 40mm and 50mm (sphere Optimet Multi Facet Dental Scanner diameters of 19.5mm) and comparing the sphere center position 3 spheres metrology jig measurement set-up and distance with the nominal values. The metrology jig is certified by a CMM machine (with 1µm accuracy). Results of 40 measurement tests on 10 3 spheres jig different Scanners show average deviations of 7.9µ from the nominal dimensions (with Std. of 5.26µm). The average deviations of distances between the spheres were 6.67µm (30mm distance), 7.93µm (40mm distance) and 9.10µm (50mm distance). The maximum deviation from the nominal distance values in all tests (120 test scans by 10 different scanners) was 19µm.

3 A multi facet scan test performed by the Dental Scanner on the metrology sphere with a nominal radius of 19.5mm (tolerance of ±2µm) with average deviations from a sphere surface of ±5µm with Std. of diameter of 6µm. The actual measured radius is , 10µm from nominal value. Deviation map metrology sphere Green color indicate deviations within ±10µm Accuracy in dental objects scans (free form) Ten different Gypsum dies (anterior and posterior teeth) as well as a titanium die were scanned by Optimet's Dental Scanner. The accuracy was analysed by comparing and calculating the deviations between the scan results (surface) to accurate scans performed by a high accuracy industrial scanner (having accuracies within ±4µm). The average deviations of the scan results of the dies surface (on all the 10 test dies) were ±7µm with Std. of 8µm (deviations are measured from the nominal values as scanned by the high accuracy industrial scanner). 97.1% of the surface area of all dies was within ±15µm from nominal dimension values. The deviation varied (according to the specific test die) between 95.5% 98.5% of the scan surface was within ± 15µm the different dies showed that on average Reproducibility analysis performed by comparing scans of the dies by different Optimet scanners showed accuracies with average deviations of -6.7µm / +5.8µm with a Std of 8.2µm (1σ = 15µm), similar to the accuracy analysis above. Deviation Maps - 3 dental test samples measured and compared to high accuracy industrial scans. (green color represents surfaces deviations up to ±19µm deviations in scale). An additional accuracy test was performed on a Titanium die (in a tooth shape) in height of 10mm with edges in radiuses (simulating a sharp crown margin line) between 0.03mm - 0.1mm. The scan was performed with no spray. Average deviations from the accurate nominal values were within -9µm / +7µm. 93% of the surface area was within ±15µm of the nominal surface (as measured by the high accuracy industrial scanner after deducting the reference measurement uncertainty), including in the sharp margin area. See below for a color deviation map and deviation distr ibution chart. Deviation map Ti die Green color indicate deviations within ±15µm From nominal values Comparative Profile along the Titanium die Measurement Vs. solid Deviation distribution chart Ti die

4 Accuracy results on Implant bridges The design and manufacturing of Implant bridges require specifically very high accuracies. Special algorithms embedded in the Dental Scanner software enable measurement of the implants accurate position and inclination angle. This high accuracy 3D scan data enable design of the implant bridge framework or bar as well as an over-denture in CAD/CAM procedures. A comprehensive accuracy test performed included scans of multiple implant bridge cases on cast models with 3-6 implants (with distances of up to 45mm between implants). The implant seats exact position (implant interface seat center) and angle was extracted by the Dental Scanner software and then compared to high precision measurements performed on a CMM measurement system (accuracy within 1µm). The accuracy analysis on the exact implant seats ZXY position and inclination angle (of the implant), show deviations within ±5µm. Note that the most critical parameters are the Z (height) and inclination angles (of the implants). 3D view of deviation map of implant bridge seats including angulations. Green color indicate deviations within ±4µm Deviation colour maps (below) show two examples of the deviation analysis of the implant seat in comparison to the nominal values (as measured by the CMM machine). In addition, the chart below shows the accuracy in different test cases within ±4.5µm as well as the reproducibility of the test results. Deviation map - implant bridge seats Green color indicate deviations within ±4µm Orange color represent ±6-8µm Deviation chart - implant bridge seats scanning accuracies multiple test cases result Average deviations from nominal position within ±4.5µm

5 General Metrology Applications The Multi Facet Scanner is capable of 3D scanning different objects with complicated geometries for general metrology and reverse engineering applications. Optimet s Conoscopic sensors have standard interchangeable objective lenses from 16mm up to 250mm lens reaching accuracies up to sub-micron range. The Scanner can be provided for metrology applications with lens objective from 25mm-100mm allowing flexible resolutions and working ranges for different purposes. 3-D Multi facet measurement: a cutting tool and a plastic part measured by the multi facet scanner. Measurements performed in one set-up and include holes and other geometries. Summary Dental and CAD/CAM procedures and tolerances should aim for reaching adequate clinical results in planning, design and manufacturing of dental restorations. The accuracy results and analysis in this paper represent the accuracies in digital data acquisition as part of the CAD/CAM process. Optimet CAD/CAM Dental Scanner metrological accuracy and reproducibility were tested and analyzed on different objects and cases. The accuracy results on dental die cases as well as verified metrology jigs are within 15µm. Optimet s Dental Scanner operational principle keeps the object being measured in a Cartesian coordinate system (XY) with no angulations or rotation of the object during the measurement. The volumetric accuracy as shown in the tests is therefore kept at the same level of accuracy 15µm with no accuracy degradation in the scanning volume. The accuracy results on multiple implant bridges cases (implants seats including height and angulations) were within a range of 5µm with similar results in reproducibility tests. References: 1. G.Y. Sirat; Conoscopic holography. Basis principles and physical basis J. Opt. Soc. Am. ( 1992) 2. Redlich M, Weinstock T, Abed Y, Schneor R, Holdstein Y, Fischer A " A new system for scanning, measuring and analyzing dental casts based on a 3D holographic sensor" Orthod Craniofac Res 2008;11: I Alvarez, J M Enguita, M Frade, J Marina and G Ojea "On-Line Metrology with Conoscopic Holography: BeyondTriangulation" Sensors 2009, 9, DIN , Grundlagen der Messtechnik Teil 1:Grundbegriffe Fundamentals of metrology Part 1: Basic terminology (German and English) 5. DIN 32877, Optoelectronic measurement of form, profile and distance, August Eurolab, European Federation of National Associations of Measurement, Testing and Analytical Laboratories, Technical Report No. 1/2006 August International Organization of Standards "Guide to the Expression of Uncertainty in Measurement"

DS6000. Advanced Dental Scanner & CAD Software. Crowns, Bridges Wax-ups Implant Bridges and Bars High Accuracy Scan & Design

DS6000. Advanced Dental Scanner & CAD Software. Crowns, Bridges Wax-ups Implant Bridges and Bars High Accuracy Scan & Design DS6000 Advanced Dental Scanner & CAD Software Crowns, Bridges Wax-ups Implant Bridges and Bars High Accuracy Scan & Design Inlay/Onlays DICOM Viewer Custom Abutments Virtual Articulator CAM Module DS6000

More information

PCB Component Placement Inspection

PCB Component Placement Inspection Executive Summary PCB Component Placement Inspection Optimet s ConoProbe Mark10 HD with a 50 mm focal length lens was used to inspect PCB component placement. The PCB board inspected contained both individual

More information

Enter into the world of unprecedented accuracy! IScan D104 Series

Enter into the world of unprecedented accuracy! IScan D104 Series Enter into the world of unprecedented accuracy! IScan D104 Series 2 Three scanners, one goal: Precisely fitting restorations The D104 Series of 3D scanning systems is the 5th generation of dental scanners

More information

New features DWOS 3.2

New features DWOS 3.2 New features DWOS 3.2 Version 1.1 Septembre 13 th 2012 Summary This document describes how to use the new features and improvements of the release 3.2.0 of the Dental Wings Open Software (DWOS). Order

More information

Optical Digitizing by ATOS for Press Parts and Tools

Optical Digitizing by ATOS for Press Parts and Tools Optical Digitizing by ATOS for Press Parts and Tools Konstantin Galanulis, Carsten Reich, Jan Thesing, Detlef Winter GOM Gesellschaft für Optische Messtechnik mbh, Mittelweg 7, 38106 Braunschweig, Germany

More information

PRECISION, RELIABILITY, SIMPLICITY. simeda. customized. & solutions. Digital performance

PRECISION, RELIABILITY, SIMPLICITY. simeda. customized. & solutions. Digital performance PRECISION, RELIABILITY, SIMPLICITY simeda customized & solutions Digital performance 2 Simeda Scan The high-precision scanner for implantology! With, a company specialised in industrial metrology, Simeda

More information

Implant Parts. A Radford Heath Guide http://www.radfordheath.com 1

Implant Parts. A Radford Heath Guide http://www.radfordheath.com 1 Implant Parts A Radford Heath Guide http://www.radfordheath.com 1 Disclaimer The information given in this document has been provided in good faith for basic information purposes only and the information

More information

Optical 3D-measurement and its use in minting. Alicona - Mr. Dipl.-Ing. Christian JANKO Münze Österreich - Mr. Ing. Alfred GNADENBERGER 29.1.

Optical 3D-measurement and its use in minting. Alicona - Mr. Dipl.-Ing. Christian JANKO Münze Österreich - Mr. Ing. Alfred GNADENBERGER 29.1. Optical 3D-measurement and its use in minting Alicona - Mr. Dipl.-Ing. Christian JANKO Münze Österreich - Mr. Ing. Alfred GNADENBERGER 29.1.2015 Index: 1. Introduction to Alicona 2. 3D measurement: principle

More information

DESS. Screws. Tijuana Ventas: (664) 685 6294/95 hirambogarin@dabocorp.com. For all major implant systems!! www.dabocorp.com

DESS. Screws. Tijuana Ventas: (664) 685 6294/95 hirambogarin@dabocorp.com. For all major implant systems!! www.dabocorp.com Screws Screws for definitive use. Made from medical grade 5 ELI Titanium Integrity and soundness guaranteed by stress tests of up to 150% of their nominal torques. Their designs ensure a perfect fit with

More information

USAGE OF SELECTIVE LASER MELTING TECHNOLOGY CAPABILITIES IMPLEMENTATION FOR TEETH IMPLANTS METAL CONSTRUCTION IN TU-SOFIA

USAGE OF SELECTIVE LASER MELTING TECHNOLOGY CAPABILITIES IMPLEMENTATION FOR TEETH IMPLANTS METAL CONSTRUCTION IN TU-SOFIA USAGE OF SELECTIVE LASER MELTING TECHNOLOGY CAPABILITIES IMPLEMENTATION FOR TEETH IMPLANTS METAL CONSTRUCTION IN TU-SOFIA George TODOROV, Nikolay NIKOLOV, Victor MITOV, Yavor SOFRONOV Technical University

More information

Making Better Medical Devices with Multisensor Metrology

Making Better Medical Devices with Multisensor Metrology Making Better Medical Devices with Multisensor Metrology by Nate J. Rose, Chief Applications Engineer, Optical Gaging Products (OGP) Multisensor metrology is becoming a preferred quality control technology

More information

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster Jesse Garant, JG&A Metrology Center Traditional Metrology and Inspection Tactile Devices (Touch Probe)

More information

CAD / CAM Dr. P. V. Madhusuthan Rao Department of Mechanical Engineering Indian Institute of Technology, Delhi Lecture No. # 12 Reverse Engineering

CAD / CAM Dr. P. V. Madhusuthan Rao Department of Mechanical Engineering Indian Institute of Technology, Delhi Lecture No. # 12 Reverse Engineering CAD / CAM Dr. P. V. Madhusuthan Rao Department of Mechanical Engineering Indian Institute of Technology, Delhi Lecture No. # 12 Reverse Engineering So what we will do in today s lecture is basically take

More information

ATLANTIS abutments as individual as your patients

ATLANTIS abutments as individual as your patients ATLANTIS abutments as individual as your patients ATLANTIS the freedom of unlimited possibilities Discover why ATLANTIS CAD/CAM abutments are the natural choice for patientspecific, cement-retained implant

More information

Simpl Implant Abutments with Atlantis CAD CAM Technology

Simpl Implant Abutments with Atlantis CAD CAM Technology Simpl Implant Abutments with Atlantis CAD CAM Technology Don t change your implant... change your abutment. See more patients in less time... and save money in the process! See how we ve made it Simpl.

More information

Laboratory Solutions. Optimization By Design

Laboratory Solutions. Optimization By Design Laboratory Solutions Optimization By Design Digital Dentistry Solutions Laboratory Solutions Business Optimization With BellaTek Digital Dentistry Solutions Optimize the solutions you give to your customers

More information

Implants in your Laboratory: Abutment Design

Implants in your Laboratory: Abutment Design 1/2 point CDT documented scientific credit. See Page 41. Implants in your Laboratory: Abutment Design By Leon Hermanides, CDT A patient s anatomical limitations have the greatest predictive value for successful

More information

BASIC INFORMATION ON THE STRAUMANN VARIOBASE ABUTMENT. Straumann Variobase Abutment

BASIC INFORMATION ON THE STRAUMANN VARIOBASE ABUTMENT. Straumann Variobase Abutment BASIC INFORMATION ON THE STRAUMANN VARIOBASE ABUTMENT Straumann Variobase Abutment 1 The ITI (International Team for Implantology) is academic partner of Institut Straumann AG in the areas of research

More information

Digital impression solutions. 3M True Definition Scanner. Precise. impressions. in a jiffy

Digital impression solutions. 3M True Definition Scanner. Precise. impressions. in a jiffy Digital impression solutions 3M True Definition Scanner Precise impressions in a jiffy True Definition Scanner Digital impressioning redefined Easy handling. Desired result. With the 3M True Definition

More information

Laboratory Designed Abutment. Optimization By Design

Laboratory Designed Abutment. Optimization By Design The Laboratory Designed Abutment Optimization By Design Table Of Contents Instructions For Scanning The Laboratory Designed Abutment.... 1 Pouring The Model And Cast... 1 Starting The Scanner.... 1 Calibrating

More information

Injection moulding and modelling on a micro scale

Injection moulding and modelling on a micro scale Injection moulding and modelling on a micro scale Technology Update Injection moulding and welding of plastics 11 November 2014 Research Projects (National / European) Micro/Nano/Multimaterial Manufacturing

More information

CAD/CAM technology supporting successful implant therapy

CAD/CAM technology supporting successful implant therapy CAD/CAM technology supporting successful implant therapy Suheil M. Boutros, DDS, MS, Manuel Fricke, DT Modern implantology opens up new treatment options for individuals with only minimal or no remaining

More information

State-of-the-art for your prostheses

State-of-the-art for your prostheses WHERE CREATIVITY meets technology E CREATECH EXCELLENCE State-of-the-art technology for your prostheses 1 CREATECH EXCELLENCE What is createch Excellence? Createch Excellence is OUR RANGE OF SOLUTIONS

More information

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven CT Traceability - Calibration and Accuracy Prof. Wim Dewulf, Group T - KU Leuven Outline Introduction: terminology and procedures Voxel size calibration Edge offset calibration Conclusions Outline Introduction:

More information

Implant solutions. 3M True Definition Scanner. Precise. implant impressions. with incredible speed

Implant solutions. 3M True Definition Scanner. Precise. implant impressions. with incredible speed Implant solutions 3M True Definition Scanner Precise implant impressions with incredible speed True Definition Scanner You know in advance that it will be as desired The desired impression fast. The precise

More information

ATLANTIS abutments design guide CAD/CAM patient-specific abutments

ATLANTIS abutments design guide CAD/CAM patient-specific abutments ATLANTIS abutments design guide CAD/CAM patient-specific abutments Contents Introduction 4 This manual helps you to explore all the benefits of ATLANTIS CAD/CAM patient-specific abutments. It gives you

More information

Milled Bar Overdentures-Protocol and Procedures

Milled Bar Overdentures-Protocol and Procedures Milled Bar Overdentures-Protocol and Procedures Dennis Urban, CDT When full implant dentures requiring a milled bar is prescribed, the proper protocol and procedure are of extreme importance. The proper

More information

PROSTHETIC PROCEDURE. for HG IMPLANT SYSTEM

PROSTHETIC PROCEDURE. for HG IMPLANT SYSTEM PROSTHETIC PROCEDURE for HG IMPLANT SYSTEM PROSTHETIC PROCEDURE for HG IMPLANT SYSTEM HG Implant System Contents Cement retained restoration Rigid abutment When abutment reduction is unnecessary When abutment

More information

Arm2. Arm Arm22. Articulated Arm. machines MEASURING. tridimensional measuring FRATELLI ROTONDI

Arm2. Arm Arm22. Articulated Arm. machines MEASURING. tridimensional measuring FRATELLI ROTONDI Arm Arm22 2 Arm2 tridimensional measuring Articulated Arm MEASURING machines FRATELLI ROTONDI Arm2 MEASURING Arm2Arm2 SOFTWARE Software in WINDOWS environment: sura tridimesio - elements with definite

More information

ENGINEERING METROLOGY

ENGINEERING METROLOGY ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring

More information

PRODUCT BROCHURE SPATIAL ANALYZER. Powerful, traceable and easy-to-use metrology and analysis software

PRODUCT BROCHURE SPATIAL ANALYZER. Powerful, traceable and easy-to-use metrology and analysis software PRODUCT BROCHURE SPATIAL ANALYZER Powerful, traceable and easy-to-use metrology and analysis software SPATIAL ANALYZER SpatialAnalyzer (SA) is a powerful, traceable and easy to use metrology and analysis

More information

Restorative Guidelines

Restorative Guidelines Restorative Guidelines Contents Restorative Guidelines 4.1 Neoss Implant System 4.2 4.2 Esthetiline Solution 4.3 4.3 Provisional Abutments 4.8 4.4 Impression Techniques Implant Level 4.12 4.5 NeoLink

More information

Pipeline External Corrosion Analysis Using a 3D Laser Scanner

Pipeline External Corrosion Analysis Using a 3D Laser Scanner Pipeline Technology Conference 2013 Pipeline External Corrosion Analysis Using a 3D Laser Scanner Pierre-Hugues ALLARD, Charles MONY Creaform, www.creaform3d.com 5825 rue St-Georges, Lévis (QC), Canada,

More information

Automotive Applications of 3D Laser Scanning Introduction

Automotive Applications of 3D Laser Scanning Introduction Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc

More information

3D SCANNING: A NEW APPROACH TOWARDS MODEL DEVELOPMENT IN ADVANCED MANUFACTURING SYSTEM

3D SCANNING: A NEW APPROACH TOWARDS MODEL DEVELOPMENT IN ADVANCED MANUFACTURING SYSTEM 3D SCANNING: A NEW APPROACH TOWARDS MODEL DEVELOPMENT IN ADVANCED MANUFACTURING SYSTEM Dr. Trikal Shivshankar 1, Patil Chinmay 2, Patokar Pradeep 3 Professor, Mechanical Engineering Department, SSGM Engineering

More information

Types of 3D Scanners and 3D Scanning Technologies.

Types of 3D Scanners and 3D Scanning Technologies. Types of 3D Scanners and 3D Scanning Technologies. There are many types of 3D scanners and 3D scanning technologies. Some are ideal for short range scanning while others are better suited for mid or long

More information

3Shape TRIOS. 3Shape TRIOS. Digital Impression Solution

3Shape TRIOS. 3Shape TRIOS. Digital Impression Solution 3Shape TRIOS 3Shape TRIOS Digital Impression Solution Technology designed the way you work 3Shape TRIOS Next-generation digital impressions More and more dentists are using digital impression taking systems

More information

Application Report. Propeller Blade Inspection Station

Application Report. Propeller Blade Inspection Station 34935 SE Douglas Street, Suite, Snoqualmie, WA 9874 Ph: 425-396-5577 Fax: 425-396-586 Application Report Propeller Blade Inspection Station Prepared By Kyle Johnston, Ph. D. Metron Systems Inc.5.5 3 2.5

More information

All-on-4 treatment concept with NobelSpeedy Groovy

All-on-4 treatment concept with NobelSpeedy Groovy All-on-4 treatment concept with NobelSpeedy Groovy Product overview Immediate Function for high patient satisfaction Immediately loaded fixed provisional prosthesis on the day of surgery. Immediate improvement

More information

Taking the Mystique out of Implant Dentistry. Dr. Michael Weinberg B.Sc., DDS, FICOI

Taking the Mystique out of Implant Dentistry. Dr. Michael Weinberg B.Sc., DDS, FICOI Taking the Mystique out of Implant Dentistry Dr. Michael Weinberg B.Sc., DDS, FICOI What is Restorative Implant Dentistry? Restorative implant dentistry involves taking a few simple mechanical principles

More information

PureForm Ceramic System. Emulating nature. The perfect disguise.

PureForm Ceramic System. Emulating nature. The perfect disguise. PureForm Ceramic System Emulating nature. The perfect disguise. The new shape of ceramics. The PureForm Ceramic System is offered in six configurations which are designed according to four basic tooth

More information

CHAPTER 10 RESTS AND PREPARATIONS. 4. Serve as a reference point for evaluating the fit of the framework to the teeth.

CHAPTER 10 RESTS AND PREPARATIONS. 4. Serve as a reference point for evaluating the fit of the framework to the teeth. CHAPTER 10 RESTS AND DEFINITIONS A REST is any rigid part of an RPD framework which contacts a properly prepared surface of a tooth. A REST PREPARATION or REST SEAT is any portion of a tooth or restoration

More information

Khaldoon G. Abu Afifeh, B.D.S, M. Sc in Prosthodontics

Khaldoon G. Abu Afifeh, B.D.S, M. Sc in Prosthodontics Khaldoon G. Abu Afifeh, B.D.S, M. Sc in Prosthodontics Diagnosis Diagnosis Diagnosis Implantology Is a Prosthetically Driven Entity. Information Before Treatment is Diagnosis Information after Treatment

More information

Dental Laboratories. Procera provides you access to our multi-million dollar production facility

Dental Laboratories. Procera provides you access to our multi-million dollar production facility Dental Laboratories provides you access to our multi-million dollar production facility minimal investment maximum rewards strong, beautiful, proven products are famous for their excellent strength and

More information

platform shifting Nobel Biocare is shifting gingival beauty to the next level

platform shifting Nobel Biocare is shifting gingival beauty to the next level platform shifting Nobel Biocare is shifting gingival beauty to the next level Nobel Biocare is shifting gingival beauty to the next level Platform Shifting is the latest in Nobel Biocare s offering of

More information

5-Axis Test-Piece Influence of Machining Position

5-Axis Test-Piece Influence of Machining Position 5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,

More information

TRIOS Digital Impression Solution

TRIOS Digital Impression Solution Digital Impression Solution 5 reasons to choose TRIOS is more than an impression Superior scanning technologies Flexible hardware configurations Widest range of indications Open and flexible for any lab

More information

WAX-UP AND CERAMIC EXTENSIVE COURSE Dr. Dario Adolfi Dr. Ivan Ronald Huanca

WAX-UP AND CERAMIC EXTENSIVE COURSE Dr. Dario Adolfi Dr. Ivan Ronald Huanca WAX-UP AND CERAMIC EXTENSIVE COURSE Dr. Dario Adolfi Dr. Ivan Ronald Huanca Duration: 6 meses STEP 1: WAX-UP OF FOUR UPPER POSTERIOR TEETH with Dr. Ivan Ronald Huanca The objective of this course s step

More information

PreciTrack3D GmbH. PRECI 3D TRACK PRECI 3D GUN PRECI 3D SCAN PRECI 3D ROBOGUN PRECI 3D CONFOCAL PRECI 3D REF Tube Inspection & Measurements PRODUCTS

PreciTrack3D GmbH. PRECI 3D TRACK PRECI 3D GUN PRECI 3D SCAN PRECI 3D ROBOGUN PRECI 3D CONFOCAL PRECI 3D REF Tube Inspection & Measurements PRODUCTS HIGHLIGHTS 3D Real Time Tracking Quality Control & Inspection Reverse Engineering ID 2D 3D Scanning PRODUCTS PRECI 3D TRACK PRECI 3D SCAN PRECI 3D CONFOCAL PRECI 3D REF Tube Inspection & Measurements PRECI

More information

Summary. DWOS 3.8.0 New features. Version 1.2. May 08, 2014. Summary 1. Latest Additions 3. Orthodontic Archiving 3

Summary. DWOS 3.8.0 New features. Version 1.2. May 08, 2014. Summary 1. Latest Additions 3. Orthodontic Archiving 3 DWOS 3.8.0 New features Version 1.2 May 08, 2014 Summary The present document contains step- by-step instructions and images to guide the users through the major new features and improvements of the release

More information

Another Implant Option for Missing Teeth with Challenging Symmetry Patrick Gannon, DDS and Luke Kahng, CDT

Another Implant Option for Missing Teeth with Challenging Symmetry Patrick Gannon, DDS and Luke Kahng, CDT Another Implant Option for Missing Teeth with Challenging Symmetry Patrick Gannon, DDS and Luke Kahng, CDT Introduction A 58 year old male had been missing teeth #7=12 for approximately 28 years. During

More information

Tizian TM Creativ RT. Abutment Designer. for the CAD-software version 3.932. Instruction Manual

Tizian TM Creativ RT. Abutment Designer. for the CAD-software version 3.932. Instruction Manual Tizian TM Creativ RT Abutment Designer for the CAD-software version 3.932 Instruction Manual Schütz Dental GmbH 61191 Rosbach, Germany Tel.: +49 6003 / 814-0 Fax: +49 6003 / 814-906 Version: 12/2010 Contents

More information

MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT

MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT C. Reinhart, C. Poliwoda, T. Guenther, W. Roemer, S. Maass, C. Gosch all Volume Graphics GmbH, Heidelberg, Germany Abstract:

More information

Encode Restorative System Procedure & Laboratory Manual

Encode Restorative System Procedure & Laboratory Manual Procedure & Laboratory Manual Encode Virtual Abutment Encode Abutment Blank Encode Abutment 1/2 Milled Encode Abutment Completed ENCODE RESTORATIVE SYSTEM Simply Impressive With the introduction of the

More information

Lava. CNC 500 Milling Machine. Redefine. Productivity

Lava. CNC 500 Milling Machine. Redefine. Productivity Lava CNC 500 Milling Machine Redefine Productivity 2 3 4 1 5 6 1 2 3 3-axis speed and precision. 5-axis milling when you need it. A workpiece magazine that s not a lot of work. A worktool magazine with

More information

NARROW DIAMETER implant

NARROW DIAMETER implant ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball

More information

INDUSTRIAL VISION. Don Braggins, UK Industrial Vision Association

INDUSTRIAL VISION. Don Braggins, UK Industrial Vision Association CMM 3_7 PROGRESS_Layout 1 05/11/2010 15:58 Page 45 Don Braggins, UK Industrial Vision Association INDUSTRIAL VISION AND MICRO MANUFACTURING MAIN IMAGE: Figure 4. Effects on an image of varying the angle

More information

0 INTRODUCTION. Fig. 1. Atos II optical scanner schematics

0 INTRODUCTION. Fig. 1. Atos II optical scanner schematics Strojniški vestnik - Journal of Mechanical Engineering 57(011)11, 86-833 Paper received: 07.07.010 DOI:10.5545/sv-jme.010.15 Paper accepted: 03.08.011 Possibilities of Using Three-Dimensional Optical Scanning

More information

Information on the product

Information on the product Information on the product Information on the product -Implantat.doc Page 1 of 15 This document composes of the following items 1. Name of the product 2. Raw materials 3. Structure and dimensions 3.1.

More information

The LOCATOR concept. Simplicity and versatility for prosthesis fixation

The LOCATOR concept. Simplicity and versatility for prosthesis fixation The concept Simplicity and versatility for prosthesis fixation The concept Experience the freedom in prosthesis fixation Simple and secure fixation of implant-supported prostheses is essential for successful

More information

Application Example: Quality Control of Injection-Molded Parts

Application Example: Quality Control of Injection-Molded Parts Application Example: Quality Control of Injection-Molded Parts Measuring Systems: ATOS Keywords: Optical 3D measuring technology, mold try-out, first article inspection, form and dimension inspection of

More information

all digital all options The right milling option High precision and productivity

all digital all options The right milling option High precision and productivity all digital all options The innovative FULL CONTOUR milling system ZIRCONIA The right milling option High precision and productivity Customized digital solutions for every laboratory Preparation Scanners

More information

Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment.

Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment. Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment. Dr. med. dent. David McFadden, Dallas County, USA Initial situation (single X-ray) Tooth 16

More information

Removing chips is a method for producing plastic threads of small diameters and high batches, which cause frequent failures of thread punches.

Removing chips is a method for producing plastic threads of small diameters and high batches, which cause frequent failures of thread punches. Plastic Threads Technical University of Gabrovo Yordanka Atanasova Threads in plastic products can be produced in three ways: a) by direct moulding with thread punch or die; b) by placing a threaded metal

More information

procedures & products NOBELESTHETICS including Procera

procedures & products NOBELESTHETICS including Procera procedures & products NOBELESTHETICS including Procera First from Nobel Biocare. NOBELPERFECT, (NP, RP, WP), NOBELDIRECT (NP, RP, WP), Brånemark System, NOBELREPLACE and NOBELSPEEDY Implants. A complete

More information

Application Example: Reverse Engineering

Application Example: Reverse Engineering Application Example: Reverse Engineering Use of optical measuring technology in the ceramics industry Measuring system: ATOS Keywords: Reverse Engineering, Tool and Moldmaking, Quality Assurance, Ceramic

More information

Choosing the right type of abutment

Choosing the right type of abutment 50 Producing custom implant abutments using CAD/CAM Choosing the right type of abutment S. KHALILOVA 1, F. KISTLER 2, S. ADLER 3, S. WEISS 3, S. KISTLER 2 AND J. NEUGEBAUER 2,4 Rapid developments in the

More information

Contents. Cement retained restoration. Screw retained restoration. Overdenture retained restoration. TS Implant System. 70 ComOcta Gold Abutment

Contents. Cement retained restoration. Screw retained restoration. Overdenture retained restoration. TS Implant System. 70 ComOcta Gold Abutment Contents TS Implant System Cement retained restoration Screw retained restoration 06 Cement-retained bridges with the Solid abutment system (non- 72 Screw retained crown with the ComOcta Gold abutment

More information

NRC Publications Archive Archives des publications du CNRC

NRC Publications Archive Archives des publications du CNRC NRC Publications Archive Archives des publications du CNRC Digital 3D Imaging and Modeling: A Metrological Approach Beraldin, Jean-Angelo Publisher s version / Version de l'éditeur: Time Compression Technologies

More information

your talent. our technology. the perfect fit.

your talent. our technology. the perfect fit. your talent. our technology. the perfect fit. itero - exclusively distributed in Europe by The digital revolution Changing the Face of Dentistry Digital technology ensures a more accurate impression from

More information

Machine Tool Inspection & Analyzer solutions

Machine Tool Inspection & Analyzer solutions Machine Tool Inspection & Analyzer solutions Introduction IBS Precision Engineering machine tool inspection series offers a range of products to provide peace of mind and control to machine operators by

More information

ZEISS T-SCAN Automated / COMET Automated 3D Digitizing - Laserscanning / Fringe Projection Automated solutions for efficient 3D data capture

ZEISS T-SCAN Automated / COMET Automated 3D Digitizing - Laserscanning / Fringe Projection Automated solutions for efficient 3D data capture ZEISS T-SCAN Automated / COMET Automated 3D Digitizing - Laserscanning / Fringe Projection Automated solutions for efficient 3D data capture ZEISS 3D Digitizing Automated solutions for efficient 3D data

More information

Luigi Vito Stefanelli,D.Eng.,DDS

Luigi Vito Stefanelli,D.Eng.,DDS LYON, 27-28/09/2013 Luigi Vito Stefanelli,D.Eng.,DDS Assistant Researcher, Department of Oral Sciences, Sapienza University of Rome, Italy, School of Dentistry. Private practitioner dental office Stasi

More information

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements DMS 680 Universal length measuring system gauge inspection for»iso 9000«requirements Universal length measuring system DMS 680 Wide range of applications For periodic inspection of gauges, reference gauges,

More information

Full Crown Module: Learner Level 1

Full Crown Module: Learner Level 1 Full Crown Module Restoration / Tooth # Full Gold Crown (FGC) / 30 Extensions: Porcelain Fused to Metal (PFM) / 12 All Ceramic / 8 Learner Level 1 Mastery of Tooth Preparation Estimated Set Up Time: 30

More information

DWOS Lava Edition 5.0 new features

DWOS Lava Edition 5.0 new features DWOS Lava Edition 5.0 new features Version 1.0 April 21, 2015 This document presents the new features and significant improvements brought with the DWOS Lava 5.0.0 release. Descriptions, instructions and

More information

STL GENERATION INTRODUCTION TO STL. Written by Greta D Angelo

STL GENERATION INTRODUCTION TO STL. Written by Greta D Angelo STL GENERATION Written by Greta D Angelo In this section: An introduction on.stl files How to make good.stl files How to make.stl on different softwares (Solidworks, PTC Creo, Rhinoceros 5, Autodesk Inventor)

More information

Full Mouth Restoration with Screw-Retained Zirconia Bridges

Full Mouth Restoration with Screw-Retained Zirconia Bridges Full Mouth Restoration with Screw-Retained Zirconia Bridges A DENTIST-LAB RELATIONSHIP THAT WORKS Dentist: Steven P. Stern, DMD Windsor Dental Center New Windsor, NY Dental Laboratory: InnoDDS Dental Laboratory

More information

DENTAL CAD/CAM 3D PRINTERS

DENTAL CAD/CAM 3D PRINTERS DENTAL CAD/CAM 3D PRINTERS Production of accurate, highly detailed dental prostheses, precision working models, drill guides and orthodontic thermoforming models Enter the Digital Dentistry Era ENHANCE

More information

páêçå~=aéåí~ä=`^al`^j=póëíéã qá_~ëé

páêçå~=aéåí~ä=`^al`^j=póëíéã qá_~ëé kéï=~ë=çñw MSKOMNS páêçå~=aéåí~ä=`^al`^j=póëíéã qá_~ëé lééê~íáåö=fåëíêìåíáçåë båöäáëü=erpf Operating Instructions = Table of contents Sirona Dental Systems GmbH Table of contents 1 Symbols used... 3 2

More information

Precise. Powerful. Pure.

Precise. Powerful. Pure. DATRON D1pure 2 Precise. Powerful. Pure. 3 DATRON Worldwide references p. 32 DATRON DentalCAM p.20 Scanners p.28 CAD p.28 4 Dental materials and Dental milling tools p. 24 DATRON D1 pure p. 7 We Understand

More information

THE OPTICAL CMM 3D SCANNERS FOR INDUSTRIAL APPLICATIONS

THE OPTICAL CMM 3D SCANNERS FOR INDUSTRIAL APPLICATIONS METROLOGY SOLUTIONS THE OPTICAL CMM 3D SCANNERS FOR INDUSTRIAL APPLICATIONS PORTABLE 3D MEASUREMENT SOLUTIONS The arm-free handheld MetraSCAN 3D line-up of optical CMM 3D scanning systems and the C-Track

More information

páêçå~=aéåí~ä=`^al`^j=póëíéã på~åmçëí

páêçå~=aéåí~ä=`^al`^j=póëíéã på~åmçëí kéï=~ë=çñw= MRKOMNP páêçå~=aéåí~ä=`^al`^j=póëíéã på~åmçëí lééê~íáåö=fåëíêìåíáçåë båöäáëü=erpf Operating Instructions = Sirona Dental Systems GmbH Table of contents 1 Symbols used... 3 2 Product description...

More information

Example survey types/use

Example survey types/use Measured surveys 3 rd ed 2014-2 Survey accuracy banding In previous editions of this guidance accuracy has been determined relative to the scale at which a survey map or product will be output. While this

More information

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment by Dr. Ronald Jung and Master Dental Technician Xavier Zahno Initial situation

More information

CMMs and PROFICIENCY TESTING

CMMs and PROFICIENCY TESTING CMMs and PROFICIENCY TESTING Speaker/Author: Dr. Henrik S. Nielsen HN Metrology Consulting, Inc. HN Proficiency Testing, Inc. Indianapolis, Indiana, USA hsnielsen@hn-metrology.com Phone: (317) 849 9577;

More information

Procedures & Products. NobelEsthetics

Procedures & Products. NobelEsthetics Procedures & Products NobelEsthetics First from Nobel Biocare. NobelPerfect, (NP, RP, WP), NobelDirect (NP, RP, WP), Brånemark System, NobelReplace and NobelSpeedy Implants. A complete assortment with

More information

7Series, 5Series & 3Series scanners. DWOS 3.0 user guide. Version 1.0

7Series, 5Series & 3Series scanners. DWOS 3.0 user guide. Version 1.0 7Series, 5Series & 3Series scanners DWOS 3.0 user guide Version 1.0 May 20, 2011 11.2 Implant module The DWOS software implant module provides the tools to design single or multiple custom abutment designs

More information

Nobel Clinician - Quick Guide

Nobel Clinician - Quick Guide 1220 E. Birch St. #201 Brea, CA 92821 (800)750-5004 Haupt Dental Lab Inc. Nobel Clinician - Quick Guide Step #1 - Diagnostics Make impressions of both the upper and lower arches using custom trays. Care

More information

A CAD MODELLING SYSTEM AUTOMATION FOR REVERSE ENGINEERING APPLICATIONS

A CAD MODELLING SYSTEM AUTOMATION FOR REVERSE ENGINEERING APPLICATIONS A CAD MODELLING SYSTEM AUTOMATION FOR REVERSE ENGINEERING APPLICATIONS Jafar Jamshidi, Antony R. Mileham and Geraint W. Owen The University of Bath, UK J.Jamshidi@bath.ac.uk, A.R.Mileham@bath.ac.uk, G.W.Owen@bath.ac.uk

More information

Methods of tolerancing injection-molded parts for illumination systems

Methods of tolerancing injection-molded parts for illumination systems Methods of tolerancing injection-molded parts for illumination systems Mark E. Kaminski *a and R. John Koshel b a Breault Research Organization, Inc., 64 East Grant Road, Suite 35, Tucson, AZ 85715; b

More information

Implant Bar Overdenture Utilizing Locator Attachments

Implant Bar Overdenture Utilizing Locator Attachments Utilizing Locator Attachments Step-by-Step Restorative Protocol Implant Bar Overdentures offer a removable implant solution for edentulous patients desiring a stable and esthetic prosthesis that improves

More information

Mercury TM 1500P PCB-Mount Digital Encoders

Mercury TM 1500P PCB-Mount Digital Encoders Mercury TM 1500P PCB-Mount Digital Encoders Factory Set Resolution to 0.50μm Reflective Linear and Rotary Encoders Sensor the size of a Dime Resolution Factory Set: x4, x8, x20, or x40 Linear: 5μm, 2.5μm,

More information

ATLANTIS ISUS the benchmark for CAD/CAM suprastructures. For all major implant systems

ATLANTIS ISUS the benchmark for CAD/CAM suprastructures. For all major implant systems ATLANTIS ISUS the benchmark for CAD/CAM suprastructures For all major implant systems Patient satisfaction meets clinical benefits ATLANTIS delivers leading CAD/CAM patient-specific restorative solutions

More information

Dental System. Product Catalog

Dental System. Product Catalog Dental System Product Catalog Please see price list for more details June 2015 Dental System Product Catalog Table of Contents Contents Page - 3Shape 3D Lab scanners 3-9 o D500 3D scanner with Dental System

More information

STEP-BY-STEP INSTRUCTIONS ON THE PROSTHETIC PROCEDURES. Straumann Anatomic IPS e.max Abutment

STEP-BY-STEP INSTRUCTIONS ON THE PROSTHETIC PROCEDURES. Straumann Anatomic IPS e.max Abutment STEP-BY-STEP INSTRUCTIONS ON THE PROSTHETIC PROCEDURES Straumann Anatomic IPS e.max Abutment The ITI (International Team for Implantology) is academic partner of Institut Straumann in the areas of research

More information

Prosthetic Manual. Implant System. Contact Us: Smarter thinking. Simpler design. Phone: 866-902-9272 781-328-3490. Fax: 866-903-9272 781-328-3400

Prosthetic Manual. Implant System. Contact Us: Smarter thinking. Simpler design. Phone: 866-902-9272 781-328-3490. Fax: 866-903-9272 781-328-3400 Contact Us: Implant System Phone: 866-902-9272 781-328-3490 Fax: 866-903-9272 781-328-3400 Prosthetic Manual Mail: Keystone Dental, Inc. 144 Middlesex Turnpike Burlington, MA 01803 USA www.keystonedental.com

More information

CROWN ABUTMENT TOOTH (54) (75) (73) (21) (22) (30) Shogo YAMAMOTO, Shinjuku-ku. GC Corporation, ltabashi-ku (JP) Foreign Application Priority Data

CROWN ABUTMENT TOOTH (54) (75) (73) (21) (22) (30) Shogo YAMAMOTO, Shinjuku-ku. GC Corporation, ltabashi-ku (JP) Foreign Application Priority Data US 20070190493A1 (19) United States (12) Patent Application Publication (10) Pub. No.: US 2007/0190493 A1 YAMAMOTO et al. (43) Pub. Date: Aug. 16, 2007 (54) (75) (73) (21) (22) (30) DENTAL PROSTHESIS,

More information

VERSATILE AND EASY-TO-USE 3D LASER SCANNERS > >

VERSATILE AND EASY-TO-USE 3D LASER SCANNERS > > VERSATILE AND EASY-TO-USE 3D LASER SCANNERS > > A WORLD OF APPLICATIONS Z Corporation, producers of the industry s fastest, easiest-to-use and most versatile 3D scanners, makes 3D scanning ideal for a

More information