PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

Size: px
Start display at page:

Download "PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT"

Transcription

1 RP&AD /Remote QA/ 01 PROFORMA FOR ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT BHABHA ATOMIC RESEARCH CENTRE Radiological Physics & Advisory Division CT & CRS Building, Anushaktinagar Mumbai

2 1 1. General Information ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.1 Name and Address of the Institution: (Include Pin Code) Fax: Tel.: Staff (a) Radiotherapist(s) Sr. No. Name Qualification(s) Experience (b) Medical Physicist(s) / Radiological Safety Officer: Sr. No. Name Qualification(s) Experience 1.3 Treatment Unit Type : LDR / MDR / HDR / PDR Make, Model & Number : Maximum capacity of the Unit : MBq/GBq (mci/ci) Radionuclide : 137 Cs / 192 Ir / 60 Co 1

3 2 Physical dimensions of : Length (mm): ; Dia. (mm): source(s) No. of sources : Total & individual activity : µgy/h at 1m (mci) of sources (as quoted by the supplier) : µgy/h at 1m (mci) Dummy sources provided? : Yes / No No. of channels : Max. no. of channels used : Simultaneously Method of source transfer : Pneumatic / Drive cable / Stepping Motor/ between applicator & safe Treatment mode : Fixed / Programmable/ Pulsed /... Method of source retraction : Gravity/ Back-up Battery/ Pneumatic / in the event of power failure Special features (if any) : Nurse's station provided? : Yes / No Plan approved on (date) : Source(s) loaded on (date) : Source(s) loaded by : 2

4 3 2. QA Tests : Mechanical 2.1 Functioning of sensors like : Satisfactory / No pressure, torque, optical 2.2 Coupling between guide- : Satisfactory / No tube and unit 2.3 Coupling between guide- : Satisfactory / No tube and applicator 2.4 Coupling between drive : Satisfactory / No cable and source 2.5 Time taken to drive the source to : sec "ON" position 2.6 Time taken to withdraw the source to : sec "OFF" position 2.7 Source guide free of kinks? : Yes/No 2.8 Mechanical integrity of applicators & : Acceptable / Not acceptable position of shield, if any 3. QA Tests : Electrical 3.1 Interlocks Treatment Room Door(s) Guide Tube & Treatment Unit Applicator & Guide Tube Emergency stop button to interrupt the irradiation 3.2 Source Safe Display Treatment "OFF" indicator Treatment "ON" indicator 3

5 4 3.3 Control Console Display Treatment "OFF" indicator Treatment "ON" indicator Source ON/OFF indicator Door interlock indicator Retrieval of the source indicator Retrieval of the dummy indicator 3.4 Control Console Function Power switch Parameter reset switch Treatment `ON switch Treatment `OFF switch Emergency `OFF switch Treatment and elapsed time display counters Multi-channel indexer function and sequence displays Back-up system(s) 4

6 5 4. QA Tests : Radiation (attach relevant data for evaluation) 4.1 Leakage and Contamination ( Limit: 200 Bq) (Using a wet swipe measure the maximum removable activity on the surface of the source / source holder) Removable activity on Source / Source holder : Bq Contamination check done on (date) : 4.2 Positioning Accuracy and Uniformity (Tolerance: ± 1 mm) (use autoradiograph / radiograph wherever required) Coincidence between dummy : mm and active source positions Accuracy of source position within the : mm applicator Reproducibility of dummy source position : mm 4.3 Temporal Accuracy (Tolerance: ± 1 %) Timer error : Timer linearity : End error : sec ('or' %) 4.4 Source Strength Verification Quoted source strength (as on ) : µgy. h -1 at 1m ( GBq / Ci ) Measured source strength (as on ) : µgy. h -1 at 1m ( GBq / Ci ) Deviation between specified : % and measured source strength (Tolerance: ± 3%) Deviation between measured : % mean and individual source strength (Tolerance: ± 5%) In a separate sheet explain the procedure and the instrument(s) used. In case of multiple sources, give the average source strength and the variation among individual sources. 5

7 6 5. QA Tests: Treatment Planning System 5.1 Digitizer and Localization Software: (Tolerance: 3% or 1mm) Accuracy of digitization of : Point co-ordinates Accuracy of data transfer from : orthogonal radiographs 5.2 Calculation Algorithm: Source specification required for TPS : mci / mg.ra.eq./ MBq / Ref. Air. Kerma Rate Initial activity quoted by the supplier : Current activity (with date) : Factor used to convert quoted source strength for TPS input : Agreement of source decay correction : Agreement between TPS and : published/manual calculation for single source, at relevant points Agreement for multiple source loading : (Specify the procedure and the result) (Tolerance: 3% or 1 mm) 6. QA Tests: Radiation Safety 6.1 Gamma Zone Monitor: Provided? Working? : Yes / No : Yes / No 6.2 CCTV Working? : Yes / No 6.3 Unit Leakage Measure the dose rate on top, bottom and four sides of the container at 5 cm from the surface and one meter from the centre of source storage for source 'OFF condition: Measured dose rate (µgy.h -1 ) at 5cm : from the surface of the source storage (Tolerance: 100 µgy.h -1 ) Measured dose rate (µgy.h -1 ) at 1 m : from the centre of the source storage (Tolerance: 10 µgy.h -1 ) Note: The above tolerance limits are applicable when the unit is loaded with the maximum capacity. 6

8 Survey of the Installation: In the Figure below, mark the occupancy on all sides, above and below the installation and measure the maximum dose rates at these locations, with the applicator loaded with maximum source activity and kept at the centre of the treatment table. Minimum locations recommended for radiation protection survey of the installation are: A, A, B, B Just outside the four walls of the installation at the level of source 'S' C Roof of the installation, at the level of source 'S' D Just outside entrance door E &F At the level of air conditioners G - Entrance to control panel (S : Source position ; CP : Control Panel ) F E A.C. A A.C. B S B F l o o r L e v e l R o o f L e v e l C D C.P A Section Across AA G Fig. 1- Typical Layout of a Remote Afterloading Brachytherapy Facility 7

9 Source Activity (GBq/Ci) during survey : Instrument(s) used for survey : Measured dose rate (µgyh -1 ) at positions marked in Fig. 1: A A B B C D E F G Note: The quoted dose rate should be the maximum measured in the specified direction. 7. Instruments Used for Measurement Sr.No. Type Make &Model Cal. Factor (& Ambient conditions) Date of Calibration Calibrated at Checked by Measured by Signature : Signature : Name : Name : Designation : Designation : Date : Date : : Instrument(s) used for survey : 8

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM Purpose: To provide licensees and registrants with a

More information

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry

More information

Checklist for submission of application form for Registration

Checklist for submission of application form for Registration Checklist for submission of application form for Registration *Incomplete submission may cause delay in processing of the application. Name of the Hospital/Institution: City State Sr.No. Checkpoints Status

More information

HyperAccess Access Control System

HyperAccess Access Control System Control System We manufacture, an advanced PC based access control solution that will give you control over who has access to your building. With you can control access on hundreds of doors for up to 10,000

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

User's Guide. Pinless Moisture Meter. Model MO257

User's Guide. Pinless Moisture Meter. Model MO257 User's Guide Pinless Moisture Meter Model MO257 Introduction Congratulations on your purchase of the Extech MO257 Pinless Moisture Meter. The pinless moisture sensor monitors the moisture in wood and other

More information

HDR Brachytherapy 1: Overview of QA. Disclosures: Learning Objectives 7/23/2014

HDR Brachytherapy 1: Overview of QA. Disclosures: Learning Objectives 7/23/2014 HDR Brachytherapy 1: Overview of QA Bruce Libby Department of Radiation Oncology University of Virginia Health System Disclosures: Nondisclosure agreement with Varian Brachytherapy Shareholder- Varian,

More information

Lessons learned from the radiological accident in Mayapuri, New Delhi, India

Lessons learned from the radiological accident in Mayapuri, New Delhi, India Lessons learned from the radiological accident in Mayapuri, New Delhi, India Rajoo Kumar Atomic Energy Regulatory Board Government of India Mumbai, India www.aerb.gov.in 1 Contents About Mayapuri, New

More information

Monitoring of Intravenous Drip Rate

Monitoring of Intravenous Drip Rate Monitoring of Intravenous Drip Rate Vidyadhar V. Kamble, Prem C. Pandey, Chandrashekar P. Gadgil, and Dinesh S. Choudhary Abstract A drip rate meter, for monitoring intravenous infusion, is developed using

More information

Accreditation Is Coming

Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

REMOTE AFTERLOADING TECHNOLOGY

REMOTE AFTERLOADING TECHNOLOGY AAPM REPORT NO. 41 REMOTE AFTERLOADING TECHNOLOGY Published for the American Association of Physicists in Medicine by the American Institute of Physics REMOTE AFTERLOADING TECHNOLOGY A REPORT OF AAPM TASK

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510 User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Temperature & Humidity SMS Alert Controller

Temperature & Humidity SMS Alert Controller Temperature & Humidity Alert Controller METERS 3 simple steps starting the unit: Insert the SIM card Plug in the sensors connectors Connect the AC power cord. Specifications: AC 90~260V Auto Select Internal

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material)

Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material) Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material) Table of Content Receiving a Radioactive Package Constancy Test Daily Survey Weekly Wipes Accuracy Test Linearity Geometric

More information

MS9000SAN Rework Station

MS9000SAN Rework Station MS9000SAN Rework Station MS9000SAN is the all-round rework system which almost all SMD can be reworked. And the original ITTS auto profiler system is operates of the system easily and exactly. Many kinds,

More information

MINIMAT-EC-Servo Screwdriver Spindles

MINIMAT-EC-Servo Screwdriver Spindles Screwdriving technology Automation Air motors Air tools Screwdriver Spindles electric MINIMAT-EC-Servo Screwdriver Spindles Maximum Flexibility and process Control Straight Spindle Form - Torque range

More information

TABLE 1: Wiring Terminals. Connect to... 1C 1H 2C 2H 1H1C 2H1C 2H2C 3H2C

TABLE 1: Wiring Terminals. Connect to... 1C 1H 2C 2H 1H1C 2H1C 2H2C 3H2C Installation TURN OFF POWER TO THE SYSTEM AT THE MAIN POWER PANEL TO AVOID ELECTRICAL SHOCK. Installation should be carried out by an electrician or a qualified technician. 1.1 Find a Location for the

More information

SERVICE INSTRUCTION R410A. WALL MOUNTEDtype INVERTER SPLIT TYPE ROOM AIR CONDITIONER. Models Indoor unit Outdoor unit

SERVICE INSTRUCTION R410A. WALL MOUNTEDtype INVERTER SPLIT TYPE ROOM AIR CONDITIONER. Models Indoor unit Outdoor unit SERVICE INSTRUCTION SPLIT TYPE ROOM AIR CONDITIONER WALL MOUNTEDtype INVERTER Models Indoor unit Outdoor unit ASYG07LECA ASYG09LECA ASYG12LECA ASYG14LECA AOYG07LEC AOYG09LEC AOYG12LEC AOYG14LEC R410A CONTENTS

More information

TS510 & TS500. Installation & User Guide. Compatible Equipment

TS510 & TS500. Installation & User Guide. Compatible Equipment Installation & User Guide Compatible Equipment TS510 REM - Remote Keypad 9040 - Loudspeaker DC54/58 - Digital Communicator SD1+ - Speech Dialler 496525 Issue A 1 of 10 TS510 and TS500 Overview Introduction

More information

ODYSSEY. Security System Owner s Manual. Kit No. 08E51-SHJ-100 08E55-SHJ-100. 2004 American Honda Motor Co., Inc. - All Rights Reserved.

ODYSSEY. Security System Owner s Manual. Kit No. 08E51-SHJ-100 08E55-SHJ-100. 2004 American Honda Motor Co., Inc. - All Rights Reserved. Kit No. 08E5-SHJ-00 08E55-SHJ-00 Security System Owner s Manual ODYSSEY 004 American Honda Motor Co., Inc. - All Rights Reserved. Contents Introduction... 3 Emergency Disarming During the Panic Alarm Activation...

More information

Cronos -1 Gamma Object / Tool Monitors

Cronos -1 Gamma Object / Tool Monitors Features Counting chamber volume: 42.9 L (1.5 cu. ft.) Robust, ergonomic, easy-to-use and decontaminate Photon energy range from 50 kev to 2 MeV Four or optional six large surface area plastic scintillator

More information

FLOW CALCULATOR INSTRUCTION MANUAL MESURES BAMOPHOX 759 26-06-2007 759 M1 02 E MES FLOW CALCULATOR 759-02/1

FLOW CALCULATOR INSTRUCTION MANUAL MESURES BAMOPHOX 759 26-06-2007 759 M1 02 E MES FLOW CALCULATOR 759-02/1 BAMOPHOX 759 E - M FLOW CALCULATOR INSTRUCTION MANUAL MESURES 22, Rue de la Voie des Bans - Z.I. de la Gare - 95100 ARGENTEUIL Tél : (33) 01 30 25 83 20 - Web : www.bamo.fr Fax : (33) 01 34 10 16 05 -

More information

Regulatory Guide. Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0

Regulatory Guide. Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0 Regulatory Guide Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0 Federal Authority for Nuclear Regulation (FANR) P.O. Box 112021, Abu Dhabi, United Arab Emirates regulation@fanr.gov.ae

More information

Operation Manual. Electrical Release Control System E-CO2. for Gas Fire Extingushing Systems. according FSS-Code Chapter 5

Operation Manual. Electrical Release Control System E-CO2. for Gas Fire Extingushing Systems. according FSS-Code Chapter 5 Operation Manual Electrical Release Control System E-CO2 for Gas Fire Extingushing Systems according FSS-Code Chapter 5 02 20 Kidde Marine Brandschutzsysteme All rights reserved. No part of this document

More information

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Fernanda B. C. Nonato, Claudinei T. Cescon and Linda V. E. Caldas, Instituto de Pesquisas Energéticas e Nucleares,

More information

Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains

Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains Drayton Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains Invensys Controls Europe Technical Helpline: +44 (0) 845 130 7722 www.draytoncontrols.co.uk Installation

More information

Portable Air Conditioner

Portable Air Conditioner Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE

More information

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 CHAPTER 227. RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT, X-RAY GAUGING EQUIPMENT, ELECTRON MICROSCOPES AND X-RAY CALIBRATION SYSTEMS Sec. 227.1.

More information

RVL470. Heating Controller. Building Technologies HVAC Products. Series B

RVL470. Heating Controller. Building Technologies HVAC Products. Series B 2 522 Heating Controller Series B RVL470 Multifunctional heating controller for use in residential and non-residential buildings; suitable for weather-dependent flow temperature control of heating zones

More information

Material test chamber with mechanical convection

Material test chamber with mechanical convection FP series 400 Material test chambers Material test chamber with mechanical convection A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and

More information

3-Component Measuring System

3-Component Measuring System Force 3-omponent Measuring System for utting Force Measurement During Turning Modular system for measurement of cutting forces when turning outside and inside diameters on turret lathes. Machine adapters

More information

BB-18 Black Body High Vacuum System Technical Description

BB-18 Black Body High Vacuum System Technical Description BB-18 Black Body High Vacuum System Technical Description The BB-18 Black Body is versatile and is programmed for use as a fixed cold target at 80 K or variable target, at 80 K- 350 K no extra cost. The

More information

Atomic Alarm Clock Instruction Manual

Atomic Alarm Clock Instruction Manual Atomic Alarm Clock Instruction Manual Introduction: Congratulations on purchasing the Atomic Alarm Clock with WWVB Radio controlled time, calendar, 12/24 hours display, time zone with US map, indoor temperature

More information

Radiation Monitoring System RMS

Radiation Monitoring System RMS Radiation Monitoring System RMS VF RMS radiation monitoring system is a modular and standardized system for the radiological monitoring. Variety of radiation monitors, display and alarm units can be connected

More information

EXTERNAL DIMENSIONS AND PARTS

EXTERNAL DIMENSIONS AND PARTS FX2N-4DA SPECIAL FUNCTION BLOCK USER S GUIDE JY992D65901A This manual contains text, diagrams and explanations which will guide the reader in the correct installation and operation of the FX2N-4DA special

More information

Drayton Digistat +2RF/+3RF

Drayton Digistat +2RF/+3RF /+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer

More information

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Atomic Energy Regulatory Board Mumbai 400 094 1 DEFINITIONS 1. Quality Assurance (QA): Planned and systematic actions

More information

SOLARCARE SERIES PRODUCT AND APPLICATION GUIDE

SOLARCARE SERIES PRODUCT AND APPLICATION GUIDE SOLARCARE SERIES PRODUCT AND APPLICATION GUIDE for solar energy management LEATEC Delivering Solutions for Energy Management SOLAR ENERGY DATA CENTER BUILDING 4 to8 String Monitoring with 0.% Accuracy

More information

API Marine. BMS Presentation. Ballast Measurement Systems

API Marine. BMS Presentation. Ballast Measurement Systems API Marine BMS Presentation Ballast Measurement Systems BMS4 Typical Configuration BMS4 Typical Configuration BMS4 Extended Configuration BMS4 System (24 tanks) Air regulator Terminal block Drain/Air Filter

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Homologation Directive for TT04 Vehicle Tracking Systems

Homologation Directive for TT04 Vehicle Tracking Systems Homologation Directive for TT04 Vehicle Tracking Systems VERSION 3.0 (31 March 2014) - Requirements and testing methods for vehicle tracking systems Issued by the Insurance Bureau for Vehicle Crime (Stichting

More information

Hygro-Thermometer + InfraRed Thermometer Model RH101

Hygro-Thermometer + InfraRed Thermometer Model RH101 User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative

More information

COMBIMASS. Technical Data COMBIMASS eco-bio +

COMBIMASS. Technical Data COMBIMASS eco-bio + COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments

More information

Transponder reader CÉLIANE TECHNICAL GUIDE

Transponder reader CÉLIANE TECHNICAL GUIDE CÉLIAE TECHICAL GUIDE Description The transponder reader is a device that activates when the appropriate badge is placed in front of it (1-2 cm). The badge is the transponder. The signal generated by placing

More information

Model SETR-50 and SETR-51 Trim Tab Control

Model SETR-50 and SETR-51 Trim Tab Control Model SETR-50 and SETR-51 Trim Tab Control Pictured above is the SETR-50 with black switches on a gray background. The SETR-51 is identical except for the color, wherein it has black switches on a black

More information

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for

More information

Selection, use and maintenance of portable monitoring instruments

Selection, use and maintenance of portable monitoring instruments HSE information sheet Selection, use and maintenance of portable monitoring instruments Ionising Radiation Protection Series No 7 Introduction The Ionising Radiations Regulations 1999 (IRR99) 1 require

More information

T7079A,B Solid State Remote Temperature Controllers

T7079A,B Solid State Remote Temperature Controllers T7079A,B Solid State Remote Temperature Controllers FEATURES PRODUCT DATA Switch selection of heat or cool mode. Temperature sensing up to 400 feet. Does not require field calibration. 0K NTC temperature

More information

APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMMAND: PERMIT NUMBER: TYPE OF ACTION:

APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMMAND: PERMIT NUMBER: TYPE OF ACTION: APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMM: PERMIT NUMBER: TYPE OF ACTION: NAVY MARINE CPS PUBLIC HEALTH CENTER REVISED NOVEMBER 2007 REVIEWER/DATE: SECOND REVIEW/DATE:

More information

Vehicle Tracking Service & Fleet Management

Vehicle Tracking Service & Fleet Management Vehicle Tracking Service & Fleet Management License No: BTRC/LL/Vehicle Tracking (07) AKASH/2010-5 Ref No: 00120101708 Akash Technology 102, Sankar West Dhanmondi, Ground Floor, Dhaka- 1207 Phone: 02-9113201

More information

User Manual. Video Doorphone CDV-1020AQ

User Manual. Video Doorphone CDV-1020AQ User Manual Video Doorphone CDV-1020AQ 513-11, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Int l Business Dept. Tel. : +82-31-7393-540~550 Fax. : +82-31-745-2133 Web site : www.commax.com

More information

HERZ-Thermal Actuators

HERZ-Thermal Actuators HERZ-Thermal Actuators Data Sheet 7708-7990, Issue 1011 Dimensions in mm 1 7710 00 1 7710 01 1 7711 18 1 7710 80 1 7710 81 1 7711 80 1 7711 81 1 7990 00 1 7980 00 1 7708 11 1 7708 10 1 7708 23 1 7709 01

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510

User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510 User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

CruzPro VAF110. AC Volts, Amps, Frequency, kw Monitor

CruzPro VAF110. AC Volts, Amps, Frequency, kw Monitor CruzPro VAF110 AC Volts, Amps, Frequency, kw Monitor Table of Contents Introduction............................ 3 Specifications........................... 4 Installation..............................5

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Auto Dialer. Manual E-921APQ E-921GPQ

Auto Dialer. Manual E-921APQ E-921GPQ Troubleshooting: Auto dialer will not arm/disarm Auto dialer will not dial out Unit doesn t respond to a call-back Difficulty in activating room monitor by telephone remote control Make sure that you have

More information

Electro-Pneumatic Bus Door Control Mechanism

Electro-Pneumatic Bus Door Control Mechanism Electro-Pneumatic Bus Door Control Mechanism Application and Feature 1. Pneumatic Door Mechanism actuates the open/close movements of bus door through two double acting cylinders and two linkages with

More information

Centro-Matic Automatic Lubrication Systems System Controls

Centro-Matic Automatic Lubrication Systems System Controls Selecting the right controls for your automated lubrication system is one of the last steps in the design process. Several different models may be chosen to control power-operated pumps, depending on the

More information

SATEX SA 230. - Manual. Weegtechniek holland b.v. PRECISION BALANCE. Patroonsweg 23-27 3892 DA Zeewolde HOLLAND

SATEX SA 230. - Manual. Weegtechniek holland b.v. PRECISION BALANCE. Patroonsweg 23-27 3892 DA Zeewolde HOLLAND - Manual SATEX SA 230 PRECISION BALANCE Weegtechniek holland b.v. Patroonsweg 23-27 3892 DA Zeewolde HOLLAND Tel. + 31 36 522 20 30 Fax. + 31 36 522 20 60 Email info@satex.nl Website www.satex.nl page

More information

BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEM EURO ALARM WIRELESS BURGLAR ALARM SYSTEM OASIS

BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEM EURO ALARM WIRELESS BURGLAR ALARM SYSTEM OASIS BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEM EURO ALARM WIRELESS BURGLAR ALARM SYSTEM OASIS BURGLAR ALARM SYSTEM EURO ALARM Six zones with the following connection capabilities: One

More information

WEATHER STATION FOR SOLAR FARM MONITORING

WEATHER STATION FOR SOLAR FARM MONITORING WEATHER STATION FOR SOLAR FARM MONITORING SOLAR FARM MONITORING SYSTEM: Measures global, horizontal, & background irradiance. Measures wind speed, wind direction, ambient temperature, and relative humidity.

More information

Daily Operator Checks. Monthly Maintenance

Daily Operator Checks. Monthly Maintenance CUSTOMER DATE OF INSPECTION SERIAL # RSF # UI Rev # PLC Project # SERVICE REPRESENTATIVE Daily Operator Checks Verify that all safety guards are in place. Verify that all safety signs are in proper location

More information

Quality Assurance Testing of AtomLab Dose Calibrators

Quality Assurance Testing of AtomLab Dose Calibrators (Manufacturer s* Instructions) Introduction The following procedures are the manufacturer s instructions for qualifying AtomLab 100, 100+, 200, 300, 400, and 500 dose calibrators in accordance with NRC

More information

SERVICE INSTRUCTION R410A. WALL MOUNTEDtype SPLIT TYPE ROOM AIR CONDITIONER INVERTER. Models Indoor unit Outdoor unit AOU 9RLFW AOU12RLFW AOU15RLS

SERVICE INSTRUCTION R410A. WALL MOUNTEDtype SPLIT TYPE ROOM AIR CONDITIONER INVERTER. Models Indoor unit Outdoor unit AOU 9RLFW AOU12RLFW AOU15RLS SERVICE INSTRUCTION SPLIT TYPE ROOM AIR CONDITIONER WALL MOUNTEDtype INVERTER Models Indoor unit Outdoor unit ASU 9RLF ASURLF ASU5RLS AOU 9RLFW AOURLFW AOU5RLS R40A CONTENTS. DESCRIPTION OF EACH CONTROL

More information

SECTION 13850 DETECTION AND ALARM

SECTION 13850 DETECTION AND ALARM SECTION 13850 DETECTION AND ALARM PART 1 GENERAL 1.01 SUMMARY A. Section Includes 1. Control Panel 2 Associated Equipment B. Products Installed But Not Supplied Under This Section 1. Section 16140 - Wiring

More information

User Manual of WT 493

User Manual of WT 493 User Manual of WT 493 Introduction Congratulations on your purchase of this smart unit. This unit can be used as a radio to catch the latest news and enjoy the beautiful music; it can also be used as an

More information

Wespot. SecNurse A NEW GENERATION TECHNICAL AIDS USER S GUIDE

Wespot. SecNurse A NEW GENERATION TECHNICAL AIDS USER S GUIDE Wespot SecNurse A NEW GENERATION TECHNICAL AIDS USER S GUIDE EN This manual is published by OPTEX, without any warranty. Improvements and changes to this manual necessitated by typographical errors, inaccuracies

More information

Captus 3000 Thyroid Uptake And Well System

Captus 3000 Thyroid Uptake And Well System (201) 825-9500 / (800) 631-3826 THRYOID UPTAKE SYSTEMS 51 Captus 3000 Thyroid Uptake And Well System The Ultimate... User Friendly... Intuitive Performing Thyroid Uptake measurements has never been easier.

More information

OPERATING INSTRUCTIONS SECURITY SYSTEM KIT NO.: 08E51-EP4-101. 2004 American Honda Motor Co., Inc. - All Rights Reserved. 1

OPERATING INSTRUCTIONS SECURITY SYSTEM KIT NO.: 08E51-EP4-101. 2004 American Honda Motor Co., Inc. - All Rights Reserved. 1 OPERATING INSTRUCTIONS SECURITY SYSTEM KIT NO.: 8E5-EP4-24 American Honda Motor Co., Inc. - All Rights Reserved. Contents Introduction... 3 Emergency Disarming During Alarming... 4 During the Entry Delay

More information

CITY OF PHILADELPHIA DEPARTMENT OF LICENSES AND INSPECTIONS ANNUAL CERTIFICATION FOR FIRE ALARM SYSTEMS

CITY OF PHILADELPHIA DEPARTMENT OF LICENSES AND INSPECTIONS ANNUAL CERTIFICATION FOR FIRE ALARM SYSTEMS CITY OF PHILADELPHIA DEPARTMENT OF LICENSES AND INSPECTIONS ANNUAL CERTIFICATION FOR FIRE ALARM SYSTEMS PROPERTY ADDRESS (BRT Address Required) TESTING CONTRACTOR (Name and Address) License No. ANNUAL

More information

GSM AD05 Slave GSM Auto Dialer- Instruction Manual

GSM AD05 Slave GSM Auto Dialer- Instruction Manual GSM AD05 Slave GSM Auto Dialer- Instruction Manual Please read these instructions before you start the installation Features LCD display Programmable 9 x 32 digit phone numbers for each trigger. 10 second

More information

contents Leading the way in intelligent automation INELS Wireless? Why Click Accessories Complete control

contents Leading the way in intelligent automation INELS Wireless? Why Click Accessories Complete control Wireless Control Systems Accessories Leading the way in intelligent automation contents Range Overview Typical Installations Transmitters Combined Modules Accessories Programming Switching & Dimmer Functions

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

DS1104 R&D Controller Board

DS1104 R&D Controller Board DS1104 R&D Controller Board Cost-effective system for controller development Highlights Single-board system with real-time hardware and comprehensive I/O Cost-effective PCI hardware for use in PCs Application

More information

Incubator with mechanical convection

Incubator with mechanical convection BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,

More information

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION Calibration Guide PREPARATION Adventurer Pro Adventurer Pro balances have two calibration methods: Span and Linearity. Some models have Internal Calibration (InCal ) for Span calibration. Ensure that the

More information

Model 5511 Filler Controller User s Manual Version 1.1 October 2011

Model 5511 Filler Controller User s Manual Version 1.1 October 2011 Thompson Scale Company WEIGHING SYSTEMS & PACKAGING MACHINERY CONTROLS 2758 Bingle Road Houston, Texas 77055 Phone: 713/932-9071 Fax: 713/932-9379 www.thompsonscale.com Model 5511 Filler Controller User

More information

3-phase, bidirectional energy meter

3-phase, bidirectional energy meter Data sheet www.sbc-support.com 3-phase, bidirectional energy meter with S0 pulse output Bidirectional energy meter with S0-interface. The S0 interface is a hardware interface for the transmission of measured

More information

AD425S Automatic door operator

AD425S Automatic door operator EC8666 Statement Automatic Door Operator is a reliable and quiet door opening and closing device. This product is ideal for enabling access to areas for people with disabilities or simply for convenience

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

ICP-CC488. Installation Guide ICP-CC488 Control Panel

ICP-CC488. Installation Guide ICP-CC488 Control Panel ICP-CC488 EN Installation Guide ICP-CC488 Control Panel ICP-CC488 Installation Guide Notices EN 2 Copyright Notice Unless otherwise indicated, this publication is the copyright of Bosch Security Systems,

More information

GSM HOME SECURITY SYSTEM

GSM HOME SECURITY SYSTEM Cell /Mobile phone home security system GSM HOME SECURITY SYSTEM Model : GSM-120 TABLE OF CONTENTS 1. FEATURES... 1 2. APPLICATION... 2 3. SPECIFICATIONS... 3 4. FRONT PANEL & LAYOUT DESCRIPTION...6 5.

More information

Users Manual FA106 / FA306 version 1.0. Field Analyzer. FA106 and FA306. Users Manual

Users Manual FA106 / FA306 version 1.0. Field Analyzer. FA106 and FA306. Users Manual Field Analyzer FA106 and FA306 Users Manual 1 Contents Contents...2 Preface...3 Chapter 1 Basic Functions...4 1.1 Turning on/off...4 1.2 Choosing the required measuring mode...4 1.3 Picture of front panel...5

More information

9452/9453 Installation and User Guide

9452/9453 Installation and User Guide 9452/9453 Installation and User Guide Compatible Equipment 9425 Remote Keypad 9040 Internal Sounder 660 Speech Communicator 8440 4-Channel Minicom 496330 Issue 1 1 of 10 9452/3 Introduction The 9452 and

More information

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407780 Product Page Integrating Sound Level Datalogger

More information

How To Program An Autodialer

How To Program An Autodialer GJD HYL005 GSM Autodialer Instruction Manual Please read these instructions before you start the installation Features: LCD display. Programmable 9 x 32 digit phone numbers for each trigger. 10 second

More information

Galleon 6-10K Online UPS. Voltronic Power Technology Corporation

Galleon 6-10K Online UPS. Voltronic Power Technology Corporation Galleon 6-10K Online UPS Voltronic Power Technology Corporation Features Galleon online UPS series delivers optimal power quality, provides superior power management, resolves utility power protection

More information

WA Manager Alarming System Management Software Windows 98, NT, XP, 2000 User Guide

WA Manager Alarming System Management Software Windows 98, NT, XP, 2000 User Guide WA Manager Alarming System Management Software Windows 98, NT, XP, 2000 User Guide Version 2.1, 4/2010 Disclaimer While every effort has been made to ensure that the information in this guide is accurate

More information

Operating Manual for the Electronic Built-in Interval Timer. Micro II (Countdown Timer)

Operating Manual for the Electronic Built-in Interval Timer. Micro II (Countdown Timer) Operating Manual for the Electronic Built-in Interval Timer Micro II (Countdown Timer) Note: This document has been designed for our OEM customers. They can use it as supporting material when creating

More information

Machine Safety Switches

Machine Safety Switches Machine Switches SI-LS31H Series 31 mm Limit-Switch-Style with Hinged Lever Actuators Features Limit switch design (EN 50047) For use on doors or flaps Positive opening safety contacts (IEC 60947-5-1)

More information

Integrated Custom Systems

Integrated Custom Systems Integrated Custom Systems Compact panels Standard Manifolds & Assemblies Electronic pressure control Custom manifolds & Assemblies Automated test benches S t r e a m l i n e y o u r p r o c e s s Complete

More information

SENTRON Switching and Protection Devices Air Circuit Breakers

SENTRON Switching and Protection Devices Air Circuit Breakers Siemens AG 9 SENTRON Switching and rotection Devices Air Circuit reakers 5 5/ Introduction 3WL Air Circuit reakers 3WL Air Circuit reakers/ Non-Automatic Air Circuit reakers up to 63 A (AC) 5/6 General

More information