Zebrafish needle EMG: a new tool for high-throughput drug screens

Size: px
Start display at page:

Download "Zebrafish needle EMG: a new tool for high-throughput drug screens"

Transcription

1 J Neurophysiol 114: , First published July 15, 2015; doi: /jn Zebrafish needle EMG: a new tool for high-throughput drug screens Sung-Joon Cho, 1 Tai-Seung Nam, 2 Donghak Byun, 3 Seok-Yong Choi, 4 Myeong-Kyu Kim, 2 * and Sohee Kim 1,3 * 1 Department of Medical System Engineering, Gwangju Institute of Science and Technology, Gwangju, Korea; 2 Department of Neurology, Chonnam National University Medical School, Gwangju, Korea; 3 School of Mechatronics, Gwangju Institute of Science and Technology, Gwangju, Korea; and 4 Department of Biomedical Sciences, Chonnam National University Medical School, Gwangju, Korea Submitted 2 June 2015; accepted in final form 15 July 2015 Cho SJ, Nam TS, Byun D, Choi SY, Kim MK, Kim S. Zebrafish needle EMG: a new tool for high-throughput drug screens. J Neurophysiol 114: , First published July 15, 2015; doi: /jn Zebrafish models have recently been highlighted as a valuable tool in studying the molecular basis of neuromuscular diseases and developing new pharmacological treatments. Needle electromyography (EMG) is needed not only for validating transgenic zebrafish models with muscular dystrophies (MD), but also for assessing the efficacy of therapeutics. However, performing needle EMG on larval zebrafish has not been feasible due to the lack of proper EMG sensors and systems for such small animals. We introduce a new type of EMG needle electrode to measure intramuscular activities of larval zebrafish, together with a method to hold the animal in position during EMG, without anesthetization. The silicon-based needle electrode was found to be sufficiently strong and sharp to penetrate the skin and muscles of zebrafish larvae, and its shape and performance did not change after multiple insertions. With the use of the proposed needle electrode and measurement system, EMG was successfully performed on zebrafish at 30 days postfertilization (dpf) and at 5 dpf. Burst patterns and spike morphology of the recorded EMG signals were analyzed. The measured single spikes were triphasic with an initial positive deflection, which is typical for motor unit action potentials, with durations of 10 ms, whereas the muscle activity was silent during the anesthetized condition. These findings confirmed the capability of this system of detecting EMG signals from very small animals such as 5 dpf zebrafish. The developed EMG sensor and system are expected to become a helpful tool in validating zebrafish MD models and further developing therapeutics. needle electromyography; zebrafish; microneedle; electrode; motor unit action potential RECENTLY, IT HAS BEEN RECOGNIZED that using small vertebrates such as zebrafish embryos or larvae is an efficient and economical way to study the fundamental pathomechanisms of neuromuscular diseases with subsequent therapy development (Gibbs et al. 2013; Lieschke and Currie 2007; Santoriello and Zon 2012; Dowling et al. 2009; White et al. 2011; Hortopan et al. 2010; Baraban et al. 2007; Baraban et al. 2013). One of the most attractive aspects of using the zebrafish model is that mass screening of chemicals in a large population is possible, since they have fast growth cycles, generate a large number of offspring, and are convenient to observe the entire body due to * M.-K. Kim and S. Kim contributed equally to this work. Address for reprint requests and other correspondence: S. Kim, Dept. of Medical System Engineering, Gwangju Institute of Science and Technology, 123 Cheomdan-gwagiro, Buk-gu, Gwangju , Korea (Republic of) ( soheekim@gist.ac.kr). their optical transparency and ease of facilitating transgenic disease models. Electromyography (EMG) is the most effective diagnostic tool for neuromuscular disorders, including motor neuron diseases, radiculopathies, peripheral neuropathies, and myopathies (Griggs et al. 2012). Needle EMG to detect motor unit action potentials (MUAPs) in skeletal muscles is preferred to surface EMG for the accurate diagnosis of neuromuscular diseases, monitoring of the progress of such diseases, and assessment of the responses to specific treatments. The morphology and firing pattern of MUAPs through needle EMG not only reflect the nature of the underlying diseases but also provide information about the associated pathophysiological mechanisms. Thus, needle EMG may not only be needed for validating transgenic zebrafish models with muscular dystrophies, but also for assessing the efficacy of therapeutics. However, needle EMG on embryonic or larval zebrafish to record skeletal muscle activities has not been reported to date. This is presumably due to the lack of EMG sensors and immobilization methods suitable for such small animals, although there have been reports on EMG in juvenile and adult zebrafish (Liu and Westerfield 1988; Gabriel et al. 2008; Kyriakatos et al. 2011). To overcome the aforementioned technical challenges and to extend EMG to larval zebrafish, we introduce a new EMG sensor and measurement system that can be used repeatedly on very small vertebrates by adapting silicon-based microfabrication technologies. First, the fabrication of the silicon microneedle electrode is explained, followed by the description of the entire measurement system, including the manipulators to precisely handle the small animals. Next, the electrical characterization of the microneedle electrode is performed, and, finally, the recorded EMG signals from larval zebrafish are demonstrated. MATERIALS AND METHODS Electrode fabrication. Figure 1 illustrates the process flow for the fabrication of the EMG electrode. The needle electrode was fabricated based on silicon, which can provide durability and reusability of the electrode. Conventional methods of generating silicon needles using a dicing machine and wet etching were adopted (Byun et al. 2013). A 2-mm-thick silicon wafer was cut orthogonally with a depth of 1,700 m to create square shanks, using a dicing saw machine (NDS-200; AM Technology, Asan, Korea). Grooves with a depth of 50 m and a width of 200 m were created between the shanks (Fig. 1A). To convert the square silicon shanks into sharp conical needles, isotropic wet etching was performed using an etching solution of HF (48%) and HNO 3 (69%) in a volume ratio of 1:19, as described previously (Byun /15 Copyright 2015 the American Physiological Society 2065

2 2066 ZEBRAFISH NEEDLE EMG Fig. 1. Fabrication processes of the Si-based microneedle electrode. Dicing a 2-mm-thick Si wafer to create shanks and grooves (A), wet etching to convert shanks into conical needles (B), separation of needles by pushing the groove (C), deposition of Cr/Au/Ir in thicknesses of 50/200/500 nm (D), selective deposition of parylene C on all areas except the needle tip (E), and wiring of the needle electrode and a connection pad on a dummy Si chip (F). et al. 2013) (Fig. 1B). Subsequently, the individual needles were physically separated by applying force to the grooves (Fig. 1C). As adhesion layers, Cr and Au were deposited in a thickness of 50 and 200 nm, respectively, and 500-nm-thick iridium was deposited to lower the electrical impedance of the needles (Fig. 1D). Parylene C was deposited on the needles using a chemical vapor deposition system (PDS 2010; Specialty Coating Systems, Indianapolis, IN) for electrical insulation and improved biocompatibility. Part of the needle base was masked using tape (3M Korea, Seoul, Korea) during parylene C deposition, for the purpose of later electrical connections. Parylene C was etched on the metalized tip of the needles by oxygen plasma through reactive ion etching (Vita-E; FemtoScience, Suwon, Korea) by employing a customized etch mask (Fig. 1E), which resulted in an exposed tip length of 100 m. The needle base was attached to a dummy silicon chip in a dimension of mm, using adhesive epoxy (Metal Adhesive; 3M, Maplewood, MN). A small region of the needle base that had not been coated with parylene C and the copper pad on the dummy silicon chip were soldered together using copper wire (Fig. 1F). To close the current circuit of the Si microneedle electrode, a reference electrode was required, which needs to be in contact with the skin of the animal. Depending on the diameter of the larval body, coil-shaped electrodes with different diameters were used as the reference as well as ground electrodes. For 30 days postfertilization (dpf) larvae, a stainless steel spiral coil with a diameter of 2 mm was used. A spring was cut by 5 cm using a nipper, and both ends were straightened for electrical connection to the measurement equipment. For 5 dpf larvae, silver-plated copper wire with a diameter of 250 m was wound manually around a 1-mm-diameter cylinder and used as the reference and ground electrodes, as shown in Fig. 2B. Electrical characterization. To characterize the electrical properties of the fabricated needle electrodes, electrochemical impedance spectroscopy was performed using an electrochemical measurement system (Reference 600; Gamry, Warminster, PA) in the frequency range from 1 to 10 khz. The needle electrode was inserted in 0.8% agarose gel (Agarose-LE; USB, Cleveland, OH), and a platinum electrode was used as a counter electrode. The 10-mV AC voltage with no DC offset was applied to the electrode. The impedance of the needle electrodes was measured and compared with that of a commercially available monopolar EMG electrode (Technomed Disposable Monopolar Needle Electrode; Cadwell, Kennewick, WA). Animal preparation. Wild-type zebrafish (AB strain) were obtained from the Zebrafish International Resource Center (Eugene, OR), maintained under a 14:10-h light-dark cycle, and staged in dpf as per standard criteria (Kimmel et al. 1995). Animal studies were approved by the Chonnam National University Medical School Institutional Animal Care and Use Committee. Zebrafish at 30 dpf, roughly 1 mm wide and up to 8 mm long, and zebrafish at 5 dpf, 0.2 mm wide and 3.9 mm long on average, were used in EMG recordings. To hold the larva and position the reference electrode in contact with the skin of the animal, we injected the larva in a coil-shaped electrode as shown in Fig. 2, using a pipette with a diameter of 2 mm (Citotest, Nanjing, China). With this approach, the motion of the larva could be restricted without using anesthetic agents that could paralyze the muscles of the animal. To prevent the dehydration of the fish skin, water was injected along with the larva. The skin of the fish remained wet, since water was present due to the surface tension of water that filled the space between the coil turns. Additionally, water was dropped on the skin surface occasionally during recordings. Micromanipulation. A customized insertion system was developed to manipulate the small animal and the delicate needle electrode on a micro scale, as shown in Fig. 2. The system consisted of two arms: a subject holder and an electrode manipulator. The animal was placed on the subject holder attached to a rotational stage with angular resolution of 0.1, whereas the microneedle electrode was attached to the electrode manipulator including X-, Y-, and Z-directional stages with resolution of motion of 0.1 m. The backside of the dummy chip of the needle electrode was bonded to a 3-mm-thick severed rubber magnet (Sheet Type NBR Magnet; Woosung Magnets, Gimhae, Korea) using epoxy, and the rubber magnet was attached to a neodymium magnet (Woosung Magnets) that was affixed to the electrode Fig. 2. Customized micromanipulation system of the electromyography (EMG) needle electrode. A: schematic diagram of the entire system. The electrode manipulator holds the working needle electrode using magnets while the subject holder stabilizes the coil-shaped reference electrode where zebrafish is contained. The electrode manipulator includes precise X-, Y-, and Z-stages, and the subject holder includes a rotational stage. The arrangement of the coilshaped electrode and zebrafish was along the Z-direction, whereas the direction of needle insertion was parallel to the X-direction. B: photographs of the microneedle electrode and the coil-shaped electrode where a 5 days postfertilization (dpf) larva was contained. The black-headed arrow in the image on the right indicates the direction of needle insertion.

3 ZEBRAFISH NEEDLE EMG 2067 Fig. 3. Images of a fabricated needle electrode. Microscopic image of a needle after wet etching, with a length of 1,300 m and a base diameter of 70 m (A), and scanning electron microscopic image of an iridium-coated tip of the needle electrode with parylene C insulation on all areas except the tip (B). A length of 100 m on the needle tip was exposed. manipulator arm. This method generated no electromagnetic noise or mechanical vibration and simplified the attachment and detachment of the electrode without the need for sophisticated procedures. EMG recording. Under direct visual guidance, the needle electrode was inserted in the axial myomeres of a larva along the lateral side using the manipulator while the needle electrode and the coil-shaped reference electrode were connected to a commercially available EMG amplifier (The EMG SpikerBox; Backyard Brains, Ann Arbor, MI) with gain of up to 1,000. EMG signals from six larvae at 30 dpf and seven larvae at 5 dpf were recorded for up to 3 min. The signals were bandpass filtered with a frequency range from 150 to 5 khz and were displayed and stored on a laptop using NeuronRecorder software (Backyard Brains). In addition, they were transformed to sound waves and monitored using a speaker. For more quantitative analysis, duration of EMG bursts, number of spikes per burst, and interburst interval were analyzed using Spike2 software (Cambridge Electronic Design, Cambridge, UK). Fast-Fourier transform analysis was performed using MATLAB (MathWorks, Natick, MA) to display the power spectrum of the recorded data in frequency domain. Additionally, zebrafish muscle activities were monitored before and after application of 0.02% tricaine (MS-222; Sigma Aldrich, St. Louis, MO). Tricaine was applied by dropping it on the animal that had been restricted within the coil-shaped electrode and removed by rinsing it from the animal with fish-system water. RESULTS Electrode fabrication and impedance measurement. Figure 3 shows the images of the fabricated needles with an average length of 1,300 m and a base diameter of about 70 m (Fig. 3A). An average length of 100 m with SD of m was exposed on the needle tips (Fig. 3B). Figure 4 shows the impedance of the fabricated electrodes coated with iridium compared with the impedance of the commercial monopolar electrode. The relationship between electrode impedance and electrode diameter or tip exposure area has been studied previously (Fontes 2013; Negi et al. 2012; Negi et al. 2010; Cogan 2008; Grubbs and Worley 1983). For all electrodes, the impedance increases with the decrease of effective recording area, and a trade-off between sensitivity and selectivity always remains (Negi et al. 2010). The iridium-coated needle electrodes showed stable impedance in all frequency ranges. At 100 Hz, the electrode impedance was 749 k on average, with an SD of 25 k. This was four times higher than but on the same order of the impedance of the commercial electrode, which was 187 k at 100 Hz. Electrode insertion. The coil-shaped electrode containing a larva was placed on the subject holder, as shown in Fig. 2. During the EMG procedure, all larvae remained inside the coil-shaped electrode without slipping or escaping. The microneedle electrode was inserted in the larva using the precision manipulator. With the X-, Y-, and Z-directional stages and rotational stage, precise manipulation of the animal and the needle electrode was possible in all degrees of freedom of motion with micrometer-range precision. The needle electrode was sufficiently sharp and rigid to penetrate the skin and muscle of the larvae. Additionally, the needle electrode displayed no damage upon multiple rounds of insertions, confirming the durability and reusability of the electrode. EMG recording. EMG signals were successfully measured from larval zebrafish at 5 dpf (n 7) and 30 dpf (n 6) (data not shown) for up to 3 min. Figure 5 shows the representative examples of the recorded muscle activities of 5 dpf zebrafish along with quantitative analysis on burst characteristics. Figure 5A displays acceptably stable recordings over 100 s, with magnified views to show burst patterns. In Fig. 5B, 77% of bursts lasted for 20 ms, and the bursts lasting longer than 40 ms were only 10% of the recorded data while the longest burst detected was 80 ms, which was also observed in the magnified views of Fig. 5A. This burst duration is much shorter than that of adult zebrafish, which lasted up to 120 ms when they struggled in a stressful environment (Liu and Westerfield 1988). The detected burst durations can be explained by the number of spikes in a burst as shown in Fig. 5C and the fact that the duration of single spikes was measured to be 10 ms as shown in Fig. 6. The analysis of interburst interval in Fig. 5D shows that bursts obtained from larvae appeared to be in a more irregular fashion compared with bursts from adult zebrafish. In the literature (Liu and Westerfield 1998), adult zebrafish showed an interburst interval of 20 ms during regular swimming, and the majority of interburst interval was shorter than 40 ms. In Fig. 6, the recorded EMG signals are shown with magnified time scales to better present the signal morphology. The morphology of the measured signals was triphasic, with an initial positive deflection followed by a negative phase, which is typical for MUAPs obtained through needle EMG in humans (Trontelj et al. 2004; Preston and Shapiro 2013). The duration of MUAPs was found to be 10 ms, which is consistent with the findings in humans. Because the insertional activity when the needle quickly moves through muscles is known to last for a few hundreds of milliseconds according to human studies Fig. 4. Impedance of the fabricated needle electrodes and that of a commercial monopolar electrode for comparison. Bars represent the SDs (n 5). At 100 Hz, the electrode impedance was 749 k on average, with an SD of 25 k, which was four times higher than but on the same order of the impedance of the commercial electrode.

4 2068 ZEBRAFISH NEEDLE EMG Fig. 5. Spontaneous EMG recordings from 5 dpf zebrafish along with quantitative analysis on burst characteristics obtained from 372 bursts. A: recorded EMG data for 100 s where the ordinate is in arbitrary units, and the magnified views to show the burst patterns with the minimum interval between bursts of 20 ms and the longest burst duration of 80 ms. Percentage of occurrences depending on burst duration (B), no. of spikes/burst (C), and interburst interval (D). (Preston and Shapiro 2013), we excluded the possibility that the measured activities were caused by needle movements. The frequency analysis showed that the recorded signals were concentrated in the frequency range 1 khz. For comparison, the frequency content of needle EMG signals in humans is reported to be 2 khz (Trontelj et al. 2004). It is noteworthy that the frequency range of zebrafish EMG signals is not available in the literature. After recording, the fish soon started swimming again when they had been returned to water, suggesting that the extent of muscle damage inflicted by electrode insertion was minimal. In another set of experiments using tricaine to observe the effect of anesthetization, EMG signals disappeared upon tricaine application and recovered after it had been washed out, validating that the measured signals indeed reflected spontaneous muscle activities. DISCUSSION In this study, a new type of microneedle electrode for EMG recording from very small vertebrates such as larval zebrafish was introduced. The microneedle electrode was fabricated by mechanical dicing and wet etching of 2-mm-thick silicon. With the use of the developed manipulator system, the insertion of the fabricated EMG needle in zebrafish larvae was possible. The needle electrodes were durable, sufficiently sharp and rigid to penetrate the skin and muscles of larvae, and maintained their shapes and function after multiple rounds of insertions. The most significant challenge when using this EMG measurement system was to hold the animal in position during the EMG procedure. Initially, we had attempted to immobilize the zebrafish larvae in an agarose block and to put the reference and ground electrodes in contact with the agarose. However, this method elicited significant noise whose amplitude was higher than that of the EMG signal itself. Subsequently, our second attempt involved using the coil-shaped electrode in contact with the skin of the animal to function as both the reference and ground electrodes. This electrode also acted as a fixation apparatus during larval EMG, thereby eliminating the need for anesthesia. First-needle EMG on larval zebrafish. Although the amplitude of the measured muscle activities could not be specified, the analysis of the waveform, burst pattern, spike duration, and frequency content of the measured signals confirmed the feasibility of measuring spontaneous EMG signals using the microneedle electrode and the coil-shaped electrode introduced

5 ZEBRAFISH NEEDLE EMG 2069 Fig. 6. Examples of spontaneous EMG recordings from 5 dpf zebrafish with magnified time scales. The ordinate is in arbitrary units. The displayed recordings are for 3 s, and the magnified views are for 50 ms. The magnified waveforms show the typical triphasic waveform of motor unit action potentials with an initial positive deflection followed by a negative phase and with durations of 10 ms. in the present study. We believe that the differences in EMG signals from adult and larval zebrafish are caused by different swimming or activity styles of adults and larvae. Larvae tend to do single tail flicks, and they come to a total standstill before initiating the next tail flick while adults flick more frequently and initiate the next tail flick before total standstill when they try to swim (Muller et al. 2000). Also, swimming activity increases with age and size. Larvae tend to rest while juveniles and adults swim all the time. It is known that, with 5 dpf larvae, cyclic swimming occurred rarely and only after strong startle stimuli (Batty 1984; Muller et al. 2004). Benefits of the proposed technique. Previously, needle EMG for zebrafish had been performed only in juveniles or adults (Liu and Westerfield 1998; Gabriel et al. 2008; Kyriakatos et al. 2011), which used hooked platinum wires in a diameter of 50 m or glass suction electrodes in a diameter of about 20 m in combination with skin removal to expose the muscle fibers. In those studies, the fish were anesthetized and immobilized using glue or nylon mesh. However, these immobilization methods are difficult to apply to larval zebrafish due to their small dimensions. Our proposed method to restrict the animal within the coil-shaped electrode resolved this challenge in holding the larva in position during EMG. Also, there have been attempts to obtain electrophysiological signals from the muscles of embryonic zebrafish by single cell electrophysiology using in vivo patch-clamp techniques (Hirata et al. 2007; Hirata et al. 2004, 2004; Buss and Drapeau 2000; Gabriel et al. 2009) although the nature of the signals from single cells is different from that of EMG signals. The use of glass suction electrodes for EMG in adult zebrafish and in vivo patch recording from embryonic zebrafish entailed anesthetization and skin preparation to access the underlying muscles. In contrast, the needle electrode and measurement system proposed in the present study enabled the performance of EMG on zebrafish larvae without anesthetization, immobilization, and skin preparation of the animal, which thus is expected to accelerate the whole recording sessions. Limitations of the proposed technique. The proposed needle EMG technique presents a few limitations. The amplitude of EMG signals could not be specified due to the limitation of the EMG acquisition system used in the present study. However, this can be resolved if a more accurate EMG amplifier is employed such as the clinical EMG equipment. In this study, we performed the recordings for up to 3 min to make sure that the larvae were alive during and after EMG recordings. This duration seems to be sufficient, since it is even longer than the duration of clinical needle EMG tests (Heatwole et al. 2013; Nandedkar 2013; Drost et al. 2015). However, longer recordings would be possible in future studies, which may include investigations on respiration maintenance of the larvae within the water droplet formed inside the coil-shaped electrode. Another limitation of our proposed method is that swimming activities of larvae are hardly observable during EMG recordings because the animals are located inside the coil-shaped electrode. As a result, it is hard to compare the EMG data with swimming activities. In addition, with the use of the proposed technique, EMG signals from zebrafish younger than 5 dpf have not been measured. To measure EMG signals from younger animals such as 1 dpf embryos, the silicon-based EMG sensor needs to become smaller in diameter, which fabrication, however, is challenging. Significance and perspective. With the use of the proposed EMG system, intramuscular activities, including MUAPs, were successfully captured for the first time from zebrafish at an age as early as 5 dpf. We also demonstrated that, by adjusting the diameter of the coil-shaped electrode, EMG signals from zebrafish larvae of various sizes could be measured. Prospectively, the capability of performing EMG on larval zebrafish as small as 5 pdf provides a new electrophysiological means that is essential for validating zebrafish models with neuromuscular diseases, investigating fundamental pathomechanisms, and further assessing the efficacy of therapeutics. Eventually, the proposed EMG system is expected to facilitate high-throughput drug screens using larval zebrafish with neuromuscular diseases. ACKNOWLEDGMENTS We thank Keonghwan Oh and Youngseok Kim for preparation of coilshaped electrodes and help with experimental setup.

6 2070 ZEBRAFISH NEEDLE EMG GRANTS This research was supported by grants from the Basic Science Research Program of the National Research Foundation (NRF-2014R1A1A ), the institute of Medical System Engineering of the Gwangju Institute of Science and Technology, and the Chonnam National University Hospital Biomedical Research Institute (CRI ), Korea. DISCLOSURES No conflicts of interest, financial or otherwise, are declared by the authors. AUTHOR CONTRIBUTIONS Author contributions: S.-J.C., T.-S.N., D.B., S.-Y.C., and S.K. performed experiments; S.-J.C. and S.K. analyzed data; S.-J.C. and D.B. prepared figures; S.-J.C. drafted manuscript; T.-S.N., S.-Y.C., M.-K.K., and S.K. conception and design of research; T.-S.N. and S.K. interpreted results of experiments; T.-S.N., S.-Y.C., M.-K.K., and S.K. edited and revised manuscript; S.K. approved final version of manuscript. REFERENCES Baraban SC, Dinday MT, Castro PA, Chege S, Guyenet S, Taylor MR. A large-scale mutagenesis screen to identify seizure-resistant zebrafish. Epilepsia 48: , Baraban SC, Dinday MT, Hortopan GA. Drug screening in Scn1a zebrafish mutant identifies clemizole as a potential Dravet syndrome treatment. Nat Comm 4: 2410, Batty R. Development of swimming movements and musculature of larval herring (Clupea harengus). J Exp Biol 110: , Buss RR, Drapeau P. Physiological properties of zebrafish embryonic red and white muscle fibers during early development. J Neurophysiol 84: , Byun D, Cho SJ, Kim S. Fabrication of a flexible penetrating microelectrode array for use on curved surfaces of neural tissues. J Micromechan Microeng 23: , Cogan S. Neural Stimulation and Recording Electrodes. Annu Rev Biomed Eng 10: , Dowling JJ, Vreede AP, Low SE, Gibbs EM, Kuwada JY, Bonnemann CG, Feldman EL. Loss of myotubularin function results in T-tubule disorganization in zebrafish and human myotubular myopathy. PLoS Genet 5: e , Drost G, Stunnenberg BC, Trip J, Borm G, McGill KC, Ginjaar IHB, van der Kooi AW, Zwarts MJ, van Engelen BGM, Faber CG, Stegeman DF, Lateva Z. Myotonic discharges discriminate chloride from sodium muscle channelopathies. Neuromuscul Disord 25: 73 80, Fontes M. Electrodes for bio-application: recording and stimulation. J Physics Conf Ser 421: , Gabriel JP, Mahmood R, Kyriakatos A, Söll I, Hauptmann G, Calabrese RL, El Manira A. Serotonergic modulation of locomotion in zebrafish: endogenous release and synaptic mechanisms. J Neurosci 29: , Gabriel JP, Mahmood R, Walter AM, Kyriakatos A, Hauptmann G, Calabrese RL, El Manira A. Locomotor pattern in the adult zebrafish spinal cord in vitro. J Neurophysiology 99: 37 48, Gibbs EM, Horstick EJ, Dowling JJ. Swimming into prominence: the zebrafish as a valuable tool for studying human myopathies and muscular dystrophies. Fed Eur Biochem Soc J 280: , Griggs RC, Jozefowicz RF, Aminoff MJ. Approach to the patient with neurologic disease. In: Cecil Medicine, edited by Goldman L, Ausiello D. New York, NY: Elsevier, Grubbs D, Worley D. New technique for reducing the impedance of silversilver chloride electrodes. Med Biol Eng Comput 21: , Heatwole CR, Statland JM, Logigian EL. The diagnosis and treatment of myotonic disorders. Muscle Nerve 47: , Hirata H, Watanabe T, Hatakeyama J, Sprague SM, Saint-Amant L, Nagashima A, Cui WW, Zhou W, Kuwada JY. Zebrafish relatively relaxed mutants have a ryanodine receptor defect, show slow swimming and provide a model of multi-minicore disease. Development 134: , Hirata H, Saint-Amant L, Waterbury J, Cui W, Zhou W, Li Q, Goldman D, Granato M, Kuwada JY. Accordion, a zebrafish behavioral mutant, has a muscle relaxation defect due to a mutation in the ATPase Ca2 pump SERCA1. Development 131: , Hortopan GA, Dinday MT, Baraban SC. Spontaneous seizures and altered gene expression in GABA signaling pathways in a mind bomb mutant zebrafish. J Neurosci 30: , Kimmel CB, Ballard WW, Kimmel SR, Ullmann B, Schilling TF. Stages of embryonic development of the zebrafish. Dev Dyn 203: , Kyriakatos A, Mahmood R, Ausborn J, Porres CP, Bueschges A, Manira AE. Initiation of locomotion in adult zebrafish. J Neurosci 31: , Lieschke PD, Currie GJ. Animal models of human disease: zebrafish swim into view. Nat Rev Genet 8: , Liu DW, Westerfield M. Function of identified motoneurones and co-ordination of primary and secondary motor systems during zebrafish swimming. J Physiol 403: 73 89, Muller U. Swimming of larval zebrafish: ontogeny of body waves and implications for locomotory development. J Exp Biol 207: , Muller U, Stamhuis E, Videler J. Hydrodynamics of unsteady fish swimming and the effects of body size: comparing the flow fields of fish larvae and adults. J Exp Biol 203: , Nandedkar SD. Emerging techniques in the electrodiagnostic laboratory. Phys Med Rehabil 5: S115 S122, Negi S, Bhandari R, Rieth L, Solzbacher F. In vitro comparison of sputtered iridium oxide and platinum-coated neural implantable microelectrode arrays. Biomed Materials 5: , Negi S, Bhandari R, Solzbacher F. Morphology and electrochemical properties of activated and sputtered iridium oxide films for functional electrostimulation. J Sensor Technol 2: , Preston DC, Shapiro BE. Electromyography and Neuromuscular Disorders: Clinical-Electrophysiological Correlations. New York, NY: Elsevier, Santoriello LI, Zon C. Hooked! Modeling human disease in zebrafish. J Clin Invest 122: , Trontelj JV, Jabre J, Mihelin M. Needle and wire detection techniques. In: Electromyography: Physiology, Engineering, and Non-invasive Applications, edited by Merletti R, Parker PA. New York, NY: Wiley, White RM, Cech J, Ratanasirintrawoot S, Lin CY, Rahl PB, Burke CJ, Langdon E, Tomlinson ML, Mosher J, Kaufman C, Chen F, Long HK, Kramer M, Datta S, Neuberg D, Granter S, Young RA, Morrison S, Wheeler GN, Zon LI. DHODH modulates transcriptional elongation in the neural crest and melanoma. Nature 471: , 2011.

Laboratory Guide. Anatomy and Physiology

Laboratory Guide. Anatomy and Physiology Laboratory Guide Anatomy and Physiology TBME04, Fall 2010 Name: Passed: Last updated 2010-08-13 Department of Biomedical Engineering Linköpings Universitet Introduction This laboratory session is intended

More information

Fundamentals of Electromyography. Amanda Peltier, MD MS Department of Neurology

Fundamentals of Electromyography. Amanda Peltier, MD MS Department of Neurology Fundamentals of Electromyography Amanda Peltier, MD MS Department of Neurology Importance of EMG Studies Diagnosis Localization Assist in further testing (i.e. identify potential biopsy sites) Prognosis

More information

How to measure absolute pressure using piezoresistive sensing elements

How to measure absolute pressure using piezoresistive sensing elements In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article

More information

Piezoelectric Driven Non-toxic Injector for Automated Cell Manipulation

Piezoelectric Driven Non-toxic Injector for Automated Cell Manipulation Medicine Meets Virtual Reality 18 J.D. Westwood et al. (Eds.) IOS Press, 2011 2011 The authors. All rights reserved. doi:10.3233/978-1-60750-706-2-231 231 Piezoelectric Driven Non-toxic Injector for Automated

More information

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is: 4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing

More information

DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT

DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT BACKGROUND AND PRODUCT DESCRIPTIONS Fracture Diagnostics provides state-of-the-art test equipment, utilizing the Rising Step Load testing technique.

More information

Lapping and Polishing Basics

Lapping and Polishing Basics Lapping and Polishing Basics Applications Laboratory Report 54 Lapping and Polishing 1.0: Introduction Lapping and polishing is a process by which material is precisely removed from a workpiece (or specimen)

More information

Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator

Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator Kenji MAWATARI, Koich SAMESHIMA, Mitsuyoshi MIYAI, Shinya MATSUDA Abstract We developed a new inkjet head by applying MEMS

More information

Author: Dr. Society of Electrophysio. Reference: Electrodes. should include: electrode shape size use. direction.

Author: Dr. Society of Electrophysio. Reference: Electrodes. should include: electrode shape size use. direction. Standards for Reportin ng EMG Data Author: Dr. Roberto Merletti, Politecnico di Torino, Italy The Standards for Reporting EMG Data, written by Dr. Robertoo Merletti, are endorsed by the International Society

More information

Silicon-On-Glass MEMS. Design. Handbook

Silicon-On-Glass MEMS. Design. Handbook Silicon-On-Glass MEMS Design Handbook A Process Module for a Multi-User Service Program A Michigan Nanofabrication Facility process at the University of Michigan March 2007 TABLE OF CONTENTS Chapter 1...

More information

ZEBRAFISH SYSTEMS SYSTEMS ZEBRAFISH. Fast, automatic analyses

ZEBRAFISH SYSTEMS SYSTEMS ZEBRAFISH. Fast, automatic analyses ZEBRAFISH SYSTEMS ZEBRAFISH SYSTEMS Fast, automatic analyses Wide-scale evaluation possibilities ( averaged ECG, QT, QTv, HRV) Distortion-free ECG signal transfer Social interaction examinations Measurements

More information

GE Healthcare. Quick Guide PDF PROOF. Neuromuscular Transmission (NMT)

GE Healthcare. Quick Guide PDF PROOF. Neuromuscular Transmission (NMT) GE Healthcare Quick Guide Neuromuscular Transmission (NMT) WHAT is Neuromuscular Transmission (NMT)? Neuromuscular Transmission (NMT) is the transfer of an impulse between a nerve and a muscle in the neuromuscular

More information

Tuesday 20 May 2014 Morning

Tuesday 20 May 2014 Morning Tuesday 20 May 2014 Morning AS GCE PHYSICS B (ADVANCING PHYSICS) G491/01 Physics in Action *1203458796* Candidates answer on the Question Paper. OCR supplied materials: Data, Formulae and Relationships

More information

Formation of Oriented Fibers Using Injection of PEO Solutions inside Electric Fields Defined by Two Parallel Suspended Electrodes

Formation of Oriented Fibers Using Injection of PEO Solutions inside Electric Fields Defined by Two Parallel Suspended Electrodes 06 (43)-AF:Modelo-AF 8/20/11 6:41 AM Page 122 Formation of Oriented Fibers Using Injection of PEO Solutions inside Electric Fields Defined by Two Parallel Suspended R. Furlan 1, J. A. M. Rosado 2, A. N.

More information

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist. Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist. Abstract. I have been asked to put together a detailed article on a SWR meter. In this article I will deal

More information

Electrophysiological Recording Techniques

Electrophysiological Recording Techniques Electrophysiological Recording Techniques Wen-Jun Gao, PH.D. Drexel University College of Medicine Goal of Physiological Recording To detect the communication signals between neurons in real time (μs to

More information

Axial and Radial Leaded Multilayer Ceramic Capacitors for Automotive Applications Class 1 and Class 2, 50 V DC, 100 V DC and 200 V DC

Axial and Radial Leaded Multilayer Ceramic Capacitors for Automotive Applications Class 1 and Class 2, 50 V DC, 100 V DC and 200 V DC Axial and Radial Leaded Multilayer Ceramic Capacitors for Automotive Applications Class 1 and Class 2, 5 V DC, 1 V DC and 2 V DC DESIGNING For more than 2 years Vitramon has supported the automotive industry

More information

STM and AFM Tutorial. Katie Mitchell January 20, 2010

STM and AFM Tutorial. Katie Mitchell January 20, 2010 STM and AFM Tutorial Katie Mitchell January 20, 2010 Overview Scanning Probe Microscopes Scanning Tunneling Microscopy (STM) Atomic Force Microscopy (AFM) Contact AFM Non-contact AFM RHK UHV350 AFM/STM

More information

ELECTROMYOGRAPHY (EMG), NEEDLE, NERVE CONDUCTION STUDIES (NCS) AND QUANTITATIVE SENSORY TESTING (QST)

ELECTROMYOGRAPHY (EMG), NEEDLE, NERVE CONDUCTION STUDIES (NCS) AND QUANTITATIVE SENSORY TESTING (QST) AND QUANTITATIVE SENSORY TESTING (QST) Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan. This Medical

More information

Cumbria Designs T-1. SSB/CW Filter kit (4.9152MHz) User Manual

Cumbria Designs T-1. SSB/CW Filter kit (4.9152MHz) User Manual Cumbria Designs T-1 SSB/CW Filter kit (4.9152MHz) User Manual CONTENTS 1 INTRODUCTION 2 2 CIRCUIT DESCRIPTION 2 3 ASSEMBLY 2 4 TESTING 4 The Steading Stainton PENRITH Cumbria CA11 0ES UK 1 Introduction

More information

Welding of Plastics. Amit Mukund Joshi. (B.E Mechanical, A.M.I.Prod.E)

Welding of Plastics. Amit Mukund Joshi. (B.E Mechanical, A.M.I.Prod.E) Welding of Plastics Amit Mukund Joshi (B.E Mechanical, A.M.I.Prod.E) Introduction Mechanical fasteners, adhesives, and welding processes can all be employed to form joints between engineering plastics.

More information

EMG and the Electrodiagnostic Consultation for the Family Physician

EMG and the Electrodiagnostic Consultation for the Family Physician EMG and the Electrodiagnostic Consultation for the Family Physician Stephanie Kopey, D.O., P.T. 9/27/15 The American Association of Neuromuscular & Electrodiagnostic Medicine (AANEM) Marketing Committee

More information

Electrodiagnostic Testing

Electrodiagnostic Testing Electrodiagnostic Testing Electromyogram and Nerve Conduction Study North American Spine Society Public Education Series What Is Electrodiagnostic Testing? The term electrodiagnostic testing covers a

More information

ECEN 1400, Introduction to Analog and Digital Electronics

ECEN 1400, Introduction to Analog and Digital Electronics ECEN 1400, Introduction to Analog and Digital Electronics Lab 4: Power supply 1 INTRODUCTION This lab will span two lab periods. In this lab, you will create the power supply that transforms the AC wall

More information

A Remote Plasma Sputter Process for High Rate Web Coating of Low Temperature Plastic Film with High Quality Thin Film Metals and Insulators

A Remote Plasma Sputter Process for High Rate Web Coating of Low Temperature Plastic Film with High Quality Thin Film Metals and Insulators A Remote Plasma Sputter Process for High Rate Web Coating of Low Temperature Plastic Film with High Quality Thin Film Metals and Insulators Dr Peter Hockley and Professor Mike Thwaites, Plasma Quest Limited

More information

Active noise control in practice: transformer station

Active noise control in practice: transformer station Active noise control in practice: transformer station Edwin Buikema 1 ; Fokke D. van der Ploeg 2 ; Jan H. Granneman 3 1, 2, 3 Peutz bv, Netherlands ABSTRACT Based on literature and extensive measurements

More information

Rapid Prototyping and Development of Microfluidic and BioMEMS Devices

Rapid Prototyping and Development of Microfluidic and BioMEMS Devices Rapid Prototyping and Development of Microfluidic and BioMEMS Devices J. Sasserath and D. Fries Intelligent Micro Patterning System Solutions, LLC St. Petersburg, Florida (T) 727-522-0334 (F) 727-522-3896

More information

Fabrication of Complex Circuit Using Electrochemical Micromachining on Printed Circuit Board (PCB)

Fabrication of Complex Circuit Using Electrochemical Micromachining on Printed Circuit Board (PCB) 5 th International & 26 th All India Manufacturing Technology, Design and Research Conference (AIMTDR 2014) December 12 th 14 th, 2014, IIT Guwahati, Assam, India Fabrication of Complex Circuit Using Electrochemical

More information

Product Introduction MyoMuscle. Telemyo DTS-System. Telemetry system for EMG and Biomechanical Sensors. www.noraxon.com

Product Introduction MyoMuscle. Telemyo DTS-System. Telemetry system for EMG and Biomechanical Sensors. www.noraxon.com Product Introduction MyoMuscle Telemyo DTS-System Telemetry system for EMG and Biomechanical Sensors DTS Device Family 3 major device categories can be used for the Direct Transmission System technology

More information

The Calculation of G rms

The Calculation of G rms The Calculation of G rms QualMark Corp. Neill Doertenbach The metric of G rms is typically used to specify and compare the energy in repetitive shock vibration systems. However, the method of arriving

More information

Coating Technology: Evaporation Vs Sputtering

Coating Technology: Evaporation Vs Sputtering Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information

More information

ERGONOMIC FIELD ASSESSMENT OF BUCKING BARS DURING RIVETING TASKS

ERGONOMIC FIELD ASSESSMENT OF BUCKING BARS DURING RIVETING TASKS PROCEEDINGS of the HUMAN FACTORS AND ERGONOMICS SOCIETY 49th ANNUAL MEETING 2005 1354 ERGONOMIC FIELD ASSESSMENT OF BUCKING BARS DURING RIVETING TASKS Michael J. Jorgensen and Muthukurappan Viswanathan

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

Ion Beam Sputtering: Practical Applications to Electron Microscopy

Ion Beam Sputtering: Practical Applications to Electron Microscopy Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a

More information

Electrospun jets launched from polymeric bubbles

Electrospun jets launched from polymeric bubbles Electrospun jets launched from polymeric bubbles J.S. Varabhas a, S. Tripatanasuwan b, G.G. Chase a,*, D.H. Reneker b a Department of Chemical and Biomolecular Engineering, The University of Akron, Akron,

More information

SpeedLight 2D. for efficient production of printed circuit boards

SpeedLight 2D. for efficient production of printed circuit boards laser direct imaging SpeedLight 2D laser direct imaging platform for efficient production of printed circuit boards MANZ AG /// Manz SpeedLight 2D /// 2 History of the development of Manz SpeedLight 2D

More information

Anodes and Misc Equipment

Anodes and Misc Equipment Anodes and Misc Equipment Application: Platinised Titanium Anodes Platinised titanium anodes are recommended for use in the following electrolytic processes:- Precious metal electroplating - e.g. Au, Pt,

More information

DualBeam Solutions for Electrical Nanoprobing

DualBeam Solutions for Electrical Nanoprobing DualBeam Solutions for Electrical Nanoprobing Richard J. Young, Technologist Peter D. Carleson, Product Marketing Engineer Electrical testing by physically probing device structures has grown more challenging

More information

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS Objective In this experiment you will study the i-v characteristics of an MOS transistor. You will use the MOSFET as a variable resistor and as a switch. BACKGROUND

More information

Surface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student

Surface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student 1 Surface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student Abstract-- As the film decreases in thickness the requirements of more

More information

Nano Meter Stepping Drive of Surface Acoustic Wave Motor

Nano Meter Stepping Drive of Surface Acoustic Wave Motor Proc. of 1st IEEE Conf. on Nanotechnology, Oct. 28-3, pp. 495-5, (21) Maui, Hawaii Nano Meter Stepping Drive of Surface Acoustic Wave Motor Takashi Shigematsu*, Minoru Kuribayashi Kurosawa*, and Katsuhiko

More information

Photolithography. Class: Figure 12.1. Various ways in which dust particles can interfere with photomask patterns.

Photolithography. Class: Figure 12.1. Various ways in which dust particles can interfere with photomask patterns. Photolithography Figure 12.1. Various ways in which dust particles can interfere with photomask patterns. 19/11/2003 Ettore Vittone- Fisica dei Semiconduttori - Lectio XIII 16 Figure 12.2. Particle-size

More information

Neuro imaging: looking with lasers in the brain

Neuro imaging: looking with lasers in the brain Neuro imaging: looking with lasers in the brain Aim: To image life cells, label free, with cellular resolution in deep tissue Marloes Groot Vrije Universiteit Amsterdam Faculteit Exacte Wetenschappen Natuurkunde

More information

Impedance Matching and Matching Networks. Valentin Todorow, December, 2009

Impedance Matching and Matching Networks. Valentin Todorow, December, 2009 Impedance Matching and Matching Networks Valentin Todorow, December, 2009 RF for Plasma Processing - Definition of RF What is RF? The IEEE Standard Dictionary of Electrical and Electronics Terms defines

More information

Generator Stator Protection, under/over voltage, under /over frequency and unbalanced loading. Ramandeep Kaur Aujla S.NO 250447392

Generator Stator Protection, under/over voltage, under /over frequency and unbalanced loading. Ramandeep Kaur Aujla S.NO 250447392 1 Generator Stator Protection, under/over voltage, under /over frequency and unbalanced loading By Ramandeep Kaur Aujla S.NO 250447392 ES 586b: Theory and applications of protective relays Department of

More information

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter

More information

INFRARED PARTS MANUAL

INFRARED PARTS MANUAL INFRARED PARTS MANUAL PIR325 FL65 GLOLAB CORPORATION Thank you for buying our Pyroelectric Infrared components. The goal of Glolab is to produce top quality electronic kits, products and components. All

More information

MRI SAFETY JEOPARDY (Technologist Edition) Questions

MRI SAFETY JEOPARDY (Technologist Edition) Questions MRI SAFETY JEOPARDY (Technologist Edition) Questions For MRI Safety Week, 2009 Provided, Inc. www.mednovus.com PowerPoint MRI Safety Quiz available for free download from http://mrimetaldetector.com/media/proving_grounds/mri_safety_week.html

More information

TechCut 4 Precision Low Speed Saw

TechCut 4 Precision Low Speed Saw Product Brochure TechCut 4 Precision Low Speed Saw 3" - 6" Blade Range Digital Speed Display 1-Micron Sample Indexing Spring-Loaded Dressing Stick Attachment All Aluminum & Stainless Steel Construction

More information

Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge

Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge , Firenze, Italy Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge A. Bulletti, L. Capineri B. Dunn ESTEC Material and Process

More information

Standards for surface electromyography: the European project "Surface EMG for non-invasive assessment of muscles (SENIAM)

Standards for surface electromyography: the European project Surface EMG for non-invasive assessment of muscles (SENIAM) Standards for surface electromyography: the European project "Surface EMG for non-invasive assessment of muscles (SENIAM) D.F. Stegeman 1,3, H.J. Hermens 2 1 Institute of Neurology, Department of Clinical

More information

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.

More information

Femtosecond laser-induced silicon surface morphology in water confinement

Femtosecond laser-induced silicon surface morphology in water confinement Microsyst Technol (2009) 15:1045 1049 DOI 10.1007/s00542-009-0880-8 TECHNICAL PAPER Femtosecond laser-induced silicon surface morphology in water confinement Yukun Han Æ Cheng-Hsiang Lin Æ Hai Xiao Æ Hai-Lung

More information

I-PEX MHF4 Micro Coaxial Connector

I-PEX MHF4 Micro Coaxial Connector PRODUCT SPECIFICATION I-PEX MHF4 Micro Coaxial Connector Plug P/N 20448-001R-81 Receptacle P/N 20449-001E RF Connector, RF Cable & Antenna Manufacturer E-mail: sales@wellshow.com.tw Tel: +886-2-24239376

More information

Supporting information

Supporting information Supporting information Ultrafast room-temperature NH 3 sensing with positively-gated reduced graphene oxide field-effect transistors Ganhua Lu 1, Kehan Yu 1, Leonidas E. Ocola 2, and Junhong Chen 1 * 1

More information

3nd Biennial Contemporary Clinical Neurophysiological Symposium October 12, 2013 Fundamentals of NCS and NMJ Testing

3nd Biennial Contemporary Clinical Neurophysiological Symposium October 12, 2013 Fundamentals of NCS and NMJ Testing 3nd Biennial Contemporary Clinical Neurophysiological Symposium October 12, 2013 Fundamentals of NCS and NMJ Testing Peter D. Donofrio, M.D. Professor of Neurology Vanderbilt University Medical Center

More information

Miniaturizing Flexible Circuits for use in Medical Electronics. Nate Kreutter 3M

Miniaturizing Flexible Circuits for use in Medical Electronics. Nate Kreutter 3M Miniaturizing Flexible Circuits for use in Medical Electronics Nate Kreutter 3M Drivers for Medical Miniaturization Market Drivers for Increased use of Medical Electronics Aging Population Early Detection

More information

Interference in the ECG and its elimination

Interference in the ECG and its elimination Interference in the ECG and its elimination General points The signal voltages in ECG recording are known to be very small and are in the millivolt range, i.e. they have amplitudes of only a few thousands

More information

The quadrature signals and the index pulse are accessed through five 0.025 inch square pins located on 0.1 inch centers.

The quadrature signals and the index pulse are accessed through five 0.025 inch square pins located on 0.1 inch centers. Quick Assembly Two and Three Channel Optical Encoders Technical Data HEDM-550x/560x HEDS-550x/554x HEDS-560x/564x Features Two Channel Quadrature Output with Optional Index Pulse Quick and Easy Assembly

More information

Introduction to VLSI Fabrication Technologies. Emanuele Baravelli

Introduction to VLSI Fabrication Technologies. Emanuele Baravelli Introduction to VLSI Fabrication Technologies Emanuele Baravelli 27/09/2005 Organization Materials Used in VLSI Fabrication VLSI Fabrication Technologies Overview of Fabrication Methods Device simulation

More information

Sense it! Connect it! Bus it! Solve it! EncoderS

Sense it! Connect it! Bus it! Solve it! EncoderS Sense it! Connect it! Bus it! Solve it! EncoderS Incremental encoders Incremental encoders use electrical pulses to measure rotation speed or position. The dual-channel incremental encoders of the Ri series,

More information

The Electrical Control of Chemical Reactions E3-1

The Electrical Control of Chemical Reactions E3-1 Experiment 3 The Electrical Control of Chemical Reactions E3-1 E3-2 The Task In this experiment you will explore the processes of oxidation and reduction, in which electrons flow between materials, and

More information

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b DIODE CIRCUITS LABORATORY A solid state diode consists of a junction of either dissimilar semiconductors (pn junction diode) or a metal and a semiconductor (Schottky barrier diode). Regardless of the type,

More information

Supporting Information

Supporting Information Supporting Information Wiley-VCH 2007 69451 Weinheim, Germany Methanol Behavior in Direct Methanol Fuel Cells Younkee Paik, Seong-Soo Kim, and Oc Hee Han * Experimental Section Preparation of MEA: Standard

More information

MCE380: Measurements and Instrumentation Lab. Chapter 9: Force, Torque and Strain Measurements

MCE380: Measurements and Instrumentation Lab. Chapter 9: Force, Torque and Strain Measurements MCE380: Measurements and Instrumentation Lab Chapter 9: Force, Torque and Strain Measurements Topics: Elastic Elements for Force Measurement Dynamometers and Brakes Resistance Strain Gages Holman, Ch.

More information

Investigation of the Residual-charge Technique for Diagnosis of Water-tree Deteriorated Cross-linked Polyethylene Cable

Investigation of the Residual-charge Technique for Diagnosis of Water-tree Deteriorated Cross-linked Polyethylene Cable Investigation of the Residual-charge Technique for Diagnosis of Water-tree Deteriorated Cross-linked Polyethylene Cable Dr. Uchida Katsumi*, Youichi Katou*, Dr. Hiroyuki Kon, Dr. Kazuo Watanabe and Takenori

More information

Physiological Basis of the BOLD Signal. Kerstin Preuschoff Social and Neural systems Lab University of Zurich

Physiological Basis of the BOLD Signal. Kerstin Preuschoff Social and Neural systems Lab University of Zurich Physiological Basis of the BOLD Signal Kerstin Preuschoff Social and Neural systems Lab University of Zurich Source: Arthurs & Boniface, 2002 From Stimulus to Bold Overview Physics of BOLD signal - Magnetic

More information

Development of High-Speed High-Precision Cooling Plate

Development of High-Speed High-Precision Cooling Plate Hironori Akiba Satoshi Fukuhara Ken-ichi Bandou Hidetoshi Fukuda As the thinning of semiconductor device progresses more remarkably than before, uniformity within silicon wafer comes to be strongly required

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

ELECTRODIAGNOSTIC MEDICINE

ELECTRODIAGNOSTIC MEDICINE ELECTRODIAGNOSTIC MEDICINE OBJECTIVES BEGINNER Patient Care Perform a comprehensive electrodiagnostic evaluation of each patient and to provide a concise diagnosis and plan for further treatment Describe

More information

Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser

Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser V.Nassisi #, G.Caretto #, A. Lorusso #, D.Manno %, L.Famà %, G.Buccolieri %, A.Buccolieri %, U.Mastromatteo* # Laboratory of Applied

More information

Why silicon MEMS? MEMS@KTH. Silicon is a strong material... Photolithography. Micromachining. Dicing and packaging

Why silicon MEMS? MEMS@KTH. Silicon is a strong material... Photolithography. Micromachining. Dicing and packaging Why silicon MEMS? MEMS@KTH Small Identical Large volumes (low cost per unit) School of Electrical Engineering Royal Institute of Technology Silicon is a strong material... Photolithography 10 µm thick

More information

State of the art in reactive magnetron sputtering

State of the art in reactive magnetron sputtering State of the art in reactive magnetron sputtering T. Nyberg, O. Kappertz, T. Kubart and S. Berg Solid State Electronics, The Ångström Laboratory, Uppsala University, Box 534, S-751 21 Uppsala, Sweden D.

More information

INTEGRATED CIRCUITS DATA SHEET. TDA7000 FM radio circuit. Product specification File under Integrated Circuits, IC01

INTEGRATED CIRCUITS DATA SHEET. TDA7000 FM radio circuit. Product specification File under Integrated Circuits, IC01 INTEGRATED CIRCUITS DATA SHEET File under Integrated Circuits, IC01 May 1992 GENERAL DESCRIPTION The is a monolithic integrated circuit for mono FM portable radios, where a minimum on peripheral components

More information

Operator Quick Guide EC SENSOR

Operator Quick Guide EC SENSOR Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra

More information

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT 4 Understand single-phase alternating current (ac) theory Single phase AC

More information

Lead & Magnet Wire Connection Methods Using the Tin Fusing Method Joyal A Division of AWE, Inc.

Lead & Magnet Wire Connection Methods Using the Tin Fusing Method Joyal A Division of AWE, Inc. Lead & Magnet Wire Connection Methods Using the Tin Fusing Method Joyal A Division of AWE, Inc. Abstract The technology for connecting lead and magnet wires for electric motors and electro mechanical devices

More information

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

Measurement of Charge-to-Mass (e/m) Ratio for the Electron Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic

More information

Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise

Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise TDK EMC Technology Product Section Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise TDK Shonai Corporation Satoru Saito Reduce Emission Noise from Cables Even if an electronic

More information

Mylar polyester film. Electrical Properties. Product Information. Dielectric Strength. Electrode Size. Film Thickness

Mylar polyester film. Electrical Properties. Product Information. Dielectric Strength. Electrode Size. Film Thickness Product Information Mylar polyester film Electrical Properties Mylar offers unique design capabilities to the electrical industry due to the excellent balance of its electrical properties with its chemical,

More information

OPTIMIZING OF THERMAL EVAPORATION PROCESS COMPARED TO MAGNETRON SPUTTERING FOR FABRICATION OF TITANIA QUANTUM DOTS

OPTIMIZING OF THERMAL EVAPORATION PROCESS COMPARED TO MAGNETRON SPUTTERING FOR FABRICATION OF TITANIA QUANTUM DOTS OPTIMIZING OF THERMAL EVAPORATION PROCESS COMPARED TO MAGNETRON SPUTTERING FOR FABRICATION OF TITANIA QUANTUM DOTS Vojtěch SVATOŠ 1, Jana DRBOHLAVOVÁ 1, Marian MÁRIK 1, Jan PEKÁREK 1, Jana CHOMOCKÁ 1,

More information

Rotary actuators Rotary wing drives Series RAK. Brochure

Rotary actuators Rotary wing drives Series RAK. Brochure Brochure 2 Rotary wing drives, angle of rotation: 30-270 Rotary wing drive, double-acting axis geometry: single 4 Rotary wing drives, angle of rotation: 90-270 Rotary wing drive, double-acting axis geometry:

More information

Comparison of results and test methods using the Micro IRHD and the Micro Shore rubber hardness measurement instruments

Comparison of results and test methods using the Micro IRHD and the Micro Shore rubber hardness measurement instruments Comparison of results and test methods using the Micro and the Micro Shore rubber hardness measurement instruments R. Morgans 1 BSc, S. Lackovic 2 BSc, PhD, P. Cobbold 2 1. School of Engineering, University

More information

Tungsten Tip Preparation

Tungsten Tip Preparation 1 Tungsten Tip Preparation Andra Mandru, August 2011 General aspects and observations: When performing STM measurements, the stability of the tip plays a crucial role. Therefore, a sharp tip which comes

More information

Fabrication and Characterization of Schottky Diode

Fabrication and Characterization of Schottky Diode Fabrication and Characterization of Schottky Diode Arnab Dhabal Acknowledgements I would like to express my greatest gratitude to the people who have helped and supported me in this project. I wish to

More information

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control

More information

Improved Contact Formation for Large Area Solar Cells Using the Alternative Seed Layer (ASL) Process

Improved Contact Formation for Large Area Solar Cells Using the Alternative Seed Layer (ASL) Process Improved Contact Formation for Large Area Solar Cells Using the Alternative Seed Layer (ASL) Process Lynne Michaelson, Krystal Munoz, Jonathan C. Wang, Y.A. Xi*, Tom Tyson, Anthony Gallegos Technic Inc.,

More information

ElectroMagnetic Induction. AP Physics B

ElectroMagnetic Induction. AP Physics B ElectroMagnetic Induction AP Physics B What is E/M Induction? Electromagnetic Induction is the process of using magnetic fields to produce voltage, and in a complete circuit, a current. Michael Faraday

More information

Frequency Response of Filters

Frequency Response of Filters School of Engineering Department of Electrical and Computer Engineering 332:224 Principles of Electrical Engineering II Laboratory Experiment 2 Frequency Response of Filters 1 Introduction Objectives To

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager Technique and Safety of performing EEG/fMRI measurements by Pierluigi Castellone, Electronics Engineer Brain Products General Manager Contents of the presentation Why recording simultaneous EEG and fmri?

More information

A Practical Guide to Free Energy Devices

A Practical Guide to Free Energy Devices A Practical Guide to Free Energy Devices Electrolysis Patents No. 16: Last updated: 30th September 2006 Author: Patrick J. Kelly The major difficulty in using Stan s low-current Water Fuel Cell (recently

More information

Conductivity of silicon can be changed several orders of magnitude by introducing impurity atoms in silicon crystal lattice.

Conductivity of silicon can be changed several orders of magnitude by introducing impurity atoms in silicon crystal lattice. CMOS Processing Technology Silicon: a semiconductor with resistance between that of conductor and an insulator. Conductivity of silicon can be changed several orders of magnitude by introducing impurity

More information

RADIANT PLASMA 4700 Plasma Spark Generator

RADIANT PLASMA 4700 Plasma Spark Generator RADIANT PLASMA 4700 Plasma Spark Generator Installation Guide / User Manual A S P A R K O F F R E S H A I R Aquapulser.com Contents 1 Introduction 2 1.1 About the Product....................................

More information

Differentiation = Making specialized cells

Differentiation = Making specialized cells Differentiation = Making specialized cells What is a stem cell? Screws: JM www.logodesignweb.com/stockphoto Differentiation = Making specialized cells What is a stem cell? What the photo shows A lump of

More information

Torque motors. direct drive technology

Torque motors. direct drive technology Torque motors direct drive technology Why Direct Drive Motors? Fast and effective Direct-drive technology in mechanical engineering is defined as the use of actuators which transfer their power directly

More information

Through-mask Electro-etching for Fabrication of Metal Bipolar Plate Gas Flow Field Channels

Through-mask Electro-etching for Fabrication of Metal Bipolar Plate Gas Flow Field Channels 991 Downloaded 23 Dec 21 to 24.16.113.125. Redistribution subject to ECS license or copyright; see http://www.ecsdl.org/terms_use.jsp ECS Transactions, 33 (1) 991-16 (21) 1.1149/1.3484593 The Electrochemical

More information

Lecture 4: Jan 12, 2005

Lecture 4: Jan 12, 2005 EE516 Computer Speech Processing Winter 2005 Lecture 4: Jan 12, 2005 Lecturer: Prof: J. Bilmes University of Washington Dept. of Electrical Engineering Scribe: Scott Philips

More information

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force

More information

understanding medium frequency induction melting furnace and its components

understanding medium frequency induction melting furnace and its components understanding medium frequency induction melting furnace and its components 9 Spruce Street, Jersey City, NJ 07306 USA sales@electroheatinduction.com INDUCTION MELTING FURNACE AND ITS COMPONENTS The purpose

More information