Overview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC

Size: px
Start display at page:

Download "Overview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC"

Transcription

1 Overview of Medical Device Controls in the US By Nandini Murthy, MS, RAC 18

2 controls are a regulatory requirement for medical devices. In the US, compliance with the design controls section of 21 Code of Federal Regulations (CFR) Part of the Quality System Regulation (QSR), 1 which applies to the design and development of new products, and changes to existing devices, is mandatory for both investigational and commercial medical devices. Because the regulation must apply to so many different types of devices, it does not prescribe in detail how a manufacturer must develop a specific device. Rather, the regulation is somewhat flexible; instead of specifics, it provides the framework that all manufacturers must follow by requiring that they develop and follow procedures and fill in the details that are appropriate to a given device. For example, the design control regulation requires documentation of device specifications. Different companies utilize terms such as Functional Requirement Specifications (FRS), Product Requirements Definition (PRD), etc., in their procedures. The Need for Control Requirements for Medical Devices Developing a new medical device from concept to market introduction is a complex process, particularly if the device and associated manufacturing processes use software, which presents the possibility of subtle errors. Devices typically have multiple components, some developed in-house, some procured from vendors. Then, the manufacturer assembles all these various parts into a finished medical device. Engineering teams developing such devices include members with specific functional expertise such as materials, software development and electronics, all of whom have to collaborate on designing the device. In addition to the complexity of the design process, device testing has to be adequate to meet regulatory requirements (US Food and Drug Administration (FDA), applicable international or domestic standards such as ISO, IEC and AAMI). This means regulatory strategy has to be developed early in design and development. Because devices are designed with intent, many can be tested exhaustively on the bench in simulated worst-case clinical conditions. Depending upon their technological risk and intended use, some devices may not need animal or clinical testing. Without thorough planning, communication and program control, it is difficult to develop a device that is error-free and or that includes all device features. Redesigning or retesting a device due to inadequate planning is a costly task and, in today s economic climate, funding such an effort could be a challenge. Controls provide a logical framework to approach device design and development, which are detailed in the article below. Figure 1 shows an overview of the design control process. History File A History File (DHF) should be initiated for each product developed. It contains and/ or references all of the documents and records noted below, necessary to establish compliance with the design control requirements. The following documents are typically maintained in the DHF and, are discussed in further detail below: design and development plans design inputs design review meeting minutes design outputs verification and validation test plans design verification protocols and results Most of these are living documents that evolve with device development, i.e., they undergo constant rethinking and modification as the project progresses. From Concept to and Development Plan At the outset of a new project, a product concept is outlined in a set of marketing requirements. The marketing requirements make a business case for proceeding with device development and outline the wish list for device features and functionality. It is a high-level description of the device, addressing the following basic questions: 2 What is the real need for the new device? Is there a market opportunity for the device? Where will the new device be used? In the hospital, the doctor s office, in consumers homes? Who will use the new device? Nurses, physicians, patients? How will the new device be used? Is it an accessory to another device? How long will the new device be used? Once the company decides to move forward with the project, a project plan is drawn up to guide and control the design and development of the device. Approval of the plan signifies the start of design controls implementation and a DHF is initiated at this time. The project plan (sometimes also called the design and development plan) is typically a project timeline or Gantt chart that is broad and complete, capturing tasks from conception through project completion, along with responsibilities and authority for the identified design and development activities. Risk management activities may be part of the plan. The project plan addresses team interfaces. For example, if contractors are responsible for developing software, developing manufacturing molds or conducting electrical safety testing, such activities are referenced and included in the plan. The interfaces also include interactions with Regulatory Focus 19

3 Figure 1. Control Overview & Development Planning Input Output Transfer Changes and Development Plan Develop Product Requirements, Specifications Implementation: Drawings, Mock Ups, Prototypes Tests (production units) Manufacturing Change Control System Safety Risk Analysis Part Specifications Reports Revalidation Testing Risk Assessment (FMEA) Clinical Trials Test Planning Unit Test & Verification Tests Significant Changes - Iterative Requirements & Risk Analysis Implementation Release manufacturing, research and development, marketing, quality assurance, regulatory or other internal functions. Some of these interfaces and responsibilities may be revised as the project advances. Most project plans identify a project manager who is tasked with coordinating this multi-functional input and ensuring compliance with design control requirements. Input input, the next step of the design and development process, involves translating user and marketing needs into a set of engineering requirements that can be tested, i.e., turning qualitative requirements into quantitative requirements. input also addresses requirements of regulations and standards, device packaging, shipping, installation and field service. This step focuses on what the design is intended to do while carefully avoiding specific solutions, especially at the start of the development phase. Besides engineering criteria, physician or nurse feedback may be obtained to capture routine clinical practice or procedures and experience with similar devices to better understand the clinical environment in which the device will be used and, if the technology is disruptive, to identify potential challenges to clinical implementation. A safety risk analysis related to intended use in the clinical environment may also be conducted. A design review is a formal, comprehensive, systematic, multi-function team evaluation and documentation of the design inputs or outputs to evaluate the adequacy of the design requirements, design capability and manufacturability, 20

4 and to identify problem areas. A design review may be used for design selection, for identification and resolution of problems and for design transfer to manufacturing. reviews are held at major decision points in product development as noted in the and Development Plan, and may be supplemented by routine team meetings. Output output is the translation of design input requirements into device specifications and production processes. It is important to realize that a design input requirement may have more than one design output. For example, a design input that states the device must alert a user via an audible alarm within one meter of the device may have specifications on both the decibel level and nature of the alarm (e.g., intermittent or continuous beep). outputs are documented and expressed in terms that can be verified and validated against the design input requirements. outputs evolve with the project. After preliminary design specifications are drafted, physical design of the device begins. Parts may be ordered and tested for acceptance before being approved as part of the design. Firms have to ensure that once parts are approved, purchasing control requirements of the QSR, such as supplier qualification and auditing, are complied with. Then prototypes of the device are assembled and preliminary testing is conducted to confirm design selection. If necessary, parts may be changed at this phase. All of this engineering work is documented. If there are significant challenges in meeting some design inputs, or revisions must be made to ensure proper functioning, the design input documents are updated using a formal document change procedure that captures the change history. In addition to document control, firms need to have supplier control and materials control procedures in place to support the design and development process. During this phase, a design risk assessment is conducted. Many methodologies are available. One is failure mode and effects analysis (FMEA). The FMEA methodology is a systematic process to identify potential material and component failures that might cause the device to fail, to identify and eliminate potential causes for these failures and to identify and reduce the failure s impacts. Verification and Test Plan A verification and validation (V&V) test plan is developed to ensure the device design meets all of the parameters in the design input phase. The design characteristics that mitigate risk as noted in the risk assessment are all tested in the V&V phase. There must be traceability from the V&V test plan back to the design input and design output/risk assessment. It is helpful to have a traceability matrix, which is a table where the direct, sequential relationship between the design input, design output/risk assessment and V&V test plan are noted. This traceability matrix is typically very detailed. Verification and validation testing must be prospectively planned and documented before actual testing commences and are typically captured in test protocols. Note that the actual V&V testing is also of the design output phase and more than one design review may be held during this phase. Verification verification testing confirms that the design output meets the design input requirements. verification involves reviewing, inspecting, testing and auditing the components, the final device, processes and documents to ensure they conform to design requirements. Testing can be performed on prototype devices. As product development proceeds, verification may be repeated as warranted when significant changes are made to the design. The verification effort takes into account the extremes of the design specifications (for example, the range of dimensional specifications of a coronary stent to 22

5 account for differences in human anatomy and the degree of arterial stenosis). When design verification testing shows that certain design requirements cannot be met, the requirements may be changed if marketing needs and the device s intended purpose are still fulfilled. A design review may be conducted to ensure that the updated design complies with the product requirements and raises no new questions of risk. validation follows successful design verification; it is intended to ensure that the products from initial production lots (or their equivalents), when used under actual or simulated conditions, function as intended. 3 The use of devices from production lots is meant to encompass expected variation in components, materials, manufacturing processes and the use environment. Methods of design validation include in vivo studies, historical database searches; literature searches; risk analysis, where appropriate; and review of labels, labeling, packaging and other historical product information. For certain products, a human clinical trial is required as part of the design validation activity to confirm that the product functions as intended in the user environment. The product tested is representative of the final, commercial product. Based on clinical trial feedback, design changes may be necessary. In general, if a product fails to meet the design input requirements during the design validation effort, it will be necessary to revise manufacturing processes (which may require re-validation) or the design (which may require reverification and reconstruction of design outputs). Transfer The purpose of design transfer is to ensure product designs are correctly transferred into manufacturing specifications. As the design output is being created, detailed part specifications, assembly instructions and quality control test instructions are being developed. They are all part of what is termed a Device Master Record (DMR), a set of documents that are used directly for production. Changes changes may be initiated because of factors such as customer feedback (suggestions for improvement, complaints), corrective or preventive actions, market changes and technological advances. changes are documented, tested, assessed for risk, reviewed and approved before implementation. The review of such changes includes the evaluation of the effect of the changes on products already in commercial or clinical use. The evaluation and documentation are to be in direct proportion to the nature and significance of the change, and may include risk analysis. The design changes are reviewed to determine whether prior design verification or validation results are impacted. If the design change is likely to affect the performance of the device under actual use conditions, new validation work, including clinical trials, may be necessary. Conclusion The design control regulations provide a comprehensive and systematic framework for device development. Regular design reviews are useful for periodically confirming that device development continues to meet clinical and market needs. Confirming that design outputs align with design inputs reduces the risk of omitting an important feature. The design control process is iterative. When new information comes up as the project progresses, the regulations require that prior documents be reviewed and revised as necessary. As noted earlier, redesigning the device after considerable time and effort has been expended is a costly exercise. control regulations are a regulatory requirement, but when used properly serve as good business practice. References CFR, Part 820, Quality System Regulation 2. Medical Device Quality Systems Manual: A Small Entity Compliance Guide 3. Control Guidance for Medical Device Manufacturers, 03/11/1997 Author Nandini Murthy, MS, RAC, is a regulatory consultant advising companies on regulatory strategy, regulatory planning, regulatory due diligence, FDA submissions, clinical study design and quality system development. Murthy has authored several PMA, modular PMA, HDE, IDE and 510(k) applications for cardiology, neurology, radiology, imaging, diagnostic device and IVD companies. She has also designed and executed clinical trials including the REMATCH, SPECTACL and BrainGate trials and has established quality systems to ISO and FDA Quality System requirements. Previously, Murthy served on the management team of several companies. Regulatory Focus 23

How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters

How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters environmental failure analysis & prevention health technology development How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters Kevin L. Ong, Ph.D., P.E. Managing

More information

Quality System: Design Control Procedure - Appendix

Quality System: Design Control Procedure - Appendix Quality System: Design Control Procedure - Appendix Page 1 of 10 Quality System: Design Control Procedure - Appendix CORP Medical Products Various details have been removed, indicated by [ ] 1. Overview

More information

IVD Regulation Overview. Requirements to Assure Quality & Effectiveness

IVD Regulation Overview. Requirements to Assure Quality & Effectiveness IVD Regulation Overview Requirements to Assure Quality & Effectiveness CLIAC Jan. 2002 Statutory and Regulatory Requirements Statute: Food, Drug, and Cosmetic Act Food and Drugs Act of 1906 Food and Drug

More information

QUALITY MANAGEMENT SYSTEM REQUIREMENTS General Requirements. Documentation Requirements. General. Quality Manual. Control of Documents

QUALITY MANAGEMENT SYSTEM REQUIREMENTS General Requirements. Documentation Requirements. General. Quality Manual. Control of Documents Chapter j 38 Self Assessment 729 QUALITY MANAGEMENT SYSTEM REQUIREMENTS General Requirements 1. Establishing and implementing a documented quality management system 2. Implementing a documented quality

More information

Quality Management System Manual

Quality Management System Manual Effective Date: 03/08/2011 Page: 1 of 17 Quality Management System Manual Thomas C. West Eric Weagle Stephen Oliver President ISO Management General Manager Representative Effective Date: 03/08/2011 Page:

More information

Mobile Medical Application Development: FDA Regulation

Mobile Medical Application Development: FDA Regulation Mobile Medical Application Development: FDA Regulation Mobile Medical Applications: Current Environment Currently over 100,000 mobile health related applications are available for download; there is an

More information

Medical Device Training Program 2015

Medical Device Training Program 2015 Medical Device Training Introduction Supplementary training and education is often overlooked by medical device professionals until it is triggered by an upcoming FDA or Notified Body and/or ISO 13485

More information

Specialties Manufacturing. Talladega Castings & Machine Co., Inc. ISO 9001:2008. Quality Manual

Specialties Manufacturing. Talladega Castings & Machine Co., Inc. ISO 9001:2008. Quality Manual Specialties Manufacturing Talladega Castings & Machine Co., Inc. ISO 9001:2008 This document is the property of TMS and may not be reproduced, wholly, or in part, without the express consent of TMS. Rev.

More information

Camar Aircraft Products Co. QUALITY MANUAL Revision D

Camar Aircraft Products Co. QUALITY MANUAL Revision D QUALITY MANUAL Revision D Gujll'y Manual Introduction The purpose of this manual is to describe the Quality Assurance Program implemented by Camar Aircraft Products Co. (hereafter referred to as C.A.P.C.)

More information

CORPORATE QUALITY MANUAL

CORPORATE QUALITY MANUAL Corporate Quality Manual Preface The following Corporate Quality Manual is written within the framework of ISO 9001:2008 Quality System by the employees of CyberOptics. CyberOptics recognizes the importance

More information

ISO 9001:2008 Audit Checklist

ISO 9001:2008 Audit Checklist g GE Power & Water ISO 9001:2008 Audit Checklist Organization Auditor Date Page 1 Std. 4.1 General s a. Are processes identified b. Sequence & interaction of processes determined? c. Criteria for operation

More information

ISO 9001:2000 AUDIT CHECKLIST

ISO 9001:2000 AUDIT CHECKLIST ISO 9001:2000 AUDIT CHECKLIST No. Question Proc. Ref. Comments 4 Quality Management System 4.1 General Requirements 1 Has the organization established, documented, implemented and maintained a quality

More information

How To Write Software

How To Write Software 1 Medical Device Software - Software Life Cycle Processes IEC 62304 2 Credits John F. Murray Software Compliance Expert U.S. Food and Drug Administration Marcie R. Williams Medical Device Fellow Ph.D.

More information

How To Validate Software

How To Validate Software General Principles of Software Validation; Final Guidance for Industry and FDA Staff Document issued on: January 11, 2002 This document supersedes the draft document, "General Principles of Software Validation,

More information

PHARMACEUTICAL QUALITY SYSTEM Q10

PHARMACEUTICAL QUALITY SYSTEM Q10 INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE PHARMACEUTICAL QUALITY SYSTEM Q10 Current Step

More information

Guidance for Industry. Q10 Pharmaceutical Quality System

Guidance for Industry. Q10 Pharmaceutical Quality System Guidance for Industry Q10 Pharmaceutical Quality System U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Biologics Evaluation

More information

What methods are used to conduct testing?

What methods are used to conduct testing? What is testing? Testing is the practice of making objective judgments regarding the extent to which the system (device) meets, exceeds or fails to meet stated objectives What the purpose of testing? There

More information

ISO 9001 Quality Systems Manual

ISO 9001 Quality Systems Manual ISO 9001 Quality Systems Manual Revision: D Issue Date: March 10, 2004 Introduction Micro Memory Bank, Inc. developed and implemented a Quality Management System in order to document the company s best

More information

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers.

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers. Supplier Quality Standard 1.0 Purpose The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers. These expectations

More information

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 globalsales@maetrics.com

More information

CENTRIS CONSULTING. Quality Control Manual

CENTRIS CONSULTING. Quality Control Manual CENTRIS CONSULTING Quality Control Manual ISO 9001:2008 Introduction Centris Consulting developed and implemented a Quality Management System in order to document the company s best business practices,

More information

Correspondence between ISO 13485:2003 and the US Quality System Regulation

Correspondence between ISO 13485:2003 and the US Quality System Regulation Correspondence between ISO 13485:2003 and the US Quality System Regulation Correspondence between ISO 13485:2003 and the US Quality System Regulation 1 Scope 1.1 General This International Standard specifies

More information

Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables

Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables (1987) Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables 1. General 1.1 Introduction 1.1.1 The present guidelines are to serve as a supplement to the IACS

More information

Quality Management Systems Manual

Quality Management Systems Manual Washington Division Quality Management Systems Manual ISO 13485:2003 DOCUMENT NUMBER 01-QM-MED REVISION LEVEL 01 PAGE 1 of 40 This manual describes the quality management systems structure at GM Nameplate

More information

Section 5: Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP)

Section 5: Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP) Section 5: Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP) 1 Quality Systems Regulation (QSR) 2 Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP) Sets of checks

More information

Micro Plastics, Inc. Quality Manual

Micro Plastics, Inc. Quality Manual ISO 9001:2008 11 Industry Lane Flippin, Arkansas 72634 QM-001-2008-F Page 2 of 39 Introduction Micro Plastics, Inc. developed and implemented a Quality Management System in order to document the company

More information

MINIMUM AUTOMOTIVE QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR SUB-TIER SUPPLIERS

MINIMUM AUTOMOTIVE QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR SUB-TIER SUPPLIERS MINIMUM AUTOMOTIVE QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR SUB-TIER SUPPLIERS CONTENTS 1. CONTROL PLANS 2. PROCESS APPROACH 3. PERFORMANCE 4. INTERNAL AUDITING 5. CONTROL OF NON-CONFORMING PRODUCT 6.

More information

Quality Manual ISO9001:2008

Quality Manual ISO9001:2008 Quality Manual ISO9001:2008 Connor Winfield Corp. Date: 4/26/10 Revision: 3 Quality Manual Quality Manual Connor Winfield GENERAL Section 0.1 Section Rev.: 2 Rev. Date: 4/26/10 Section Page 1 Index and

More information

Design Controls: Are They Worth the Effort?

Design Controls: Are They Worth the Effort? Design Controls: Are They Worth the Effort? Compliance-Alliance conducted a survey to measure the effects of FDA s design control regulation on the industry. Most respondents believe the controls have

More information

Risk Assessment for Medical Devices. Linda Braddon, Ph.D. Bring your medical device to market faster 1

Risk Assessment for Medical Devices. Linda Braddon, Ph.D. Bring your medical device to market faster 1 Risk Assessment for Medical Devices Linda Braddon, Ph.D. Bring your medical device to market faster 1 My Perspective Work with start up medical device companies Goal: Making great ideas into profitable

More information

ISO 9001:2000 Gap Analysis Checklist

ISO 9001:2000 Gap Analysis Checklist ISO 9001:2000 Gap Analysis Checklist Type: Assessor: ISO 9001 REQUIREMENTS STATUS ACTION/COMMENTS 4 Quality Management System 4.1 General Requirements Processes needed for the quality management system

More information

FDA Software Validation-Answers to the Top Five Software Validation Questions

FDA Software Validation-Answers to the Top Five Software Validation Questions Whitepaper FDA Software Validation-Answers to the Top Five Software Validation Questions Author: Penny Goss, Penny Goss Technical Solutions The FDA (Food and Drug Administration) and IEC (International

More information

FINAL DOCUMENT. Guidelines for Regulatory Auditing of Quality Management Systems of Medical Device Manufacturers Part 1: General Requirements

FINAL DOCUMENT. Guidelines for Regulatory Auditing of Quality Management Systems of Medical Device Manufacturers Part 1: General Requirements GHTF/SG4/N28R4:2008 FINAL DOCUMENT Title: Guidelines for Regulatory Auditing of Quality Management Systems of Medical Device Manufacturers Authoring Group: GHTF Study Group 4 Endorsed by: The Global Harmonization

More information

AP1000 European 18. Human Factors Engineering Design Control Document

AP1000 European 18. Human Factors Engineering Design Control Document 18.2 Human Factors Engineering Program Management The purpose of this section is to describe the goals of the AP1000 human factors engineering program, the technical program to accomplish these goals,

More information

QUALITY MANUAL ISO 9001. Quality Management System

QUALITY MANUAL ISO 9001. Quality Management System Page 1 of 20 QUALITY MANUAL ISO 9001 Quality Management System Printed copies are not controlled unless marked "CONTROLLED". Upon receipt of this document, discard all previous copies. Page 2 of 20 Approval

More information

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Quality Management System MANUAL SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Doc. No. G5500 Rev. 9.2 Status : APPROVED Effective: 9/25/2015 Page 2 of 23 Quality Manual Table of Contents

More information

FINE LOGISTICS. Quality Manual. Document No.: 20008. Revision: A

FINE LOGISTICS. Quality Manual. Document No.: 20008. Revision: A FINE LOGISTICS Quality Manual Document No.: 20008 Revision: A 20008 Rev. A FINE LOGISTICS, Quality Manual Page 1 of 24 Quality Manual: Table of contents Number Section Page 1. GENERAL 3 1.1 Index and revision

More information

Combination Products. Presented by: Karen S. Ginsbury For: IFF March 2014. PCI Pharma

Combination Products. Presented by: Karen S. Ginsbury For: IFF March 2014. PCI Pharma Combination Products Presented by: Karen S. Ginsbury For: IFF March 2014 Types of products Biological and medical device (freeze dried + syringe dual volume) Medical device and plasma devised product (syringe)

More information

UNCONTROLLED COPY FOR REFERENCE ONLY

UNCONTROLLED COPY FOR REFERENCE ONLY CLOVER MACHINE AND MFG. 800 MATHEW ST. #101 SANTA CLARA, CA 95050 727-3380 727-7015 fax REVISION: DATE: PAGE 1 OF 45 QUALITY POLICY MANUAL DISTRIBUTION LIST: President Purchasing Manager Vice President

More information

RTP s NUCLEAR QUALITY ASSURANCE PROGRAM

RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP operates under one quality program, whether you purchase products that are commercial grade, nuclear safety-related or industrial safety compliant (IEC 61508).

More information

Row Manufacturing Inc. Quality Manual ISO 9001:2008

Row Manufacturing Inc. Quality Manual ISO 9001:2008 Row Manufacturing Inc. Quality Manual ISO 9001:2008 Row Manufacturing 210 Durham Drive Athens, Alabama 35611 Phone:256.232.4151 Fax:256.232.4133 Page 2 of 33 This Page intentionally left Blank Page 3 of

More information

SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE

SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE Company Name Street Address City, State, Zip code Phone Number Fax Company Website Email Address ORGANIZATION NAME PHONE NUMBER EMAIL ADDRESS President/CEO General Manager Engineering Manager Production

More information

Quality Assurance QUALITY ASSURANCE PLAN

Quality Assurance QUALITY ASSURANCE PLAN Revision 2 Page 1 of 40 QUALITY ASSURANCE PLAN PLAN APPROVALS: Jeff Shouse Signature on File DIRECTOR OF QUALITY ASSURANCE DIRECTOR OF QUALITY ASSURANCE (signature) DATE Rodney Baltzer Signature on File

More information

Off-the-Shelf Software: A Broader Picture By Bryan Chojnowski, Reglera Director of Quality

Off-the-Shelf Software: A Broader Picture By Bryan Chojnowski, Reglera Director of Quality Off-the-Shelf Software: A Broader Picture By Bryan Chojnowski, Reglera Director of Quality In the past decade, there has been a sea change in the business software domain. Many companies are no longer

More information

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Document issued on: March 27, 2006 For questions regarding this document, contact Everette Beers, Ph.D. at 240-276-4200

More information

ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan)

ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan) ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan) Part 1: Project-Specific Quality Plan Part 2: Company Quality Manual Part 3: Submittal Forms Part 4:

More information

The Prophotonix (UK) Ltd Quality manual

The Prophotonix (UK) Ltd Quality manual The Prophotonix (UK) Ltd Quality manual Date: March 2014 Revision: D Sparrow lane, Hatfield Broad Oak, Herts, UK, CM22 7BA Tel: +44 (0)1279 717170 Fax: +44 (0)1279 717171 e-mail: info@prophotonix.com Page

More information

QUALITY ASSURANCE MANUAL JPM OF MISSISSIPPI, INC.

QUALITY ASSURANCE MANUAL JPM OF MISSISSIPPI, INC. QUALITY ASSURANCE MANUAL JPM OF MISSISSIPPI, INC. Hattiesburg, MS Revision E 01/19/11 Revised to ISO 9001:2008 on July 9, 2009 JPM OF MISSISSIPPI, INC. MANAGEMENT QUALITY POLICY It is the goal of JPM of

More information

ORACLE CONSULTING GROUP

ORACLE CONSULTING GROUP ORACLE CONSULTING GROUP 9 Golder Ranch Rd., Ste. 1 Tucson, Arizona 9 Web Site: E-mail: 20-2-0 20-2-0 (FAX) CONSULTING MEMORANDUM QUALITY SYSTEM INSPECTION TECHNIQUE

More information

G & R Labs. Quality Assurance Manual. G&R Labs. 2996 Scott Boulevard Santa Clara CA 95054. Light Meters and Calibration

G & R Labs. Quality Assurance Manual. G&R Labs. 2996 Scott Boulevard Santa Clara CA 95054. Light Meters and Calibration G & R Labs Light Meters and Calibration Quality Assurance Manual G&R Labs 2996 Scott Boulevard Santa Clara CA 95054 QM01 QUALITY MANUAL ISO9001:2008 Manual Contents QM 01 Manual Contents 2 QM 02 Copy Holder

More information

Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca

Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca Software Regulated as a Medical Device Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca Therapeutic Products Directorate Director General S. Sharma

More information

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Quality Management System MANUAL SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Doc. No. G5500 Rev. 9.1 Status : APPROVED Effective: 12/11/2014 Page 2 of 23 Quality Manual Table of Contents

More information

Introduction to AS 9100

Introduction to AS 9100 Introduction to AS 9100 Rev. C 1 Questions we will cover today: What is AS9100? What does a company need to do to Register to AS9100 Rev C? What are the requirements? Section 4 General Requirements Section

More information

ISO 13485:201x What is in the new standard?

ISO 13485:201x What is in the new standard? ISO 13485:201x What is in the new standard? Eric Finegan, Quality Mgr, BTE Technologies, Inc. 2015-09-10 1 Presentation Slides This slide deck is the presentation performed on 2015-09-10. A more detailed

More information

QUALITY POLICY MANUAL Document: 01-090000 Revision: E Effective Date: January 15, 2010

QUALITY POLICY MANUAL Document: 01-090000 Revision: E Effective Date: January 15, 2010 Section i INTRODUCTION QUALITY POLICY STATEMENT: The purpose of our business is to provide continuity of electrical power to keep businesses in business. We do this by helping clients solve their power

More information

Usability of Medical Applications Ved Line Kagenow Svenstrup, lks@delta.dk

Usability of Medical Applications Ved Line Kagenow Svenstrup, lks@delta.dk Usability of Medical Applications Ved Line Kagenow Svenstrup, lks@delta.dk What is usability? The user, rather than the system, at the center of the process. Risk of operating errors that can cause injury

More information

OPERATIONAL STANDARD

OPERATIONAL STANDARD 1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.

More information

Your Software Quality is Our Business. INDEPENDENT VERIFICATION AND VALIDATION (IV&V) WHITE PAPER Prepared by Adnet, Inc.

Your Software Quality is Our Business. INDEPENDENT VERIFICATION AND VALIDATION (IV&V) WHITE PAPER Prepared by Adnet, Inc. INDEPENDENT VERIFICATION AND VALIDATION (IV&V) WHITE PAPER Prepared by Adnet, Inc. February 2013 1 Executive Summary Adnet is pleased to provide this white paper, describing our approach to performing

More information

ISO 9001 (2000) QUALITY MANAGEMENT SYSTEM ASSESSMENT REPORT SUPPLIER/ SUBCONTRACTOR

ISO 9001 (2000) QUALITY MANAGEMENT SYSTEM ASSESSMENT REPORT SUPPLIER/ SUBCONTRACTOR Page 1 of 20 ISO 9001 (2000) QUALITY MANAGEMENT SYSTEM ASSESSMENT REPORT SUPPLIER/ SUBCONTRACTOR SUPPLIER/ SUBCONTRACTOR NAME: ADDRESS: CITY AND STATE: ZIP CODE: SUPPLIER/MANUFACTURER NO PHONE: DIVISION:

More information

Summary of GE Healthcare's Quality Management System (QMS) Covering BioProcess chromatography media, equipment, software, and spare parts

Summary of GE Healthcare's Quality Management System (QMS) Covering BioProcess chromatography media, equipment, software, and spare parts GE Healthcare Summary of GE Healthcare's Quality Management System (QMS) Covering BioProcess chromatography media, equipment, software, and spare parts Table of content General requirements 3 Documentation

More information

How To Know If A Mobile App Is A Medical Device

How To Know If A Mobile App Is A Medical Device The Regulation of Medical Device Apps Prepared for West of England Academic Health Science Network and University of Bristol June 2014 June 2014 1 Table of Contents 1 Purpose...3 2 Scope...3 3 The Regulation

More information

Comparison between FDA QSR and ISO 13485

Comparison between FDA QSR and ISO 13485 Comparison between FDA QSR and ISO 13485 Most countries in the world including the Europe, for the conformity assessment of medical devices to be used by their countrymen, assess not only whether the product

More information

Quality Assurance Manual

Quality Assurance Manual Quality Assurance Manual Revised April 30, 2014 2004, 2005, 2006, 2007, 2008, 2009, 2010, 2011, 2012, 2013, 2014 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Lucy Malby Business Development Manager eg technology Ltd

Lucy Malby Business Development Manager eg technology Ltd Lucy Malby Business Development Manager eg technology Ltd Design review, engineering prototype and design freeze Pre- clinical unit verification, clinical unit validation Engineering Prototype What is

More information

INTRODUCTION. This book offers a systematic, ten-step approach, from the decision to validate to

INTRODUCTION. This book offers a systematic, ten-step approach, from the decision to validate to INTRODUCTION This book offers a systematic, ten-step approach, from the decision to validate to the assessment of the validation outcome, for validating configurable off-the-shelf (COTS) computer software

More information

GxP Process Management Software. White Paper: Ten Most Common Reasons for FDA 483 Observations and Warning Letter Citations

GxP Process Management Software. White Paper: Ten Most Common Reasons for FDA 483 Observations and Warning Letter Citations GxP Process Management Software : Ten Most Common Reasons for FDA 483 Observations and Warning Letter Citations Most FDA violations involve one of the following: Not having procedures in a regulated area

More information

Title: Rio Tinto management system

Title: Rio Tinto management system Standard Rio Tinto management system December 2014 Group Title: Rio Tinto management system Document No: HSEC-B-01 Standard Function: Health, Safety, Environment and Communities (HSEC) No. of pages: 23

More information

Revision Date Author Description of change. 10 07Jun13 Mark Benton Removed Admin. Manager from approval

Revision Date Author Description of change. 10 07Jun13 Mark Benton Removed Admin. Manager from approval Page 2 of 15 Document Revision History Revision Date Author Description of change 10 07Jun13 Mark Benton Removed Admin. Manager from approval 12Feb13 Mark Benton 08 01Oct12 Mark Benton 07 8/30/2012 Refer

More information

ICH guideline Q10 on pharmaceutical quality system

ICH guideline Q10 on pharmaceutical quality system September 2015 EMA/CHMP/ICH/214732/2007 Committee for Human Medicinal Products Step 5 Transmission to CHMP May 2007 Transmission to interested parties May 2007 Deadline for comments November 2007 Final

More information

QUALITY MANUAL 3 KENDRICK ROAD WAREHAM, MA 02571 508-295-9591 FAX 508-295-6752

QUALITY MANUAL 3 KENDRICK ROAD WAREHAM, MA 02571 508-295-9591 FAX 508-295-6752 QUALITY MANUAL 3 KENDRICK ROAD WAREHAM, MA 02571 508-295-9591 FAX 508-295-6752 Section 1. REVISION STATUS Revision 01 Initial Issue 02/13/95 Revision 02 ECO No.Q-0032 05/07/96 Revision 03 ECO No.97171

More information

Medical Device Software

Medical Device Software Medical Device Software Bakul Patel Senior Policy Advisor 1 Overview Medical devices and software Oversight principles and Current approach Trends, Challenges and opportunities Addressing challenges 2

More information

ISO 9000 Introduction and Support Package: Guidance on the Documentation Requirements of ISO 9001:2008

ISO 9000 Introduction and Support Package: Guidance on the Documentation Requirements of ISO 9001:2008 Document: ISO/TC 176/SC 2/N 525R2 ISO 9000 Introduction and Support Package: 1 Introduction Two of the most important objectives in the revision of the ISO 9000 series of standards have been a) to develop

More information

Corrective and Preventive Action Background & Examples Presented by:

Corrective and Preventive Action Background & Examples Presented by: Corrective and Preventive Action Background & Examples Presented by: Kimberly Lewandowski-Walker Food and Drug Administration Division of Domestic Field Investigations Office of Regulatory Affairs Overview

More information

Forth Engineering (Cumbria) Limited QUALITY MANUAL. Quality Manual Issue 4 Updated April 2012. Authorised by: Managing Director.

Forth Engineering (Cumbria) Limited QUALITY MANUAL. Quality Manual Issue 4 Updated April 2012. Authorised by: Managing Director. Quality Manual Issue 4 Forth Engineering (Cumbria) Limited QUALITY MANUAL Copy Number: 1 The information contained in this Manual is the property of Forth Engineering (Cumbria) Limited and must not be

More information

ISO 9001 : 2008 QUALITY MANAGEMENT SYSTEM AUDIT CHECK LIST INTRODUCTION

ISO 9001 : 2008 QUALITY MANAGEMENT SYSTEM AUDIT CHECK LIST INTRODUCTION INTRODUCTION What auditors should look for: the items listed in these headings that the ISO requirement is met that the requirement is met in the manner described in the organization's documentation Page

More information

LinkTech LLC 3000 Bunsen Ave. #A Ventura, CA 93003

LinkTech LLC 3000 Bunsen Ave. #A Ventura, CA 93003 D LT-422-001 Page 1 of 12 QUALITY MANUAL LinkTech LLC 3000 Bunsen Ave. #A Ventura, CA 93003 Based on ISO 9001:2008 D LT-422-001 Page 2 of 12 INTRODUCTION: This Quality Manual establishes the quality system

More information

Eagle Machining, Inc. Quality Management System

Eagle Machining, Inc. Quality Management System Eagle Machining, Inc. Quality Management System 1 of 10310 Antoine Drive Bldg D, Houston, Texas 77086 BUSINESS OPERATING MANUAL (QUALITY MANUAL) Revision Date: 08/01/2014 Approved By: Joseph Vu Date: 08/01/2014

More information

QUALITY MANAGEMENT SYSTEM MANUAL

QUALITY MANAGEMENT SYSTEM MANUAL The online version of this document is controlled. Therefore, all printed versions of this document are unofficial copies. QUALITY MANAGEMENT SYSTEM MANUAL 6901 Charles Street Towson, Maryland 21204 Manual

More information

Quality Management System Manual

Quality Management System Manual Quality Management System Manual This manual has been reviewed and approved for use by: Jack Zazulak President, Aurora Machine Limited March 07, 2011 Date - Copyright Notice - This document is the exclusive

More information

arenasolutions.com Whitepaper Reduce Regulatory Delays by Formalizing Design Control Processes

arenasolutions.com Whitepaper Reduce Regulatory Delays by Formalizing Design Control Processes Whitepaper Reduce Regulatory Delays by Formalizing Design Control Processes Reduce Regulatory Delays by Formalizing Design Control Processes The path to market for a medical device is fraught with potential

More information

ISO 9001:2008 Quality Management System Requirements (Third Revision)

ISO 9001:2008 Quality Management System Requirements (Third Revision) ISO 9001:2008 Quality Management System Requirements (Third Revision) Contents Page 1 Scope 1 1.1 General. 1 1.2 Application.. 1 2 Normative references.. 1 3 Terms and definitions. 1 4 Quality management

More information

Draft Template. Quality Manual. For Biotechnology Quality Management System v1b

Draft Template. Quality Manual. For Biotechnology Quality Management System v1b Draft Template For Biotechnology Quality Management System v1b Printed copies are uncontrolled Page 1 of 16 Table of contents: Section Page 4.0 Biotechnology Quality Management System. 3-4 5.0 Management

More information

RAC (US) Examination Study Checklist

RAC (US) Examination Study Checklist RAC (US) Examination Study Checklist Instructions: Use this checklist to track your progress when preparing for the RAC (US) certification examination. When you begin your studying, each task statement

More information

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR STUDIO on the Cloud Data Management Software DECISION SUMMARY

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR STUDIO on the Cloud Data Management Software DECISION SUMMARY A. DEN Number: DEN140016 EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR STUDIO on the Cloud Data Management Software B. Purpose for Submission: DECISION SUMMARY De novo request for adjunct data management

More information

Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices

Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices Experts in medical device quality systems, compliance, regulatory affairs, auditing and training www.nsf.org The Right People.

More information

AS9100 Quality Manual

AS9100 Quality Manual Origination Date: August 14, 2009 Document Identifier: Quality Manual Revision Date: 8/5/2015 Revision Level: Q AS 9100 UNCONTROLLED IF PRINTED Page 1 of 17 1 Scope Advanced Companies (Advanced) has established

More information

Quality Manual TABLE OF CONTENTS APPROVAL SIGNATURE PAGE 1.1 1 AMENDMENT RECORD 1.2 2 SCOPE 2.0 3 EXCLUSIONS 2.1 3

Quality Manual TABLE OF CONTENTS APPROVAL SIGNATURE PAGE 1.1 1 AMENDMENT RECORD 1.2 2 SCOPE 2.0 3 EXCLUSIONS 2.1 3 TABLE OF CONTENTS DESCRIPTION SECTION PAGE INTRODUCTION 1.0 1 APPROVAL SIGNATURE PAGE 1.1 1 AMENDMENT RECORD 1.2 2 SCOPE 2.0 3 EXCLUSIONS 2.1 3 CORPORATE POLICY 3.0 3 QUALITY MANAGEMENT SYSTEM 4.0 4 GENERAL

More information

AS 9100 Rev C Quality Management System Manual. B&A Engineering Systems, Inc. 3554 Business Park Drive, Suite A-1 Costa Mesa, CA 92626

AS 9100 Rev C Quality Management System Manual. B&A Engineering Systems, Inc. 3554 Business Park Drive, Suite A-1 Costa Mesa, CA 92626 AS 9100 Rev C Quality Management System Manual B&A Engineering Systems, Inc. 3554 Business Park Drive, Suite A-1 Costa Mesa, CA 92626 Doc. No. AS9100C Rev E Effective Date: 01 JAN 2013 Page 2 of 45 CONTROLLED

More information

FINAL DOCUMENT. Global Harmonization Task Force

FINAL DOCUMENT. Global Harmonization Task Force FINAL DOCUMENT Global Harmonization Task Force Title: Quality management system Medical Devices Guidance on corrective action and preventive action and related QMS processes Authoring Group: Study Group

More information

INTEGRATED MANAGEMENT SYSTEM MANUAL IMS. Based on ISO 9001:2008 and ISO 14001:2004 Standards

INTEGRATED MANAGEMENT SYSTEM MANUAL IMS. Based on ISO 9001:2008 and ISO 14001:2004 Standards INTEGRATED MANAGEMENT SYSTEM MANUAL IMS Based on ISO 9001:2008 and ISO 14001:2004 Standards Approved by Robert Melani Issue Date 30 December 2009 Issued To Management Representative Controlled Y N Copy

More information

What We Are..! www.ardent-cro.com

What We Are..! www.ardent-cro.com Your Trusted CRO! Regus, Level-2, Connaught Place, Bund Garden Road, Pune-411001, MH, India. Phone: 020-65-31-31-71 Email: ardent@ardent-cro.com Web: What We Are..! Ardent Clinical Research Services is

More information

Quality by Design (QbD) Overview

Quality by Design (QbD) Overview Quality by Design (QbD) Overview Gary Warren Director, Haemostasis and Thrombosis R&D October, 2015 CSL Behring Pty Ltd Broadmeadows, Victoria What is Quality by Design (QbD)? QbD is: A Quality System

More information

Medical Software Development. International standards requirements and practice

Medical Software Development. International standards requirements and practice Medical Software Development International standards requirements and practice Food and Drug Administration What? A public health agency Why? Protect American consumers How? By enforcing the Federal Food,

More information

Implementation of a Quality Management System for Aeronautical Information Services -1-

Implementation of a Quality Management System for Aeronautical Information Services -1- Implementation of a Quality Management System for Aeronautical Information Services -1- Implementation of a Quality Management System for Aeronautical Information Services Chapter IV, Quality Management

More information

General Principles of Software Validation; Final Guidance for Industry and FDA Staff

General Principles of Software Validation; Final Guidance for Industry and FDA Staff General Principles of Software Validation; Final Guidance for Industry and FDA Staff Document issued on: January 11, 2002 This document supersedes the draft document, "General Principles of Software Validation,

More information

QUALITY MANAGEMENT SYSTEM Corporate

QUALITY MANAGEMENT SYSTEM Corporate Page 1 of 12 4 Quality Management System 4.1 General Requirements The Peerless Pump Quality Management System shall include: Documented statements of a quality policy and of quality objectives; A quality

More information

Waters Quality System Documentation

Waters Quality System Documentation (x) POLICY - Policy ( ) PROCED - Procedure ( ) WKINST Work Instructions 1 of 39 TABLE OF CONTENTS 1. General... 3 1.1 Introduction... 3 1.3 Process Approach... 4 2. Permissible Exclusions And Items Not

More information

QMS Operational Procedure QOP-42-02

QMS Operational Procedure QOP-42-02 QMS Operational Procedure QOP-42-02 DEVICE MASTER RECORD Issued by: Quality Assurance Eff. Date: 00/00/00 Rev.: A Pg. 1 of 4 ISO 13485 requires that for each type or model of medical device the organization

More information