AN EXPERIMENTAL MODEL TO MEASURE REAL INTRAOCULAR PRESSURE (IOP) DURING LASIK

Size: px
Start display at page:

Download "AN EXPERIMENTAL MODEL TO MEASURE REAL INTRAOCULAR PRESSURE (IOP) DURING LASIK"

Transcription

1 ARCH SOC ESP OFTALMOL 2007; 82: ORIGINAL ARTICLE AN EXPERIMENTAL MODEL TO MEASURE REAL INTRAOCULAR PRESSURE (IOP) DURING LASIK MODELO EXPERIMENTAL PARA MEDIR LA PRESIÓN INTRAOCULAR (PIO) DURANTE LASIK HERNÁNDEZ-VERDEJO JL 1, BOLÍVAR-DE-MIGUEL G 2, ROMÁN-GUINDO JM 3, TEUS-GUEZALA MA 4 ABSTRACT Purpose: To design a model to measure real IOP during lasik in pig eyes. Methods: A reusable blood pressure transducer was used. Each eye was placed on a stand with sufficient support for the surgical procedure. The globes were inflated with 5% glucosated solution through the optic nerve to attain an IOP of mmhg. A 27-gauge catheter was then inserted from the pars plana to the anterior chamber cavity. Real IOP was transmitted via the catheter liquid column to an external sensor. A suction ring was applied and a flap was created. During the procedure, IOP was registered with the ML110 Bridge Amplifier connected to a baro transducer. Results: Values of real IOP were measured in freshly enucleated pig eyes during the lasik procedure. The model was able to instantaneously register the changes in IOP induced by the application of a suction ring and the creation of a flap in the pig eyes with good reproducibility. At the end of the procedure the IOP values remained stable. Conclusions: We have developed an experimental model to observe the course of suctioning and chan- RESUMEN Objetivo: Diseñar un modelo para medir la presión intraocular real (PIO) durante el procedimiento LASIK (Laser Assisted in situ Keratomileusis) en ojos de cerdo. Métodos: En este experimento se utilizó un transductor de presión sanguínea. Cada ojo se colocó sobre una superficie con soporte como para realizar el procedimiento quirúrgico. Los ojos se insuflaron con una solución glucosada al 5% introducida a través del nervio óptico con el fin de conseguir una presión intraocular de inicio de entre 8-20 mmhg. En ese momento, un catéter de 27G se insertó desde la pars-plana hacia la cámara anterior. La PIO real se transmitió vía la columna líquida del catéter hacia un sensor externo. Después se colocó el anillo de succión y se realizó el corte. Durante todo el proceso, la PIO fue registrada mediante un amplificador ML 110 Bridge conectado al baro-transductor. Resultados: Se midieron los valores de PIO real, en siete ojos de cerdo enucleados, durante el procedimiento LASIK. El modelo fue capaz de registrar instantáneamente los cambios de PIO inducidos por Received: 31/5/06. Accepted: 25/7/07. VISSUM Ophthalmological Corporation. Madrid. Spain. 1 Graduate in Optics and Optometrics. VISSUM Madrid. EU in Optics. Complutense University. Madrid. Spain. 2 Graduate in Medicine. Prince of Asturias University Hospital. Alcalá de Henares. Madrid. Spain. 3 Graduate in Medicine. VISSUM Madrid. 4 Ph.D. in Medicine. VISSUM Madrid. Prince of Asturias Uni versity Hospital. Alcalá de Henares. Madrid. Spain. Communication partially presented at the LXXXII Congress of S.E.O. (A Coruña 2006). Correspondence: José Luis Hernández Verdejo VISSUM Corporación Oftalmológica C/. Santa Hortensia, Madrid Spain jlhernan@opt.ucm.es

2 HERNÁNDEZ-VERDEJO JL, et al. ges in IOP simultaneously. This model seems to be accurate in obtaining IOP values from the anterior chamber cavity during lasik (Arch Soc Esp Oftalmol 2007; 82: ). Key words: Real IOP, Lasik, manometry, porcine eye, anterior chamber. la aplicación del anillo de succión y la creación del flap corneal con buena reproductibilidad. Al final del procedimiento, los valores de PIO resultantes permanecían estables. Conclusiones: Hemos desarrollado un modelo experimental para observar simultáneamente durante la succión del ojo y la creación del flap los cambios de presión intraocular que se producen. Este modelo parece ofrecer con precisión los valores de PIO en la cámara anterior durante el procedimiento LASIK. Palabras clave: PIO real, lasik, manometría, ojo cerdo, cámara anterior. INTRODUCTION Having a model with which IOP (intraocular pressure) can be measured in a sensitive and immediate manner when ocular surgery is performed is something useful that allows recording pressure changes induced by surgical procedures. These experiments cannot be carried out with humans for ethical reasons, thus they require developing animal models presenting sufficient anatomical similarities to human eyes. Some authors have measured IOP during corneal refractive surgery. Bissen-Miyajima H et al. (1) developed an experimental model to observe the course of suction and changes in IOP simultaneously. Kasetsuwan N et al (2) measured actual IOP during keratomileusis and determined the variability of the flap thickness using different adjustments to the suction pressures. Sachs HG et al. (3) evaluated IOP during LASIK procedure in pig eyes using two different microkeratome systems. In these studies the measurement system used, with some variations, was through cannulation of the vitreous chamber connecting it to a pressure sensor. Pressure values, therefore, were recorded from an area in the eye not directly affected by the surgical procedure. This fact may give rise to certain questions about the reliability of the measurements given that the vitreous maintains a viscous consistency which hinders mobility through the liquid column used in these measurement systems. For this reason and in order to overcome the natural disadvantages of the models previously mentioned, we decided to design a model in which IOP values were recorded directly in the anterior chamber of the eye, a space occupied by the aqueous humor, with less consistency than vitreous and more easily moved. SUBJECTS, MATERIAL AND METHOD In this experimental model we used enucleated pig eyes and we evaluated prospectively the changes in IOP that arose from application of the suction ring until the flap creation was finalized with a phemtosecond laser (IntraLase Corp. Irvine, CA). We used seven anatomically intact pig eyes, free from corneal damages, the latter having been verified with a slit lamp. The eyes were insufflated with a 5% glucose solution through the optic nerve (similar to the experiment conducted by Kasetsuwan et al.) in order to obtain an IOP between 15 and 20 mmhg, confirmed with Perkins applanation tonometry (Clement Clarke, Essex, England), and placed on a surface with enough support to be able to conduct the surgical procedure. We measured existing IOP in anterior chamber using a 27G catheter (Set ref: BD Valu-Set ) inserted intrasclerally such that it would not interfere with the application of the suction ring, and ensuring that no end of the catheter touched any intraocular structure throughout the surgical procedure. We obtained the pressure measurements through a reusable blood pressure transducer (MLT0380 Reusable BP Transducer, Power Lab, AD Instruments, Racine, WI). The transducer is an external 542 ARCH SOC ESP OFTALMOL 2007; 82:

3 Actual IOP in LASIK vascular pressure sensor connected (in our experiment IOP of the anterior chamber) through a catheter-fluid column present within a tube of silicone connected to the transducer. The transducer was adjusted according to the manufacturer s instructions in order to ensure precise adherence, precise recording of IOP and elimination of any possibility of air presence in the system. The transducer in its turn was connected to an amplifier (ML110 Bridge Amplifier, AD Instruments) that instantly showed the existing intraocular pressure. At that moment, we applied the suction ring and proceeded to create the corneal flap. All surgeries were conducted on the same day by the same experienced surgeon (JMR). During the procedure we continuously recorded IOP, right from the moment the suction ring was applied until the final moment of the cut conducted with the phemtosecond laser. The IOP values were also recorded before and after the procedure with «Perkins» applanation tonometry. Considering that the high pressure levels present during the experiment could facilitate the evacuation path of the fluid present in the eye, we assumed that if the pressure level, after surgery, was not less than 6 mmhg it would mean there was no fluid leak that could invalidate the experiment. Average time to complete suction was 40 SD 9.75 sec (range, sec) and average time required to create the flap was SD SD 5.66 sec (range, sec), total average time being SD SD sec. The average level of IOP for the seven eyes before placing the suction ring and starting the experiment was SD 3.43 mmhg reaching an average value at the end of the procedure of 8.85 SD 2.11 mmhg. DISCUSSION We designed a model that allows measuring intraocular pressure during surgical procedures on the eye when subject to LASIK surgery. The use of invasive manometry is a usual way of measuring intraocular pressure especially in animal studies. This method is presented as the most realistic way to record pressure used as a reference to assess the precision of other non-invasive means of measurement. When reviewing existing literature, in some papers the cannulation necessary to record with RESULTS We evaluated actual IOP during surgery in seven enucleated pig eyes. The average IOP obtained during the procedure was SD mmhg throughout the time of eye suction and SD mmhg during flap creation. Table I. The table below shows the mean IOP values, standard deviation and ranges in mmhg for each eye for the full duration of the experiment, i.e., from the placement of the suction ring to the end of the flap creation Fig. 1: This figure shows average values and standard deviation for each eye separately of IOP, expressed in mmhg, reached during suction phase. Eye Mean SD Range Fig. 2: This figure shows average values and standard deviation for each eye separately of IOP, expressed in mmhg, reached during the flap creation phase. ARCH SOC ESP OFTALMOL 2007; 82:

4 HERNÁNDEZ-VERDEJO JL, et al. manometry was conducted through the vitreous chamber while in others it was done through the anterior chamber. In our opinion, both means of measurement can show considerable differences. The studies by Bissen-Miyajima H et al. (1) as well as those by Kasetsuwan N et al (2) measure IOP using a sensor inserted in the vitreous chamber. Likewise, in an experiment conducted by Flaxel CJ et al. (4) in human cadaver eyes they proceeded in the same manner through cannulation of the vitreous chamber for manometric recording in order to prove the relation of the suction using during refractive LASIK surgery and predisposing retinal lesions with regard to a change in the axial longitude of the eye during surgery. However, other studies conducted cannulation through anterior chamber, such as in the experiments by Lim JI et al. (5) where they compared the precision of pneumotonometry and tonometry by means of Tono-Pen, in both cases using as reference the manometry pressure measurement. Likewise, in another publication, Pasaglia CL et al. (6) also used the same cannulation procedure of the anterior chamber to gauge non-invasive measurements conducted with the Ton-Pen KL in the enucleated eyes of newborn kittens. We must take into account that in experiments carried out with cannulation of the vitreous chamber, since a glucose solution is insufflated through the optic nerve, a certain level of vitreous liquefaction is achieved, facilitating movement of the fluid. Nonetheless, this situation certainly makes it more difficult to assume that the values obtained can be reproduced as they depend to a great extent on the level of viscosity achieved during preparation of the eye. Our experiment holds greater similarities with the studies with cannulation in the anterior chamber, which, in our opinion, makes recording of the values obtained more reliable. Another important aspect that provides validity to the proposed method is the fact that the final intraocular pressure is maintained within normal values, which could justify the lack of fluid leaks during the procedure, something which could have invalidated the experiment. With our model of cannulation of the anterior chamber we were capable of recording IOP values both in the suction phase (40 SD 9.75 sec) and in the cutting phase (52.85 SD 5.66 sec). When analyzing existing fluctuations in IOP, considered as the standard deviation present in each set of measurements, it is worth highlighting a greater stability present in the cutting phase compared to the suction phase. This fact can be explained as a result of the non-manipulation of the eye by the surgeon at this phase; and on the contrary, a possible source of the existing fluctuations during the suction phase. Even taking into account the advantages provided by this invasive method of measuring IOP during certain surgical procedures, our design is not exempt from disadvantages which should be taken into account for future improvements. On the one hand, cannulation of the anterior chamber intrasclerally can be somewhat complicated and can lead to additional difficulties for the surgeon when placing the suction ring. Also, another consideration to take into account is that the IOP measurement obtained may not be the one actually existing in other areas of the eye, although the physical properties of fluids indicate that the pressure exercised upon a point in a watertight compartment is directly transmitted towards all the points of that compartment. Again, the fact that the IOP at the end of our experiment was maintained within normal levels leads us to believe the eye acts as such a compartment throughout the experiment, thus transmitting the IOP values to all the points inside the eye. Recently, Morris, CA et al. (7) compared the precision of non-invasive rebound tonometry with the IOP values obtained through cannulation in an experiment with animals related to the hypotensive effect of topical prostaglandin, demonstrating there were no statistically significant differences between either forms of measurements. Taking into account that the pressure measurement with cannulation is shown to date as the best way for an actual measurement of IOP, it is logical to believe that if we also measure it in the anterior chamber of the eye, thus avoiding the disadvantages previously mentioned of doing it in the vitreous chamber, this might prove to be the best way to note the actual pressure inside the eye. REFERENCES 1. Bissen-Miyajima H, Suzuki S, Ohashi Y, Minami K. Experimental observation of intraocular pressure changes during microkeratome suctioning in laser in situ keratomileusis. J Cataract Refract Surg 2005; 31: Kasetsuwan N, Pangilinan RT, Moreira LL, DiMartino DS, Shah SS, Schallhorn SC, et al. Real time intraocular 544 ARCH SOC ESP OFTALMOL 2007; 82:

5 Actual IOP in LASIK pressure and lamellar corneal flap thickness in keratomileusis. Cornea 2001; 20: Sachs HG, Lohmann CP, Op de Laak JP. Intraocular pressure in sections with 2 microkeratomes in vitro. Ophthalmologe 1997; 94: Flaxel CJ, Choi YH, Sheety M, Oeinck SC, Lee JY, McDonnell PJ. Proposed mechanism for retinal tears after LASIK: an experimental model. Ophthalmology 2004; 111: Lim JI, Blair NP, Higginbothan EJ, Farber MD, Shaw WE, Garretson BR. Assessment of intraocular pressure in vitrectomized gas-containing eyes. A clinical a manometric comparison of the Tono-Pen to the pneumotonometer. Arch Ophthalmol 1990; 108: Passaglia CL, Guo X, Chen J, Troy JB. Tono-Pen XL calibration curves for cats, cows and sheep. Vet Ophthalmol 2004; 7: Morris CA, Crowston JG, Lindsey JD, Danias J, Weinreb RN. Comparison of invasive and non-invasive tonometry in the mouse. Exp Eye Res 2006; 82: ARCH SOC ESP OFTALMOL 2007; 82:

COMPARISON OF DYNAMIC CONTOUR TONOMETRY (PASCAL ) WITH PNEUMOTONOMETRY AND GOLDMANN TONOMETRY

COMPARISON OF DYNAMIC CONTOUR TONOMETRY (PASCAL ) WITH PNEUMOTONOMETRY AND GOLDMANN TONOMETRY ARCH SOC ESP OFTALMOL 2007; 82: 337-342 ORIGINAL ARTICLE COMPARISON OF DYNAMIC CONTOUR TONOMETRY (PASCAL ) WITH PNEUMOTONOMETRY AND GOLDMANN TONOMETRY COMPARACIÓN DEL TONÓMETRO PASCAL CON EL NEUMOTONÓMETRO

More information

How do you adjust for IOP in a LASIK patient?

How do you adjust for IOP in a LASIK patient? How do you adjust for IOP in a LASIK patient? Sunita Radhakrishnan, M.D Glaucoma Center of San Francisco Glaucoma Research and Education Group Principle of applanation tonometry Imbert Fick law Applanation

More information

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US JON DISHLER, MD DENVER, COLORADO, USA INTRODUCTION AND STUDY OBJECTIVES This article summarizes the first US

More information

Profiles of Intraocular Pressure in Human Donor Eyes during Femtosecond Laser Procedures A Comparative Study

Profiles of Intraocular Pressure in Human Donor Eyes during Femtosecond Laser Procedures A Comparative Study Cornea Profiles of Intraocular Pressure in Human Donor Eyes during Femtosecond Laser Procedures A Comparative Study Clemens Strohmaier, Christian Runge, Orang Seyeddain, Martin Emesz, Christian Nischler,

More information

Case Report Laser Vision Correction on Patients with Sick Optic Nerve: A Case Report

Case Report Laser Vision Correction on Patients with Sick Optic Nerve: A Case Report Case Reports in Ophthalmological Medicine Volume 2011, Article ID 796463, 4 pages doi:10.1155/2011/796463 Case Report Laser Vision Correction on Patients with Sick Optic Nerve: A Case Report Ming Chen

More information

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay Minoru Tomita, MD, PhD 1,2 1) Shinagawa LASIK, Tokyo,

More information

Retreatment by Lifting the Original Laser in Situ Keratomileusis Flap after Eleven Years

Retreatment by Lifting the Original Laser in Situ Keratomileusis Flap after Eleven Years Retreatment by Lifting the Original Laser in Situ Keratomileusis Flap after Eleven Years Hassan Hashemi, MD 1,2 Mehrdad Mohammadpour, MD 3 Abstract Purpose: To describe a case of successful laser in situ

More information

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening GALILEI G4 Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software

More information

INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the

INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the INTRACOR An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the 1 Dr Luis Ruiz Presbyopia treatment with INTRACOR Luis

More information

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM 1 BoydVision TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM Risks and Side Effects... 2 Risks Specific to PRK... 3 Risks Specific to LASIK... 4 Patient Statement of Consent... 5 Consent for Laser Eye

More information

Long-Term Outcomes of Flap Amputation After LASIK

Long-Term Outcomes of Flap Amputation After LASIK Long-Term Outcomes of Flap Amputation After LASIK Priyanka Chhadva BS, Florence Cabot MD, Anat Galor MD, Sonia H. Yoo MD Bascom Palmer Eye Institute, University of Miami Miller School of Medicine Miami

More information

Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery

Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery Anterior Lamellar Keratoplasty With a Microkeratome: A Method for Managing Complications After Refractive Surgery Farhad Hafezi, MD; Michael Mrochen, PhD; Franz Fankhauser II, MD; Theo Seiler, MD, PhD

More information

New and Improved Femtosecond Laser Applications. Karl Stonecipher, MD Wavefront Congress 2008

New and Improved Femtosecond Laser Applications. Karl Stonecipher, MD Wavefront Congress 2008 New and Improved Femtosecond Laser Applications Karl Stonecipher, MD Wavefront Congress 2008 When birds don t fly, neither should you. When cows bunch together in a field, a storm is coming. When ants

More information

Bascom Palmer Eye Institute. LASIK and Your Vision Correction

Bascom Palmer Eye Institute. LASIK and Your Vision Correction Bascom Palmer Eye Institute LASIK and Your Vision Correction For Many Patients, the Wait is Over The Vision Correction Center at Bascom Palmer For many patients who want to see near, far and everywhere

More information

Cataract Testing. What a Patient undergoes prior to surgery

Cataract Testing. What a Patient undergoes prior to surgery Cataract Testing What a Patient undergoes prior to surgery FINANCIAL DISCLOSURE I have no financial interest or relationships to disclose What do most Technicians find to be the most mundane yet very important

More information

laser eye care center

laser eye care center laser eye care center Providing Increased Visual Freedom from Glasses and Contacts What is LASIK? Welcome to the Laser Eye Care Center Eyesight is one of our most precious resources. Keeping our eyes healthy

More information

Femto-LASIK. Pulsewidth: Ultrashort-pulse micro- machining can make sub- wavelength holes. micromachining

Femto-LASIK. Pulsewidth: Ultrashort-pulse micro- machining can make sub- wavelength holes. micromachining All-laser laser LASIK (Femto( Femto-LASIK) Femto-LASIK 台 大 眼 科 王 一 中 IntraLase 2/1 Perfect Vision Ziemer (DaVinci) Carl Zeiss Meditec Pulsewidth: Femtosecond laser (Nd:Glass)) 153 nm (near infrared) Each

More information

BSM Connection elearning Course

BSM Connection elearning Course BSM Connection elearning Course Scope of the Eye Care Practice 2008, BSM Consulting All Rights Reserved. Table of Contents OVERVIEW...1 THREE O S IN EYE CARE...1 ROUTINE VS. MEDICAL EXAMS...2 CONTACT LENSES/GLASSES...2

More information

Alexandria s Guide to LASIK

Alexandria s Guide to LASIK Alexandria s Guide to LASIK A Community Service Project sponsored by: Wallace Laser Center Your Guide To A Successful LASIK Procedure The word LASIK is actually an acronym for Laser Assisted In-Situ Keratomileusis.

More information

LASIK SURGERY OUTCOMES, VOLUME AND RESOURCES

LASIK SURGERY OUTCOMES, VOLUME AND RESOURCES MOH Information Paper: 2006/17 LASIK SURGERY OUTCOMES, VOLUME AND RESOURCES By Dr. Ganga Ganesan 1 I INTRODUCTION LASIK stands for Laser-Assisted In Situ Keratomileusis and is a surgical procedure that

More information

Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD

Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD Post LASIK Ectasia Gina M. Rogers, MD and Kenneth M. Goins, MD October 6, 2012 Chief Complaint: Decreasing vision after laser- assisted in- situ keratomileusis (LASIK) History of Present Illness: This

More information

CONSENT FOR CATARACT SURGERY REQUEST FOR SURGICAL OPERATION / PROCEDURE AND ANAESTHETIC

CONSENT FOR CATARACT SURGERY REQUEST FOR SURGICAL OPERATION / PROCEDURE AND ANAESTHETIC CONSENT FOR CATARACT SURGERY REQUEST FOR SURGICAL OPERATION / PROCEDURE AND ANAESTHETIC Your doctor has indicated that the condition of your eye appears stable and your cataract surgery and/or implantation

More information

Long-term stability of the posterior cornea after laser in situ keratomileusis

Long-term stability of the posterior cornea after laser in situ keratomileusis ARTICLE Long-term stability of the posterior cornea after laser in situ keratomileusis Joseph B. Ciolino, MD, Stephen S. Khachikian, MD, Michael J. Cortese, OD, Michael W. Belin, MD PURPOSE: To study long-term

More information

Vision Correction Surgery Patient Information

Vision Correction Surgery Patient Information Vision Correction Surgery Patient Information Anatomy of the eye: The eye is a complex organ composed of many parts, and normal vision requires these parts to work together. When a person looks at an object,

More information

How To Know If You Can See Without Glasses Or Contact Lense After Lasik

How To Know If You Can See Without Glasses Or Contact Lense After Lasik The LASIK experience I WHO CAN HAVE LASIK? To be eligible for LASIK you should be at least 21 years of age, have healthy eyes and be in good general health. Your vision should not have deteriorated significantly

More information

Overview of Refractive Surgery

Overview of Refractive Surgery Overview of Refractive Surgery Michael N. Wiggins, MD Assistant Professor, College of Health Related Professions and College of Medicine, Department of Ophthalmology Jones Eye Institute University of Arkansas

More information

Shinagawa LASIK Center. Examination rooms

Shinagawa LASIK Center. Examination rooms Examination rooms Since was established in October 2004 and from that time through March 2010 we have performed over 700,000 cases. in Tokyo (70% of operations) and 3 branch clinics in Osaka, Nagoya, and

More information

Minimally Invasive Surgery: Femtosecond Lasers and Other Innovative Microsurgical Techniques

Minimally Invasive Surgery: Femtosecond Lasers and Other Innovative Microsurgical Techniques Minimally Invasive Surgery: Femtosecond Lasers and Other Innovative Microsurgical Techniques Julio Narváez MD Associate Professor of Ophthalmology Loma Linda University Non-Refractive Applications of Femtosecond

More information

The cataract laser technology of tomorrow is here for you today.

The cataract laser technology of tomorrow is here for you today. The cataract laser technology of tomorrow is here for you today. See inside to learn about all your exciting new options Thanks to new advancements, this is a great time to have cataract surgery Laser

More information

Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps

Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps Lingo Y. Lai, MD William G. Zeh, MD Clark L. Springs, MD The authors

More information

Effect of Preoperative Keratometric Power on Intraoperative Complications in LASIK in 34,099 Eyes

Effect of Preoperative Keratometric Power on Intraoperative Complications in LASIK in 34,099 Eyes Effect of Preoperative Keratometric Power on Intraoperative Complications in LASIK in 34,099 Eyes J. Carlos Albelda-Vallés, MD; Clara Martin-Reyes, MD; Francisco Ramos, MD; Jaime Beltran, MD; Fernando

More information

LASIK. What is LASIK? Eye Words to Know. Who is a good candidate for LASIK?

LASIK. What is LASIK? Eye Words to Know. Who is a good candidate for LASIK? 2014 2015 LASIK What is LASIK? LASIK (laser in situ keratomileusis) is a type of refractive surgery. This kind of surgery uses a laser to treat vision problems caused by refractive errors. You have a refractive

More information

Uniquely Safe. predictably better for our patients. enhancement, may be significantly reduced.

Uniquely Safe. predictably better for our patients. enhancement, may be significantly reduced. Uniquely Safe Clinical Support: Six different studies verify the improved safety of flap creation with the INTRALASE FS laser when compared to traditional microkeratomes. Clinical studies validate the

More information

Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy

Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy Keratorefractive Surgery for Post-Cataract Refractive Surprise Moataz El Sawy Departmentof Ophthalmology, Faculty of Medicine,MenoufiyaUniversity, Egypt mfelsawy@yahoo.co.uk Abstract: Purpose: To evaluate

More information

ALTERNATIVES TO LASIK

ALTERNATIVES TO LASIK EYE PHYSICIANS OF NORTH HOUSTON 845 FM 1960 WEST, SUITE 101, Houston, TX 77090 Office: 281 893 1760 Fax: 281 893 4037 INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) INTRODUCTION This information

More information

Cataract Surgery in Small Eyes Richard S. Hoffman, MD Clinical Associate Professor of Ophthalmology Oregon Health & Science University No Financial Interests Anatomic Classification Short AC depth with

More information

Wavefront technology has been used in our

Wavefront technology has been used in our Wavefront Customized Ablations With the WASCA Asclepion Workstation Sophia I. Panagopoulou, BSc; Ioannis G. Pallikaris, MD ABSTRACT PURPOSE: WASCA (Wavefront Aberration Supported Cornea Ablation) is a

More information

Tucson Eye Care, PC. Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens

Tucson Eye Care, PC. Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens Tucson Eye Care, PC Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens INTRODUCTION This information is provided so that you may make an informed decision about having eye

More information

OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment

OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment Shinagawa LASIK Center OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment Minoru Tomita, MD, Ph.D 1) Executive Medical Director at Shinagawa LASIK Center, Tokyo, Japan 2) Clinical Professor

More information

&855,&8/809,7$( 1993/94-1995/96 First three years in Physics University of Valladolid, Spain. College of Optometry. University of Houston, Texas USA.

&855,&8/809,7$( 1993/94-1995/96 First three years in Physics University of Valladolid, Spain. College of Optometry. University of Houston, Texas USA. &855,&8/809,7$( '(02*5$3+,&,1)250$7,21 1$0(Sergio Barbero Briones 3/$&(2)%,57+: Madrid, SPAIN '$7(2)%,57+ December 31, 1976 (PDLODGGUHVV: sergibarbero@yahoo.es ('8&$7,21$1'75$,1,1* (GXFDWLRQ 1993/94-1995/96

More information

Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical,

Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical, sight see OLYMPIA EYE & LASER CENTRE Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical, At the Olympia Eye & Laser Centre, our vision is to improve

More information

LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY

LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY Thi-Qar Medical Journal (TQMJ): Vol(4) No(4):1(14-21) SUMMARY: LASIK SURGERY IN AL- NASSIRYA CITY A CLINICOSTATISTICAL STUDY Dr. Ali Jawad AL- Gidis (M.B.Ch.B., D.O., F.I.C.O.)* Background: LASIK which

More information

Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking

Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking Kay Lam, MD, Dan B. Rootman, MSc, Alejandro Lichtinger, and David S. Rootman, MD, FRCSC Author affiliations:

More information

Epi-LASIK Surgery. The procedure of choice for surface ablation.

Epi-LASIK Surgery. The procedure of choice for surface ablation. Epi-LASIK Surgery The procedure of choice for surface ablation. powered by Moria Preservation of Corneal Biomechanical Integrity, minimizing the risks of ectasia Post-LASIK CH decrease (Reichert ORA) 0-1

More information

PATIENT CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)

PATIENT CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) INTRODUCTION: You have been diagnosed with myopia (nearsightedness) or hyperopia (farsightedness) with or without astigmatism, or astigmatism alone. Myopia is a result of light entering the eye and focusing

More information

ReLEx smile Minimally invasive vision correction Information for patients

ReLEx smile Minimally invasive vision correction Information for patients ReLEx smile Minimally invasive vision correction Information for patients Seeing is living Our eyes are our most important sensory organ. The human brain obtains over 80 % of its information via the sense

More information

Femtosecond Lasers in LASIK Surgery

Femtosecond Lasers in LASIK Surgery Femtosecond Lasers in LASIK Surgery Dr Chan Tat Keong Senior Consultant Refractive Surgery Service Singapore National Eye Centre Disclosure Speaker has no financial interest in the products to be discussed

More information

Laser Refractive Cataract Surgery with the LenSx Laser

Laser Refractive Cataract Surgery with the LenSx Laser Laser Refractive Cataract Surgery with the LenSx Laser a Novartis company LenSx Laser 2 3 An Evolving Definition of Cataract Average Age of US Cataract Patient is Projected to Decline Today: earlier diagnosis

More information

EFFECT OF MYOPIC LASIK ON RETINAL NERVE FIBER LAYER THICKNESS- IS IT SAFE OR UNSAFE?

EFFECT OF MYOPIC LASIK ON RETINAL NERVE FIBER LAYER THICKNESS- IS IT SAFE OR UNSAFE? 24. Glaucoma: Imaging EFFECT OF MYOPIC LASIK ON RETINAL NERVE FIBER LAYER THICKNESS- IS IT SAFE OR UNSAFE? Chief Author: Dr. Amit porwal 1 Co-Authors: Dr. Kavita Porwal 2, Dr. Puja Rai 1 1. Choithram Netralaya,

More information

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) USING INTRALASE TM BLADE-FREE TECHNOLOGY

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) USING INTRALASE TM BLADE-FREE TECHNOLOGY EYE PHYSICIANS OF NORTH HOUSTON 845 FM 1960 WEST, SUITE 101, Houston, TX 77090 Office: 281 893 1760 Fax: 281 893 4037 INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) USING INTRALASE TM BLADE-FREE

More information

Bitoric Laser In Situ Keratomileusis for the Correction of Simple Myopic and Mixed Astigmatism

Bitoric Laser In Situ Keratomileusis for the Correction of Simple Myopic and Mixed Astigmatism Bitoric Laser In Situ Keratomileusis for the Correction of Simple Myopic and Mixed Astigmatism Arturo S. Chayet, MD, 1 Miguel Montes, MD, 1 Laura Gómez, MD, 1 Xavier Rodríguez, MD, 1 Nora Robledo, OD,

More information

Spectral domain OCT used to view and quantify choroidal vascular congestion in new subretinal fluid following scleral buckling.

Spectral domain OCT used to view and quantify choroidal vascular congestion in new subretinal fluid following scleral buckling. Spectral domain OCT used to view and quantify choroidal vascular congestion in new subretinal fluid following scleral buckling. Robert Gizicki MD, Mohamed Haji, Flavio Rezende Retina Service, Hôpital Maisonneuve-Rosemont,

More information

iocutouchtm for ipad Contents of Videos and Still Images Anatomy 906. Normal Eye and Orbit - no labels 907. Normal Eye and Orbit - with labels

iocutouchtm for ipad Contents of Videos and Still Images Anatomy 906. Normal Eye and Orbit - no labels 907. Normal Eye and Orbit - with labels iocutouchtm for ipad Contents of Videos and Still Images Anatomy 906. Normal Eye and Orbit - no labels 907. Normal Eye and Orbit - with labels Normal Eye 474. Normal Eye overview 476. Cornea - overview

More information

Original Articles. Laser in situ Keratomileusis to Correct Residual Myopia After Cataract Surgery

Original Articles. Laser in situ Keratomileusis to Correct Residual Myopia After Cataract Surgery Original Articles Laser in situ Keratomileusis to Correct Residual Myopia After Cataract Surgery Maria J. Ayala, MD, PhD; Juan J. Pérez-Santonja, MD; Alberto Artola, MD, PhD; Pascual Claramonte, MD; Jorge

More information

LASIK RESULTS OF INTRALASE FEMTOSECOND LASER TO CORRECT MYOPIA: EFFICACY, SECURITY AND PREDICTABILITY

LASIK RESULTS OF INTRALASE FEMTOSECOND LASER TO CORRECT MYOPIA: EFFICACY, SECURITY AND PREDICTABILITY ARCH SOC ESP OFTALMOL 2007; 82: 423-428 ORIGINAL ARTICLE LASIK RESULTS OF INTRALASE FEMTOSECOND LASER TO CORRECT MYOPIA: EFFICACY, SECURITY AND PREDICTABILITY RESULTADOS DE LASIK REALIZADO CON LÁSER DE

More information

Providing Optimal Optics For Your Astigmatic Cataract Patients. While the cornea remains relatively stable and prolate throughout life

Providing Optimal Optics For Your Astigmatic Cataract Patients. While the cornea remains relatively stable and prolate throughout life Providing Optimal Optics For Your Astigmatic Cataract Patients David I. Geffen, OD, FAAO Why keep the crystalline lens? While the cornea remains relatively stable and prolate throughout life Unless we

More information

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State University Columbus, Ohio Refractive Errors Myopia (nearsightedness)

More information

How To See With An Cl

How To See With An Cl Deciding on the vision correction procedure that s right for you is an important one. The table below provides a general comparison of the major differences between Visian ICL, LASIK and PRK. It is NOT

More information

IntraLase and LASIK: Risks and Complications

IntraLase and LASIK: Risks and Complications No surgery is without risks and possible complications and LASIK is no different in that respect. At Trusted LASIK Surgeons, we believe patients can minimize these risks by selecting a highly qualified

More information

Consent for LASIK (Laser In Situ Keratomileusis) Retreatment

Consent for LASIK (Laser In Situ Keratomileusis) Retreatment Consent for LASIK (Laser In Situ Keratomileusis) Retreatment Please read the following consent form very carefully. Please initial at the bottom of each page where indicated. Do not sign this form unless

More information

Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure

Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure ORIGINAL ARTICLE Early Transient Visual Acuity Loss After LASIK Due to Steroid-induced Elevation of Intraocular Pressure Joseph Frucht-Pery, MD; David Landau, MD; Frederik Raiskup, MD; Faik Orucov, MD;

More information

LASIK Surgery. One Use-Plus. The Moria Option for SBK

LASIK Surgery. One Use-Plus. The Moria Option for SBK LASIK Surgery One Use-Plus The Moria Option for SBK One Use-Plus Accuracy and predictability equivalent to Femto-SBK One Use-Plus SBK Flap thickness using ultrasound pachymetry 1 Flap thickness with Femto-SBK

More information

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Document issued on: March 27, 2006 For questions regarding this document, contact Everette Beers, Ph.D. at 240-276-4200

More information

Curriculum for the Glaucoma EBO-EGS Subspecialty exam/diploma (Glaucoma-medical)

Curriculum for the Glaucoma EBO-EGS Subspecialty exam/diploma (Glaucoma-medical) Curriculum for the Glaucoma EBO-EGS Subspecialty exam/diploma (Glaucoma-medical) This curriculum will concentrate on: 1) General requirements for the candidate 2) Requirements for the institution and the

More information

Corneal intrastromal implantation surgery for the treatment of moderate and high myopia

Corneal intrastromal implantation surgery for the treatment of moderate and high myopia TECHNIQUE Corneal intrastromal implantation surgery for the treatment of moderate and high myopia Albert Daxer, MD, PhD I describe a corneal intrastromal implantation technique that uses a new type of

More information

femtosecond laser platform Exceptional versatility without compromise

femtosecond laser platform Exceptional versatility without compromise Introducing the VICTUS femtosecond laser platform Exceptional versatility without compromise FEMTOSECOND TECHNOLOGY that empowers Introducing VICTUS the first femtosecond laser capable of exceptional performance

More information

INFORMED CONSENT TO HAVE LASIK

INFORMED CONSENT TO HAVE LASIK A Division of Scott & Christie and Associates INFORMED CONSENT TO HAVE LASIK This information is to help you make an informed decision about having Laser Assisted Intrastromal Keratomileusis (LASIK), an

More information

CENTRO OFTALMOLOGICO GUSTAVO TAMAYO BOGOTA COLOMBIA LASIK XTRA GUSTAVO TAMAYO MD CLAUDIA CASTELL MD PILAR VARGAS MD

CENTRO OFTALMOLOGICO GUSTAVO TAMAYO BOGOTA COLOMBIA LASIK XTRA GUSTAVO TAMAYO MD CLAUDIA CASTELL MD PILAR VARGAS MD CENTRO OFTALMOLOGICO GUSTAVO TAMAYO BOGOTA COLOMBIA LASIK XTRA GUSTAVO TAMAYO MD CLAUDIA CASTELL MD PILAR VARGAS MD From: BOGOTA LASER REFRACTIVE INSTITUTE Bogota, Colombia DISCLOSURES FOR GUSTAVO TAMAYO

More information

One Use-PlusSBK versus LDV Femto Laser Clinical evaluation

One Use-PlusSBK versus LDV Femto Laser Clinical evaluation One Use-PlusSBK versus LDV Femto Laser Clinical evaluation Nikica Gabric, Maja Bohac, Iva Dekaris Eye Clinic Svjetlost Zagreb, Croatia MORIA ONE USE-PLUS SBK MICROKERATOME Automatic linear microkeratome

More information

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University. IOL Power Calculation After Myopic LASIK Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University. SUPERVISORS Prof. Dr. Mohammad El-Hifnawy Professor of Ophthalmology Faculty

More information

Refractive Surgery. Common Refractive Errors

Refractive Surgery. Common Refractive Errors Refractive Surgery Over the last 25 years developments in medical technology and Refractive Surgery allow almost all need for glasses and contact lenses to be eliminated. Currently there are a number of

More information

Initial clinical experience with the FS200 Femto and EX500 excimer

Initial clinical experience with the FS200 Femto and EX500 excimer Initial clinical experience with the FS200 Femto and EX500 excimer lasers for LASIK ASCRS, San Diego 2011 A. John Kanellopoulos, MD Professor NYU Medical School, NY Director, Laservision.gr Institute,

More information

Underestimation of Intraocular Pressure in Eyes After Laser In Situ Keratomileusis

Underestimation of Intraocular Pressure in Eyes After Laser In Situ Keratomileusis Underestimation of Intraocular Pressure in Eyes After Laser In Situ Keratomileusis Ako Arimoto*, Kimiya Shimizu, Nobuyuki Shoji, Kikuko Enomoto and Makio Kohara* *Department of Ophthalmology, Musashino

More information

Richard S. Hoffman, MD. Clinical Associate Professor of Ophthalmology Oregon Health & Science University

Richard S. Hoffman, MD. Clinical Associate Professor of Ophthalmology Oregon Health & Science University Zeiss Mel 80 and Visumax Refractive Laser Systems Richard S. Hoffman, MD Clinical Associate Professor of Ophthalmology Oregon Health & Science University No Financial Interest ZEISS Workstation CRS-Master

More information

Nick Strouthidis MBBS MD PhD FRCS FRCOphth FRANZCO CONSULTANT OPHTHALMIC SURGEON

Nick Strouthidis MBBS MD PhD FRCS FRCOphth FRANZCO CONSULTANT OPHTHALMIC SURGEON TUBES/SHUNTS TUBES/SHUNTS Implantation of a glaucoma drainage device (GDD - also known as a tube or aqueous shunt) works by diverting aqueous from the front of the eye via a tube to a drainage plate stitched

More information

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement) Mark Packer, M.D. Informed Consent for Refractive Lens Exchange (Clear Lens Replacement) This surgery involves the removal of the natural lens of my eye, even though it is not a cataract. The natural lens

More information

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) Lasik Center 2445 Broadway Quincy, IL 62301 217-222-8800 INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) INTRODUCTION This information is being provided to you so that you can make an informed

More information

To date, several million patients have been treated worldwide. So why not discover the benefits The Eye Hospital can bring to your life.

To date, several million patients have been treated worldwide. So why not discover the benefits The Eye Hospital can bring to your life. L a s e r E y e S u r g e r y I N F O R M A T I O N 1 Welcome Imagine the freedom of being able to do away with glasses and contact lenses. You too, may be suitable for laser eye surgery, freeing you from

More information

REFRACTIVE ERROR AND SURGERIES IN THE UNITED STATES

REFRACTIVE ERROR AND SURGERIES IN THE UNITED STATES Introduction REFRACTIVE ERROR AND SURGERIES IN THE UNITED STATES 150 million wear eyeglasses or contact lenses 2.3 million refractive surgeries performed between 1995 and 2001 Introduction REFRACTIVE SURGERY:

More information

Descemet s Stripping Automated Endothelial Keratoplasty (DSAEK)

Descemet s Stripping Automated Endothelial Keratoplasty (DSAEK) Descemet s Stripping Automated Endothelial Keratoplasty (DSAEK) John D. Goosey, MD Introduction DSAEK is a corneal transplant technique where the unhealthy, diseased, posterior portion of a patient s cornea

More information

Accelerated Refractive Performance

Accelerated Refractive Performance Accelerated Refractive Performance Get There at the Speed of WaveLight Designed to accommodate your refractive technology goals now and into the future, the WaveLight Workstation is a faster way to get

More information

EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com

EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com ejpmr, 2015,2(3), 436-440 EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com Tumram et al. SJIF Impact Factor 2.026 Research Article ISSN 3294-3211 EJPMR CLINICAL OUTCOME OF TORIC IOL

More information

Delivers On Its Promise

Delivers On Its Promise Supplement to March 2013 Laser Cataract Surgery Delivers On Its Promise Our Experience With the Catalys Precision Laser System BY PAUL MANN, MD WILLIAM F. WILEY, MD ROBERT P. RIVERA, MD PROFESSOR H. BURKHARD

More information

Technology Brief. Femtosecond lasers for cataract surgery

Technology Brief. Femtosecond lasers for cataract surgery HealthPACT Health Policy Advisory Committee on Technology Australia and New Zealand Technology Brief Femtosecond lasers for cataract surgery February 2012 State of Queensland (Queensland Health) 2012 This

More information

refractive surgery a closer look

refractive surgery a closer look 2011-2012 refractive surgery a closer look How the eye works Light rays enter the eye through the clear cornea, pupil and lens. These light rays are focused directly onto the retina, the light-sensitive

More information

Ulllted States Patent [19] [11] Patent Number: 6,055,990

Ulllted States Patent [19] [11] Patent Number: 6,055,990 US006055990A Ulllted States Patent [19] [11] Patent Number: 6,055,990 Thompson [45] Date of Patent: May 2, 2000 [54] POLYMERIZING GEL 4,969,912 11/1990 Kelman etal...... 623/66 INTRAKERATOPHAKIA-PGI 5,4,408

More information

Surgical Coding Errors & English 101. Riva Lee Asbell. Fort Lauderdale, FL

Surgical Coding Errors & English 101. Riva Lee Asbell. Fort Lauderdale, FL Surgical Coding Errors & English 101 Riva Lee Asbell Fort Lauderdale, FL INTRODUCTION Many surgical coding mistakes result from misinterpretation of CPT (Current Procedural Terminology) wording. When the

More information

Consumer s Guide to LASIK

Consumer s Guide to LASIK Consumer s Guide to LASIK A Community Service Project brought to you by Price Vision Group Your Guide To A Successful LASIK Procedure The purpose of this educational guide is to help prospective patients

More information

Diego Fernando Suárez Sierra, MD Fellow Cornea and Refractive Surgery Fellow Lens and Ocular Surface Vejarano Laser Vision Center

Diego Fernando Suárez Sierra, MD Fellow Cornea and Refractive Surgery Fellow Lens and Ocular Surface Vejarano Laser Vision Center Corneal crosslinking with riboflavin and ultraviolet light before or after subepithelial keratectomy laser-assisted (LASEK) in patients with thin corneas. Diego Fernando Suárez Sierra, MD Fellow Cornea

More information

How To Implant A Keraring

How To Implant A Keraring Corneal Remodeling Using the Keraring A variety of thicknesses, arc lengths, and optical zone sizes allows tailoring of the procedure to the individual patient. BY DOMINIQUE PIETRINI, MD; AND TONY GUEDJ

More information

Dynamic Contour Tonometry

Dynamic Contour Tonometry Chapter Dynamic Contour Tonometry Evelin Schneider, Hartmut E. Kanngiesser, Christoph Kniestedt Core Messages Intraocular pressure is a major risk factor in the pathogenesis of glaucoma Goldmann applanation

More information

Curriculum Vitae. Irwin Y. Cua, M.D. Present Positions:

Curriculum Vitae. Irwin Y. Cua, M.D. Present Positions: Curriculum Vitae Name: Irwin Y. Cua, M.D. Present Positions: Affiliate Consultant, Vision Laser Center, St. Luke s Medical Center Affiliate Consultant, Section of Cornea and External Disease, St. Luke

More information

The «three-in-one» Microkeratome for traditional LASIK, EPI-LASIK, and anterior and posterior Lamellar Keratoplasty

The «three-in-one» Microkeratome for traditional LASIK, EPI-LASIK, and anterior and posterior Lamellar Keratoplasty The «three-in-one» Microkeratome for traditional LASIK, EPI-LASIK, and anterior and posterior Lamellar Keratoplasty Ziemer Group We develop and produce diagnostic and surgical products for the ophthalmic

More information

The Laser Eye Center s surgeons are sub-specialized in both cornea and refractive surgery, and are among the region s most experienced surgeons.

The Laser Eye Center s surgeons are sub-specialized in both cornea and refractive surgery, and are among the region s most experienced surgeons. Laser Eye Center 1 About Us The Laser Eye Center at AUBMC is a state-of-the-art, continuously updated facility with a mission to provide high-precision refractive surgery to correct visual errors. Staffed

More information

Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday?

Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday? Financial Disclosure Arturo Chayet, MD Tijuana, BC Mexico Perry S. Binder, MS, MD San Diego CA USA I have the following financial interests or relationships to disclose: AMO/IntraLase Corporation - C Acufocus

More information

OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction

OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction American Academy of Ophthalmology Annual Meeting Orlando, Florida, 21-25 October 2011 OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction Weeks Professor

More information

Eye Associates Custom LASIK With IntraLASIK Correction Of Nearsightedness, Farsightedness, and Astigmatism Using IntraLase TM Technology

Eye Associates Custom LASIK With IntraLASIK Correction Of Nearsightedness, Farsightedness, and Astigmatism Using IntraLase TM Technology Eye Associates Custom LASIK With IntraLASIK Correction Of Nearsightedness, Farsightedness, and Astigmatism Using IntraLase TM Technology INDICATIONS AND PROCEDURE This information is being provided to

More information

LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth?

LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth? JKAU: Med. Sci., Vol. 18 No. 4, pp: 29-36 (2011 A.D. / 1432 A.H.) DOI: 10.4197/Med. 18-4.3 LASIK To Improve Visual Acuity in Adult Neglected Refractive Amblyopic Eyes: Is It Worth? Ali M. El-Ghatit, MD,

More information

The Efficacy of Multi-Zone Cross-Cylinder Method for Astigmatism Correction

The Efficacy of Multi-Zone Cross-Cylinder Method for Astigmatism Correction Korean J Ophthalmol Vol. 18:29-34, 2004 The Efficacy of Multi-Zone Cross-Cylinder Method for Astigmatism Correction Seong Joo Shin, MD, Hae Young Lee, MD Department of Ophthalmology, Seoul Adventist Hospital,

More information

Risks and Limitations of LASIK Procedure

Risks and Limitations of LASIK Procedure Drs. Fine, Hoffman & Packer, LLC 1550 Oak Street, Suite #5 Eugene, OR 97401 541-687-2110 From Drs. Fine, Hoffman, & Packer Risks and Limitations of LASIK Procedure Infection, serious injury, or even death,

More information