REPORT ORGANIZATION ABSTRACT I. INTRODUCTION CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE

Size: px
Start display at page:

Download "REPORT ORGANIZATION ABSTRACT I. INTRODUCTION CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE"

Transcription

1 E S T I M A T I O N O F T H E I M A G I N G D O S E F O R T H E C Y B E R K N I F E R O B O T I C R A D I O S U R G E R Y S Y S T E M REPORT ORGANIZATION ABSTRACT I. INTRODUCTION II. III. IV. CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE CALCULATION UNCERTAINTIES V. SUMMARY

2 A B S T R A C T The use of image guided radiation therapy (IGRT) has seen large scale acceptance as a method for accurate pre-treatment patient setup. Techniques such as cone-beam CT (CBCT) and imaging are used in several of the conventional gantry based radiation delivery systems. The Robotic Radiosurgery System uses stereoscopic X-ray imaging not only for patient set up, but also to continually track and correct for any target motion throughout the treatment delivery process. This paper presents an estimation of the dose delivered by the imaging system. This is compared with published data describing the imaging dose associated with gantry based IGRT devices. I. I N T R O D U C T I O N The System consists of a six-degrees-of-freedom 1 robotic manipulator which positions a compact 6 MV linac to accurately treat lesions anywhere in the body. 1,2,3 For target localization, the system uses a pair of amorphous silicon flat panel detectors arranged in a stereoscopic geometry with two ceiling-mounted kilovoltage X-ray sources at 45 degrees to the vertical, and the corresponding, floor-mounted flat panel detectors. The imaging system field of view, at the point where the central axes of these beams intersect (the imaging isocenter), is approximately 15 x 15 cm. In this paper we use the methodology described in the AAPM Task Group 75 Report (TG-75) 4 to obtain the effective dose 5 for kv X-ray imaging as utilized by the System to allow comparison with other image guidance modalities. It should be noted that the imaging system dose calculations presented here apply to all versions of the System, up to and including the recently released VSI System. Figure 1. VSI Robotic Radiosurgery System with stereoscopic in-floor flat panel detectors and ceiling mounted kv X-ray sources. 1 1 The equivalent of three Cartesian axes and three Euler angles are provided by the 6 rotational joints of the robot.

3 I I. C A L C U L A T I O N O F E F F E C T I V E D O S E P E R I M A G E A C Q U I S I T I O N The data in Table 1 show the surface dose per single image (averaged for the two imagers) as measured at the imaging isocenter using optically stimulated luminescent dosimeters (OSLDs), mounted on water equivalent slab phantoms to simulate backscatter. The ceiling mounted X-ray sources can achieve peak voltage of 150 kvp (depending upon imaging frequency) and have approximately 2.5 mm of aluminum inherent filtration. Table 1. Measured surface dose per single System image. kvp ma ms mas Dose (mgy) To compare the dose between different image guidance modalities we use the effective, dose, E (msv), which can be obtained from these absorbed dose measurements using the TG-75 method as follows: E = D*A*F where, D = absorbed dose (mgy) at the center of the field, A = X-ray beam area (cm 2 ) and F is a conversion factor of dosearea product to effective dose. Tables of F values are provided in TG-75 for a variety of anatomical target locations, beam orientations, and X-ray techniques. The X-ray technique used during treatment depends upon the target anatomy and patient size. A limited survey of sites was conducted to establish the range of kv, ma and exposure time settings used for a variety of treatment sites, and from these ranges the median values were used to calculate the typical effective dose per image pair according to the TG-75 method and the dose measurements reported in Table 1. Absorbed dose per mas varies approximately linearly with kvp within the kvp range presented in Table 1, and therefore linear interpolation was used to estimate the dose per mas for kvp settings not included in the measurements. The results of these calculations are shown in Table 2. It should be noted that the maximum available kv was increased to 150 kv only recently, and therefore there is little current user experience with kvp settings above about 125 kv, which might bias the kv settings in Table 2 for deep lesions and larger patients. Table 2. Calculated Effective Dose per image pair E, for typical System treatments. Anatomy kvp mas D (mgy) per image A (cm 2 ) F (msv/mgy cm 2 x10-5 ) E (msv) per image pair Head Chest Pelvis

4 In comparison, the effective doses for other image guidance modalities are listed in Table 3 (with the exception of the final row, these data points are taken from TG-75; the source references are listed for completeness): Table 3. Effective Dose for other image guidance modalities including imaging, diagnostic kv CT scanning and kv cone beam CT scanning. Modality Anatomy Effective Dose (msv) Reference Orthogonal MV Portal Image Pair (assuming 2MU per image) Orthogonal MV Portal Image Pair (assuming 2MU per image) Chest 8.6 [6] Male Pelvis 1.3 [6] Standard CT Head [7] [8][9][10] Standard CT Chest [7] [8][9][10] Standard CT Pelvis [7] [8][9][10] kv CBCT Head 3.0 [11] kv CBCT Chest 8.1 [11] kv CBCT Pelvis 3.0 (low dose mode) 23.0 (standard mode) [12] I I I. C A L C U L A T I O N O F E F F E C T I V E D O S E P E R T R E A T M E N T C O U R S E The calculations presented in Table 4 compare the total effective dose delivered by the X-ray image guidance system during a hypofractionated treatment course with the effective dose delivered by a diagnostic (or planning) CT scan of the same region, and by imaging or kv CBCT image guidance techniques. Unlike other image guidance techniques, the continual intra-fraction tracking and correction provided by the System requires X-ray image pairs to be acquired repeatedly throughout each fraction rather than just once for initial patient set-up. The range in typical numbers of image pairs acquired during treatment for various treatment sites was obtained from the limited survey of users described earlier, and the calculations presented in Table 4 are based on the median of those ranges (the total number of image acquisitions for each technique is stated in the table). For Synchrony treatments this number includes images to build the correlation model. All calculations are based on the data presented in Tables 2 and 3. Table 4. A comparison of effective doses delivered by image guidance methods during an entire hypofractionated treatment course. (Head) 1) Single-fraction intracranial SRS # images E (msv) # images E (msv) # images E (msv) # images E (msv) No data (Chest) 2) Hypofractionated (3 fractions) Lung SBRT # images E (msv) # images E (msv) # images E (msv) # images E (msv) (Pelvis) 3) Hypofractionated (4 fractions) Prostate SBRT # images E (msv) # images E (msv) # images E (msv) # images E (msv) (low dose) 92 (standard) 3

5 Table 5 shows a similar comparison for a conventionally fractionated prostate treatment, in which the image guidance dose resulting from a delivery of Robotic IMRT treatment is compared with that from set-up only IGRT technologies, assuming a 20 minute treatment fraction with 15 image pairs per fraction (with five minute patient set-up), which is approximately an imaging period of 60 s. Daily CBCT is included as an IGRT option in this table, but the relatively high effective dose from such a guidance strategy means that this is unlikely to be a viable option. Therefore, a hybrid IGRT strategy involving CBCT set-up at fractions 1-3 and weekly after that, with used every day 12 is also included for comparison. Table 5. A comparison of effective doses delivered by image guidance methods during an entire conventionally fractionated treatment course. 4) Conventionally Fractionated (N=40) Prostate IMRT CBCT set-up + # images E (msv) # images E (msv) # images E (msv) # images E (msv) # images E (msv) (low dose) 920 (standard) 10 (CBCT) + 40 (MV EPID) 82.8 (low dose) 283 (standard) I V. C A L C U L A T I O N U N C E R T A I N T I E S The uncertainties in these calculations are significant. Firstly, the X-ray technique and imaging frequency is dependent on the target anatomy, patient size and to some extent the patient behavior. While the calculations presented here represent a typical case, the data presented in Tables 1 and 2 enable these estimates to be varied as kv, mas, or imaging frequency are altered. Secondly, the F factors included in TG-75 do not correspond to the exact beam geometry or X-ray spectrum employed in the System. However, effective dose calculations made using these factors have been shown to be in reasonable agreement (within ~20%) of results calculated by Monte Carlo simulation of the actual imaging system. 13 Finally these F factors, taken from the Le Heron data, 14 were generated using Monte Carlo simulation to calculate the dose delivered to organs at risk (OARs) in an anthropomorphic mathematical phantom. The physical doses to each OAR were then multiplied by organ specific tissue weighting factors and summed to give the effective dose, using the data provided in ICRP This method has since been revised in ICRP with some significant changes made to tissue weighting factors, and therefore the TG-75 method does not provide an estimate of the effective dose that is consistent with the latest ICRP guidance. Unfortunately the raw dose to organ data is not provided by Le Heron, so it is not possible to convert. Recent calculations of the imaging system effective doses have shown that using ICRP 103 tissue weighting factors rather than ICRP 60 factors results in no significant change for head imaging, an increase of ~70% for thorax imaging, and a decrease of ~55% for pelvic imaging. 17,18 While this represents a significant uncertainty in the absolute values of effective dose given in sections II and III of this report, the relative comparisons between alternative imaging system doses for each treatment site presented here are likely to remain valid since all of these calculations are based on the same (ICRP 60) tissue weighting factors. V. S U M M A R Y In this paper we have provided data on the imaging dose delivered by the system for different treatment modalities and anatomical sites, by using in house measurements of surface dose to estimate the effective dose using the methodology described in TG-75. This will enable clinicians to make informed decisions regarding the dose associated with kv imaging during typical treatments and compare it with other commonly used modalities such as CBCT and imaging. 4

6 REFERENCES 1. C. Antypas and E. Pantelis, Performance evaluation of a G4 image-guided robotic stereotactic radiosurgery system, Phys. Med. Biol (2008). 2. W. Kilby, C. R. Maurer Jr, C. Amies et al., Platforms for image-guided and adaptive radiation therapy. In: R. D. Timmerman, L. Xing, eds. Image-Guided and Adaptive Radiation Therapy. Philadelphia: Lippincott Williams &Wilkins. pp (2010). 3. W. Kilby, J. R. Dooley, G. Kuduvalli, S. Sayeh, C. R. Maurer Jr, The Robotic Radiosurgery System in 2010, Technology in Cancer Research and Treatment (accepted). 4. M. J. Murphy et al., The management of imaging dose during image-guided radiotherapy: Report of the AAPM Task Group 75, Med. Phys. 34, (2007). 5. W. Jacobi, The concept of effective dose: A proposal for the combination of organ doses, J. Radiat. Environ. Biophys. 12, (1975). 6. S. P. Waddington and A. L. McKensie, Assessment of effective dose from concomitant exposures required in verification of the target volume in radiotherapy, Br. J. Radiol. 77, (2004). 7. M. Galanski, H. D. Nagel, and G. Stamm, Expositions-dosis bei CT-untersuchungen: Ergebnisse einer bundesweiten umgfrage, Fortschr. Rontgenstr. 172, M164 M168 (2000). 8. E. G. Fiberg, Norwegian Radiation Protection Authority, Department of Radiation Protection and Nuclear Safety, Østerås, Norway, (2000). 9. R. Smith-Bindman, J. Lipson, R. Marcus, K-P. Kim, M. Mahesh, R. Gould, A. Berrington de González, D. L. Miglioretti, Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer, Arch. Intern. Med (2009). 10. A. Berrington de González, M. Mahesh, K-P. Kim, M. Bhargavan, R. Lewis, F. Mettler, C. Land, Projected cancer risks from computed tomography scans performed in the United States in 2007, Arch. Intern. Med (2009). 11. M. K. Islam, T. G. Purdie, B. D. Norrlinger, H. Alasti, D. J. Moseley, M. B. Sharpe, J. H. Siewerdsen, and D. A. Jaffray, Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy, Med. Phys. 33, (2006). L R. Potts and M. Oatey, An investigation of the concomitant doses from cone beam CT and CT simulation in radiotherapy WC 2009, IFMBE Proc 25/I, (2009). 13. P. Francescon, Cyberknife: A users point of view, Rad Oncol 73 (sup1), S9-10 (2004). 14. J. C. Le Heron, Estimation of effective dose to the patient during medical x-ray examinations from measurements of the dose-area product, Phys Med Biol 37, (1992). 15. ICRP-60, The International Commission on Radiological Protection, The 1990 recommendations of the international commission on radiological protection, Annals of the ICRP 21 (1-3) (1990). 16. ICRP-103, The International Commission on Radiological Protection, Recommendations of the international commission on radiological protection, Annals of the ICRP 37 (2-4) (2007). 17. P. Francescon, Effective dose delivered by the Cyberknife image guidance system, Focus on Radiosurgery (Accuray Inc) 7, 1-3 (2009). 18. P. Francescon, private communication (2009). 5

7 The information provided in this report is intended for background and educational purposes only as it relates to the System. The information contained herein is not intended, nor should it be construed, as advocating the acquisition or purchase of a System. Accuray Incorporated reserves the right to update or change any specifications without notice. Please direct any questions or comments to the contact information below. Accuray Worldwide Headquarters 1310 Chesapeake Terrace Sunnyvale, CA USA Tel: Toll Free: , ext 4337 Fax: sales@accuray.com Accuray Europe Tour Atlantique 25 e 1 Place de la Pyramide Paris La Défense Cedex France Tel: Fax: Accuray Asia Ltd. Suites , Tower 6 The Gateway, Harbour City 9 Canton Road, T.S.T. Hong Kong Tel: Fax: Accuray Japan K.K. Daini Tekko Building 6F Marunouchi, Chiyoda-ku Tokyo Japan Tel: Fax: Accuray Incorporated. All Rights Reserved. Accuray, the stylized logo,, Synchrony, Xsight, Xchange and RoboCouch are among the trademarks and/or registered trademarks of Accuray Incorporated in the United States and other countries A

02 CyberKnife: Treatment Delivery

02 CyberKnife: Treatment Delivery TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options

More information

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment. DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering

More information

TOMOTHERAPY H SERIES VERSATILE, EFFICIENT AND EFFECTIVE

TOMOTHERAPY H SERIES VERSATILE, EFFICIENT AND EFFECTIVE TOMOTHERAPY H SERIES VERSATILE, EFFICIENT AND EFFECTIVE The TomoTherapy H Series VERSATILE, EFFICIENT AND EFFECTIVE FOR THE RANGE OF RADIATION ONCOLOGY PATIENTS Treat the Cases You Need To, THE WAY YOU

More information

What is the CyberKnife System?

What is the CyberKnife System? CYBERKNIFE Robotic Radiosurgery System PATIENT BROChURE 1 What is the CyberKnife System? have stabilizing frames bolted to their head or limit their breathing during treatment to minimize movement of the

More information

Radiation Protection in Radiotherapy

Radiation Protection in Radiotherapy Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France a-lisbona@nantes.fnclcc.fr Radiation therapy The lecture is oriented to

More information

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT

More information

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

Gamma Knife and Axesse Radiosurgery

Gamma Knife and Axesse Radiosurgery Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery

More information

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

1. Provide clinical training in radiation oncology physics within a structured clinical environment. Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,

More information

CBCT for Prone Breast. Todd Jenkins, MS, DABR Nash Cancer Treatment Center

CBCT for Prone Breast. Todd Jenkins, MS, DABR Nash Cancer Treatment Center CBCT for Prone Breast Todd Jenkins, MS, DABR Nash Cancer Treatment Center Disclosures No outside funding or support Disclosures Techniques likely apply across vendors Prone Breast Technique Rationale

More information

Gated Radiotherapy for Lung Cancer

Gated Radiotherapy for Lung Cancer Gated Radiotherapy for Lung Cancer Steve B. Jiang, Ph.D. Depart Of Radiation Oncology University of California San Diego sbjiang@ucsd.edu radonc.ucsd.edu/research/cart Two Types of Gating Internal gating

More information

IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect Vincent.bossier@iba-group.com. Protect, Enhance and Save Lives

IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect Vincent.bossier@iba-group.com. Protect, Enhance and Save Lives Vincent Bossier System Architect Vincent.bossier@iba-group.com IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT

More information

Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care

Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require

More information

Horizon Blue Cross Blue Shield of New Jersey 2012 Radiation Therapy Payment Rules

Horizon Blue Cross Blue Shield of New Jersey 2012 Radiation Therapy Payment Rules 55875 55876 55876 Description Transperineal placement of needles or catheters into prostate for interstitial radioelement application, with or wout cytosocopy 55920 Placement of interstitial device(s)

More information

Protons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D.

Protons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D. Protons vs. CyberKnife UC SF Protons vs. CyberKnife UC SF Alexander R. Gottschalk, M.D., Ph.D. Associate Professor and Director of the CyberKnife Radiosurgery Program Department of Radiation Oncology University

More information

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients 대한방사선종양학회지 2005;23(3):169~175 Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients Department of Radiation Oncology, College of Medicine, Ewha Womans University

More information

Our Department: structure and organization

Our Department: structure and organization EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled

More information

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology

More information

Cyberknife Information Guide. Prostate Cancer Treatment

Cyberknife Information Guide. Prostate Cancer Treatment Cyberknife Information Guide Prostate Cancer Treatment CYBERKNIFE INFORMATION GUIDE PROSTATE CANCER TREATMENT As a patient recently diagnosed with localized prostate cancer, it is important that you familiarize

More information

Complex 2015 Changes to Radiation Oncology Coding

Complex 2015 Changes to Radiation Oncology Coding Complex 2015 Changes to Radiation Oncology Coding The Centers for Medicare & Medicaid Services (CMS) issued its Final Rule on October 31 outlining the codes it would recognize in calendar year (CY) 2015.

More information

Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose

Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Monitoring Program for Computed Tomography, Digital Radiography, and Nuclear Medicine by Olav Christianson Medical Physics

More information

First Three Years After Project Proton Therapy Facility:

First Three Years After Project Proton Therapy Facility: 39. Project Related Utilization and Financial Information. Proton Therapy Facility: Unit of Service Delivery Projected Utilization FY2017 FY2018 FY2019 Estimated Patients 448 670 808 Average Revenue Per

More information

Innovative RT SBRT. The variables with REQ in superscript are required.

Innovative RT SBRT. The variables with REQ in superscript are required. The variables with REQ in superscript are required. The variables with a are single-select variables; only one answer can be selected. The variables with a are multi-select variables; multiple answers

More information

Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system

Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system

More information

A dosimetry technique for measuring kilovoltage cone-beam CT dose on a linear accelerator using radiotherapy equipment

A dosimetry technique for measuring kilovoltage cone-beam CT dose on a linear accelerator using radiotherapy equipment JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 A dosimetry technique for measuring kilovoltage cone-beam CT dose on a linear accelerator using radiotherapy equipment Daniel Scandurra,

More information

Use of lead shielding for adult chest CT. patient & radiographer experiences

Use of lead shielding for adult chest CT. patient & radiographer experiences Use of lead shielding for adult chest CT Dose measurements and patient & radiographer experiences Previous published work Doshi S, Negus I, Oduko J, Foetal radiation dose from CT pulmonary angiography

More information

QA of intensity-modulated beams using dynamic MLC log files

QA of intensity-modulated beams using dynamic MLC log files 36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,

More information

Implementation Date: April 2015 Clinical Operations

Implementation Date: April 2015 Clinical Operations National Imaging Associates, Inc. Clinical guideline PROSTATE CANCER Original Date: March 2011 Page 1 of 5 Radiation Oncology Last Review Date: March 2015 Guideline Number: NIA_CG_124 Last Revised Date:

More information

Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies

Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies Presented by Susan Granucci, Healthcare Reimbursement Specialist Topics 1. Intensity Modulated Radiation Therapy (IMRT) 2. Tomotherapy

More information

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu

More information

A Patient s Guide to the Calypso System for Breast Cancer Treatment

A Patient s Guide to the Calypso System for Breast Cancer Treatment A Patient s Guide to the Calypso System for Breast Cancer Treatment Contents Introduction How Radiation Works To Treat Cancer Radiation Treatment for Breast Cancer Special Concerns for Treating the Left

More information

Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT

Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Wen Li, Andrew Vassil, Lama Mossolly, Qingyang Shang, Ping Xia Department of Radiation Oncology Why

More information

ADROTERAPIA A MEDAUSTRON

ADROTERAPIA A MEDAUSTRON ADROTERAPIA A MEDAUSTRON Antonio Carlino Qualified Medical Physicist EBG MedAustron GmbH, Wiener Neustadt, Austria Outline Introduction and milestones Building and construction Treatment beam lines and

More information

External dosimetry Dosimetry in new radiotherapeutic techniques

External dosimetry Dosimetry in new radiotherapeutic techniques External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau 44805 Saint-Herblain France a-lisbona@nantes.fnclcc.fr Objective : To describe

More information

SEAAPM Symposium The Practice of Quality Assurance in an Era of Change

SEAAPM Symposium The Practice of Quality Assurance in an Era of Change SEAAPM Symposium The Practice of Quality Assurance in an Era of Change Directors: Dan Bourland and Ingrid Marshall Vendors, breaks, and food next door Thank you to SEAAPM EXCOM, Commercial Partners, Faculty,

More information

Stereotactic Radiosurgery & Stereotactic Body Radiation Therapy - Billing Basics. Presented: June 19, 2013 AAMD Annual Meeting San Antonio, TX

Stereotactic Radiosurgery & Stereotactic Body Radiation Therapy - Billing Basics. Presented: June 19, 2013 AAMD Annual Meeting San Antonio, TX Stereotactic Radiosurgery & Stereotactic Body Radiation Therapy - Billing Basics Presented: June 19, 2013 AAMD Annual Meeting San Antonio, TX Presenters Kelli Weiss, RT(R)(T) Executive Director Tamara

More information

MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015)

MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015) MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015) CODE CODE TYPE Description 0330 REVCODE RADIOLOGY/THERAPEUTIC - GENERAL CLASSIFICATION 0333 REVCODE RADIOLOGY/THERAPEUTIC

More information

Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth

Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth Executive Office of Health and Human s Office of Medicaid www.mass.gov/masshealth May 2015 TO: FROM: RE: Radiation Oncology Centers Participating in Daniel Tsai, Assistant Secretary and Director of (2015

More information

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY B. Wall, (Co-ordinator) 1, D. Hart 1, H. Mol 2, A. Lecluyse 2, A. Aroua 3, P. Trueb 3, J. Griebel 4, E. Nekolla 4, P. Gron 5, H. Waltenburg 5, H. Beauvais-March

More information

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review

More information

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular

More information

Patient Exposure Doses During Diagnostic Radiography

Patient Exposure Doses During Diagnostic Radiography Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University

More information

The Canadian National System for Incident Reporting in Radiation Treatment (NSIR-RT) Taxonomy March 2, 2015 V2

The Canadian National System for Incident Reporting in Radiation Treatment (NSIR-RT) Taxonomy March 2, 2015 V2 The Canadian National System for Incident Reporting in Radiation Treatment (NSIR-RT) Taxonomy March 2, 2015 V2 Taxonomy Data Category Number Description Data Fields and Menu Choices 1. Impact 1.1 Incident

More information

The Challenge of CT Dose Records

The Challenge of CT Dose Records The Challenge of CT Dose Records Kimberly Applegate, MD, MS, FACR Professor of Radiology and Pediatrics Emory University *financial disclosures: -Springer Textbook contracts -AIM advisory board for patient

More information

Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179

Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179 Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179 Jean-Pierre Bissonnette a) Task Group 179, Department of Radiation Physics, Princess Margaret

More information

On Board Cone Beam CT for Treatment Planning in Image Guided Radiotherapy

On Board Cone Beam CT for Treatment Planning in Image Guided Radiotherapy On Board Cone Beam CT for Treatment Planning in Image Guided Radiotherapy By Fouad Abdulaziz Abolaban A thesis submitted to the University of Surrey for the award of the degree of Doctor of Philosophy.

More information

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.

More information

Diagnostic Exposure Tracking in the Medical Record

Diagnostic Exposure Tracking in the Medical Record Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand

More information

Secondary Neutrons in Proton and Ion Therapy

Secondary Neutrons in Proton and Ion Therapy Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles

More information

HITACHI Proton Beam Therapy System

HITACHI Proton Beam Therapy System HITACHI Proton Beam Therapy System Masumi Umezawa, M.S. Hitachi, Ltd., Hitachi Research Laboratory HITACHI Proton Beam Therapy System Contents 1. Overview of HITACHI 2. HITACHI Proton Beam Therapy and

More information

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada

More information

A New Tool to Help Fight Cancer Tomotherapy

A New Tool to Help Fight Cancer Tomotherapy A New Tool to Help Fight Cancer Tomotherapy JEFFREY S. ESHLEMAN, M.D. Lancaster Radiology Associates, Ltd. ABSTRACT Advances in radiation therapy for cancer are reviewed, with emphasis on TomoTherapy Hi

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

Proton Therapy for Prostate Cancer

Proton Therapy for Prostate Cancer Proton Therapy for Prostate Cancer Andrew K. Lee, MD, MPH Director, Proton Therapy Center Associate Professor Department of Radiation Oncology M.D. Anderson Cancer Center Randomized studies showing benefit

More information

Radiosurgery for Benign Spinal Lesions

Radiosurgery for Benign Spinal Lesions Radiosurgery for Benign Spinal Lesions Iris C. Gibbs, M.D. Associate Professor, Radiation Oncology Co-Director Cyberknife Radiosurgery Program Stanford University Sean Sachdev, B.S.; Robert L. Dodd, M.D.,

More information

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical

More information

STABILITY OF MARKERS USED FOR REAL-TIME TUMOR TRACKING AFTER PERCUTANEOUS INTRAPULMONARY PLACEMENT

STABILITY OF MARKERS USED FOR REAL-TIME TUMOR TRACKING AFTER PERCUTANEOUS INTRAPULMONARY PLACEMENT doi:10.1016/j.ijrobp.2010.12.026 Int. J. Radiation Oncology Biol. Phys., Vol. 81, No. 3, pp. e75 e81, 2011 Copyright Ó 2011 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/$ - see front

More information

Accreditation Is Coming

Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt

More information

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of

More information

Proton Therapy System A NEW WAY TO DO WHAT YOU DO BEST FIGHT CANCER

Proton Therapy System A NEW WAY TO DO WHAT YOU DO BEST FIGHT CANCER Proton Therapy System A NEW WAY TO DO WHAT YOU DO BEST FIGHT CANCER Particle Therapy is one of my top initiatives as CEO of. We are leveraging all of our existing and future technologies from image guidance,

More information

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

Acknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group

Acknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group Cone Beam CT guided Radiotherapy Jan-Jakob Sonke Acknowledgements NKI-AVL: Marcel van Herk, Jose Belderbos, Suzanne van Beek, Anja Betgen, Josien de Bois, ~Rianne de Jong, Michel Frenay, Danny Minkema,

More information

There must be an appropriate administrative structure for each residency program.

There must be an appropriate administrative structure for each residency program. Specific Standards of Accreditation for Residency Programs in Radiation Oncology 2015 VERSION 3.0 INTRODUCTION The purpose of this document is to provide program directors and surveyors with an interpretation

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Multi-slice Helical CT Scanning of the Chest

Multi-slice Helical CT Scanning of the Chest Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.

More information

Assessing Radiation Dose: How to Do It Right

Assessing Radiation Dose: How to Do It Right Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine

More information

AK IMRT Würzburg 26./27.03.2009. Roland Kramer, Conmedica GmbH, Schriesheim

AK IMRT Würzburg 26./27.03.2009. Roland Kramer, Conmedica GmbH, Schriesheim AK IMRT Würzburg 26./27.03.2009 Roland Kramer, Conmedica GmbH, Schriesheim 2 AK IMRT Würzburg 26./27.03.2009 Roland Kramer, Conmedica GmbH, Schriesheim OVERVIEW 3 Company Background AlignRT GateCT and

More information

Specifications Varian Treatment Streamlined Treatment Delivery Management Application. Specifications

Specifications Varian Treatment Streamlined Treatment Delivery Management Application. Specifications Specifications Varian Treatment Streamlined Treatment Delivery Management Application i Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment

More information

Key words: treatment planning, quality assurance, 3D treatment planning

Key words: treatment planning, quality assurance, 3D treatment planning American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical

More information

Clinical Trials and Radiation Treatment. Gerard Morton Odette Cancer Centre Sunnybrook Research Institute University of Toronto

Clinical Trials and Radiation Treatment. Gerard Morton Odette Cancer Centre Sunnybrook Research Institute University of Toronto Clinical Trials and Radiation Treatment Gerard Morton Odette Cancer Centre Sunnybrook Research Institute University of Toronto What I will cover.. A little about radiation treatment The clinical trials

More information

2015 ASTRO INVESTOR MEETING. October 20, 2015

2015 ASTRO INVESTOR MEETING. October 20, 2015 2015 ASTRO INVESTOR MEETING October 20, 2015 Agenda 7:30AM Welcome Dow Wilson Our Cancer Care Vision: The Next Big Advances Kolleen Kennedy Particle Therapy: Building Momentum Dow Wilson Q & A 9:00AM Booth

More information

FOR IMMEDIATE RELEASE

FOR IMMEDIATE RELEASE FOR IMMEDIATE RELEASE The PROBEAT-RT Proton Beam Therapy System Incorporating Real-time Tumor-tracking is Approved for Commercial Manufacturing and Sales under Japan s Pharmaceutical Affairs Law Treatments

More information

Gestión global de la dosis en TC. Sistema de registro y gestión

Gestión global de la dosis en TC. Sistema de registro y gestión Gestión global de la dosis en TC. Sistema de registro y gestión IV Jornada de Protección Radiológica Hospitalaria SARH FEA Radiofísica Hospital Virgen de las Nieves. Granada jalmansa.lopez@gmail.com 22

More information

Particle Therapy for Lung Cancer. Bradford Hoppe MD, MPH Assistant Professor University of Florida bhoppe@floridaproton.org

Particle Therapy for Lung Cancer. Bradford Hoppe MD, MPH Assistant Professor University of Florida bhoppe@floridaproton.org Particle Therapy for Lung Cancer Bradford Hoppe MD, MPH Assistant Professor University of Florida bhoppe@floridaproton.org Content Rationale for Particle Therapy in Lung Cancer Proof of Principle Treatment

More information

Challenges in small field MV photon dosimetry

Challenges in small field MV photon dosimetry Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland maria.aspradakis@ksl.ch Why are

More information

RISK OF SECONDARY MALIGNANT NEOPLASMS FROM PROTON THERAPY AND INTENSITY-MODULATED X-RAY THERAPY FOR EARLY-STAGE PROSTATE CANCER

RISK OF SECONDARY MALIGNANT NEOPLASMS FROM PROTON THERAPY AND INTENSITY-MODULATED X-RAY THERAPY FOR EARLY-STAGE PROSTATE CANCER doi:10.1016/j.ijrobp.2009.01.001 Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 2, pp. 616 622, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front

More information

Quality Assurance of accelerators; the technologists responsibility

Quality Assurance of accelerators; the technologists responsibility Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section

More information

RADIOTHERAPY FOR BREAST CANCER: HOW CAN IT BENEFIT FROM ADVANCING TECHNOLOGY?

RADIOTHERAPY FOR BREAST CANCER: HOW CAN IT BENEFIT FROM ADVANCING TECHNOLOGY? RADIOTHERAPY FOR BREAST CANCER: HOW CAN IT BENEFIT FROM ADVANCING TECHNOLOGY? *Tomas Kron, 1,2 Boon Chua 2,3 1. Department of Physical Sciences, Peter MacCallum Cancer Centre, Melbourne, Australia 2. Sir

More information

Radiation Therapy. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements

Radiation Therapy. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements PRIMARY CERTIFICATION DIDACTIC AND CLINICAL COMPETENCY REQUIREMENTS EFFECTIVE JANUARY 2014 Radiation Therapy 1. Introduction Candidates for certification and registration are required to meet the Professional

More information

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning Karin M. Andersson,

More information

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015 INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design

More information

Quality Assurance in Stereotactic. Radiotherapy

Quality Assurance in Stereotactic. Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Clinical Education A comprehensive and specific training program. carry out effective treatments from day one

Clinical Education A comprehensive and specific training program. carry out effective treatments from day one Proton Therapy Clinical Education A comprehensive and specific training program carry out effective treatments from day one Forewarned is forearmed Although over 100,000 patients have been treated in proton

More information

Department of Radiation Oncology

Department of Radiation Oncology Department of Radiation Oncology Welcome to Radiation Oncology at Emory Clinic Every member of Emory Clinic Department of Radiation Oncology strives to provide the highest quality of care for you as our

More information

Dose Tracking System:

Dose Tracking System: Dose Tracking System: A Paradigm Shift in Patient Dose Monitoring Andrew Kuhls-Gilcrist, Ph.D. XRVL Clinical Science Manager Toshiba America Medical Systems, Inc. 2 Dose Tracking System: A Paradigm Shift

More information

Clinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION

Clinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION Clinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION Days: Monday-Friday Times: Full Time Location: One of the participating cancer clinics Program Director: George

More information

THE DOSIMETRIC EFFECTS OF

THE DOSIMETRIC EFFECTS OF THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,

More information

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry

More information

Radiosurgery 8/16/2011. Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago

Radiosurgery 8/16/2011. Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago Radiosurgery The use of radiation as a surgical tool Small volumes of tissues

More information

FAQ About Prostate Cancer Treatment and SpaceOAR System

FAQ About Prostate Cancer Treatment and SpaceOAR System FAQ About Prostate Cancer Treatment and SpaceOAR System P. 4 Prostate Cancer Background SpaceOAR Frequently Asked Questions (FAQ) 1. What is prostate cancer? The vast majority of prostate cancers develop

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information