Electronic apex locator: A comprehensive literature review Part I: Different generations, comparison with other techniques and different usages

Size: px
Start display at page:

Download "Electronic apex locator: A comprehensive literature review Part I: Different generations, comparison with other techniques and different usages"

Transcription

1 Systematic Review Electronic apex locator: A comprehensive literature review Part I: Different generations, comparison with other techniques and different usages Hamid Mosleh, Saber Khazaei 1, Hamid Razavian 2, Armita Vali, Farzad Ziaei Dental Students Research Center, 2 Department of Endodontics, Torabinejad Dental Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, 1 Department of Research, School of Dentistry, Kermanshah University of Medical Sciences, Kermanshah, Iran ABSTRACT Introduction: To compare electronic apex locators (EAL) with others root canal determination techniques and evaluate other usage of this devices. Materials and Methods: Tooth apex, Dental instrument, Odontometry, Electronic medical, and Electronic apex locator were searched as primary identifiers via Medline/PubMed, Cochrane library, and Scopus data base up to 30 July Original articles that fulfilled the inclusion criteria were selected and reviewed. Results: Out of 402 relevant studies, 183 were selected based on the inclusion criteria. In this part, 108 studies are presented. Under the same conditions, no significant s could be seen between different EALs of one generation. The application of EALs can result in lower patient radiation exposure, exact diagnosing of fractures, less perforation, and better retreatment. Conclusions: EALs were more than other techniques in root canal length determination. Key words: Dental instrument, electronic apex locator, electronic medical, odontometry, tooth apex Introduction Ideal pulp treatment is defined as the removal of infected pulp and cleaning, shaping, and disinfecting the root canal system. [1] Subsequently, a three-dimensional filling can be provided. To achieve this goal, an essential stage is the assessment of the correct length of the root. [1,2] Working length is defined as the distance between the coronal/incisal reference point and the area that has been prepared and at which the filled canal should end. [1] Quick Response Code: Access this article online Website: DOI: / Commonly, the minor apical foramen or apical isthmus is considered the end of the area for canal preparation and filling. The minor apical foramen is the border line between the dental pulp and periodontal area, which is approximately mm from the anatomic apex. [1,3] Failure to determine the root canal length can result in both over- and underestimation of the root canal length. Overestimated working length can result in preparation beyond the apical isthmus, which can damage the peri-apical region. [4] Underestimated working length and inadequate debridement can cause unsuccessful treatment and dissatisfaction of both the patient and dentist. [5] Due to the pivotal role of working length determinations in root canal therapy, several methods have been introduced as follows. A: Tactile sensation and using the mean canal length and the application of paper cones are examples of experimental methods that are used by some clinicians due to their simplicity and relative efficiency. [6] These Corresponding Author: Dr. Hamid Razavian, Department of Endodontics, School of Dentistry, Isfahan University of Medical Sciences, Hezar Jerib St, Isfahan , Iran. razavian@dnt.mui.ac.ir 84 Dental Hypotheses Jul-Sep 2014 / Vol 5 Issue 3

2 techniques can be in in some patients, however, due to open apex teeth and apical curvature. [6] B: Radiography is a common method for determining the apical isthmus, which is mm shorter than radiographic apex. [2] The radiographic method has its advantages, such as direct observation of the root canal system and the canal curvature and of the existence of peri-apical lesions, [2] but radiography cannot determine the apical isthmus, because it provides a twodimensional picture of a three-dimensional object. [2,7] In addition, a disadvantage of radiation is that it can be dangerous to both patients and dental staff. C: Owing to the advantages of electronic apex locators (EAL), such as the elimination of radiographic obstacles and EAL s accuracy and convenience, the application of EAL has developed. [3,8] The principal design and development of the early apex locators dates back to Suzuki (1942) [9] whom investigated on dogs and found out that the electrical resistance between the periodontal membrane and the oral mucosa was a constant value. This point was introduced into clinical practice by Sunada [10] (1962) which almost measured the electrical resistance between oral mucosa and periodontal ligament. Over the last decade, different versions of EAL have been released. Table 1 shows and compares the different versions of these devices, based on the functions of each generation. The aim of the present study was to investigate and compare the accuracy of EALs in determining the working length through root canal therapy. Materials and Methods English language only. Systematic reviews, case reports, letter to editors, editorials, and congress abstracts were excluded. The title and abstract of each article were reviewed by three of the authors (HM, AV, and FZ), and articles that fulfilled the inclusion criteria were selected. Data extraction Extraction of data from studies and assessment of validity was independently performed by two authors (HM and AV) and checked by a third author (FZ). In the case of disagreement between evaluators, it reassessed by discussion between two reviewers and a final consensus was agreed on (HR and SKh). Figure 1 provides information on the number of papers identified through the search strategy. Information of the authors, their institutions, and result of primary studies were removed before assessment of the validity. Information on the first author, year of publication, study design, study population and sample size, and the outcome measurements (main results) were extracted. Results Out of 402 articles, 183 studies were reviewed and 108 studies were selected for this part. The studies were categorized as follows. Comparison of different EALs In this part of the present study, 38 articles, consisting of 14 in vitro, [11-24] 9 ex vivo, [25-33] and 15 in vivo studies [34-48] were reviewed. Of the in vitro studies, four articles did not report significant s between various devices [12,13,16,19] [Table 2]. All of the ex vivo studies showed significant s between different devices, except for the study by Comin Chiaramonti et al. [31] and Baginska et al. [32] [Table 2]. Among the Search strategy Electronic searches were performed using tooth apex, dental instrument, odontometry, electronic medical, and electronic apex locator as keywords. Moreover, cross-references were screened to identify further study that probably missed through the search strategy. Electronic databases Electronic searches were performed in Medline/PubMed, Cochrane library, and Scopus data base up to 30 July Four hundred and two articles were found. Inclusion criteria The inclusion criteria were articles, clinical trials in humans and cohort and case-control studies in the Figure 1: Flow chart of selected articles Jul-Sep 2014 / Vol 5 Issue 3 Dental Hypotheses 85

3 Table 1: Different generations of electrical apex and their operation base [86] Generation Operation base Device samples First Measurement of electrical resistance Root canal meter (Onuki medical Co.tokyo, Japan) Endodontic meter(onuki medical Co.tokyo, Japan) Dentometer (Dahlin ectromedicine, Copenhagen, Denmark) Endo Radar (Electronica liarre, Imola, Italy) Second Measurements of electrical impedance Sono-Explorer (Hayashi Dental Supply, Tokyo, Japan) Endo Cater (Yamaura Seisokushu, Tokyo, Japan) Digipex (Mada Equipment Co.Carlstadt, NJ, USA) Exact-A-Pex (Ellman International Hewlett, NY, USA) Formatron IV (Parkell Dental, Farmingdale, NY, USA) Endodontic Meter S II (Onuki Medical Co., Tokyo, Japan) Sono-Explorer Mark II (Hayashi Dental Supply, Tokyo, Japan). Sono-Explorer Mark II Junior (Hayashi Dental Supply, Tokyo, Japan) Fifth Using two different frequencies at the same time in order to measure the or ratio between two currents Using two or more non-simultaneous continuous frequencies in order to measure the or ratio between two currents Measures the capacitance and resistance of the circuit separately Endex/Apit (Osada Electrica Co. Tokyo, Japan) Root ZX (J.Morita, Tokyo, Japan) Neosono Ultima EZ (Satelec Inc, Mount Lourel, Nj, USA) TCM Endo V (Nouvag Ag, Goldach, Switzerland) Apex Pointer (MicroMega, Besanc on, France) Dat Apex (Dentsply Maillerfer, Ballaiques, Switzerland) Just or Justy II (Yoshida Co.Yokyo, Japan) Mark V Plus (Moyco/Union Broach, Bethpage, NY, USA) Apex pointer Endy 5000 (Loser, Leverkusen, Germany) Mini Apex Locator (Sybron Endo, Anaheim, CA, USA) Dentaport ZX (J. Morita MFG Corp., Osaka, Japan) Endo Analyzer Model 8005 (Analytic/Endo, Orange, CA, USA) Apex Finder AFA (Analytic Technologies, Redmond, WA) Mark V Plus (Moyco/Union Broach, Bethpage, New York, USA) Endox (Co. Lysis, Milan, Italy). Endy (Loser, Leverkusen, Germany) Apex Finder (Endo Analyzer 8001; Analytic Technology, Redmond, WA, USA) Foramatron D10 (Parkell Electronic Division, Farmingdale, New York, USA). Bingo1020/Raypex4 (Foroum Engineering Technologies Rishon Lezion, Israel) Raypex4 (VDW, Munich, Germany) Element Dianostic Unit & Apex Locator (Sybron Endo, Anaheim, CA, USA) Neosono MC (Amadent Medical and Dental, Co., Cherry Hill, New Jersey, USA). Propex (Dentsply Maillerfer, Ballaiques, Switzerland) Novapex (Foroum Engineering Technologies Rishon Lezion, Israel) Apex NRG XFR (Medic NRG Ltd, Tel Aviv, Israel) Apex DSP (Septodont, Saint-Maur des Fosse s, Cedex, France) AFA Apex Finder, Model 7005 (Ana-lytic Endodontics, Orange, CA) ipex (NSK Ltd, Tokyo, Japan) Romi Apex D-30 (Romidan LTD, Kiryat-ono, Israel) Propex II (Dentsply Maillerfer, Ballaiques, Switzerland) Top of Form Bottom of Form Apex Locator Joypex 5 (Henan, CBD Neihuan Road, Zhengzhou, China) I-ROOT (E-Magic Finder)(S-Denti SEoul, South Korea) Raypex 5 (VDW, Munich, Germany) in vivo studies, Welk et al. [34] and Arora et al. [47] found a significant between EALs [Table 2]. Comparison of EALs with different working length determination methods Thirty-one studies evaluated different methods of working length determination. Among these studies, eleven compared different EALs and conventional and digital radiography as three different methods of working length determination. [7,35-46,49-59] Among the studies that compared EALs to radiographic method, six studies did not report any significant s, [7,35,36,43,53,54] three studies concluded that radiography method was more, [49,55,59] and the remainder reported greater accuracy with the EAL [Table 3]. Three studies that compared digital radiography to EAL showed that EAL was more [37,54,60] [Table 3]. Shanmugaraj et al. [6] compared three methods of measuring root canal length (apex locator, radiography, tactile sense) and reported that EAL was the most reliable method, but Ounsi et al. [61] showed that EAL and radiography method have 86 Dental Hypotheses Jul-Sep 2014 / Vol 5 Issue 3

4 Table 2: Comparison of different electronic apex locators (EALs) in working length determination Authors Type of study Publication year Sample (n) Types of studied ELA Comin Chiaramonti Ex vivo teeth Bingo 1020 et al. [31] Propex Baginska et al. [32] Ex vivo teeth Raypex 5, Apex D.S.P and Locapex Five Somma et al. [89] In vivo teeth Dentaport ZX, Raypex 5, ProPex II Paludo et al. [48] In vivo root Apex and ipex Canals Jung et al. [15] In vitro teeth Root ZX, I-Root Stober et al. [90] In vivo root Stober et al. [87] In vivo root Miletic et al.[91] In vivo root Raypex 5 Mini Apex Locator Root ZX and ipex Silveira et al. [88] In vivo teeth Root ZX, Novapex devasconcelos Ex vivo teeth Root ZX, RomiApex et al. [28] D-30, and Ipex Guise et al. [17] In vitro teeth Root ZX II, Elements AL, Precision AL Aim of study Main study result Type of generation Comparing the accuracy of EALs in Dentaport ZX, RomiApex A-15 and Raypex 5 D Assuncao et al. [33] Ex vivo teeth Root ZX-II, Novapex, Mini AL Pascon et al. [92] In vivo root DentaPort ZX, Raypex 5 Higa et al. [13] In vitro teeth Justylll, Dentaport, E-Magic Finder Siu et al. [93] In vivo teeth Root ZX II, Apex NRG XFR, Mini Apex Locator Pascon et al. [25] Ex vivo teeth Dentaport ZX, Raypex 5 Elements Diagnostic Unit and A L Ebrahim et al. [16] In vitro teeth Dentaport ZX, ProPex, Foramatron D10, Apex NRG, Apit 7 D Assuncao et al. [18] Invitro teeth Mini AL, Root ZX II Bernardes et al. [12] Invitro teeth Root ZX, Elements Diagnostic Unit and A L, RomiAPEX D-30 There weren t significant s and both devices were Root ZX was more Root ZX II was more Root ZX II was more Justylll was more Fifth Elements Diagnostic Unit and A L was not than others Dentaport ZX, ProPex, Foramatron D10 were more (Continued ) Jul-Sep 2014 / Vol 5 Issue 3 Dental Hypotheses 87

5 Table 2: (Continued) Authors Type of study Publication year Sample (n) Types of studied ELA Wrbas et al. [94] In vivo teeth Root ZX Raypex5 Stavrianos et al. [24] In vivo root Venturi et al. [26] Ex vivo root Dentaport ZX,RayPex 4 Root ZX, Apex Finder Topuz et al. [19] In vitro teeth TCM Endo V Root ZX Ebrahim et al. [27] Ex vivo teeth Root ZX, Foramatron D10, Apex NRG and Apit 7 Plotino et al. [30] Ex vivo teeth Root ZX, Elements Diagnostic Unit and A L, ProPex D Assunco et al. [20] In vitro teeth Novapex Root-ZX Hor et al. [29] Ex vivo teeth Justy II Raypex4 Haffner et al. [95] In vivo teeth Root ZX Endy Justy II EndoxLysis ElAyouti et al. [14] Ex vivo root Raypex 4, Apex pointer, Root ZX Venturi et al. [96] Invivo teeth Apex Finder, Root ZX Aim of study Main study result Type of generation evaluating the accuracy of the apex-locating function of the TCM Endo V and to compare the results to these of the Root ZX to compare the accuracy of the Root- ZX and Novapex electronic apex locators (EALs) in locating the apical foramen Lucena-Martin et al. [11] In vitro teeth Justy II, Root ZX, NeosonoUltima EZ Hoer et al. [46] In vivo teeth Justy II, Endy 5000 Welk et al. [34] In vivo teeth Root ZX Endo Analyzer Model 8005 De Moor et al. [21] In vitro single canal teeth Apex Finder AFA Model 7005, Apex-Finder, Neosono Ultima EZ and Apit 2 Comparing operation of different apex locator devices The accuracy and operator dependency of four electronic canal length measuring devices were compared under a set of specified conditions Root ZX was more TCM Endo V proved to be as reliable as Root ZX but the use of the device to determine the working length was not easy as Root ZX Root ZX and Foramatron D10 were more ProPex was not than others 88 Dental Hypotheses Jul-Sep 2014 / Vol 5 Issue 3 Root-ZX and Novapex are useful and devices for the apical foramen location Raypex4 was more Root ZX was more Root ZX was more Apex Finder AFA Model 7005 was the most (Continued )

6 Table 2: (Continued) Authors Type of study Publication year Sample (n) Weiger et al. [22] In vitro teeth Root ZX Apit Lauper et al. [97] Ex vivo root Arora et al. [47] In vivo root Pallares and Faus [98] Ex vivo root Nahmias et al. [23] In vitro single root teeth Types of studied ELA Apit and Odontometer ENDEX and RCM Mark II Odontometer and Endo Cater Sono-Explorer, C. L. Meter and Neosono-D Aim of study Main study result Type of generation Two apex locators were compared regarding their ability to ly locate the apical constriction in the presence of various canal fluids at different meter readings. Evaluated the accuracy of EAL in length measurement Comparison the accuracy of EALs in canal length measurement in present of different contents(vital pulp, necrotic pulp, pus/ exudates, sodium hypochlorite, and water) Compared the accuracy of EAL in Compared the accuracy of EAL in In the presence of NaOCl, Root ZX provides the most EWL measurements Apit was more Accurate ENDEX was higher accuracy than the RCM Mark II First and Endo Cater was First and higher accuracy than Second the Odontometer All EALs were Second and same accuracy and significantly were more than tactile method. In the study by Subramaniam et al. [62], there were no significant s among conventional radiography, digital radiography, and tactile sense in primary teeth. Janner et al. [63] compared cone beam computed tomography to EAL in root canal length determination and concluded that both two methods were comparable [Table 3]. Use of EAL in root perforations, fractures, and apical foramen widening Ten studies reported other uses of EALs, such as identification of root perforation sites and the location of horizontal and vertical fractures. EALs were only in horizontal fracture diagnosis [64,65] although Topez et al. [66] reported that EALs were in both vertical and horizontal root fractures. Furthermore, Goldberg et al. [67] studied the consistency of EALs in teeth with simulated horizontal root fractures, and they showed that EALs were and consistent. Several studies showed that EALs were able to detect perforation sites [Table 4]. [68-71] There were five studies that evaluated EALs ability to control apical foramen widening with rotary files, and they both concluded that use of EALs in root that were prepared with rotary instruments was not sufficiently to control apical extensions. [72-75] Hoer et al. [76] evaluated the ability of the Justy II and Endy 5000 devices in determining the canal length of 79 teeth (93 ). They showed that these devices can specify the sites of the minor and major apical foramen, but they cannot determine apical constriction with sufficient accuracy. Also, Oishi et al. [77] investigated the accuracy of EALs in determining apical constriction, and they were [Table 4]. Use of EAL in root canal retreatment Six articles were published on this topic. [78-81] Two of them evaluated the accuracy of EALs before and after canal filling and showed that, in most cases, EALs were in root canal retreatment. [78,80] In the study by Aggarwal et al. [80], the accuracy of Root ZX and Protaper devices was evaluated in the retreatment of filled with: 1) gutta-percha+ zinc oxide ogenol sealer; 2) gutta-percha+ AH plus sealer; and 3) Resilon+ Epiphany sealer. Both devices showed high accuracy in the first and second treatments, and no significant s were reported in the presence of different filling materials. Uzun et al. [79] studied the precision of two apex locators (tri auto ZX TCM, locating handpieces endo apex) in root canal retreatment with root-end-cured teeth evaluated. They demonstrated that both devices could be used for determining apical area, but for root canal retreatment, accuracy of 0.0 mm is required, which these devices could not accomplish [Table 4]. Jul-Sep 2014 / Vol 5 Issue 3 Dental Hypotheses 89

7 Table 3: Comparison of electronic apex locators (EALs) and different other techniques in working length determination Authors Type of study Publication Sample (n) year Types of studied ELA Aim of study Orosco et al. [60] In vivo teeth Root ZX Comparison of EAL and radiography Main study result Conventional and EAL were comparable and better than digital of Kishor [99] In vitro teeth Foramatron D10 Comparison of EAL and radiography, EAL are more radiovisography in determination of working length Chougule et al. [58] In vivo primary teeth Dentaport ZX Comparison of EAL and radiography EAL are more Zand et al. [100] In vivo/ teeth Root ZX Comparison of EAL and radiography Ex vivo Jarad et al. [36] In vivo teeth Raypex 5 Comparison of EAL and radiography Parekh and Ex vivo patients Root ZX Comparison of EAL and radiography Radiography and EAL Taluja[101] can be used together Janner et al. [63] In vivo patients Root ZX Compared CBCT method to EAL Two methods were comparable Neena et al. [54] In vivo teeth Comparison of EAL and radiography in determination of working length Singh et al. [57] In vivo teeth Propex II Comparison of EAL and radiography EALs are more Real et al. [51] In vitro root Root ZX Comparison of EAL and radiography EALs are more Just II Elements Diagnostic Patino-Marin In vivo root Root ZX and Comparison of EAL and radiography EALs are more et al. [45] primary teeth ProPex Kqiku et al. [7] Ex vivo teeth Root ZX Comparison of EAL and radiography Vieyra et al. [38] In vivo/in vitro teeth (693 root ) Root ZX, Elements- Diagnostic, Precision AL and Raypex 5 Mancini et al. [44] Ex vivo teeth Endex, Propexll and Root ZX Cianconi et al. [37] Ex vivo teeth Endex ProPex II Root ZX Sharma and In vivo/ Arora [102] Ex vivo Vieyra et al. [46] In vivo teeth Root ZX and Elements- Diagnostic Comparison of EAL and other methods in determination of working length Comparison of EAL and other methods in determination of working length Comparison of EAL and radiography EALs are more EALs are more EALs are more teeth Root ZX Comparison of EAL and radiography EALs are more Comparison of EAL and radiography EALs are more Javidi et al. [40] In vitro teeth Root ZX Comparison of EAL and radiography Radiography and EAL can be used together Kim et al. [43] In vivo teeth Root ZX Evaluating EAL with or without radiography in determination of working length Krajczar et al. [39] In vitro teeth ProPex Comparison of EAL and radiography EAL are more Hassanien et al. [50] In vivo patients Root ZX Comparison of EAL and radiography EAL are more in determination of working Shanmugaraj et al. [6] In vivo/in vitro teeth Foramatron-IV Comparison of EAL and other methods in determination of working length Smadi et al. [51] In vivo root Tri Auto ZX Comparison of EAL and radiography Subramaniam In vitro teeth Formatron D10 Comparison of EAL and other et al. [62] methods in determination of working length EAL are more Use of EAL could reduce need for additional radiography (Continued ) 90 Dental Hypotheses Jul-Sep 2014 / Vol 5 Issue 3

8 Table 3: (Continued) Authors Type of study Publication Sample (n) year Types of studied ELA Aim of study Main study result ElAyouti et al. [41] In vitro teeth Root ZX Comparison of EAL and radiography EAL are more Brunton et al. [42] In vitro teeth - Effect of apex locator on reduction X-ray exposure Use of EAL reduced patients X-ray exposure Martinez-Lozano In vitro teeth Apit EM-S3 Comparison of EAL and et al. [35] radiography in determination of working length Saad et al. [103] In vivo teeth Root ZX Effect of apex locator combining Root ZX and a digital imaging system (RadioVisioGraphy]on reduction X-ray exposure This technique was useful Ounsi et al. [61] In vitro teeth Endex Comparison of EAL and radiography, tactile sense method in determination of working length Himel et al. [55] In vivo root Formatron IV Comparing the accuracy of EALs and radiography method in length measurement Hembrough In vivo maxillary et al. [49] molar teeth Frank et al. [56] In vivo root canal Sono-Explorer Mark III Endex Trope et al. [59] In vivo root Sono-Explorer Mark III Murakami et al. [104] In vivo infected Sono-Explorer Compared the accuracy of EAL and radiography method in length measurement Compared EAL with radiography method in Tactile sense method was in, but other two methods were same EALs weren t EAL is useful only combined with radiography and couldn t replace it EAL was comparable with radiography method Comparison of EAL and radiography EAL wasn t accuracy same as radiography To retrospectively assess the Use of the Sonosuccess of endodontic treatment Explorer that had been guided by audiometric aided successful (electronic) measurement. treatment of infected root EAL was Stavrianos et al. [105] In vivo teeth Raypex 5 Evaluated the accuracy of EAL in Ounsi et al. [106] In vitro single root Root ZX Evaluated the accuracy of EAL in teeth In vivo Vajrabhaya Root ZX Evaluated the accuracy of EAL in et al. [107] Single root teeth Shabahang In vivo root Root ZX Evaluated the accuracy of EAL in et al. [108] Wu et al. [109] In vivo single root Sono-Explorer Evaluated the accuracy of EAL in teeth type Y-III Ricard et al. [110] In vivo teeth RCM Mark II Evaluated the accuracy of EAL in McDonald et al. [111] In vivo teeth Endocater Evaluated the accuracy of EAL in detecting apical constriction Berman et al. [112] In vivo mature Neosono-D Evaluated the accuracy of EAL in and 5 immature root Busch et al. [113] In vitro teeth (46 vital and 26 necrotic) Sono-Explorer Evaluated the accuracy of EAL in Root ZX couldn t detect apical constriction and should only use to detecting major foramen In clinical acceptable range its In clinical acceptable range its In clinical acceptable range its In clinical acceptable range its It was EAL only in mature canal was EAL was in both groups Use of EAL in patients with pacemakers Wilson et al. [82] investigated the operation of the Endo Analyzer Model 8005 in patients with pacemakers and cardioverter/defibrillator devices. They demonstrated that there was no interference between the apex locator and pacemaker function. Discussion Different generations of EALs, with improved functions and greater clinical applications, have entered the market over these past few decades. Our results indicate that there are no significant s between different Jul-Sep 2014 / Vol 5 Issue 3 Dental Hypotheses 91

9 Table 4: Other usage of electronic apex locators (EALs) Authors Type of study Publication year Sample (n) Types of studied ELA Aggarwal et al. [80] In vitro teeth Root ZX and ProPex Stavrianos et al. [114] In vitro teeth Dentaport ZX, RayPex 5, Endo Master and Bingo-1020 Aim of study Application of apex locator devices in root canal retreatment Application of apex locator devices in root canal retreatment Alves et al. [78] Ex vivo teeth Tri Auto ZX Application of apex locator devices in root canal retreatment Goldberg et al. [115] In vitro teeth ProPex, NovApex, and Root ZX Uzun et al. [79] Ex vivo teeth TCM Endo V and Tri Auto ZX Uzun et al. [81] In vitro root TCM Endo V and resected teeth Tri Auto ZX Application of apex locator devices in root canal retreatment Evaluation the accuracy of apex locator device along with rotary files in root canal in retreatment Evaluation the accuracy of Fadel et al. [74] In vivo single root Root ZX II premolar Jakobson et al. [73] In vivo teeth Root ZX II Evaluation of apex locator ability control apical fromen widening with rotary files Main study result Both devices had high accuracy in retreatment All device were reliable, but Dentaport ZX and Endo Master were more In most cases EAL was in retreatment devices had high accuracy in retreatment Devices should be used with caution These devices are apex locator device along with not appropriate for rotary files in root canal length retreating. measurement in retreatment Use of apex locator in Not appropriate controlled canal widening EAL with rotary instruments was not to controlling apical extension Felippe et al. [72] Ex vivo single root Root ZX II Use of apex locator in Not appropriate teeth controlled canal widening Campbell et al. [75] In vitro teeth Tri Auto ZX to examine the apical extent of rotary canal Instrumentation with the automatic apical instrumentation and the ability reverse feature set to maintain apical constriction at 1 consistently with the Tri Auto ZX at different automated settings approximated the apical constriction; however, the constriction was frequently enlarged Goldberg et al. [67] In vitro teeth ProPex NovApex Root ZX Elements AL Topuz et al. [66] In vitro teeth TCM Endo V and Tri Auto ZX Ebrahim et al. [65] In vitro teeth Root ZX, Foramatron D10, Apex NRG al Kadi et al. [116] In vitro teeeth Propex and Raypex-4 Evaluation apex locatordevice operation in diagnosis root fractures All devices are reliable Evaluation apex locator Both devices identified device operation in diagnosis different root fracture in root fractures an acceptable range Evaluation apex locatordevice Device works ly operation in diagnosis root in teeth with horizontal fractures root fractures Evaluation apex locatordevice operation in diagnosis root fractures Device works ly in teeth with horizontal and vertical root fractures Azabal et al. [64] In vitro teeth Justy II Evaluation apex locatordevice Device works ly operation in diagnosis root fractures in teeth with horizontal root fractures Hoer et al. [76] In vitro/ In vivo root Justy II, Endy 5000 Evaluation apex locators ability in determination apical constriction Oishi et al. [77] In vitro teeth ROOT ZX Evaluation apex locators ability in detremination apical constriction Pratten and Mc In vitro 1996 Apit Evaluation apex locators Donald [117] ability in detremination apical constriction Devices cannot determine apical constriction Device can determine apical constriction Device can determine apical constriction (Continued ) 92 Dental Hypotheses Jul-Sep 2014 / Vol 5 Issue 3

10 Table 4: (Continued) Authors Type of study Publication year Sample (n) Types of studied ELA Aim of study Keller et al. [118] In vivo Endocater Evaluated the ability of EAL in detecting apical constriction and cemento dentinal junction Zmener et al. [71] In vitro teeth Tri Auto ZX Detection and measurement of endodontic root perforations using a newly designed apex-locating handpiece Kaufman et al. [70] In vitro teeth with perforation in middle third Fuss et al. [68] In vitro teeth with perforation in midle third Root ZX, Sono Explorer Mark II Junior and Apit III Sono Explorer Mark 2 Junior and Apit 2 Evaluated the ability of EALs in locatin perforation Evaluated the ability of EALs in locatin perforation Main study result Device was not The Tri Auto ZX detected and measured endodontic root perforations within a range of clinically acceptable variations Regardless of the perforation size all EALs were Both device were Hulsmann et al. [69] In vivo teeth Exact-A-Pex Evaluated the ability of EAL in Was control apical bridge formation in the treatment of teeth with incomplete root formation EALs of the same generation. The first generation apex locator was supplied by single frequency of direct current in order to measurement of electrical resistance. Pain and discomfort were often felt with using this type of apex locator. [83] The second generation apex locator known as impedance apex locators was measured opposition to the flow of alternating current or impedance. [84] The disadvantage of this generation is that electro-conductive materials in canal affect on its accuracy. [83] The third generation apex locator (frequency dependent apex locators) was supplied by two frequencies to measure the impedance in the canal. The disadvantage of this generation sensitivity to canal fluid and the machine needs a fully charged battery. [85] The fourth generation apex locator measures the impedance characteristics using more than two frequencies. [3] The disadvantage includes needing to perform in relatively dry or in partially dried. [84] The fifth generation apex locator was developed in 2003 which measure the capacitance and resistance of the circuit separately. [86] Many studies compared the ability of various generations of EALs in determining root canal length. Most of these studies showed that EALs were for canal length measurement, within a clinically acceptable range of ± 0.5. Some studies indicated that the most recent generation of these devices had enhanced accuracy, better patient acceptance and greater ease of use for dentists. [31,87,88] but other studies mentioned that some EALs of the third generation were more than those of the fourth generation. [20,27,28] Although most of the previous studies reported that EALs were more, compared to radiography, some of the studies noted no significant s between the two methods due to small sample sizes. However, a recent randomized, controlled clinical trial study showed no significant s between these two methods. [2] To consider the advantages of conventional radiography, such as the ability to observe the root canal system and the canal curvature directly and to determine the existence of peri-apical lesions, the decision of which method to use should be different in each case. It should be noted that EALs could decrease the patient s radiation exposure. [42,51] There is controversy in the diagnosis of the horizontal and vertical root fractures by EALs. Some studies have reported that EALs have the capacity to diagnose horizontal and lateral root fractures, [66,67] and others studies have indicated that horizontal fractures and perforation sites can be better diagnosed by EALs than vertical fractures. [64,65] Few studies have investigated the ability of apex locators to detect root fractures and perforations. Due to limited information on this subject, a general conclusion could not be achieved. More studies are required on this subject. The present review has some limitations. First, only relevant articles were searched in Medline/PubMed, Cochrane library, and Scopus, which might have restricted the results. Second, our keywords were limited to Tooth apex, Dental instrument, Odontometry, Electronic medical, and Electronic apex locator to focus on EALs. More prospective, randomized clinical Jul-Sep 2014 / Vol 5 Issue 3 Dental Hypotheses 93

11 trials are needed to determine various conditions that affect EALs accuracy. Conclusion The results of the present study showed that EAL is an appropriate technique for root canal length measurements. References 1. Renner D, Grazziotin-Soares R, Gavini G, Barletta FB. Influence of pulp condition on the accuracy of an electronic foramen locator in posterior teeth: An in vivo study. Braz Oral Res 2012;26: Ravanshad S, Adl A, Anvar J. Effect of working length measurement by electronic apex locator or radiography on the adequacy of final working length: A randomized clinical trial. J Endod 2010;36: Duran-Sindreu F, Stober E, Mercade M, Vera J, Garcia M, Bueno R, et al. Comparison of in vivo and in vitro readings when testing the accuracy of the Root ZX apex locator. J Endod 2012;38: Stoll R, Urban-Klein B, Roggendorf MJ, Jablonski-Momeni A, Strauch K, Frankenberger R. Effectiveness of four electronic apex locators to determine distance from the apical foramen. Int Endod J 2010;43: Ozsezer E, Inan U, Aydin U. In vivo evaluation of ProPex electronic apex locator. J Endod 2007;33: Shanmugaraj M, Nivedha R, Mathan R, Balagopal S. Evaluation of working length determination methods: An in vivo / ex vivo study. Indian J Dent Res 2007;18: Kqiku L, Stadtler P. Radiographic versus electronic root canal working length determination. Indian J Dent Res 2011;22: Mello-Moura AC, Moura-Netto C, Araki AT, Guedes-Pinto AC, Mendes FM. Ex vivo performance of five methods for root canal length determination in primary anterior teeth. Int Endod J 2010;43: Suzuki K. Experimental study on iontophoresis. J Jap Stomatol 1942;16: Sunada I. New method for measuring the length of the root canal. J Dent Res 1962;41: Lucena-Martin C, Robles-Gijon V, Ferrer-Luque CM, de Mondelo JM. In vitro evaluation of the accuracy of three electronic apex locators. J Endod 2004;30: Bernardes RA, Duarte MA, Vasconcelos BC, Moraes IG, Bernardineli N, Garcia RB, et al. Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104:e Higa RA, Adorno CG, Ebrahim AK, Suda H. Distance from file tip to the major apical foramen in relation to the numeric meter reading on the display of three different electronic apex locators. Int Endod J 2009;42: ElAyouti A, Kimionis I, Chu AL, Lost C. Determining the apical terminus of root-end resected teeth using three modern apex locators: A comparative ex vivo study. Int Endod J 2005;38: Jung IY, Yoon BH, Lee SJ, Lee SJ. Comparison of the reliability of 0.5 and APEX mark measurements in two frequencybased electronic apex locators. J Endod 2011;37: Ebrahim AK, Wadachi R, Suda H. In vitro evaluation of the accuracy of five different electronic apex locators for determining the working length of endodontically retreated teeth. Aust Endod J 2007;33: Guise GM, Goodell GG, Imamura GM. In vitro comparison of three electronic apex locators. J Endod 2010;36: D Assuncao FL, de Albuquerque DS, Salazar-Silva JR, de Queiroz Ferreira LC, Bezerra PM. The accuracy of root canal measurements using the Mini Apex Locator and Root ZX-II: An evaluation in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104:e Topuz O, Uzun O, Tinaz AC, Sadik B. Accuracy of the apex locating function of TCM Endo V in simulated conditions: A comparison study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;103:e D Assuncao FL, de Albuquerque DS, de Queiroz Ferreira LC. The ability of two apex locators to locate the apical foramen: An in vitro study. J Endod 2006;32: De Moor RJ, Hommez GM, Martens LC, De Boever JG. Accuracy of four electronic apex locators: An in vitro evaluation. Endod Dent Traumatol 1999;15: Weiger R, John C, Geigle H, Lost C. An in vitro comparison of two modern apex locators. J Endod 1999;25: Nahmias Y, Aurelio JA, Gerstein H. An in vitro model for evaluation of electronic root canal length measuring devices. J Endod 1987;13: Stavrianos C, Vladimirov SB, Vangelov LS, Papadopoulos C, Bouzala A. Evaluation of the accuracy of electronic apex locators Dentaport ZX and Ray Pex 4 under clinical conditions. Folia Med (Plovdiv) 2007;49: Pascon EA, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani MA. An ex vivo comparison of working length determination by 3 electronic apex locators. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009;108:e Venturi M, Breschi L. A comparison between two electronic apex locators: An ex vivo investigation. Int Endod J 2007;40: Ebrahim AK, Wadachi R, Suda H. Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters. Aust Dent J 2006;51: de Vasconcelos BC, do Vale TM, de Menezes AS, Pinheiro- Junior EC, Vivacqua-Gomes N, Bernardes RA, et al. An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2010;110:e Hor D, Krusy S, Attin T. Ex vivo comparison of two electronic apex locators with different scales and frequencies. Int Endod J 2005;38: Plotino G, Grande NM, Brigante L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. Int Endod J 2006;39: Comin Chiaramonti L, Menini M, Cavalleri G. A comparison between two fourth generation apex locators. Minerva Stomatol 2012;61: Baginska J, Ewa L, Bartlomiej B, Anna K. An ex vivo comparison of three electronic apex locators. Health Med BA 2012;10: D Assuncao FL, Albuquerque DS, Salazar-Silva JR, Dos Santos VC, Sousa JC. Ex vivo evaluation of the accuracy and coefficient of repeatability of three electronic apex locators using a simple mounting model: A preliminary report. Int Endod J 2010;43: Dental Hypotheses Jul-Sep 2014 / Vol 5 Issue 3

12 34. Welk AR, Baumgartner JC, Marshall JG. An in vivo comparison of two frequency-based electronic apex locators. J Endod 2003;29: Martinez-Lozano MA, Forner-Navarro L, Sanchez-Cortes JL, Llena-Puy C. Methodological considerations in the determination of working length. Int Endod J 2001;34: Jarad FD, Albadri S, Gamble C, Burnside G, Fox K, Ashley JR, et al. Working length determination in general dental practice: A randomised controlled trial. Br Dent J 2011;211: Cianconi L, Angotti V, Felici R, Conte G, Mancini M. Accuracy of three electronic apex locators compared with digital radiography: An ex vivo study. J Endod 2010;36: Vieyra JP, Acosta J. Comparison of working length determination with radiographs and four electronic apex locators. Int Endod J 2011;44: Krajczar K, Marada G, Gyulai G, Toth V. Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root of extracted upper molars. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106: Javidi M, Moradi S, Rashed R, Raziee L. In vitro comparative study of conventional radiography and Root ZX apex locator in determining root canal working length. N Y State Dent J 2009;75: ElAyouti A, Weiger R, Lost C. The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working length. J Endod 2002;28: Brunton PA, Abdeen D, MacFarlane TV. The effect of an apex locator on exposure to radiation during endodontic therapy. J Endod 2002;28: Kim E, Marmo M, Lee CY, Oh NS, Kim IK. An in vivo comparison of working length determination by only root- ZX apex locator versus combining root-zx apex locator with radiographs using a new impression technique. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;105:e Mancini M, Felici R, Conte G, Costantini M, Cianconi L. Accuracy of three electronic apex locators in anterior and posterior teeth: An ex vivo study. J Endod 2011;37: Patino-Marin N, Zavala-Alonso NV, Martinez-Castanon GA, Sanchez-Benavides N, Villanueva-Gordillo M, Loyola- Rodriguez JP, et al. Clinical evaluation of the accuracy of conventional radiography and apex locators in primary teeth. Pediatr Dent 2011;33: Vieyra JP, Acosta J, Mondaca JM. Comparison of working length determination with radiographs and two electronic apex locators. Int Endod J 2010;43: Arora RK, Gulabivala K. An in vivo evaluation of the ENDEX and RCM Mark II electronic apex locators in root with different contents. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1995;79: Paludo L, Souza SL, So MV, Rosa RA, Vier-Pelisser FV, Duarte MA. An in vivo radiographic evaluation of the accuracy of Apex and ipex electronic Apex locators. Braz Dent J 2012;23: Hembrough JH, Weine FS, Pisano JV, Eskoz N. Accuracy of an electronic apex locator: A clinical evaluation in maxillary molars. J Endod 1993;19: Hassanien EE, Hashem A, Chalfin H. Histomorphometric study of the root apex of mandibular premolar teeth: An attempt to correlate working length measured with electronic and radiograph methods to various anatomic positions in the apical portion of the canal. J Endod 2008;34: Smadi L. Comparison between two methods of working length determination and its effect on radiographic extent of root canal filling: A clinical study [ISRCTN ]. BMC Oral Health 2006;6: Fouad AF, Reid LC. Effect of using electronic apex locators on selected endodontic treatment parameters. J Endod 2000;26: Real DG, Davidowicz H, Moura-Netto C, Zenkner Cde L, Pagliarin CM, Barletta FB, et al. Accuracy of working length determination using 3 electronic apex locators and direct digital radiography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;111:e Neena I, Ananthraj A, Praveen P, Karthik V, Rani P. Comparison of digital radiography and apex locator with the conventional method in root length determination of primary teeth. J Indian Soc Pedod Prev Dent 2011;29: Himel VT, Cain C. An evaluation of two electronic apex locators in a dental student clinic. Quintessence Int 1993;24: Frank AL, Torabinejad M. An in vivo evaluation of Endex electronic apex locator. J Endod 1993;19: Singh SV, Nikhil V, Singh AV, Yadav S. An in vivo comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators. Indian J Dent Res 2012;23: Chougule RB, Padmanabhan MY, Mandal MS. A comparative evaluation of root canal techniques in primary teeth. Pediatr Dent 2012;34: Trope M, Rabie G, Tronstad L. Accuracy of an electronic apex locator under controlled clinical conditions. Endod Dent Traumatol 1985;1: Orosco FA, Bernardineli N, Garcia RB, Bramante CM, Duarte MA, Moraes IG. In vivo accuracy of conventional and digital radiographic methods in confirming root canal working length determination by Root ZX. J Appl Oral Sci 2012;20: Ounsi HF, Haddad G. In vitro evaluation of the reliability of the Endex electronic apex locator. J Endod 1998;24: Subramaniam P, Konde S, Mandanna DK. An in vitro comparison of root canal measurement in primary teeth. J Indian Soc Pedod Prev Dent 2005;23: Janner SF, Jeger FB, Lussi A, Bornstein MM. Precision of endodontic working s: A pilot investigation comparing cone-beam computed tomography scanning with standard measurement techniques. J Endod 2011;37: Azabal M, Garcia-Otero D, de la Macorra JC. Accuracy of the Justy II Apex locator in determining working length in simulated horizontal and vertical fractures. Int Endod J 2004;37: Ebrahim AK, Wadachi R, Suda H. Accuracy of three different electronic apex locators in detecting simulated horizontal and vertical root fractures. Aust Endod J 2006;32: Topuz O, Uzun O, Tinaz AC, Bodrumlu E, Gorgul G. Accuracy of two apex-locating handpieces in detecting simulated vertical and horizontal root fractures. J Endod 2008;34: Goldberg F, Frajlich S, Kuttler S, Manzur E, Briseno- Marroquin B. The evaluation of four electronic apex locators in teeth with simulated horizontal oblique root fractures. J Endod 2008;34: Fuss Z, Assooline LS, Kaufman AY. Determination of location of root perforations by electronic apex locators. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996;82: Hulsmann M, Pieper K. Use of an electronic apex locator in the treatment of teeth with incomplete root formation. Endod Dent Traumatol 1989;5: Kaufman AY, Fuss Z, Keila S, Waxenberg S. Reliability of different electronic apex locators to detect root perforations in vitro. Int Endod J 1997;30: Zmener O, Grimberg F, Banegas G, Chiacchio L. Detection and measurement of endodontic root perforations using a newly Jul-Sep 2014 / Vol 5 Issue 3 Dental Hypotheses 95

13 designed apex-locating handpiece. Endod Dent Traumatol 1999;15: Felippe WT, Felippe MC, Reyes Carmona J, Crozoe FC, Alvisi BB. Ex vivo evaluation of the ability of the ROOT ZX II to locate the apical foramen and to control the apical extent of rotary canal instrumentation. Int Endod J 2008;41: Jakobson SJ, Westphalen VP, da Silva Neto UX, Fariniuk LF, Picoli F, Carneiro E. The accuracy in the control of the apical extent of rotary canal instrumentation using Root ZX II and ProTaper instruments: An in vivo study. J Endod 2008;34: Fadel G, Piasecki L, Westphalen VP, Silva Neto UX, Fariniuk LF, Carneiro E. An in vivo evaluation of the auto apical reverse function of the Root ZX II. Int Endod J 2012;45: Campbell D, Friedman S, Nguyen HQ, Kaufman A, Keila S. Apical extent of rotary canal instrumentation with an apexlocating handpiece in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1998;85: Hoer D, Attin T. The accuracy of electronic working length determination. Int Endod J 2004;37: Oishi A, Yoshioka T, Kobayashi C, Suda H. Electronic detection of root canal constrictions. J Endod 2002;28: Alves AM, Felippe MC, Felippe WT, Rocha MJ. Ex vivo evaluation of the capacity of the Tri Auto ZX to locate the apical foramen during root canal retreatment. Int Endod J 2005;38: Uzun O, Topuz O, Tinaz C, Nekoofar MH, Dummer PM. Accuracy of two root canal devices integrated into rotary endodontic motors when removing gutta-percha from root-filled teeth. Int Endod J 2008;41: Aggarwal V, Singla M, Kabi D. An in vitro evaluation of performance of two electronic root canal devices during retreatment of different obturating materials. J Endod 2010;36: Uzun O, Topuz O, Tinaz AC, Alacam T. Apical accuracy of two apex-locating handpieces in root canal retreatments of root-end resected teeth. J Endod 2007;33: Wilson BL, Broberg C, Baumgartner JC, Harris C, Kron J. Safety of electronic apex locators and pulp testers in patients with implanted cardiac pacemakers or cardioverter/defibrillators. J Endod 2006;32: Suchde RV, Talim ST. Electronic ohmmeter. An electronic device for the determination of the root canal length. Oral Surg Oral Med Oral Pathol 1977;43: Muthu MS, Sivakumar N. Accuracy of electronic apex locator in length determination in the presence of different irrigants: An in vitro study. J Indian Soc Pedod Prev Dent 2006;24: Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod 1994;20: Gordon MP, Chandler NP. Electronic apex locators. Int Endod J 2004;37: Stober EK, Duran-Sindreu F, Mercade M, Vera J, Bueno R, Roig M. An evaluation of root ZX and ipex apex locators: An in vivo study. J Endod 2011;37: Silveira LF, Petry FV, Martos J, Neto JB. In vivo comparison of the accuracy of two electronic apex locators. Aust Endod J 2011;37: Somma F, Castagnola R, Lajolo C, Paterno Holtzman L, Marigo L. In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II. Int Endod J 2012;45: Stober EK, de Ribot J, Mercade M, Vera J, Bueno R, Roig M, et al. Evaluation of the Raypex 5 and the Mini Apex Locator: An in vivo study. J Endod 2011;37: Miletic V, Beljic-Ivanovic K, Ivanovic V. Clinical reproducibility of three electronic apex locators. Int Endod J 2011;44: Pascon EA, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani MA. An in vivo comparison of working length determination of two frequency-based electronic apex locators. Int Endod J 2009;42: Siu C, Marshall JG, Baumgartner JC. An in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files. J Endod 2009;35: Wrbas KT, Ziegler AA, Altenburger MJ, Schirrmeister JF. In vivo comparison of working length determination with two electronic apex locators. Int Endod J 2007;40: Haffner C, Folwaczny M, Galler K, Hickel R. Accuracy of electronic apex locators in comparison to actual length an in vivo study. J Dent 2005;33: Venturi M, Breschi L. A comparison between two electronic apex locators: An in vivo investigation. Int Endod J 2005;38: Lauper R, Lutz F, Barbakow F. An in vivo comparison of gradient and absolute impedance electronic apex locators, J Endod 1996;22: Pallares A, Faus V. An in vivo comparative study of two apex locators. J Endod 1994;20: Kishor KM. Comparison of working length determination using apex locator, conventional radiography and radiovisiography: An in vitro study. J Contemp Dent Pract 2012;13: Zand V, Mokhtari H, Lotfi M, Reyhani MF, Sohrab A, Tehranchi P, et al. Accuracy of working length determination with root ZX apex locator and radiography: An in vivo and ex vivo study. Afr J Biotechnol 2013;10: Parekh V, Taluja C. Comparative study of periapical radiographic techniques with apex locator for endodontic working length estimation: An ex vivo study. J Contemp Dent Pract 2011;12: Sharma MC, Arora V. Determination of working length of root canal. Med J Armed Forces India 2010;66: Saad AY, al-nazhan S. Radiation dose reduction during endodontic therapy: A new technique combining an apex locator (Root ZX) and a digital imaging system (RadioVisioGraphy). J Endod 2000;26: Murakami M, Inoue S, Inoue N. Clinical evaluation of audiometric control root canal treatment: A retrospective case study. Quintessence Int 2002;33: Stavrianos C, Vasiliadis L, Stavrianou I, Kafas P. Clinical evaluation of the accuracy of an electronic tooth apex locator. J Med Sci 2007;7: Ounsi HF, Naaman A. In vitro evaluation of the reliability of the Root ZX electronic apex locator. Int Endod J 1999;32: Vajrabhaya L, Tepmongkol P. Accuracy of apex locator. Endod Dent Traumatol 1997;13: Shabahang S, Goon WW, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod 1996;22: Wu YN, Shi JN, Huang LZ, Xu YY. Variables affecting electronic root canal measurement. Int Endod J 1992;25: Ricard O, Roux D, Bourdeau L, Woda A. Clinical evaluation of the accuracy of the Evident RCM Mark II Apex Locator. J Endod 1991;17: McDonald NJ, Hovland EJ. An evaluation of the Apex Locator Endocater. J Endod 1990;16: Berman LH, Fleischman SB. Evaluation of the accuracy of the Neosono-D electronic apex locator. J Endod 1984;10: Busch LR, Chiat LR, Goldstein LG, Held SA, Rosenberg PA. Determination of the accuracy of the Sono-Explorer for establishing endodontic measurement control. J Endod 1976;2: Dental Hypotheses Jul-Sep 2014 / Vol 5 Issue 3

14 114. Stavrianos CH, Vladimirov ST, Vasiliadis L, Stavrianou I, Panayotov I, Pamporakis P. In vitro evaluation of the precision of four different electronic apex locators in determining the working length of teeth after removing root canal obturation materials. Res J Med Sci 2008;2: Goldberg F, Marroquín BB, Frajlich S, Dreyer C. In vitro evaluation of the ability of three apex locators to determine the working length during retreatment. J Endod 2005; 31: al Kadi H, Sykes LM, Vally Z. Accuracy of the Raypex-4 and Propex apex locators in detecting horizontal and vertical root fractures: An in vitro study. SADJ 2006;61: Pratten DH, McDonald NJ. Comparison of radiographic and electronic working lengths. J Endod 1996;22: Keller ME, Brown CE Jr, Newton CW. A clinical evaluation of the Endocater an electronic apex locator. J Endod 1991;17: Cite this article as: Mosleh H, Khazaei S, Razavian H, Vali A, Ziaei F. Electronic apex locator: A comprehensive literature review - Part I: Different generations, comparison with other techniques and different usages. Dent Hypotheses 2014;5: Source of Support: Nil. Conflict of Interest: None declared. Jul-Sep 2014 / Vol 5 Issue 3 Dental Hypotheses 97

Review. Electronic Apex Locators A Review. Aqeel Khalil Ebrahim, Reiko Wadachi and Hideaki Suda

Review. Electronic Apex Locators A Review. Aqeel Khalil Ebrahim, Reiko Wadachi and Hideaki Suda J Med Dent Sci 2007; 54: 125 136 Review Electronic Apex Locators A Review Aqeel Khalil Ebrahim, Reiko Wadachi and Hideaki Suda Pulp Biology and Endodontics, Department of Restorative Sciences, Graduate

More information

Clinical Efficacy of Three Different Electronic Apex Locators in Comparison with Radiographic Working Length Determination

Clinical Efficacy of Three Different Electronic Apex Locators in Comparison with Radiographic Working Length Determination Research Article Clinical Efficacy of Three Different Electronic Apex Locators in Comparison with Radiographic Working Length Determination Ayca Yilmaz, (contact author) aycasimsek@gmail.com Selçuk Gokyay

More information

Root canal treatment by means of electronic conductometry without radiographic transoperative verification: a report of five cases

Root canal treatment by means of electronic conductometry without radiographic transoperative verification: a report of five cases original article Root canal treatment by means of electronic conductometry without radiographic transoperative verification: a report of five cases Norberto J. BROON 1 Alvaro CRUZ 2 Claudia A. PALAFOX-SÁNCHEZ

More information

Accuracy of Different Methods of Working Length Determination in Endodontics

Accuracy of Different Methods of Working Length Determination in Endodontics IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861. Volume 12, Issue 4 (v.- Dec. 2013), PP 25-38 Accuracy of Different Methods of Working Length Determination

More information

Electronic apex locators

Electronic apex locators REVIEW Electronic apex locators M. P. J. Gordon & N. P. Chandler Department of Oral Rehabilitation, School of Dentistry, University of Otago, Dunedin, New Zealand Abstract Gordon MPJ, Chandler NP. Electronic

More information

Effect of using electronic apex locator on the radiographic extent of root canal. filling and the exposure to radiation during endodontic therapy.

Effect of using electronic apex locator on the radiographic extent of root canal. filling and the exposure to radiation during endodontic therapy. Effect of using electronic apex locator on the radiographic extent of root canal filling and the exposure to radiation during endodontic therapy. L. Smadi BDS, Mdent Sci, FDS RCS Ed Assistant Professor,

More information

Electronic Apex Locators: An Evidence-Based Study

Electronic Apex Locators: An Evidence-Based Study Electronic Apex Locators: An Evidence-Based Study M. Dagenais, S. Hooper, B. Kong, S. Prusky, J. Yip, and H. Zohoor A b s t r a c t The purpose of this evidence-based review of the literature was to determine

More information

The Benefits of Using Arap Locator in Root Canal Treatment

The Benefits of Using Arap Locator in Root Canal Treatment UNSURPASSED ACCURACY UNRIVALED RELIABILITY 1 1. Data on file. Unsurpassed accuracy 1 with convenient, easy-to-read display Fully automatic multifrequency measurements Compact design with small footprint

More information

Electronic Apex Locators

Electronic Apex Locators Electronic Apex Locators This material is provided for information and education purposes only. No doctor/patient relationship has been established by the use of this site. No diagnosis or treatment is

More information

Comparative analysis of two electronic apex locators in working length determination at endodontic therapy an in vitro study

Comparative analysis of two electronic apex locators in working length determination at endodontic therapy an in vitro study ISSN: Printed version: 1806-7727 Electronic version: 1984-5685 RSBO. 2011 Jan-Mar;8(1):25-9 Original Research Article Comparative analysis of two electronic apex locators in working length determination

More information

14 Electronic Apex Locators and

14 Electronic Apex Locators and 14 Electronic Apex Locators and Conventional Radiograph in Working Length Measurement Gevik Malkhassian, Andres Plazas, and Yosef Nahmias Introduction The success of endodontic treatment is highly dependent

More information

Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex

Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex doi: 10.1111/j.1365-2591.2006.01095.x Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex G. Plotino, N. M. Grande, L. Brigante, B. Lesti &

More information

Evaluation of electrical impedance ratio measurements in accuracy of electronic apex locators

Evaluation of electrical impedance ratio measurements in accuracy of electronic apex locators Research article ISSN 2234-7658 (print) / ISSN 2234-7666 (online) Evaluation of electrical impedance ratio measurements in accuracy of electronic apex locators Pil-Jong Kim, Hong- Gee Kim, Byeong- Hoon

More information

Tooth anatomy risk factors influencing root canal working length accessibility

Tooth anatomy risk factors influencing root canal working length accessibility Int J Oral Sci (2011) 3: 135-140. www.ijos.org.cn ORIGINAL ARTICLE Tooth anatomy risk factors influencing root canal working length accessibility Lu Tang 1,2, Tuo-qi Sun 1,2, Xiao-jie Gao 1, Xue-dong Zhou

More information

Accuracy of electronic apex locators to detect root canal perforations with inserted metallic posts: an ex vivo study

Accuracy of electronic apex locators to detect root canal perforations with inserted metallic posts: an ex vivo study Marroquín et al. Head & Face Medicine (2014) 10:57 DOI 10.1186/s13005-014-0057-2 HEAD & FACE MEDICINE RESEARCH Accuracy of electronic apex locators to detect root canal perforations with inserted metallic

More information

Clinical Accuracy of Ipex Apex Locator for Measurement of Root Canal Length of Primary Molars

Clinical Accuracy of Ipex Apex Locator for Measurement of Root Canal Length of Primary Molars Original Article Clinical Accuracy of Ipex Apex Locator for Measurement of Root Canal Length of Primary Molars M. Karami Nogorani 1, M. Zare Jahromi 2, Z. Dehghan 3, R. Talaei 4. 1 Associate Professor,

More information

Oftentimes, as implant surgeons, we are

Oftentimes, as implant surgeons, we are CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the

More information

A comparative assessment of the accuracy of electronic apex locator (Root ZX) in the presence of commonly used irrigating solutions

A comparative assessment of the accuracy of electronic apex locator (Root ZX) in the presence of commonly used irrigating solutions Journal section: New technologies in Dentistry Publication Types: Research doi:10.4317/jced.51230 http://dx.doi.org/10.4317/jced.51230 (Root ZX) in the presence of commonly used irrigating solutions Osama

More information

Removal of a fractured instrument with a new extractor: clinical cases

Removal of a fractured instrument with a new extractor: clinical cases Clinical Removal of a fractured instrument with a new extractor: clinical cases Dominique Martin 1 Introduction The fracture of a root canal instrument during endodontic treatment is a surgical accident

More information

Jamia Millia Islamia: Performa for CV of Faculty/ Staff Members

Jamia Millia Islamia: Performa for CV of Faculty/ Staff Members Curriculum Vitae Brief Profile: 1-2 paragraphs (not exceeding 500 words) 1. Name: Vivek Aggarwal 2. Designation: Assistant Professor 3. Office Address: Department of Conservative Dentistry & Endodontics,

More information

Dentaport ZX Reliable Apex Location Perfect Endodontics Curing with High-Powered Parallel Precision

Dentaport ZX Reliable Apex Location Perfect Endodontics Curing with High-Powered Parallel Precision Dentaport ZX Reliable Apex Location Perfect Endodontics Curing with High-Powered Parallel Precision Thinking ahead. Focused on life. Thinking ahead. Focused on life. 4 Key product features include the

More information

Location of the apical foramen and its relationship with foraminal file size

Location of the apical foramen and its relationship with foraminal file size original article Location of the apical foramen and its relationship with foraminal file size Ronaldo Araújo Souza, DDS, MSc, PhD 1 José Antônio Poli de Figueiredo, DDS, MSc, PhD 2 Suely Colombo, DDS 1

More information

Instruction manual. Safe and efficient NiTi rotary system. Fulfilling the biological requirement for successful endodontics

Instruction manual. Safe and efficient NiTi rotary system. Fulfilling the biological requirement for successful endodontics Instruction manual Safe and efficient NiTi rotary system Fulfilling the biological requirement for successful endodontics FKG REF. - 99.7AA.50.04B.AN - 06/2010 BioRaCe kit 1 Educational CD 1 Endo Stand

More information

Victims Compensation Claim Status of All Pending Claims and Claims Decided Within the Last Three Years

Victims Compensation Claim Status of All Pending Claims and Claims Decided Within the Last Three Years Claim#:021914-174 Initials: J.T. Last4SSN: 6996 DOB: 5/3/1970 Crime Date: 4/30/2013 Status: Claim is currently under review. Decision expected within 7 days Claim#:041715-334 Initials: M.S. Last4SSN: 2957

More information

Endodontic treatment with MTA apical plugs: a case report

Endodontic treatment with MTA apical plugs: a case report 325 Journal of Oral Science, Vol. 49, No. 4, 325-329, 2007 Case Report Endodontic treatment with MTA apical plugs: a case report Pari Ghaziani 1), Navid Aghasizadeh 2) and Mahshid Sheikh-Nezami 1) 1) Department

More information

BioRace NiTi system: biologically desirable apical sizes safely and efficiently

BioRace NiTi system: biologically desirable apical sizes safely and efficiently BioRace NiTi system: biologically desirable apical sizes safely and efficiently Dr Martin Trope and Dr Gilberto Debelian explore the BioRace NiTi system, designed to achieve an apical preparation that

More information

Removal of a silver cone by using clinical microscope and ultrasound: Case report

Removal of a silver cone by using clinical microscope and ultrasound: Case report original article Removal of a silver cone by using clinical microscope and ultrasound: Case report Maíra do Prado 1 Marcos Cesar Pimenta de Araújo 2 Heloísa Carla Gusman 3 abstract Introduction: The retreatment

More information

C-shaped root canal system in mandibular second molars in a Chinese population evaluated by cone-beam computed tomography

C-shaped root canal system in mandibular second molars in a Chinese population evaluated by cone-beam computed tomography doi:10.1111/j.1365-2591.2011.01896.x C-shaped root canal system in mandibular second molars in a Chinese population evaluated by cone-beam computed tomography Q. Zheng 1, L. Zhang 1, X. Zhou 2, Q. Wang

More information

We make a point of ensuring safety.

We make a point of ensuring safety. CAD/CAM SYSTEMS INSTRUMENTES HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS INSTRUMENTS FOR EFFICIENT ROOT CANAL TREATMENT We make a point of ensuring safety. T h e D e n t a l C o m p a n y ENDODONTIC

More information

Analysis of forces developed during root canal filling by different operators

Analysis of forces developed during root canal filling by different operators original article Analysis of forces developed during root canal filling by different operators Maria Rosa Felix de Sousa G. GUIMARÃES, DDS, MSc 1 Henner Alberto GOMIDE, DDS, MSc, PhD 2 Maria Antonieta

More information

What Parents Should Know about the Safety of Dental Radiology.

What Parents Should Know about the Safety of Dental Radiology. What Parents Should Know about the Safety of Dental Radiology. There are many different types of x-ray images (pictures) that can be taken of children in the dental office to assist in diagnosis. These

More information

Key words maxillary sinus, posterior teeth roots. Introduction. Tikrit Journal for Dental Sciences 1(2012)81-88

Key words maxillary sinus, posterior teeth roots. Introduction. Tikrit Journal for Dental Sciences 1(2012)81-88 Tikrit Journal for Dental Sciences 1(2012)81-88 The Relation of Maxillary Posterior Teeth Roots to the Maxillary Sinus Floor Using Panoramic and Computed Tomography Imaging in a Sample of Kurdish People

More information

RICHMOND CROWN - FOR RESTORATION OF BADLY MUTILATED POSTERIOR TEETH - A CASE REPORT

RICHMOND CROWN - FOR RESTORATION OF BADLY MUTILATED POSTERIOR TEETH - A CASE REPORT RICHMOND CROWN - FOR RESTORATION OF BADLY MUTILATED POSTERIOR TEETH - A CASE REPORT 1. Dr. Yadav Chakravarthy, MDS* Professor and Head of The Department. Department of Conservative Dentistry and Endodontics

More information

Use of electronic apex locator in the impossibility of radiographic visualization of the root apex: report of two cases

Use of electronic apex locator in the impossibility of radiographic visualization of the root apex: report of two cases ISSN: Electronic version: 1984-5685 RSBO. 2013 Oct-Dec;10(4):402-6 Case Report Article Use of electronic apex locator in the impossibility of radiographic visualization of the root apex: report of two

More information

Endodontic treatment outcome is related to treatment expectations of the operator

Endodontic treatment outcome is related to treatment expectations of the operator Endodontic Treatment Outcome: Survey of Oral Health Care Professionals Ilan Rotstein, DDS,* Robert Salehrabi, DDS,* and Jane L. Forrest, BSDH, MS, EdD Abstract This study assessed the opinion of oral health

More information

Root ZX II Endodontic Predictability Apex Locator with Low Speed Handpiece. Thinking ahead. Focused on life.

Root ZX II Endodontic Predictability Apex Locator with Low Speed Handpiece. Thinking ahead. Focused on life. Root ZX II Endodontic Predictability Apex Locator with Low Speed Handpiece Thinking ahead. Focused on life. Root ZX II Key Features Endodontic Predictability As the world s best selling apex locator, Root

More information

A new cone beam computerized tomography system for use in endodontic surgery

A new cone beam computerized tomography system for use in endodontic surgery doi:10.1111/j.1365-2591.2006.01198.x CASE REPORT A new cone beam computerized tomography system for use in endodontic surgery T. Tsurumachi 1 & K. Honda 2 1 Department of Endodontics, Division of Advanced

More information

Diagnosis, Treatment and Prognosis of Cracked Teeth Iowa Dental Association Annual Session May 6, 2011 Dr. Bruce C. Justman

Diagnosis, Treatment and Prognosis of Cracked Teeth Iowa Dental Association Annual Session May 6, 2011 Dr. Bruce C. Justman Diagnosis, Treatment and Prognosis of Cracked Teeth Iowa Dental Association Annual Session May 6, 2011 Dr. Bruce C. Justman Longitudinal Fractures Types of Longitudinal Fractures Craze Lines Cusp Fracture

More information

EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE SYSTEM

EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE SYSTEM www.fob.usp.br/jaos or www.scielo.br/jaos J Appl Oral Sci. 2008;16(5):350-4 EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE

More information

Veraviewepocs 3D R100 & F40

Veraviewepocs 3D R100 & F40 Veraviewepocs 3D R100 & F40 Innovative 3D Reuleaux Full Arch FOV Thinking ahead. Focused on life. Veraviewepocs 3D R100 A New Frontier in X-ray Diagnostics Veraviewepocs 3D R100 has changed the shape of

More information

High-tech solutions for a better quality of life

High-tech solutions for a better quality of life Follow-up report High-tech solutions for a better quality of life IDS 2015: Innovative new products and enhancements by Morita For almost 100 years now, the family-owned traditional Japanese company Morita

More information

Non-surgical management of a large periapical lesion using a simple aspiration technique: a case report

Non-surgical management of a large periapical lesion using a simple aspiration technique: a case report doi:10.1111/j.1365-2591.2010.01719.x CASE REPORT Non-surgical management of a large periapical lesion using a simple aspiration technique: a case report M. Fernandes & I. De Ataide Goa Dental College &

More information

Develop a specialist who is capable of correlation of basic sciences and clinical sciences, and challenge the requirements for certification.

Develop a specialist who is capable of correlation of basic sciences and clinical sciences, and challenge the requirements for certification. Course Specification Faculty : Dentistry Department : Endodontics Program Specification: Diploma Degree A-Basic Information 1-Programme Title: Diploma in Endodontics 2-Departments (s): Endodontics 4-Coordinator:

More information

Tough Choice for Young First Permanent molars: To Do Pulp Treatment or to Extract?

Tough Choice for Young First Permanent molars: To Do Pulp Treatment or to Extract? Tough Choice for Young First Permanent molars: To Do Pulp Treatment or to Extract? Jung-Wei Chen, DDS, MS, PhD Professor and Program Director Advanced Education Program in Pediatric Dentistry Loma Linda

More information

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y. 11. The Effect of Abutment Tooth Connection with Extracoronal Attachment using the Three Dimensional Finite Element Method - Part 2. The Construction of Finite Element Model from CT Data - Ando A., Nakamura

More information

Analysis of pulp prognosis in 603 permanent teeth with uncomplicated crown fracture with or without luxation

Analysis of pulp prognosis in 603 permanent teeth with uncomplicated crown fracture with or without luxation Dental Traumatology 2014; doi: 10.1111/edt.12099 Analysis of pulp prognosis in 603 permanent teeth with uncomplicated crown fracture with or without luxation Chao Wang, Man Qin, Yue Guan Department of

More information

Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose!

Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Dale A. Miles BA, DDS, MS, FRCD(C) Diplomate, American Board of Oral and Maxillofacial Radiology It used to be that panoramic images

More information

Veraviewepocs 3D R100 & F40

Veraviewepocs 3D R100 & F40 Veraviewepocs 3D R100 & F40 Innovative 3D Reuleaux Full Arch FOV Thinking ahead. Focused on life. Veraviewepocs 3D R100 A New Frontier in X-ray Diagnostics Veraviewepocs 3D R100 has changed the shape of

More information

ENDODONTOLOGY INTRODUCTION. Original Research ABSTRACT

ENDODONTOLOGY INTRODUCTION. Original Research ABSTRACT Original Research Radiovisiography and ultrasound real time imaging in the differential diagnosis of periapical lesions of endodontic origin A comparative in vivo study SHILPA H. BHANDI * SUBHASH T.S.

More information

In vitro evaluation of shape changes in curved artificial root canals prepared with two rotary systems

In vitro evaluation of shape changes in curved artificial root canals prepared with two rotary systems original article In vitro evaluation of shape changes in curved artificial root canals prepared with two rotary systems Benito André Silveira Miranzi, MSc, PhD 1 Almir José Silveira Miranzi, DDS 2 Luis

More information

4-1-2005. Dental Clinical Criteria and Documentation Requirements

4-1-2005. Dental Clinical Criteria and Documentation Requirements 4-1-2005 Dental Clinical Criteria and Documentation Requirements Table of Contents Dental Clinical Criteria Cast Restorations and Veneer Procedures... Pages 1-3 Crown Repair... Page 3 Endodontic Procedures...

More information

An In Vivo Study of Working Length Determination with a New Apex Locator

An In Vivo Study of Working Length Determination with a New Apex Locator Brazilian Dental Journal (2014) 25(1): 17-21 http://dx.doi.org/10.1590/0103-6440201302326 An In Vivo Study of Working Length Determination with a New Apex Locator Key Fabiano Souza Pereira 1, Pedro Gregol

More information

The efficacy of gutta-percha removal using ProFiles

The efficacy of gutta-percha removal using ProFiles Blackwell Science, Ltd The efficacy of gutta-percha removal using ProFiles J. J. Ferreira, J. S. Rhodes & T. R. Pitt Ford Department of Conservative Dentistry, Guy s, King s and St. Thomas Dental Institute,

More information

Postendodontic Tooth Restoration - Part I: The Aim and the Plan of. the procedure.

Postendodontic Tooth Restoration - Part I: The Aim and the Plan of. the procedure. Postendodontic Tooth Restoration - Part I: The Aim and the Plan of the Procedure Sanja egoviê 1 Nada GaliÊ 1 Ana Davanzo 2 Boæidar PaveliÊ 1 1 Department of Dental Pathology School of Dental Medicine University

More information

INTRAOSSEOUS ANESTHESIA

INTRAOSSEOUS ANESTHESIA ENDODONTICS: Colleagues for Excellence Winter 2009 Bonus Material F INTRAOSSEOUS ANESTHESIA Intraosseous Anesthesia With the Stabident and X Tip Systems The intraosseous injection allows placement of a

More information

Perforations are unfortunate complications that occur in the course of endodontic

Perforations are unfortunate complications that occur in the course of endodontic Management of Perforations: Four Cases from Two Private Practices with Medium- to Long-term Recalls Marga Ree, DDS, MSc,* and Richard Schwartz, DDS Abstract Introduction: Perforation repair is a fairly

More information

Micro-CT evaluation of residual material in canals filled with Activ GP or GuttaFlow following removal with NiTi instruments

Micro-CT evaluation of residual material in canals filled with Activ GP or GuttaFlow following removal with NiTi instruments doi:10.1111/j.1365-2591.2009.01659.x Micro-CT evaluation of residual material in canals filled with Activ GP or GuttaFlow following removal with NiTi instruments M. J. Roggendorf 1, M. Legner 2, J. Ebert

More information

Users Group Meeting 2011 Northeastern Society of Orthodontists 89th Annual Meeting November 11-14, 2010 Hilton Montréal Bonaventure Montréal, Quebec, Canada 14 Lecture Schedule Saturday continued w Doctor

More information

Maxillary sinusitis of odontogenic origin: cone-beam volumetric computerized tomography aided diagnosis

Maxillary sinusitis of odontogenic origin: cone-beam volumetric computerized tomography aided diagnosis Maxillary sinusitis of odontogenic origin: cone-beam volumetric computerized tomography aided diagnosis Umadevi P. Nair, DMD, MDS, a and Madhu K. Nair, DMD, MS, PhD, b Gainesville, Florida UNIVERSITY OF

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

Root Canal Retreatment: 2. Practical Solutions T.R. PITT FORD AND J.S. RHODES

Root Canal Retreatment: 2. Practical Solutions T.R. PITT FORD AND J.S. RHODES E N D O D O N T I CE SN D O D O N T I C S Root Canal Retreatment: 2. Practical Solutions T.R. PITT FORD AND J.S. RHODES Abstract: Root canal retreatment is often the preferred method of treating a tooth

More information

Lee et al. Maxillofacial Plastic and Reconstructive Surgery (2015) 37:30 DOI 10.1186/s40902-015-0030-4

Lee et al. Maxillofacial Plastic and Reconstructive Surgery (2015) 37:30 DOI 10.1186/s40902-015-0030-4 Lee et al. Maxillofacial Plastic and Reconstructive Surgery (2015) 37:30 DOI 10.1186/s40902-015-0030-4 RESEARCH Open Access Assessment of the proximity between the mandibular third molar and inferior alveolar

More information

Supervisors: Dr. Farhan Raza Khan

Supervisors: Dr. Farhan Raza Khan 1 Presenter: Dr. Sana Ehsen Supervisors: Dr. Farhan Raza Khan 2 A dental implant (also known as an endosseous implant or fixture) is a surgical component that interfaces with the bone of the jaw to support

More information

Residency Competency and Proficiency Statements

Residency Competency and Proficiency Statements Residency Competency and Proficiency Statements 1. REQUEST AND RESPOND TO REQUESTS FOR CONSULTATIONS Identify needs and make referrals to appropriate health care providers for the treatment of physiologic,

More information

Endodontic Treatment Failure Consecutive to Unsystematic Radiographic Examination

Endodontic Treatment Failure Consecutive to Unsystematic Radiographic Examination Endodontic Treatment Failure Consecutive to Unsystematic Radiographic Examination Oscar Alonso-Ezpeleta DDS 1, Jenifer Martín-González DDS 2, Milagros Martín-Jiménez DDS 2, Juan J Segura-Egea MD, PhD,

More information

1 The Single Tooth Implant. The Ultimate Aesthetic Challenge

1 The Single Tooth Implant. The Ultimate Aesthetic Challenge 1 The Single Tooth Implant The Ultimate Aesthetic Challenge by Daniel G. Pompa, D.D.S. 2 Before starting any Maxillary Anterior Single Implant, or any case in the esthetic zone: TAKE A PHOTO OF YOUR PATIENT

More information

Friday 29 th April 2016

Friday 29 th April 2016 8.00 9.00 Registration and coffee/bread 9.00 9.45 1. Status of Dental Traumatology worldwide. Is prevention realistic? Lars Andersson Presently almost all countries have published data on the dental trauma

More information

Influence of cone beam computed tomography on dens invaginatus treatment planning

Influence of cone beam computed tomography on dens invaginatus treatment planning case report Influence of cone beam computed tomography on dens invaginatus treatment planning Daniel de Almeida DECURCIO, DDS, MSc, PhD 1 Julio Almeida SILVA, DDS, MSc, PhD 1 Rafael de Almeida DECURCIO,

More information

AAE and AAOMR Joint Position Statement

AAE and AAOMR Joint Position Statement INTRODUCTION AAE and AAOMR Joint Position Statement The following statement was prepared by the Special Committee to Revise the Joint American Association of Endodontists/American Academy of Oral and Maxillofacial

More information

Periodontal Screening and Recording: Early Detection of Periodontal Diseases

Periodontal Screening and Recording: Early Detection of Periodontal Diseases Periodontal Screening and Recording: Early Detection of Periodontal Diseases Tanya Villalpando Mitchell, RDH, MS Continuing Education Units: 1 hour Online Course: www.dentalcare.com/en-us/dental-education/continuing-education/ce53/ce53.aspx

More information

Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers

Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers Dubravka KnezoviÊ-ZlatariÊ Asja»elebiÊ Biserka LaziÊ Department of Prosthodontics School of Dental Medicine University

More information

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S.

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S. Bitewing Radiography B.E. DIXON B.D.S., M.Sc., D.P.D.S. Main Indications Detection of Dental Caries Monitoring progression of caries Assessment of existing restorations Assessment of Periodontal status

More information

Revascularization of Immature Permanent Teeth With Apical Periodontitis: New Treatment Protocol?

Revascularization of Immature Permanent Teeth With Apical Periodontitis: New Treatment Protocol? JOURNAL OF ENDODONTICS Printed in U.S.A. Copyright 2004 by The American Association of Endodontists VOL. 30, NO. 4, APRIL 2004 CLINICAL RESEARCH Revascularization of Immature Permanent Teeth With Apical

More information

TRI Product NewsFlash. December 2015

TRI Product NewsFlash. December 2015 TRI Product NewsFlash December 2015 Study Overview 2015 Dear Partners Year in, year out, we are screening all major scientific journals to ensure that our TRI Performance Concept still reflects the latest

More information

Endodontics. Colleagues for Excellence. Access Opening and Canal Location

Endodontics. Colleagues for Excellence. Access Opening and Canal Location Endodontics Colleagues for Excellence Spring 2010 Access Opening and Canal Location Published for the Dental Professional Community by the American Association of Endodontists Cover artwork: Rusty Jones,

More information

Use of ultrasound Doppler to determine tooth vitality in a discolored tooth after traumatic injury: its prospects and limitations

Use of ultrasound Doppler to determine tooth vitality in a discolored tooth after traumatic injury: its prospects and limitations Case report ISSN 2234-7658 (print) / ISSN 2234-7666 (online) http://dx.doi.org/10.5395/rde.2014.39.1.68 Use of ultrasound Doppler to determine tooth vitality in a discolored tooth after traumatic injury:

More information

USI Master Policy Information

USI Master Policy Information Policy I.D. Gender A.B. (#327) M 86 United of Omaha $ 1,000,000.00 A.B. (#430) - (#436) M 86 Metlife $ 2,000,000.00 A.G. #1 (#371), (#610), (#624) M Conseco Life $ 3,125,000.00 10/Apr/10 A.G. #2 (#380),

More information

Tooth avulsion-a Dental emergency in children: A Review

Tooth avulsion-a Dental emergency in children: A Review American Journal of Advances in Medical Science www.arnaca.com eissn: 2347-2766 Review Article Tooth avulsion-a Dental emergency in children: A Review Debapriya Pradhan 1, Prasant MC 2, Varsha H Tambe

More information

Endo Easy Efficient Sonic Powered Endo Irrigation

Endo Easy Efficient Sonic Powered Endo Irrigation Endo Easy Efficient Sonic Powered Endo Irrigation www.vdw-dental.com Naturally fascinating: gentle, yet powerful In our goal to provide easy and safe products and systems for endodontics which fulfil

More information

Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation

Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation Clinical Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation Kazuto Makigusa 1 Abstract Histological analysis of the biological width surrounding primate

More information

Ridge Reconstruction for Implant Placement

Ridge Reconstruction for Implant Placement Volume 1, No. 5 July/August 2009 The Journal of Implant & Advanced Clinical Dentistry Ridge Reconstruction for Implant Placement 2 Hours of CE Credit Oral Implications of Cancer Chemotherapy Immediate

More information

The main goal of endodontic therapy is to eradicate all vital and necrotic tissue, microorganisms,

The main goal of endodontic therapy is to eradicate all vital and necrotic tissue, microorganisms, Efficacy and Safety of Various Active Irrigation Devices When Used with Either Positive or Negative Pressure: An In Vitro Study Augusto Malentacca, MD, DDS,* Umberto Uccioli, DDS, Dario Zangari, DDS,*

More information

Dr. Little received his doctorate degree in dentistry from UT Health at San Antonio Dental

Dr. Little received his doctorate degree in dentistry from UT Health at San Antonio Dental Implant Solutions for the Implant Patient: Diagnosis and Treatment Planning for Predictable Results David Little, DDS 6961 U.S. Highway 87 East San Antonio, TX 78263 Phone: (210)648-4411 Fax: (210) 648-6498

More information

Epidemiologic survey of traumatic dental injuries in children seen at the Federal University of Rio de Janeiro, Brazil

Epidemiologic survey of traumatic dental injuries in children seen at the Federal University of Rio de Janeiro, Brazil Pediatric Dentistry Pediatric Dentistry Epidemiologic survey of traumatic dental injuries in children seen at the Federal University of Rio de Janeiro, Brazil Marina Alvine de Jesus (a) Lívia Azeredo A.

More information

Humana Health Plans of Florida. Important:

Humana Health Plans of Florida. Important: Humana Health Plans of Florida Important: Dental discount membership in Florida is determined by viewing the member s ID card and verifying that the Humana Logo and Medicare name is listed with an effective

More information

DATING YOUR GUILD 1952-1960

DATING YOUR GUILD 1952-1960 DATING YOUR GUILD 1952-1960 YEAR APPROXIMATE LAST SERIAL NUMBER PRODUCED 1953 1000-1500 1954 1500-2200 1955 2200-3000 1956 3000-4000 1957 4000-5700 1958 5700-8300 1959 12035 1960-1969 This chart displays

More information

The WaveOne single-file reciprocating system

The WaveOne single-file reciprocating system I clinical technique _ WaveOne The WaveOne single-file reciprocating system Authors_ Dr Julian Webber, UK; Drs Pierre Machtou & Wilhelm Pertot, France; Drs Sergio Kuttler, Clifford Ruddle & John West,

More information

Mani Silk: A new and novel means of predictable canal shaping

Mani Silk: A new and novel means of predictable canal shaping technique_ instrumentation I Mani Silk: A new and novel means of predictable canal shaping Author_Rich Mounce, DDS _Introduction Irrespective of how a root canal is shaped, the goals of canal shaping remain

More information

Taylor Dental Assisting School Course Description

Taylor Dental Assisting School Course Description Taylor Dental Assisting School Course Description Entry Level Dental Assisting The Entry Level Dental Assisting Course is divided into Twenty Six (26) modules of four hours each. Total time is One Hundred

More information

Periapical radiography

Periapical radiography 8 Periapical radiography Periapical radiography describes intraoral techniques designed to show individual teeth and the tissues around the apices. Each film usually shows two to four teeth and provides

More information

How To Train To Be A Dentist

How To Train To Be A Dentist Bilaga 1. Sökstrategier Abbreviations * Wildcard indicating a variable number of characters (including none) /AE adverse effects (MeSH Subheading) /CL classification (MeSH Subheading) /CO complications

More information

Dental Extractions, Antibiotics and Curettage First, Do no Harm

Dental Extractions, Antibiotics and Curettage First, Do no Harm Global Journal of Medical research: J Dentistry and Otolaryngology Volume 14 Issue 1 Version 1.0 Year 2014 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc.

More information

The Treatment of Traumatic Dental Injuries

The Treatment of Traumatic Dental Injuries The Recommended Guidelines of the American Association of Endodontists for The Treatment of Traumatic Dental Injuries 2013 American Association of Endodontists Revised 9/13 The Recommended Guidelines of

More information

Multidisciplinary Approach to Delayed Treatment of Traumatic Teeth Injuries Involving Extrusive Luxation, Avulsion and Crown Fracture

Multidisciplinary Approach to Delayed Treatment of Traumatic Teeth Injuries Involving Extrusive Luxation, Avulsion and Crown Fracture Ó Operative Dentistry, 2014, 39-6, 566-571 Clinical Technique/Case Report Multidisciplinary Approach to Delayed Treatment of Traumatic Teeth Injuries Involving Extrusive Luxation, Avulsion and Crown Fracture

More information

EndoActivator from Tulsa Dental

EndoActivator from Tulsa Dental EndoActivator from Tulsa Dental Irrigation-EndoActivator http://endoactivator.com/video3.html EndoActivator System did not enhance the removal of smear layer as compared with conventional Max-I-Probe irrigation

More information

Endodontics. Colleagues for Excellence. in Endodontics. Published for the Dental Professional Community by the American Association of Endodontists

Endodontics. Colleagues for Excellence. in Endodontics. Published for the Dental Professional Community by the American Association of Endodontists Endodontics Colleagues for Excellence Winter 2011 Summer 2011 Root Cone Canal Beam-Computed Irrigants and Tomography Disinfectants in Endodontics Published for the Dental Professional Community by the

More information

This paper summarizes current trends in the

This paper summarizes current trends in the Clinical Diagnosis of Dental Caries: A North American Perspective Stephen F. Rosenstiel, B.D.S., M.S.D. Abstract: This paper summarizes current trends in the clinical diagnosis of occlusal caries in response

More information

Teeth and Dental Implants: When to save, and when to extract.

Teeth and Dental Implants: When to save, and when to extract. Teeth and Dental Implants: When to save, and when to extract. One of the most difficult decisions a restorative dentist has to make is when to refer a patient for extraction and placement of dental implants.

More information

How To Treat An Altered Sensation After Endodontic Treatment

How To Treat An Altered Sensation After Endodontic Treatment Vol. 112 No. 1 July 2011 ENDODONTOLOGY Editor: Larz S.W. Spångberg Medico-legal aspects of altered sensation following endodontic treatment: a retrospective case series Navot Givol, DMD, a Eyal Rosen,

More information

Future Trends in Airline Pricing, Yield. March 13, 2013

Future Trends in Airline Pricing, Yield. March 13, 2013 Future Trends in Airline Pricing, Yield Management, &AncillaryFees March 13, 2013 THE OPPORTUNITY IS NOW FOR CORPORATE TRAVEL MANAGEMENT BUT FIRST: YOU HAVE TO KNOCK DOWN BARRIERS! but it won t hurt much!

More information