1 Forensic Applications of the Determination of Benzodiazepines in Blood Samples by Microcolumn Cleanup and High-Performance Liquid Chromatography with Reductive Mode Electrochemical Detection J.B.F. Lloyd and D.A. Parry Home Office Forensic Science Laboratory, Gooch Street North, Birmingham B5 6QQ, United Kingdom l Abstract I Recently described microcolumn cleanup and highperformance liquid chromatography with electrochemical reductive mode detection have facilitated the determination of benzodiazepines in contaminated and degraded blood samples. Examples from an extensive application of the procedure to British forensic science casework are given here. Introduction Recently introduced procedures based on high-performance liquid chromatography (HPLC) with electrochemical reductive mode detection and on microcolumn cleanup enable the determination of benzodiazepines and their metabolites in small samples of degraded and contaminated bloods, which are typical of forensic science work (l). Depending on the actual benzodiazepine, detection limits and recovery values are in the respective ranges of ng / ml and , and selectivities greatly exceed those of the HPLC techniques usually applied to this type of sample. Examples drawn from eighteen months' experience with these procedures are given in this report. Experimental The procedures and the usual validating data are given in detail elsewhere (1). Briefly, centrifugally filtered blood samples (usually, 100/zL), diluted with 250 #L of w/v sodium octyl sulfate in aqueous 0.1M disodium hydrogen phosphate to suppress protein binding, were extracted with 5 mg of loose Porapak-T (Millipore-Waters). The extractant was transferred to a microcolumn prepacked with an equal amount of clean Porapak-T. From the column the benzodiazepines were eluted in 60 #L water/acetonitrile (ca. 1:2 v/v, depending on the activity of the adsorbent). The eluate was chromatographed on ODS-Hypersil (Shandon), 3/~m, 150 x 4.5 mm, at 40~ in deoxygenated methanol/l-propanol/aqueous phosphate (0.02M, ph 6), 100:8:80 v/v/v (1). Detection was at a pendent mercury drop electrode (EG&G PARC; model 310, modified (2)), which was held, usually, at V vs. Ag/AgCI. The injection volume of the deoxygenated microcolumn eluate was 10 #L. Coefficients of the within-day variation of the retention times were in the range ~ questioned peaks were identified with reference to standards included in accompanying chromatograms. The variation in retention times apparent in the figures reflects the variation seen in 18 months' work with several columns (each prepared from the same batch of packing material). The eluent composition was varied slightly to accommodate changes in column selectivity with age (1). For work within the quoted detection limits or to reduce the setup time, the mercury electrode was preceded by a coulometric detector (Environmental Sciences Associates, model 5010). The porous carbon working electrodes were held at 0 V (proprietary reference electrode). This arrangement removed the more readily reduced traces of impurities in the effluent stream and allowed oxidative detection of other compounds. We strongly emphasize that, along with some other straightforward precautions (l), the most efficient and practicable way to deoxygenate an HPLC eluent for electrochemical detection is to maintain the eluent reservoir under a slow reflux while the chromatograph is in use. Quantitative analyses were calibrated against spiked hemolyzed human blood. The internal standard was either a benzodiazepine selected to occupy a vacant position in the chromatograms or 6-nitroquinoline when the use of a benzodiazepine was impracticable, for example, in the examination of previously uncharacterized samples or a collection of samples in which a considerable variety of benzodiazepines and metabolites was present. Results and Discussion To summarize the results and to place them in perspective, details from 18 months' casework are given in Table I. The quoted concentrations of parent drug include the equivalent amount of the common metabolites, which are also detected. Benzodiazepines were found in 82 of the 118 samples that were analyzed. The total of the cases column, 143, exceeds the number of samples because more than one benzodiazepine had been ingested in some cases. Most of the samples (73~ were taken post-mortem; the others were mainly from subjects involved Reproduction (photocopying) of editorial content of this journal is prohibited without publisher's permission. 163

2 in motoring offenses. Except for temazepam the two sets of data do not significantly differ, and all have been tabulated together, therefore. The means of the respective temazepam concentrations are 2878 ng/ml (N = 26) and 438 ng/ml (N = 9). Two of the postmortem results exceed 10 #g/ml. Alcohol and other drugs were also involved in these deaths. In most instances the tabulated results exceed the range expected from therapeutic plasma levels (3) because of the influence of overdose cases. The differences are particularly marked if account of the hemolyzed state of the samples is taken. The conventionally quoted plasma concentrations include benzodiazepines extensively bound to serum albumin (4) together with any remaining unbound drug. The liberation of the erythrocyte contents, which contains negligible amounts of bound drug, lowers the measured concentrations of hemolyzed samples. The techniques were designed to give distinguishable peaks on a single chromatogram of all the benzodiazepines and their metabolites important to British casework. A chromatogram of these compounds, together with the 6-nitroquinoline used as an internal standard, is shown in Figure 1. The compounds are listed in the figure caption. Common to all of these benzodiazepines, and to the benzodiazepines in general with the exception of clobazam (unencountered in any casework), is an azomethine group which, occasionally with some others, is responsible for the reductive response. Straightforward modification of the chromatography conditions permits application to the benzodiazepines apart from those specifically considered here (1). In contrast to clinical samples, the blood samples encountered in forensic science are often severely degraded, and in addition to the benzodiazepines, their metabolites, and their degradation products, exogenous contaminants may be present. An example from a road traffic case is shown in Figure 2 (chromatogram A), where there are six peaks assigned to the benzodiazepines and metabolites with concentrations varying from 45 ng/ml (oxazepam) to 398 ng/ml (diazepam). The chromatogram includes a peak thought to be due to an uncharacterized metabolite (seen only in such cases), an internal standard peak, and a peak due to contamination from the particular type of septum fitted to the sample container. This last peak often Qccurs but does not interfere with any of the benzodiazepines. The other relatively weak and early emerging peaks occur in various amounts from blood samples generally. A negative chromatogram from a sample taken post-mortem is also shown in Figure 2 (chromatogram B). As well as an examination of the variation of the HPLC response with the detector electrode potential (1), the identifications may be confirmed by direct mass spectrometric examination of the microcolumn eluates (5), or the presence of benzodiazepines generally may be detected by a broadly specific radioimmunoassay (RIA) screening procedure (6). A comparison between the R1A and the HPLC results from a variety of cases positive by both techniques (21 samples) is shown in double logarithmic coordinates in Figure 3, for which the HPLC results have been converted to the diazepam units of the RIA by means.of cross-reactivity factors. Although the RIA gives only a semiquantitative indication of the presence of a drug (6), which is shown by the substantial scatter from a one-to-one relationship (broken line) in Figure 3, there is a clearly evident correlation between the two sets (r = 0.865, N = 21). This is particularly so in cases involving diazepam (r = 0.985, N = 6), with which the RIA is calibrated for screening purposes. With the inclusion of negative results from either technique, 28 comparisons were made. There was only one instance in which HPLC revealed no benzodiazepine in an RIA-positive sample (70 ng/ml diazepam equivalents). Possibly the RIA result was due to residual uncharacterized metabolites, because for most benzodiazepines the techniques are comparable in detection limits. In the original comparison (6) the applied HPLC (ultraviolet absorbance detection) and gas chromatography (electron capture and alkali flame ionization detection) failed to confirm 9 f na Table I. Summary of Results of Analyses for Benzodiazepines in 118 Casework Blood Samples Concentration in blood Parent (nglml)* Therapeutic range in bonzodlazeplne Cases Mean Range plasma (nglml)*" Chlordiazepoxide Clonazepam Oiazepam Flurazepam Lorazepam Nitrazepam xazepam Temazepam Triazolam None found 36 9 Includes metabolites expressed as parent compound *9 Based on ref t t o 9 rain. Figure 1. Chromatogram of a standard mixture: 1 = 7=aminonitrazepam; 2 = 7-acetamidonitrazepam; 3 = 6-nitroquinoline; 4 = demoxepam; 5 = nitrazepam; 6 = triazolam; 7 = Iorazepam; 8 = oxazepam; 9 = Ioprazolam; 10 = temazepam; 11 = Iormetazepam; 12 = chlordiazepoxide; 13 = desmethyldiazepam; 14 = diazepam. Each peak is due to a 10-ng amount, except for peak 3 (5 ng). 164

3 Journal of Analytical Toxicology, Vol 13, May/June 1989 the presence of benzodiazepines in 19 of 55 RIA-positive samples. A direct comparison of!;he performance of the reductive and the UV detection techniques in the HPLC of 14 samples has demonstrated that the UV technique may often be vitiated by irrelevant sample components, whereas the reductive technique is unaffected (l). Examples of a positive HPLC result in the face of a negative or weak response in the RIA have occurred only in nitrazepam cases. In these the major chromatographic peak was usually due to 7-aminonitrazepam, which exhibits negligible cross-reactivity in the RIA. In Table I the concentrations given as equivalent to nitrazepam are largely due to the amino compound. For these 19 cases the mean nitrazepam and aminonitrazepam levels were 57 and 704 ng/ml respectively, the nitrazepam figure mainly being due to two high values, 241 and 512 ng/ml. The latter was from a sample taken post-mortem, and the other was from a defendant in a road traffic case. Usually little or no nitrazepam was present. Examples are given in Figure 4. Chromatogram A shows the presence of the aminonitrazepam (525 ng/ml) and the complete absence of nitrazepam. The RIA result was negative, but there was independent evidence of the ingestion of nitrazepam. Chromatogram B, which exhibits an offscale peak of the aminonitrazepam (1500 ng/ml), is from a similar case. Chromatogram C is from a case in which both the aminonitrazepam and nitrazepam were found (460 and 512 ng/ml respectively); the corresponding R1A response was augmented considerably by the diazepam and the desmethyldiazepam also present (5 and 116 ng/ml respectively; this part of the chromatogram is not included in C). The effect of a less cathodic potential on the peaks of the last-mentioned sample is illustrated in D. The result is characteristic of the peak assignments (1). From clinical work, the major metabolites of nitrazepam are the 7-amino and 7-acetamido derivatives, which occur in blood at levels roughly comparable to that of any remaining nitrazepam (7). That only the amino compound was consistently present in the samples of interest here is apparently due to reactions occurring post-mortem or during storage. This was shown in some brief experiments conducted with a variety of drugfree blood samples to which nitrazepam and its metabolites were individually added. It was found that at room temperatures over several days nitrazepam was lost and replaced in various amounts with the 7-amino derivative. The effect was suppressed in samples containing fluoride preservative or stored at 4 ~ The 7-acetamido compound was similarly lost, but the 7-amino compound was stable under the conditions. Our previous paper shows that the determination of 7-aminonitrazepam by HPLC t na 7 A J, O 10 Figure 2. Casework example of a positive blood sample from a defendant in a road traffic case (A) and of a negative blood sample taken post-mortem (8) 1 = 7-aminonitrazepam; 2 = uncharacterized metabolite; 3 = 6-nitro- min. B r I I I I I I I I I I 0 10 quinoline (internal standard); 4 = nitrazepam; 5 = oxazepam; 6 = temazepam; 7 = septum component; 8 = desmethyldiazepam; 9 = diazepam 165

4 ) 100 lop 1P N J" L // /J L J~ J,J JJ f I 1 10./ D" T," I. N RIA ng ml "I 9 0"" D J T j," M / I s "N T D 0 T D, I I Figure 3. Comparison of HPLC and RIA casework results. The HPLC results have been converted to the diazepam equivalents of the RIA. Each point indicates the component responsible for the main RIA crossreactivity. D = diazepam; L = Iorazepam; M = desmethyldiazepam; N = nitrazepam; 0 = oxazepam; T = temazepam. The broken line indicates the response required by a one-to-,~ne relationship; correlation coefficients are given in the text. T with UV detection is subject to severe interference by components of degraded blood samples (1). This, coupled with the present results and with the insensitivity of the RIA screening to the aminonitrazepam, explains the low detection rate sometimes seen in earlier nitrazepam casework. Apparently the circumstances of temazepam cases in forensic science also give rise to differences from the clinical results, which indicated that the desmethyl metabolite (oxazepam) should appear only in small amounts (8). Thus, in 35 temazepam cases (Table I), none of which showed evidence of separate oxazepam ingestion or that the temazepam was the product of diazepam metabolism, the level of oxazepam, expressed as temazepam equivalents relative to the remaining temazepam, varied over the range < ~ Chromatograms exemplifying extremes of the range are given in Figure 5. In these cases the respective concentrations of temazepam and oxazepam were 760 and 430 ng/ml (chromatogram A) and 5.7/zg/mL and 50 ng/ml (chromatogram B). To date, no common factor within cases at either extreme has emerged. Both compounds were shown to be stable in the samples received. Flunitrazepam has not occurred in any material to which the new techniques have been applied. In view of the recently reported difficulty of detecting this compound and its metabolites (9), however, some experiments were made on spiked blood samples. We concluded that this group of compounds may be [ t t 2! JL_ I i I 270 sec. Figure 4. Three nitrazepam cases (A-C) and an example (D) of the effect of a modification of the detector potential from V to V vs. Ag/AgCI in case C. The same scale expansion was used in C and D; otherwise it was arbitrarily varied aminonitrazepam; 2 = 6-nitroquinoline (internal standard); 3 = nitrazepam. Concentrations are given in the text. 166

5 [ Irl.8 na 1 1!! 420 sec. I I I! 420 sec. Figure 5. Chromatograms from blood samples taken post-mortem in two temazepam overdoses (A,B). 1 = 6-nitroquinoline (internal standard); 2 = oxazepam; 3 = temazepam. Concentrations are given in the text. Figure 6. Chromatograms from blood spiked with flunitrazepam and two metabolites (A) and unspiked blood (B). 1 = 7-aminoflunitrazepam (20 ng/ml); 2 --= 6-nitroquinoline (internal standard); 3 = desmethylflunitrazepam (10 ng/ml); 4 = flunitrazepam (10 ng/ml). The eluent given in the experimental was modified by the omission of the l-propanol: determined similarly to the other benzodiazepines. An example is given in Figure 6. Hemolyzed human blood was spiked at 10 ng/ml with flunitrazepam and desmethylflunitrazepam and at 20 ng/ml with aminoflunitrazepam (chromatogram A). The unspiked blood is also shown (chromatogram B). Both samples included the usual internal standard. To remove flunitrazepam from the nitrazepam position, the eluent was modified by the omission of 1-propanol. For the interpretation of results from actual casework, account of degradation processes similar to those of nitrazepam cases may be needed. The new cleanup and detection techniques are likely to be relevant to other drugs and toxic materials. An obvious example is the alkyl nitrate vasodilators and their metabolites. These techniques were developed and are routinely used for this type of compound--in the form of traces of explosives and firearm residues (10). Polychlorinated compounds are another example, an instance of which occurred in a benzodiazepine case, shown in Figure 7. In addition to temazepam (1200 ng/ml) and oxazepam (280 ng/ml), the sample gave a peak due to trichloroethanol (3 #g/ml) from the metabolism of chloral hydrate. Other possible applications may be found in the extensive literature of polarography. Used with other detection techniques, the cleanup procedure could be widely applicable. (It should be noted that the detection technique exploited here is not a polarographic technique: each chromatogram is detected at a single mercury drop held at a constant potential.) References 1. J.B.F. Lloyd and D,A. Parry. Detection and determination of common benzodiazepines and their metabolites in blood samples of forensic science interest: Microcolumn cleanup and high- 3 na i! 0 sec. 420 Figure 7. Chromatogram from the postmortem blood of a subject prescribed chloral hydrate and temazepam. 1 = trichloroethanol; 2 = 6-nitroquinoUne (internal standard); 3 = oxazepam; 4 = temazepam. Concentrations are given in the text. 167

6 performance liquid chromatography with reductive electrochemical detection at a pendent mercury drop electrode. J. Chromatogr. 449: (1988). 2. J.B.F. Lloyd. High-performance liquid chromatography of organic explosives components with electrochemical detection at a pendent mercury drop electrode. J. Chromatogr. 257: (1983). 3. M.D. Osselton. Toxicological tables: A compendium of pharmacological, therapeutic and toxicological data on 136 drugs and chemicals in humans. Bull. InL Assoc. Forensic Toxicol. 17: (1983). 4. E.M. Sellers, C.A. Naranjo, V. Khouw, and D.J. Greenblatt. Binding of Benzodiazepines to Plasma Proteins. In Pharmacology of Benzodiazepines, E. Usdin, P. Skolnick, J.F, Tallman, D. Greenblatt, and S.M. Paul, Eds. MacMillan, London, 1982, pp S.R. Blunt, J.B.F. Lloyd, and D.A. Parry. Unpublished work. 6. C.P. Goddard, A. Stead, P.A. Mason, B. Law, and A.C. Moffat. An iodine-125 radioimmunoassay for the direct detection of benzodiazepines in blood and urine. Analyst 111: (1986). 7. K. Tada, A. Miyahira, and T. MorojL Liquid chromatographic assay of nitrazepam and its main metabolites in serum, and its application to pharmacokinetic study in the elderly. J. Liq. Chromatogr. 10: (1987). 8. H.J. Schwarz. Pharmacokinetics and metabolism of temazepam in man and several animal species. Br. J. C/in. Pharmacol. 8: 23S-29S (1979). 9. B. Heyndrickx. Fatal intoxication due to flunitrazepam. J. Anal. Toxicol. 11 : 278 (t987). 10. J.B.E Lloyd. Liquid chromatography with electrochemical detection of explosives and firearms propellant traces. Anal. Proc. 24: (1987). 168

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories. Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine

More information

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,

More information

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10

More information

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS Therapeutic Drug Monitoring of Antiretroviral Drugs with PLC-M Ursula Gutteck-Amsler, Katharina M. Rentsch Abstract Prospective and retrospective studies have provided some evidence of the clinical and

More information

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting

More information

SUCRALOSE. White to off-white, practically odourless crystalline powder

SUCRALOSE. White to off-white, practically odourless crystalline powder SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA

More information

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

FATAL INTOXICATION BY CIBENZOLINE (CIPRALAN )

FATAL INTOXICATION BY CIBENZOLINE (CIPRALAN ) FATAL INTOXICATION BY CIBENZOLINE (CIPRALAN ) N. RICHARD, Marie H. GHYSEL, M. SAVART Laboratoire de Police Scientifique, Lille, France ABSTRACT: We report here a fatal case of intoxication by an antiarrythmic

More information

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for

More information

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische

More information

How To Test For Contamination In Large Volume Water

How To Test For Contamination In Large Volume Water Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup

Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup Application Note 25570306 Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup Keywords Gel Permeation Chromatography GPC AutoPrep 2000 Pesticides RapidVap N2 System

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas

More information

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,

More information

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications TECHNICAL NOTE Toxicology Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

Overview. Purpose. Methods. Results

Overview. Purpose. Methods. Results A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines

More information

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA

More information

Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine

Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine Application Note Authors Peter JW Stone and Kevin McCann Agilent Technologies, Inc. Santa Clara,

More information

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 Simultaneous determination of Telmisartan

More information

New drugs of abuse? A case study on Phenazepam & Methoxetamine

New drugs of abuse? A case study on Phenazepam & Methoxetamine New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied

More information

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview

More information

LC-MS/MS for Chromatographers

LC-MS/MS for Chromatographers LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,

More information

6. ANALYTICAL METHODS

6. ANALYTICAL METHODS 45 6.1 BIOLOGICAL MATERIALS No methods were located that are routinely used for the detection of BCEE in biological materials. Norpoth et al. (1986) reported a method for measuring thiodiglycolic acid

More information

Gel Filtration Standard

Gel Filtration Standard Gel Filtration Standard Instruction Manual Catalog # 151-1901 Table of Contents Section 1 Introduction... 1 1.1 Instructions... 1 1.2 Recommended Volume of Standard... 2 1.3 Shelf Life... 5 1.4 Storage...

More information

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen

More information

IMPURITIES IN NEW DRUG PRODUCTS

IMPURITIES IN NEW DRUG PRODUCTS INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE IMPURITIES IN NEW DRUG PRODUCTS Q3B(R2) Current

More information

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

CLEAN-UP PROCESS FOR MASS SPECTRAL STUDY OF AMPHETAMINES IN PUTREFIED BODY MATERIALS

CLEAN-UP PROCESS FOR MASS SPECTRAL STUDY OF AMPHETAMINES IN PUTREFIED BODY MATERIALS CLEAN-UP PROCESS FOR MASS SPECTRAL STUDY OF AMPHETAMINES IN PUTREFIED BODY MATERIALS Kenji HARA, Seiichi KASHIMURA, Masayuki KASHIWAGI, Tomoko HAMANAKA, Mitsuyoshi KAGEURA Department of Forensic Medicine,

More information

Interlaboratory Studies on the Analysis of Hair for Drugs of Abuse: Results from the Fourth Exercise

Interlaboratory Studies on the Analysis of Hair for Drugs of Abuse: Results from the Fourth Exercise Interlaboratory Studies on the Analysis of Hair for Drugs of Abuse: Results from the Fourth Exercise Lorna T. Sniegoski and Michael J. Welch Analytical Chemistry Division, Chemical Science and Technology

More information

Anion chromatography using on-line recycled eluents

Anion chromatography using on-line recycled eluents Journal of Chromatography A, 1089 (2005) 82 86 Anion chromatography using on-line recycled eluents Takashi Yokoyama, Hitoe Maekubo, Asami Sakai, Michio Zenki Department of Chemistry, Faculty of Science,

More information

CUT-OFF CONCENTRATIONS FOR DRUGS OF ABUSE IN SALIVA FOR DUI, DWI OR OTHER DRIVING-RELATED CRIMES

CUT-OFF CONCENTRATIONS FOR DRUGS OF ABUSE IN SALIVA FOR DUI, DWI OR OTHER DRIVING-RELATED CRIMES CUT-OFF CONCENTRATIONS FOR DRUGS OF ABUSE IN SALIVA FOR DUI, DWI OR OTHER DRIVING-RELATED CRIMES Vina SPIEHLER 1, Dene BALDWIN 2, Christopher HAND 2 1 DABFT, Newport Beach, United States of America 2 Cozart

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

ABRF 2004 Poster# P55-S Displacement Chromatography Effects Can Cause Highly Selective Sampling of Peptides During Solid Phase Extraction Cleanup

ABRF 2004 Poster# P55-S Displacement Chromatography Effects Can Cause Highly Selective Sampling of Peptides During Solid Phase Extraction Cleanup ABRF 2004 Poster# P55-S Displacement Chromatography Effects Can Cause Highly Selective Sampling of Peptides During Solid Phase Extraction Cleanup by Andrew J. Alpert PolyLC Inc./ 9151 Rumsey Road, ste.

More information

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for

More information

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Application of Structure-Based LC/MS Database Management for Forensic Analysis Cozette M. Cuppett and Michael P. Balogh Waters Corporation, Milford, MA USA Antony Williams, Vitaly Lashin and Ilya Troisky Advanced Chemistry Development, Toronto, ntario, Canada verview Application

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT CRYOCOOLER CONTAMINATION STUDY S.W.K. Yuan, D.T.Kuo, and A.S. Loc EI Technologies, Cryocooler Group Sylmar, CA 91342 ASTRACT Eliminating or reduction of contamination is essential for the success of cryocoolers

More information

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex

More information

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,

More information

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Vitamin B 12 - A review of analytical methods for use in food. Government Chemist Programme Report

Vitamin B 12 - A review of analytical methods for use in food. Government Chemist Programme Report Vitamin B 12 - A review of analytical methods for use in food Government Chemist Programme Report March 2015 Vitamin B12 - A review of analytical methods for use in food Author: Paul Lawrance Contact point:

More information

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler

More information

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application

More information

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo

More information

Appropriate utilization of drug tests for pain management patients

Appropriate utilization of drug tests for pain management patients Appropriate utilization of drug tests for pain management patients Gwen McMillin, PhD, DABCC (CC, TC) Medical Director, Toxicology, ARUP Laboratories Associate Professor (clinical), University of Utah

More information

Scientific Working Group for Forensic Toxicology (SWGTOX) Standard Practices for Method Validation in Forensic Toxicology

Scientific Working Group for Forensic Toxicology (SWGTOX) Standard Practices for Method Validation in Forensic Toxicology Scientific Working Group for Forensic Toxicology (SWGTOX) Standard Practices for Method Validation in Forensic Toxicology 1. Introduction... 2 2. Definitions... 2 3. When to Validate Methods... 4 4. Method

More information

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS

More information

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Carmen T. Santasania and David S. Bell SUPELCO, 595 North Harrison Road, Bellefonte, PA 16823 T404045 GUS Abstract

More information

Analytical Specifications RIVAROXABAN

Analytical Specifications RIVAROXABAN Page 1 of 9 ANALYTE NAME AND STRUCTURE - RIVAROXABAN SYNONYMS Xarelto CATEGORY Anticoagulant TEST CODE PURPOSE Therapeutic Drug Monitoring GENERAL RELEVANCY BACKGROUND Xarelto (rivaroxaban) is an orally

More information

Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary

Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary Application Note Increasing the activity of monoclonal antibody isoforms by MCSGP Category Matrix Method Keywords Countercurrent chromatography, FPLC Antibodies MCSGP FPLC, Biobetters, MCSGP, countercurrent

More information

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.

More information

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Derek Hillbeck, Thermo Fisher Scientific, Runcorn, UK Application Note 2999 Key

More information

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:

More information

A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System

A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System Kent Johnson Fortes lab, Wilsonville Oregon List of drug of abuse candidates for

More information

AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT

AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT AppNote 7/2009 Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS Fred D. Foster, John R. Stuff, Edward A. Pfannkoch Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum,

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C

More information

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction

More information

SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS

SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS Ingrid J. BOSMAN 1, Douwe DE BOER 2, Robert A. A. MAES 1 1 Department of Human

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic

More information

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1 GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,

More information

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -

More information

Purification of reaction mixtures using flash chromatography.

Purification of reaction mixtures using flash chromatography. Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?

More information

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND

More information

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer

More information

Affi-Prep Protein A Matrix Instruction Manual

Affi-Prep Protein A Matrix Instruction Manual Affi-Prep Protein A Matrix Instruction Manual Catalog Numbers 156-0005 156-0006 Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules, CA 94547 LIT-230 Rev B Table of Contents Section 1 Introduction...1

More information

Introduction. The following definitions may help you better understand the components of the data report.

Introduction. The following definitions may help you better understand the components of the data report. Introduction This document is designed to help our clients understand the quality control requirements and limitations of data reporting. There are three sections to this document. The first section will

More information

What Does the Drug Test Tell Us

What Does the Drug Test Tell Us What Does the Drug Test Tell Us QUALITATIVE TOXICOLOGY TESTING for CLINICAL MANAGEMENT of the PATIENT Background The qualitative drug test is often referred to as the drug screen - a misnomer Testing performed

More information

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary Application Note Separation of three monoclonal antibody variants using MCSGP Category Matrix Method Keywords Analytes ID Continuous chromatography, Biochromatography; FPLC Protein A-purified monoclonal

More information

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs

More information

A Novel Approach to Low Volume Sample Preparation

A Novel Approach to Low Volume Sample Preparation A Novel Approach to Low Volume Sample Preparation Tony Edge, Jon Bardsley, Ken Meadows, Luisa Pereira International Symposium on Chromatography Salzburg, Austria September 2014 1 The world leader in serving

More information

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated

More information

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection 52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis

More information

I. ACID-BASE NEUTRALIZATION, TITRATION

I. ACID-BASE NEUTRALIZATION, TITRATION LABORATORY 3 I. ACID-BASE NEUTRALIZATION, TITRATION Acid-base neutralization is a process in which acid reacts with base to produce water and salt. The driving force of this reaction is formation of a

More information

Quantitation of Drugs in Dried Bloodstains. Thomas Meyer Anna Gomenyuk Nadiah Lester

Quantitation of Drugs in Dried Bloodstains. Thomas Meyer Anna Gomenyuk Nadiah Lester Quantitation of Drugs in Dried Bloodstains Thomas Meyer Anna Gomenyuk Nadiah Lester Objective To find the quantitative value of drugs in dried bloodstains Currently Cloth With Dried Bloodstain Containing

More information

How To Use An Acquity Qda Detector

How To Use An Acquity Qda Detector Mass-Directed Isolation of a Synthetic Peptide Using the ACQUITY QDa Detector Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY QDa Detector

More information

MASS SPECTROMETRIC IDENTIFICATION OF SOME SULPHUR CONTAINING PHENALKYLAMINE DESIGNER DRUGS *

MASS SPECTROMETRIC IDENTIFICATION OF SOME SULPHUR CONTAINING PHENALKYLAMINE DESIGNER DRUGS * MASS SPECTROMETRIC IDENTIFICATION OF SOME SULPHUR CONTAINING PHENALKYLAMINE DESIGNER DRUGS * Ingrid J. BOSMAN 1, Douwe DE BOER 2, Edwin B. SIDERIUS 1, Lesseps J. A. L. DOS REYS 2, Robert A. A. MAES 1 1

More information

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection Application Note 9 The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection INTRODUCTION The accurate measurement of the amount of sugar in final molasses

More information

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PARACETAMOL ORAL SUSPENSION (September 2010)

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PARACETAMOL ORAL SUSPENSION (September 2010) September 2010 RESTRICTED DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PARACETAMOL ORAL SUSPENSION (September 2010) DRAFT FOR COMMENTS This document was provided by a quality control expert and

More information

Chemical analysis service, Turner s Green Technology Group

Chemical analysis service, Turner s Green Technology Group Chemical analysis service, Turner s Green Technology Group Hourly costs, Academic collaborations leading to co-publications: 627 SEK/hr (excl. VAT) Hourly costs, Industry/Academic collaborations, no publications:

More information

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,

More information

Title 17 California Code of Regulations

Title 17 California Code of Regulations Title 17 California Code of Regulations s 1215. Authority. Chapter 5 Sections 436.50-436.63 of Part 1 of Division 1 of the Health and Safety Code. Note: Authority cited: Sections 102 and 208, Health and

More information

Opiates in Urine by SAMHSA GC/MS

Opiates in Urine by SAMHSA GC/MS application Note Gas Chromatography/ Mass Spectrometry Author Timothy D. Ruppel PerkinElmer, Inc. Shelton, CT 06484 USA Opiates in Urine by SAMHSA GC/MS Introduction The United States Department of Health

More information

Teriflunomide is the active metabolite of Leflunomide, a drug employed since 1994 for the treatment of rheumatoid arthritis (Baselt, 2011).

Teriflunomide is the active metabolite of Leflunomide, a drug employed since 1994 for the treatment of rheumatoid arthritis (Baselt, 2011). Page 1 of 10 ANALYTE NAME AND STRUCTURE TERIFLUNOMIDE Teriflunomide TRADE NAME Aubagio CATEGORY Antimetabolite TEST CODE PURPOSE Therapeutic Drug Monitoring GENERAL RELEVANCY BACKGROUND sclerosis. The

More information