Other than the obvious effect of reduced audibility of conversational

Size: px
Start display at page:

Download "Other than the obvious effect of reduced audibility of conversational"

Transcription

1 666 JSLHR, Volume 40, , June 1997 Comparison of the enefit Provided by Well-Fit Linear Hearing ids and Instruments With utomatic Reductions of Low-Frequency Gain Larry E. Humes Department of Speech and Hearing Sciences Indiana University loomington Laurel. Christensen Department of Communication Disorders LSU Medical Center New Orleans, L Fred H. ess ndrea Hedley-Williams Division of Hearing and Speech Sciences Vanderbilt University School of Medicine Nashville, TN In this clinical study, 110 patients seen at three different clinical facilities were fit binaurally with linear, in-the-canal (ITC) hearing aids. ll patients were new hearing aid users. Each of the hearing aids was equipped with an adjustable control that could be set by one of the audiologists (udiologist ) at each site to convert it from a linear instrument to an experimental nonlinear one with automatic reduction of low-frequency gain at high input levels (or base increase at low levels, ILL). oth the patient and the audiologist performing the outcome testing at each site (udiologist ) were blind as to the present setting of the hearing aid. Each participant was enrolled in the study for a total of 12 weeks, with the hearing aid set to either the linear or ILL-processing mode of operation for the first 8 weeks and the opposite setting for a subsequent 4-week period. In summary, this was a prospective, doubleblind, crossover study of 110 new hearing-aid users. Outcome measures focused on hearing-aid benefit and included both objective and subjective measures. Objective measures were derived from scores on the Northwestern University uditory Test No. 6 (NU-6) and the Connected Speech Test (CST) obtained for all possible combinations of two speech presentation levels (60 and 75 d SPL), two types of background noise (cafeteria noise and multitalker babble), and two signal-to-noise ratios (+5 and +10 d). Subjective outcome measures included magnitude estimation of listening effort (MELE), the abbreviated form of the Hearing id Performance Inventory (HPI), and estimations of hearing-aid usage based on daily-use logs kept by the participants. ll of these measures were used to evaluate the benefit provided by linear amplification and the benefit resulting from the experimental ILL processing. Participant preferences for the experimental ILL-processing scheme or linear processing were also examined by using a paired-comparison task at the end of the study. Results were analyzed separately for three subgroups of patients (mild, moderate, severe) formed on the basis of their average hearing loss at 500, 1000, 2000, and 4000 Hz. In all three subgroups, significant improvement in performance was observed for linear amplification and for ILL processing when compared to unaided performance. There were no significant differences in aided performance, however, between linear processing and the experimental ILL processing. KEY WORDS: hearing aids, benefit, linear processing, ILL-processing, speech understanding in noise Other than the obvious effect of reduced audibility of conversational level speech, it has long been observed that sensorineural hearing loss interferes with communication in noisy backgrounds. 666 Journal of Journal Speech, of Speech, Language, Language, and Hearing and Hearing Research Research /97/ , merican Speech-Language-Hearing ssociation

2 Humes: enefit From Linear Hearing ids 667 For the vast majority of listeners with sensorineural hearing loss seen by audiologists and hearing instrument specialists over the past few decades, the conventional approach to rehabilitation has been to fit the person having sensorineural hearing loss with a linear, head-worn hearing aid with output limiting accomplished via peak clipping (ess & Humes, 1995; Hawkins & Naidoo, 1993). To improve the communication performance of persons wearing such instruments in noise a variety of modifications to the basic linear, peak-clipping instrument has been investigated. High-pass filtering in a hearing aid has been advanced as one technique to minimize the effect of background noise on speech communication. High-pass filtering can be accomplished in a hearing aid in two ways. First, the hearing aid can be set to have a static high-pass response. t all times, the hearing aid s frequency response is set to minimize low-frequency gain and to maximize high-frequency gain. The principal drawback to this approach is that many listeners do not like the sound quality of such an amplifying system in quieter environments (Punch & eck, 1986). Further, there is useful speech information in the lower frequencies that should be provided to the listener, if at all possible. This low-frequency information should be restricted only if there is a drawback to its presence (such as excessive upward spread of masking or undesirable sound quality). The second manner in which low-frequency energy can be attenuated is via adaptive high-pass filtering (Fabry & Walden, 1990). In this approach, the hearing aid is set to have a relatively broad response in quieter environments. s the overall sound level in the environment increases above a certain threshold level, the hearing aid automatically shifts to a more high-pass response. The assumption behind such an approach is that the listener is likely to prefer a broader response if there is a minimal amount of noise present. If the overall level of the listening environment rises, there is a high likelihood that a significant amount of this energy is low-frequency noise (Ono, Kanzaki, & Mizoi, 1983). Therefore, the hearing aid automatically reduces gain for this undesirable low-frequency sound and concentrates the response of the hearing aid to the more important high-frequency region. This automatic low-frequency reduction in gain at high input levels is implemented through different methods in several models of commercially available hearing aids. It has been variously termed automatic signal processing (SP), adaptive frequency response (FR), and bass increase at low levels (ILL). Some commercial devices incorporate adaptive high-pass filtering in single-channel analog devices, whereas others make use of single- or multichannel compression to realize ILL processing. Despite the prevailing use of linear hearing aids over the past several decades, there have been no comprehensive, large-scale, prospective clinical studies of the benefits provided by such instruments. Such information is critically important as alternatives to linear amplification, such as ILL-processing, multichannel compression, and wide-dynamic-range compression, are developed and evaluated. When studies have been conducted comparing linear and nonlinear systems, it has typically been the case that the hearing aids under investigation differ in more than just their linearity. In view of the foregoing, the purpose of this study was to evaluate the clinical performance of a linear ITC hearing aid with and without activation of an experimental ILL-based processing option. The hearing aid, with and without the ILL processing activated, was evaluated using objective and subjective tests of device performance and measures of participant preference. s described in more detail below, a total of 110 persons with sensorineural hearing loss participated in this prospective study conducted at three different clinical sites. Protocol Overview of Design This was a prospective, double-blind, crossover, controlled study. The study population was divided into three groups according to their hearing loss: (1) mild, defined as a four-frequency (500, 1000, 2000, and 4000 Hz) pure tone average (4PT) of d HL; (2) moderate, 4PT of d HL; and (3) severe, 4PT of d HL. Each of the participants was tested with the same hearing aid in two different circuit configurations: ILL option and linear. Randomly, one-half of the participants were dispensed the ILL circuit active first, and the other one-half received the linear circuit active first. fter an appropriate adjustment period (discussed below), the participant s performance with the first circuit configuration was assessed. fter that assessment, the participant was switched to the second circuit configuration and was given an appropriate adjustment period. The participant then returned for an assessment of performance with the second circuit configuration. ssignment of initial aided condition (ILL vs. linear) was randomized such that an equal number of participants was included in each group. Further, the study was double blind. The participants did not know which circuit type was being tested first. t each test site, one audiologist () was responsible for fitting the hearing aids. second audiologist (), without knowledge of which circuit type a participant was using for any given test session, performed the performance testing. The participants and udiologist were also blind regarding the hearing aid s manufacturer. This was accomplished by having no manufacturer s labeling or markings on the hearing aid

3 668 JSLHR, Volume 40, , June 1997 casing or accompanying hardcopy materials, as well as by channeling all communications between the manufacturer and test site through udiologist. This study was carried out at three independent audiology research facilities: (1) the Division of Hearing and Speech Sciences, Vanderbilt University School of Medicine, Nashville, Tennessee; (2) the Department of Communication Disorders, Louisiana State University Medical Center, New Orleans, Louisiana; and (3) the Department of Otolaryngology, University of Iowa Hospitals and Clinics, Iowa City, Iowa. These facilities were all in close physical and administrative proximity to an active clinical hearing aid dispensing program. Investigators at all research facilities obtained the appropriate Institutional Review oard approval for the use of human participants at their facilities. Participants The participant was eligible for entry into the study if the participant: (a) was at least 18 years of age or older; (b) had a flat or gently sloping sensorineural hearing loss (slope 0 15 d/octave from 250 Hz to 4000 Hz, no interoctave change of greater than 20 d), consistent with the fitting guidelines of the hearing aid; (c) had a hearing loss that was symmetrical (within 10 d at octave and halfoctave frequencies from 250 Hz to 4000 Hz); (d) had thresholds that ranged between 25 and 85 d HL at octave and inter-octave intervals from 250 Hz to 4000 Hz, inclusively; (e) had received medical clearance by a physician to use hearing aids binaurally; (f) had read, understood, and signed the Informed Consent form; (g) had normal tympanograms bilaterally; and (h) resided within a reasonable distance of the test facility. The participant was ineligible for entry into the study if the participant: (a) was a prior hearing aid user, although the participant could have participated in a prior 30-day trial of amplification (but not within the previous 6 months); (b) had any indication of a medically or surgically treatable ear disorder; (c) had an external auditory canal with a size and shape that was inconsistent with an appropriately sized device; (d) had known fluctuating or rapidly progressive hearing loss; (e) had any cognitive, medical, or language-based condition that limited his/her ability to complete all test procedures; (f) was currently participating in any other clinical trials; (g) had any known disease or condition, other than sensorineural hearing loss, that might affect hearing or cognition; or (h) was taking any medication(s) that could affect hearing or cognition. Participants could also be discontinued from the study for one of the following reasons: (a) development of any of the exclusion criteria during the course of the study, such as medical/surgical ear condition, fluctuating hearing loss, disease/condition affecting hearing/cognition, or VIII-nerve tumor; (b) inability or unwillingness to maintain wearing schedule; (c) inability or unwillingness to make or keep clinic appointments during study; (d) death or disability; (e) relocation of participant; (f) over 3 weeks cumulative off-the-ear time; (g) change in participant s thresholds at two or more test frequencies by more than 10 d, as compared to the initial baseline audiogram; or (h) a missed evaluation visit. For their participation in the study, the participants received their hearing aids at no charge. ll clinical services provided during the study were also provided at no charge to the participant. The mean air-conduction audiograms for the left (squares) and right (circles) ears for each of the three hearing-loss subgroups are provided in Figure 1. Standard deviations are also given in the upper portion of Figure 1. The mean audiogram for each subgroup shows a bilaterally symmetrical, gently sloping configuration that varies in severity across subgroups. Table 1 shows the distribution of participants by site, severity of hearing loss, and gender. ge ranges are also provided. s can be seen, most of the participants with severe hearing loss were tested at the University of Iowa and were mostly men, whereas those with mild hearing loss were fairly evenly distributed across all three sites and were mostly women. Figure 1. Mean right (circles) and left (squares) air-conduction thresholds (NSI, 1989) for the mild, moderate, and severe hearing-loss participant groups. N indicates the number of participants in each group. Corresponding standard deviations for each group are plotted at the top of the audiogram.

4 Humes: enefit From Linear Hearing ids 669 Table 1. Detailed descriptions of study participants by clinical site, participant hearing loss severity, and participant gender. Hearing loss Site # Males # Females ge range (y) Mild LSU Vanderbilt Iowa Figure 2. Illustration of the effects of hearing-aid circuit setting (linear vs. ILL) on the frequency response measured in accordance with NSI S Frequency responses are shown for both the linear (solid lines) and ILL (dashed lines) settings. Each panel depicts the frequency responses for each setting obtained for a different input level (50 90 d SPL). Results shown are for the highest gain linear circuit included in this study. Moderate LSU Vanderbilt Iowa Severe LSU Vanderbilt Iowa Hearing id ll participants were fit with binaural, custom, inthe-canal (ITC) hearing aids manufactured by Dahlberg, Inc. The shells of the hearing aids were unvented (which was considered to be generally appropriate given the degree and configuration of the hearing losses) and were fabricated from standard materials and with mediumlength canals. The test hearing aids included an experimental ILL circuit that reduced the amount of gain in the low frequencies as the input intensity increased. manually adjustable potentiometer on the device, accessible only to the audiologist, allowed the ILL circuit to be deactivated. When the circuit was deactivated, the device acted as a linear amplifier the most common circuit currently used on the hearing aid market. This particular linear circuit, however, made use of a patented Class output stage that, according to the manufacturer, incorporated low-distortion peak clipping, reduced low-frequency circuit noise, and low current drain. When the ILL circuit was active, the threshold for activation of the ILL processing was set at the lowest possible level (about 60 d SPL). This was essentially a single-channel analog device with adaptive highpass filtering with a variable threshold for activation of the adaptive filtering. When the potentiometer was adjusted to one extreme position, inputs as low as 60 d SPL would activate the adaptive high-pass filtering and maximum reduction of low frequencies occurred, whereas adjustment to the other extreme position resulted in deactivation of the adaptive filtering (linear processing). Figure 2 illustrates the effect of these two extreme potentiometer settings on the frequency response of the hearing aid as a function of input level. The biggest reductions in low-frequency gain are obtained for the ILL circuit for inputs of 70 and 80 d. Note that activation of the ILL circuit actually reduces the gain in the mid and high frequencies by about 8 10 d, compared to the linear setting at an input level of

5 670 JSLHR, Volume 40, , June d and, to a lesser extent, at 70 d. lthough results shown in Figure 2 are only for the hearing aid circuit with the highest gain, identical patterns of results were observed for the circuit with lowest gain, although the reduction of gain in the mid- and high-frequency regions was less pronounced (4 6 d) for this circuit. The relative effects of ILL processing shown in this figure, moreover, are independent of volume-control setting. During the study, the hearing aids were marked only with a serial number no brand identification information. t study completion, they were returned to the manufacturer for proper labeling before they were returned to the participants. ll hearing aids were tested electroacoustically in accordance with NSI S before dispensing and at all post-fit evaluations. The gain, frequency response, and maximum output of each hearing aid was individually selected by the manufacturer on the basis of audiometric characteristics of each test ear. The gain and frequency response were selected to meet the targets prescribed by the revised National coustics Laboratory (NL-R) fitting procedure (yrne & Dillon, 1986) when set for the linear mode of operation. The desired frequency response was confirmed via probe-tube measurement techniques at hearing-aid delivery. Consistent with the recommendations of Mueller (1992), the real-ear insertion gain of the hearing aid was within 10 d at all octave frequencies below 4000 Hz and within 15 d at 4000 Hz. The mean target gain curves based on the NL-R formula and the mean real-ear insertion responses for the linear setting obtained from the participants for a 60-d- SPL input are shown for each ear and hearing-loss subgroup in Figure 3. There is excellent agreement between the prescribed (target) gain and the measured insertion gain in all cases from 250 Hz to 4000 Hz, but a consistent underamplification of about 8 12 d at 6000 Hz. SSPL90 targets were derived from loudness discomfort levels (LDLs) measured using procedures described by Hawkins, Walden, Montgomery, and Prosek (1987). The maximum output of the device was selected to be no more than 10 d below the SSPL90 target maximum at octave frequencies from 500 Hz though 4000 Hz. Hearing id Evaluations No evaluation of hearing aid benefit took place until after a 2-month adjustment period with the first circuit. This adjustment period is consistent with recent evidence of the maturation of hearing aid benefit or acclimitization (Cox & lexander, 1992; Gatehouse, 1992, 1993), allowing for plateau performance, although there is some debate about the size and significance of this effect (Turner, Humes, entler, & Cox, 1996). Nonetheless, it was assumed that it takes some finite amount of time for a new hearing aid user to get used to listening to an amplified signal. It may not be until several weeks after being fit with a new hearing aid that some users can make maximal use of the new auditory information. Therefore, in order to increase the likelihood of achieving optimal performance with the hearing aids in this study, the user was allowed to adapt to its use. longer period of acclimitization (8 weeks) was implemented for the trial with the first circuit because it was anticipated that there would be a greater need for adjustment to the use of amplification initially that included familiarization with the insertion/ removal of the device, operation of controls, replacement of batteries, and adjustment to the new sound quality. During the adjustment period, no changes in the response or casing of the hearing aid were allowed, except in cases of reported physical or loudness discomfort or acoustic feedback. In typical clinical practice, the audiologist may make fine-tuning adjustments in the participant s hearing aid during the first several weeks. These adjustments are typically due to sound-quality preferences on the part of the participant. However, these adjustments are not predictable and will vary from participant to participant depending on factors that are not fully understood. The NL-R prescriptive rules are well developed, well supported in the literature, and generally accepted by audiologists. Therefore, in an attempt to provide consistent responses from participant to participant, variations from the NL-R prescribed hearing aid responses were not permitted during the study. The first evaluation session took place no more than 2 weeks after the end of the 2-month adjustment period. fter the first circuit had been evaluated, the participant s hearing aids were changed to the alternative circuit. The participant was then given a 1-month adjustment period with the second circuit. The second evaluation session took place no more than 2 weeks after the end of this 1- month adjustment period. Follow-up phone and mail contacts were made in cases in which participants did not keep scheduled evaluation sessions. Speech levels vary across a broad range in typical communicative environments (Pearsons, ennett, & Fidell, 1977). In this study, testing was performed at two different speech levels, 60 and 75 d SPL, which represent normal and raised speech (Pearsons et al., 1977). Further, the degree of low-frequency amplification in the experimental ILL circuit varies inversely with the overall input level. Evaluating nonlinear circuits at more than one level has been recommended by others (e.g., French-St. George, Engebretsen, & O Connell, 1992). y testing at these two levels, we were able to assess ILL-circuit performance aspects and aided benefit across a range of typical input levels. variety of background noise types has been used in past evaluations of automatic low-frequency reduc-

6 Humes: enefit From Linear Hearing ids 671 Figure 3. Mean target real-ear insertion gain generated by the NL-R formula (squares) compared to the mean real-ear insertion gain measured (circles) for a 60-d input. Vertical bars above and below circles represent one standard deviation above and below the mean. Left-hand panels and unfilled symbols are for left ear, whereas right-hand panels and filled symbols are for right ear. Top panels present data for group with mild hearing loss, middle panels contain data for group with moderate hearing loss, and bottom panels present data for group with severe hearing loss. tion amplification. The performance of automatic lowfrequency reduction amplification has been demonstrated to vary with background noise type (Fabry, 1991). Specifically, as the noise shifts from predominantly lowfrequency energy to more broadband energy, the beneficial effects of the circuit tend to diminish. lso, the use of steady-state narrow-band, speech, or white noise does not replicate real-world listening environments. Therefore, in this study, two noise types were used: cafeteria noise (uditec) and multitalker speech babble (uditec). Spectrally, cafeteria noise tends to be dominated by lowfrequency energy, but continues to have some higher

7 672 JSLHR, Volume 40, , June 1997 frequency content because of the presence of some competing speech. abble tends to have the same long-term characteristics as test speech material. y using both cafeteria noise and speech babble, we assessed the effects of the ILL circuit and the benefit provided by both ILLprocessing and linear hearing aids in realistic, yet spectrally different, background conditions. Procedures efore the beginning of this study, investigators from each clinical site met with the study sponsor (the manufacturer) to review the protocol and data-handling issues. Identical materials, protocols, and forms were used at all sites. n independent study monitor familiar with medical-device clinical field studies was retained by the study sponsor to monitor the study procedures at each site; this person made several visits to the sites throughout the duration of the study to assure uniformity and adherence to the protocol. t the patient s first visit to the facility, referred to as the recruitment visit, information on the participant s history, demographics, eligibility, unaided hearing performance, and hearing-aid fit was collected and evaluated. Ear impressions were obtained at this visit, and the hearing aids were ordered. The investigator also explained the Physician Release Form and the Informed Consent Form to the participant during this visit. The second visit, the dispensing visit, took place within 3 weeks of the recruitment visit. t this visit, the participant returned the signed Informed Consent and Physician Release Forms, and before the hearing aids were dispensed udiologist confirmed the participant s hearing thresholds measured previously. Next, udiologist evaluated the hearing aids in the testbox, measured the real-ear performance, and gave a hearing-aid-use log to the participant with an explanation of its use. Participants were instructed to wear their hearing aids at least 5 hours per day, 5 days per week. udiologist then obtained unaided speech-recognition scores, along with magnitude estimates of listening effort (see below), from the participant. The third visit to the facility represented the first aided evaluation of performance and took place no more than 2 weeks after the 2-month acclimatization period (time from dispensing visit to the first aided evaluation visit). t this visit, hearing sensitivity was again measured by udiologist to confirm the stability of hearing thresholds. udiologist evaluated the hearing aids in the testbox and collected the hearing-aid-use log from the participant. In addition, the Hearing id Performance Inventory (HPI; Walden, Demorest, & Hepler, 1984) was administered in its abbreviated form (Schum, 1993) using a pencil-andpaper format, and a new hearing-aid-use log was given to the participant. udiologist then obtained aided speech-recognition and listening-effort scores. Once these measures were completed, the hearing aid was switched to the alternate setting (linear or ILL) by udiologist. The fourth visit to the facility represented the second and final aided evaluation of performance and took place no more than 2 weeks after the 1-month adjustment period (time from the first aided evaluation visit to the second aided evaluation visit). t this visit, hearing sensitivity was again measured by udiologist to confirm the stability of hearing thresholds. udiologist evaluated the hearing aids in the testbox and collected the hearing-aid-use log from the participant. In addition, the abbreviated HPI was again administered. udiologist then obtained aided speech-recognition and listening-effort scores. Finally, a series of pairedcomparison measurements was performed by the participant with the assistance of udiologist. Table 2 provides a summary of the sequence of measurements peformed during this study. It also indicates which audiologist ( [not blind] or [blind]) at each site was involved in the measurements. Explanations of hearing-aid function, use, care, and communication strategies were provided by udiologist. Outcome Measures Taped speech-recognition tests (NU-6; Tillman & Carhart, 1966 / Connected Speech Test, CST; Cox, lexander, Gilmore, & Pusakulich, 1988) were administered at +5 and +10 d signal-to-noise ratios (S/N) in babble and cafeteria noise at speech presentation levels of 60 and 75 d SPL. Oral responses were used for both tests. ll scores for both the NU-6 and CST tests were based on 50 items. Speech and noise were presented from a single loudspeaker located at zero-degrees azimuth and one meter from the center of the participant s head. ll sound pressure levels refer to those measured in the sound field using the method of substitution (Skinner, 1988). magnitude-estimate of listening effort (MELE) was also developed and used in this study. In this task, the listener rated the difficulty of listening to speech in cafeteria noise and babble backgrounds. The listener heard a 10-sentence encyclopedia-style passage from the CST. Four passages were presented at 60 and 75 d SPL in both noise backgrounds at +5 and +10 d S/N (32 total passages). fter each passage, the listener rated ease of listening on a scale, with 100 representing extremely easy listening. In each listening condition, the ratings from the last three passages were averaged and the first in each condition was discarded as practice. fter completion of the second round of objective and subjective measures of performance, the participant was then placed in the sound field and presented with

8 Humes: enefit From Linear Hearing ids 673 Table 2. Summary of visits and forms used in the study. Session Form/Data description udiologist 1st visit: Recruitment Participant History Form Order Form Pure tone thresholds Loudness discomfort levels Speech discrimination Tympanometry Hearing aid order information Impressions Eligibility Criteria (reviewed participant eligibility) Physician Release (explained & gave to participant) Informed Consent (explained and gave to participant) 2nd visit: Dispensing Eligibility Criteria (reconfirmed eligibility and signed) Performance Testing form Pure tone thresholds NU-6 monosyllabic word list score a Connected Speech Test score a Listening Effort Test score a Hearing Inventory (HHIE) Hearing id NSI Performance Hearing id Use Log (explained & gave to participant) 3rd visit: 1st Evaluation Performance Testing form Pure tone thresholds NU-6 monosyllabic word list scores a Connected Speech Test scores a Listening Effort Test scores a Hearing id Performance Inventory (HPI) Hearing id NSI Performance Hearing id Use Log (explained & gave to participant) 4th visit: 2nd Evaluation Performance Testing form Pure tone thresholds NU-6 monosyllabic word list scores a Connected Speech Test scores a Listening Effort Test scores a Hearing id Performance Inventory (HPI) Hearing id NSI Performance Paired Comparison Testing a a These tests were performed at +5 and +10 d S/N; at 60 and 75d SPL; and in cafeteria and babble noise. continuous discourse (the Rainbow Passage recording available on the Q/MSS Speech udiometry, Volume 1 compact disk) at the same levels and signal-to-noise ratios described above. The audiologist then switched the circuit back and forth from the ILL to the linear setting. The participant heard a 15- to 20-second passage in one circuit setting, followed by a second 15- to 20- second passage with the other circuit active. The order of which circuit was active first was randomly varied. The participant then chose which circuit he/she preferred to use in that noise condition. The participant used the following rating scale to indicate preference: (1) Circuit significantly better than circuit ; (2) Circuit moderately better than circuit ; (3) Circuit slightly better than circuit ; (4) Circuit slightly better than circuit ; (5) Circuit moderately better than circuit ; (6) Circuit significantly better than circuit. The participant made one practice, then three test ratings, at each speech level, in both noise backgrounds and at each signal-to-noise ratio. The results from the three test ratings were averaged in each listening condition. Result total of 115 participants was enrolled in the study, but 1 was not dispensed a hearing aid and 4

9 674 JSLHR, Volume 40, , June 1997 were discontinued. Of the 4 discontinued participants, 1 was discontinued because closer evaluation revealed that the participant did not meet all of the inclusion criteria and should not have been enrolled, whereas the other 3 were lost to follow-up (did not return and could not be reached). Of the 110 remaining participants, none missed a scheduled visit. The results described below are for the 110 who were enrolled and completed the entire study. Group Data Linear vs. ILL-Processing NU-6 Test Recall that the NU-6 test was performed at several signal-to-noise ratios, speech levels, and in two different types of noise background at the dispensing visit (unaided) and the first and second evaluations (aided). Figure 4 provides a summary of the mean aided NU-6 percent-correct data for the 110 participants with mild, moderate, or severe hearing loss, respectively. Table 3 provides the corresponding standard deviations for the mean values appearing in Figure 4. Figure 4. Mean percent-correct scores on the NU-6 test for the linear (cross-hatched bars) and ILL (unfilled bars) settings of the hearing aid. Panels are organized from top to bottom according to hearing loss group such that participants with mild hearing loss are in the top two panels, those with moderate hearing loss are in the middle two panels, and those with severe hearing loss are in the bottom two panels. ll left-hand panels contain data for the 60-d- SPL presentation level, whereas all right-hand panels contain data for the 75-d-SPL presentation level. Within each panel, results for babble background at +5 and +10 d signal-to-noise ratio, followed by cafeteria-noise background at +5 and +10 d signalto-noise ratio, are plotted along the x-axis as one moves from left to right. Connected Speech Test The Connected Speech Test (CST) was also performed for the same combinations of signal-to-noise ratio, speech level, and types of background noise used with the NU-6 at the dispensing visit (unaided) and at the first and second evaluations (aided). Figure 5 provides a summary of the mean aided CST percent-correct data for the 110 participants with mild, moderate, or severe hearing loss, respectively. Table 4 provides the corresponding standard deviations for the mean CST values appearing in Figure 5. NOVs on NU-6 and CST Scores Percent-correct scores for the NU-6 test and the CST were arcsine transformed to stabilize the error variance (Kirk, 1968), and the difference scores between the linear and ILL settings were computed. n analysis of variance (NOV) with a between-subject factor of hearing-loss group and four bivalent repeated-measures or within-subject variables (test materials, signal-to-noise ratio, noise type, and speech level) was performed first and showed no statistically significant (p <.05) effects of hearing-loss group (either main effects or interactions). Consequently, a second NOV was performed on the difference scores (linear vs. ILL) for all 110 participants and the four repeated-measures, within-participant variables without regard to hearing-loss group. The results of this NOV are summarized in Table 5 and indicate that there was no difference in performance between the linear and ILL settings (i.e., the linear-minus-ill difference scores did not differ significantly from zero). Moreover,

10 Humes: enefit From Linear Hearing ids 675 Figure 5. Mean percent-correct scores on the CST for the linear (cross-hatched bars) and ILL (unfilled bars) settings of the hearing aid. Panels are organized from top to bottom according to hearing loss group such that participants with mild hearing loss are in the top two panels, those with moderate hearing loss are in the middle two panels, and those with severe hearing loss are in the bottom two panels. ll left-hand panels contain data for the 60-d-SPL presentation level, whereas all right-hand panels contain data for the 75-d-SPL presentation level. Within each panel, results for babble background at +5 and +10 d signal-to-noise ratio, followed by cafeteria-noise background at +5 and +10 d signalto-noise ratio, are plotted along the x-axis as one moves from left to right. none of the other variables (test materials, noise type, signal-to-noise ratio, or speech level) significantly affected the linear-minus-ill difference scores, either individually or in combination. Magnitude Estimation of Listening Effort Test (MELE) The MELE test required the listener to rate, from 0 to 100, the ease of listening to speech in cafeteria noise and in babble backgrounds. The same connected-speech passages used in the CST served as the speech stimulus. s with the NU-6 and CST tests, the MELE test was performed at +5 and +10 d S/N in babble and cafeteria noise, with the speech signal presented at 60 and 75 d SPL at the dispensing visit (unaided) and the first and second evaluation visits (aided). Figure 6 provides a summary of the MELE data for the 110 participants in this study with mild, moderate, or severe hearing loss, respectively. Table 6 provides the standard deviations corresponding to the mean MELE data appearing in Figure 6. Ratings, which ranged from 0 to 100, were repeated a total of four times per condition, with the first rating discarded and the remaining three averaged for each subject. Preliminary analyses conducted on the linear-minus-ill difference scores, as used in the analysis of the NU-6 and CST results, found significant heterogeneity of variance for the MELE difference scores. Consequently, analyses of variance were performed on the raw scores rather than difference measures for the MELE test (variances of raw scores were found to be homogeneous across conditions). n analysis of variance performed on the MELE ratings, with hearing-aid setting (linear vs. ILL) as a variable, is summarized in Table 7. For the most part, this analysis indicated that there was no significant effect of the hearing-aid setting on MELE ratings. That is, the main effect of circuit setting (linear vs. ILL) was not significant and only one of the 15 interactions involving this variable was significant (hearing loss group x noise type x setting). The presence of this significant interaction involving hearing loss group, however, confounds a simple interpretation of the effect of hearing-aid setting. From visual inspection of the data in Figure 6, this threeway interaction involving the hearing-aid setting appears to be due to the higher MELE scores for the subjects with severe hearing loss listening in babble for the ILL hearing-aid setting and the reversal of this trend under the same conditions for the listeners with mild hearing loss. The presence of several significant two-way interactions (noise type x signal-to-noise ratio, noise type x presentation level, and signal-to-noise ratio x presentation level), none involving the hearing-aid setting variable, illustrates the importance of assessing MELE across a range of listening conditions.

11 676 JSLHR, Volume 40, , June 1997 Table 3. Standard deviations for NU-6 percent-correct scores for participants with mild, moderate, or severe hearing loss. Processing condition and presentation level ILL Linear Stimulus condition 60 d SPL 75 d SPL 60 d SPL 75 d SPL Mild hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Moderate hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Severe hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Order Effect The results were analyzed for effects of the order in which the hearing-aid setting was first dispensed to a participant. One might hypothesize that participants will perform best with the hearing-aid setting (linear vs. ILL) to which they first became accustomed, in part because it was their first experience, but also because they wore the hearing aid with that setting for 8 weeks, versus 4 Table 4. Standard deviations for CST percent-correct scores for participants with mild, moderate, and severe hearing loss. Processing condition and presentation level ILL Linear Stimulus condition 60 d SPL 75 d SPL 60 d SPL 75 d SPL Mild hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Moderate hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Severe hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise

12 Humes: enefit From Linear Hearing ids 677 Table 5. Summary of NOV results for ILL-linear differences for all 110 participants combined (all p values >.05). Effect F value (df = 1, 109) Constant (ILL-linear difference > or < 0) 1.07 Material (NU-6, CST) 0.30 Noise (cafeteria noise or babble) 0.10 SNR (signal-to-noise ratio, +5 or +10 d) 0.66 Sound level (sound level, 60 or 75 d SPL) 0.14 Material by noise 0.09 Material by SNR 0.48 Material by sound level 0.22 Noise by SNR 0.55 Noise by sound level 0.20 SNR by sound level 0.06 Material by noise by SNR 0.41 Material by noise by sound level 3.52 Material by SNR by sound level 1.43 Noise by SNR by sound level 0.86 Material by noise by SNR by sound level 0.02 Figure 6. Mean ratings on the MELE test for the linear (cross-hatched bars) and ILL (unfilled bars) settings of the hearing aid. Panels are organized from top to bottom according to hearing loss group such that participants with mild hearing loss are in the top two panels, those with moderate hearing loss are in the middle two panels, and those with severe hearing loss are in the bottom two panels. ll lefthand panels contain data for the 60-d-SPL presentation level, whereas all right-hand panels contain data for the 75-d-SPL presentation level. Within each panel, results for babble background at +5 and +10 d signal-to-noise ratio, followed by cafeteria-noise background at +5 and +10 d signal-to-noise ratio, are plotted along the x-axis as one moves from left to right. weeks for the second setting. To address this issue, six separate NOVs were performed: one for each of the three benefit measures (NU6, CST, and MELE) in each of two noise types (cafeteria and babble). significant (p <. 01) main effect of the initial hearing-aid setting was not observed, and no significant interactions with the initial setting were obtained in any of these NOVs. Hearing id Performance Inventory (HPI) The HPI questionnaire was used to identify the effect of the hearing aids on the participant s communication abilities in daily life. Each participant completed this questionnaire at the first and second evaluation visits, ranking the benefit provided by the hearing aid from Very Helpful (1) to Hinders Performance (5). Table 8 provides a summary of the observed HPI scores (means and standard deviations) for all hearing loss groups. nalyses using Wilcoxon Signed Ranks Tests for each participant group showed no significant difference in HPI scores for linear or ILL settings (Wilcoxon Z = 1.10, p = 0.27; Z = 1.59, p = 0.11; Z = 1.53, p = 0.12; for mild, moderate, and severe hearing-loss groups, respectively). The grand mean scores across all hearing loss groups were 2.10 for the ILL setting and 2.02 for the linear setting, indicating that both circuits were generally considered to be helpful to the participants. (On the HPI, a rating of 1 is associated with a label of very helpful and a rating of 2 with a label of helpful. ) The difference in grand mean HPI scores for the linear and ILL settings, collapsed across all three subject groups, however, approached statistical significance (Z = 2.36, p = 0.02).

13 678 JSLHR, Volume 40, , June 1997 Table 6. Standard deviations for MELE ratings for completed participants with mild, moderate, and severe hearing loss. Processing condition and presentation level ILL Linear Stimulus condition 60 d SPL 75 d SPL 60 d SPL 75 d SPL Mild hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Moderate hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Severe hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Paired-Comparison Test fter completion of the second round of objective and subjective measures of performance at the second evaluation visit, the participant was placed in the sound field and presented with continuous discourse to determine participant preference for either the ILL or linear setting. The test was performed at +5 and +10 d S/N in babble and cafeteria noise, with the speech signal presented at 60 and 75 d SPL. During this test the audiologist switched the circuit back and forth from the ILL setting to the linear setting randomly. For each of the eight listening conditions (2 noise types x 2 S/N values x 2 sound levels), the participant chose the circuit he/she preferred. Each listening condition was repeated four times, and the mean preference rating for the final three ratings was computed (first rating discarded as practice). For this task, the hearing-aid settings were rated on a scale of 1 to 6 by the participants, with ratings of 1, 2, and 3 representing varying degrees of preference for the ILL setting and ratings of 4, 5, and 6 representing varying degrees of preference for the linear setting. mean value of 3.49 or below, therefore, meant that the participant preferred the ILL setting, and a value of 3.51 or above meant that the participant preferred the linear setting. The overall mean result for the pairedcomparison test (3.57) showed no group preference for either the linear or ILL setting (z = , p = 0.442). Hearing-id Use Table 9 provides a summary of the average reported hearing-aid wear times for participants with mild, moderate, and severe hearing loss. Participants with severe hearing loss reported wearing their hearing aids longer (about 9 hours/day) than participants in the mild and moderate groups (about 7 hours/day). Group Data Unaided vs. ided Performance Figures 7 and 8 provide summaries of the mean data for the NU-6 and CST with regard to unaided and aided conditions for the 110 participants of this study with mild, moderate, or severe hearing loss, respectively. Table 10 provides the standard deviations for the unaided scores appearing in these two figures. Since aided performance with the linear and ILL circuits was found to be equivalent in the previous analyses, the linear setting was selected arbitrarily to represent aided listening, and the standard deviations for these data appeared previously in Tables 3 and 4. The less well-established MELE measures are not considered in detail here as measures of benefit. Rather, the focus is on the more conventional measures of benefit (NU-6, CST, and HPI). When a repeated-measures NOV was calculated for the aided-minus-unaided differences collapsed across all three subject groups (for arcsine-transformed NU-6 and CST data), differences were found to be significantly

14 Humes: enefit From Linear Hearing ids 679 Table 7. Summary of NOV for MELE results. Effect F (df) HL group (mild, moderate, severe) 2.22 (2,107) Noise (cafeteria noise, babble) (1,107)* HL group by noise 0.33 (2,107) Setting (ILL, linear) 0.09 (1,107) HL group by setting 0.29 (2,107) Snr (+5, +10 d) (1,107)* HL group by snr 1.24 (2,107) Sndlev (60, 75 d SPL) 5.92 (1,107) HL group by sndlev 3.84 (2,107) Noise by setting 1.44 (1,107) HL group by noise by setting 7.77 (2,107)* Noise by snr (1,107)* HL group by noise by snr 0.62 (2,107) Noise by sndlev (1,107)* HL group by noise by sndlev 1.06 (2,107) Setting by snr 0.01 (1,107) HL group by setting by snr 0.36 (2,107) Setting by sndlev 1.29 (1,107) HL group by setting by sndlev 0.98 (2,107) Snr by sndlev 9.45 (1,107)* HL group by snr by sndlev 1.90 (2,107) Noise by setting by snr 1.86 (1,107) HL group by noise by setting by snr 1.66 (2,107) Noise by setting by sndlev 1.51 (1,107) HL group by noise by setting by sndlev 0.56 (2,107) Noise by snr by sndlev 5.73 (1,107) HL group by noise by snr by sndlev 0.60 (2,107) Setting by snr by sndlev 3.94 (1,107) HL group by setting by snr by sndlev 1.80 (2, 107) Noise by setting by snr by sndlev 0.11 (1,107) HL group by noise by setting by snr by sndlev 0.18 (2,107) *p <.01 greater than zero (F = , df = 1,109, p <.01). In particular, aided scores were higher than unaided scores, which led to statistically significant benefit. To analyze the effects of each independent variable on the aidedminus-unaided difference scores, a second mixed-model NOV was performed. The results are summarized in Table 11. s can be seen in this table, main effects of test material (NU-6, CST) and noise type (cafeteria, babble) were significant (p <. 01) and unconfounded by Table 8. Summary of means and standard deviations (in parentheses) for the bbreviated Hearing id Performance Inventory (HPI) for participants with mild, moderate, or severe hearing loss. Hearing aid mode Hearing-loss group ILL setting Linear setting Mild 2.21 (0.56) 2.16 (0.55) Moderate 2.01 (0.53) 1.95 (0.47) Severe 2.11 (0.50) 1.88 (0.49) any interactions with other independent variables. idedminus-unaided differences were greater in cafeteria noise than in babble and for the CST than for the NU-6 test. In addition to these main effects, the three-way interaction of hearing-loss group, signal-to-noise ratio, and presentation level was also significant. From visual inspection of Figures 7 and 8, this three-way interaction appears to be attributable to the effect of signal-to-noise ratio on the aided-minus-unaided difference at various presentation levels across the three groups. In particular, benefit on both the NU-6 test (Figure 7) and the CST (Figure 8) increases as signal-to-noise ratio increases from +5 to +10 d, except for the groups with mild or moderate hearing loss at the highest presentation level (75 d). Finally, it should be noted that, given the lack of any significant differences in aided scores between the linear and ILL circuits described previously, significant benefit (aided scores minus unaided scores) exists for both circuits, even though the linear circuit was arbitrarily selected as the circuit for the aided condition in these analyses of the benefit provided by the hearing aid. Individual Data Figure 9 provides scatterplots of the individual data for all 110 participants on the NU-6 test. In the top panel, percent-correct scores for the linear setting of the hearing aid are plotted against the corresponding percent-correct score for the ILL setting. The solid lines represent the diagonal (equality of the two scores) and 95% critical-difference boundaries from Thornton and Raffin (1978) for a test consisting of 50 items. Note that the vast majority of the data points plotted fall within the 95% critical differences, with approximately equal numbers of points falling below the boundaries as above them. In contrast, the bottom panel of Figure 9 shows the aided NU-6 scores for the linear condition plotted as a function of the unaided scores. Notice that a much greater number of data points falls above the upper 95%-critical-difference boundary than below the lower boundary. Thus, these individual data are consistent with the analyses of the group results in that little difference was observed between the linear and ILL settings (top panel), Table 9. Mean reported wearing time (in hours per day) for participants by hearing loss group for ILL and linear hearing-aid settings. Hearing aid mode Hearing-loss group ILL setting Linear setting Mild Moderate Severe

15 680 JSLHR, Volume 40, , June 1997 Figure 7. Mean percent-correct scores on the NU-6 test for the unaided (unfilled bars) and aided (linear; cross-hatched bars) settings of the hearing aid. Panels are organized from top to bottom according to hearing loss group such that participants with mild hearing loss are in the top two panels, those with moderate hearing loss are in the middle two panels, and those with severe hearing loss are in the bottom two panels. ll left-hand panels contain data for the 60-d-SPL presentation level, whereas all right-hand panels contain data for the 75-d-SPL presentation level. Within each panel, results for babble background at +5 and +10 d signal-to-noise ratio, followed by cafeteria-noise background at +5 and +10 d signal-to-noise ratio, are plotted along the x-axis as one moves from left to right. Figure 8. Mean percent-correct scores on the CST for the unaided (unfilled bars) and aided (linear; cross-hatched bars) settings of the hearing aid. Panels are organized from top to bottom according to hearing loss group such that participants with mild hearing loss are in the top two panels, those with moderate hearing loss are in middle the two panels, and those with severe hearing loss are in the bottom two panels. ll left-hand panels contain data for the 60-d-SPL presentation level, whereas all right-hand panels contain data for the 75-d-SPL presentation level. Within each panel, results for babble background at +5 and +10 d signal-tonoise ratio, followed by cafeteria-noise background at +5 and +10 d signal-to-noise ratio, are plotted along the x-axis as one moves from left to right.

16 Humes: enefit From Linear Hearing ids 681 Table 10. Standard deviations for unaided NU-6 and unaided CST scores for participants with mild, moderate, and severe hearing loss. Test material and presentation level NU-6 NU-6 CST CST Stimulus condition 60 d SPL 75 d SPL 60 d SPL 75 d SPL Mild hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Moderate hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise Severe hearing loss group +5 d S/N in abble d S/N in abble d S/N in Cafeteria Noise d S/N in Cafeteria Noise but many individuals demonstrated significantly higher aided scores than unaided scores (bottom panel). The latter conclusion appears to be particularly true for unaided NU-6 scores less than or equal to 40%. Figure 10 provides a comparable set of scatterplots for the CST percent-correct scores. The top panel again shows comparison of aided scores for the linear and ILL conditions, whereas the bottom panel compares aided (linear) to unaided performance. lthough there is more scatter than in the NU-6 data, the same trends are apparent. Specifically, the two aided conditions do not appear to differ systematically (top panel), whereas aided scores are frequently significantly greater than unaided scores (bottom panel). gain, because there were no significant differences between aided scores for the linear and ILL circuits, scatterplots equivalent to those in Figures 8 and 9 would have resulted if the results from the ILL circuit, rather than the linear circuit, had been selected to represent aided listening in these figures. voluminous amount of data was gathered in the course of this study not only in terms of sample size, but in terms of the number of outcome measures obtained. Speech-recognition performance alone was measured in 16 different conditions. How do these data compare with those in the literature? Unfortunately, there are no norms available for these measures under the same combinations of noise type, speech level, and signal-to-noise ratio. To gain some insight into the validity of these results, Speech Intelligibility Index (SII; NSI S3.79 draft V3.1) calculations were performed. The SII is essentially an updated version of the rticulation Index (I) and is sensitive to factors manipulated in this study, including noise spectrum, noise level, speech level, hearing loss, and hearing-aid gain. SII calculations were performed only for the babble background for which it could be reasonably assumed that the speech and babble had the same long-term average spectrum. Moreover, it was assumed that the average speech spectrum and the normal vocal effort included in the SII standard were reasonably representative of the 60-d SPL speech level for both the NU-6 and CST talkers. Figures 11 and 12 depict the mean results from this study for the NU-6 test and the CST, respectively, plotted as a function of the SII. estfitting polynomials fit to the data in each figure (solid lines) indicate that the data are consistent with SII predictions in that performance increases monotonically with the SII. In the absence of adequate norms, this analysis assists in establishing the validity of these data. Discussion No statistically significant differences in performance were found between the ILL-processing and linear settings for any of the outcome measures used in this clinical investigation. This is consistent with a number of previous reports from smaller scale laboratory studies (see review by Fabry, 1991). Essentially, in situations in which the speech and noise are broadband stimuli, ILL

17 682 JSLHR, Volume 40, , June 1997 Table 11. Summary of NOV on aided-unaided differences. Effect F (df) HL group (mild, moderate, severe) 8.42 (2,107)* Material (NU-6, CST) (1,107)* HL group by material 0.73 (2,107) Noise (cafeteria noise, babble) (1,107)* HL group by noise 4.17 (2,107) Snr (+5, +10 d) (1,107)* HL group by snr 1.18 (2,107) Sndlev (60, 75 d SPL) (1,107)* HL group by sndlev (2,107)* Material by noise 0.28 (1,107) HL group by material by noise 0.84 (2,107) Material by snr 0.51 (1,107) HL group by material by snr 0.34 (2,107) Material by sndlev 0.83 (1,107) HL group by material by sndlev 2.62 (2,107) Noise by snr 4.35 (1,107) HL group by noise by snr 1.38 (2,107) Noise by sndlev 2.28 (1,107) HL group by noise by sndlev 4.45 (2,107) Snr by sndlev 9.31 (1,107)* HL group by snr by sndlev (2,107)* Material by noise by snr 0.65 (1,107) HL group by material by noise by snr 0.10 (2,107) Material by noise by sndlev 1.68 (1,107) HL group by material by noise by sndlev 1.43 (2,107) Material by snr by sndlev 0.25 (1,107) HL group by material by snr by sndlev 0.48 (2,107) Noise by snr by sndlev 2.39 (1,107) HL group by noise by snr by sndlev 0.09 (2,107) Material by noise by snr by sndlev 0.30 (1,107) HL group by material by noise by snr by sndlev 1.12 (2,107) Figure 9. Scatterplots of NU-6 percent-correct scores for individual participants. The top panel plots the aided score for the linear setting against the aided score for the ILL setting, whereas the bottom panel plots aided score for the linear setting against the unaided score. Triangles represent scores obtained in babble, whereas squares represent those obtained in cafeteria noise. There are a total of 880 datapoints in each panel (440 for babble and 440 for cafeteria noise) for the 110 participants in the study. Solid lines in each panel represent diagonal and 95% critical-difference boundaries above and below the diagonal (Thornton & Raffin, 1978). *p <.01 processing will provide level-dependent changes in gain that affect both the speech and the noise equivalently and do not improve the signal-to-noise ratio in any frequency region. Were the noise predominantly low-frequency in nature, however, one might expect different results. Perhaps of greater importance, however, is the consistent observation, across several outcome measures, that there was significant benefit provided by these hearing aids when operating in either the linear or ILLprocessing mode compared to unaided listening. For both the NU-6 and CST measures, for example, the difference between unaided and aided scores was significantly greater than zero. Moreover, the participants reported using their hearing aids 7 9 hours per day, on average, supporting a self-perceived benefit associated with the use of their hearing aids. That is, they would probably not wear the hearing aids much if they failed to perceive significant benefit. Recall, however, that participants were instructed to wear their hearing aids at least 5 hours per day, 5 days per week, as participants in this study. The extent to which these instructions could have biased the participants estimates of hearing-aid use is unclear. The validity of the estimates of hearing-aid use is reinforced, however, by the fact that participants with severe hearing loss reported greater daily usage than those with milder hearing loss and that many participants

18 Humes: enefit From Linear Hearing ids 683 Figure 10. Scatterplots of CST percent-correct scores for individual participants. The top panel plots the aided score for the linear setting against the aided score for the ILL setting, whereas the bottom panel plots the aided score for the linear setting against the unaided score. Triangles represent scores obtained in babble, whereas squares represent those obtained in cafeteria noise. There are a total of 880 datapoints in each panel (440 for babble and 440 for cafeteria noise) for the 110 participants in this study. Solid lines in each panel represent diagonal and 95% critical-difference boundaries above and below the diagonal (Thornton & Raffin, 1978). Figure 11. Scatterplot of mean NU-6 scores obtained in babble for each hearing-loss subgroup plotted as a function of the Speech Intelligibility Index (SII). ll unfilled symbols represent unaided listening conditions, and filled symbols represent aided listening. The solid line represents the best-fitting 3rd-order polynomial, whereas the dotted lines represent 95% prediction intervals, or population confidence intervals, above and below the best-fitting polynomial. The best-fitting polynomial accounts for 85.2% of the variance. reported daily usage that was either greater or less than the minimum usage targeted in the audiologist s instructions. Finally, the HPI provided additional support for significant benefit in that, on average, participants rated their hearing aids as being helpful. Thus, this large-scale study documents significant benefit provided by well-fit linear amplification, including significant improvements in speech-recognition performance in noise. Moreover, equivalent benefit was measured for the linear and ILLprocessing circuits. lthough ILL processing did not result in aided performance exceeding that achieved with well-fit linear amplification, both circuits provided significant, measurable, and equivalent benefit across a wide range of measures and conditions. It is important to recognize the importance of the term well-fit in the preceding sentence. The real-ear gain measured in these participants was in close agreement with that prescribed by the NL-R procedure out to a frequency of 4000 Hz, thereby making a wide portion of the speech spectrum audible. The importance of the region from 2000 to 4000 Hz to the understanding of speech in noise by hearing-impaired listeners has been emphasized previously (e.g., Lee & Humes, 1993). Moreover, output limiting was adjusted so that SSPL90 was maintained below an uncomfortable level for each patient. lthough this may not have had a direct bearing on the measures of objective benefit, it most likely had a significant impact on the amount of time the hearing aids were worn by the participants. wide range of listening conditions was sampled in this study: a total of 16 measures of speech recognition in noise, resulting in 16 objective measures of benefit

19 684 JSLHR, Volume 40, , June 1997 Figure 12. Scatterplot of mean CST scores obtained in babble for each hearing-loss subgroup plotted as a function of the Speech Intelligibility Index (SII). ll unfilled symbols represent unaided listening conditions, and filled symbols represent aided listening. The solid line represents the best-fitting 2nd-order polynomial, whereas the dotted lines represent 95% prediction intervals, or population confidence intervals, above and below the best-fitting polynomial. The best-fitting polynomial accounts for 80.6% of the variance. per participant. When attempting to determine individual differences in the benefit provided by the hearing aids, the large number of measures poses some challenges in defining benefit for individual participants. For example, should one consider a participant to have conclusively demonstrated significant benefit only when showing significant benefit on all 16 speech-recognition measures or is 8 out of 16 adequate? Perhaps one should only care that the participant showed significant improvement on any one condition of the 16 sampled. That one condition, for instance, could prove to be a highly important and frequently encountered listening situation for the participant. Our approach to determining the number of individual participants who demonstrated significant benefit was the following. First, the data were grouped by test material (NU-6, CST), hearing-loss category (mild, moderate, severe), and speech presentation level (60, 75 d SPL). For each of these subgroups, a total of four listening conditions remained (two SNRs and two noise types). The percentage of participants showing significant (p <.05) benefit on at least two of these four conditions was then determined. To evaluate significance, the individual scores were converted to rationalized arcsine units (RUs; Studebaker, 1985), which enable the use of a constant- RU difference score as a criterion for significance. For the NU-6 test, a RU difference of 18.1 was used to establish significance (Studebaker, 1985), whereas Cox et al. (1988) recommend a 15.5 RU difference-score criterion for the CST. Table 12 provides the results of this analysis. Each entry in the table represents the percentage of participants showing significant improvements for two of the four conditions included in each subset of conditions represented in the table. It is apparent from the values in this table that about 60 80% of the participants showed significant improvement in at least half of the conditions when speech is presented at normal conversational levels (60 d). This statement applies across test materials and hearing-loss categories, although there appear to be some systematic variations in this percentage with hearing-loss category. t the raised speech level (75 d), a lower percentage of each hearing-loss group demonstrated significant improvement than at normal conversational levels. t the higher speech level, moreover, the effects of speech material (NU-6, CST) and hearing-loss category on the observed percentages are more apparent. It is probably not too surprising that fewer participants demonstrated significant benefit for raised speech levels than for normal levels. Recall that benefit is the relative difference between unaided and aided performance. t higher speech levels, listeners especially those having milder impairments will obtain higher unaided scores and consequently will have less room for improvement when the hearing aid is worn. Even so, approximately 20 60% of listeners with moderate or severe hearing loss demonstrated significant benefit for speech recognition in noise in this study at the higher speech presentation level. cknowledgments This work was supported, in part, by research contracts provided by Dahlberg, Inc. to the investigators at each of the clinical sites. The first author was not one of the clinical investigators for the study, but was hired by the study sponsor as a paid consultant while the study was underway to oversee the data collection, analyses, and internal reporting. The authors would like to express appreciation for their support to all those involved in this project at Dahlberg, Inc. especially Melanie Raska and Tom Scheller (presently with another hearing-aid company); those at ausch & Lomb Incorporated especially Eric nkerud, Heather ornemann, and Tom Crescuillo; and those involved at the participating clinical sites especially Tara Thomas. In addition, two individuals at the University of Iowa deserve special mention for their significant contributions to this project. First, Don Schum, presently working for another hearing-aid manufacturer, played a major role in the design and development of the clinical protocol used in this study. Second, aron Parkinson, who assumed Don s on-site responsibilities at the University of Iowa when Don left Iowa. Finally, preparation of this work for publication was supported, in part, by a grant from the National Institute on ging to the first author and, in part, by the Retirement Research Foundation.

20 Humes: enefit From Linear Hearing ids 685 Table 12. Percentage of participants in each hearing-loss group who showed significant (p <.05) benefit (aided scores > unaided scores) on at least half of the conditions (two of four) tested for the specified test material and speech presentation level. References Presentation level Hearing-loss group 60 d SPL 75 d SPL NU-6 test material Mild Moderate Severe CST test material Mild Moderate Severe merican National Standards Institute. (1987). Specification of Hearing id Characteristics (NSI S ). New York: uthor. merican National Standards Institute. (1989). merican National Standard Specification for udiometers (NSI S ). New York: uthor. merican National Standards Institute. (in press). merican National Standard Methods for the Calculation of the Speech Intelligibility Index (NSI S3.79-draft Version 3.1). New York: uthor. ess, F. H., & Humes, L. E. (1995). udiology: The fundamentals, 2nd edition. altimore: Williams & Wilkins. yrne, D., & Dillon, H. (1986). The National coustic Laboratories (NL) new procedure for selecting the gain and frequency response of a hearing aid. Ear and Hearing, 7, Cox, R. M., & lexander, G. C. (1992). Maturation of hearing aid benefit: Objective and subjective measurements. Ear and Hearing, 13, Cox, R., lexander, G., Gilmore, C., & Pusakalich, K. M. (1988). Use of the Connected Speech Test (CST) with hearing impaired listeners. Ear and Hearing, 9, Fabry, D.. (1991). Programmable and automatic noise reduction in hearing aids. In G.. Studebaker, F. H. ess, & L.. eck (Eds)., The Vanderbilt Hearing-id Report II (pp ). Parkton, MD: York Press. Fabry, D., & Walden,. (1990). Noise reduction hearing aids. sha, 32, French-St. George, M., Engebretsen,., & O Connell, M. (1992). ehavioral assessment of CID s benchtop (version 2) digital hearing aid: Noise reduction. Journal of the merican cademy of udiology, 3, Gatehouse, S. (1992). The timecourse and magnitude of peripheral acclimatization to frequency responses: Evidence from monaural fitting of hearing aids. Journal of the coustical Society of merica, 92, Gatehouse, S. (1993). The role of perceptual acclimatization in the selection of frequency responses for hearing aids. Journal of the merican cademy of udiology, 4, Hawkins, D.., & Naidoo, S. V. (1993). Comparison of sound quality and clarity with asymmetrical peak clipping and output limiting compression. Journal of the merican cademy of udiology, 4, Hawkins, D.., Walden,., Montgomery,., & Prosek, R. (1987). Description and validation of an LDL procedure designed to select SSPL90. Ear and Hearing, 8, Kirk, R. E. (1968). Experimental design: Procedures for the behavioral sciences. elmont, C: Wadsworth. Lee, L. W., & Humes, L. E. (1993). Evaluation of a speechreception threshold model for hearing-impaired listeners. Journal of the coustical Society of merica, 93, Mueller, H. G. (1992). Insertion gain measurements. In H. G. Mueller, D.. Hawkins, & J. L. Northern (Eds.), Probe microphone measurements (pp ). San Diego: Singular Publishing Group, Inc. Ono, H., Kanzaki, J., & Mizoi, K. (1983). Clinical results of hearing aid noise-level-controlled selective amplification. udiology, 22, Pearsons, K. S., ennett, R. L., & Fidell, S. (1977). Speech Levels in Various Noise Environments (U.S. EP Report # EP-600/ ). Punch, J., & eck, L.. (1986). Relative effects of lowfrequency amplification on syllable recognition and speech quality. Ear and Hearing, 7, Schum, D. (1993). Test-retest reliability of a shortened version of the Hearing id Performance Inventory. Journal of the merican cademy of udiology, 4, Skinner, M. (1988). Hearing aid evaluation. Englewood Cliffs, NJ: Prentice Hall. Studebaker, G.. (1985). rationalized arcsine transform. Journal of Speech and Hearing Research, 28, Thornton,., & Raffin, M. J. M. (1978). Speech discrimination scores modeled as a binomial variable. Journal of Speech and Hearing Research, 21, Tillman, T. W., & Carhart, R. (1966). n Expanded Test for Speech Discrimination Utilizing CNC Monosyllabic Words (Northwestern University uditory Test No. 6, Technical Report, SM-TR-66-55). rooks ir Force ase, TX: USF School of erospace Medicine. Turner, C. W., Humes, L. E., entler, R.., & Cox, R. M. (1996). review of past research on changes in hearing aid benefit over time. Ear and Hearing, 17, 14S 25S. Walden,. E., Demorest, M. E., & Hepler, E. L. (1984). Self-report approach to assessing benefit derived from amplification. Journal of Speech and Hearing Research, 27, Received July 10, 1996 ccepted January 10, 1997 Contact author: Larry E. Humes, PhD, Department of Speech and Hearing Sciences, Indiana University, loomington, IN 47405

Prescriptive Fitting Algorithms: New and Improved! Michelle Mattingly, Au.D. and Lisa Perhacs, Au.D.

Prescriptive Fitting Algorithms: New and Improved! Michelle Mattingly, Au.D. and Lisa Perhacs, Au.D. 1 Prescriptive Fitting Algorithms: New and Improved! Michelle Mattingly, Au.D. and Lisa Perhacs, Au.D. TM 2 What is the goal of a prescriptive algorithm? Optimize audibility Optimize speech intelligibility

More information

The NAL Percentage Loss of Hearing Scale

The NAL Percentage Loss of Hearing Scale The NAL Percentage Loss of Hearing Scale Anne Greville Audiology Adviser, ACC February, 2010 The NAL Percentage Loss of Hearing (PLH) Scale was developed by John Macrae of the Australian National Acoustic

More information

Does premium listening require premium hearing aids?

Does premium listening require premium hearing aids? Does premium listening require premium hearing aids? Effectiveness of basic and premium hearing aids on speech understanding and listening effort outcomes. Jani Johnson, Jingjing Xu, Robyn Cox Presented

More information

Benefit Acclimatization in Elderly Hearing Aid Users

Benefit Acclimatization in Elderly Hearing Aid Users J Am Acad Audiol 7 : 428-441 (1996) Benefit Acclimatization in Elderly Hearing Aid Users Robyn M. Cox*t Genevieve C. Alexandert Izel M. Taylor' Ginger A. Gray* Abstract A previous study from this laboratory

More information

Long-Term Findings from the NIDCD/VA Hearing Aid Clinical Trial. Gene W. Bratt, Ph.D, Chair gene.bratt@med.va.gov

Long-Term Findings from the NIDCD/VA Hearing Aid Clinical Trial. Gene W. Bratt, Ph.D, Chair gene.bratt@med.va.gov Long-Term Findings from the NIDCD/VA Hearing Aid Clinical Trial Gene W. Bratt, Ph.D, Chair gene.bratt@med.va.gov Demographics and Audiometrics CSP #418-A Gene W. Bratt, Ph.D. Study Chair Barbara F. Peek,

More information

PURE TONE AUDIOMETRY Andrew P. McGrath, AuD

PURE TONE AUDIOMETRY Andrew P. McGrath, AuD PURE TONE AUDIOMETRY Andrew P. McGrath, AuD Pure tone audiometry is the standard behavioral assessment of an individual s hearing. The results of pure tone audiometry are recorded on a chart or form called

More information

Maturation of Hearing Aid Benefit: Objective and Subjective Measurements

Maturation of Hearing Aid Benefit: Objective and Subjective Measurements Amplification and Aural Rehabilitation Maturation of Hearing Aid Benefit: Objective and Subjective Measurements Robyn M. Cox, PhD; Genevieve C. Alexander, MA Department of Audiology and Speech Pathology,

More information

NEW YORK STATE MEDICAID PROGRAM HEARING AID PRIOR APPROVAL GUIDELINES

NEW YORK STATE MEDICAID PROGRAM HEARING AID PRIOR APPROVAL GUIDELINES NEW YORK STATE MEDICAID PROGRAM HEARING AID PRIOR APPROVAL GUIDELINES TABLE OF CONTENTS Section I - Purpose Statement... 2 Section II Instructions for Obtaining Prior Approval... 3 Prior Approval Form

More information

A Hearing Aid Primer

A Hearing Aid Primer A Hearing Aid Primer Includes: Definitions Photographs Hearing Aid Styles, Types, and Arrangements WROCC Outreach Site at Western Oregon University This hearing aid primer is designed to define the differences

More information

The Accuracy of 0 db HL as an Assumption of Normal Hearing

The Accuracy of 0 db HL as an Assumption of Normal Hearing The Accuracy of 0 db HL as an Assumption of Normal Hearing Introduction An operating assumption of diagnostic audiology is that hearing level of a young adult with no known hearing loss or history of noise

More information

5th Congress of Alps-Adria Acoustics Association NOISE-INDUCED HEARING LOSS

5th Congress of Alps-Adria Acoustics Association NOISE-INDUCED HEARING LOSS 5th Congress of Alps-Adria Acoustics Association 12-14 September 2012, Petrčane, Croatia NOISE-INDUCED HEARING LOSS Davor Šušković, mag. ing. el. techn. inf. davor.suskovic@microton.hr Abstract: One of

More information

How To Test If A Cell Phone With Selective Amplification Improves Speech Recognition And Hearing Perception

How To Test If A Cell Phone With Selective Amplification Improves Speech Recognition And Hearing Perception J Am Acad Audiol 20:109 118 (2009) Evaluation of Cellular Phone Technology with Digital Hearing Aid Features: Effects of Encoding and Individualized Amplification DOI: 10.3766/jaaa.20.2.4 Carol L. Mackersie*

More information

Wisconsin Department of Health & Family Services Division of Disability and Elder Services Bureau of Aging & Long Term Care Resources

Wisconsin Department of Health & Family Services Division of Disability and Elder Services Bureau of Aging & Long Term Care Resources Waiver Wise Technical Assistance for the Community Options Program Waiver COP-W Wisconsin Department of Health & Family Services Division of Disability and Elder Services Bureau of Aging & Long Term Care

More information

Early vs. Late Onset Hearing Loss: How Children Differ from Adults. Andrea Pittman, PhD Arizona State University

Early vs. Late Onset Hearing Loss: How Children Differ from Adults. Andrea Pittman, PhD Arizona State University Early vs. Late Onset Hearing Loss: How Children Differ from Adults Andrea Pittman, PhD Arizona State University Heterogeneity of Children with Hearing Loss Chronological age Age at onset Age at identification

More information

Edited by Victor de Andrade Page 1

Edited by Victor de Andrade Page 1 Guidelines for Hearing Aid Selection and Fitting for Adults This document serves to function as a guideline for hearing care professionals who select and dispense hearing aids to adult patients. Its purpose

More information

HEARING SCREENING FOR CHILDREN

HEARING SCREENING FOR CHILDREN PURPOSE Audiologic (hearing) screening identifies those persons who are likely to have hearing impairments or disorders that may interfere with body function/structure and/or activity/participation as

More information

Everybody has the right to

Everybody has the right to Everybody has the right to good hearing The Guide TO Hearing satisfaction for more people Beltone was founded on the act of helping a friend to enjoy life more. Since 19, we have provided knowledge, tools,

More information

Using Your Fitting Software This guide provides comprehensive, task-based information about all the fitting software features.

Using Your Fitting Software This guide provides comprehensive, task-based information about all the fitting software features. Gravity Fitting Software User's Manual part #: S0273-01 Rev A Using Your Fitting Software This guide provides comprehensive, task-based information about all the fitting software features. You may access

More information

11. Analysis of Case-control Studies Logistic Regression

11. Analysis of Case-control Studies Logistic Regression Research methods II 113 11. Analysis of Case-control Studies Logistic Regression This chapter builds upon and further develops the concepts and strategies described in Ch.6 of Mother and Child Health:

More information

A PROFESSIONAL PRACTICE PROFILE

A PROFESSIONAL PRACTICE PROFILE A PROFESSIONAL PRACTICE PROFILE FOR HEARING HEALTH PROFESSIONALS The International Hearing Society has adopted the following practice profile as a comprehensive declaration of dispensing characteristics

More information

ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING

ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING Dennis P. Driscoll, P.E. and David C. Byrne, CCC-A Associates in Acoustics, Inc. Evergreen, Colorado Telephone (303)

More information

Magnitude of Hearing Loss and Open Ear Fittings

Magnitude of Hearing Loss and Open Ear Fittings Magnitude of Hearing Loss and Open Ear Fittings Elizabeth Talmage, AuD 1 Mia Rosenfeld, PhD 2 Ginny Alexander, MS 3 1 Bay Pines VA Healthcare System, Bay Pines, FL 2 VA Tennessee Valley Healthcare System,

More information

NUCLEAR SECURITY - GATB, WCT, PREP, Vision and Hearing

NUCLEAR SECURITY - GATB, WCT, PREP, Vision and Hearing NUCLEAR SECURITY - GATB, WCT, PREP, Vision and Hearing 1. GENERAL APTITUDE TEST BATTERY (GATB) Description: The GATB is a timed pencil and paper test which evaluates arithmetic, problem solving, spatial

More information

Auditory measures of selective and divided attention in young and older adults using single-talker competition

Auditory measures of selective and divided attention in young and older adults using single-talker competition Auditory measures of selective and divided attention in young and older adults using single-talker competition Larry E. Humes, Jae Hee Lee, and Maureen P. Coughlin Department of Speech and Hearing Sciences,

More information

Workshop Perceptual Effects of Filtering and Masking Introduction to Filtering and Masking

Workshop Perceptual Effects of Filtering and Masking Introduction to Filtering and Masking Workshop Perceptual Effects of Filtering and Masking Introduction to Filtering and Masking The perception and correct identification of speech sounds as phonemes depends on the listener extracting various

More information

Minimum Hearing Loss Threshold (MHLT)

Minimum Hearing Loss Threshold (MHLT) Minimum Hearing Loss Threshold (MHLT) Table of Contents Exemption Criteria to the Minimum Hearing Loss Threshold... 2 Guidelines for the Administration of the Minimum Hearing Loss Threshold 3 Criterion

More information

SOUTHERN UTAH UNIVERSITY S WRITTEN HEARING CONSERVATION PROGRAM September 24, 2015

SOUTHERN UTAH UNIVERSITY S WRITTEN HEARING CONSERVATION PROGRAM September 24, 2015 SOUTHERN UTAH UNIVERSITY S WRITTEN HEARING CONSERVATION PROGRAM September 24, 2015 1.0 PURPOSE: The objective of the Southern Utah University hearing conservation program is to minimize occupational hearing

More information

S ilman, Gelfand, and Silverman (1984)

S ilman, Gelfand, and Silverman (1984) J Am Acad Audiol 4: 285-295 (1993) Monaural Hearing Aid Effect : Case Presentations Raymond M. Hurley* Abstract A phenomenon associated with a monaural hearing aid fitting has previously been identified.

More information

By Michael Block, PhD

By Michael Block, PhD By Michael Block, PhD IHS offers a diversity of options for obtaining continuing education credit: seminars and classroom training, institutional courses, online studies, and distance learning programs.

More information

Convention Paper Presented at the 118th Convention 2005 May 28 31 Barcelona, Spain

Convention Paper Presented at the 118th Convention 2005 May 28 31 Barcelona, Spain Audio Engineering Society Convention Paper Presented at the 118th Convention 25 May 28 31 Barcelona, Spain 6431 This convention paper has been reproduced from the author s advance manuscript, without editing,

More information

ASU Speech & Hearing Clinic Hearing Aid Services

ASU Speech & Hearing Clinic Hearing Aid Services ASU Speech & Hearing Clinic Hearing Aid Services Thank you for considering the ASU Speech & Hearing Clinic for your hearing healthcare needs. The purchase and fitting of hearing instruments is a very important

More information

Interference to Hearing Aids by Digital Mobile Telephones Operating in the 1800 MHz Band.

Interference to Hearing Aids by Digital Mobile Telephones Operating in the 1800 MHz Band. Interference to Hearing Aids by Digital Mobile Telephones Operating in the 1800 MHz Band. Reference: EB968 Date: January 2008 Author: Eric Burwood (National Acoustic Laboratories) Collaborator: Walter

More information

SPEECH AUDIOMETRY. @ Biswajeet Sarangi, B.Sc.(Audiology & speech Language pathology)

SPEECH AUDIOMETRY. @ Biswajeet Sarangi, B.Sc.(Audiology & speech Language pathology) 1 SPEECH AUDIOMETRY Pure tone Audiometry provides only a partial picture of the patient s auditory sensitivity. Because it doesn t give any information about it s ability to hear and understand speech.

More information

Hearing Tests And Your Child

Hearing Tests And Your Child HOW EARLY CAN A CHILD S HEARING BE TESTED? Most parents can remember the moment they first realized that their child could not hear. Louise Tracy has often told other parents of the time she went onto

More information

What Audio Engineers Should Know About Human Sound Perception. Part 2. Binaural Effects and Spatial Hearing

What Audio Engineers Should Know About Human Sound Perception. Part 2. Binaural Effects and Spatial Hearing What Audio Engineers Should Know About Human Sound Perception Part 2. Binaural Effects and Spatial Hearing AES 112 th Convention, Munich AES 113 th Convention, Los Angeles Durand R. Begault Human Factors

More information

Hearing Tests And Your Child

Hearing Tests And Your Child How Early Can A Child s Hearing Be Tested? Most parents can remember the moment they first realized that their child could not hear. Louise Tracy has often told other parents of the time she went onto

More information

The Effects of Hearing Impairment and Aging on Spatial Processing

The Effects of Hearing Impairment and Aging on Spatial Processing The Effects of Hearing Impairment and Aging on Spatial Processing Helen Glyde, 1 3 Sharon Cameron, 1,2 Harvey Dillon, 1,2 Louise Hickson, 1,3 and Mark Seeto 1,2 Objectives: Difficulty in understanding

More information

The Disability Tax Credit Certificate Tip sheet for Audiologists

The Disability Tax Credit Certificate Tip sheet for Audiologists The Disability Tax Credit Certificate Tip sheet for Audiologists Developed by: The Canadian Academy of Audiology (CAA) & Speech- Language and Audiology Canada (SAC) Purpose of This Document The Canada

More information

The Effect of Network Cabling on Bit Error Rate Performance. By Paul Kish NORDX/CDT

The Effect of Network Cabling on Bit Error Rate Performance. By Paul Kish NORDX/CDT The Effect of Network Cabling on Bit Error Rate Performance By Paul Kish NORDX/CDT Table of Contents Introduction... 2 Probability of Causing Errors... 3 Noise Sources Contributing to Errors... 4 Bit Error

More information

Convention Paper Presented at the 112th Convention 2002 May 10 13 Munich, Germany

Convention Paper Presented at the 112th Convention 2002 May 10 13 Munich, Germany Audio Engineering Society Convention Paper Presented at the 112th Convention 2002 May 10 13 Munich, Germany This convention paper has been reproduced from the author's advance manuscript, without editing,

More information

Step-by-Step RECD Guide

Step-by-Step RECD Guide Precision Audiometric Instruments www.medrx-usa.com Step-by-Step RECD Guide The RECD task involves 4 steps: 1 - Complete Calibration of the Speakers and Probe Tube 2 - Measure an Ear Response 3 - Perform

More information

Stanford University. Hearing Conservation Program

Stanford University. Hearing Conservation Program Stanford University Hearing Conservation Program April 2006 Stanford University HEARING CONSERVATION PROGRAM CONTENTS PAGE 1.0 INTRODUCTION...1 1.1 Program Description...1 1.2 Scope...1 2.0 DEFINITIONS

More information

Dr. Abdel Aziz Hussein Lecturer of Physiology Mansoura Faculty of Medicine

Dr. Abdel Aziz Hussein Lecturer of Physiology Mansoura Faculty of Medicine Physiological Basis of Hearing Tests By Dr. Abdel Aziz Hussein Lecturer of Physiology Mansoura Faculty of Medicine Introduction Def: Hearing is the ability to perceive certain pressure vibrations in the

More information

A Microphone Array for Hearing Aids

A Microphone Array for Hearing Aids A Microphone Array for Hearing Aids by Bernard Widrow 1531-636X/06/$10.00 2001IEEE 0.00 26 Abstract A directional acoustic receiving system is constructed in the form of a necklace including an array of

More information

Hearing Aid FAQs from the American Academy of Audiology 2006 American Academy of Audiology www.audiology.org

Hearing Aid FAQs from the American Academy of Audiology 2006 American Academy of Audiology www.audiology.org Hearing Aid FAQs from the American Academy of Audiology 2006 American Academy of Audiology www.audiology.org How do I determine if I am a candidate for a hearing aid? The critical variable is whether you

More information

Sound Pressure Measurement

Sound Pressure Measurement Objectives: Sound Pressure Measurement 1. Become familiar with hardware and techniques to measure sound pressure 2. Measure the sound level of various sizes of fan modules 3. Calculate the signal-to-noise

More information

Stream Boost: All About That Bass

Stream Boost: All About That Bass Carreen Pederson, M.A., & Alyson Gruhlke, Au.D. Stream Boost is an automatic feature that activates hearing aid settings optimized for high-quality streamed audio. Stream Boost settings are not part of

More information

S-822. Noise and Hearing Conservation RISK MANAGEMENT

S-822. Noise and Hearing Conservation RISK MANAGEMENT RISK MANAGEMENT Noise and Hearing Conservation Resources and information for developing Patient handling programs in health care Employers with workplaces that have excessive levels of noise need to implement

More information

4. Have you been exposed to loud noise or listened to music with headphone in the past 24 hours? [ ] Yes [ ] No [ ] Refused [ ] Don t know

4. Have you been exposed to loud noise or listened to music with headphone in the past 24 hours? [ ] Yes [ ] No [ ] Refused [ ] Don t know PhenX Measure: Audiogram Hearing Test (#200100) PhenX Protocol: Audiogram Hearing Test (#200101) Date of Interview/Examination (MM/DD/YYYY): Preliminary Activities 1. Does the examinee have hearing aids

More information

Studies of Hearing-Aid Outcome Measures in Older Adults: A Comparison of Technologies and an Examination of Individual Differences

Studies of Hearing-Aid Outcome Measures in Older Adults: A Comparison of Technologies and an Examination of Individual Differences Studies of Hearing-Aid Outcome Measures in Older Adults: A Comparison of Technologies and an Examination of Individual Differences Larry E. Humes, Ph.D., 1 Jayne B. Ahlstrom, M.S., 2 Gene W. Bratt, Ph.D.,

More information

PROFESSIONAL BOARD FOR SPEECH, LANGUAGE AND HEARING PROFESSIONS STANDARDS OF PRACTICE IN AUDIOLOGY

PROFESSIONAL BOARD FOR SPEECH, LANGUAGE AND HEARING PROFESSIONS STANDARDS OF PRACTICE IN AUDIOLOGY The Health Professions Council of South Africa PROFESSIONAL BOARD FOR SPEECH, LANGUAGE AND HEARING PROFESSIONS STANDARDS OF PRACTICE IN AUDIOLOGY December 2002 The following are considered to be standard

More information

Humulin (LY041001) Page 1 of 1

Humulin (LY041001) Page 1 of 1 (LY041001) These clinical study results are supplied for informational purposes only in the interests of scientific disclosure. They are not intended to substitute for the FDA-approved package insert or

More information

CHAPTER 11 NOISE AND HEARING CONSERVATION PROGRAM

CHAPTER 11 NOISE AND HEARING CONSERVATION PROGRAM CHAPTER 11 NOISE AND HEARING CONSERVATION PROGRAM INTRODUCTION This program contains information on the effects, evaluation, and control of noise. For assistance in evaluating a noise problem, contact

More information

GUIDELINES FOR HEARING AID FITTING FOR ADULTS

GUIDELINES FOR HEARING AID FITTING FOR ADULTS GUIDELINES FOR HEARING AID FITTING FOR ADULTS By Rony Ganguly MSc, Clinical Scientist (Audiology) and Peter Ferguson FISHAA, FSHAA, Fd.Sc.HAA., BSc (Pharm), MPSI Introduction The aim of these protocols

More information

DAILY BIOLOGIC CHECK OF AUDIOMETERS

DAILY BIOLOGIC CHECK OF AUDIOMETERS DAILY BIOLOGIC CHECK OF AUDIOMETERS An audiometer must be checked before each day of use or if a problem is suspected with the equipment during use. After the audiometer is turned on, the operator should

More information

Optimizing IP3 and ACPR Measurements

Optimizing IP3 and ACPR Measurements Optimizing IP3 and ACPR Measurements Table of Contents 1. Overview... 2 2. Theory of Intermodulation Distortion... 2 3. Optimizing IP3 Measurements... 4 4. Theory of Adjacent Channel Power Ratio... 9 5.

More information

The impact of using real ear measures to calculate prescriptive targets on hearing aid follow-up visits

The impact of using real ear measures to calculate prescriptive targets on hearing aid follow-up visits Washington University School of Medicine Digital Commons@Becker Independent Studies and Capstones Program in Audiology and Communication Sciences 2013 The impact of using real ear measures to calculate

More information

Pediatric Fitting Guide

Pediatric Fitting Guide Pediatric Fitting Guide October 2013 / Phonak Target 3.2 This guide aims to provide a detailed introduction to Phonak Target Junior Mode for hearing care professionals who fit children of all ages. Introduction...

More information

Unilateral (Hearing Loss in One Ear) Hearing Loss Guidance

Unilateral (Hearing Loss in One Ear) Hearing Loss Guidance Unilateral (Hearing Loss in One Ear) Hearing Loss Guidance Indiana s Early Hearing Detection and Intervention Program Before universal newborn hearing screening, most children with unilateral hearing loss

More information

Acoustic Terms, Definitions and General Information

Acoustic Terms, Definitions and General Information Acoustic Terms, Definitions and General Information Authored by: Daniel Ziobroski Acoustic Engineer Environmental and Acoustic Engineering GE Energy Charles Powers Program Manager Environmental and Acoustic

More information

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few

More information

Audiologist and Hearing Aid Dispenser. Provider Manual

Audiologist and Hearing Aid Dispenser. Provider Manual Audiologist and Hearing Aid Provider Manual Provider 1 TABLE OF CONTENTS Chapter I. General Program Policies Chapter II. Member Eligibility Chapter IV. Billing Iowa Medicaid Appendix III. Provider-Specific

More information

PURE TONE AUDIOMETER

PURE TONE AUDIOMETER PURE TONE AUDIOMETER V. Vencovský, F. Rund Department of Radioelectronics, Faculty of Electrical Engineering, Czech Technical University in Prague, Czech Republic Abstract Estimation of pure tone hearing

More information

Audiometry and Hearing Loss Examples

Audiometry and Hearing Loss Examples Audiometry and Hearing Loss Examples An audiogram shows the quietest sounds you can just hear. The red circles represent the right ear and the blue crosses represent the left ear. Across the top, there

More information

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS 1. Bandwidth: The bandwidth of a communication link, or in general any system, was loosely defined as the width of

More information

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS Objective In this experiment you will study the i-v characteristics of an MOS transistor. You will use the MOSFET as a variable resistor and as a switch. BACKGROUND

More information

SmartFocus Article 1 - Technical approach

SmartFocus Article 1 - Technical approach SmartFocus Article 1 - Technical approach Effective strategies for addressing listening in noisy environments The difficulty of determining the desired amplification for listening in noise is well documented.

More information

10. Analysis of Longitudinal Studies Repeat-measures analysis

10. Analysis of Longitudinal Studies Repeat-measures analysis Research Methods II 99 10. Analysis of Longitudinal Studies Repeat-measures analysis This chapter builds on the concepts and methods described in Chapters 7 and 8 of Mother and Child Health: Research methods.

More information

January 26, 2009 The Faculty Center for Teaching and Learning

January 26, 2009 The Faculty Center for Teaching and Learning THE BASICS OF DATA MANAGEMENT AND ANALYSIS A USER GUIDE January 26, 2009 The Faculty Center for Teaching and Learning THE BASICS OF DATA MANAGEMENT AND ANALYSIS Table of Contents Table of Contents... i

More information

Hearing Devices Policy and Administration Manual

Hearing Devices Policy and Administration Manual Hearing Devices Policy and Administration Manual Assistive Devices Program, Ministry of Health and Long-Term Care Table of Amendments This page will list all substantive changes to policies and procedures

More information

AN-007 APPLICATION NOTE MEASURING MAXIMUM SUBWOOFER OUTPUT ACCORDING ANSI/CEA-2010 STANDARD INTRODUCTION CEA-2010 (ANSI) TEST PROCEDURE

AN-007 APPLICATION NOTE MEASURING MAXIMUM SUBWOOFER OUTPUT ACCORDING ANSI/CEA-2010 STANDARD INTRODUCTION CEA-2010 (ANSI) TEST PROCEDURE AUDIOMATICA AN-007 APPLICATION NOTE MEASURING MAXIMUM SUBWOOFER OUTPUT ACCORDING ANSI/CEA-2010 STANDARD by Daniele Ponteggia - dp@audiomatica.com INTRODUCTION The Consumer Electronics Association (CEA),

More information

MAPS - Bass Manager. TMH Corporation 2500 Wilshire Blvd., Suite 750 Los Angeles, CA 90057 ph: 213.201.0030 fx: 213.201.0031 www.tmhlabs.

MAPS - Bass Manager. TMH Corporation 2500 Wilshire Blvd., Suite 750 Los Angeles, CA 90057 ph: 213.201.0030 fx: 213.201.0031 www.tmhlabs. MAPS - Bass Manager TMH Corporation 2500 Wilshire Blvd., Suite 750 Los Angeles, CA 90057 ph: 213.201.0030 fx: 213.201.0031 www.tmhlabs.com Thankyou for purchasing the TMH Bass Manager. By this purchase

More information

Introduction Bone Anchored Implants (BAI), Candidacy and Pre-Operative Testing for Adult Patients

Introduction Bone Anchored Implants (BAI), Candidacy and Pre-Operative Testing for Adult Patients Introduction Bone Anchored Implants (BAI), Candidacy and Pre-Operative Testing for Adult Patients Developed by Hakanssonand his colleagues in Sweden in the late 1970s 3 Components Sound Processor (#1)

More information

Pure-Tone Assessment and Screening of Children with Middle-Ear Effusion

Pure-Tone Assessment and Screening of Children with Middle-Ear Effusion J Am Acad Audiol 5 : 173-182 (1994) Pure-Tone Assessment and Screening of Children with Middle-Ear Effusion Shlomo Silman* t Carol A. Silvermantt Daniel S. Arick Abstract The purpose of this prospective

More information

Audio Tone Control Using The TLC074 Operational Amplifier

Audio Tone Control Using The TLC074 Operational Amplifier Application Report SLOA42 - JANUARY Audio Tone Control Using The TLC74 Operational Amplifier Dee Harris Mixed-Signal Products ABSTRACT This application report describes the design and function of a stereo

More information

A Professional Practice Profile for Hearing Health Professionals

A Professional Practice Profile for Hearing Health Professionals A Professional Practice Profile for Hearing Health Professionals Hearing instrument dispensing includes several professions that overlap. There are traditional hearing aid dispensers, board certified hearing

More information

Audio Examination. Place of Exam:

Audio Examination. Place of Exam: Audio Examination Name: Date of Exam: SSN: C-number: Place of Exam: The Handbook of Standard Procedures and Best Practices for Audiology Compensation and Pension Exams is available online. ( This is a

More information

A Game of Numbers (Understanding Directivity Specifications)

A Game of Numbers (Understanding Directivity Specifications) A Game of Numbers (Understanding Directivity Specifications) José (Joe) Brusi, Brusi Acoustical Consulting Loudspeaker directivity is expressed in many different ways on specification sheets and marketing

More information

UCSC HEARING CONSERVATION PROGRAM Environment, Health & Safety (EH&S) August 2007

UCSC HEARING CONSERVATION PROGRAM Environment, Health & Safety (EH&S) August 2007 UCSC HEARING CONSERVATION PROGRAM Environment, Health & Safety (EH&S) August 2007 Table of Contents Purpose Applicable Regulations and Standards Responsibilities Deans and Directors Department Heads, Managers,

More information

Understanding Hearing Loss 404.591.1884. www.childrensent.com

Understanding Hearing Loss 404.591.1884. www.childrensent.com Understanding Hearing Loss 404.591.1884 www.childrensent.com You just found out your child has a hearing loss. You know what the Audiologist explained to you, but it is hard to keep track of all the new

More information

Guide for families of infants and children with hearing loss

Guide for families of infants and children with hearing loss With early detection, Early Intervention can begin! Guide for families of infants and children with hearing loss Birth to 3 2008 Cover photograph Geneva Marie Durgin was born January 20, 2007. She lives

More information

USER GUIDE. Thank you for choosing Ovation Hearing and congratulations on taking control of your hearing health.

USER GUIDE. Thank you for choosing Ovation Hearing and congratulations on taking control of your hearing health. Ovation Lift USER GUIDE Thank you for choosing Ovation Hearing and congratulations on taking control of your hearing health. While no hearing device can restore normal hearing health or prevent or improve

More information

SEMI-IMPLANTABLE AND FULLY IMPLANTABLE MIDDLE EAR HEARING AIDS

SEMI-IMPLANTABLE AND FULLY IMPLANTABLE MIDDLE EAR HEARING AIDS Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan. This Medical Coverage Guideline must be read in its

More information

1000+ AUDIOMETER OPERATING INSTRUCTIONS

1000+ AUDIOMETER OPERATING INSTRUCTIONS 1000+ AUDIOMETER OPERATING INSTRUCTIONS AMBCO Model 1000+ is a microprocessor controlled pure tone air conduction audiometer with automated screening test feature. INDEX A. Getting Started Patient Instructions

More information

GUIDELINES ON THE ACOUSTICS OF SOUND FIELD AUDIOMETRY IN CLINICAL AUDIOLOGICAL APPLICATIONS

GUIDELINES ON THE ACOUSTICS OF SOUND FIELD AUDIOMETRY IN CLINICAL AUDIOLOGICAL APPLICATIONS NOTE Although care has been taken in preparing the information supplied by the British Society of Audiology, the BSA does not and cannot guarantee the interpretation and application of it. The BSA cannot

More information

Guidelines for Hearing Aid Fitting for Adults

Guidelines for Hearing Aid Fitting for Adults ASHA Ad Hoc Committee on Hearing Aid Selection and Fitting Reference this material as: American Speech-Language-Hearing Association. (1998). for Hearing Aid Fitting for Adults []. Available from www.asha.org/policy.

More information

CONVENTIONAL AND DIGITAL HEARING AIDS

CONVENTIONAL AND DIGITAL HEARING AIDS CONVENTIONAL AND DIGITAL HEARING AIDS Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan. This Medical

More information

Hearing Protection Standard OSHA Standard 29 CFR 1910.95

Hearing Protection Standard OSHA Standard 29 CFR 1910.95 Hearing Protection Standard OSHA Standard 29 CFR 1910.95 The purpose of this standard is protect employees over exposure to noise and to prevent hearing loss. This standard also states that employees must

More information

Portions have been extracted from this report to protect the identity of the student. RIT/NTID AURAL REHABILITATION REPORT Academic Year 2003 2004

Portions have been extracted from this report to protect the identity of the student. RIT/NTID AURAL REHABILITATION REPORT Academic Year 2003 2004 Portions have been extracted from this report to protect the identity of the student. Sessions: 9/03 5/04 Device: N24 cochlear implant Speech processors: 3G & Sprint RIT/NTID AURAL REHABILITATION REPORT

More information

4.2. Desktop Fitting Guide. Content. August 2015

4.2. Desktop Fitting Guide. Content. August 2015 4.2 August 2015 Desktop Fitting Guide This guide provides you with a detailed introduction to latest hearing instrument fitting with Phonak Target. www.phonakpro.com/target_guide Please also find the [News]

More information

Once your baby has good head control and can turn towards something interesting, a more advanced behavioural procedure can be used.

Once your baby has good head control and can turn towards something interesting, a more advanced behavioural procedure can be used. How do we test the hearing of babies and children? An audiologist will select from a range of different tests to evaluate your child s hearing. The choice of test depends on the information that is needed

More information

Measuring the Outcomes of a National Rehabilitation Program :

Measuring the Outcomes of a National Rehabilitation Program : J Am Acad Audiol 10 : 67-79 (1999) Measuring the Outcomes of a National Rehabilitation Program : Normative Data for the Client Oriented Scale of Improvement (COSI) and the Hearing Aid User's Questionnaire

More information

Yamaha Power Amplifier. White Paper

Yamaha Power Amplifier. White Paper Yamaha Power Amplifier White Paper August 2008 Table of Contents 1. About EEEngine...2 1.1. Introduction...2 1.2. Explanation of different amplifier topologies...2 2. Yamaha technology...5 2.1. Dual mono-amplifier

More information

REGULATIONS FOR THE DEGREE OF MASTER OF SCIENCE IN AUDIOLOGY (MSc[Audiology])

REGULATIONS FOR THE DEGREE OF MASTER OF SCIENCE IN AUDIOLOGY (MSc[Audiology]) 224 REGULATIONS FOR THE DEGREE OF MASTER OF SCIENCE IN AUDIOLOGY (MSc[Audiology]) (See also General Regulations) Any publication based on work approved for a higher degree should contain a reference to

More information

UC Santa Barbara Hearing Conservation Program Manual

UC Santa Barbara Hearing Conservation Program Manual UCSB Hearing Conservation Program Manual Page 1 of 15 UC Santa Barbara Hearing Conservation Program Manual Rev. August 2014 Program Manager: Nick Nieberding Title: Industrial Hygiene Specialist Email:

More information

C HAPTER TEN. Electroacoustic Verification Measures with Modern Hearing Instrument Technology. Susan D. Scollie and Richard C. Seewald.

C HAPTER TEN. Electroacoustic Verification Measures with Modern Hearing Instrument Technology. Susan D. Scollie and Richard C. Seewald. C HAPTER TEN Electroacoustic Verification Measures with Modern Hearing Instrument Technology Susan D. Scollie and Richard C. Seewald Introduction Modern-day hearing instruments offer a wide variety of

More information

COMPARATIVE STUDY OF ANALOG AND DIGITAL HEARING AIDS

COMPARATIVE STUDY OF ANALOG AND DIGITAL HEARING AIDS COMPARATIVE STUDY OF ANALOG AND DIGITAL HEARING AIDS A Thesis Submitted to the Graduate Faculty of the Louisiana State University and Agricultural and Mechanical College in partial fulfillment of the requirements

More information

Estimation of Loudness by Zwicker's Method

Estimation of Loudness by Zwicker's Method Estimation of Loudness by Zwicker's Method Loudness is one category in the list of human perceptions of sound. There are many methods of estimating Loudness using objective measurements. No method is perfect.

More information

hi HealthInnovations Methods for Prescribing Hearing Device Gain: Reliability and Accuracy

hi HealthInnovations Methods for Prescribing Hearing Device Gain: Reliability and Accuracy hi HealthInnovations Methods for Prescribing Hearing Device Gain: Reliability and Accuracy Dianne J. Van Tasell, Ph.D. Dianne Van Tasell is Adjunct Professor in the Department of Speech, Language, and

More information

DIAGNOSTIC TESTING GUIDELINES for Audiology

DIAGNOSTIC TESTING GUIDELINES for Audiology DIAGNOSTIC TESTING GUIDELINES for Audiology In 1999, the Illinois legislature passed the Hearing Screening for Newborns Act. By December 31, 2002, hospitals delivering babies were required to provide hearing

More information