817 Bioscan. Bioscan. Instructions for Use CH-9101 Herisau/Switzerland Internet

Size: px
Start display at page:

Download "817 Bioscan. Bioscan. Instructions for Use 8.817.1003. CH-9101 Herisau/Switzerland E-Mail info@metrohm.com Internet www.metrohm."

Transcription

1 Bioscan CH-9101 Herisau/Switzerland Internet Instructions for Use

2

3 CH-9101 Herisau/Switzerland Internet Instructions for Use / up

4

5 Contents Contents 1 Introduction Instrument description Parts and controls Information about the Instructions for Use Organization Notation and pictograms Safety notes Electrical safety General precautionary rules Installation Setting up the instruments Packaging Check Location Arrangement of the instruments Mains connection Setting the mains voltage Fuses Mains cable and mains connection On/Off switching Capillary connections Connection of the Bottle Rack (option) Connection of the pulsation absorber Connection of the injection valve Mounting the column Connection of the flow cell Connecting the waste tubing Rinsing the tubing Rinsing the column System for sugar analysis MIC Electrical connections Settings in «IC Net 2.1» Connecting a sample changer System for the analysis of anions and sugars MIC Electrical connections Capillary connections Configuration in «IC Net 2.1» Preparing the for the analysis System switch-on Start pump Set operating mode I

6 Contents 3 Operation Handling the Operation using «IC Net 2.1» icon window Basic principles Introduction Measuring conditions Pulsed amperometric detection Optimization of the PAD parameters Optimization of the measuring potential Notes Maintenance Faults Practical notes on ion chromatography Separating columns High-pressure pump Eluents Connections Maintenance and servicing General information Shutdown Cleaning the working electrode Changing separating columns Faults and malfunctions Malfunctions and their rectification Instrument test with the dummy cell Validation / GLP Appendix Technical data Standard equipment Warranty and conformity Warranty EU declaration of conformity Certificate of conformity and system validation Index II

7 Contents List of illustrations Fig. 1: Front of the... 3 Fig. 2: Rear of the... 4 Fig. 3: Interior of the (with permanently attached accessories).. 6 Fig. 4: Setting the mains voltage Fig. 5: Connections at the injection valve of the Valve Unit Fig. 6: Connections at flowcell Fig. 7: Connecting up a simple modular IC system with the MIC Fig. 8: Connecting up a simple MIC 8 system including a sample changer Fig. 9: Connections with a as second detector with a modular IC system MIC Fig. 10: Connections of injection valves A and B in the IC Separation Center ( ) Fig. 11: Potentials applied during pulsed amperometric detection (PAD) Fig. 12: Example of a hydrodynamic voltammogram of a substance (A) with additional presentation of the measured values for the pure eluent (B) Fig. 13: Example of a scan of a substance (A) with additional presentation of the scan of the pure eluent (B) III

8 Contents IV

9 1.1 Instrument description 1 Introduction 1.1 Instrument description The Bioscan is a PC-controlled measuring instrument for the sensitive analysis of carbohydrates by ion chromatography using pulsed amperometric detection. The compact housing of the contains several IC system components: Detector flow-through cell with three-electrode arrangement for amperometric detection in the Pulse, DC and Scan modes. Column compartment the perfect insulation of the housing creates not only stable thermal conditions for the separating columns, but also shields the system from electromagnetic interference; in addition to the column it also contains the detection cell, pulsation absorber and preheating capillary. Oven amperometric determinations require extremely stable thermal conditions. The built-in column compartment oven ensures that all important components can be set exactly to a temperature from 10 C above room temperature to 60 C with a stability of 0.1 C. All components that come into contact with the eluent and the sample, except the electrodes, are metal-free. Signal converter the contains its own analog/ digital converter for the detector signal. This means that complete control of the instrument is possible from a PC via an RS232 interface. Together with the 709 IC Pump and the 812 Valve Unit, the forms a complete IC system for carbohydrate analysis. Operation is via a PC connected to the RS232 interface and uses the included «IC Net 2.1» control and evaluation program. This software fulfills all the demands which are placed today on a modern integration software: 1-point or multi-point calibration, internal or external standard, selectable algorithms for non-linear calibration, numerous Integration modes with selectable parameters and integration events, various peak recognition methods, peak editor, free scaling, superimposition of several chromatograms, post-treatment of chromatograms, high-performance GLP-conform report generator with output interfaces for monitor, printer and external databases. 1

10 1 Introduction 2

11 1.2 Parts and controls 1.2 Parts and controls Bioscan Fig. 1: Front of the 1 Door to interior 3 Feed through for tubing e.g. connection of injection valve and preheating capillary or connection to waste 2 Feed through for tubing e.g. connection of 709 IC Pump and pulsation absorber 3

12 1 Introduction Fig. 2: Rear of the 4

13 1.2 Parts and controls 4 Mounting screw for fastening the rear panels 5 / 6 5 Detachable rear panel access to top part of the inner compartment 6 Detachable rear panel access to bottom part of the inner compartment 7 Analog signal (REC.) AD/DA-converted signal (not used with IC Net; see chapter 6.1 for details) 8 Analog signal (Int.) Original signal (not used with IC Net; see chapter 6.1 for details) 9 Manual valve (not used with IC Net) 10 Model plate with technical data serial number 12 Lower connections for remote lines (not used with IC Net) 13 Outlet for spilled liquid for discharge of spilled liquid from the inner compartment 14 RS232 interface connection of the PC; see chapter 6.1 for details 15 Upper connections for remote lines only Aux 1, Aux 2 as well as 1 and 2 are used with IC Net 16 Fuse holder changing the fuses, see section Mains connection plug mains connection, see section Mains switch to switch instrument on and off: I = ON 0 = OFF 11 Electric valve (not used with IC Net) 5

14 1 Introduction Metrosep Carb Fig. 3: Interior of the (with permanently attached accessories) 6

15 1.2 Parts and controls 19 Electrode cable (red) connection for working electrode 20 Electrode cable (blue) connection for auxiliary electrode 21 Column connection capillary PEEK capillary 27 Connection for elektrode cable connection '1' 28 Connection capillary connection 709 IC Pump pulsation absorber 29 Connection capillary connection pulsation absorber injection valve 22 Measuring cell holder 30 Pulsation absorber Au - Flow cell Electrode cable (black) connection for reference electrode 25 IC column e.g. Metrosep Carb 1 ( ) for carbohydrate analysis 31 Preheating capillary , length: 3 m 32 Oven 33 Oven heating fan 26 Column holder 34 Connection capillary connection flow cell waste 7

16 1 Introduction 1.3 Information about the Instructions for Use Please read through these Instructions for Use carefully before you put the into operation. The Instructions for Use contain information and warnings to which the user must pay attention in order to assure safe operation of the instruments Organization These Instructions for Use for the provide a comprehensive overview of the startup procedure, operation, fault rectification and technical specifications of these instruments. The Instructions for Use are organized as follows: Section 1 Section 2 Section 3 Section 4 Section 5 Section 6 Introduction General description of instruments, parts and controls and safety notes Installation Setting up, mains connection, attachment of accessories, connection to IC system Operation Detailed description of the operation Basics Information about the pulsed amperometric detection Notes Maintenance Faults Notes on ion chromatography, maintenance, fault rectification, diagnostic tests, validation Appendix Technical data, standard equipment, options, warranty, declarations of conformity, index To find the required information on the instrument, you will find it an advantage to use either the Table of contents or the Index at the back. As a supplement to the Instructions for Use, the Metrohm Monograph "Ion chromatography" is also supplied. This provides an introduction to the theoretical fundamentals and general information on separating columns and sample pretreatment. The Metrohm Monograph "Practical Ion Chromatography" is a practical textbook, which provides an illustrative introduction to the basic principles of ion chromatography and also describes 22 experiments covering the whole world of ion chromatography. IC Application Notes for different applications of Metrohm IC systems are also supplied. 8

17 1.3 Information about the Instructions for Use Notation and pictograms The following notations and pictograms (symbols) are used in these Instructions for Use: <PARAM> Key 'Range' Parameter or entry value 35 Part or control of Part or control of 709 Hazard This symbol draws attention to a possible danger to life or of injury if the associated directions are not followed correctly. Warning This symbol draws attention to possible damage to instruments or instrument parts if the associated directions are not followed correctly. Caution This symbol marks important information. First read the associated directions before you continue. Comment This symbol marks additional information and tips. 9

18 1 Introduction 1.4 Safety notes Electrical safety While electrical safety in the handling of the is assured in the context of the specifications EN / IEC , the following points should be noted: Mains connection Setting of the mains voltage, checking the mains fuse and the mains connection must be effected in accordance with the instructions in section 2.4. Opening the If the is connected to the power supply, the instrument must not be opened (apart from the front door) nor must parts be removed from it, otherwise there is a danger of coming into contact with components which are live. Hence, always disconnect the instrument from all voltage sources before you open it and ensure that the mains cable is disconnected from mains connection 17! Protection against static charges Electronic components are sensitive to static charging and can be destroyed by discharges. Before you touch any of the components inside the, you should earth yourself and any tools you are using by touching an earthed object (e.g. housing of the instrument or a radiator) to eliminate any static charges which exist General precautionary rules Handling of solvents Check all lines of the IC system periodically for possible leaks. Follow the relevant instructions regarding the handling of flammable and/or toxic solvents and their disposal. Never block drain opening for spilled liquids On the bottom of the interior directly below the front door there is a drain opening 1. Spilled liquids can flow directly to the outlet 13 on the back. Due to safety consideration, take care that these openings may never be blocked. 10

19 2.1 Setting up the instrument 2 Installation The is an electrochemical detector and can be included in a modular IC system in two different ways: 1) The as the only detector in an IC system. Apart from the 709 IC Pump, you also require an 812 Valve Unit with one injection valve. In this MIC-8 modular IC system the is intended for the analysis of sugars only. 2) The is included in an IC system as the second detector. An example of this is given in Section 2.5 where an IC system (MIC-9) is described for the simultaneous determination of sugars and anions with chemical suppression. These modular IC system configurations (MIC-1, MIC-2, etc.) are possible instrument combinations that have been properly tested. Under you will find further examples; these are continually updated. Following the general information about the installation of the instruments, Sections 2.4 and 2.5 describe the two configurations mentioned above including their control by the «IC Net 2.1» software. 2.1 Setting up the instrument Packaging The is supplied together with the separately packed accessories in special packaging containing shock-absorbing foam linings designed to provide excellent protection. The actual instruments are packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all this special packaging as only it can assure transport of the instruments free from damage Check After receipt, immediately check whether the shipment is complete and has arrived without damage (compare with delivery note and list of accessories in section 6.2). In the case of transport damage, see instructions in section "Warranty". 11

20 2 Installation Location Position the instruments in the laboratory at a location convenient for operation, free from vibrations and protected against a corrosive atmosphere and contamination by chemicals. The same applies to all other components of the IC system. To avoid disturbing temperature influences on the insulated column compartment, the entire system including pump and eluent reservoir must be protected against direct sunlight Arrangement of the instruments If the is used as a single detector (see chapter 2.4), the 709 IC Pump, 812 Valve Unit and are best stacked on top of one another in this order. In a two-channel system (see chapter 2.5), the optimum arrangement (1, 2 or 3 towers) depends on the laboratory space available. In general, the IC pumps should be set up at the very bottom and the IC detectors at the very top. 2.2 Mains connection Follow the instructions below for connecting to the power supply. If the instrument is operated with a mains voltage set wrongly and/or wrong mains fuse, there is a danger of fire! Setting the mains voltage Before switching on the for the first time, check that the mains voltage set on the instrument (see Fig. 4) matches the local mains voltage. If this is not the case, you must reset the mains voltage on the instrument as follows: 1 Disconnect mains cable Disconnect mains cable from mains connection plug Remove fuse holder Using a screwdriver, loosen fuse holder 16 below the mains connection plug and take out completely. 3 Check and change fuse if necessary Carefully take the fuses installed for the desired mains voltage out of fuse holder and check their specifications: V 3.15 A (slow-blow) Metrohm No. U V 1.6 A (slow-blow) Metrohm No. U

21 2.2 Mains connection 4 Insert fuse Change fuses if necessary and reinsert in fuse holder. 5 Install fuse holder Depending on the necessary mains voltage, insert fuse holder in the so that the corresponding mains voltage range can be read normally and the adjacent white arrow points to the white bar imprinted below the fuse holder (see Fig. 4) V V Mains switch Mains connection plug V V 16 Fuse holder V V Fig. 4: Setting the mains voltage Fuses Two fuses 3.15 A/slow-blow for V or 1.6 A/slow-blow for V are installed in fuse holder 16 of the as standard. Ensure that the instrument is never put into operation with fuses of a different type as this could cause a fire! For checking or changing fuses, proceed as described in section

22 2 Installation Mains cable and mains connection Mains cable The instrument is supplied with one of three mains cables with plug SEV 12 (Switzerland, ) with plug CEE(7), VII (Germany, ) with plug NEMA 5-15 (USA, ) which are three-cored and fitted with a plug with an earthing pin. If a different plug has to be fitted, the yellow/green lead (IEC standard) must be connected to protective earth (protection class 1). Any break in the earthing inside or outside the instrument can make it a hazard! Mains connection Plug the mains cable into mains connection plug 17 of the (see Fig. 2) On/off switching of the instruments The is switched on and off using mains switch 18. Do not switch the IC components on before all cable connections have been established. 14

23 2.3 Capillary connections 2.3 Capillary connections The capillary connections described below are necessary for operating the in a simple IC system for carbohydrate analysis (MIC- 8). The electronic connections and control by IC Net 2.1 are described in Section 2.4. Section 2.5 contains the description of a combination of this IC system with an IC system for the determination of anions with chemical suppression (MIC-9). The necessary changes to the capillary connections are described there Connection of the Bottle rack (option) The optionally available Bottle rack for supply vessels can be placed on top of the. The tubing connections to the 709 IC Pump are described in the corresponding leaflet Connecting the pulsation absorber To protect the column material from any pressure shocks which the injection may cause we recommend the insertion of the MF Pulsation absorber between the high-pressure pump and the injection valve. This is done as follows (see Fig. 3): 1 Connect the pulsation absorber Place pulsation absorber 30 on the floor in the interior of the. 2 Connection to pump Connect the PEEK Filter unit described in Section of the Instructions for Use to connection 23 of the 709 Pump by using PEEK capillary 22. Lead PEEK capillary 28 or 43 from connection piece 47 of the filter unit outwards through opening 2 of the Bioscan and attach it to one of the 52 connections on the top side of the pulsation absorber. 3 Connection to injection valve Connect PEEK capillary 29 to the second connection 30 of the pulsation absorber, lead it out through 2 of the Bioscan and connection it to the injection valve on the 812 Valve Unit (connection 17). The pulsation absorber is filled with isopropanol and must be rinsed with eluent before a separating column is connected (see Section 2.3.7). The Pulsation absorber can be operated in both directions. 15

24 2 Installation Connection of the injection valve Connect all capillaries to the injection valve according to the following description: Fig. 5: Connections at the injection valve of valve unit 42 Connection for syringe tubing PTFE tubing with coupling ( ) and syringe ( , without needle) 43 Connection for aspirating tubing PTFE tubing for aspirating sample 44 Connection for sample loop 20 µl PEEK sample loop ( ) 45 Eluent outlet Connection to preheating capillary 31 (PEEK, ) 46 Eluent inlet PEEK capillary connection 29 to pulsation absorber 47 Connection for sample loop 20 µl PEEK sample loop ( ) Mounting the column The IC separating column (e.g. Metrosep Carb 1, ) is first mounted in the as follows (see Fig. 3): 1 Connect column to the preheating capillary Remove the caps from column 25. Screw eluent outlet of preheating capillary 31 on to the inlet end of the column (observe flow direction). 2 Connect column to flow-through measuring cell Screw outlet end of column 25 on to eluent inlet 41 of the flow-through cell using the KELF pressure screw. 3 Fix column Fix the column in column holder

25 2.3 Capillary connections Connection of the flowcell The Flowcell ( ) is a part of the standard equipment of the. It is already mounted and ready to use Fig. 6: Connections at flowcell 35 Check marks For correct orientation of the top and bottom part of the flow-cell 39 Auxiliary electrode connection for blue electrode cable Mounting screws (4x) 40 Reference electrode connection for black electrode cable Outlet for eluent Connection of PEEK capillary by KELF pressure screw. PEEK pressure screws may damage the threads of the cell. 41 Inlet for eluent Connection of PEEK capillary by KELF pressure screw. PEEK pressure screws may damage the threads of the cell. 38 Working electrode connection for red electrode cable 19 1 Before starting up check all screws 36 Use the Allen key provided to loosen all four of the 36 screws by one rotation. Then slightly tighten the screws again, e.g. by holding the Allen key between the thumb and index finger only. Finally tighten all four screws by a quarter rotation. Always tighten opposing screws, i.e. tighten one screw and then the opposite screw. Avoid overtightening the screws as this could damage the Distance piece! 17

26 2 Installation 2 Connect the electrode cable The included Cable for connecting the measuring cell is attached to connections 38, 39 and 40 of the measuring cell using the three connectors 19, 20 and 24 (see Fig. 6). The other end of the cable is inserted into connection '1' of the 817 with connector 27 (see Fig 3). 3 Connect the column Screw column connection capillary 21 of the column on to eluent inlet 41 of the flow-through measuring cell using the Pressure screw. 4 Connect the waste container In order to ensure an adequate backpressure, screw a PEEK capillary (approx. 1 m) on to eluent outlet 37 of the measuring cell. Lead the end through opening 3 into a sufficiently large waste container and attach it there. 5 Mount the measuring cell Attach the flow-through measuring cell to the holder Bioscan which is provided for it (see Fig. 3). Rotate the holder and the cell so that the cell outlet is located as high as possible. In this way it is easier for any air bubbles that may occur to escape from the cell. Never switch on the flow-through cell when this is (a) not being rinsed through with a conducting eluent at the time, (b) not completely connected up, or (c) when its outside is moist so that so a short circuit could occur between the connections of working electrode 38 and auxiliary electrode 39! Connecting the waste tubing The has an outlet 13 on its rear panel for discharged liquids to which the waste tubing can be attached. Proceed as follows: 1 Connect the waste tubing Attach matching tubing (e.g. Silicone tubing, optional) into outlet 13 (see Fig. 2). 2 Lead the waste tubing to a drain Place the other end of the waste tubing in a drain and attach it there. 18

27 2.3 Capillary connections Rinsing the tubing The pulsation absorber is supplied permanently mounted and is filled with isopropanol. In order to remove this and any air that may be present in the capillaries and the measuring cell all the eluent-carrying tubing or capillary connections must first be installed; the column and the flow cell must not be connected during the beginning of the process. Proceed as follows: 1 Deconnect preheating capillary from column Deconnect the eluent outlet of preheating capillary 31 from the column and fix it at a waste bottle. 2 Set injection valve to Inject Press the <INJECT> key on the 812 Valve unit, which should not yet be connected to the. The green LED in the key lights up to show that the injection valve is in the Inject position. 3 Vent 709 IC Pump before start up Proceed according to Section 2.7 of the Instructions for Use of the 709 IC Pump. 4 Rinse with dist. H 2 O Immerse the aspiration capillary of the 709 IC Pump in distilled or deionized water. Switch on the 709 IC Pump, rinse the IC system for approx. 10 min. and switch off the 709 IC Pump. 5 Rinse with eluent Immerse the aspiration capillary of the 709 IC Pump in the eluent to be used (e.g. 100 mm NaOH). Switch on the 709 IC Pump, rinse the IC system for approx. 10 min. and switch off the 709 IC Pump. 6 Connect flowcell and rinse with eluent Connect the preheating capillary 31 to flowcell using column connection capillary 21 and Coupling. Switch on the 709 IC Pump, rinse the IC system for approx. 10 min. with eluent and switch off the 709 IC Pump. 7 Remove the coupling Remove the Coupling between column connection capillary 21 and preheating capillary 31. Remount the column between column connection capillary 21 and preheating capillary 31 as show in Fig Rinsing the column In order to rinse or precondition the column the connection capillary 21 to flowcell must be unscrewed from the column. Proceed as described in point 3 and 4 in section

28 2 Installation 2.4 System for sugar analysis MIC Electrical connections The MIC-8 system with manual sample injection is described below. Of course, you can also install a Metrohm sample changer. This is described in Section 2.5 or in the Instructions for Use of the sample changer. In order to connect a sample changer and control it completely you need a further free COM port on your PC. However, we recommend the use of the 762 IC Interface, to which you can then also connect the 709 IC Pump to be controlled by IC Net 2.1. The electrical connections of the MIC-8 system, which consists of the, the 812 Valve unit and the 709 IC Pump, are made according to Figure 7: 817 AUX 1 AUX PC Ground Fill Inject 709 Fig. 7: Connecting up a simple modular IC system with the 817 Bioscan MIC 8 20

29 2.4 System for sugar analysis MIC Settings in «IC Net 2.1» 1 System As in this configuration only the is to be directly controlled by IC Net then only this instrument is to be installed as the detector with the help of the system assistant New system wizard. 2 Method You should first select the standard method supplied for use with your particular column (e.g. for Metrosep Carb 1: CARB1_250.mtw). You can save any alterations under a different file name. 3 Data source Select: Cell( 817 BioScan) as the signal source. 4 Save system Then save the system (e.g. 817_MIC-8.smt). The following system window should now appear: 5 Connect system and workplace Under SYSTEM / Control / Connect to workplace connect the new system with the workplace. At the same time the symbol bar. will appear in the symbol 6 Check measuring cell If the is switched on without the flow-through cell being constantly rinsed by a conducting eluent (e.g. 0.1 M NaOH) then the measuring cell must not be switched on! You should now check this in the instrument window of the 817 Bioscan (accessed with a double-click with the mouse on instrument symbol) under Manual / Cell. 7 Check start mode Make sure that in this system the is standing under Start with determination under Setup / Start mode.. 21

30 2 Installation Connecting a sample changer If the MIC-8 system described above is to be operated with a Metrohm sample changer then the following connections and settings must be made: 817 AUX 1 AUX CELL OFF CELL ON + 24 V GROUND PC 766/788/ Ground Fill Inject Fig. 8: Connecting a simple MIC 8 system including a sample changer The IC Sample Processor 766, the IC Filtration Sample Processor 788 or the Compact Autosampler 813 can be used as the sample changer. In this combination of the MIC-8 system the is operated as the "Master". The start impulse for the instrument methods to the connected sample changer is transferred via the remote connection ( , optional). You can then select any instrument method at the sample changer that is described as being for control with the PC as master. Information about this is given in the corresponding Instructions for Use. If a second COM port on the PC or a 762 IC Interface is available for an RS232 connection then the 766 IC Sample Processor or the 788 IC Filtration Sample Processor can be connected up via this connection. In this way you can generate your own time programs in IC Net 2.1 for these sample changers. Details are given in the Instructions for Use for the particular instrument. 22

31 2.4 System for sugar analysis MIC-8 Creating the time program In order to control the 766 IC Sample Processor and the 788 IC Filtration Processor the following program is created in the Bioscan: In this way shortly after the injection valve has been switched to 'Fill' at the 812 Valve unit, the first remote line 'AUX 1' (see Fig. 6) will be switched briefly to 'ON'. This signal will be scanned at the start of the suitable sample changer programs. Terminate the program entry by activating the program with Enable and store it in the system with Save. The control of the 813 Compact Autosampler is carried out similarly. The only thing to be done here is to set the remote lines to the required times as described above for each of the three working steps: autosampler start, pump start and pump stop in order for a brief impulse to be transferred. In order for the connection Autosampler to function properly the program must first be started at the sample changer and then the "Sample queue" in IC Net

32 2 Installation 2.5 System for the analysis of anions and sugars MIC Electrical connections The combination of an anion system with chemical suppression with the IC system for sugar analysis is described as an example of a complete IC system for the simultaneous determination of anions and sugars. As ideally the two sample loops for the different separation pathways should be connected by as short a piece of capillary as possible for sample injection, a 733 IC Separation Center with two injectors ( ) is used. This means that the 812 Valve Unit is no longer required for sample injection for the Bioscan. The MSM Suppressor Module, which is frequently located in the 733 IC Separation Center, is then replaced by the 753 IC Suppressor Module which is used instead of the 752 IC Pump Unit. The electrical connection of the system, consisting for example of two 709 IC Pumps,, 732 IC Detector, 733 IC Separation Center (2 injectors), 753 IC Suppressor Module, 762 IC Interface and 766 IC Sample Processor is made according to the Figure 9: B A 762 PC MSM Fig. 9: Connections with a Bioscan 817 as second detector with a modular IC system MIC 9 24

33 2.5 System for the analysis of anions and sugars MIC Capillary connections This section describes the capillary connections required for the combination of the IC system for sugar analysis (MIC-8) with an anion system with chemical suppression (e.g. MIC-2). It is assumed that the capillary connections for the anion system are already known. Arrange the necessary components as shown in Fig. 9. Please also observe the recommendations about the arrangement of the IC modules given in Section The capillary connections for the correspond to a large extent to the connections described in Section 2.3. Only the connection of the injection valve to the 812 Valve Unit is not necessary. Instead the second injection valve B of the 733 IC Separation Center is connected for injecting the sample for the sugar determination (see below). For the components for determination of the anions the typical arrangement of the capillaries for such an anion system is assumed to be present. The necessary alterations concern the injection valve A in the 733 IC Separation Center, which must be connected for injection valve B so that the two sample loops can be filled simultaneously. In addition the MSM Suppressor Module connections must be transferred to the 753 IC Suppressor Module if the 752 IC Pump Unit has previously been used with the MSM Module contained in the 733 IC Separation Center. The necessary modifications are described below. Injections valves A and B are designated as follows: Injection valve A: sample injection for anion determination Injection valve B: sample injection for sugar determination All the connections for the two injection valves are shown in Figure 10 on the following page. 1 Connection injection valve A injection valve B For simultaneously filling the two sample loops as short as possible piece of PEEK capillary is attached to sample outlet 48 of injection valve A and sample inlet 55 of injection valve B. 2 Connection pulsation absorber injection valve B Connect PEEK capillary 29 to pulsation absorber 52 in the Bioscan, lead it out through opening 2 or 3 of the Bioscan, and in through opening 27 or 28 of the 733 IC Separation Center and connect it to injection valve B (eluent inlet 58). 3 Connection injection valve B preheating capillary Connect PEEK capillary at injection valve B (eluent outlet 57), lead it out through opening 27 or 28 of the 733 IC Separation Center and in through opening 2 or 3 of the Bioscan and use a PEEK coupling to connect it to preheating capillary

34 2 Installation A B Fig. 10: Connections of injection valves A and B in the IC Separation Center ( ) Injection Valve A - Anions - 48 Sample outlet PEEK capillary connection to injection valve B 49 Sample inlet PEEK capillary connection to sample processor (18 in Instructions for Use for 766 IC Sample Processor) 50 Connection for sample loop PEEK sample loop for anion analysis 51 Eluent outlet PEEK capillary connection to anion separation column 52 Eluent inlet PEEK capillary connection, e.g. to pulsation absorber for anion eluent 53 Connection for sample loop PEEK sample loop for anion analysis Injection Valve B - Carbohydrates - 54 Sample outlet PTFE tubing connection to waste 55 Sample inlet PEEK capillary connection to injection valve A 56 Connection for sample loop 20 µl PEEK sample loop ( ) 57 Eluent outlet Connection of preheating capillary 31 (PEEK, ) 58 Eluent inlet PEEL capillary connection 29 to pulsation absorber 59 Connection for sample loop 20 µl PEEK sample loop ( ) 26

35 2.5 System for the analysis of anions and sugars MIC Configuration in IC Net 2.1 The complete IC system is installed in IC Net using the New system wizard: IC Interface IC Detector IC Pump 27

36 2 Installation IC Pump IC Sample Processor IC Separation Center Suppressor Module 28

37 2.5 System for the analysis of anions and sugars MIC-9 9 Second data recorder 10 Anion method and data source 11 System overview 12 Final settings In the system window of the 753 Suppressor Module the option Start pump with hardware should be activated under Manual. 29

38 2 Installation 13 Bioscan method and data source The Bioscan method and corresponding data source must be selected in the Setup menu which can be accessed by a righthand mouse click on the second data recorder symbol. 14 Connect system with workplace Under SYSTEM / Control / Connect to workplace connect the new system with the workplace. At the same time the symbol appears in the symbol bar. 15 Check measuring cell If the is switched on without the flow-through cell being constantly rinsed by a conducting eluent (e.g. 0.1 M NaOH) then the measuring cell must not be switched on! You should now check this in the instrument window of the 817 Bioscan (accessed with a double-click with the mouse on instrument symbol) under Manual / Cell. 16 Check connections You can now check the electronic connections to all modules. This is done by double-clicking with the left-hand mouse key on each instrument symbol. If the connection is functioning properly then the current status of the instrument parameters will be shown. Remarks The event-line for the suppressor of the 753 Suppressor Module is only set up as a "dummy connection" (Step 8). The real command for switching the suppressor further is made by the 766 Sample Changer via a remote line. For the installation described above the option System face / Stacked recorder icons must be activated in IC Net 2.1 under Options / Global preferences. This is why only a symbol for the data recorder is visible in the system overview. With a right-hand mouse click and the offered Setup option the setting possibilities described for both recorders man (Step 10 and Step 13) can be accessed. 30

39 2.5 System for the analysis of anions and sugars MIC-9 Creating the time program 1. Input of the program for the 766 IC Sample Processor (Master): 001 Ctrl INIT 002 Move sample 003 Lift work 004 Ctrl ZERO Ctrl FILL A Ctrl STEP MSM Ctrl FILL B/Step Pump 120 s 009 Ctrl INJECT A Ctrl INJECT B 1 Initialize remote interface Move needle to sample position Move lift with needle to working height Trigger Autozero on 732 IC Detector Switch injection valve A on 733 to "Fill" Switch on 753 Suppressor-Module Switch injection valve B on 733 to "Fill" Fill sample loops A + B with sample for 120 s Switch injection valve A on 733 to "Inject" Switch injection valve B on 733 to "Inject" 1. Terminate the program entry by activating the program with Enable and store it in the system with Save. 2. Finally check that in the IC system the 766 Sample Processor is entered under Start with inject under Setup / Start mode. Remarks If only one of the detection pathways is to be used in this IC system then the modules which are not required can be removed from the system (remove recorder and instrument symbols with the right-hand mouse key and Unlink) and save the system under a new name. However, please note that in this instrument configuration the recording of the Bioscan data nevertheless begins only when the injection valve A on the 733 IC Separation Center is switched from Fill to Inject, as there is only a corresponding connection to the 762 IC Interface from this valve (see Fig. 9). This means that even if only sugar analysis is to be carried out, the 733 IC Separation Center and the 732 IC Detector must still form part of the system described above. The program described above for the 766 IC Sample Processor must then remain unchanged. 31

40 2 Installation 2.6 Preparing the for the analysis System switch-on This section describes the necessary steps for preparing the to carry out an analysis. A more detailed description of the operation of the using IC Net 2.1 can be found in Section 3 and in the Instructions for Use of the Metrodata IC Net 2.1 software. 1 Mains on Start up the individual instruments and the IC Net program in the following sequence: 1. Computer 2. IC instruments 3. IC Net Check measuring cell For safety reasons check the status of the measuring cell in the instrument window of the Bioscan (accessed by a double-click with the mouse on the instrument symbol) under Manual / Cell. If the is switched on without the flow-through measuring cell being constantly rinsed through a conducting eluent (e.g. 0.1 M NaOH) or correctly connected up then the measuring cell must not be switched on! 3 Check the external control of the 709 IC Pump Depending on the control of the 709 IC Pump (MIC-8: manual; combination with 762 IC Interface: external) the setting of the external control of the IC Pump should be checked with key 8 <EXT.> Start pump 1 Column/ Eluent With the Metrosep Carb 1 ( ) column use 0.1 M NaOH (quality) as the eluent for sugar analysis. This should be made up and degassed freshly every day; it should at least be degassed before use each day. 2 Flow rates / Pressure The Metrosep Carb 1 column should not be subjected to large variations in pressure. For this reason each time that the eluent is used it should be pumped for a period of 10 minutes both before and after use at 0.3 ml/min. For normal operation a flow rate of1.0 ml/min is used. Set a maximum pressure P max of 15 MPa. At 1 ml/min and with 0.1 M NaOH this should establish a pressure of approx MPa with the Metrosep Carb 1. 32

41 2.6 Preparing the for the analysis Set operating mode 1 Manual control of the 817Bioscan In the instrument window of the Bioscan (double-click with the mouse on the instrument symbol) select Mode: Pulse for pulse amperometric detection. The cell must remain deactivated for as long no eluent flows through the flow-through measuring cell. The oven is set to e.g. 32 C. In general this temperature should be at least 10 C higher than the ambient room temperature. 33

42 2 Installation 2 Parameters for the pulse mode Set these method parameters for the pulse mode. Please note that this menu can only be accessed when the option Pulse has been selected for Mode under Manual (see above). Finally transfer the current parameters to Bioscan with <Set> and use System / Save to save them in the system file (*.smt). The IC system with the is now ready for use. 34

43 3.1 Handling the 3 Operation 3.1 Handling the General Amperometric detection takes place with a flowing current and therefore with a chemical conversion of the analyte. The course of a chemical reaction depends directly on various physical parameters, among other things. In order to obtain optimum measuring conditions (e.g. stable baseline or reproducible signals) it is necessary to take the following points into consideration. Further information can be found in Section 4.2. Constant temperature The allows the interior of the instrument to be kept at a constant temperature of up to max. 60 C. The lower limit is the ambient temperature plus 10 C. Under IC Net you should activate the appropriate oven-operating mode in the 817 method parameters and define the temperature (see Section 3.2.2) Constant ph The characteristic current/potential curves of all electrochemically reactive substances depend very strongly on the ph of the ph eluent used. This means that you should prepare the eluent very carefully and check its ph at regular intervals. Pulsation-free eluent flow Changes to the eluent flow should be avoided as far as is possible. The Pulsation Absorber is available for this purpose; it is mounted between the 709 IC Pump and the injection valve as described in Section Never make dry measurements The flow-through measuring cell must not be switched on if it is not being rinsed constantly by the eluent or if the measuring cell is not connected up properly! 35

44 3 Operation 3.2 Operation using «IC Net 2.1» icon The 817 icon is available in the system window if an has been installed with the New system wizard or by using the Setup/New devices/install new device option of the SYSTEM window (see section 2.4.2). If the system is connected and the 817 icon is clicked with the right-hand mouse button then the following menu appears: Open Unlink Opens the window for parameter settings (this window can also be opened by double-clicking the icon). Deletes the 817 icon from the system. 36

45 3.2 Operation using «IC Net 2.1» window 817 icon / Open The window for parameter settings is opened by selecting this menu option with the right-hand mouse button or by double-clicking the 817 icon in the SYSTEM window. It consists of the four tabs Manual, Method parameters, Program, and Links. Manual The Manual tab of the window is only available for a connected system. Mode DC Pulse Scan Selection of detector mode: In the DC mode a constant potential is applied; the analytes are oxidized or reduced in accordance with their electrochemical properties. The Pulse mode works with three different potentials that are applied cyclically. This frees the electrode surface from any adhering reaction products during each sweep and an activated surface is produced for the next measurement. The Scan mode allows potential/current curves to be recorded. 37

46 3 Operation Cell <On> <Off> Manual control of amperometric detector cell: Switches on detector cell. Switches off detector cell. Attention Switch on the detector cell only if a conducting eluent is being pumped through it continuously! Oven <On> <Off> Temperature Manual control of column oven (heater): Switches on oven. Switches off oven. Temperature of column oven. Range: C (> room temp C) Control Manual setting of 4 remote lines. <Set> Sets remote lines to selected states. Aux1, Aux2, Cont1, Cont2 Setting of 4 remote lines. Selection: 0, 1, * Valve <Fill> <Inject> Analog output <Zero> Actual Mode Cell Oven Control Manual switching of the injection valve, e.g. at the 812 Valve Unit: Switches injection valve to "Fill" position (control line Cont1 is set to 0). Switches injection valve to "Inject" position (control line Cont1 is set to 1). Manual setting of the analog output signal to zero. For RS232 data the full-scale range (see next page) is shifted until the signal is located in the center of this range again (s. also Chapter 6.1). Triggers the autozero function. Display of actual values. Display of current detector mode. Display of current cell state. Display of current oven state. Display of current control line settings. 38

47 3.2 Operation using «IC Net 2.1» Method parameters DC mode The DC mode subtab of the Method parameters tab on the 817 Bioscan window is used to set the parameters for the DC mode. This window is only available if Mode has been set to DC on the Manual tab. Method data Cell potential Electric potential applied to the amperometric detector cell in V Range: V. Oven Switches column oven on/off and sets operating temperature. Range: C (> room temp C) Range Selection of measuring range: pa pa. na na. µa µa. Full scale The full-scale range (operating range) sets the desired sensitivity for the analog output. The possible values of the full-scale range depend on the preset measuring Range. For RS232 data (IC Net 2.1), always the maximum settings should be used. 39

48 3 Operation Filter Offset Electronic damping of the analog output (REC) and RS232 signal. Selection: 0.1, 0.2, 0.5, 1, 2, 5 s Offset of the zero point. Selection: % (in steps of 10 %) Actual Cell potential Display of current cell potential in V Oven Display of current oven state. Range Display of current measuring range. Full scale Display of current full-scale range. Filter Display of current damping constant. Offset Display of current offset. <Set> Sends current parameters immediately to the. Parameters are not stored in the system file (*.smt) as long as the file is not saved. 40

49 3.2 Operation using «IC Net 2.1» Pulse mode The Pulse mode subtab of the Method parameters tab on the window is used to set the parameters for the pulse mode. This window is only available if Mode has been set to Pulse on the Manual tab. Method data Cell Definition of three different working potentials that are applied to the amperometric detector cell cyclically. The set potentials are graphically displayed if the Method option is enabled above the graphic window. E1, E2, E3 Potentials. Range: V. t1, t2, t3 Time duration to apply the potentials. Range: s Sample time Measuring time at the end of working potential E1. Selection: 20, 40, 60, 80, 100 ms Oven Switches column oven on/off and sets operating temperature. Range: C (> room temp C) Range Selection of measuring range: pa pa. na na. µa µa. In the Pulse mode, always the µa range should be selected. 41

50 3 Operation Full scale Offset The full-scale range (operating range) sets the desired sensitivity for the analog output. The possible values of the full-scale range depend on the preset measuring Range. For RS232 data (IC Net 2.1), always the maximum settings should be used. Offset of the zero point. Selection: % (in steps of 10 %) Actual E1, E2, E3 Display of current potentials in V. t1, t2, t3 Display of current time durations for potentials in s. Sample time Display of current sample time in ms. Oven Range Full scale Offset Display of current oven state. Display of current measuring range. Display of current full-scale range. Display of current offset. Cycle potentials Graphical display of cycle potentials. Method Actual <Set> Display of potentials set in the Method data frame. Display of current potentials. Sends current parameters immediately to the. Parameters are not stored in the system file (*.smt) as long as the file is not saved. 42

51 3.2 Operation using «IC Net 2.1» Scan mode Parameters The Scan mode subtab of the Method parameters tab on the window is used for parameter setting and curve display for the scan mode. This window is only available if Mode has been set to Scan on the Manual tab. Method data Cell potential Definition of start and end potential for a user-defined potential sweep that is applied to the amperometric detector cell. E1 Start potential for sweep. Range: V. E2 End potential for sweep. Range: V. Scan cycle Definition of scan cycle: full double scan E1... E2... E1 half single scan E1... E2 Selection: full, half Scan rate Definition of scan rate in mv/s. Selection: 1, 2, 5, 10, 20, 50 mv/s Oven Switches column oven on/off and sets operating temperature. Range: C (> room temp C) Range Selection of measuring range: pa pa. na na. µa µa. 43

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator

More information

Portable Air Conditioner

Portable Air Conditioner Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE

More information

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam

More information

ChromQuest 5.0 Chromatography Data System

ChromQuest 5.0 Chromatography Data System ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks

More information

804 Ti Stand. Manual 8.804.8001EN

804 Ti Stand. Manual 8.804.8001EN 804 Ti Stand Manual 8.804.8001EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 804 Ti Stand Manual 8.804.8001EN 06.2010 ek Teachware

More information

SUB Universal water baths SBB Boiling baths JB, PB Unstirred water baths

SUB Universal water baths SBB Boiling baths JB, PB Unstirred water baths SUB Universal water baths SBB Boiling baths JB, PB Unstirred water baths Operating instructions Universal water baths SUB6: 6 litres SUB14: 14 litres SUB28: 28 litres SUB36: 36 litres Boiling baths SBB6:

More information

Operating Instructions Electronics module VEGADIF 65

Operating Instructions Electronics module VEGADIF 65 Operating Instructions Electronics module VEGADIF 65 Document ID: 36132 Differential pressure Contents Contents 1 About this document 1.1 Function.................................. 3 1.2 Target group..............................

More information

Brake module AX5021. Documentation. Please read this document carefully before installing and commissioning the brake module!

Brake module AX5021. Documentation. Please read this document carefully before installing and commissioning the brake module! Documentation Brake module AX5021 Please read this document carefully before installing and commissioning the brake module! Version : 1.2 : 2012.03.05 Date Article-no. : TDmlAX-5021-0000-0200 Page 2/8

More information

Electronic Pressure Switch EDS 300

Electronic Pressure Switch EDS 300 Electronic Pressure Switch EDS 300 User manual Page 2 of 16 Content 1. Functions of the EDS 300...3 2. Mounting...3 3. Operating keys on the membrane keypad...4 4. Digital display...4 5. Output function...5

More information

Installation and safety instructions for AC/DC built-in devices

Installation and safety instructions for AC/DC built-in devices The device type and date of manufacture (week/year) can be found on the device rating plate. In the event of any queries about the device, please quote all the details given on the rating plate. For further

More information

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting

More information

VENTIsupport Version 1

VENTIsupport Version 1 Description and Operating Instructions VENTIsupport Version 1 Compliance and titration software Content 1. Description.............................2 1.1 Purpose............................2 1.2 Labelling............................2

More information

Rousseau 10. User Guide

Rousseau 10. User Guide Rousseau 10 User Guide Base station overview ➀ LED signals White, permanent light: - Base station ready, you can register a handset Slow flashing white light: - Paging call in progress - Active call (you

More information

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05 RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,

More information

Monitoring Network DMN

Monitoring Network DMN Monitoring Network DMN User Manual Table of contents Table of contents... 2 1. Product features and capabilities... 3 2. System requirements... 5 3. Getting started with the software... 5 3-1 Installation...

More information

Quickstart. English Deutsch Français

Quickstart. English Deutsch Français Type T-CUT Tubular/pipe modules for micro and ultrafiltration Tubular-/Rohrmodule für die Mikro- und Ultrafiltration Modules tubulaires pour micro- et ultrafiltration Quickstart English Deutsch Français

More information

FCC COMPLIANCE STATEMENT FOR AMERICAN USERS

FCC COMPLIANCE STATEMENT FOR AMERICAN USERS FCC COMPLIANCE STATEMENT FOR AMERICAN USERS This equipment has been tested and found to comply with the limits for a CLASS A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed

More information

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011 Instruction Manual 2in1 LAN Tester & Multimeter Model: LA-1011 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.

More information

OPL BASIC. Dosing System for Professional Laundry machines. Contents

OPL BASIC. Dosing System for Professional Laundry machines. Contents OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries

More information

ORGANIC SAMPLE PREPARATION

ORGANIC SAMPLE PREPARATION ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,

More information

GenStore Archive Cabinet

GenStore Archive Cabinet GenStore Archive Cabinet Setup and Maintenance Manual Version A April 2014 Two GenStore Archives shown mounted on optional table stand. 2 Contents Getting Started... 4 Using this Manual...4 Safety Precautions...4

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CCS Customer Support : 1-800-474-2147 Voltage rating: 115V~60Hz Power rating : 1400W For product inquiries or support

More information

FT-100 Fire Training Fog Machine User Manual English

FT-100 Fire Training Fog Machine User Manual English FT-100 Fire Training Fog Machine User Manual English 2015 Antari Lighting and Effects Ltd. User Manual - English Safety Information Please read the following safety information carefully before operating

More information

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510 User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Infrarot-Bewegungsmelder IP44

Infrarot-Bewegungsmelder IP44 Infrarot-Bewegungsmelder IP44 infrared motion sensors IP44 ODA (weiß) slim ODA (schwarz) slim 95174 96000 ODA (weiß) ODA (schwarz) 95175 96001 Betriebsanleitung User s Manual User s Manual infrared motion

More information

Multi-Protocol decoder 76 200 with Load regulation

Multi-Protocol decoder 76 200 with Load regulation Multi-Protocol decoder 76 2 with Load regulation For locomotives with universal motors on digital layouts operating in the DCC and Motorola data format. Features 76 2 Load regulated multi-protocol decoder

More information

Keysight N4877A Clock Data Recovery and Demultiplexer 1:2

Keysight N4877A Clock Data Recovery and Demultiplexer 1:2 Keysight N4877A Clock Data Recovery and Demultiplexer 1:2 Getting Started Getting Started with the Keysight N4877A Clock Data Recovery and Demultiplexer 1:2 You only need a few minutes to get started

More information

ELECTRIC CENTRAL HEATING FLOW BOILER

ELECTRIC CENTRAL HEATING FLOW BOILER ELECTRIC CENTRAL HEATING FLOW BOILER EKCO.T Used product can t be treated as general communal waste. Disassembled appliance has to be delivered to the collection point of electrical and electronic equipment

More information

ÄKTAexplorer, ÄKTApurifier

ÄKTAexplorer, ÄKTApurifier GE Healthcare ÄKTAexplorer, ÄKTApurifier and ÄKTAmicro Operating Instructions Original instructions Table of Contents Table of Contents 1 Introduction... 1.1 Important user information... 1.2 Regulatory

More information

MOUNTING AND OPERATING INSTRUCTIONS

MOUNTING AND OPERATING INSTRUCTIONS MOUNTING AND OPERATING INSTRUCTIONS CF 51/55 1. Delivery CF 51/55-1 calculator with battery (optionally with mains power supply) - 1 wall mounting bracket - package with material for sealing, screws, wall

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MN12CES / MN10CESWW Voltage rating: 120V~60Hz Power rating : 1100W (MN12CES) Power rating : 900W (MN10CESWW) Customer

More information

SVC400P/SVC800P. 4/8 Camera Live Tracking Vehicle DVR Installation Manual. Version 1.0

SVC400P/SVC800P. 4/8 Camera Live Tracking Vehicle DVR Installation Manual. Version 1.0 SVC400P/SVC800P 4/8 Camera Live Tracking Vehicle DVR Installation Manual Version 1.0 1. MAIN FEATURES... 2 2. PRODUCT OVERVIEW... 2 3. DIMENSIONS... 4 4. PACKAGE CONTENTS... 5 5. MOUNTING AND ENVORNMENTAL

More information

Oil and Coolant Circulating Heating System. Model - OCSM

Oil and Coolant Circulating Heating System. Model - OCSM Oil and Coolant Circulating Heating System Model - OCSM Installation & Operation Manual 216280-000 REV 2 Identifying Your System The HOTSTART heating system is designed to heat fluids for use in marine

More information

Online UPS. PowerWalker VFI 1000R/1U

Online UPS. PowerWalker VFI 1000R/1U Online UPS PowerWalker VFI 1000R/1U Manual Uninterruptible Power Supply System Table of Contents 1. Important Safety Warning... 1 1-1. Transportation... 1 1-2. Preparation... 1 1-3. Installation... 1 1-4.

More information

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss heating

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss heating MAKING MORN LIVING POSSIBLE Danfoss heating living connect Installation and User Guide Contents 1.0 System overview... 3 2.0 Overview of display and control buttons... 3 3.0 Installation - step by step...

More information

Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS

Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Technical Information sheet 0001/07/Rev01 ENG FUNCTION Series S001 and S002 circulation units are applied to the primary circuit of solar

More information

How to read this guide

How to read this guide How to read this guide The following shows the symbols used in this Quick start guide with descriptions and examples. Symbol Description Example P oint Reference Caution [ ] This symbol explains information

More information

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P2.6.3.2. Thermodynamic cycle Heat pump

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P2.6.3.2. Thermodynamic cycle Heat pump Heat Thermodynamic cycle Heat pump LD Physics Leaflets P2.6.3.2 Investigating the function of the expansion valve of the heat pump Objects of the experiment g To study the operational components of the

More information

HP UPS R1500 Generation 3

HP UPS R1500 Generation 3 HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications

More information

R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide

R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide (=7ÔWÌ) 1307.7039.62 06 Test & Measurement Quick Start Guide This Quick Start Guide describes the following converter

More information

USER MANUAL Stand Alone Power Supply PSQ 2909 / PSQ 3909 / PSQ 4909 PSQ 2920 / PSQ 3920 / PSQ 4920

USER MANUAL Stand Alone Power Supply PSQ 2909 / PSQ 3909 / PSQ 4909 PSQ 2920 / PSQ 3920 / PSQ 4920 USER MANUAL Stand Alone Power Supply PSQ 2909 / PSQ 3909 / PSQ 4909 PSQ 2920 / PSQ 3920 / PSQ 4920 [This page intentionally left blank] Warning for Your Protection 1. Read these instructions. 2. Keep these

More information

Hydraulic control unit series 0086-372-01

Hydraulic control unit series 0086-372-01 Technical Product Information No. 1290 EN Hydraulic control unit series 0086-372-01 Contents Page About this Technical Product Information (TPI) 2 The ORTLINGHAUS numbering system 2 About the product 3

More information

WE-350 Series ¼ Turn Electric Actuator

WE-350 Series ¼ Turn Electric Actuator WE-350 Series ¼ Turn Electric Actuator Operation and Installation Manual Pg 1 (Rev. 020113) Table of Contents 1.0 General 1.1 Pre-Installation Inspection 1.2 Storage 1.3 Features & General Information

More information

Model 1756 Test Lead Kit

Model 1756 Test Lead Kit Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Model 1756 Test Lead Kit Gerneral Purpose Test Lead Information Description These test leads allow you

More information

FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION

FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION Caution! UV-radiation of this device is in the range of UV-A (320-400 nm). Direct exposure to eyes shall therefore be avoided. UV protection glasses shall

More information

REGTRONIC OPERATING INSTRUCTIONS

REGTRONIC OPERATING INSTRUCTIONS REGTRONIC OPERATING INSTRUCTIONS 1.FEATURES Electrical connection: M12 8-pin connector Preset pressure range 0.05-10 bar with possible full scale regulation. 10 100 mbar adjustable deadband Supply pressure:

More information

Congratulations! Thank you!

Congratulations! Thank you! TM-47 ORDERCODE D1370 Congratulations! You have bought a great, innovative product from DAP Audio. The DAP Audio Microphone range brings excitement to any venue. Whether you want simple plug-&-play action

More information

Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001

Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001 Express5800/120Ed Rack Mount Kit Installation Procedures PN: 455-01607-001 Proprietary Notice and Liability Disclaimer The information disclosed in this document, including all designs and related materials,

More information

Manual for Fire Suppression & Methane Detection System

Manual for Fire Suppression & Methane Detection System Manual for Fire Suppression & Methane Detection System Fogmaker North America Post address: 150 Gordon Dr Exton, PA 19341 Delivery address: 150 Gordon Dr Exton, PA 19341 Tel: 610-265-3610 Fax: 610-265-8327

More information

Antenna Splitter ASA 1. Instruction manual

Antenna Splitter ASA 1. Instruction manual Antenna Splitter ASA 1 Instruction manual Contents Important safety instructions... 2 The ASA 1 active antenna splitter... 4 Delivery includes... 4 Operating elements... 5 Putting the ASA 1 into operation...

More information

CG-1171-20 PTFE Diaphragm Liquid Transfer Pump

CG-1171-20 PTFE Diaphragm Liquid Transfer Pump 3800 N. Mill Road Vineland, NJ 08360 Phone: 800.843.1794 Fax: 800.922.4361 CG-1171-20 PTFE Diaphragm Liquid Transfer Pump This is the safety alert symbol. It is used to alert you to potential personal

More information

VIEW. SLX300 SpeakerLinX IP Zone. Amplifier Installation and Setup Guide. AVoIP

VIEW. SLX300 SpeakerLinX IP Zone. Amplifier Installation and Setup Guide. AVoIP VIEW SLX300 SpeakerLinX IP Zone Amplifier Installation and Setup Guide TM AVoIP ClearOne 5225 Wiley Post Way Suite 500 Salt Lake City, UT 84116 Telephone 1.800.283.5936 1.801.974.3760 Tech Sales 1.800.705.2103

More information

Bubble King. User Manual

Bubble King. User Manual Bubble King User Manual TABLE OF CONTENTS 1. Before You Begin... 3 What Is Included... 3 Unpacking Instructions... 3 Claims... 3 Text Conventions... 3 Icons... 3 Document Information... 3 Product at a

More information

HERZ-Thermal Actuators

HERZ-Thermal Actuators HERZ-Thermal Actuators Data Sheet 7708-7990, Issue 1011 Dimensions in mm 1 7710 00 1 7710 01 1 7711 18 1 7710 80 1 7710 81 1 7711 80 1 7711 81 1 7990 00 1 7980 00 1 7708 11 1 7708 10 1 7708 23 1 7709 01

More information

Material test chamber with mechanical convection

Material test chamber with mechanical convection FP series 400 Material test chambers Material test chamber with mechanical convection A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and

More information

Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46. 1 Index... 2

Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46. 1 Index... 2 0932 Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46 1 Index 1 Index... 2 2 Description of the Inputs / Outputs... 4 2.1 Inputs / Module OMRON ID201 / ID211...4 2.2 Outputs / Module OMRON OC211...5

More information

TMS TANK MANAGEMENT SYSTEM

TMS TANK MANAGEMENT SYSTEM TMS TANK MANAGEMENT SYSTEM Page 1 of 9 Operating Instructions GENERAL The Tank Management System is a bespoke design to control, monitor and accommodate efficient storage and dispensing of TMS. FUNCTIONS

More information

Installation and User Instructions Unvented Electric Storage Water Heater Models: S10UNV, S15UNV.

Installation and User Instructions Unvented Electric Storage Water Heater Models: S10UNV, S15UNV. Installation and User Instructions Unvented Electric Storage Water Heater Models: S10UNV, S15UNV. MULTIPOINT Please read and understand these instructions before starting work. Please leave this leaflet

More information

http://www.coastalamusements.com

http://www.coastalamusements.com Nexus is a Registered Trademark of Coastal Amusements. OWNERS MANUAL 1935 Swarthmore Avenue, Lakewood, NJ 08701 (732)-905-6662 FAX (732)-905-6815 http://www.coastalamusements.com GAME DESCRIPTION Nexus

More information

Business Audio System: Music & Messaging MP3 Player. by Grace Digital Audio. User Guide. Model No. GDI-USBM10

Business Audio System: Music & Messaging MP3 Player. by Grace Digital Audio. User Guide. Model No. GDI-USBM10 Business Audio System: Music & Messaging MP3 Player by Grace Digital Audio User Guide Model No. GDI-USBM10 User Guide Contents Introduction 2 Safety & General Use Information 2 Features 3 Set Up & Operation

More information

Ingenico. User Guide 5100M. Secure transaction and payment solutions

Ingenico. User Guide 5100M. Secure transaction and payment solutions User Guide Ingenico 5100M Secure transaction and payment solutions Ingenico 5100 M Contents 1 Presentation... 5 1.1 Overview of Ingenico 5100... 6 1.2 Keyboard details and functionality... 7 2 Use...

More information

Mobile Data Power Model: MDP-25

Mobile Data Power Model: MDP-25 Mobile Data Power Model: MDP-25 Topic Section Features... 2 Operational Features Summary... 2 Back-up Battery Power Internal Charger Voltage Spike Protection RF Noise Filtering Warning of Imminent Loss

More information

FTDI VCP DRIVER (free) (WIN/MAC/LINUX) http://www.ftdichip.com/drivers/vcp.htm

FTDI VCP DRIVER (free) (WIN/MAC/LINUX) http://www.ftdichip.com/drivers/vcp.htm 002 - CONNECTING THE PRINTER Now that you have an idea what 3D printing entails, we can continue and connect the printer to your computer. First make sure you have a computer with a decent amount of RAM

More information

To reduce the risk of electrical shocks, fire, and related hazards:

To reduce the risk of electrical shocks, fire, and related hazards: Owner s Manual 1 Contents 1. Safety Notes 5 2. Introduction 6 3. Features 7 4. Installation Guide 8 5. Front Panel Explained 9 6. Rear Panel Explained 10 7. Software Control Panel 12 8. Advanced User Tips

More information

Instruction manual. Electrobloc EBL 100-2

Instruction manual. Electrobloc EBL 100-2 Instruction manual 0 Manual Electrobloc EBL 100-2 Electrobloc EBL 100-2 Type no. 911.503 Table of contents 1 Introduction..........................................2 2 Safety information.....................................2

More information

Operating instructions Platform/floor scales

Operating instructions Platform/floor scales KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Platform/floor scales KERN

More information

Vision Engravers and Routers PRE-Installation Guide. 2015 Vision Engraving & Routing Systems

Vision Engravers and Routers PRE-Installation Guide. 2015 Vision Engraving & Routing Systems Vision Engravers and Routers PRE-Installation Guide Revised: 8/19/2015 Vision Engravers and Routers PRE-Installation Guide All rights reserved. No parts of this work may be reproduced in any form or by

More information

Mitos P-Pump. product datasheet

Mitos P-Pump. product datasheet product datasheet page Product description 2 Main benefits 3 Why choose the Mitos P-Pump? 3 Closed-loop flow control 4 Product specifications 5 Accessories - overview 6 Mitos P-Pump Basic 7 Mitos P-Pump

More information

NEWTECH PRO 1-2-3 KVA ONLINE UNINTERRUPTIBLE POWER SUPPLY USER MANUAL

NEWTECH PRO 1-2-3 KVA ONLINE UNINTERRUPTIBLE POWER SUPPLY USER MANUAL NEWTECH PRO 1-2-3 KVA ONLINE UNINTERRUPTIBLE POWER SUPPLY USER MANUAL info@tuncmatik.com / www.tuncmatik.com Table of Contents 1. Important Safety Warning 2 1-1. Transportation 2 1-2. Preparation 2 1-3.

More information

B150 Series Commercial Brewer. Direct Plumb Kit

B150 Series Commercial Brewer. Direct Plumb Kit B150 Series Commercial Brewer Welcome Keurig Authorized Dealer! Direct Plumb Kit This manual provides: Installation of the Direct Plumb Kit (DPK) that allows you to connect your customer s B150 brewer

More information

MultiDim MKII V.2 ORDERCODE 50386

MultiDim MKII V.2 ORDERCODE 50386 MultiDim MKII V.2 ORDERCODE 50386 Congratulations! You have bought a great, innovative product from Showtec. The Showtec MultiDim brings excitement to any venue. Whether you want simple plug-&-play action

More information

Manual Ranging MultiMeter

Manual Ranging MultiMeter Owner s Manual Manual Ranging MultiMeter Model 82345 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product.! Safety! Operation! Maintenance!

More information

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of the blood 7 Procedure to test glucose 7 Procedure to

More information

Protocol: HPLC (amino acids)

Protocol: HPLC (amino acids) Page 1 of 6 1. Chemicals Composition M MW gram volume (ml) product code 0-pthaldialdehyde C 8H 6O 2 134.13 Sigma P0657 Perchloric acid HClO 4 3.5 100.46 Dilute 70% stock solution 1:1 with demi to obtain

More information

COMPUTHERM Q7 Programmable, digital room thermostat. Operating Instructions

COMPUTHERM Q7 Programmable, digital room thermostat. Operating Instructions COMPUTHERM Q7 Programmable, digital room thermostat Operating Instructions GENERAL DESCRIPTION OF THE THERMOSTAT The COMPUTHERM Q7 type switched-mode room thermostat is suitable to regulate the overwhelming

More information

TraceTek TTDM Series Leak Detection and Location Modules Replacement Parts Installation Instructions

TraceTek TTDM Series Leak Detection and Location Modules Replacement Parts Installation Instructions TraceTek TTDM Series Leak Detection and Location Modules Replacement Parts Installation Instructions TRACETEK TraceTek TTDM Replacement Parts General Information These instructions detail the steps to

More information

SA-9600 Surface Area Software Manual

SA-9600 Surface Area Software Manual SA-9600 Surface Area Software Manual Version 4.0 Introduction The operation and data Presentation of the SA-9600 Surface Area analyzer is performed using a Microsoft Windows based software package. The

More information

WB/WBS 3, 4, 5, 7, 8, 11, 14, 30

WB/WBS 3, 4, 5, 7, 8, 11, 14, 30 INSTRUCTION MANUAL FOR WB/WBS 3, 4, 5, 7, 8, 11, 14, 30 WATER BATH PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com

More information

SHOWER WATER HEATER MODEL BS 35 / 45 / 60 BS 35 E / 45 E / 60 E OPERATION AND INSTALLATION INSTRUCTIONS

SHOWER WATER HEATER MODEL BS 35 / 45 / 60 BS 35 E / 45 E / 60 E OPERATION AND INSTALLATION INSTRUCTIONS SHOWER WATER HEATER MODEL BS 35 / 45 / 60 BS 35 E / 45 E / 60 E OPERATION AND INSTALLATION INSTRUCTIONS 2 This water heater must be installed (water and electrical installation), commissioned and serviced

More information

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm 1. How to prepare straws for pronuclear stages and embryos (use straws with plug, 66.5 mm) Straws are

More information

IMPORTANT INSTRUCTIONS & OPERATING MANUAL. Houston 50 Inch Electric Wall Mounted Fireplace Black / White

IMPORTANT INSTRUCTIONS & OPERATING MANUAL. Houston 50 Inch Electric Wall Mounted Fireplace Black / White IMPORTANT INSTRUCTIONS & OPERATING MANUAL Houston 50 Inch Electric Wall Mounted Fireplace Black / White Model Number:MFE5050BK Model Number:MFE5050WH Read these instructions carefully before attempting

More information

AXIS 291 1U Video Server Rack Installation Guide

AXIS 291 1U Video Server Rack Installation Guide AXIS 291 1U Video Server Rack Installation Guide About This Document This document describes how to install Axis blade video servers in the AXIS 291 1U Video Server Rack. Updated versions of this document

More information

64-CHANNEL INTEGRATED AMPLIFIER

64-CHANNEL INTEGRATED AMPLIFIER Panasonic 64-CHANNEL INTEGRATED AMPLIFIER Operating Instructions P/N: SU-MED640E Caution for AC main lead FOR YOUR SAFETY, PLEASE READ THE FOLLOWING TEXT CAREFULLY This product is equipped with 2 types

More information

User Manual. Interface Adapter IA1 / Computer Interface EFOY 600 / EFOY 900 / EFOY 1200 / EFOY 1600

User Manual. Interface Adapter IA1 / Computer Interface EFOY 600 / EFOY 900 / EFOY 1200 / EFOY 1600 User Manual Interface Adapter IA1 / Computer Interface EFOY 600 / EFOY 900 / EFOY 1200 / EFOY 1600 2 1. Introduction 1. Introduction 1.1 Introduction Thank you for choosing an EFOY product. Please read

More information

Drayton Digistat +2RF/+3RF

Drayton Digistat +2RF/+3RF /+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer

More information

Tyan Computer. Transport PX22. Service Engineer s Manual

Tyan Computer. Transport PX22. Service Engineer s Manual Tyan Computer Transport PX22 Service Engineer s Manual 1 Precaution To read through the user manual, check all assembly and follow setup process before any operation on this server To keep paper clips,

More information

Visions become real. Temperature control and cooling systems. The innovative and reliable solution

Visions become real. Temperature control and cooling systems. The innovative and reliable solution Temperature control and cooling systems The innovative and reliable solution With our PulseTemp equipment you have the option of up to 72 cooling circuits in the impulse cooling, temperature control, temperature

More information

How To Use A Power Supply Unit (Upu)

How To Use A Power Supply Unit (Upu) BRAVER UPS (Uninterruptible Power System) User s Manual Safety CAUTION! This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user replaceable parts.

More information

Meaco 30L and Meaco 40L dehumidifier instruction manual

Meaco 30L and Meaco 40L dehumidifier instruction manual Meaco 30L and Meaco 40L dehumidifier instruction manual Please read this instruction manual before using the dehumidifier and keep safe for future reference SAFETY INSTRUCTIONS PLEASE READ ALL INSTRUCTIONS

More information

HCS-3300/3302/3304 USB Remote Programmable Laboratory Grade Switching Mode Power Supply

HCS-3300/3302/3304 USB Remote Programmable Laboratory Grade Switching Mode Power Supply 1. INTRODUCTION HCS-3300/3302/3304 USB Remote Programmable Laboratory Grade Switching Mode Power Supply User Manual This family of efficient, upgraded SMPS with small form factor, auto cross over CV CC,

More information

XPanel V2. Remote Control Panel. User Manual. XILICA Audio Design

XPanel V2. Remote Control Panel. User Manual. XILICA Audio Design XPanel V2 Remote Control Panel User Manual XILICA Audio Design Important Safety Instructions 1. READ THESE INSTRUCTIONS All the safety and operating instructions should be read before the product is operated.

More information

How To Use A Power Supply On A Powerline 2.2 (Ai)

How To Use A Power Supply On A Powerline 2.2 (Ai) KNX/EIB Product documentation Issue: 05.08.2010 629x1220 Push button sensor 3 comfort 1-gang Push button sensor 3 comfort 2-gang (1+1) Push button sensor 3 comfort 3-gang Push button sensor 3 comfort 4-gang

More information

Electronic Pressure Switch EDS 3000

Electronic Pressure Switch EDS 3000 For Your Information Electronic Pressure Switch EDS 3000 User Manual Revision: November 01, 2002 1. EDS 3000 Functions The EDS 3000 unit offers the following functions: Display pressure in psi, MPa, or

More information

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss Heating Solutions

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss Heating Solutions MAKING MODERN LIVING POSSIBLE living connect Installation and User Guide Danfoss Heating Solutions 2 living connect Thank you for buying a Danfoss product Danfoss Heating Solutions 3 Contents 1. System

More information

2 (. 3 4 5 . -,.. -,. 3. : 1 2 3 > 4, :!! 1,, 17 Lt MAX 2 3 4,, > 5.,,, 6 :, :1)! 2)! 7,, 1, :,. 2,, Bottom of steam generator 7 8 9 10 11 12 automatically and this phenomenon will continue for

More information

RISH CON - Hz. Salient Features : Application : Product Features: FREQUENCY TRANSDUCER

RISH CON - Hz. Salient Features : Application : Product Features: FREQUENCY TRANSDUCER Application : The RISH CON - Hz transducer is used for frequency measurement. The output signal is proportional to measured frequency and is either load independent DC Current or load independent DC Voltage.

More information

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions Description The LS-1 Series of tungsten halogen light sources are versatile, white-light lamps optimized for use in the

More information

CL90i. 77-021 Please read these instructions before operating the product. 3 - Beam Self-Leveling Cross Line Laser

CL90i. 77-021 Please read these instructions before operating the product. 3 - Beam Self-Leveling Cross Line Laser 3 - Beam Self-Leveling Cross Line Laser CL90i 77-01 Please read these instructions before operating the product Self-Leveling GB D F I E PT NL DK SE FIN NO PL GR CZ RU HU SK SI BG RO EE LV LT Contents

More information