Design and Evaluation of Robotic Steering of a Flexible Endoscope

Size: px
Start display at page:

Download "Design and Evaluation of Robotic Steering of a Flexible Endoscope"

Transcription

1 Design and Evaluation of Robotic Steering of a Flexible Endoscope Jeroen Ruiter, Esther Rozeboom, Mascha van der Voort, Maarten Bonnema, Ivo Broeders Abstract In current flexible endoscopy the physician faces problems in manipulating the endoscope. A lot of experience is required to master the procedure. This paper describes the design of an add-on robotic module that improves the user interface of traditional endoscopes and allows a single physician to operate it easily. We identified critical user aspects of traditional endoscopes that need to be copied in a robotic setup. In our design the physician uses a remote control that is connected to a light drive system. It allows manipulation of the robotic endoscope in space. We conducted an experiment to judge the usability of our system. Results indicate that robotic steering, using a position-controlled touchpad or a ratecontrolled joystick increases efficiency and satisfaction. I I. INTRODUCTION N flexible endoscopy the interior surfaces of the gastrointestinal, reproductive and respiratory tracts are assessed. The physician uses a flexible endoscope with a camera at the steerable distal tip that is introduced in the natural body openings. Experience is necessary to master the procedure of inserting, steering, and executing small interventions with a flexible endoscope. Performing a colonoscopy to inspect the large intestine is one of the most demanding diagnostic endoluminal procedures performed in the clinic (Fig. 1). According to [1]-[2] about 100 to 200 procedures are required to reach the level to perform the procedure safely and within reasonable time. No revolutionary changes have occurred in endoscope technology during the last five decades. Main usability problems are related to the control section at the proximal end of the flexible endoscope. To obtain single person endoscope control, the control section is designed for the left hand, while the right hand manipulates the flexible shaft. Because of the configuration of control elements the physician often faces handling problems. For instance, approximately 20% of the physicians are using both hands for the control section, while an assistant manipulates the shaft according to spoken instructions [3]. This is undesirable since introducing the flexible endoscope into the tortuous and elastic colon is a delicate task that requires interpretation of force feedback information to support Manuscript received January 31, This work was supported by the Dutch Ministry of Economic Affairs and the Province of Overijssel, within the Pieken in de Delta (PIDON) initiative. Jeroen Ruiter (phone: +31(0) , fax +31(0) , j.g.ruiter@utwente.nl), Mascha van der Voort and Maarten Bonnema are with the Laboratory of Design, Production and Management, Department of Engineering Technology, University of Twente, Enschede, The Netherlands. Jeroen Ruiter is also with DEMCON advanced mechatronics, Oldenzaal, The Netherlands. Esther Rozeboom and Ivo Broeders are with the Laboratory of Minimal Invasive Surgery & Robotics, Department of Technical Medicine, University of Twente, The Netherlands. Fig. 1. Colonoscopy procedure: The physician uses a flexible endoscope to inspect the large intestine of a patient. steering. Inefficient steering may lead to time loss and excessive stretching of the intestinal wall, leading to increased patient discomfort [4]. At present there are no flexible endoscopes available that can be controlled in an intuitive and user-friendly way by one person. A long term solution is to redesign the control section. However, despite the limited usability, current flexible endoscopes are widely used systems with qualities like cleanability, maneuverability, and good image quality [5]. We propose an add-on robotic module positioned on a traditional endoscope. The physician uses a remote control to actuate the bendable tip. Key factor is that tip steering and tip actuation are mechanically decoupled and computer intelligence is integrated. Robotic steering has the potential to improve usability, preserve current endoscope qualities, and prevent high costs related to replacement of endoscopic equipment. Acceptance is expected to be high since our robotic setup fits to the current workflow and infrastructure. Allemann et al. [6] have developed a system with a game joystick to control a motorized traditional endoscope. In their evaluation both novices and experienced physicians required significantly more time to complete a given task when using a joystick compared to conventional controls. They concluded that possibly the limited maneuverability of the endoscope positioned in the setup is responsible for the disappointing results. Zhang et al. [7] performed a comparable experiment with a joystick controller and a motorized endoscope with a fixed position in the setup. They concluded that the time required to finish the process relies on the degree of familiarity with the robot system. After 3-5 test runs an expert in flexible endoscopy performs equal in both techniques. Nevertheless, the lack of proprioceptive

2 feedback in robot supported manipulation was indicated to decrease the effectiveness of the system. Reilink et al. [8] conducted an experiment with a six degrees of freedom haptic controller, coupled to a stationary motorized traditional endoscope. Experts appeared faster when using the conventional steering method compared to motorized steering methods. Students who had done flexible endoscopy training showed no significant differences. In all above work endoscope handling opportunities were inferior to the current manual steering design. We propose a hybrid setup in which mobile as well as stationary use of the robotic endoscope is possible. The control section of the robotic endoscope can be manipulated freely to resolve for instance shaft looping inside the lumen. The experimental setup in the studies mentioned was not designed to be implemented in clinical practice. Our robotic endoscope is based on the clinical workflow and integrates medical and technical state of the art. This paper discusses the design and evaluation of such a robotic flexible endoscope. In Section II shortcomings of the current user interface are discussed. Section III describes the design considerations for our system. In Section IV the design is specified and shown. Section V contains the results of the experiments that are conducted. Finally, Section VI concludes and provides directions for further work. II. CURRENT USER INTERFACE SHORTCOMINGS This section discusses the current problems related to physician-instrument interaction. The left hand steers the distal tip by turning two navigation wheels on the control section (Fig. 2). The control of the tip orientation is not very intuitive because the navigation wheels are arranged in the same plane while the bendable portion will bend in two perpendicular directions. Single handed operation of the wheels is difficult due to size, position and force requirements, especially with small hands. In a survey of U.S. gastroenterology fellows, 41% of the respondents considered their hands too small for a standard endoscope s control section [9]. Some endoscopists release the grip of the right hand on the shaft and use it to turn the smaller outer wheel. The shaft position is maintained by trapping it between the physician s thigh and the examination table [10]. Other physicians use torque steering as an alternative technique. They turn the large navigation wheel, while the small wheel is locked in neutral position, and additionally torque the shaft of the endoscope [11]. The control section also contains manual operated valves to insufflate air or CO2 into the lumen, to rinse the camera lens, to suck fluids and gas out of the lumen, and some programmable switches for functions like taking a picture or switching to narrow band imaging. This complicates single handed operation even further [10]. As a consequence, manipulation of flexible endoscopes is associated with awkward body rotations, bending of the knees, and a variety of arm movements. These movements are in the clinic sometimes referred to as the endoscopy Fig. 2. Control section endoscope: (1) Programmable switches, (2) Navigation wheels, (3) Valves for insufflation, rinsing, and suction, (4) Steerable tip with camera. dance and are especially performed during difficult parts of the procedure. The prevalence of musculoskeletal complaints has been shown to be higher for endoscopists than for other medical specialties [12]. Steering the navigation wheels and operating the control buttons of the control section require repetitive, extreme and prolonged wrist and finger flexion or extension. Manipulation of the shaft of a flexible endoscope is associated with awkward wrist, shoulder, and neck postures. In a questionnaire under colonoscopists concerning work related injuries, 226 out of the 608 respondents reported physical complaints obtained by performing colonoscopy. Most injuries were related to torquing the shaft and turning the dials [3]. It can thus be concluded that current endoscope handling is not ergonomic and user friendly. Physicians have learned to overcome the drawbacks, but at the expense of personal well-being. III. DESIGN DIRECTIONS FOR ROBOTIC STEERING In this section the opportunities for robotics for intuitive and user-friendly single person endoscope handling are discussed. Many alternative endoscopes have been developed to improve colonoscopy physically and technically for the operator and make it more comfortable for patients. However, none of them is commercial available. Gaglia et al. [13] highlight technical innovations of new endoscopic devices. All described systems are designed to be less skill dependent compared to the user interface of conventional endoscopes, but all of them are also based on a new design of the endoscope. As stated in the introduction, we believe that acceptance is higher if conventional endoscopes can be used. The human machine interface of the robotic endoscope should allow the physician to operate cooperatively with the robot. Thus, ergonomics and integration into the clinical workflow are essential elements of a successful design [14].

3 Fig. 3. Stationary (left) versus mobile use of the robotic endoscope In the clinic the robotic module and the flexible endoscope are connected during preparation. The mechanical interfaces between the clean endoscope, the drive unit, and the user interface are critical. Even in case of non-sterile endoluminal interventions the patient should be protected against cross contamination. After the procedure, the assistant dismantles the robotic system for cleaning or replacement of disposable parts. The endoscope will be cleaned or disinfected according to the current clinical workflow. Direct manipulation of the endoscope handle may be required to pass difficult parts of the lumen. We propose a hybrid setup that is configurable during the procedure. In one configuration the endoscope including the add-on robotic module is positioned in a docking station and the physician holds the remote control in one hand and the shaft in the other hand. In the other configuration the physician carries the robotic endoscope with the remote control that is directly coupled to the control section of the endoscope, as shown in Fig. 3. The former is easy to carry while the latter allows for extra maneuverability of the endoscope. Possibly this is beneficial during insertion of the endoscope. If necessary during a procedure the setup can also be changed to conventional steering by taking the endoscope in a few seconds out of the robot. The remote control should be intuitive and suitable for single handed use. It ought to reduce musculoskeletal complaints of the operator and has to be operated close to the patient. All input controls of the control section of the current endoscope should be included in the remote control. During insertion the physician should be able to actuate insufflation, suction or rinsing while steering the tip. In current practice the flow is controlled by repeatedly actuation of the buttons (digital input). In the robotic setup flow could be controlled proportional (analog input). In particular precise (limited) insufflation contributes to successful insertion [15] and minimal patient discomfort [16]. The remote control can be operated while wearing disposable gloves and should allow for left as well as right handed use. The delicate task of shaft manipulation, that requires precise interpretation of force feedback information, can always be done with the dominant hand. This optimizes the setup for the about 10% left handed physicians [17]. Different kind of input controllers that are used for computer applications (e.g. mouse, joystick, haptic controller) are used for clinical systems as well. New innovations are related to speech, gaze, and gesture control. The clinical application of these innovative techniques is limited, mainly because of safety issues, like limited accuracy and robustness. The input controller that steers the tip is a critical component with regard to usability and intuitive use. It should be able to manipulate the tip from -180 to +180 in left-right, up-down and combined (diagonal) directions. The controller should allow for fast large movements, precise small movements and stabilization of the bendable tip in a preferred position. For instance during inspection of the lumen, the tip should follow a smooth circular path. Thereby providing images of the entire mucosal surface within reasonable time and allowing for precise camera positioning to inspect suspicious areas. In current practice the physician relates the forces required to turn the navigation wheels of a traditional endoscope to the shape of the tip and shaft inside the body. The navigation wheels transmit actuation forces to the tip by means of flexible Bowden cables. The force increases by friction in accordance with the degree of flexion of the endoscope. This helps the physician to estimate the flexion of the tip, interaction forces of the tip with tissue, and shaft loops that need to be straightened [18]. Force information from the navigation wheels need to be fed back to the physician in the robotic setup. Ideally this would be haptic feedback to achieve a sense of transparency but a haptic controller with at least 2 degrees of freedom that can be integrated with a small remote control is not available. Vision could provide an appropriate sensory substitute in the robotic setup. Indication bars reflecting force information are shown on the monitor. We have to verify in our setup if vision can adequately replace haptic information. The setup limits the number of controllers suitable to steer the tip. A thumb joystick as seen in gamepad controllers or a touchpad like in laptops are the most likely options. Position and rate control are the two common transfer functions. In position control the input device indicates the desired position of the end effector, whereas in rate control the input device indicates the desired end effector velocity. There is no upfront evidence in literature to choose between these alternatives [19]. In our setup a thumb joystick combines best with rate control. It allows the physician to use the full bending range of the tip of the endoscope. Additionally, rate control can freeze the tip in a preferred position when releasing the joystick. The joystick, with spring loaded return-to-center functionality, returns to its initial position and sets the speed to zero. A touchpad can be best combined with incremental position control. Like in mouse navigation, clutching allows the physician to use the full manipulation range. Lifting the finger fixates the tip of the endoscope into position. In our experiment we try to determine the best control option. Above considerations are implemented in our design of a robotic endoscope, as described in the next section.

4 Fig. 4. System overview Fig. 4 IV. DESIGN In Fig. 4 our system is depicted. The configuration is designed to obtain a light robotic endoscope that can be manipulated freely by the operator. For that reason all heavy components, like motors, are placed in a stationary positioned motor unit that is connected through a flexible transmission to a compact and light mobile drive unit. If the two motors for navigation wheel actuation would be positioned in the mobile drive unit it would add about 0.7 kg to the weight. The motor unit is placed on the endoscopy cart and the generic mobile drive unit connects with a dedicated interface to the navigation wheels of each individual type of endoscope. On top of the interface unit the holder of the remote control is positioned. Table 1 contains an overview of some general specifications of the designed mobile drive unit. TABLE 1 SPECIFICATIONS OF THE DESIGNED MOBILE DRIVE UNIT Maximum torque on wheels 1.5 Nm Range of motion 360 (omnidirectional) Maximum angular velocity 2π rad/s Weight mobile drive unit 0.92 kg A. Drive system Antagonistic cable pairs between the stationary motor unit and the mobile drive unit actuate the navigation wheels of the endoscope (Fig. 5). The cables are pretensioned to prevent backlash and delay in control. The outer sheath of the Bowden cables at the load side are supported by load cells to measure the applied force to the navigation wheels. The drive system is self-locking so the position of the tip of the endoscope (camera position) is maintained when the controller of the remote control is not actuated. Two optical encoders are added to the load side to be able to improve control. In the current setup these are not in use. Two DC servo motors were selected for actuation. The motors, motor controllers and power supply are all integrated in the motor unit box. The main program is computed on an external computer. Fig. 5. Mobile drive unit: (1) Optical disc encoders, (2) Bowden cables, (3) Force sensors. Fig. 6. Exploded view coupling mechanism: (1) Drive unit, (2) Remote control, (3) Holder remote control, (4) Plug, (5) Sterile interface, (6) Docking station, (7) Sterile bag, (8) Endoscope. B. Coupling mechanism The drive system cannot be sterilized. A sterile interface couples the drive unit to the clean endoscope to prevent cross contamination. If preferred the drive unit is sealed in a bag as shown in Fig. 6. The interface is first connected to the endoscope. It is locked by an endoscope specific plug that bridges the valves of insufflation, rinsing, and suction on the endoscope so these can be controlled with the remote control. This assembly is subsequently connected with the mobile drive unit by threaded knobs. Torque between drive unit and interface is transferred with a pin hole connection. A configurable holder for the remote control is integrated with the interface. The physician is able to position it to personal preferences to comfortably hold and carry the robotic endoscope. If preferred the robotic endoscope is positioned in a docking station on a pole cart and the remote control can be detached from the holder. The docking station allows axial rotation of the shaft of the endoscope that is induced by the physician during the procedure.

5 Fig. 7. Joystick and touchpad remote controllers: (1) Thumb joystick, (2) Buttons for insufflation, rinsing, and suction, (3) Programmable switches, (4) USB cable to computer, (5) Touchpad, (6) Hold-to-run safety switch. Fig. 9. Experimental setup feedback to support control of the robotic endoscope. These are integrated in a single monitor (Fig. 4). The endoscopic images that visualize patient tissue are most important and take up most surface of the monitor. The endoscope and its imaging unit determine the characteristics of these images. Often these are provided in high definition. One fourth of the monitor is reserved for additional feedback provided by the robotic system. The flexion of the endoscope tip is shown in a bending diagram. The diagram shows a bar in a white circle that extends from the center into the direction that the tip is moving. The direction and length of the bar are an indication for tip direction and flexion respectively. Additional bar indicators provide information about the torque required to turn the navigation wheels. This relates to the shape of the shaft and interaction of the endoscope tip with tissue. Flow information of insufflation, rinsing, and suction is also fed back by bar indicators. V. EVALUATION Fig. 8. Endoscopy cart: (1) Solenoid valve suction, (2) Motors Bowden cables, (3) Solenoid valves insufflation and rinsing, (4) Water container, (5) Pump unit, (6) Air/CO 2 gas cylinder. C. Remote control The remote control allows single handed control of all available functionality of a traditional endoscope, as shown in Fig. 7. A thumb joystick as well as a touchpad can be integrated as input controller to steer the tip. Push buttons are arranged to operate all valves and switches of the control section of the current endoscope. Buttons for proportional insufflation, rinsing, and suction are pressure sensitive by means of an underlying force sensing resistor. Flow is controlled with solenoid pinch valves that are positioned in the motor unit (Fig. 8). A hold-to-run safety switch needs to be pressed during operation of the robotic endoscope to prevent unintended actuation of input controls. D. Feedback information The operator is provided with several sources of visual A. Experimental setup This section describes the experiments conducted to determine the optimal settings of our robotic endoscope and to assess its intuitiveness and user-friendliness. We compared conventional steering of the tip to robotic steering to obtain knowledge about the best input controller and the required maneuverability of the endoscope. The tested setups were: 1) Conventional steering with navigation wheels. We use this method as a reference for the robotic setups. 2) Joystick steering with stationary endoscope. The endoscope including the drive unit is positioned in a docking station and the subject only holds the remote control, as shown in Fig. 3. 3) Touchpad steering with stationary endoscope. 4) Joystick steering with mobile endoscope. The subject carries the endoscope with the remote control that is directly coupled to the control section of the endoscope, as shown in Fig. 3 and Fig. 9. 5) Touchpad steering with mobile endoscope. A standard flexible colonoscope (Exeria II CF-H180AL, Olympus, Tokyo, Japan) and imaging unit (Exeria II CLV-

6 TABLE 2 QUANTATIVE RESULTS EXPERIMENT Joystick - stationary Touchpad - stationary Setup Conventional Joystick - (n=24) scope (n=12) scope(n=12) mobile scope (n=12) Insertion time (s) a 178 (75) 145 (50) 188 (94) 150 (70) 108 (68) Workload (max. 25) a 16 (4) 13 (4) 10 (4) 11 (4) 15 (4) Detection rate (%) a 68 (16) 86 (22) 71 (22) 64 (26) 71 (19) a Values are represented as median (standard deviation) Touchpad - mobile scope (n=12) 180, Olympus, Tokyo, Japan) were used for all experimental conditions. Subjects, without experience in handling an endoscope and without medical background, were asked to perform 2 colonoscopy tasks on an anatomical model (M40, Kyoto Kagaku, Kyoto, Japan). The absence of experience enabled testing of intuitiveness. First, subjects had to advance and steer the endoscope up to the point where the colon starts, the cecum. Secondly the endoscope had to be retracted to the rectum while inspecting the mucosal surfaces for lesions, represented by 7 prepositioned red blocks sized 2x2x1 mm. It was too time consuming to test all setups on all individual participants. For that reason the population of 24 subjects (aged years, 7 women and 17 men) was divided over 2 groups. One group tested setup 1, 2 and 5. The other group tested setup 1, 3 and 4. This way all participants experienced both input controllers and both settings of endoscope handling. Each of the six possible orders of the three conditions was performed equally often to correct for learning effects and fatigue. For each setup 5 minutes of practice time was available and the opportunity to ask for advice on usage. An easy bowel configuration was chosen in which all subjects could complete the task. Our focus was to test the steering usability of the endoscope. Future experiments by experienced physicians will be more challenging to test also the (clinical) usability of features like insufflation, suction and force feedback information that are typically required in difficult procedures. These functionalities were not available in this novices experiment. Usability is defined by the International Standardisation Organisation (ISO) as: the extent to which a product can be used by specific users to achieve goals with effectiveness, efficiency, and satisfaction in a specified context of use [20]. The three factors are widely accepted to concern distinct measures and were evaluated in the experiment [21]. In our experiment the following dependent variables were measured: - Full insertion into the cecum (effectiveness) - Detection rate of lesions (effectiveness) - Insertion time (efficiency) - Subjective workload analysis measuring mental and physical demand, performance, effort and frustration. Based on assigning scores to a modified NASA Task Load Index, [22] (efficiency) - Rank interfaces according to preference (satisfaction) - Questionnaire by interview (satisfaction) B. Results and discussion The quantitative results of the experiment are depicted in Table 2. The results indicate that robotic steering by novices improves insertion time and workload experience. Compared to the conventional method, joystick steering with stationary endoscope and touchpad steering with mobile endoscope had significantly faster insertion times. The workload scoring of all robotic setups are significantly better than the conventional method, except for touchpad steering with mobile endoscope. The detection rate was not significantly affected by robotic steering. All subjects performed full insertion into the cecum. Extensive maneuvering of the endoscope shaft was not required during insertion. For that reason nothing conclusive can be said on the necessity of a mobile scope. Future experiments with physicians in a challenging procedure should provide more knowledge. In the interviews all participants complain about the additional weight that needs to be carried in the mobile endoscope setup. One might consider free manipulation only in awkward circumstances such as looping of the shaft. During easy parts of the procedure the scope is docked. Despite the additional weight, novices appreciated the ergonomics and work posture of all robotic setups more than of the conventional setup. The conventional steering method would be the first choice for one subject, and last in 16 of 24 subjects. Subjects picked the joystick control (mobile and stationary) first in 15 of 24 cases. Although subjects prefer joystick to touchpad control, data on performance does not endorse that outcome. Possibly, the reduced proprioceptive feedback in touchpad control limits the feeling of being in control. In addition, subjects tend to roll the thumb during touchpad control instead of only moving the tip of the thumb over the touch surface. In this case, tip movement will not occur as expected, since the center of the touched surface is not moved as intended. In the mobile setup, rolling the thumb is restricted by the additional weight and the imposed position of the hand with respect to the endoscope, possibly explaining the faster insertion time of the mobile compared to the stationary setup. A system that encourages steering with the tip of the thumb will likely lead to improved touchpad control. Almost all participants thought that the joystick as well as the touchpad controller were too sensitive, but they valued the intuitiveness of operation and experienced no delay in tip response. Previous experience with game controllers or touch interfaces did not influence the outcome. Some

7 suggested that up-down in joystick control should be reversed to copy flight control. Sensitivity as well as updown direction could be made adaptable to comply with user preferences. Force feedback information from the navigation wheels was not available in the robotic setup in our experiment. However, the bending diagram, as described in Section IV.D, was shown during the experiment. We estimated that feedback on the extent to which the tip is bent was essential even in an easy bowel configuration. Novices appreciated this substitute for haptic feedback very much. VI. CONCLUSION AND FUTURE WORK A robotic system is built that allows ergonomic single person control while preserving current endoscope qualities. Acceptance is expected to be high since our robotic setup can easily be implemented in the current clinical workflow. We showed that robotic steering by novices, using touchpad or joystick control, increases efficiency and satisfaction. The effectiveness was not significantly affected by robotic steering. Our results did not show a clear preference for a position-controlled touchpad or a rate-controlled joystick. We will perform additional experiments in which we will critically look at the type of input controller and the accompanying control algorithm. Breaking the mechanical linkage and integrating computer intelligence between operator and end effector provides opportunities for improved usability. However, we also identified critical user aspects of traditional flexible endoscopes that are related to mechanical interfaces, like force feedback. These are probably required in a robotic setup, especially in challenging procedures. We will work on optimizing functionality proposed in Section IV that was not yet implemented (e.g. insufflation) or facilities that could not be well tested in this experimental setup (e.g. mobile versus stationary endoscope). Expert testing is required to test performance in clinically relevant advanced procedures. REFERENCES [1] P. S. Tassios, S. D. Ladas, I. Grammenos, K. Demertzis, and S. A. Raptis, Acquisition of competence in colonoscopy: the learning curve of trainees, Endoscopy, vol. 31, pp , Nov [2] G. C. Harewood, Relationship of colonoscopy completion rates and endoscopist features, Digestive Diseases and Sciences, vol. 50, pp , Jan [3] A. S. Liberman, I. Shrier, and P. H. Gordon, Injuries sustained by colorectal surgeons performing colonoscopy, Surgical endoscopy, vol. 19, pp , Oct [4] W. H. Kim, Y. J Cho, J. Y. Park, P. K. Min, J. K. Kang, and I. S. Park, Factors affecting insertion time and patient discomfort during colonoscopy, Gastrointestinal Endoscopy, vol. 52, pp , June [5] B. Bardou, F. Nageotte, P. Zanne, and M. de Mathelin, Design of a telemanipulated system for transluminal surgery, in Proc. of 31st Annual International Conference of the IEEE EMBS, Minnesota, USA, 2009, pp [6] P. Allemann, L. Ott, M. Asakuma, N. Masson, S. Perretta, B. Dallemagne, D. Coumaros, M. De Mathelin, L. Soler, and J. Marescaux, Joystick interfaces are not suitable for robotized endoscope applied to NOTES, Surgical Innovation, vol. 16, pp , June [7] S. H. Zhang, D. X. Wang, Y. R. Zhang, Y. H. Wang, Y.G. Wang, X.P. Ma, The human machine interface implementation for the robot assisted endoscopic surgery system. In Proc. of the International Workshop on Robot and Human Interactive Communication, Berlin, Germany, 2002, pp [8] R. Reilink, S. Stramigioli, A. M. L. Kappers, and S. Misra, Evaluation of flexible endoscope steering using haptic guidance, The International Journal of Medical Robotics and Computer Assisted Surgery, vol. 7, pp , June [9] D. Cohen, J. R. Naik, L. J. Tamariz, R. D. Madanick, The perception of gastroenterology fellows towards the relationship between hand size and endoscopic training, Digestive Diseases and Sciences, vol. 53, pp , July [10] C. Williams, Insertion technique, In Colonoscopy: Principles and Practice, 2nd ed., J. D. Waye, D. K. Rex, and C.B. Williams, 2nd edition, Chichester: Wiley-Blackwell, 2009, pp [11] A. K. Shergill, K. R. McQuaid, and D. Rempel, Ergonomics and GI endoscopy, Gastrointestinal Endoscopy, vol. 70, pp , July [12] S. L. Hansel, M. D. Crowell, D. S. Pardi, E. P. Bouras, and J. K. DiBaise, Prevalence and impact of musculoskeletal injury among endoscopists: a controlled pilot study, Journal of Clinical Gastroenterology, vol. 43, pp , May [13] A. Gaglia, I. S. Papanikolaou, and W. Veltzke-Schlieker, New endoscopy devices to improve population adherence to colorectal cancer prevention programs, World Journal of Gastrointestinal Endoscopy, vol. 2, pp , July [14] G. D. Hager, A. M. Okamura, P. Kazanzides, L. L. Whitcomb, G. Fichtinger, and R. H. Taylor, Surgical and interventional robotics: Surgical assistance systems, IEEE Robotics and Automation Magazine, vol. 15, pp , Dec [15] I. C. Roberts-Thomson and Edward Teo, Colonoscopy: Art or science?, Journal of Gastroenterology and Hepatology, vol. 24, pp , Feb [16] M. Bretthauer, G. S. Hoff, E. Thiis-Evensen, G. Huppertz-Hauss, and E. Skovlund, Air and carbon dioxide volumes insufflated during colonoscopy, Gastrointestinal Endoscopy, vol. 58, pp , Aug [17] C. Hardyck and L. F. Petrinovich, Left handedness, Psychological Bulletin, vol. 84, pp , May [18] O. Körner and R. Männer, Implementation of a haptic interface for a virtual reality simulator for flexible endoscopy, in Proc. of 11th International Symposium on Haptic Interfaces for Virtual Environment and Teleoperator Systems, Los Angeles, USA, 2003, pp [19] S. Zhai, Human performance in six degree of freedom input control, Ph.D. dissertation, Dept. of Industrial Engineering, University of Toronto, [20] Ergonomic requirements for office work with visual display terminals (VDTs) - Part 11: Guidance on usability, ISO :1998(E). [21] K. Hornbæk, Current practice in measuring usability: Challenges to usability studies and research, International Journal of Human- Computer Studies, vol. 64, pp , Feb [22] S. G. Hart and L. E. Staveland, Development of NASA-TLX (Task Load Index): Results of empirical and theoretical research, In Human Mental Workload, P. A. Hancock and N. Meshkati, Amsterdam: North Holland Press, 1988,

PENTAX Medical i10 Series HD+ Endoscopes Ergonomic Design, Superior Function, Quality Care

PENTAX Medical i10 Series HD+ Endoscopes Ergonomic Design, Superior Function, Quality Care PENTAX Medical i10 Series Endoscopes Ergonomic Design, Superior Function, Quality Care PENTAX Medical i10 Series Endoscopes Raising the Standards of Clinical Acceptance Engineered and designed in partnership

More information

UNIT II Robots Drive Systems and End Effectors Part-A Questions

UNIT II Robots Drive Systems and End Effectors Part-A Questions UNIT II Robots Drive Systems and End Effectors Part-A Questions 1. Define End effector. End effector is a device that is attached to the end of the wrist arm to perform specific task. 2. Give some examples

More information

Scooter, 3 wheeled cobot North Western University. PERCRO Exoskeleton

Scooter, 3 wheeled cobot North Western University. PERCRO Exoskeleton Scooter, 3 wheeled cobot North Western University A cobot is a robot for direct physical interaction with a human operator, within a shared workspace PERCRO Exoskeleton Unicycle cobot the simplest possible

More information

Medical Robotics. Control Modalities

Medical Robotics. Control Modalities Università di Roma La Sapienza Medical Robotics Control Modalities The Hands-On Acrobot Robot Marilena Vendittelli Dipartimento di Ingegneria Informatica, Automatica e Gestionale Control modalities differ

More information

Unit 24: Applications of Pneumatics and Hydraulics

Unit 24: Applications of Pneumatics and Hydraulics Unit 24: Applications of Pneumatics and Hydraulics Unit code: J/601/1496 QCF level: 4 Credit value: 15 OUTCOME 2 TUTORIAL 4 DIRECTIONAL CONTROL VALVES The material needed for outcome 2 is very extensive

More information

Multi-Touch Control Wheel Software Development Kit User s Guide

Multi-Touch Control Wheel Software Development Kit User s Guide Multi-Touch Control Wheel Software Development Kit User s Guide V3.0 Bulletin #1204 561 Hillgrove Avenue LaGrange, IL 60525 Phone: (708) 354-1040 Fax: (708) 354-2820 E-mail: instinct@grayhill.com www.grayhill.com/instinct

More information

Introduction. Basis and preliminary investigations

Introduction. Basis and preliminary investigations RSI prevention by changing computer input devices - Study by TU Darmstadt investigates the pen tablet as an ergonomic alternative to the computer mouse - Introduction Users of computer work stations often

More information

Industrial Robotics. Training Objective

Industrial Robotics. Training Objective Training Objective After watching the program and reviewing this printed material, the viewer will learn the basics of industrial robot technology and how robots are used in a variety of manufacturing

More information

2003 ACCORD - Automatic Transmission Removal

2003 ACCORD - Automatic Transmission Removal 2003 ACCORD - Automatic Transmission Removal Special Tools Required Engine support hanger, A and Reds AAR-T-12566 Engine hanger balancer bar VSB02C000019 Front subframe adapter VSB02C000016 These special

More information

CIM Computer Integrated Manufacturing

CIM Computer Integrated Manufacturing INDEX CIM IN BASIC CONFIGURATION CIM IN ADVANCED CONFIGURATION CIM IN COMPLETE CONFIGURATION DL CIM A DL CIM B DL CIM C DL CIM C DL CIM B DL CIM A Computer Integrated Manufacturing (CIM) is a method of

More information

Robot coined by Karel Capek in a 1921 science-fiction Czech play

Robot coined by Karel Capek in a 1921 science-fiction Czech play Robotics Robot coined by Karel Capek in a 1921 science-fiction Czech play Definition: A robot is a reprogrammable, multifunctional manipulator designed to move material, parts, tools, or specialized devices

More information

Pneumatically Driven Robot System with Force Perception for Minimally Invasive Surgery

Pneumatically Driven Robot System with Force Perception for Minimally Invasive Surgery Pneumatically Driven Robot System with Force Perception for Minimally Invasive Surgery Tokyo Institute of Technology Kotaro Tadano Daisuke Haraguchi Kenji Kawashima Research Overview on Surgical Robots

More information

Universal Exoskeleton Arm Design for Rehabilitation

Universal Exoskeleton Arm Design for Rehabilitation Journal of Automation and Control Engineering Vol. 3, No. 6, December 215 Universal Exoskeleton Arm Design for Rehabilitation Siam Charoenseang and Sarut Panjan Institute of Field Robotics, King Mongkut

More information

GYN 56 1.0 11/2015-E. GynTrainer. the virtual platform for diverse and risk-free training of gynecological diagnosis and therapy

GYN 56 1.0 11/2015-E. GynTrainer. the virtual platform for diverse and risk-free training of gynecological diagnosis and therapy GYN 56 1.0 11/2015-E GynTrainer the virtual platform for diverse and risk-free training of gynecological diagnosis and therapy Virtual Reality (VR) in Medical Training and Further Education at KARL STORZ

More information

EVALUAT ING ACADEMIC READINESS FOR APPRENTICESHIP TRAINING Revised for ACCESS TO APPRENTICESHIP

EVALUAT ING ACADEMIC READINESS FOR APPRENTICESHIP TRAINING Revised for ACCESS TO APPRENTICESHIP EVALUAT ING ACADEMIC READINESS FOR APPRENTICESHIP TRAINING for ACCESS TO APPRENTICESHIP SCIENCE SKILLS SIMPLE MACHINES & MECHANICAL ADVANTAGE AN ACADEMIC SKILLS MANUAL for The Construction Trades: Mechanical

More information

Force/position control of a robotic system for transcranial magnetic stimulation

Force/position control of a robotic system for transcranial magnetic stimulation Force/position control of a robotic system for transcranial magnetic stimulation W.N. Wan Zakaria School of Mechanical and System Engineering Newcastle University Abstract To develop a force control scheme

More information

Hemostasis Solutions Boston Scientific is committed to improving patient care in the management of gastrointestinal bleeding.

Hemostasis Solutions Boston Scientific is committed to improving patient care in the management of gastrointestinal bleeding. Hemostasis Solutions Boston Scientific is committed to improving patient care in the management of gastrointestinal bleeding. Through innovation and continuous educational support, we offer a wide range

More information

Equipment loan arrangements and loan form

Equipment loan arrangements and loan form Safety and Health Services Equipment loan arrangements and loan form Document control information Published document name: Work-equip-loan-gn Date issued: February 2016 Version: 4.1 Previous review dates:

More information

Design Aspects of Robot Manipulators

Design Aspects of Robot Manipulators Design Aspects of Robot Manipulators Dr. Rohan Munasinghe Dept of Electronic and Telecommunication Engineering University of Moratuwa System elements Manipulator (+ proprioceptive sensors) End-effector

More information

Closed-Loop Motion Control Simplifies Non-Destructive Testing

Closed-Loop Motion Control Simplifies Non-Destructive Testing Closed-Loop Motion Control Simplifies Non-Destructive Testing Repetitive non-destructive testing (NDT) applications abound, and designers should consider using programmable motion controllers to power

More information

Development of Docking System for Mobile Robots Using Cheap Infrared Sensors

Development of Docking System for Mobile Robots Using Cheap Infrared Sensors Development of Docking System for Mobile Robots Using Cheap Infrared Sensors K. H. Kim a, H. D. Choi a, S. Yoon a, K. W. Lee a, H. S. Ryu b, C. K. Woo b, and Y. K. Kwak a, * a Department of Mechanical

More information

dspace DSP DS-1104 based State Observer Design for Position Control of DC Servo Motor

dspace DSP DS-1104 based State Observer Design for Position Control of DC Servo Motor dspace DSP DS-1104 based State Observer Design for Position Control of DC Servo Motor Jaswandi Sawant, Divyesh Ginoya Department of Instrumentation and control, College of Engineering, Pune. ABSTRACT This

More information

Is it possible to work comfortably and productively on a tablet?

Is it possible to work comfortably and productively on a tablet? Is it possible to work comfortably and productively on a tablet? An overview of tablet use for professionals Introduction The tablet is no longer simply a new gadget, as professional use of this technology

More information

Servo Motors (SensorDAQ only) Evaluation copy. Vernier Digital Control Unit (DCU) LabQuest or LabPro power supply

Servo Motors (SensorDAQ only) Evaluation copy. Vernier Digital Control Unit (DCU) LabQuest or LabPro power supply Servo Motors (SensorDAQ only) Project 7 Servos are small, relatively inexpensive motors known for their ability to provide a large torque or turning force. They draw current proportional to the mechanical

More information

HAND INSTRUMENTS. HiQ+ Monopolar Hand Instruments Needle Holders Suction and Irrigation Monopolar HF Electrodes. THE HiQ+ SERIES

HAND INSTRUMENTS. HiQ+ Monopolar Hand Instruments Needle Holders Suction and Irrigation Monopolar HF Electrodes. THE HiQ+ SERIES HAND INSTRUMENTS HiQ+ Monopolar Hand Instruments Needle Holders Suction and Irrigation Monopolar HF Electrodes THE HiQ+ SERIES CONTENTS FORM FOLLOWS FUNCTION FOLLOWS YOUR NEEDS HiQ+ Monopolar and Standard...4

More information

FRC WPI Robotics Library Overview

FRC WPI Robotics Library Overview FRC WPI Robotics Library Overview Contents 1.1 Introduction 1.2 RobotDrive 1.3 Sensors 1.4 Actuators 1.5 I/O 1.6 Driver Station 1.7 Compressor 1.8 Camera 1.9 Utilities 1.10 Conclusion Introduction In this

More information

SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM AND A NEW METHOD OF RESULTS ANALYSIS 1. INTRODUCTION

SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM AND A NEW METHOD OF RESULTS ANALYSIS 1. INTRODUCTION Journal of Machine Engineering, Vol. 15, No.1, 2015 machine tool accuracy, metrology, spindle error motions Krzysztof JEMIELNIAK 1* Jaroslaw CHRZANOWSKI 1 SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM

More information

FUNDAMENTALS OF ROBOTICS

FUNDAMENTALS OF ROBOTICS FUNDAMENTALS OF ROBOTICS Lab exercise Stäubli AULINAS Josep (u1043469) GARCIA Frederic (u1038431) Introduction The aim of this tutorial is to give a brief overview on the Stäubli Robot System describing

More information

COMPUTER-RELATED MUSCLE, TENDON, AND JOINT INJURIES

COMPUTER-RELATED MUSCLE, TENDON, AND JOINT INJURIES CHAPTER ELEVEN COMPUTER-RELATED MUSCLE, TENDON, AND JOINT INJURIES To reduce the risk of pain in your neck and shoulders, stay within these recommended ranges of movement: Neck Flexion: 0 o -15 o (bending

More information

Targus Wireless RF Mouse USER S GUIDE. Making Your Mobile Life Easier.

Targus Wireless RF Mouse USER S GUIDE. Making Your Mobile Life Easier. Targus Wireless RF Mouse Visit our Web site at: www.targus.com Features and specifications are subject to change without notice. 2004 Targus Group International and Targus, Inc.. 400-0111-001B USER S GUIDE

More information

Encoders for Linear Motors in the Electronics Industry

Encoders for Linear Motors in the Electronics Industry Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy

More information

ERGONOMICS. Improve your ergonomic intelligence by avoiding these issues: Awkward postures Repetitive tasks Forceful exertions Lifting heavy objects

ERGONOMICS. Improve your ergonomic intelligence by avoiding these issues: Awkward postures Repetitive tasks Forceful exertions Lifting heavy objects ERGONOMICS The goal of ergonomics is to reduce your exposure to work hazards. A hazard is defined as a physical factor within your work environment that can harm your body. Ergonomic hazards include working

More information

Operation of the overrun braking system.

Operation of the overrun braking system. Operation of the overrun braking system. Fig 1 The overrun device can be described as the control device of the overrun brake system. A draw-bar force is produced at the coupling point by reducing the

More information

Design of a Robotic Arm with Gripper & End Effector for Spot Welding

Design of a Robotic Arm with Gripper & End Effector for Spot Welding Universal Journal of Mechanical Engineering 1(3): 92-97, 2013 DOI: 10.13189/ujme.2013.010303 http://www.hrpub.org Design of a Robotic Arm with Gripper & End Effector for Spot Welding Puran Singh *, Anil

More information

HYFLAM D3: Report of Scenarios 08/02/2012

HYFLAM D3: Report of Scenarios 08/02/2012 08/02/2012 Author: Daniel Greenwald Reviewers: Hugo Elias, Rich Walker, Jan Gries, Jackie Duggan 1/13 Table of Contents 1 Introduction...2 2 Experiments...2 2.1 Opening and Closing a Plastic Jar...3 2.2

More information

CS-525V: Building Effective Virtual Worlds. Input Devices. Robert W. Lindeman. Worcester Polytechnic Institute Department of Computer Science

CS-525V: Building Effective Virtual Worlds. Input Devices. Robert W. Lindeman. Worcester Polytechnic Institute Department of Computer Science CS-525V: Building Effective Virtual Worlds Input Devices Robert W. Lindeman Worcester Polytechnic Institute Department of Computer Science gogo@wpi.edu Motivation The mouse and keyboard are good for general

More information

APPLIED PNEUMATICS AND HYDRAULICS H TUTORIAL HYDRAULIC AND PNEUMATIC CYLINDERS. This work covers part of outcome 2 of the standard Edexcel module.

APPLIED PNEUMATICS AND HYDRAULICS H TUTORIAL HYDRAULIC AND PNEUMATIC CYLINDERS. This work covers part of outcome 2 of the standard Edexcel module. APPLIED PNEUMATICS AND HYDRAULICS H TUTORIAL HYDRAULIC AND PNEUMATIC CYLINDERS This work covers part of outcome 2 of the standard Edexcel module. The material needed for outcome 2 is very extensive so

More information

Linear modules Lifting units Rotary modules Grip modules Inductive proximity switches Plug connectors

Linear modules Lifting units Rotary modules Grip modules Inductive proximity switches Plug connectors 20000 Linear modules Lifting units Rotary modules Grip modules Inductive proximity switches Plug connectors 23000 22000 21000 20000 09000 08000 07000 06000 05000 04000 03000 02000 01000 823 Notes 824 Technical

More information

// ipq-view Digital web viewing for your quality assurance. Be inspired. Move forward.

// ipq-view Digital web viewing for your quality assurance. Be inspired. Move forward. // ipq-view Digital web viewing for your quality assurance. Be inspired. Move forward. // EXCELLENT PRESENTATION. RELIABLE QUALITY. Competition in the printing industry is hard, and the requirements are

More information

Multi-Touch Control Wheel Software Development Kit User s Guide

Multi-Touch Control Wheel Software Development Kit User s Guide Multi-Touch Control Wheel Software Development Kit User s Guide v1.0 Bulletin #1204 561 Hillgrove Avenue LaGrange, IL 60525 Phone: (708) 354-1040 Fax: (708) 354-2820 E-mail: instinct@grayhill.com On the

More information

Frequently Asked Questions

Frequently Asked Questions Frequently Asked Questions Basic Facts What does the name ASIMO stand for? ASIMO stands for Advanced Step in Innovative Mobility. Who created ASIMO? ASIMO was developed by Honda Motor Co., Ltd., a world

More information

2010 APPLICATION REFERENCE GUIDE

2010 APPLICATION REFERENCE GUIDE 2010 APPLICATION REFERENCE GUIDE INDEX QUALITY POLICY 2 ERGO ASSIST STANDARD 4-13 CUSTOM 14-18 END EFFECTORS 19-24 LIFT ASSIST 25-41 FIXTURED TOOLS 42-53 FIXTURES 54-63 ASSEMBLY TABLES 64-72 CARTS/ DUNNAGE

More information

A Remote Maintenance System with the use of Virtual Reality.

A Remote Maintenance System with the use of Virtual Reality. ICAT 2001 December 5-7, Tokyo, JAPAN A Remote Maintenance System with the use of Virtual Reality. Moez BELLAMINE 1), Norihiro ABE 1), Kazuaki TANAKA 1), Hirokazu TAKI 2) 1) Kyushu Institute of Technology,

More information

Submitted By: Submitted To: XYZ ABCD ******** Mechatronics

Submitted By: Submitted To: XYZ ABCD ******** Mechatronics Submitted To: ABCD Submitted By: XYZ ******** Mechatronics END EFFECTORS In robotics, an end effector is the device at the end of a robotic arm, designed to interact with the environment. The exact nature

More information

Computer Workstation Ergonomic Self Evaluation

Computer Workstation Ergonomic Self Evaluation Computer Workstation Ergonomic Self Evaluation Use this guidance to perform a preliminary evaluation of your workstation, and make any adjustments that are necessary to achieve a correct neutral posture.

More information

Development of Easy Teaching Interface for a Dual Arm Robot Manipulator

Development of Easy Teaching Interface for a Dual Arm Robot Manipulator Development of Easy Teaching Interface for a Dual Arm Robot Manipulator Chanhun Park and Doohyeong Kim Department of Robotics and Mechatronics, Korea Institute of Machinery & Materials, 156, Gajeongbuk-Ro,

More information

Design and Implementation of a 4-Bar linkage Gripper

Design and Implementation of a 4-Bar linkage Gripper IOSR Journal of Mechanical and Civil Engineering (IOSR-JMCE) e-issn: 2278-1684,p-ISSN: 2320-334X, Volume 11, Issue 5 Ver. IV (Sep- Oct. 2014), PP 61-66 Design and Implementation of a 4-Bar linkage Gripper

More information

HYDRAULIC TABLE CART 500-LB.

HYDRAULIC TABLE CART 500-LB. HYDRAULIC TABLE CART 500-LB. OWNER S MANUAL WARNING: Read carefully and understand all MACHINE ADJUSTMENT AND OPERATION INSTRUCTIONS before operating. Failure to follow the safety rules and other basic

More information

INTRODUCTION TO SERIAL ARM

INTRODUCTION TO SERIAL ARM INTRODUCTION TO SERIAL ARM A robot manipulator consists of links connected by joints. The links of the manipulator can be considered to form a kinematic chain. The business end of the kinematic chain of

More information

Fact Sheet: Occupational Overuse Syndrome (OOS)

Fact Sheet: Occupational Overuse Syndrome (OOS) Fact Sheet: Occupational Overuse Syndrome (OOS) What is OOS? Occupational Overuse Syndrome (OOS) is the term given to a range of conditions characterised by discomfort or persistent pain in muscles, tendons

More information

Repetitive Strain Injury (RSI)

Repetitive Strain Injury (RSI) Carpal Tunnel Syndrome and Other Musculoskeletal Problems in the Workplace: What s the Solution? by Richard N. Hinrichs, Ph.D. Dept. of Kinesiology Arizona State University Repetitive Strain Injury (RSI)

More information

General Guidelines. Neck Stretch: Side. Neck Stretch: Forward. Shoulder Rolls. Side Stretch

General Guidelines. Neck Stretch: Side. Neck Stretch: Forward. Shoulder Rolls. Side Stretch Stretching Exercises General Guidelines Perform stretching exercises at least 2 3 days per week and preferably more Hold each stretch for 15 20 seconds Relax and breathe normally Stretching is most effective

More information

A guide for employees

A guide for employees A guide for employees This booklet has been developed as a guide to assist in the prevention of muscular discomfort and strain that can sometimes be associated with computer use. The information provided

More information

CE801: Intelligent Systems and Robotics Lecture 3: Actuators and Localisation. Prof. Dr. Hani Hagras

CE801: Intelligent Systems and Robotics Lecture 3: Actuators and Localisation. Prof. Dr. Hani Hagras 1 CE801: Intelligent Systems and Robotics Lecture 3: Actuators and Localisation Prof. Dr. Hani Hagras Robot Locomotion Robots might want to move in water, in the air, on land, in space.. 2 Most of the

More information

COUNTERBALANCE VALVES

COUNTERBALANCE VALVES COUNTERBALANCE VALVES Introduction They are modulating valves which allow free flow into the actuator and then block the reverse flow until they feel a pilot pressure inversely proportional to the load

More information

Development of Combined Automatic Blade Control for Snow-Removing Grader

Development of Combined Automatic Blade Control for Snow-Removing Grader Technical Papers Yukihisa Sakai In snowy regions, improving the efficiency of snow-removing machines has been called for so as to reduce the time and cost involved in clearing the roads of snow. As a means

More information

Electronic Manual Gearbox

Electronic Manual Gearbox Service. Self-Study Programme 221 Electronic Manual Gearbox Design and Function Taking the Lupo as the basis, Volkswagen has developed the world's first 3 L car that will also go into volume production.

More information

CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS

CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS E. Batzies 1, M. Kreutzer 1, D. Leucht 2, V. Welker 2, O. Zirn 1 1 Mechatronics Research

More information

Vibrations can have an adverse effect on the accuracy of the end effector of a

Vibrations can have an adverse effect on the accuracy of the end effector of a EGR 315 Design Project - 1 - Executive Summary Vibrations can have an adverse effect on the accuracy of the end effector of a multiple-link robot. The ability of the machine to move to precise points scattered

More information

Recommended Forklift Free System Features for Proper Safety and Ergonomic Operation

Recommended Forklift Free System Features for Proper Safety and Ergonomic Operation When re-designing material handling flow to address restricted use of forklifts (Forklift Free), equipment ergonomics and safety play a key role. Following is a list of equipment and corresponding features

More information

microscope OP-Dent-5

microscope OP-Dent-5 microscope OP-Dent-5 OP-Dent-5 ERGONOMICS AND VERSATILITY The OP-Dent-5 microscope manufactured by OP- TOMIC fulfils the most demanding requirements in the dental field. The OP-Dent-5 microscope offers

More information

DEMOLITION Photocopy this profile and distribute it as widely as possible!

DEMOLITION Photocopy this profile and distribute it as widely as possible! Musculoskeletal hazards and controls DEMOLITION Photocopy this profile and distribute it as widely as possible! Musculoskeletal disorders (MSDs), such as chronic back pain or shoulder problems, often take

More information

Linear Motion System: Transport and positioning for demanding applications

Linear Motion System: Transport and positioning for demanding applications Linear Motion System: Transport and positioning for demanding applications 2 The Perfect Concept for a variety of applications The Linear Motion System (LMS) from Rexroth is a unique technical solution

More information

Integration of a Robotic Arm with the Surgical Assistant Workstation Software Framework

Integration of a Robotic Arm with the Surgical Assistant Workstation Software Framework Integration of a Robotic Arm with the Surgical Assistant Workstation Software Framework Release 1.7 Jessie Young 1, Haytham Elhawary 2 and Aleksandra Popovic 2 July 21, 2011 1 Center for Computer-Integrated

More information

MMVR: 10, January 2002. On Defining Metrics for Assessing Laparoscopic Surgical Skills in a Virtual Training Environment

MMVR: 10, January 2002. On Defining Metrics for Assessing Laparoscopic Surgical Skills in a Virtual Training Environment On Defining Metrics for Assessing Laparoscopic Surgical Skills in a Virtual Training Environment Shahram Payandeh, Alan J. Lomax, John Dill, Christine L. Mackenzie and Caroline G. L. Cao Schools of Engineering

More information

Robotics and Automation Blueprint

Robotics and Automation Blueprint Robotics and Automation Blueprint This Blueprint contains the subject matter content of this Skill Connect Assessment. This Blueprint does NOT contain the information one would need to fully prepare for

More information

Tracking devices. Important features. 6 Degrees of freedom. Mechanical devices. Types. Virtual Reality Technology and Programming

Tracking devices. Important features. 6 Degrees of freedom. Mechanical devices. Types. Virtual Reality Technology and Programming Tracking devices Virtual Reality Technology and Programming TNM053: Lecture 4: Tracking and I/O devices Referred to head-tracking many times Needed to get good stereo effect with parallax Essential for

More information

This week. CENG 732 Computer Animation. Challenges in Human Modeling. Basic Arm Model

This week. CENG 732 Computer Animation. Challenges in Human Modeling. Basic Arm Model CENG 732 Computer Animation Spring 2006-2007 Week 8 Modeling and Animating Articulated Figures: Modeling the Arm, Walking, Facial Animation This week Modeling the arm Different joint structures Walking

More information

Introduction to Medical Imaging (5XSA0)

Introduction to Medical Imaging (5XSA0) Introduction to Medical Imaging (5XSA) Medical image acquisition and analysis: Endoscopy 1 Endoscopy: Introduction (1) Endo-scopy: looking inside (the body) Minimally invasive procedure. Mostly through

More information

Active Vibration Isolation of an Unbalanced Machine Spindle

Active Vibration Isolation of an Unbalanced Machine Spindle UCRL-CONF-206108 Active Vibration Isolation of an Unbalanced Machine Spindle D. J. Hopkins, P. Geraghty August 18, 2004 American Society of Precision Engineering Annual Conference Orlando, FL, United States

More information

Three Channel Optical Incremental Encoder Modules Technical Data

Three Channel Optical Incremental Encoder Modules Technical Data Three Channel Optical Incremental Encoder Modules Technical Data HEDS-9040 HEDS-9140 Features Two Channel Quadrature Output with Index Pulse Resolution Up to 2000 CPR Counts Per Revolution Low Cost Easy

More information

The Shape Of Things To Come. Dr. F.P. Wieringa LVO congres 7-8 feb 2013

The Shape Of Things To Come. Dr. F.P. Wieringa LVO congres 7-8 feb 2013 The Shape Of Things To Come Dr. F.P. Wieringa LVO congres 7-8 feb 2013 Long as you ve got your health, Bradley W. Schenck 2004 f.wieringa@tno.nl Medical technology innovation eco-system Hospitals Universities

More information

Robot Task-Level Programming Language and Simulation

Robot Task-Level Programming Language and Simulation Robot Task-Level Programming Language and Simulation M. Samaka Abstract This paper presents the development of a software application for Off-line robot task programming and simulation. Such application

More information

How To Control Gimbal

How To Control Gimbal Tarot 2-Axis Brushless Gimbal for Gopro User Manual V1.0 1. Introduction Tarot T-2D gimbal is designed for the Gopro Hero3, which is widely used in film, television productions, advertising aerial photography,

More information

Cost effective ergonomic improvements

Cost effective ergonomic improvements Innovative Ergonomic Solutions in Manufacturing Cost effective ergonomic improvements John Schlyer 316-526 526-4661 Spirit AeroSystems, Inc. Innovative Ergonomic Solutions in Manufacturing Ergonomic improvements

More information

Belt Tensioning Methods for Small Package Conveyors What s the best solution?

Belt Tensioning Methods for Small Package Conveyors What s the best solution? Belt Tensioning Methods for Small Package Conveyors What s the best solution? Industrial White Paper By: Michael A. Hosch, P.E. Dorner Mfg. Corp. 975 Cottonwood Avenue Hartland, WI 53029 USA Phone: 800

More information

Digital Single Axis Controller

Digital Single Axis Controller Digital Single Axis Controller The concept of intelligent hydraulic drive a vision? DSAC (Digital Single Axis Controller) Do you need hydraulic servo drive for your machine/system? What would it be like

More information

Proper Workstation Setup

Proper Workstation Setup Proper Workstation Setup A Step by Step Guide to help you organize your workstation and reduce the risk of injury Step 1 Review Work Habits Working in a stationary position for long periods will increase

More information

ROBOT END EFFECTORS SCRIPT

ROBOT END EFFECTORS SCRIPT Slide 1 Slide 2 Slide 3 Slide 4 An end effector is the business end of a robot or where the work occurs. It is the device that is designed to allow the robot to interact with its environment. Similar in

More information

How To Choose The Right End Effector. For Your Application

How To Choose The Right End Effector. For Your Application How To Choose The Right End Effector For Your Application TABLE OF CONTENTS INTRODUCTION... 3 DIFFERENT END EFFECTORS ON THE MARKET... 3 ROBOT GRIPPERS... 3 ROBOTIC TOOLS... 7 HOW TO CHOOSE?... 8 HOW MUCH

More information

Lecture 3: Teleoperation

Lecture 3: Teleoperation ME 328: Medical Robotics Spring 2015 Lecture 3: Teleoperation Allison Okamura Stanford University Announcements 1. I am not Allison. 2. You re a huge class. So, we found a 2nd CA! meet Jeesu Baek jeesu@stanford.edu

More information

VEHICLE SPEED CONTROL SYSTEM

VEHICLE SPEED CONTROL SYSTEM PL VEHICLE SPEED CONTROL SYSTEM 8H - 1 VEHICLE SPEED CONTROL SYSTEM TABLE OF CONTENTS page DESCRIPTION AND SPEED CONTROL SYSTEM...1 SPEED CONTROL SERVO-PCM OUTPUT....2 SPEED CONTROL SWITCHES PCM INPUT...2

More information

Fig 1 Power Transmission system of Tractor

Fig 1 Power Transmission system of Tractor POWER TRANSMISSION SYSTEM Transmission is a speed reducing mechanism, equipped with several gears (Fig. 1). It may be called a sequence of gears and shafts, through which the engine power is transmitted

More information

Yes No Correction Date. Seated Position Do you know how to adjust your chair? Is your chair adjusted for support & comfort?

Yes No Correction Date. Seated Position Do you know how to adjust your chair? Is your chair adjusted for support & comfort? UC Davis Health System - Computer Workstation Self Evaluation Assessment Date: Employee Name: Employee ID#: Department: Supervisor: Union Affiliation: Date of Hire (if new hire) Reason for Assessment (circle

More information

Frequently Asked Questions

Frequently Asked Questions Frequently Asked Questions Basic Facts What does the name ASIMO stand for? ASIMO stands for Advanced Step in Innovative Mobility. Who created ASIMO? ASIMO was developed by Honda Motor Co., Ltd., a world

More information

Tool Turrets and Tool Discs

Tool Turrets and Tool Discs Our other products Automatic Tool Changer Chucking Cylinders Tool Discs Indexing Tables Tool Turrets and Tool Discs Pragati Automation Pvt. Ltd. 1, IV Phase, 11th Cross, Peenya Industrial Area Bangalore

More information

ECE 495 Project 3: Shocker Actuator Subsystem and Website Design. Group 1: One Awesome Engineering

ECE 495 Project 3: Shocker Actuator Subsystem and Website Design. Group 1: One Awesome Engineering ECE 495 Project 3: Shocker Actuator Subsystem and Website Design Group 1: One Awesome Engineering Luquita Edwards Evan Whetsell Sunny Verma Thomas Ryan Willis Long I. Executive Summary The main goal behind

More information

The Back Saver. Ergonomic Lifting Guide For Roll Handling

The Back Saver. Ergonomic Lifting Guide For Roll Handling The Back Saver Ergonomic Lifting Guide For Roll Handling Since 1990, Alum-a-lift has conceived, designed, developed and manufactured ergonomic lifting devices; Devices to lift, lower, move, position and

More information

CURVETM Image Guided Surgery

CURVETM Image Guided Surgery CURVETM Image Guided Surgery THE ULTIMATE COMMAND AND CONTROL CENTER FOR INFORMATION-GUIDED SURGERY Features building benefits, Curve Image Guided Surgery challenges conventional navigation utility. Super

More information

Considering Bariatric Surgery? Learn about minimally invasive da Vinci Surgery

Considering Bariatric Surgery? Learn about minimally invasive da Vinci Surgery Considering Bariatric Surgery? Learn about minimally invasive da Vinci Surgery The Condition: Obesity Obesity is defined as having a body mass index (BMI) of 30 or greater. Obesity is a serious medical

More information

IMPORTANT SAFETY RULES TO FOLLOW

IMPORTANT SAFETY RULES TO FOLLOW WARNING FLOOR & CARPET CLEANER Any piece of equipment can be dangerous if not operated properly. YOU are responsible for the safe operation of this equipment. The operator must carefully read and follow

More information

PART 2 FORKLIFT HYDRAULIC SYSTEM

PART 2 FORKLIFT HYDRAULIC SYSTEM PART 2 FORKLIFT HYDRAULIC SYSTEM Chapter 1 Description and Operation Component Locations & Circuit Layouts 1 Hydraulic Pump 11 Control Valve 14 Valve Section Oil Flows 15 Anti-Cavitation Valve 22 Velocity

More information

Computer Graphics in Medicine

Computer Graphics in Medicine Computer Graphics in Medicine Loren Block For my presentation, I will discuss several different papers that are related to graphics and imaging in the medical field. The main areas that the papers cover

More information

A descriptive definition of valve actuators

A descriptive definition of valve actuators A descriptive definition of valve actuators Abstract A valve actuator is any device that utilizes a source of power to operate a valve. This source of power can be a human being working a manual gearbox

More information

HealthandSafetyOntario.ca. Hazards. Introduction. Legislation

HealthandSafetyOntario.ca. Hazards. Introduction. Legislation Manual Materials handling Introduction Manual materials handling operations are carried out in most workplaces. Each handling task poses unique demands on the worker. However, workplaces can help workers

More information

Range of Motion. A guide for you after spinal cord injury. Spinal Cord Injury Rehabilitation Program

Range of Motion. A guide for you after spinal cord injury. Spinal Cord Injury Rehabilitation Program Range of Motion A guide for you after spinal cord injury Spinal Cord Injury Rehabilitation Program This booklet has been written by the health care providers who provide care to people who have a spinal

More information

Subminiature Load Cell Model 8417

Subminiature Load Cell Model 8417 w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...

More information

67273 EN V40/14 NOUVAG AG, all rights reserved Subject to change. Laparoscopic Hysterectomy Morcellator

67273 EN V40/14 NOUVAG AG, all rights reserved Subject to change. Laparoscopic Hysterectomy Morcellator 67273 EN V40/14 NOUVAG AG, all rights reserved Subject to change Laparoscopic Hysterectomy Morcellator Laparoscopic Hysterectomy Nowadays newly developed medical devices make hysterectomy possible by a

More information

Ergonomics in Agriculture

Ergonomics in Agriculture Ergonomics in Agriculture Although, agriculture has become more mechanized in the last century, there are still many small farm operations where planting and harvesting activities are done by hand-a person

More information

UOM Robotics and Industrial Automation Laboratory

UOM Robotics and Industrial Automation Laboratory UOM Robotics and Industrial Automation Laboratory Department of Industrial and Manufacturing Engineering, RIAL Mission o to be a Centre of Excellence in Malta for research, development, teaching and outreach

More information